Field Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in San Francisco, USA, and Kisumu, Kenya Le D, 1 De Pinto S, 1 Efros L, 1 Lee S, 1 Bui N, 1 Crow M, 1 Mosqueda F, 1 Tran A, 1 Waheed S, 1 Swaim M, 1 Yaneza D, 1 Corr N, 1 Bhatia S, 1 Wu A, 1 Mo D, 1 Tilahun H, 1 Lu B, 1 Shea E, 2 Ong C, 2 Mimba E, 3 Akoth B, 3 Oyaro B, 3 Angira F, 3 Wu A, 2 Zeh C, 3 Chen R, 1 Bush-Donovan C, 1 Bornheimer S 1 1 BD Biosciences, San Jose, CA, USA, 2 University of California at San Francisco, San Francisco, CA, USA, and 3 KEMRI/CDC Research and Public Health Collaboration, Kisumu, Kenya Presented at AIDS 2014- Melbourne, Australia
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Field Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in San Francisco, USA, and Kisumu, Kenya
Le D,1 De Pinto S,1 Efros L,1 Lee S,1 Bui N,1 Crow M,1 Mosqueda F,1 Tran A, 1 Waheed S,1 Swaim M,1 Yaneza D,1 Corr N,1 Bhatia S,1 Wu A,1 Mo D,1 Tilahun H,1 Lu B,1 Shea E,2 Ong C,2 Mimba E,3 Akoth B,3 Oyaro B,3 Angira F,3 Wu A,2 Zeh C,3 Chen R,1 Bush-Donovan C,1 Bornheimer S1
1BD Biosciences, San Jose, CA, USA, 2University of California at San Francisco, San Francisco, CA, USA, and 3KEMRI/CDC Research and Public Health Collaboration, Kisumu, Kenya
Presented at AIDS 2014- Melbourne, Australia
BD FACSPresto™ System Workflow
BD FACSPresto™ Performance Summary Accuracy
BD FACSPresto™ Performance Summary Accuracy – San Francisco
0 200 400 600 800 1000 1200 1400 16000
200
400
600
800
1000
1200
1400
1600
Fingerstick Deming Regression for CD4
BD FACSCalibur - BD Tritest CD4 Absolute Counts per μL
BD
FA
CS
Pre
sto
CD
4 A
bsol
ute
Cou
nts
per μ
L
R2=0.97Slope= 1.04
BD FACSPresto™ Performance SummaryPrecision
BD FACSPresto™ Performance Conclusion
In development field studies, the BD FACSPresto Near-Patient CD4 Counter provided accurate and precise clinical results for CD4, %CD4, and total Hb, using capillary or venous samples from HIV patients in hospital clinics in the USA and Kenya. The system is suitable for point-of-care CD4 enumeration.