Basic Pacing Concepts Basic Pacing Concepts Part II Part II
Dec 14, 2015
Basic Pacing ConceptsBasic Pacing ConceptsPart IIPart II
Electrical ConceptsElectrical Concepts
Voltage
Current
Impedance
Every Electrical Pacing Circuit Has Every Electrical Pacing Circuit Has the Following Characteristics:the Following Characteristics:
VoltageVoltage
Voltage is the force or “push” that causes electrons to move through a circuit
In a pacing system, voltage is:
– Measured in volts
– Represented by the letter “V”
– Provided by the pacemaker battery
– Often referred to as amplitude
CurrentCurrent
The flow of electrons in a completed circuit
In a pacing system, current is:
– Measured in mA (milliamps)
– Represented by the letter “I”
– Determined by the amount of electrons that move through a circuit
ImpedanceImpedance
The opposition to current flow
In a pacing system, impedance is:
– Measured in ohms
– Represented by the letter “R” (for numerical values)
– The measurement of the sum of all resistance to the flow of current
Voltage, Current, and Impedance Voltage, Current, and Impedance Are InterdependentAre Interdependent
The interrelationship of the three components can be likened to the flow of water through a hose
– Voltage represents the force with which . . .
– Current (water) is delivered through . . .
– A hose, or lead, where each component represents the total impedance:
The nozzle, representing the electrode
The tubing, representing the lead wire
Voltage and Current FlowVoltage and Current Flow
Spigot (voltage) turned up(high current drain)
Spigot (voltage) turned low(low current drain)
Resistance and Current FlowResistance and Current Flow
“Normal” resistance
“Low” resistance
“High” resistanceLow current flow
High current flow
Ohm’s Law is a Fundamental Ohm’s Law is a Fundamental Principle of Pacing That:Principle of Pacing That:
VV
II RRV = I X RV = I X RI = V / RI = V / RR = V / IR = V / I
Describes the relationship between voltage, current, and resistance
xx
If you reduce the voltage by half, the current is also cut in half
If you reduce the impedance by half, the current doubles
If the impedance increases, the current decreases
When Using Ohm’s Law When Using Ohm’s Law You Will Find That:You Will Find That:
Ohm’s Law Can Be Used to Find Amounts of Ohm’s Law Can Be Used to Find Amounts of Current Passing Through Pacemaker CircuitryCurrent Passing Through Pacemaker Circuitry
If: Voltage = 5 V
Impedance = 500
What will the current be?
I = V/R
I = 5 V ÷ 500 = 0.010 Amperes
0.010 x 1000 = 10 mA
In This Example, the Voltage is HalvedIn This Example, the Voltage is Halved
If: Voltage = 2.5 V
Impedance = 500
Current = ?
I = V/R
V = 2.5 V ÷ 500 = 0.005 Amperes
0.005 x 1000 = 5 mA
In This Example, the Impedance is In This Example, the Impedance is Reduced By HalfReduced By Half
If: Voltage = 5 V
Impedance = 250
Current = ?
I = V/R
I = 5 V ÷ 250 = 0.020 Amperes
0.020 x 1000 = 20 mA
Impedance Changes Affect Pacemaker Impedance Changes Affect Pacemaker Function and Battery LongevityFunction and Battery Longevity
High impedance reading reduces battery current drain and increases longevity
Low impedance reading increases battery current drain and decreases longevity
Impedance reading values range from 300 to 1,000
– High impedance leads will show impedance reading values greater than 1,000 ohms
Lead Impedance Values Will Change Due to:Lead Impedance Values Will Change Due to:
Insulation breaks
Wire fractures
An Insulation Break Around the Lead Wire An Insulation Break Around the Lead Wire Can Cause Impedance Values to FallCan Cause Impedance Values to Fall
Insulation breaks expose the wire to body fluids which have a low resistance and cause impedance values to fall
Current drains through the insulation break into the body which depletes the battery
An insulation break can cause impedance values to fall below 300
Insulation break
Decreased resistance
A Wire Fracture Within the Insulating Sheath A Wire Fracture Within the Insulating Sheath May Cause Impedance Values to RiseMay Cause Impedance Values to Rise
Impedance values across a break in the wire will increase
Current flow may be too low to be effective
Impedance values may exceed 3,000
Lead wire fracture
Increased resistance
StimulationStimulation
Stimulation ProcessStimulation Process
Time (Milliseconds)100 200 300 400 500
Phase 2
Phase 1
Phase 3
Phase 4
Tra
nsm
emb
ran
e P
ote
nti
al(M
illiv
olt
s)
-50
0
50
-100
Ph
ase
0
Threshold
Stimulation ThresholdStimulation Threshold
The minimum electrical stimulus needed to consistently capture the heart outside of the heart’s refractory period
VVI / 60
Capture Non-Capture
Amplitude
Pulse width
Two Settings Are Used to Ensure Capture:Two Settings Are Used to Ensure Capture:
Amplitude is the Amount of Voltage Amplitude is the Amount of Voltage Delivered to the Heart By the PacemakerDelivered to the Heart By the Pacemaker
Amplitude reflects the strength or height of the impulse:
– The amplitude of the impulse must be large enough to cause depolarization ( i.e., to “capture” the heart)
