Top Banner

of 40

AWS Procedure - Rev 2.pdf

Jun 02, 2018

Download

Documents

gueridi
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/10/2019 AWS Procedure - Rev 2.pdf

    1/40

    Page 1of 40

    Forward

    While it is recognized that phased array technology is becoming more widespread and

    popular within various industries, it has not yet been specifically addressed by the

    AWS. The following document was written to satisfy customer requests. Its purpose is

    to serve as a guideline to align the Phasor XS with the current methodology prescribed

    in Clause 6, Part F, of the AWS D1.1, 2008 Code.

    This is achievable due to the well-established fact that an array transducer with the

    same physical properties of size, shape, and frequency, as a conventional monolithic

    transducer, will have identical acoustic properties.

    This document has been produced in typical procedure format, for the sake of

    familiarity. As such, it is intended as a guideline only.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    2/40

    Page 2of 40

    Table of contents

    SECTION PAGE

    1. Scope 3

    2. References 3

    3. Personnel 4

    4. Equipment 4

    5. Calibration 6

    6. Examination 13

    7. Indication evaluation and reporting 18

    Annex A: dB Accuracy Procedure 23

    Annex B: Horizontal Linearity Procedure 29

    Annex C: Search Unit Check 33

    Annex D: Drawings 37

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    3/40

    Page 3of 40

    1. Scope

    1.1.This procedure describes the use of the Phasor XS utilizing the sectorial and

    A-scan views. For the detection and interpretation of discontinuities in groovetype welds and the associated heat affected zones (HAZ). Applicable part

    thickness range is from 5/16 to 8 inclusive of material thickness as prescribed

    byPart F of the AWS D1.12008 code. (herein referenced as the code)

    1.2.Limitations

    1.2.1. This procedure is not intended for the inspection of tube-to-tube (tubular)

    T, Y, K connections

    1.2.2. This procedure is not intended for the inspection of base material.

    However, discontinuities discovered in the adjacent base material shall be

    reported to the Owner / Engineer for disposition.

    1.2.3. As a supplement to the method(s) described herein, RT is recommended

    for the detection of piping porosity in ESW, and EGW weldament.

    1.2.4. For discrepancies between this procedure, and the code, the code shall

    hold precedence.

    2. References

    2.1.American Welding SocietyAWS D1.1 Structural Welding Code Steel

    2.2.American Society for Nondestructive Testing (ASNT) - SNT-TC-1A Personnel

    Qualifications

    2.3.GE Inspection Technologies (GEIT) - Phasor Owner Manual.

    (www.geinspectiontechnologies.com)

    http://www.geinspectiontechnologies.com/http://www.geinspectiontechnologies.com/http://www.geinspectiontechnologies.com/http://www.geinspectiontechnologies.com/
  • 8/10/2019 AWS Procedure - Rev 2.pdf

    4/40

    Page 4of 40

    3. Personnel

    3.1. Personnel shall be certified in accordance with the ASNT SNT-TC-1A, or be

    certified in accordance with the employers written practice. Provided that the

    employers written practice contains, as a minimum, the requirements of SNT-

    TC-1A.

    3.2.Personnel performing exams shall be a certified level I, II, or a level III who has

    taken, and passed a Level II Practical certification test.

    3.2.1. Level I personnel shall only perform exams under the direct supervision

    of a level II, or III, and shall not independently (1) conduct exams, (2)

    interpret exam results, or (3) write a report of exam results.

    3.3.All personnel conducting exams in accordance with this procedure shall

    demonstrate proficiency by successfully passing a general, specific and

    practical exam administered by a level III, or a Level II so designated by a level

    III, and applying the methodology of this procedure and the rules of AWS D1.1.

    4. Equipment

    4.1.The ultrasonic instrument shall be the Phasor XS.

    4.1.1. For compliance with Sec. 6.22 of the AWS code, reference the Phasor

    Owners Manual for the technical specifications.

    4.2.The search unit shall be comprised of transducer model # 115-000-866,

    coupled to wedge # 118-350-037 for straight beam, or the standard 60 AWS

    wedge for angle beam. The use of other search units by engineer approval is

    not addressed within this document.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    5/40

    Page 5of 40

    4.3.The International Institute of Welding (IIW) Type 1 or Type 2 ultrasonic

    reference block(s) shall be used for distance and sensitivity calibration, a

    drawing of which can be found in Annex D, Drawing 2. Other approved

    reference blocks (DSC, DS, and SC) may also be used, provided that the

    reference sensitivity of the instrument/search unit combination is adjusted to

    achieve equivalent results as that of the IIW block. Drawings for these blocks

    can be found in Annex D, Drawing 4.

