Top Banner
ARTIFICIAL LUNG
50
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Artificial Lung2

ARTIFICIAL LUNG

Page 2: Artificial Lung2

Artificial Lung

Often called blood oxygenators.Replace entirely the pulmonary gas exchange function

(when the natural organ is totally disabled or, while still sound, must be taken out of commission for a limited time to allow a surgical intervention) or to assist the deficient gas transfer capacity of the natural organ, either temporarily, with the hope that the healing process will eventually repair the diseased organ, or permanently, when irreversible lung damage leaves the patient permanently disabled.

Page 3: Artificial Lung2

Since most artificial lungs cannot be placed in the anatomical location of the natural lung, venous blood must be diverted from its normal path through the central veins, right heart, and pulmonary vascular bed and rerouted, via catheters and tubes, through an extracorporeal circuit which includes the artificial lung before being returned, by means of a pump, to the arterial system.

The procedure in which the pulmonary circulation is temporarily interrupted for surgical purposes and gas exchange is provided by an artificial lung is often referred to as extracorporeal circulation (ECC) because, for convenience sake in the operating room, the gas exchange device as well as the pumps which circulate the blood are located outside the body.

Page 4: Artificial Lung2

Gas exchange systems

Stationary film oxygenator: a bulky device in which venous blood was evenly smeared over a stack of vertical wire screen meshes in an oxygen atmosphere, flowing gently downward to accumulate in a reservoir from where blood could be returned to a systemic artery.

Problems: cumbersome dimensions, blood streaming, maintaining a constant blood-gas exchange area.

Page 5: Artificial Lung2

Gas exchange systems

Replacing the stationary film support with rotating screens or rotating discs partly immersed in a pool of blood: Allowed some control of gas transfer performance by

changing the rotational speed. Foaming and hemolysis were encountered at high disc

spinning velocity.

Page 6: Artificial Lung2

Gas exchange systems

Technical advances: The discovery that hydrophobic microporous

membranes, through which gas can freely diffuse, have a high enough surface tension to prevent plasma filtration at the moderate pressures prevailing in the blood phase of an artificial lung.

The large-scale fabrication of defect-free hollow fibers of microporous polypropylene, which can be assembled in bundles, potted and manifolded at each extremity to form an artificial capillary bed of parallel blood pathways immersed in a cylindrical hard shell through which oxygen circulates.

Page 7: Artificial Lung2

Plasma proteins were denatured at the gas interface, leading to blood trauma associated with platelet activation and aggregation, complement activation, and hemolysis.

Manifolding and blood distribution system could match the fluid dynamic efficiency of the pulmonary circulation, where a single feed vessel – the pulmonary artery – branches over a short distance and with minimal resistance to flow into millions of tiny gas exchange capillaries the size of an erythrocyte. In chemistry, hydrophobicity (from the combining form of water in

Attic Greek hydro- and for fear phobos) is the physical property of a molecule (known as a hydrophobe) that is repelled from a mass of water.[1]

Hydrophobic molecules tend to be non-polar and thus prefer other neutral molecules and nonpolar solvents. Hydrophobic molecules in water often cluster together forming micelles. Water on hydrophobic surfaces will exhibit a high contact angle.

Examples of hydrophobic molecules include the alkanes, oils, fats, and greasy substances in general. Hydrophobic materials are used for oil removal from water, the management of oil spills, and chemical separation processes to remove non-polar from polar compounds.

Page 8: Artificial Lung2

Basic principles of operation

Hollow fiber

membranes

Small polymer tubes with microporous

walls of 20 to 50 µm thickness and with

outer diameters from 200 to 400 µm.

The wall pores have characteristic sizes

typically below about 0.1 µm, and the porosity

or volume fraction of the fiber wall can vary

from about 40% to 50%.

Made from hydrophobic polymers (often

polypropylene), so that the membrane wall pores

remain gas-filled and respiratory gases can

diffuse readily across it.

In most artificial lung applications, an oxygen (O2) “sweep gas” flows through

the inside lumens of the hollow fibers, while blood flows outside the hollow

fibers through the interstitial spaces in the hollow fiber

bundle.

Oxygen diffuses down its concentration gradient across the fiber wall into blood, while carbon dioxide (CO2) diffuses

down its concentration gradient from the blood into the sweep gas flowing through the fibers

and is removed when the sweep gas exits the device.

Page 9: Artificial Lung2

Determinants of gas exchange

The overall O2 exchange rate, VO2, is related to the O2 permeance, KO2 (overall mass transfer coefficient for O2 exchange), according to

Where PO2g and PO2b = average O2 partial pressures in the sweep gas and blood phases, respectively, flowing through the artificial lung.

