Top Banner

of 11

B20M01 8-part Eye Examination.pdf

Feb 28, 2018

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    1/11

    8-Part Eye Examination BLOCK 20MODULE 0

    Ophthalmologists

    Page 1of 11

    SUMMARY/OUTLINE

    I. Distance Acuity Test

    II.

    External Exam

    III.

    Pupillary Exam

    IV.

    Motility Exam

    V.

    Visual Fields (Confrontation Test)

    VI. Tonometry

    VII.

    Ophthalmoscopy

    VIII.

    Slit-lamp Biomicroscopy

    DISTANCE VISUAL ACUITY

    1.

    Ask the patient to stand or sit at a designated testing

    distance (20 feet)

    2. Occlude the left eye (testing one eye at a time)

    3. Ask the patient to identify each letter in the chart, on

    the lines of successively smaller optotypes, until patient

    correctly identifies only half the optotypes on a line

    4. Note the corresponding acuity measurement shown at

    the line of the chart

    5.

    Repeat above steps for the left eye, with the right eye

    covered

    6. Retest acuity with the patients with low vision, e.g.

    counting fingers, hand motion, light perception, etc.

    Visual acuity can be tested either for distance or near,

    conventionally at 20 feet (6 meters) and 14 inches (33 cms)

    away, respectively, but distance acuity is the general standard

    for comparison. For diagnostic purposes visual acuity is

    always tested separately for each eye, whereas binocularvisual acuity is useful for assessing functional vision, such as

    for assessing the eligibility to drive.Vaugn Asburys General Ophthalmology 39

    thEdition

    SNELLENS CHART

    Distance visual acuity test should be the first thing to do

    during the 8 part eye exam. You should do this before

    palpating, or putting any eye medication or drugs so that

    visual acuity is not altered because of the prior tests

    administered.

    In using the snellens chart, place the patient 20 feet away

    from the chart.

    Remember to test each eye separately and be tested with

    and without corrective lenses

    But in some clinics, there is only limited space, so they

    minimize the distance by converting them to smallerdistances.

    The largest letter E is equivalent to 20/200. If the patien

    cannot read the letter E, move the patient 5 feet closer

    until the patient reaches 5 feet away. Note the distance

    where the patient is able to read the letter E. If the patient

    still cannot read at 5 feet, proceed to counting finger test.

    Some clinics may use meters. Patients usually placed 6

    meters away from the chart.

    Visual acuity is scored as a fraction (eg, 20/40). The first

    number represents the testing distance between the chart

    and the patient, and the second number represents thesmallest row of letters that the patients eye can read. Hence

    normal vision is 20/20 and 20/60 acuity indicates that the

    patients eye can only read from 20 feet letters large enough

    for a normal eye to read from 60 feet.Vaugn Asburys General Ophthalmology 39

    thEdition

    For pediatric patients, you can use figures or the tumbling

    E charts

    Other patients can use the Jaeger chart. the patient is

    scored depending on which line of sentences he can read

    J10 is the biggest and J1 is the normal acuity.

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    2/11

    Page 2of 11

    Example result

    With

    correction

    Without

    correction

    FAR SIGHT

    OS 20/150 20/20

    OD 20/300 20/20

    NEAR SIGHT

    OS J8 J1

    OD J10 J1

    COUNTING FINGERS

    By convention, ophthalmologists test first the right eye. If

    it is known that the other eye is buron maybe the patient

    complained of it before assessment, check the buron eye

    first.

    In counting fingers, place your finger 1 ft away from the

    patient and let the patient count the fingers you are

    showing to him. If the patient cannot count at 1 feet away,

    move 1 more feet away until you reach the maximum of 3

    feet. Record the distance where the patient can count the

    fingers shown to him.

    After the maximum of 3 feet and the patient cannot count

    the fingers, proceed to light perception test.

    HAND MOTION TEST

    Make sure that you move your hands against the light.

