Balances The balance is an instrument which measures the mass of a body or substance using the gravity force which acts on that body At present, there are two large groups of balances: mechanical and electronic balances.
22
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Balances

BalancesThe balance is an instrument which measures the mass of a body or substance using the gravity force which acts on that bodyAt present, there are two large groups of balances: mechanical and electronic balances.

Page 2: Balances

Spring balance

Page 3: Balances

Sliding weight balance.

Page 4: Balances

Analytical balance.

Page 5: Balances

Mechanical Balances

1.0 Spring balance

2.0 Sliding weight balance.

3.0 Analytical balance

Its function is based on a mechanical property of springs as the force exercised on a spring

This type of balance is equipped with two known weights which can be moved on setting scales

Page 6: Balances

Operation Verification Verify that the balance is leveled.Test the zero mechanism.Verify and adjust the sensitivity.a) Lock the balance.b) Place a standard weight (equivalent to the optical scale range) on

the plate. c) Position the micro setting to one (1). d) Release the balance. e) Adjust to the zero position.

Page 7: Balances

Maintenance of the mechanical balanceDaily Annually

1. Verify the level. 2. Verify the zero setting.3. Verify the sensitivity

adjustment. 4. Clean the weighing

plate.

1. Calibrate the balance and document the process.

2. Disassemble and clean the internal components.

Page 8: Balances

Electronic balancesThe weighing plate.A measuring deviceElectronic controlling parts and display

Page 9: Balances

The weighing plate.

Page 10: Balances

load cell

Page 11: Balances

Theory of operationElectronic balances are mostly used two

different technologies. Load cell / Strain gauge

and electromagnetic force restoration.Load cell principle is used for all scales with

higher limits and less accuracy like patient and baby scales, while high quality analytical balances are used electromagnetic force restoration principle.

Page 12: Balances

Load cell / Strain gauge.A load on the pan creates as force and

the load cell gets deformed. Length-sensitive resistors get

deformed and change their resistance. Amplifier creates a voltage out of the

current through the resistances and delivers a measurement voltage which corresponds to the weight.

 A Wheatstone bridge with three fixed resistors and a sensor resistor. When the bridge is in balance the output voltage is zero.

Page 13: Balances

Electromagnetic force restoration The method give more precise measurements

than the load cell principle. The coil is floating in a magnetic field which is

created by an amplifier. The amplifier delivers always the right current to keep the lever in balance, regardless of the weight on the pan.

A optical sensor detects the balance of the leaver. The needed current for this balancing is

proportional to the weight on the pan, the current is also used to measure the weight. Therefore the current flows through a precision resistor which creates a measurement voltage.

An A/D unit transforms the voltage into a digital signal. The following micro processor with an appropriate software delivers the measurement result which is finally show on the display.

Additional sensors like a displacement detector and a temperature sensor are also controlling the amplifier and ensure the accuracy of the balance.

Page 14: Balances

Microprocessor base Electronic Balance

Page 15: Balances

Power Supply

1. The input socket with the plug from the external power supply.

2. The glass fuse. 3. A capacitor. Works together

with the following coil as a filter for the DC input voltage.

4. Looks like a transformer, but is just a coil for filtering out any AC components of the DC  voltage.

5. Three stabilizers for three different DC output voltages.

6. Capacitors for stabilizing the output voltages.

Page 16: Balances
Page 17: Balances

Maintenance Check list Clean the weighing plate, Remove any contaminations.Check door open and closed is smooth.Clean the weighing chamber, externally and internally.Check the “Zero point display ” is stable. Verify that the adjustment mechanisms on the front door of

the weighing chamber works adequatelyclean, pre-weighed container for weighing (glass container

or weighing paper if possible). Note that plastic can become electromagnetically charged and is not recommended for weighing powdered or granulated chemicals.

Any spill must be cleaned immediately to avoid corrosion or contamination. Use 70% ethanol to disinfect the pan of the balance.

Page 18: Balances

Basic fault IdentificationThe weight reading is

incorrect.The balance’s reader is

unstable.Functional Error

Page 19: Balances

1.0 The weight reading is incorrect.

The balance was not adjusted to zero before the reading.

The balance is incorrectly calibrated.

The balance is not leveled.

Place the balance on zero; repeat the measurement

Place the balance on zero; repeat the measurement

Page 20: Balances

2.0 The balance’s reader is unstable.

The balance’s reader is unstable.

The front door of the balance is open.

Place the balance on a stable surface

Close the front door to measure

Page 21: Balances

Functional Error

Readings not reproducible (hysteresis).

Non-linear readings

The screen indicates an overload or negative condition without a load being applied.

The measurement cell is dirty.The measurement cell is badly

assembled.

Non-linear readingsMechanical system is in bad

condition.

Measuring cell damaged by overload.

Measuring cell is inadequately assembled.