A = total membrane area of the hollow fiber bundle.

Page 10: Artificial Lung2

Determinants of gas exchange

The overall CO2 exchange rate, VCO2, is related to the CO2 permeance, KCO2 (overall mass transfer coefficient for CO2 exchange), according to

Where PCO2g and PCO2b = average CO2 partial pressures in the sweep gas and blood phases, respectively, flowing through the artificial lung.

A = total membrane area of the hollow fiber bundle.

Page 11: Artificial Lung2

Permeance

Inverse of a diffusional resistance. Overall transfer resistance in an artificial lung device has 2 principal

components:

Km and Kb = the membrane and blood-side permeances for each gas (O2 and CO2). 1/Km = diffusional resistance for the membrane itself. 1/Kb = resistance for gas diffusing between the membrane and the flowing blood

stream.

Page 12: Artificial Lung2

Figure 5 illustrates the membrane and blood-side diffusional resistances to gas exchange in artificial lungs by showing the general gradient in CO2 partial pressure from the sweep gas to the blood pathway. Transfer resistance within the sweep gas pathway is negligible. Most of the diffusional resistance resides within a blood-side diffusional boundary layer, and secondarily within the membrane itself. The blood-side and membrane permeances dictate overall gas exchange in artificial lungs and represent serial transport processes whose resistances add directly to determine overall resistance, as in Eq. 3. As serial “resistors” the smallest permeance or largest resistance controls overall gas exchange in an artificial lung.

Page 13: Artificial Lung2

Membrane permeance

Most artificial lungs use microporous hollow fiber membranes. Gas exchange occurs by diffusion through these gas-filled

pores.The hydrophobic nature of the polymers (e.g.

polypropylene) used to make the fiber membranes prevents intrusion of blood plasma into the fiber pores under normal conditions.

Page 14: Artificial Lung2

Most artificial lungs, including standard blood oxygenators, use microporous hollow fiber membranes. Microporous hollow fibers have fixed submicron pores within the wall that are contiguous from outer to inner lumen, and gas exchange occurs by diffusion through these gas-filled pores. The polymer used does not dictate gas exchange through the membrane as much as the pore characteristics and the fiber wall porosity. In artificial lungs the hydrophobic nature of the polymers (e.g. polypropylene) used to make the fiber membranes prevents intrusion of blood plasma into the fiber pores under normal conditions. Most microporous hollow fiber membranes for artificial lungs are manufactures by Celgard (Charlotte, NC), Membranea (Germany), and Mitsubishi Rayon (Japan).

Page 15: Artificial Lung2

Membrane permeance

The membrane diffusional resistance of a microporous hollow fiber depends on the permeance, Km, of the fiber membrane.

Its effect on the overall gas exchange performance of artificial lungs is negligible compared to blood-side permeance.

Can be estimated using simple diffusion principles or measured using gas-gas test systems.

Example: If membrane permeance dictated overall gas exchange, an

artificial lung with 2 m2 membrane area perfused with blood at a PCO2 of 50 mmHg would remove CO2 at a theoretical rate of __60__ L/min.

Km = 1.47 x 10-2 ml/cm2/s/cm Hg

Page 16: Artificial Lung2

The gas exchange rate of artificial lungs is much smaller than this because overall gas exchange is dictated by diffusional boundary layers that arise on fiber surfaces n the flowing blood stream. In practice, therefore, K~Kb unless hollow fibers are coated with nonporous polymers (true membranes) to resist plasma wetting.

Page 17: Artificial Lung2

Membrane permeance

In practice, K~ Kb unless hollow fibers are coated with nonporous polymers (true membranes) to resist plasma wetting.

Plasma wetting A process in which blood plasma infiltrates the microporous walls of

hollow fibers. A common problem when extracorporeal oxygenators are used in

extended respiratory support and can lead to device failure within days.

Results primarily from phospholipids, lipoproteins and/or proteins in blood that absorb onto the fiber polymer surfaces at the plasma interface, rendering the interface hydrophilic and allowing for wetting of the pores by either partial or complete plasma infiltration.

Plasma infiltration markedly diminishes the membrane permeance, Km, because relatively rapid gas phase diffusion is replaced by diffusion through stagnant plasma within fiber pores.

Page 18: Artificial Lung2

Membrane permeance

Composite hollow fibers incorporate a thin nonporous polymer layer as a true membrane or “skin” on the microporous fiber surface.

Blocks infiltration of plasma into pores and is a key functional requirement of artificial lungs for longer-term respiratory support.