    4.4.Any suitable couplant material of glycerin type, or a cellulose gum and water

    mixture shall be used between the search unit and base material. A wetting

    agent may be added to improve consistency. Light machine oil or equivalent

    may be used for coupling on calibration blocks.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    6/40

    Page 6of 40

    5. Calibration

    5.1. General

    5.1.1. Reference Annex A and B for equipment qualification requirements.

    5.1.2. A transducer element, cable, and coupling check shall be made prior to

    an exam or series of exams, and again at the completion of an exam or

    series of exams, but not to exceed intervals of 8 hours, and at any time of

    suspected system irregularity. Reference Annex C for equipment

    programming, and guidance.

    5.1.3. The primary testing angle used for calibration shall be selected from

    table 5.1

    5.1.4. Calibration shall be performed with the Phasor set up in accordance with

    table(s) 5.2, or 5.3.

    5.1.5. Calibration intervals for sensitivity and horizontal sweep shall be made

    by the operator at the location of each weld being tested, and at intervals

    not to exceed 30 minutes as testing proceeds.

    5.1.6. Recalibration shall be made after:

    a) Change of personnel.

    b) Time intervals not to exceed a maximum of 2 hours.

    c) Any change to the search unit, or changing of the search unit.

    d) Any disruption of the electronic circuitry. (i.e. battery swap, powerfailure)

    e) Any instance of system irregularity.

    5.1.7. The use of corner reflectors shall be prohibited for calibration.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    7/40

    Page 7of 40

    5.2. Straight beam calibration

    5.2.1. With the Phasor programmed in accordance with Table 5.2, place the

    search unit on face A and adjust the leg function such that 2 plate

    thicknesses are displayed on the screen.

    5.2.2. For sensitivity, place the search unit at a location on the base material

    that is free of discontinuities. Adjust the gain control to display the first back

    reflection at 50% to 75% of amplitude. (A%A reading in the results

    window).

    5.3. Angle beam calibration

    5.3.1. A linear screen range shall be achieved to represent the actual sound

    path in metal. The following method describes the use of the IIW2

    reference block. However any suitable block described in 4.3 may be used.

    5.3.2. Using the primary testing angle, and with the Phasor programmed in

    accordance with Table 5.3, adjust the instrument to achieve a linear screen

    range as follows:

    5.3.2.1. Place the search unit at position D as depicted in Annex D

    Drawing 1.

    5.3.2.2. Adjust the leg function to display the 2 and 4 radius on the

    screen, maximizing the signal(s) amplitude. (NOTE: For this

    process, the Material Thickness function may also be adjusted.

    However, prior to examinations, the Material Thickness function

    shall be returned to the actual component thickness as prescribedin Table 5.3.)

    5.3.2.3. Place the beam cursor on the primary testing angle to ensure the

    readings in the Results windows are for this angle.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    8/40

    Page 8of 40

    5.3.2.4. In the CAL submenu (HOME > PART > CAL) adjust the gates to

    surround where the beam cursor intersects the reflection from the

    4 radius. (NOTE: the gates may also need to be narrowed.)

    Maximize the signal response from the 4 radii by adjusting the

    probe position. Adjust the gain such that the amplitude response

    reads less than 100% in the A%A results window. Adjust the

    VELOCITY S until the SA reading in the results window reads 4.

    5.3.2.5. In the same manner as described in 5.3.2.4, adjust the PROBE

    DELAY such that the SA result reads 2 for the response from the

    2 radii reflector.

    (NOTE: Alternately, the method described in the Phasor Owners

    Manual may be used to achieve a linear screen prescribed in 5.3.2.)

    5.3.3. With the Phasor programmed in accordance with table 5.3, adjust the

    DbD gain control such that the response from the 0.060hole, when the

    search unit is placed at position A from Annex D Drawing 1, is maximized

    with the primary testing angle from Table 5.1, to display a resulting

    amplitude in the A%A between 40% to 60%. The resulting DbD setting is

    the zero reference level and shall be recorded on the Ultrasonic Test

    Report Table 7.1 under Reference level (b).