    LIGHT PERCEPTION TEST

    Make sure to turn off the lights

    The patient unable to read the largest (20/200) letter on a

    Snellen chart should be moved closer to the chart until that

    letter can be read. The distance from the chart is then

    recorded as the first number. Visual acuity of 5/200 means

    that the patient can identify correctly the largest letter from a

    distance of 5 feet but not further away. An eye unable to read

    any letters is tested by the ability to count fingers. CF at 2 ft

    indicates that the eye was able to count fingers held 2 feet

    away but not farther away. If counting fingers is not possible,

    the eye may be able to detect a hand moving vertically or

    horizontally (HM, or hand motions vision). The next lower

    level of vision would be the ability to perceive light (LP, or

    light perception). An eye that is totally blindis recorded as

    having no light perception (NLP).Vaugn Asburys General Ophthalmology 39

    thEdition

    EXTERNAL EXAMINATION

    1.

    Observe the facial skin for any dermal or vascular

    changes; note any lesions or evidence of trauma

    2. Note any significant asymmetry of facial bones

    3. Note the lid position; assess effectiveness of eyelid

    closure and strength of the orbicularis muscles if

    appropriate

    4. Palpate the bony orbit for any lesion or deformity

    This is performed before studying the eye under

    magnification

    Gross inspection and palpation: lesions, growth,

    inflammatory signs (swelling, erythema, warmth,

    tenderness)

    Check for the ff:

    o position of eyelids (ptosis, lid retraction)

    o asymmetry can be quantified by measuring the

    central width (in mm) of palpebral fissure (space bet

    lower and upper lid margins)

    o abnormal motor fxn of the lids (upper lid elevation,

    forceful lid closure) may be due to neurologic o

    primary muscular abnormalities

    o malposition of the globe (proptosis) that may occur

    in orbital disease

    o bony orbital rim and periocular soft tissue

    General facial evaluation:

    o enlarged preauricular LN, sinus tenderness, tempora

    artery prominence, skin/mucous membrane

    abnormalities

    o You canauscultate for bruit at the temporal side of

    the orbit or directly at the globe. This can help youdetect carotid sinus fistula

    PUPILLARY EXAMINATION

    1. Turn off the light to decrease the room illumination

    2. Ask the patient to maintain fixation on a distance target

    3.

    Shine a bright handheld light directly into the right eye

    by approaching it from the side or from below

    4. Record the direct pupillary response to light in the right

    eye in terms of briskness of the response; observe the

    consensual reflex by noting the response to light of the

    non-illuminated pupil5. Repeat above steps for the left eye

    6. Enumerate the steps in performing the swinging flash

    light test and explain the clinical significance of relative

    afferent papillary defect (Marcus Gunn pupil)

    Assessment of pupil function should be done before any

    drops are instilled in the eye and before the cornea is

    touched (eg, applanation tension or Schirmer test).

    Examination of the pupils with a light stimulus provides

    evidence of the health of both the afferent and efferent

    systems. In addition to light, the pupils also respond to

    accommodation and convergence for clear and single vision

    at near. The pupils will constrict equally when either

    accommodation or convergence is stimulated by a nea

    object. When all 3 actionsaccommodation, convergence

    and miosisoccur simultaneously, this is called the synkinetic

    near response.

    Pupil function is evaluated with a bright penlight or other

    intense, small light in a dimly illuminated room. Pupil or iris

    abnormalities found with the naked eye can then be more

    thoroughly evaluated using the biomicroscope. The pupils are

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    3/11

    Page 3of 11

    evaluated for size, shape, direct light response, consensual

    response, and near response.

    SIZE

    In dim illumination, the average pupil diameter is 3 or 4 mm.

    This is ascertained by shining the light from below the

    patients nose so that the pupils are just visible to the

    examiner; the light is not shone directly into the patients

    eye. Pupils smaller than 2 mm are said to be miotic; pupilslarger than 6 mm are mydriatic. Miotic pupils may be caused

    by antiglaucoma medications, chronic iris inflammation, age,

    or a neurologic disorder. Abnormal mydriasis is caused by

    certain drugs, neurologic disorders, iris injury, or acute

    glaucoma. The pupils should be equal in size, although a

    small difference (1 mm) may be a normal variation. I f they are

    unequal (anisocoria), the difference between them should be

    further evaluated in both dark and bright room illumination.

    SHAPE

    Both pupils should be round. The pupils are normally

    centered or a little nasal in the iris. An eccentric pupil may be

    the result of faulty embryonic development, injury,

    intraocular surgery, or inflammation. In addition to beingeccentric, a pupil may also have an unusual shape.

