J Neurosurg / Volume 120 / April 2014 J Neurosurg 120:901–907, 2014 901 ©AANS, 2014 A LTHOUGH monitoring of intracranial pressure (ICP) remains the foundation for the care of brain-injured patients in the neurointensive care unit, advanced cerebral monitoring has been used increasingly in recent years to monitor brain oxygenation and cerebral blood flow (CBF). The most recent update of the Guidelines for the Management of Severe Traumatic Brain Injury pro- vides recommended thresholds of brain oxygenation in patients with severe traumatic brain injury (TBI). 7 Most current monitoring technologies involve the placement of an invasive monitor in the brain parenchyma, with all the attendant risks. Development of a noninvasive method to monitor cerebral physiology in a continuous and clini- cally useful fashion has long been a goal in neurocritical care. The CerOx monitor (Ornim Medical Ltd.) uses near- infrared spectroscopy and ultrasound to assess cerebral oxygenation and blood flow in a noninvasive manner. The monitor employs a new technology, ultrasound-tagged near-infrared spectroscopy (UT-NIRS), leveraging the ef- fect of ultrasound waves on light to tag it. This technol- ogy harnesses the ability of near-infrared light to measure regional oxygen saturation in combination with ultra- sound that can achieve localization via the acousto-optic effect. 25,27 We sought to assess the feasibility of using the CerOx monitor to measure regional cerebral oxygenation in the neurointensive care unit and to compare CerOx- based measurements with hemispheric measures of cere- bral oxygenation in patients with severe TBI. Assessment of a noninvasive cerebral oxygenation monitor in patients with severe traumatic brain injury Clinical article GUY ROSENTHAL, M.D., 1 ALEX FURMANOV , R.N., B.S.N., 1 EYAL ITSHAYEK, M.D., 1 YIGAL SHOSHAN, M.D., 1 AND VINEETA SINGH, M.D. 2 1 Department of Neurosurgery, Hadassah-Hebrew University Medical Center, Jerusalem, Israel; and 2 Department of Neurology, University of California, San Francisco, California Object. Development of a noninvasive monitor to assess cerebral oxygenation has long been a goal in neu- rocritical care. The authors evaluated the feasibility and utility of a noninvasive cerebral oxygenation monitor, the CerOx 3110, which uses near-infrared spectroscopy and ultrasound to measure regional cerebral tissue oxygenation in patients with severe traumatic brain injury (TBI), and compared measurements obtained using this device to those obtained using invasive cerebral monitoring. Methods. Patients with severe TBI admitted to the intensive care unit at Hadassah-Hebrew University Hospital requiring intracranial pressure (ICP) monitoring and advanced neuromonitoring were included in this study. The authors assessed 18 patients with severe TBI using the CerOx monitor and invasive advanced cerebral monitors. Results. The mean age of the patients was 45.3 ± 23.7 years and the median Glasgow Coma Scale score on ad- mission was 5 (interquartile range 3–7). Eight patients underwent unilateral decompressive hemicraniectomy and 1 patient underwent craniotomy. Sixteen patients underwent insertion of a jugular bulb venous catheter, and 18 patients underwent insertion of a Licox brain tissue oxygen monitor. The authors found a strong correlation (r = 0.60, p < 0.001) between the jugular bulb venous saturation from the venous blood gas and the CerOx measure of regional cerebral tissue saturation on the side ipsilateral to the catheter. A multivariate analysis revealed that among the physi- ological parameters of mean arterial blood pressure, ICP, brain tissue oxygen tension, and CerOx measurements on the ipsilateral and contralateral sides, only ipsilateral CerOx measurements were significantly correlated to jugular bulb venous saturation (p < 0.001). Conclusions. Measuring regional cerebral tissue oxygenation with the CerOx monitor in a noninvasive manner is feasible in patients with severe TBI in the neurointensive care unit. The correlation between the CerOx measure- ments and the jugular bulb venous measurements of oxygen saturation indicate that the CerOx may be able to provide an estimation of cerebral oxygenation status in a noninvasive manner. (http://thejns.org/doi/abs/10.3171/2013.12.JNS131089) KEY WORDS brain oxygenation non-invasive monitor jugular bulb venous monitor severe traumatic brain injury 901 Abbreviations used in this paper: CBF = cerebral blood flow; CPP = cerebral perfusion pressure; FiO 2 = fraction of inspired oxygen; GCS = Glasgow Coma Scale; ICP = intracranial pressure; MABP = mean arterial blood pressure; PbtO 2 = partial pressure of brain tissue oxygen; rCsat = regional cerebral tissue oxygen saturation; SjvO 2 = jugular bulb venous oxygen saturation; TBI = traumatic brain injury; UT-NIRS = ultrasound-tagged near-infrared spectroscopy.
7

Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

Aug 17, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

J Neurosurg / Volume 120 / April 2014

J Neurosurg 120:901–907, 2014

901

©AANS, 2014

Although monitoring of intracranial pressure (ICP) remains the foundation for the care of brain-injured patients in the neurointensive care unit, advanced

cerebral monitoring has been used increasingly in recent years to monitor brain oxygenation and cerebral blood flow (CBF). The most recent update of the Guidelines for the Management of Severe Traumatic Brain Injury pro-vides recommended thresholds of brain oxygenation in patients with severe traumatic brain injury (TBI).7 Most current monitoring technologies involve the placement of

an invasive monitor in the brain parenchyma, with all the attendant risks. Development of a noninvasive method to monitor cerebral physiology in a continuous and clini-cally useful fashion has long been a goal in neurocritical care. The CerOx monitor (Ornim Medical Ltd.) uses near-infrared spectroscopy and ultrasound to assess cerebral oxygenation and blood flow in a noninvasive manner. The monitor employs a new technology, ultrasound-tagged near-infrared spectroscopy (UT-NIRS), leveraging the ef-fect of ultrasound waves on light to tag it. This technol-ogy harnesses the ability of near-infrared light to mea sure regional oxygen saturation in combination with ultra-sound that can achieve localization via the acousto-optic effect.25,27 We sought to assess the feasibility of using the CerOx monitor to measure regional cerebral oxygenation in the neurointensive care unit and to compare CerOx-based measurements with hemispheric measures of cere-bral oxygenation in patients with severe TBI.

Assessment of a noninvasive cerebral oxygenation monitor in patients with severe traumatic brain injury

Clinical articleGuy Rosenthal, M.D.,1 alex FuRManov, R.n., B.s.n.,1 eyal Itshayek, M.D.,1 yIGal shoshan, M.D.,1 anD vIneeta sInGh, M.D.2

1Department of Neurosurgery, Hadassah-Hebrew University Medical Center, Jerusalem, Israel; and 2Department of Neurology, University of California, San Francisco, California

Object. Development of a noninvasive monitor to assess cerebral oxygenation has long been a goal in neu-rocritical care. The authors evaluated the feasibility and utility of a noninvasive cerebral oxygenation monitor, the CerOx 3110, which uses near-infrared spectroscopy and ultrasound to measure regional cerebral tissue oxygenation in patients with severe traumatic brain injury (TBI), and compared measurements obtained using this device to those obtained using invasive cerebral monitoring.

Methods. Patients with severe TBI admitted to the intensive care unit at Hadassah-Hebrew University Hospital requiring intracranial pressure (ICP) monitoring and advanced neuromonitoring were included in this study. The authors assessed 18 patients with severe TBI using the CerOx monitor and invasive advanced cerebral monitors.

Results. The mean age of the patients was 45.3 ± 23.7 years and the median Glasgow Coma Scale score on ad-mission was 5 (interquartile range 3–7). Eight patients underwent unilateral decompressive hemicraniectomy and 1 patient underwent craniotomy. Sixteen patients underwent insertion of a jugular bulb venous catheter, and 18 patients underwent insertion of a Licox brain tissue oxygen monitor. The authors found a strong correlation (r = 0.60, p < 0.001) between the jugular bulb venous saturation from the venous blood gas and the CerOx measure of regional cerebral tissue saturation on the side ipsilateral to the catheter. A multivariate analysis revealed that among the physi-ological parameters of mean arterial blood pressure, ICP, brain tissue oxygen tension, and CerOx measurements on the ipsilateral and contralateral sides, only ipsilateral CerOx measurements were significantly correlated to jugular bulb venous saturation (p < 0.001).

Conclusions. Measuring regional cerebral tissue oxygenation with the CerOx monitor in a noninvasive manner is feasible in patients with severe TBI in the neurointensive care unit. The correlation between the CerOx measure-ments and the jugular bulb venous measurements of oxygen saturation indicate that the CerOx may be able to provide an estimation of cerebral oxygenation status in a noninvasive manner.(http://thejns.org/doi/abs/10.3171/2013.12.JNS131089)

key WoRDs      •      brain oxygenation      •      non-invasive monitor      •      jugular bulb venous monitor      •      severe traumatic brain injury

901

Abbreviations used in this paper: CBF = cerebral blood flow; CPP = cerebral perfusion pressure; FiO2 = fraction of inspired oxygen; GCS = Glasgow Coma Scale; ICP = intracranial pressure; MABP = mean arterial blood pressure; PbtO2 = partial pressure of brain tissue oxygen; rCsat = regional cerebral tissue oxygen saturation; SjvO2 = jugular bulb venous oxygen saturation; TBI = traumatic brain injury; UT-NIRS = ultrasound-tagged near-infrared spectroscopy.

Page 2: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

G. Rosenthal et al.

902 J Neurosurg / Volume 120 / April 2014

MethodsResearch Subjects

Institution review board approval for the study was received from the Hadassah Medical Organization.

Eligible subjects were TBI patients, admitted to the in-tensive care unit at Hadassah-Hebrew University Hospital and requiring ICP monitoring and advanced neuromoni-toring. Inclusion criteria included a Glasgow Coma Scale (GCS) score of ≤ 8 on admission or multisystem injury requiring ICP monitoring. Intracranial pressure was moni-tored with an external ventricular drain or intraparenchy-mal sensor. Patients were treated in accordance with the Brain Trauma Foundation’s Guidelines for the Manage-ment of Severe Traumatic Brain Injury3–8 and with the fol-lowing goals: PaO2 > 100 mm Hg; PaCO2 > 35 mm Hg and < 45 mm Hg; cerebral perfusion pressure (CPP) ≥ 50 mm Hg and ≤ 70 mm Hg; central venous pressure 5–7 mm Hg; ICP < 20 mm Hg; partial pressure of brain tissue oxygen (PbtO2) > 15 mm; hemoglobin > 10 g/dl; and Na > 135 mEg/L and < 145 mEq/L. Patients who had surgical mass lesions underwent surgical intervention in accordance with the Guidelines for the Surgical Management of Severe Traumatic Brain Injury.9–13

Jugular Bulb Venous Saturation MonitoringJugular bulb venous saturation monitors were insert-

ed under ultrasound guidance. Plain skull radiographs or axial CT scans confirmed the correct placement of the catheter tip.