– The amplitude of the impulse must be sufficient to provide an appropriate pacing safety margin
Pulse Width Is the Time (Duration) Pulse Width Is the Time (Duration) of the Pacing Pulseof the Pacing Pulse
Pulse width is expressed in milliseconds (ms)
The pulse width must be long enough for depolarization to disperse to the surrounding tissue
5 V
0.5 ms 0.25 ms 1.0 ms
The Strength-Duration CurveThe Strength-Duration Curve
The strength-duration curve illustrates the relationship of amplitude and pulse width
– Values on or above the curve will result in capture
DurationPulse Width (ms)
.50
1.0
1.5
2.0
.25S
tim
ula
tio
n T
hre
sho
ld (
Vo
lts)
0.5 1.0 1.5
Capture
Clinical Usefulness of the Clinical Usefulness of the Strength-Duration CurveStrength-Duration Curve
Adequate safety margins must be achieved due to:
– Acute or chronic pacing system
– Daily fluctuations in threshold
Capture
0.5 1.0 1.5Duration
Pulse Width (ms)
.50
1.0
1.5
2.0
.25
Sti
mu
lati
on
Th
resh
old
(V
olt
s)
After Patient Safety, the Second Most Important Goal After Patient Safety, the Second Most Important Goal in Programming is to Extend Battery Lifein Programming is to Extend Battery Life
The best way to extend the service life of a battery is to lower voltage settings while maintaining adequate safety margins
– Amplitude values greater than the cell capacity of the pacemaker battery require a voltage multiplier, resulting in decreased battery longevity
Lead impedance
Amplitude and pulse width setting
Percentage paced vs. intrinsic events
Rate responsive modes programmed “ON”
Factors That Affect Battery Factors That Affect Battery Longevity Include:Longevity Include:
Electrode Design May Also Impact Electrode Design May Also Impact Stimulation ThresholdsStimulation Thresholds
Lead maturation process
Lead Maturation ProcessLead Maturation Process
Fibrotic “capsule” develops around the electrode following lead implantation
Steroid Eluting LeadsSteroid Eluting Leads
Steroid eluting leads reduce the inflammatory process and thus exhibit little to no acute stimulation threshold peaking and low chronic thresholds
Porous, platinized tipfor steroid elution
Silicone rubber plugcontaining steroid
Tines forstablefixation
Lead Maturation ProcessLead Maturation Process
Effect of Steroid on Stimulation Thresholds
Pulse Width = 0.5 msec
03 6
Implant Time (Weeks)
Textured Metal Electrode
Smooth Metal Electrode
1
2
3
4
5
Steroid-Eluting Electrode
0 1 2 4 5 7 8 9 10 11 12
Vol
ts
General Medtronic Pacemaker DisclaimerINDICATIONS
Medtronic pacemakers are indicated for rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity (Thera, Thera-i, Prodigy, Preva and Medtronic.Kappa 700 Series) or increases in activity and/or minute ventilation (Medtronic.Kappa 400 Series).
Medtronic pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include various degrees of AV block to maintain the atrial contribution to cardiac output and VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm.
9790 Programmer
The Medtronic 9790 Programmers are portable, microprocessor based instruments used to program Medtronic implantable devices.
9462
The Model 9462 Remote Assistant™ is intended for use in combination with a Medtronic implantable pacemaker with Remote Assistant diagnostic capabilities.
CONTRAINDICATIONS
Medtronic pacemakers are contraindicated for the following applications:
Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias.
Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms.
Unipolar pacing for patients with an implanted cardioverter-defibrillator because it may cause unwanted delivery or inhibition of ICD therapy.
Medtronic.Kappa 400 Series pacemakers are contraindicated for use with epicardial leads and with abdominal implantation.
WARNINGS/PRECAUTIONS
Pacemaker patients should avoid sources of magnetic resonance imaging, diathermy, high sources of radiation, electrosurgical cautery, external defibrillation, lithotripsy, and radiofrequency ablation to avoid electrical reset of the device, inappropriate sensing and/or therapy.
9462
Operation of the Model 9462 Remote Assistant™ Cardiac Monitor near sources of electromagnetic interference, such as cellular phones, computer monitors, etc. may adversely affect the performance of this device.
See the appropriate technical manual for detailed information regarding indications, contraindications, warnings, and precautions.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Medtronic Leads
For Indications, Contraindications, Warnings, and Precautions for Medtronic Leads, please refer to the appropriate Leads Technical Manual or call your local Medtronic Representative.
Caution: Federal law restricts this device to sale by or on the order of a Physician.
Note:
This presentation is provided for general educational purposes only and should not be considered the exclusive source for this type of information. At all times, it is the professional responsibility of the practitioner to exercise independent clinical judgment in a particular situation.
Continued inContinued in
Basic Pacing ConceptsBasic Pacing ConceptsPart IIIPart III