    DIRECT LIGHT RESPONSE

    In dim room illumination, the patient is instructed to look at a

    distant target (this prevents the pupillary response to a near

    stimulus). The light source is presented to each eye

    separately and slightly off center to avoid the near response.

    Each pupil should exhibit a brisk response and constrict to

    about 2 mm.

    CONSENSUAL RESPONSE

    The consensual response is the simultaneous and equal

    response of one pupil when the other pupil is being

    stimulated by direct illumination or a near target. If the

    stimulated pupil constricts normally, then the consensual

    response of the other pupil will produce equal constriction

    without direct light stimulus.

    The pupils should be symmetric, and each one should be

    examined for size, shape (circular or irregular), and reactivity

    to both light and accommodation. Pupillary abnormalities

    may be due to (1) neurologic disease, (2) intraocular

    inflammation causing either spasm of the pupillary sphincter

    or adhesions of the iris to the lens (posterior synechiae), (3)

    markedly elevated intraocular pressure causing atony of thepupillary sphincter, (4) prior surgical alteration, (5) the effect

    of systemic or eye medications, and (6) benign variations of

    normal.

    SWINGING PENLIGHT TEST FOR MARCUS GUNN PUPIL

    As a light is swung back and forth in front of the two pupils,

    one can compare the reactions to stimulation of each eye,

    which should be equal. If the neural response to stimulation

    of the left eye is impaired, the pupil response in both eyes

    will be reduced on stimulation of the left eye compared to

    stimulation of the right eye. As the light is swung from the

    right to the left eye, both pupils will begin to dilate normally

    as the light is moved away from the right eye and then not

    constrict or paradoxically widen as the light is shone into the

    left eye (since the direct response in the left eye and

    the consensual response in the right eye are reduced

    compared to the consensual response in the left eye and

    direct response in the right eye from stimulation of the righ

    eye). When the light is swung back to the right eye, both

    pupils will begin to dilate as the light is moved away from the

    left eye and then constrict normally as the light is shone into

    the right eye. This phenomenon is called a relative afferen

    pupillary defect (RAPD).Vaugn Asburys General Ophthalmology 39

    thEdition

    MOTILITY EXAMINATION

    1. Sit facing the patient. Hold finger on or small fixation

    target at eye level about 10-14 inches in front of the

    patient, with the patient looking straight ahead.

    2. Ask the patient to follow the target as you move it into

    the six cardinal fields and up and down along the midlineElevate the upper eyelid with a finger on your free to

    observe downgaze

    3. Note whether the amplitude of eye movement is norma

    or abnormal in both eyes

    4.

    Note any nystagmus that may be present

    5. Determine alignment using the Hirschberg method of

    corneal light reflection test hold a penlight in front o

    the patient eyes at a distance of approximately 2 feet

    directing the light at the midpoint between the two eyes

    of the patient; instruct the patient to look directly at the

    light; compare the position of the two corneal light

    reflections and record the estimated result

    The extraocular muscles include: the medial, inferior, and

    superior recti, the inferior oblique, and levator palpebrae

    muscles, all innervated by the oculomotor nerve (III); the

    superior oblique muscle, innervated by the trochlear nerve

    (IV); and the lateral rectus muscle, innervated by the

    abducens nerve (VI).

    The precise action of any muscle depends on the orientation

    of the eye in the orbit and the influence of the orbita

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    4/11

    Page 4of 11

    connective tissues, which regulates the direction of action

    of the extraocular muscles by acting as their functional

    mechanical origins (the active pulley hypothesis).

    Mnemonic: "SO-4, LR-6, All the rest 3" (ie Superior Oblique by

    CN 4, Lateral rectus by CN 6, and all the other EOMs by CN 3).

    TESTING OF THE VISUAL FIELD

    The patient is asked to follow a target with both eyes as it is

    moved in each of the four cardinal directions of gaze. The

    examiner notes the speed, smoothness, range, and symmetry

    of movements and observes for unsteadiness of fixation (eg,nystagmus). Impairment of eye movements can be due to

    neurologic problems (eg, cranial nerve palsy), primary

    extraocular muscular weakness (eg, myasthenia gravis), or

    mechanical constraints within the orbit limiting rotation of

    the globe (eg, orbital floor fracture with entrapment of the

    inferior rectus muscle). Deviation of ocular alignment that is

    the same amount in all directions of gaze is called comitant.