    (NOTE: The reference level may also be achieved by using only one side

    of the gain in the Phasor. DbA or DbD. In this case toggle the steps key

    to display the gain being used in the home screen. To activate the Level

    above reference function, press and hold the orange key next to the gain

    sub-menu box. In either case it is important to note the sum of all gain

    applied above reference.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    9/40

    Page 9of 40

    Table 5.1

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    10/40

    Page 10of 40

    Table 5.1 (cont.)

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    11/40

    Page 11of 40

    Table 5.2

    PROBE MENU

    (Sub-menu) (Tab) (Value)

    PRB DAT Dialog Probe Load Def

    PRB GEO

    Frequency 2.3 MHz

    Number of Elements 16

    Pitch 0.040

    Elevation 0.650

    WDGE DAT Element 1 Position Low end

    WDGE GEO

    Velocity .1086 in/s

    Wedge Front 0.000

    Offset Z 1.00 (Annex C)

    Angle 0.0 deg

    OFFSET Origin Offset 0.000

    PART MENU

    (Sub-menu) (Tab) (Value)

    PLAN Material Thickness Enter material thickness

    CAL as described in section 5.2

    SCAN MENU

    (Sub-menu) (Tab) (Value)

    ELECTRNC

    Type Linear

    Focal Point Unfocused

    Wave Type Longitudinal

    SCN PATT

    Angle Start 0.0 deg

    Angle Stop 0.0 deg

    Angle Step 1.0 deg

    APERTURE

    Number of Steps 1

    First Element 1

    Aperture Size 16

    Aperture Step 1

    DISPLAY MENU

    (Sub-menu) (Tab) (Value)

    VIEW

    View Mode Frame

    Angle corrected OnData Source Amp A

    BACKGRND Color Leg On

    RESULTS1

    Reading 1 Beam

    Readings may bedisplayed in any order

    Reading 2 A%A

    Reading 3 SA

    Reading 4 PA

    UT MENU

    (Sub-menu) (Tab) (Value)

    BaseDisplay Start Material

    Display Delay 0.000 in

    Leg 2.3

    Pulser

    Voltage 150V

    Width 220ns

    PRF Value set to maximum

    ReceiverFrequency 2MHz

    Ascan Rectify Fullwave

    GATEMODE

    Gate select Gate A

    Logic Positive

    Gate Display On

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    12/40

    Page 12of 40

    Table 5.3

    PROBE MENU

    (Sub-menu) (Tab) (Value)

    PRB DAT Dialog Probe Load Def

    PRB GEO

    Frequency 2.3 MHz

    Number of Elements 16

    Pitch 0.040

    Elevation 0.650

    WDGE DAT Element 1 Position Low end

    WDGE GEO

    Velocity .1086 in/s

    Wedge Front 0.711

    Offset Z 0.435 (Annex C)

    Angle 47.5 deg

    OFFSET Origin Offset 0.000

    PART MENU

    (Sub-menu) (Tab) (Value)

    PLAN Material Thickness Enter material thickness

    CAL as described in section 5.3

    SCAN MENU

    (Sub-menu) (Tab) (Value)

    ELECTRNC

    Type Sector

    Focal Point Unfocused

    Wave Type Shear

    SCN PATT

    Angle Start 35.0 degminimum

    Angle Stop 70.0 deg

    Angle Step 1.0 deg

    APERTURE

    Number of Steps 1

    First Element 1

    Aperture Size 16

    Aperture Step 1

    DISPLAY MENU

    (Sub-menu) (Tab) (Value)

    VIEW

    View Mode Frame

    Angle corrected OnData Source Amp A

    BACKGRND Color Leg On

    RESULTS1

    Reading 1 Beam

    Readings may be

    displayed in any order

    Reading 2 A%A

    Reading 3 SA

    Reading 4 PA

    UT MENU

    (Sub-menu) (Tab) (Value)

    BaseDisplay Start Material

    Display Delay 0.000 in

    Leg 2.3

    Pulser

    Voltage 150V

    Width 220ns

    PRF Value set to maximum

    ReceiverFrequency 2MHz

    Ascan Rectify Fullwave

    GATEMODE

    Gate select Gate A

    Logic Positive

    Gate Display On

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    13/40

    Page 13of 40

    6. Examination

    6.1.General

    6.1.1. An X line for flaw location shall be marked on the test face of the

    weldament in a direction parallel to the weld axis. The location distance

    perpendicular to the weld axis is based on the dimensional figures on the

    detail drawing and usually falls on the centerline of the butt welds, and

    always falls on the near face of the connecting member of T and corner

    welds, the face opposite Face C. (see Table 5.1 for Face descriptions.)