Brain Tissue Oxygen MonitoringPatients with severe TBI underwent insertion of a

Licox intraparenchymal brain tissue oxygen monitor (In-tegra LifeSciences Corp.) as part of standard neurointen-sive care at our institution. Licox probes were inserted over Kocher’s point into the frontal lobe of the least-in-jured hemisphere. Probes were allowed to stabilize for at least 12 hours following insertion before any oxygen challenge was performed.

Noninvasive Cerebral Oxygen Saturation MonitoringRegional cerebral tissue oxygen saturation (rCsat) was

measured using the CerOx 3110 monitor, which utilizes UT-NIRS technology; UT-NIRS is a hybrid technology based on locally modulating coherent light with a local-ized low-power ultrasound wave via the acousto-optic ef-fect.25,27 Like other NIRS-based devices, UT-NIRS illumi-nates tissue with near-infrared light. The UT-NIRS device then adds rapid, brief pulses of ultrasound into the tissue over the region of interest. The ultrasound waves induce a local, artificial modulation in the detected light intensity in a manner similar to the Doppler effect. The modulated light signal is then analyzed on the basis of the changes in light intensity induced by the ultrasound wave. In this manner, only light from a specific volume of brain tissue is selected for analysis, allowing measurement from a spe-cific predetermined region of interest. The tissue volume measured with the device is approximately 1 cm3. The UT-NIRS device uses 3 light wavelengths of light (780–830

nm) that are modulated by ultrasound as described above. A spectral analysis is then performed to calculate the satu-ration of hemoglobin within the region of interest.

The measurement system consisted of 2 separate probes placed on the skin in the frontal region (after shav-ing the head, when needed) and secured with a special ad-hesive pad (Fig. 1). Each probe illuminates the tissue with laser light at 3 wavelengths and “collects” the scattered light back to the detector. The probe also incorporates a small ultrasound transducer that provides low-power waves for inducing the UT-NIRS signal. Probes were placed bilaterally over the frontal region of the patient’s head. Patients in whom a large subgaleal hematoma in the frontal region(s) obscured the UT-NIRS signal were excluded from the study.

Physiological ChallengesAt our institution, challenge testing is performed as

part of routine clinical care of severe TBI patients who undergo advanced cerebral monitoring to assess cerebral vascular autoregulation (mean arterial blood pressure [MABP] challenge), the response to hyperoxia to test proper functioning of brain tissue oxygen probes (oxygen challenge), and cerebrovascular CO2 reactivity (hyper-ventilation challenge).

MABP ChallengeTo assess cerebral vascular autoregulation, we in-

creased MABP by initiating a noradrenaline drip to achieve a 10– to 15–mm Hg rise in MABP. In patients already treated with a noradrenaline drip, the rate of infusion was increased to achieve the desired rise of 10–15 mm Hg in MABP. This portion of the MABP challenge occurs over a 20- to 30-minute time period. Physiological variables, including systemic blood pressure, MABP, percutaneous systemic oxygen saturation, ICP, and PbtO2, were recorded

Fig. 1. Cerebral monitors, including an external ventricular drain and Licox brain tissue oxygen monitor, shown in a patient with severe TBI. The arrow indicates the noninvasive CerOx monitor attached to the shaved scalp.

Page 3: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

J Neurosurg / Volume 120 / April 2014

Use of the CerOx monitor in severe TBI

903

during the MABP challenge and during all subsequent challenges. No CSF drainage occurred during any portion of the challenges, so as not to confound interpretation of the challenges. All patients are maintained on sedation dur-ing all challenges (either with propofol and remifentanil or midazolam and morphine). An arterial blood gas test and a venous blood gas test from the jugular venous catheter are performed when the goal MABP is reached. The purpose of increasing MABP is to observe the response of CBF, ICP, and PbtO2 to an increase in pressure. In those patients in whom these parameters stay relatively constant during a rise in MABP, cerebral autoregulation is presumed to be intact. In those patients in whom a definite rise in seen in these parameters along with the increase in MABP, cere-bral autoregulation is presumed to be impaired.31 The goals of CPP are then set as recommended in the Guidelines for the Management of Severe Traumatic Brain Injury, with a lower CPP goal (50–60 mm Hg) for those patients with impaired autoregulation and a higher CPP goal (60–70 mm Hg) for those patients with preserved autoregulation.6 Pa-tients who are hemodynamically unstable or have baseline high blood pressure are deemed unsuitable for a MABP challenge. Also, patients with extremely labile ICP may be deemed unstable for MABP challenge as vasopressors may initiate a spike in ICP that may be difficult to control if autoregulation is impaired. Clinical judgment is crucial to determine which patients may safely undergo a MABP challenge to avoid harm.