    It is incomitant if the amount of deviation varies with the

    direction of gaze.Vaugn Asburys General Ophthalmology 39

    thEdition

    SIMPLE TEST OF BINOCULAR ALIGNMENT/ HIRSCHBERG

    METHOD

    Procedure:

    1. Have the patient look toward a penlight held several feet

    away. (33 cms according to Vaughan & Asburys 18th ed.

    Page 244)

    2. Note for the pinpoint light reflection, or reflex,.

    Note: In normal eyes, pinpoint light reflection, or reflex,

    should appear on each cornea and should be centered over

    each pupil if the two eyes are straight in their alignment.

    If the eye positions are convergent, such that one eye points

    inward (esotropia), the light reflex will appear temporal to

    the pupil in that eye. If the eyes are divergent, such that one

    eye points outward (exotropia), the light reflex will be

    located more nasally in that eye.Vaugn Asburys General Ophthalmology 39

    thEdition

    Landmarks to remember:

    0mm Light reflection is at the center of the pupil

    1mm Light reflection is in between the center of the pupil

    and the pupillary border

    2mm Light reflection is at the pupillary border

    3mm Light reflection is in between pupillary border and

    limbus

    4mm Light reflection is at the limbus

    KIMPSKY TEST

    The Krimsky test is essentially the Hirschberg test, but with

    prisms employed to quantitate deviation of ocular

    misalignment by determining how much prism is required to

    centre the reflex [2] The Krimsky test is advisably used fo

    patients with tropias, but not with phorias.https://en.wikipedia.org/wiki/Hirschberg_test

    COVER TEST

    - More accurate method of verifying normal ocular

    alignment.The test requires good vision in both eyes.

    Procedure:

    1. Ask the patient to gaze at a distant target with both

    eyes open. If both eyes are fixating together on the

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    5/11

    Page 5of 11

    target, covering one eye should not affect the

    position or continued fixation of the other eye.

    2. Suddenly covers one eye and carefully watches to see

    that the second eye does not move (indicating that it

    was fixating on the same target already). If the

    second eye was not identically aligned but was

    instead turned abnormally inward or outward, it

    could not have been simultaneously fixating on the

    target. Thus, it will have to quickly move to find the

    target once the previously fixating eye is covered.

    Fixation of each eye is tested in turn.

    Note!

    An abnormal cover test is expected in patients with diplopia.

    However, diplopia is not always present in many patients with

    long-standing ocular misalignment. When the test is

    abnormal, prism lenses of different power can be used to

    neutralize the refixation movement of the misaligned eye

    (prism cover test). In this way, the amount of eye deviation

    can be quantified based on the amount of

    prism power needed.

    BRUCKNER TEST

    The Brckner test is a qualitative assessment of ocula

    alignment. This test is done under dark room illumination

    and the direct ophthalmoscope aperture set on the largest

    aperture setting so as to equally illuminate both eyes. The

    examiner is viewing at about 1 meter away and observing the

    relative brightness of the fundus reflex from each eye. A

    whiter and brighter reflex is noted in the eye that is

    strabismic. To confirm a difference in color, retestmonocularly to note any changes to the reflex. The

    strabismic eye will appear whiter and brighter as a result of

    the fundus reflection emanating from outside of the

    pigmented macula region.http://apps.ketchum.edu/ceonline/courseview.a...

    VISUAL FIELDS EXAMINATION (CONFRONTATION TEST)

    1. Seat the patient and make sure the eye not being testes

    is occluded

    2. Seat facing the patient at a distance of about 1m. close

    your eye that is directly opposite the patients occluded

    eye

    3.

    Ask the patient to fixate on your nose or on your open

    eye

    4. Hold your hands stationary midway between yoursel

    and the patient is opposite quadrants about 30 degrees

    from central fixation

    5. Quickly extend then retract a finger or fingers on one

    hand in one quadrant of the monocular field asking the

    patient to state the number of fingers seen

    6.

    Repeat all four quadrants, testing at least twice per

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    6/11

    Page 6of 11

    quadrant

    7. Diagram the confrontation field if an abnormality is

    detected

    Make sure to test each eye separately. When occluding

    the other eye, dont depress the eyeball because it may

    cause blurring of vision.

