Top Banner
Field Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in San Francisco, USA, and Kisumu, Kenya Le D, 1 De Pinto S, 1 Efros L, 1 Lee S, 1 Bui N, 1 Crow M, 1 Mosqueda F, 1 Tran A, 1 Waheed S, 1 Swaim M, 1 Yaneza D, 1 Corr N, 1 Bhatia S, 1 Wu A, 1 Mo D, 1 Tilahun H, 1 Lu B, 1 Shea E, 2 Ong C, 2 Mimba E, 3 Akoth B, 3 Oyaro B, 3 Angira F, 3 Wu A, 2 Zeh C, 3 Chen R, 1 Bush-Donovan C, 1 Bornheimer S 1 1 BD Biosciences, San Jose, CA, USA, 2 University of California at San Francisco, San Francisco, CA, USA, and 3 KEMRI/CDC Research and Public Health Collaboration, Kisumu, Kenya Presented at AIDS 2014- Melbourne, Australia
6

BD FACSPresto ™ System Workflow

Feb 22, 2016

Download

Documents

Amy Leaver

Field Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in San Francisco, USA, and Kisumu, Kenya. - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: BD  FACSPresto ™ System Workflow

Field Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in San Francisco, USA, and Kisumu, Kenya

Le D,1 De Pinto S,1 Efros L,1 Lee S,1 Bui N,1 Crow M,1 Mosqueda F,1 Tran A, 1 Waheed S,1 Swaim M,1 Yaneza D,1 Corr N,1 Bhatia S,1 Wu A,1 Mo D,1 Tilahun H,1 Lu B,1 Shea E,2 Ong C,2 Mimba E,3 Akoth B,3 Oyaro B,3 Angira F,3 Wu A,2 Zeh C,3 Chen R,1 Bush-Donovan C,1 Bornheimer S1

1BD Biosciences, San Jose, CA, USA, 2University of California at San Francisco, San Francisco, CA, USA, and 3KEMRI/CDC Research and Public Health Collaboration, Kisumu, Kenya  

Presented at AIDS 2014- Melbourne, Australia

Page 2: BD  FACSPresto ™ System Workflow

BD FACSPresto™ System Workflow

Page 3: BD  FACSPresto ™ System Workflow

BD FACSPresto™ Performance Summary Accuracy

Page 4: BD  FACSPresto ™ System Workflow

BD FACSPresto™ Performance Summary Accuracy – San Francisco

0 200 400 600 800 1000 1200 1400 16000

200

400

600

800

1000

1200

1400

1600

Fingerstick Deming Regression for CD4

BD FACSCalibur - BD Tritest CD4 Absolute Counts per μL

BD

FA

CS

Pre

sto

CD

4 A

bsol

ute

Cou

nts

per μ

L

R2=0.97Slope= 1.04

Page 5: BD  FACSPresto ™ System Workflow

BD FACSPresto™ Performance SummaryPrecision

Page 6: BD  FACSPresto ™ System Workflow

BD FACSPresto™ Performance Conclusion

In development field studies, the BD FACSPresto Near-Patient CD4 Counter provided accurate and precise clinical results for CD4, %CD4, and total Hb, using capillary or venous samples from HIV patients in hospital clinics in the USA and Kenya. The system is suitable for point-of-care CD4 enumeration.