At the conclusion of the MABP challenge, the phen-ylephrine drip was discontinued (in patients who had not been receiving pressor support to maintain CPP) or re-turned to the baseline infusion rate (in patients who had been receiving phenylephrine to maintain CPP), and the MABP was allowed to return to the prechallenge level and followed by a 30-minute observation period before initiation of the oxygen challenge.

Oxygen ChallengeDuring the oxygen challenge, we increased the frac-

tion of inspired oxygen (FiO2) from baseline to 1.0 for approximately 20 minutes. Arterial and venous blood gas samples were drawn from the radial artery and jugular bulb venous catheter prior to initiating the oxygen chal-lenge (baseline) and at the end of the oxygen challenge. The FiO2 was maintained at 1.0 during subsequent chal-lenges (hyperventilation challenge) to reduce the risk of tissue hypoxia or desaturation during the hyperventila-tion challenge. Oxygen challenge was performed rou-tinely with the purpose of assessing proper functioning of the Licox brain tissue oxygen monitor as previously described.28 In addition to testing probe functionality, an oxygen challenge helps to determine the maximal PbtO2 that can be achieved at an FiO2 of 1.0, which is often an indicator of lung function.24,28

Hyperventilation ChallengeImmediately after completing the oxygen challenge,

a hyperventilation challenge was initiated to assess ce-rebral CO2 vasoreactivity, as previously described.29 The purpose of the hyperventilation challenge is to assess the

degree to which ICP can be reduced with moderate hy-perventilation without impairing brain tissue oxygenation so that an optimal point can be determined as a therapeu-tic goal in the event of a rise in ICP.31 Patients who have pulmonary complications with difficulty in oxygenation or ventilation do not undergo hyperventilation challenge, as the potential to worsen pulmonary status exists.

The respiratory rate was increased by 25% and then by increments of 2 breaths per minute with the goal of decreasing PaCO2 by 10 mm Hg. This portion of the chal-lenge also occurs over a 15- to 20-minute period. A hy-perventilation challenge was considered successful if the PaCO2 decreased by 5 mm Hg or more on the arterial blood gas test. As preset safety precautions, hyperventila-tion was terminated when PbtO2 approached 15 mm Hg or if it decreased more than 50% of the value attained during an oxygen challenge to ensure patient safety. Arte-rial and jugular bulb venous blood gases were sampled at the conclusion of the challenge. The total time course for a 3-step challenge (MABP, oxygen, and hyperventilation) was approximately 120 minutes. For each individual pa-tient a daily assessment was made to determine whether it was safe to undergo a 3-step challenge.

Statistical AnalysisMeans and standard deviations were calculated for

continuous physiological data. The values for jugular bulb venous oxygen saturation (SjvO2) and arteriovenous oxygen difference were taken directly from the venous blood draw and are thus point measurements. Values of continuously measured physiological parameters including MABP, ICP, CPP, PbtO2, and rCsat were averaged over a 1-minute pe-riod corresponding to the draw time of the venous blood gases. A 2-way ANOVA model corrected for repeated measures within the same patients was used to analyze dif-ferences at each of the challenge conditions. Linear corre-lation coefficients between CerOx monitor rCsat measure-ments and jugular bulb venous saturation were evaluated using a Bland-Altman correction for repeated measure-ments in the same patients.1,2 This correlation was tested in a multivariate analysis adjusted to other physiological parameters. The relationship between CerOx-measured rCsat and physiological parameters was evaluated using a general linear model, corrected for multiple measurements in the same patient. The level of statistical significance level was defined as p = 0.05. All analyses were carried out using SPSS version 20.01 software (IBM).

ResultsWe studied 18 patients (13 male, 5 female) with se-

vere TBI using the CerOx 3110 monitor. Patient charac-teristics are detailed in Table 1. The mean age was 45.3 ± 23.7 years (range 20–80 years), and the median GCS on admission was 5 (interquartile range 3–7). The side of surgical intervention (if performed) and the location of invasive monitors are also detailed in Table 1. Eight patients underwent unilateral decompressive hemicra-niectomy and 1 patient underwent craniotomy. Sixteen patients underwent insertion of a jugular bulb venous

Page 4: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

G. Rosenthal et al.

904 J Neurosurg / Volume 120 / April 2014

catheter, which in 5 patients was ipsilateral to the side of surgical intervention and in 3 patients contralateral to the side of surgical intervention. In those patients who did not undergo surgical intervention the jugular bulb venous catheter was placed on the right side in 6 patients and on the left side in 2 patients. Eighteen patients underwent insertion of a Licox brain tissue oxygen monitor, which was generally placed in the least-injured hemisphere per our clinical protocol. Mean values for physiological pa-rameters at the time of the challenges are presented in Table 2. Importantly, no complications associated with use of the CerOx noninvasive monitor were observed in any patients. We did not observe any skin irritation in-duced by the UT-NIRS probe placed on the scalp. We did not observe obstruction of the UT-NIRS signal by an underlying intracranial hematoma, likely because lesions causing significant mass effect were evacuated, according to the Guidelines for the Surgical Management of Severe Traumatic Brain Injury, and monitors were placed only after surgical intervention in those patients who under-went surgery.