Made either by coating an existing microporous fiber with a thin nonporous polymer (a true composite hollow fiber) or by modifying the fabrication of the microporous fiber itself to seal off pores at the surface (an asymmetric hollow fiber).

Page 19: Artificial Lung2

Membrane permeance

The membrane permeance of a composite hollow fiber is essentially dominated by the nonporous polymer layer:

αP and DP are the solubility and diffusivity of the gas within the nonporous polymer.

δ is the polymer layer thickness.The design of composite hollow fiber

membranes for artificial lungs requires a Km that does not significantly reduce overall gas exchange.

Page 20: Artificial Lung2

Nonporous polymer skin that prevents plasma wetting also diminishes membrane permeance because a nonporous polymer can present an impediment to gas diffusion. Pm = polymer permeability to specific gases.As an example, if coated or composite fibers are to

exert no more than a 5% reduction in overall gas exchange for a particular artificial lung design, then Km needs to be greater than 20 times Kb. For this reason, composite hollow fiber membranes for artificial lungs require nonporous polymers with relatively high gas permeabilities (~100 Barriers or greater) that can be coated in a continuous layer of 1 um thickness or less on microporous hollow fiber surfaces.

Page 21: Artificial Lung2

Diffusional Boundary Layers

The blood side permeance, Kb, accounts for gas movement through the diffusional boundary layers that exist adjacent to the fiber surfaces, where fluid velocity is reduced by drag forces.

αb and Db are the effective solubility and diffusion coefficient of the diffusing gas in blood. δbl is the average boundary layer thickness.

The boundary layer thickness on a flat surface grows with distance along the surface in the direction of flow according to

where ν is the kinematic viscosity Db is the species diffusion coefficient V is the bulk flow velocity past the surface.

Page 22: Artificial Lung2

The blood-side permeance of an artificial lung, Kb, accounts for gas movement through the diffusional boundary layers that exist adjacent to the fiber surfaces, where fluid velocity is reduced by drag forces. Gas molecules traverse the boundary layer by molecular diffusion before being exposed to sufficient convection by the blood flowing past fiber surfaces. The blood-side permeance can be expressed as

Where are the effective solubility and diffusion coefficient of the diffusing gas in blood and is an average boundary layer thickness. For O2 and CO2 the effective solubility accounts for increased solubility due to hemoglobin binding (for O2) or carriage as bicarbonate ion (for CO2).

The boundary layer thickness, , depends on the local interaction between diffusional and velocity fields in the flowing blood phase subjacent to the fiber surfaces of the artificial lung. The nature of these diffusional boundary layers is complex, but the simple boundary layer paradigm of laminar flow past a flat membrane surface can be instructive. Boundary layer thickness on a flat surface grows with distance along the surface in the direction of flow according to

Where v is the kinematic viscosity, Db is the species diffusion coefficient, and V is the bulk flow velocity past the surface. An important concept is that boundary layer thickness can be decreased by increasing the blood flow velocity past the fiber surfaces, and the resulting increase in gas exchange permeance varies as the square root of flow velocity. Furthermore, because boundary layers grow along the fiber surface, permeance and gas exchange are less with longitudinal flow, parallel to the fiber axes, than with transverse or cross flow, perpendicular to the fiber axes. The simple boundary layer paradigm predicts that Kb for transverse versus longitudinal flow would be Kbtran/Kblong ~ sqrt(L/d), where L and d are fiber length and diameter, respectively. Since L/d in hollow fiber bundles can vary from 100 to 1000, an appreciable mass transfer benefit exists for transverse compared to parallel blood flow through hollow fiber bundles.

Page 23: Artificial Lung2

The blood oxygenator

Membrane blood oxygenators consisting of either microporous polypropylene hollow fiber membranes or, as in one design, silicone sheets.

Blood enters the oxygenator through an inlet port and flows either along the outside of the hollow fibers or the outside of the silicone sheet. The blood is then collected in a manifolded region, flows through a heat exchanger, and then exits the device through an outlet port.

The gas, which can be pure oxygen or a mixture of oxygen and room air, enters the oxygenator through a gas inlet port, flows through the inside of the hollow fibers/silicone sheets, and exits the device via an outlet port.