    6.1.2. A Y accompanied with a weld identification number shall be clearlymarked on the base metal adjacent to the weld being examined at the left

    end on Face A (see Table 5.1 for Face descriptions.) This identification

    number serves as:

    a) Weld identification,

    b) Identification of Face A,

    c) Distance measurements and direction (+ or -) from the X line,

    d) Location measurements from weld ends or edges.

    6.1.3. A couplant described in 4.4 shall be used between the search unit and

    scanning surface.

    6.1.4. All surfaces to which the search unit(s) is applied shall be free of weld

    spatter, dirt, grease, oil, paint, loose scale, or any material/condition which

    would prevent the transmission of ultrasound, and shall have a contour

    allowing intimate coupling.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    14/40

    Page 14of 40

    6.2.Base metal examination

    6.2.1. The entire volume of base metal through which ultrasound must travel to

    examine the weld shall be examined for laminar reflectors utilizing the

    straight beam search unit calibrated in section 5.2. If any area of base

    material exhibits a total loss of back wall reflection or displays an indication

    equal to, or greater than the original back wall reflection height, and is

    located such that its presence will interfere with the transmission of the

    ultrasound for the weld inspection, its size, location, and depth from the A

    Face shall be determined and reported on the ultrasonic test report, and an

    alternate weld scanning procedure shall be used.

    6.2.1.1. When the indication is larger than the search unit a full loss of theback wall reflection will usually occur. By sliding the search unit off

    the indication until a 50% drop in amplitude from the indication is

    realized, and then by measuring to the centerline of the search

    unit, is generally reliable for determining the edges of the

    indication.

    6.2.1.2. The size evaluation of an indication which is smaller than the

    search unit must be made by sliding the search unit toward theindication until a signal begins to form. At this point, the leading

    edge of the search unit is indicative of the edge of the indication.

    6.2.2. If all or part of the weld is deemed inaccessible due to laminar

    indications, and cannot be examined in accordance with the

    requirements of table 5.1, the examination shall be conducted using one

    or more of the following alternate procedures, as necessary, to attain full

    weld coverage.

    a) Weld surfaces shall be ground flush in accordance with the

    applicable paragraph(s) of the code,

    b) Test from Faces A and B,

    c) Utilize alternate search angles.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    15/40

    Page 15of 40

    6.3. Weld examination

    6.3.1. Welds shall be tested with the search unit described in Section 4.2, and

    calibrated in Section 5.3. Following calibration, and during testing, the

    only instrument adjustment permitted is the sensitivity level (gain)

    setting.

    6.3.1.1. Sensitivity shall be increased from the reference level to scanning

    level in accordance with scanning levels described in Table 7.5A,

    or 7.5B as applicable.

    6.4. Scan patterns (see Figure 6.4)

    6.4.1. It is intended as a minimum that all welds are to be examined by passingsound through the entire volume of the weld and HAZ in two crossing

    directions, wherever practical.

    6.4.2. All butt welds shall be examined from each side of the weld axis. Corner

    and T welds shall be primarily tested from one side of the weld axis only.

    All Welds shall be examined using the applicable scanning pattern, or

    patterns as shown in Figure 6.4, as necessary to detect both longitudinal

    and transverse discontinuities.

    6.4.3. To detect longitudinal defects, the following movements may be

    combined from figure 6.4

    6.4.3.1. Scanning Movement A, Rotation angle a = +/-10deg.

    6.4.3.2. Scanning Movement B, Scanning Distance b shall be such that

    the full section of the weld is being examined with the primary

    angle from Section 5.3.

    6.4.3.3. Scanning Movement C, Progression Distance c shall be

    approximately one half the array width.

    6.4.4. To detect transverse defects, cover the full weld section with one of the

    following movements from Figure 6.4

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    16/40

    Page 16of 40

    6.4.4.1. Scanning pattern D, when weld reinforcement is ground flush.

    6.4.4.2. Scanning Pattern E, when weld reinforcement is not ground

    flush. Scanning Angle e = 15deg. maximum.