We studied the relationship between rCsat and jug-ular bulb hemispheric measures of venous saturation. Good quality jugular bulb venous data were obtained in 14 (88%) of 16 patients. Good-quality UT-NIRS data were

obtained in 14 (78%) of 18 patients. Eleven patients had both good-quality jugular bulb venous data and CerOx data that could be compared. We compared CerOx rCsat measurements to the jugular bulb venous saturation on both the ipsilateral and contralateral sides to the jugular bulb catheter. We found a strong correlation (r = 0.60, p < 0.001) between the jugular bulb venous saturation from the venous blood gas and the UT-NIRS measure of rCsat on the side ipsilateral to the catheter (Fig. 2) that remained significant when adjusted for multiple measurements in the same patient (p < 0.001). In contrast, the correlation between the jugular bulb venous saturation and the CerOx rCsat on the contralateral side was substantially weaker (r = 0.23, p = 0.004) and was not significant when adjusted for multiple measurements in the same patient (p = 0.31). We then assessed the relationship between jugular bulb venous saturation with CerOx rCsat on the ipsilateral and contralateral sides and physiological parameters that may influence cerebral perfusion including MABP, ICP, and brain tissue oxygen tension in a multivariate analysis (Table 3). Only CerOx rCsat on the side ipsilateral to the catheter was related to jugular bulb venous saturation in a statistically significant manner. The other parameters in-cluding MABP, ICP, and brain tissue oxygen tension did not reach statistical significance.

TABLE 1: Patient characteristics*

Case No.

Age (yrs), Sex

Mechanism of Injury

Initial GCS Score

Type of Intracranial Injury

ICP Monitor (type & side)

Licox Monitor Side Side of Op

Side of SjvO2 Monitor

GOS-E Score at 6 mos

1 22, M fall 3 contusion, SAH EVD, rt rt none rt 62 20, M bike accident 7 SAH, IVH, DAI EVD, rt rt none rt 43 58, M fall 9 SDH, contusion, SAH EVD, lt lt rt lt 24 21, F MVA 5 contusion, SAH, IVH,

DAIEVD, rt rt none lt 4

5 42, M MVA 3 contusion, IVH, SAH, SDH (tentorial), DAI

IPM, rt rt none rt 2

6 20, M MCA 6 contusion, SAH, IVH, DAI

EVD, rt rt none rt 1

7 79, F fall 6 SDH EVD, rt rt lt rt 28 59, M fall 3 SDH EVD, rt rt lt lt 19 23, M MVA 3 SDH, SAH EVD, rt rt lt lt 1

10 46, M MVA 3 contusion, SAH, SDH (tentorial)

IPM, rt rt none rt 3

11 72, M fall 4 SDH, contusion, SAH EVD, lt lt rt rt 112 47, F struck by falling

object 5 SDH, contusion, SAH EVD, lt lt rt none 1

13 80, F fall 3 SDH EVD, rt rt lt lt 114 78, M fall (attempted

suicide)4 contusion, SAH EVD, rt rt none none 1

15 23, M fall 7 contusion, SAH EVD, rt rt none rt 616 29, M MVA 9 contusion, SAH, SDH EVD, rt lt none lt 417 22, M MVA 7 contusion, SDH, SAH EVD, rt lt rt rt 318 74, F fall 7 SDH, contusion, SAH IPM, lt rt lt rt 1

* CerOx probes were placed over the frontal region bilaterally in all patients. DAI = diffuse axonal injury; EVD = external ventricular drain; GOS-E = Glasgow Outcome Scale–Extended; IPM = intraparenchymal monitor; IVH = intraventricular hemorrhage; MCA = motorcycle accident; MVA = motor vehicle accident; SAH = subarachnoid hemorrhage; SDH = subdural hematoma.

Page 5: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

J Neurosurg / Volume 120 / April 2014

Use of the CerOx monitor in severe TBI

905

We also studied the relationship between UT-NIRS re-gional measures of cerebral oxygenation and physiological parameters of MABP, ICP, and brain tissue oxygen tension. Since patients were stable at the time of challenges with-out extremes of hypotension or intracranial hypertension, it is not surprising that MABP and ICP were not related to cerebral oxygenation as measured by CerOx (Table 4). Interestingly, brain tissue oxygenation measured by Licox was also not significantly related to CerOx measures of oxygenation, indicating that these 2 measures of brain oxy-genation are not strongly related to each other.

DiscussionOur findings indicate that noninvasively monitoring

brain oxygenation with the CerOx monitor in the neuro-intensive care unit in severe TBI patients is feasible. The correlations between the UT-NIRS measurements and the jugular venous bulb measurements of oxygen saturation indicate that the CerOx monitor may be able to provide a noninvasive estimation of cerebral oxygenation status in brain-injured patients. Interestingly, we found that only the ipsilateral measurements of rCsat were significantly

correlated to jugular bulb venous saturation. This find-ing is compatible with the observations of Stocchetti et al. who found a difference in jugular bulb venous oxygen saturation and oxygen content of greater than 15% be-tween the hemispheres in 15 of 32 patients with severe TBI, suggesting that substantial differences in cerebral oxygenation exist between the hemispheres in patients with severe TBI and that unilateral hemispheric measures may not adequately assess overall cerebral oxygenation status.33 Our results suggest that the CerOx monitor may be able to provide a noninvasive measure of regional cere-bral oxygenation that reflects differences in oxygenation between the hemispheres. Whether the CerOx rCsat mea-surements will be useful in helping guide clinical care requires further study.