Page 24: Artificial Lung2

Artificial lungs that are used currently are membrane blood oxygenators consisting of either microporous polypropylene hollow fiber membranes or, as in one design, silicone sheets. The general anatomy of the oxygenator is similar between the two types of devices despite the differing gas exchange surfaces. Blood enters the oxygenator through an inlet port and flows either along the outside of the hollow fibers or the outside of the silicone sheet. The blood is the collected in a manifolded region, flows through a heat exchanger, and then exits the device through an outlet port. The gas, which can be pure oxygen or a mixture of oxygen and room air, enters the oxygenator through a gas inlet port, flows through the inside of the hollow fibers/silicone sheets, and exits the device via an outlet port. The key design considerations in blood oxygenators include minimizing the resistance to blood flow, reducing the priming volume, ensuring easy debubbling at setup, and minimizing blood activation and thrombogenicity.

Page 25: Artificial Lung2

The blood oxygenator

Key design considerations: minimizing the resistance to blood flow, reducing the priming volume, ensuring easy debubbling at setup, and minimizing blood activation and thrombogenicity.

Most current blood oxygenators have fiber membranes with outer diameters of 200-400µm and wall thickness of 20-50µm, total membrane surface area of 2-4m2, and blood priming volume of 135-340mL.

Page 26: Artificial Lung2

Most current blood oxygenators (Fig. 9) have fiber membranes with outer diameters of 200–400 mm and wall thickness of 20–50 mm, total membrane surface area of 2– 4 m2, and blood priming volume of 135–340 ml.[17] The hollow fibers are wound or matted within a hard plastic outer shell to produce fiber packing densities in the bundle of 40–60%, and the arrangement of the fiber bundle and blood flow patterns differ between devices.[32] For example, fibers are helically wound in the Medtronic Affinity NT oxygenator. Blood enters the device through a central core channel and is then distributed radially through the fiber bundle. Fibers in the Jostra Quadrox oxygenator are aligned so that blood flow is perpendicular to the gas pathways. Hollow fiber oxygenators with intraluminal blood flow have been designed but are rarely used due to a generally unfavorable high resistance to blood flow.

Page 27: Artificial Lung2

Rated flow

Rated flow: the flow rate through the oxygenator at which an inlet blood saturation of 70% can be oxygenated to an outlet blood saturation of 95%. Measure of the gas exchange capacity of the device.

The rated flow can range from 1 -1 1.8L/min for a neonatal oxygenator and up to 7 L/min for an adult oxygenator.

Page 28: Artificial Lung2

The fabrication or wrapping of the fiber bundle in a blood oxygenator can be important as the geometry obtained impacts diffusional boundary layers, secondary flows, and gas exchange efficiency. A blood oxygenator is often characterized by its rated flow as a measure of the gas exchange capacity of the device. Rated flow is the flow rate through the oxygenator at which an inlet blood saturation of 70% can be oxygenated to an outlet blood saturation of 95%. The rated flow can range from 1-1.8L/min for a neonatal oxygenator and up to 7 L/min for an adult oxygenator with increased gas exchange capacity.

As flow through the membrane increases, actual O2 transfer increases proportionally until the residence time of the venous return prevents complete hemoglobin saturation. At this point, the absolute O2 transfer becomes fixed, but as the flow continues to increase, a smaller percentage of the venous return becomes saturated. R represents the rated flow, which is the flow at which the blood leaving the membrane is 95% saturated.

Page 29: Artificial Lung2

Silicone membrane oxygenators vs. microporous hollow fiber oxygenators

Silicone membrane oxygenators are often used in extracorporeal membrane oxygenation for respiratory support since plasma leakage does not occur as it does in microporous hollow fiber oxygenators.

Because diffusion occurs across a nonporous silicone sheet, the thickness of these sheets was reduced to 100-200µm. Nevertheless, the gas exchange efficiency of silicone oxygenators is substantially below that of hollow fiber oxygenators.

Page 30: Artificial Lung2

Silicone membrane oxygenators are often used in extracorporeal membrane oxygenation for respiratory support since plasma leakage does not occur as it does in microporous hollow fiber oxygenators. Kolobow[33] is generally credited with developing the first spiral-wound silicone membrane oxygenators in 1963. The oxygenator contains two silicone sheets sealed around the edges, which are wound around a polycarbonate core. Gas flows within the sealed sheets and blood flows countercurrently between the spiral wraps. The surface area of silicone membrane oxygenators ranges from 0.4 to 4.5 m2 and the priming volumes range from 90 to 665 ml.[33] Because diffusion occurs across a nonporous silicone sheet, the thickness of these sheets was reduced to 100–200 mm.

Nevertheless, the gas exchange efficiency of silicone oxygenators is substantially below that of hollow fiber oxygenators. The Avecor 0800 silicone oxygenator (a descendant of the Kolobow silicone oxygenator) has an O2 transfer efficiency of 88 ml/min/m2 compared to 150 ml/ min/m2 for the Affinity hollow fiber device.[33] Them resistance to blood flow is also higher in silicone sheet oxygenators compared with hollow fiber oxygenators, and debubbling the sheet oxygenators can be more difficult.