    During the 8part eye exam, we are expected to use only

    the manual test. Automated perimetry is done in sophisticated diagnostic

    centers. Is uses a machine that accurately records the

    visual fields. The machine can also determine how much

    light the patient can see by altering the intensity of the

    light being shown to the visual fields.

    It is advisable to make a diagram as to which visual field

    is blinded because it can guide you in diagnosis

    If the patient has bitemporal hemianopia (number 2)

    what possible disease can lead to such problem knowing

    that the optic chiasm is the structure affected? Pituitary

    tumors.

    In automated perimetry:

    TONOMETRY

    1. Enumerate and differentiate the methods of measuring

    intraocular pressure

    globe can be thought of as an enclosed compartment

    through which there is a constant circulation of aqueous

    humor

    this fluid maintains the shape and a relatively uniform

    pressure within the globe tonometry is the method of measuring intraocular

    pressure using calibrated instruments.

    normal range is 10 to 21 mm Hg

    less corneal indentation is produced as intraocular

    pressure rises.

    since both methods employ devices that touch the

    patients cornea, they require topical anesthetic and

    disinfection of the instrument tip prior to use.

    with any method of tonometry, care must be taken to

    avoid pressing on the globe and artificially increasing its

    pressure.

    APPLANATION TONOMETRY

    - intraocular pressure is determined by the force required to

    flatten the cornea by a standard amount. The force required

    increases with intraocular pressures.

    - the GOLDMANN APPLANATION TONOMETERis attached to

    the slitlamp and measures the amount of force required to

    flatten the corneal apex by a standard amount.

    - the higher the intraocular pressure, the greater the force

    required.

    - Goldmann applanation tonometer is a more accurate

    method than Schiotz tonometry

    - following topical anesthesia and instillation of fluorescein

    the patient is positioned at the slitlamp and the tonometer isswung into place. To visualize the fluorescein, the cobalt blue

    filter is used with the brightest illumination setting. After

    grossly aligning the tonometer in front of the cornea, the

    examiner looks through the slitlamp ocular just as the tip

    contacts the cornea. A manually controlled counterbalanced

    spring varies the force applied by the tonometer tip.Upon

    contact, the tonometer tip flattens the central cornea and

    produces a thin circular outline of fluorescein. A prism in the

    tip visually splits this circle into two semicircles that appear

    green while viewed through the slitlamp oculars. The

    tonometer force is adjusted manually until the two

    semicircles just overlap, as shown in Figure 210. This visua

    end point indicates that the cornea has been flattened by the

    set standard amount. The amount of force required to do this

    is translated by the scale into a pressure reading in

    millimeters of mercury.

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    7/11

    Page 7of 11

    - accuracy of intraocular pressure measurement is affected by

    central corneal thickness. The thinner the cornea, the more

    easily it is indented, but the calibration of tonometers

    generally assumes a cornea of standard thickness. If the

    cornea is relatively thin, the actual intraocular pressure is

    higher than the measured value, and if the cornea is relatively

    thick, the actual intraocular pressure is lower than the

    measured value. Thus ultrasonic measurement of corneal

    thickness (pachymetry) may be helpful in assessment of

    intraocular pressure. The Pascal dynamic contour tonometer,

    a contact but non-applanating technique, measures

    intraocular pressure independent of corneal thickness.

    - other applanation tonometers are the Perkins tonometer, a

    portable mechanical device with a mechanism similar to the

    Goldmann tonometer, the Tono-Pen, a portable electronic

    applanation tonometer that is reasonably accurate but

    requires daily recalibration, and the pneumatotonometer,

    which is particularly useful when the cornea has an irregular

    surface. The Perkins tonometer and Tono-Pen are commonlyused when examination at the slitlamp is not feasible, for

    example, in emergency rooms in cases of orbital trauma with

    retrobulbar hemorrhage and in operating rooms during

    examinations under anesthesia.

    SCHIOTZ TONOMETRY

    - now rarely used, measures the amount of corneal

    indentation produced by preset weights

    - advantage of this method is that it is simple, requiring only a

    relatively inexpensive, easily portable hand-held instrument.