Not surprisingly, the UT-NIRS measurements of re-gional cerebral tissue oxygenation did not correlate with systemic and cerebral physiological parameters including MABP and ICP. This is likely due to the fact that at the time that data were acquired patients were stable in the intensive care unit without extremes of ICP, MABP, CPP, or systemic oxygenation. Even within the context of phys-iological challenges to assess cerebral autoregulation and

TABLE 2: Mean values for physiological parameters at baseline and during challenges*

Parameter Baseline MABP Challenge Oxygen Challenge Hyperventilation Challenge

MABP (mm Hg) 80 ± 15 88 ± 17† 80 ± 14 76 ± 15‡ ICP (mm Hg) 10 ± 7 10 ± 5 12 ± 6‡ 9 ± 4†CPP (mm Hg) 70 ± 17 78 ± 18† 68 ± 14 66 ± 15PbtO2 (mm Hg) 23.9 ± 10.3 25.7 ± 8.3 31.0 ± 14† 32.6 ± 12.5SjvO2 (%) 75 ± 9 78 ± 8 80 ± 8† 69 ± 10†AVDO2 (ml O2/100 g) 2.9 ± 1.2 2.7 ± 1.2 2.4 ± 1.2 4.5 ± 1.4†CerOx rCsat (%)§ ipsilateral 56 ± 12 56 ± 11 55 ± 11 51 ± 12 contralateral 55 ± 11 55 ± 10 54 ± 12 56 ± 10

* Values are presented as the mean ± SD. AVDO2 = arteriovenous oxygen difference.† p < 0.001‡ p < 0.05. § Measured using CerOx; ipsilateral and contralateral refer to the side of the jugular bulb catheter.

Fig. 2. Graph demonstrating the relationship between jugular bulb venous saturation taken from venous blood gases and the ipsilateral CerOx rCsat (r = 0.60, p < 0.001).

Page 6: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

G. Rosenthal et al.

906 J Neurosurg / Volume 120 / April 2014

cerebral CO2 vasoreactivity, the parameters of MABP, ICP, CBF, and cerebral oxygenation vary within a narrow range. This very narrow range of values likely precludes a strong correlation to MABP or ICP to cerebral oxygen-ation, which may be seen at the more extreme ranges of high and low values. Experimental studies have clearly demonstrated that the relationship between both ICP and CPP and brain oxygenation is weak unless extremes of either ICP or CPP are reached.15,18,30,32,34 The lack of a significant relationship between brain tissue oxygen ten-sion and rCsat is also anticipated. Gupta and colleagues directly compared brain tissue oxygen tension with jugu-lar bulb venous saturation during hyperventilation chal-lenges in patients with severe TBI and found that changes in brain oxygen tension did not correlate with changes in SjvO2 when patient data were pooled.17 Although both PbtO2 and rCsat measure cerebral tissue oxygenation, brain tissue oxygen tension is primarily influenced by ar-terial oxygen tension and CBF.20–23,26,29 In contrast, rCsat is a measure of regional tissue saturation that seems to be influenced primarily by the degree of oxygen extraction determined by the balance between oxygen delivery and utilization. In this regard, the rCsat measure is similar to the SjvO2 measurements that are most strongly influ-enced by the interaction between blood oxygen content and CBF, on the one hand, and cerebral oxygen utiliza-tion, on the other hand.14,16 The UT-NIRS measurements of regional tissue saturation have the potential benefit of providing a noninvasive correlate of hemispheric oxygen-ation status that may reflect the balance between oxygen delivery and utilization.

LimitationsThe current study has several limitations. First, it

is a small study that intended to assess the feasibility of using the CerOx monitor in patients with severe TBI in the neurointensive care unit. The small size of this study

precludes any assessment of the relationship of rCsat measurements to outcome. Second, comparing a hemi-spheric measurement of cerebral oxygenation such as jugular bulb venous saturation to a regional measurement of brain tissue oxygenation is problematic. Unfortunately, current technology does not provide a definitive measure-ment of regional cerebral tissue oxygen saturation that can be compared with the CerOx noninvasive measure. Last, while the CerOx monitor allows for continuous measure of rCsat, we compared the measurements ob-tained to discrete values of hemispheric venous saturation from the jugular bulb catheter. The continuous measure of jugular bulb venous saturation entails technical dif-ficulties including a relatively small amount of time for collecting good-quality data and the need for frequent recalibrations.19 While our approach substantially limits the number of comparisons, we felt that the venous gases drawn at defined physiological conditions allowed a more definitive assessment of oxygenation status.

ConclusionsMeasuring regional cerebral tissue oxygenation with

the CerOx monitor in a noninvasive manner is feasible in patients with severe TBI in the neurointensive care unit. The CerOx rCsat measures correlate to hemispheric measures of cerebral oxygenation in the ipsilateral jugu-lar bulb but not to those on the contralateral side, suggest-ing that it may be able to noninvasively identify regional differences in cerebral oxygenation. Further studies are required to determine the clinical utility of the CerOx measures of regional cerebral tissue oxygenation.