    6.4.5. For electroslag or electrogas welds, use Scanning Pattern E, with the

    search unit rotation angle e between 45deg. to 60deg.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    17/40

    Page 17of 40

    Table 6.4

    Scanning Patterns

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    18/40

    Page 18of 40

    7. Indication Evaluation and Reporting

    7.1. When a discontinuity indication appears on the screen, the maximum

    attainable reflection from the indication shall be adjusted to produceamplitude equivalent to the horizontal reference level (set in Sec. 5.3.3.).

    The instrument reading in dB shall be used as the Indication Level, a, for

    calculating the Indication Rating, d, as shown in the ultrasonic test report

    Fig 7.1.

    7.2. The Attenuation Factor, c, in the ultrasonic test report shall be attained by

    subtracting 1 inch of sound path (SA in the results window) from the

    distance to the indication and multiplying the remainder by 2, and rounded to

    the nearest dB value.

    7.3. The Indication Rating, d, in the ultrasonic test report represents the

    algebraic difference in decibels between the indication level with correction

    for attenuation as indicated by the following expression:

    Gain in dB: ac =d

    (Use: abc = d if the method noted under 5.3.3. is being used)

    NOTE: Traditionally the indication level a is the sum of all gain applied. For

    the purpose of this document the indication level is expressed as the amount of

    gain OVER the reference gain. Therefore the equation will be a c = d, and

    NOT a bc = d

    7.4. Lengths of indications with ratings more severe than Class D shall be

    determined by measuring the distance between the transducer centerline

    locations where the indication drops 50% in amplitude (6dB) below the

    rating for the applicable flaw classification. The length shall be recorded as

    discontinuity length on the ultrasonic test report.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    19/40

    Page 19of 40

    7.5. Each weld discontinuity shall be accepted or rejected on the basis of its

    indication rating, and its length in conformance with the applicable Table

    7.5A or 7.5B. Only those indications that are found to be unacceptable need

    be recorded on the test report. The exception being those welds identified in

    the contractual documents as being fracture critical. Fracture critical welds,

    which are within 6dB inclusive of the minimal unacceptable rating, shall be

    included on the test report.

    7.6. Each unacceptable discontinuity shall be indicated on the weld by a mark

    directly over the discontinuity for its entire length. The depth from the

    surface and indication rating shall be clearly noted on the adjacent basemetal.

    7.7. Retesting of welds that have been repaired shall be documented on a new

    line in the ultrasonic report, if the original is used, and shall be designated by

    an R prefix in the indication number. If additional report forms are used, the

    R designation shall prefix the report number, and / or the weld number, or

    any such manner that clearly identifies the weld as being previously

    repaired.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    20/40

    Page 20of 40

    Table 7.1

    Ultrasonic Report Form

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    21/40

    Page 21of 40

    Table 7.5A

    AWS D1.1 STRUCTURAL WELDING CODE STEEL

    UT ACCEPTANCE-REJECTION CRITERIA (Statically Loaded Nontubular Connections)

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    22/40

    Page 22of 40

    Table 7.5B

    AWS D1.1 STRUCTURAL WELDING CODE STEEL

    UT ACCEPTANCE-REJECTION CRITERIA (Cyclically Loaded Nontubular Connections)

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    23/40

    Page 23of 40

    Annex A

    dB accuracy procedure

    1.dB Accuracy Procedure.

    Note: In order to attain the required accuracy (+ 1%) in reading the indicationheight, the Display must be graduated vertically at 2% intervals at horizontalmidscreen. These graduations shall be placed on the Display between 60%and 100% of screen height. This may be done with use of a graduatedtransparent screen overlay. If this overlay is applied as a permanent part ofthe ultrasonic unit, care should be taken that the overlay does not obscurenormal testing displays.

    (1) Program the Phasor in accordance with Table 5.2. Thencouple the straight beam search unit to the block shown inDrawing D.1 and position "T".

    (2) Adjust the distance calibration so that the first 2 inch backreflection indication (hereafter called "the indication") is athorizontal mid-screen.

    (3) Adjust the calibrated gain or attenuation control so that the

    indication is exactly at or slightly above 40% screen height.(NOTE: the pulser voltage may be turned down if the gainvalue will not go low enough to establish a 40% value.)

    (4) Move the search unit toward position U, until the indicationis at exactly 40% screen height.

    (5) Increase the sound amplitude 6dB with the calibrated gainor attenuation control. The indication level theoreticallyshould be exactly 80% screen height.

    (6) Record the dB reading under "a" and actual % screenheight under "b" from step 5 on Line 1 of the "UltrasonicUnit Calibration Report", Form A.1 of this annex.