    It can be used in any clinic or emergency room setting, at the

    hospital bedside, or in the operating room, but it requires

    greater expertise and has generally been superseded by

    applanation tonometers.

    NONCONTACT TONOMETRY

    - noncontact (air-puff) tonometer is not as accurate as

    applanation tonometers.

    - small puff of air is blown against the cornea.

    - air rebounding from the corneal surface hits a pressure

    sensing membrane in the instrument.

    - does not require anesthetic drops, since no instrument

    touches the eye. Thus, it can be more easily used by

    optometrists or technicians and is useful in screening

    programs.

    OPHTHALMOSCOPY

    1.

    Position the patient about 2 feet away

    2. Turn off the light to dim the room illumination

    3. Set the focusing lens of the ophthalmoscope to zero

    4.

    Check the red reflex from a distance of 2 feet

    5.

    Approach the patients eye; the instrument is steadied

    against the patients face by resting the ulnar border of

    the hand holding the instrument against the patients

    cheek; the thumb of the free hand raises the upper lid

    6.

    Instruct the patient to stare into the distance

    7. Dial the ophthalmoscopes focusing lenses into place to

    clarify the fundus image

    8.

    Find the optic disc by following a retinal blood vessel

    9. Examine the peripapillary retina

    10. From the optic disc, follow the blood vessels outward to

    examine the four quadrants around the posterior pole

    11. Check for foveal reflex

    THE DIRECT OPHTHALMOSCOPE

    This instrument consists of a single aperture through which

    light is projected into the subjects eye and the examiner

    views the eye. It provides a magnified image (15) and a field

    of view of some 6.510 degrees. A set of corrective lenses can

    be dialled into the aperture. These enables the focal point of

    the instrument to be adjusted. The rack of lenses usually

    contains equal numbers of positive and negative spheres

    which can be dialled up to take account of the patient and/or

    examiners refractive status. If examiners wish to wear their

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    8/11

    Page 8of 11

    glasses, they can do so, and effectively they will need a zero

    lens in the eyepiece. The patients refraction must also be

    taken into account and the relevant lens dialed into place.

    With highly myopic or hypermetropic patients, their glasses

    can be left on and used to nullify the effect of the refractive

    variation. Alternatively plus and minus 10 or 20 D lenses can

    be positioned in the sight aperture to take account of very

    high hypermetropia or myopia. The size and brightness of the

    illumination spot can be varied with the appropriate controls.

    Additional features vary among the different models but

    include a slit filter, producing a vertical slit of light which can

    be used to examine contours or elevations on the fundus, a

    grid for assessing the size of a fundus lesion, and a green filter

    for red-free viewing. This latter filter will make red features,

    such as haemorrhages, stand out due to increasing contrast

    between the various shades of red and orange which reflect

    from the fundus. Some ophthalmoscopes also include a

    cobalt blue filter for use with fluorescein dye.

    The view obtained with this instrument has a narrow angle of

    view and a high magnification. The more myopic the patient,

    the more effective the magnifying effect. This is useful forexamining the optic nerve head; however the view is

    monocular and two-dimensional.

    METHOD OF USE

    1. Inform patients that you are going to look at their eye

    with a bright light and that you will have to get very close

    to their face. Instruct them to breathe normally.

    2.

    The instrument is held to the examiners eye with the

    illumination system switched on and for steadiness and

    ease of use a hand can be placed on the patients

    shoulder.

    3.

    The examiners right eye is used for the patients right

    eye and the examiners left for the patients left eye. If

    the examiner finds it difficult to close one eye, or the

    other, then it can be left open with practice the brain

    manages to ignore the image from the non-examining

    eye.

    4. The correct lens, as described above, is dialled into the

    aperture.

    5.

    The patient is asked to fix on a distant object and is told

    to maintain that fixation, regardless of whether the

    examiner gets in the way. The examiner thus knows

    roughly where the patients macula is situated and the

    optic disc will be just nasal to this.

    6.

    The examiner then points the instruments illuminationbeam into the patients pupil and obtains a red reflex

    from a distance of about half a metre and slowly moves

    towards the patient. At this point media opacities such as

    cataract can be seen as black features against the red

    reflex. The rheostat is used to adjust the brightness of

    the light for the patients comfort. I f required, the front

    of the eye, cornea, iris and lens can be examined with a

    +10 lens dialed into the instruments lens bank.