Disclosure

This work was supported by a grant from Ornim Medical Ltd. None of the authors has any financial interest in the company

or the device.Author contributions to the study and manuscript preparation

include the following. Conception and design: Rosenthal, Singh. Acquisition of data: Rosenthal, Furmanov. Analysis and interpre-tation of data: Rosenthal, Furmanov, Singh. Drafting the article: Rosenthal. Critically revising the article: all authors. Reviewed sub-mitted version of manuscript: all authors. Approved the final version of the manuscript on behalf of all authors: Rosenthal. Administra-tive/technical/material support: Furmanov, Shoshan. Study supervi-sion: Rosenthal.

References

1. Bland JM, Altman DG: Calculating correlation coefficients with repeated observations: Part 1—Correlation within sub-jects. BMJ 310:446, 1995

2. Bland JM, Altman DG: Calculating correlation coefficients with repeated observations: Part 2—Correlation between sub-jects. BMJ 310:633, 1995 (Erratum in BMJ 312:572, 1996)

3. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al: Guidelines for the management of severe traumatic brain injury. I. Blood pressure and oxygen-ation. J Neurotrauma 24 (Suppl 1):S7–S13, 2007 (Erratum in J Neurotrauma 25:276–278, 2008)

4. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al: Guidelines for the management of severe traumatic brain injury. II. Hyperosmolar therapy. J Neurotrauma 24 (Suppl 1):S14–S20, 2007 (Erratum in J Neurotrauma 25:276–278, 2008)

TABLE 3: Multivariate analysis of the relation between physiological parameters and jugular bulb venous saturation values

Parameter Coefficient p Value95% CI

Lower Bound Upper Bound

Cerox rCsat ipsilateral 0.26 <0.001 0.14 0.37 contralat −0.07 0.264 −0.19 0.05MABP 0.10 0.068 −0.01 0.20ICP −0.20 0.302 −0.57 0.18PbtO2 0.08 0.208 −0.04 0.20

TABLE 4: Correlations between physiological parameters and CerOx-measured rCsat

rCsatCorrelation Coefficients

MABP ICP PbtO2

least-injured hemisphere 0.063 0.055 0.014most-injured hemisphere 0.136 0.062 0.041

Page 7: Assessment of a noninvasive cerebral oxygenation monitor ...pdfs.semanticscholar.org/f17d/d418d59ecaa2fd8b56a88d463aa72a4d7d6f.pdfremains the foundation for the care of brain-injured

J Neurosurg / Volume 120 / April 2014

Use of the CerOx monitor in severe TBI

907

5. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al: Guidelines for the management of severe traumatic brain injury. VIII. Intracranial pressure thresholds. J Neurotrauma 24 (Suppl 1):S55–S58, 2007 (Er-ratum in J Neurotrauma 25:276–278, 2008)

6. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al: Guidelines for the management of severe traumatic brain injury. IX. Cerebral perfusion thresh-olds. J Neurotrauma 24 (Suppl 1):S59–S64, 2007 (Erratum in J Neurotrauma 25:276–278, 2008)

7. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al: Guidelines for the management of severe traumatic brain injury. X. Brain oxygen monitoring and thresholds. J Neurotrauma 24 (Suppl 1):S65–S70, 2007 (Er-ratum in J Neurotrauma 25:276–278, 2008)

8. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al: Guidelines for the manage-ment of severe traumatic brain injury. XIV. Hyperventilation. J Neurotrauma 24 (Suppl 1):S87–S90, 2007 (Erratum in J Neurotrauma 25:276–278, 2008)

9. Bullock MR, Chesnut R, Ghajar J, Gordon D, Hartl R, Newell DW, et al: Surgical management of acute epidural hematomas. Neurosurgery 58:S2-7–S2-15, Si–iv, 2006

10. Bullock MR, Chesnut R, Ghajar J, Gordon D, Hartl R, Newell DW, et al: Surgical management of acute subdural hemato-mas. Neurosurgery 58:S2-16–S2-24, Si–iv, 2006

11. Bullock MR, Chesnut R, Ghajar J, Gordon D, Hartl R, Newell DW, et al: Surgical management of depressed cranial frac-tures. Neurosurgery 58 (3 Suppl):S2-56–S2-60, Si–iv, 2006

12. Bullock MR, Chesnut R, Ghajar J, Gordon D, Hartl R, Newell DW, et al: Surgical management of posterior fossa mass le-sions. Neurosurgery 58:S2-47–S2-55, Si–iv, 2006

13. Bullock MR, Chesnut R, Ghajar J, Gordon D, Hartl R, Newell DW, et al: Surgical management of traumatic parenchymal le-sions. Neurosurgery 58:S2-25–S2-46, Si–iv, 2006

14. Feldman Z, Robertson CS: Monitoring of cerebral hemody-namics with jugular bulb catheters. Crit Care Clin 13:51–77, 1997

15. Friess SH, Ralston J, Eucker SA, Helfaer MA, Smith C, Mar-gulies SS: Neurocritical care monitoring correlates with neu-ropathology in a swine model of pediatric traumatic brain injury. Neurosurgery 69:1139–1147, 2011

16. Gopinath SP, Valadka AB, Uzura M, Robertson CS: Compari-son of jugular venous oxygen saturation and brain tissue Po2 as monitors of cerebral ischemia after head injury. Crit Care Med 27:2337–2345, 1999

17. Gupta AK, Hutchinson PJ, Al-Rawi P, Gupta S, Swart M, Kirkpatrick PJ, et al: Measuring brain tissue oxygenation compared with jugular venous oxygen saturation for monitor-ing cerebral oxygenation after traumatic brain injury. Anesth Analg 88:549–553, 1999