    (7) Move the search unit further toward position U, until theindication is exactly 40% screen height.

    (8) Repeat step 5.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    24/40

    Page 24of 40

    (9) Repeat step 6. Except, information should be applied tothe next consecutive line on the "Ultrasonic Unit CalibrationReport", Form A.1.

    (10) Repeat steps 7, 8 and 9 consecutively until the full range of

    the gain control or attenuator is reached (60 dB minimum).

    (11) Apply the information from Rows "a" and "b" to the belowreferenced equation (Section 2. Annex A) or thenomograph (Form A.2) to calculate the corrected dB.

    (12) Apply corrected dB from step 11 to Row "c".

    (13) Subtract Row "c" value from Row "a" value and apply thedifference in Row "d", dB error. These values may beeither positive or negative and should be noted with a plus

    sign "+" or a minus sign "-".

    (14) Information shall be tabulated on a form, includingminimum equivalent information as displayed on the"Ultrasonic Unit Calibration Report" (Form A.1) and the unitevaluated in accordance with instructions shown on thatform.

    (15) Form A.2 provides a relatively simple means of evaluatingdata from item (14). Instructions for this are given in (16)through (18) below.

    (16) Apply the dB information from Row "e" (Form A.1) verticallyand dB reading from Row "a" (Form A.1) horizontally as Xand Y coordinates for plotting a dB curve on Form A.2.

    (17) The longest horizontal length, as represented by the dBreading difference, which can be inscribed in a rectanglerepresenting 2 dB in height, denotes the dB range in whichthe equipment meets the Code requirements. Theminimum allowable range is 60 dB.

    (18) Equipment that does not meet this minimum requirementmay be used, provided correction factors are developedand used for discontinuity evaluation outside the instrumentacceptability linearity range, or the weld testing and thediscontinuity evaluation is kept within the acceptablevertical linearity range of the equipment.

    Note: The dB error figures (Row d) may be used ascorrection factor figures.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    25/40

    Page 25of 40

    2. Decibel Equation.The following equation shall be used to calculate decibels:

    dB1 = Row a

    dB2 = Row c

    %1 = Row b

    %2 = Is the average of Row b disregarding the first andlast three tabulations

    dB+%

    %x20=dB 1

    1

    2

    2

    log

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    26/40

    Page 26of 40

    Form A.1

    Ultrasonic Unit Calibration Report Form

    Ultrasonic Unit Model: Serial No.:

    Search Unit - Size: Type: Frequency: MHZ

    Calibration - Date: Interval: 2 Months Method:

    Block Serial No.: Data: As Found As Adjusted

    Supplemental Instructions:

    Start with the lowest dB level that you can obtain a 40 percent display height indication from. Add 6 dB's andrecord this dB reading in Row A and display height Row B as the starting points on the tabulation chart.

    After recording these values in Rows A and B, slide the transducer to obtain a new 40 percent display

    height. Without moving the transducer add 6 dBs and record the new dB reading and the new display height inRows A and B. Repeat this step as many times as the unit allows.

    Find the average % screen values from Row B by disregarding the first 3 and the last 3 tabulations. Usethis as %2in calculating the corrected reading.

    The following equation is used to calculate rowc:

    %1 is Row B

    %2 is the average of Row B disregarding thefirst and last three tabulations.

    dB1is Row A

    dB2is Row C

    The dB Error "Row D" is established by subtracting Row C" from Row A": (A - C = D).

    The Collective dB Error "Row E" is established by starting with the dB Error "Row D" nearest to 0.0,collectively add the dB Error Row D values horizontally, placing the sub-totals in Row E".

    Moving horizontally, left and right from the Average % line, find the span in which the largest and smallestCollective dB Error figures remain at or below 2 dB. Count the number of horizontal spaces of movement,subtract one, and multiply the remainder by six. This dB value is the acceptable range of the unit.

    In order to establish the acceptable range graphically, Form D-9 from AWS D1.1 may be used as follows:

    (1) Apply the collective dB Error Row E values vertically on the horizontal offset coinciding with the dB readingvalues from "Row A".

    (2) Establish a curve line passing through this series of points.

    (3) Apply a 2 dB high horizontal window over this curve positioned vertically so that the longest section iscompletely encompassed within the 2 dB Error height.

    (4) This window length represents the acceptable dB range of the unit.

    dB+%

    %x20=dB 1

    1

    2

    2

    log

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    27/40

    Page 27of 40

    Ultrasonic Unit Calibration Report Form A.1 (cont.)