    7. Following this part of the examination the lens dial is

    progressively turned towards zero to focus further back

    into the patients eye and eventually reach the retina. It

    must be stressed that the head of the ophthalmoscope

    must be held very close to the patients eye in order to

    gain the maximum field of view.Clinical Skills for the Ophthalmic Examination: Basic Procedures, 2

    ndEdition

    *For parts of the opthalmoscope: see last page

    Macula is usually at temporal side while optic disc is on

    the temporal side.

    INDIRECT OPHTHALMOSCOPY

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    9/11

    Page 9of 11

    Binocular Indirect Ophthalmoscopy (BIO) is a technique that

    provides thorough view of the retina and vitreous through a

    dilated pupil in order to evaluate the health of the interior of

    the eye and to identify structural abnormalities that may be

    associated with reduced visual acuity thereby aiding the

    diagnosis of amblyopiaoptometry.osu.edu

    Comparison Between Direct and Indirect OphthalmoscopyDIRECT INDIRECT

    Magnified image Not magnified

    Can see only small area Lets you see bigger area

    If with cataract, cannot see Ideal if with cataracts

    One handheld apparatus Uses a head gear and a

    handheld condensing lenses

    SLIT-LAMP BIOMICROSCOPY

    1. Identify the different parts of the slit-lamp biomicroscope

    2. Enumerate the different uses of the slit-lamp

    The slitlamp is a table-mounted binocular microscope with a

    special adjustable illumination source attached. A linear slit

    beam of incandescent light is projected onto the globe,

    illuminating an optical cross section of the eye. The angle of

    illumination can be varied along with the width, length, and

    intensity of the light beam. The magnification can be adjusted

    as well (normally 10 to 16 power). Since the slitlamp is a

    binocular microscope, the view is stereoscopic, or three-

    dimensional.

    Slitlamp photograph of a normal right eye. The curved slit of

    light to the right is reflected off of the cornea (C), while the

    slit to the left is reflected off of the iris (I). As the latter slit

    passes through the pupil, the anterior lens (L) is faintly

    illuminated in cross section.

    The patient is seated while being examined, and the head is

    stabilized by an adjustable chin rest and forehead strap.Vaugn Asburys General Ophthalmology 39

    thEdition

    Parts of Slit lamp biomicroscopy

    Viewing Arm.The binocular eyepieces provide stereoscopic vision and can

    be adjusted to accommodate the examiner's interpupillary

    distance. The focusing ring can be twisted to suit the

    examiner's refractive error.

    The magnification element can be adjusted with the side dial.

    Illumination Arm

    The illumination arm can be swung 180 degrees side to side

    on its pivoting bases allowing the examiner to direct the light

    beam anywhere between the nasal and temporal aspect o

    the eye examination.The dimension of the light beam can be

    varied in height and width with these levers. It can provide

    diffuse or focal illumination as an optical cross-section of the

    anterior segment.Cobalt blue, or green filters can be selected

    with this lever.

    The Patient Positioning Frame

    The patient positioning frame consist of two upright meta

    rods to which are attached a forehead strap and a chin rest.

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    10/11

    Page 10of 11

    The chin rest height can be adjusted with the knob just below

    it.

    The JoystickThe joystick allows for focusing by shifting forward,

    backward, laterally or diagonally. The joystick can also be

    rotated to lower or elevate the light beam.

    The locking screw located at the base secures the slit lamp

    from movement when it is not in use.

    Just below the slit lamp table on the left is the ON switch and

    provides high or low options in light intensity.

    Uses of Slit lamp

    1. To visualize the anterior half of the globethe

    anterior segment.

    2.

    To study the details of the lid margins and lashes,

    the palpebral and bulbar conjunctival surfaces, the

    tear film and cornea, the iris, and the aqueous can

    be studied.

    3.

    Through a dilated pupil, the crystalline lens and theanterior vitreous can be examined as well.

    4. Because the slit beam of light provides an optica

    cross section of the eye, the precise anteroposterior

    location of abnormalities can be determined within

    each of the clear ocular structures (eg, cornea, lens

    vitreous body).