18. Hemphill JC III, Knudson MM, Derugin N, Morabito D, Man-ley GT: Carbon dioxide reactivity and pressure autoregulation of brain tissue oxygen. Neurosurgery 48:377–384, 2001

19. Kiening KL, Unterberg AW, Bardt TF, Schneider GH, Lanksch WR: Monitoring of cerebral oxygenation in patients with se-vere head injuries: brain tissue PO2 versus jugular vein oxy-gen saturation. J Neurosurg 85:751–757, 1996

20. Longhi L, Valeriani V, Rossi S, De Marchi M, Egidi M, Stoc-chetti N: Effects of hyperoxia on brain tissue oxygen tension in cerebral focal lesions. Acta Neurochir Suppl 81:315–317, 2002

21. McLeod AD, Igielman F, Elwell C, Cope M, Smith M: Mea-suring cerebral oxygenation during normobaric hyperoxia: a comparison of tissue microprobes, near-infrared spectrosco-py, and jugular venous oximetry in head injury. Anesth Analg 97:851–856, 2003

22. Menzel M, Doppenberg EM, Zauner A, Soukup J, Reinert MM,

Bullock R: Increased inspired oxygen concentration as a factor in improved brain tissue oxygenation and tissue lactate levels after severe human head injury. J Neurosurg 91:1–10, 1999

23. Menzel M, Doppenberg EM, Zauner A, Soukup J, Reinert MM, Clausen T, et al: Cerebral oxygenation in patients after severe head injury: monitoring and effects of arterial hyper-oxia on cerebral blood flow, metabolism and intracranial pres-sure. J Neurosurg Anesthesiol 11:240–251, 1999

24. Oddo M, Nduom E, Frangos S, MacKenzie L, Chen I, Ma-loney-Wilensky E, et al: Acute lung injury is an independent risk factor for brain hypoxia after severe traumatic brain in-jury. Neurosurgery 67:338–344, 2010

25. Racheli N, Ron A, Metzger Y, Breskin I, Enden G, Balberg M, et al: Non-invasive blood flow measurements using ultrasound modulated diffused light. Proc SPIE 8223:82232A, 2012

26. Reinert M, Barth A, Rothen HU, Schaller B, Takala J, Seiler RW: Effects of cerebral perfusion pressure and increased fraction of inspired oxygen on brain tissue oxygen, lactate and glucose in patients with severe head injury. Acta Neurochir (Wien) 145:341–350, 2003

27. Ron A, Racheli N, Breskin I, Shecter R: A tissue mimicking phantom model for applications combining light and ultra-sound. Proc SPIE 8583:858307, 2013

28. Rosenthal G, Hemphill JC III, Sorani M, Martin C, Morabito D, Meeker M, et al: The role of lung function in brain tissue ox-ygenation following traumatic brain injury. J Neurosurg 108: 59–65, 2008

29. Rosenthal G, Hemphill JC III, Sorani M, Martin C, Morabito D, Obrist WD, et al: Brain tissue oxygen tension is more in-dicative of oxygen diffusion than oxygen delivery and metab-olism in patients with traumatic brain injury. Crit Care Med 36:1917–1924, 2008

30. Rosenthal G, Morabito D, Cohen M, Roeytenberg A, Derugin N, Panter SS, et al: Use of hemoglobin-based oxygen-carrying solution-201 to improve resuscitation parameters and prevent secondary brain injury in a swine model of traumatic brain in-jury and hemorrhage. Laboratory investigation. J Neurosurg 108:575–587, 2008

31. Rosenthal G, Sanchez-Mejia RO, Phan N, Hemphill JC III, Martin C, Manley GT: Incorporating a parenchymal thermal diffusion cerebral blood flow probe in bedside assessment of cerebral autoregulation and vasoreactivity in patients with severe traumatic brain injury. Clinical article. J Neurosurg 114:62–70, 2011

32. Scheufler KM, Röhrborn HJ, Zentner J: Does tissue oxygen-tension reliably reflect cerebral oxygen delivery and consump-tion? Anesth Analg 95:1042–1048, 2002

33. Stocchetti N, Barbagallo M, Gordon CR, Mensi F, Paparella A, Piazza P, et al: [Arterio-jugular difference of oxygen and intracranial pressure in comatose, head injured patients. I. Technical aspects and complications.] Minerva  Anestesiol 57:319–326, 1991 (Ital)

34. Teranishi K, Scultetus A, Haque A, Stern S, Philbin N, Rice J, et al: Traumatic brain injury and severe uncontrolled haemor-rhage with short delay pre-hospital resuscitation in a swine model. Injury 43:585–593, 2012

Manuscript submitted June 12, 2013.Accepted December 18, 2013.This work was presented in abstract form at the 2013 Israel

Neurosurgical Society Meeting, Herzliya, Israel.Please include this information when citing this paper: published

online January 31, 2014; DOI: 10.3171/2013.12.JNS131089.Address correspondence to: Guy Rosenthal, M.D., Department

of Neurosurgery, Hadassah-Hebrew University Medical Center, P.O. Box 12000, Kiryat Hadassah, Jerusalem, Israel 91120. email: [email protected].