    NUMBER 1 2 3 4 5 6 7 8 9 10 11 12 13

    Row A dB READING

    Row B DISPLAY HEIGHT

    Row C CORRECTED READING

    Row D dB ERROR

    Row E COLLECTIVE dB ERROR

    Accuracy Required: Minimum allowable Range is 60 dB. %2(Average) %

    Equipment is: Acceptable for Use Not Acceptable for Use Recalibration Due Date:

    Total qualified range dB to dB = dB Total error dB (From the Chart above)

    Total qualified range dB to dB = dB Total error dB (From AWS Form D-9)

    Calibrated by: Level: Location:

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    28/40

    Page 28of 40

    Form A.2

    dB Accuracy Evaluation Report Form

    Ultrasonic Unit Model: Serial No.:

    Search Unit - Size: Type: Frequency: MHZ.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    29/40

    Page 29of 40

    Annex B

    Horizontal Linearity, Internal Reflectors, and Resolution

    1. Horizontal Linearity Procedure.1.1 This qualification procedure is performed with the straight beam

    search unit, which produces longitudinal wave with a sound velocityof almost double that of shear wave. It is necessary to double theshear wave distance ranges to be used in applying this procedure.Example:

    The use of a 10 in. [250 mm] screen calibration in shear wave wouldrequire a 20 in. [500 mm] screen calibration for this qualification procedure.

    The following procedure shall be used for instrument certification:

    (1) Couple a straight beam search unit meeting the requirementsof paragraph 4.2 to the DS block (Drawing D.1 in Position U).

    (2) To attain a 20 in. [500 mm] screen calibration, adjust the firstand fifth back reflections to their proper locations with use ofthe distance calibration and zero delay adjustments.

    (3) Each indication shall be adjusted to reference level with thegain or attenuation control for horizontal location examination.

    (4) Each intermediate trace deflection location shall be correctwith +2% of the screen width. With the display width at 20 in.[500 mm] the +2% would be equivalent to +.400 in. [10 mm].

    2. Internal Reflections Procedure.(1) Calibrate the equipment in accordance with 5.3

    (2) Remove the search unit from the calibration block withoutchanging any other equipment adjustments.

    (3) Increase the calibrated gain or attenuation 20 dB more

    sensitive than reference level.

    (4) The screen area beyond 1/2 in. [12 mm] sound path andabove reference level height shall be free of any indication.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    30/40

    Page 30of 40

    3. Resolution Requirements.3.1 The combination of search unit and instrument shall resolve three holes

    in the RC resolution reference test block shown in Drawing D.1, or D.3.The search unit position(s) from Drawing D.1 are as follows:

    (1) The 45 degree shall be placed at position S.

    (2) The 60 degree shall be placed at position R.

    (3) The 70 degree shall be placed at position Q.

    3.2 The resolution shall be evaluated with the instrument controls set atnormal test settings and with indications from the holes brought tomidscreen height. Resolution shall be sufficient to distinguish at least

    the peaks of indications from the three holes. Use of the RC resolutionreference block for calibration shall be prohibited. Each combination ofinstrument search unit (shoe and transducer) shall be checked prior toits initial use. This equipment verification shall be done initially witheach search unit and UT unit combination. The verification need not bedone again provided documentation is maintained that records thefollowing items:

    (1) UT machines make, model and serial number.

    (2) Search units manufacturer, type, size, angle, and serial

    number.

    (3) Date of verification and technicians name.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    31/40

    Page 31of 40

    Form B.1

    Ultrasonic Unit Calibration Report Form B.1

    Ultrasonic Unit Model: Serial No:

    Search Unit - Size: Type: Frequency: MHZ

    DS Block Serial No: Calibration Interval: 40 hours use

    Horizontal linearity :

    INDICATION

    NUMBER

    LOCATION ACCEPTABLE

    RANGE

    ACTUAL

    LOCATION

    1 4.0" 3.6 to 4.4

    2 8.0" 7.6 to 8.4

    3 12.0" 11.6 to 12.4

    4 16.0" 15.6 to 16.4

    5 20.0" 19.6 to 20.4

    Internal Reflections:

    IIW Block SN: Reference Level: dB

    Remove search unit from block without changing any equipment settings

    Increase the calibrated gain 20 dB more sensitive than reference level

    Is the CRT screen area beyond 1/2" sound path and above reference level height free of

    any indications? Yes: No:___

    Equipment Status:

