ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES M. M. Sánchez-Lorente 1 ; P. García-Molina 2-3 ; C. Vergara-Hernández 2 ; E. Sanchis-Sánchez 2 ; J. Martos-Torres 2 ; A. Álvarez-Ordiales 4 ; & E. Balaguer-López 2-3 FP-50002 EWMA/GNEAUPP 2014 1. Conselleria de Sanitat, Valencia (SPA). 2. University of Valencia, Valencia (SPA). 3. Hospital Clínico Universitario, Valencia (SPA). 4. Hospital Universitari i Politècnic “La Fe”, Valencia (SPA). Presenting author: Mr. Pablo García-Molina
16
Embed
ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES M. M. Sánchez-Lorente 1 ; P. García-Molina 2-3 ; C. Vergara-Hernández 2 ; E. Sanchis-Sánchez.
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED)
FOR NEONATES
M. M. Sánchez-Lorente1; P. García-Molina2-3; C. Vergara-Hernández2;
E. Sanchis-Sánchez2; J. Martos-Torres2; A. Álvarez-Ordiales4;
4. Hospital Universitari i Politècnic “La Fe”, Valencia (SPA).
Presenting author: Mr. Pablo García-Molina
Background
FP-5
0002
EW
MA/
GN
EAU
PP 2
014
Existence of several neonatal support surfaces
Lack of evaluation
Interface pressure & pressure redistribution Efficacy in the prevention or treatment of PU
however…
regarding…
Aim
FP-5
0002
EW
MA/
GN
EAU
PP 2
014
• To assess the interface pressure and pressure redistribution of five devices of constant low pressure (static and mixed support surfaces).
152
43
Methods
FP-5
0002
EW
MA/
GN
EAU
PP 2
014
Study design• Laboratory study with repeated measures design.
Sample• Eight neonatal simulation dummies with different characteristics
regarding head circumference, weight and height.
Methods
FP-5
0002
EW
MA/
GN
EAU
PP 2
014
Support devices
Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D
Structure Viscoelastic“T” units
ViscoelasticBlock
ViscoelasticBlock
ViscoelasticBlock
Polyester filaments
Layers 1 2 2 2 1
Density (Kg/m3) 50 50/50 50/50 40 NA
Hardness (Kpa) 1.6 1.01/5 1.01/3.6 2.5 NA
Compression factor 2 2/2 2/2 - NA
Methods
FP-5
0002
EW
MA/
GN
EAU
PP 2
014
Variables• Mean pressure (mmHg).
• Maximum pressure (mmHg).
• Mean pressure in head region (mmHg).
• Area of contact (cm2).
Equipment• XSENSOR X3 PX100:40.40.002 device.
Procedure• See: E-poster no. 50006.
Methods
FP-5
0002
EW
MA/
GN
EAU
PP 2
014
Statistical analysis• Descriptive statistics.
• Due to a small sample size, a non-parametric test (Friedman test) was carried out to assess differences among groups. If a significant difference was found, post hoc comparisons among groups were performed using the Wilcoxon-Nemenyi-McDonald-Thompson test.
• The analysis was carried out by means of the R environment (The R Foundation for Statistical Computing).