    5. The highest magnification setting is sufficient to

    show the abnormal presence of cells within the

    aqueous, such as red or white blood cells or pigment

    granules. Aqueous turbidity, called flare, resulting

    from increased protein concentration can be

    detected in the presence of intraocular

    inflammation. Normal aqueous is optically clear

    without cells or flare.Vaugn Asburys General Ophthalmology 39

    thEdition

    Other uses of slit lamp biomicroscope:- Internet source.

    1. Routine observation of ocular adnexia

    2. Routine investigation of posterior segment

    3. Monitoring signs and symptoms of anterior segment

    conditions

    4. Further "special eye" investigations

    Definition of TermsConjugate movement: Movement of the eyes in the same

    direction at the same time.

    Deviation: Magnitude of ocular misalignment, usually

    measured in prism diopters but sometimes measured in

    degrees.

    Comitant deviation: Deviation not significantly affected by

    which eye is fixing or direction of

    gaze, typically a feature of childhood (nonparetic) strabismus.

    Incomitant deviation: Deviation varies according to which

    eye is fixing and direction of gaze,

  • 7/25/2019 B20M01 8-part Eye Examination.pdf

    11/11

    Page 11 of 11

    usually a feature of recent onset extraocular muscle paresis

    and other types of acquired

    strabismus.

    Primary deviation: Incomitant deviation measured with the

    normal eye fixing).

    Secondary deviation: Incomitant deviation measured with

    the affected eye fixing.

    Ductions: Monocular rotations with no consideration of the

    position of the othereye.

    Adduction: Inward rotation.

    Abduction: Outward rotation.

    Supraduction (elevation): Upward rotation.

    Infraduction (depression): Downward rotation.

    Fusion: Formation of one image from the two images seen

    simultaneously by the two eyes.Fusion has two aspects.

    Motor fusion: Adjustments made by the brain in innervation

    of extraocular muscles in order to

    bring both eyes into bifoveal and torsional

    alignment.

    Sensory fusion: Integration in the visual sensory

    areas of the brain of images seen with the twoeyes into one picture.

    Heterophoria (phoria): Latent deviation of the eyes held

    straight by binocular fusion.

    Esophoria: Tendency for one eye to turn inward.

    Exophoria: Tendency for one eye to turn outward.

    Hyperphoria: Tendency for one eye to deviate upward.

    Hypophoria: Tendency for one eye to deviate downward.

    (See Hypotropia.)

    Heterotropia (tropia):

    Strabismus: Manifest deviation of the eyes that cannot be

    controlled by binocular vision.

    Esotropia: Convergent manifest deviation (crossed eyes).

    Exotropia: Divergent manifest deviation (wall eyes).

    Hypertropia: Manifest deviation of one eye upward.

    Hypotropia: Manifest deviation of one eye upward. By

    convention, in the absence of specific

    causation to account for the lower position of one eye,

    vertical deviations are designated by the

    higher eye (eg, right hypertropia, not left hypotropia, when

    the right eye is higher).

    Incyclotropia: Manifest rotation of the 12 oclock meridian of

    one eye about its anteroposterior

    axis toward the midline of the head.Excyclotropia: Manifest rotation of the 12 oclock meridian of

    one eye about its anteroposterior

    axis away from the midline of the head.

    Orthophoria: The absence of any tendency of either eye to

    deviate when fusion is suspended.

    This state is rarely seen clinically. A small phoria is normal.

    Prism diopter (PD): The unit of angular measurement used to

    characterize ocular deviations. A

    1 diopter prism deflects a ray of light toward the base of the

    prism by 1 cm at 1 m. One degree of

    arc equals approximately 1.7 PD.

    Secondary deviation: The deviation measured with the

    paretic eye fixing and the

    normal eye deviating.

    Torsion: Rotation of the eye about its anteroposterior axis

    Intorsion (incycloduction): Rotation of the 12 oclock meridian

    of the eye toward the midline of

    the head.

    Extorsion (excycloduction): Rotation of the 12 oclock

    meridian of the eye away from the

    midline of the head.

    Vergences (disjunctive movements): Movement of the two

    eyes in opposite directions.

    Convergence: The eyes turn inward.

    Divergence: The eyes turn outward.

    Versions: Binocular rotations of the eyes in qualitatively the

    same direction.