    As Found As Adjusted

    Not Acceptable for Use Acceptable for Use

    Calibrated By: Level: Date:_____________

    Reviewed By: Location:____________________

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    32/40

    Page 32of 40

    Form B.2

    Ultrasonic Unit Resolution Report Form B.2

    Ultrasonic Unit Model: Serial No:___________________________

    IIW Block Serial No: Resolution Block Serial No:_____________

    RESOLUTION REQUIREMENTS:

    Longitudinal Mode - Transducer and instrument shall resolve all 3 distances

    Shear Wave Mode - Transducer and instrument shall resolve the 3 test holes

    TRANSDUCERS:

    Serial No. Size Type Frequency Angle(s)

    Equipment is: Acceptable for Use Not Acceptable for Use_________

    Performed By: Level: __ Date:_________

    Reviewed By: Location:_____________________

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    33/40

    Page 33of 40

    Annex C

    Search unit check

    1. This annex is written as a supplemental document to support this procedure anddescribes the method(s) to perform the following:

    1) Transducer element, cable, and coupling check.2) Calculate the Z offset.

    2. Due to the proximity of each element in relation to the face of the wedge (e.g. the1stelement is closer to the bottom of the wedge than the last due to the slope of the

    angle.) each individual element will have a different delay in succession fromlowest to highest. To display this, program the Phasor according to the table in thisannex.2.1. Elements that do not produce a response are dead elements.

    2.1.1. The number of dead elements shall not exceed more than 3 total in thearray, and no more than 2 adjacent for the array to be consideredacceptable.

    2.1.2. Poor coupling will display a decrease, or lack of sound transmissionthrough all or part of the wedge. In this case remove the wedge, andre-couple the probe.

    2.2. Elements that do produce a response, but appear out of sequence, or

    otherwise not in line with the rest of the slope, are wired incorrectly and notfiring in the proper sequential order.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    34/40

    Page 34of 40

    PROBE MENU

    (Sub-menu) (tab) (value)

    Probe GeometryNumber of elements 16

    Pitch 0.040Wedge Dat Elem 1 Pos Low End

    Wedge GeometryOffset Z 0

    Angle 0

    Offset Probe Delay 0.000s

    PART MENU

    (Sub-menu) (tab) (value)

    Part Velocity L .1080 in/us

    SCAN MENU

    (Sub-menu) (tab) (value)

    Electronic

    Type Linear

    Focal Point Unfocused

    Wave Type Longitudinal

    Scan Pattern Angle Start 0.0 deg

    Aperture

    Number of Steps 16

    First Element 1

    Aperture Size 1

    Aperture Step 1

    DISPLAY MENU

    (Sub-menu) (tab) (value)

    View Angle Corrected On

    Results1

    Reading 1 P%A

    Reading 2 PSA

    Reading 3 Beam

    Reading 4 SA

    UT MENU

    (Sub-menu) (tab) (value)

    Base Display Delay 0.0

    Pulser

    Voltage 150V

    Width 220ns

    Receiver Frequency 2MHz

    Gate mode TOF mode Flank

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    35/40

    Page 35of 40

    Results for a 1D probe

    The pattern created is the result of firing each individual element in sequence,and the step like slope that is created is due to the different thickness inwedge material, or delay, beneath each element.

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    36/40

    Page 36of 40

    3. Determining the Z Offset.

    Material thickness and leg functions must be adjusted to your wedge in orderto display the step pattern on the screen.

    The A gate must frame the wedge echoes.

    The gain must be high enough so that the P%A reading is the result of thefirst element being barely > 100% FSH.

    Position the beam cursor to the last element so that it is reading in the SA.

    For formulas use:

    D1 = delay of first element (PSA)D2 = delay of last element (SA)N = number of elements

    To determine the Z offset:

    Z offset = (D1 + D2) / 2

    1) Z = (9.14mm + 13.60mm) /22) Z = (22.74mm) /23) Z = 11.37mmSpecified value 11.5mm Measured value 11.37mm

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    37/40

    Page 37of 40

    Annex D

    Drawings

    Drawing D.1

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    38/40

    Page 38of 40

    Drawing D.2

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    39/40

    Page 39of 40

    Drawing D.3

  • 8/10/2019 AWS Procedure - Rev 2.pdf

    40/40

    Page 40of 40

    Drawing D.4