Page 31: Artificial Lung2

Cardiopulmonary bypass (CPB)

A procedure that allows the temporary replacement of the gas exchange function of the lungs and the blood-pumping function of the heart (bypass the heart cavities and the pulmonary circulation).

Equipment used: pump oxygenator and heart-lung machine.On the blood side – hemodilution, some degree of

hypothermia, nonpulsatile arterial perfusion at a flow rate near the resting cardiac output, and continuous recirculation of blood in an extracorporeal circuit in series with the systemic circulation of the patient.

On the gas side – oxygen or an oxygen-enriched gas mixture (with or without a low concentration of CO2) flows from a moderately pressurized source in a continuous, nonrecycling manner and is vented to the room atmosphere.

Page 32: Artificial Lung2

Oxygenators in Cardiopulmonary Bypass

Cardiopulmonary bypass (CPB) uses an external flow circuit incorporating a blood oxygenator.

Components of CPB: heat exchanger, flowmeters & blenders, suction devices, cardiotomy reservoir, pressure and temperature monitors, sampling ports, filters, tubing, and cannulae.

Page 33: Artificial Lung2

In CPB, blood is drained by gravity from the inferior/superior vena cava or the right atrium into a venous reservoir and is then pumped through the oxygenator by either a roller or centrifugal pump back into the ascending aorta.

Blood flow during CBP is kept low (2–2.4 l/m2/min) to minimize bleeding.[35] A heat exchanger is required to cool and rewarm the patient and is typically incorporated into the oxygenator. Oxygen, or a mixture of oxygen and carbon dioxide, is fed through flowmeters and blenders

into the oxygenator at flow rates of 5–10 l/min, which is 2–3 times the flow rate of blood.[36] The oxygenator must be capable of transferring up to 250 ml/min of oxygen and 200 ml/min of carbon dioxide during cardiopulmonary bypass in order to meet the metabolic needs of the patient.[36] The bypass circuit also includes suction devices that are used to maintain a blood-free surgical field. The suctioned blood is collected and filtered in a cardiotomy reservoir and is then pumped into the venous reservoir. Other components of the bypass circuit include pressure and temperature monitors, sampling ports, filters, tubing, and cannulae.

Page 34: Artificial Lung2

Typical operating conditions for a CPB

Venous reservoir Cardiotomy reservoir

Heat exchanger

Oxygenator

Blood from central veins

Blood from coronary suction

10 – 20 oC0 – 5 oCIce water

5 – 10 L/min

Blood to patient 5L/min

mmHg6753047

mmHgpO2 713pCO2 0pH2O 47

Gas

pO2 40 mmHgpCO2 45 mmHg

Typical Operating Parameters for Cardiopulmonary Bypass in an Adult

Oxygen transfer requirement 250 mL/min

CO2 elimination requirement 200 mL/min

Respiratory gas exchange ratio (respiratory quotient) 0.8

Blood flow rate 5 L/min

Gas flow rate 5-10 L/min

Gas partial pressures (in mmHg)Blood in pO2 = 40 pCO2 = 45Blood out pO2 = 100-300 pCO2 = 30-40Gas in (humidified) pO2 = 250-713 pCO2 = 0-20Gas out pO2 = 150-675 pCO2 = 10-30

Page 35: Artificial Lung2

The respiratory quotient (or RQ or respiratory coefficient), is a unitless number used in calculations of basal metabolic rate (BMR) when estimated from carbon dioxide production. Such measurements, like measurements of oxygen uptake, are forms of indirect calorimetry. It is measured using Ganong's Respirometer.

RQ = CO2 eliminated / O2 consumed

Page 36: Artificial Lung2

Complications with CPB

Activation of the coagulation cascades and thrombosis. Thrombus formation in the oxygenator can cause an

increase in resistance to blood flow and a decrease in gas transfer.

To reduce the risk of clot formation, oxygenators are designed to minimize regions of blood flow stasis, which typically promote thrombus formation.

High level of anticoagulation: increased risk of bleeding.

Oxygenators and the entire bypass circuit are now being coated with heparin in order to prevent clotting in the circuit while reducing the required amount of systemic anticoagulation.

Page 37: Artificial Lung2

There are several different heparin coatings currently available on the market.[38] The Carmeda Bioactive Surface has been used for more than a decade and utilizes a covalent end-point attachment of heparin to the surface. The BioLine Coating by Jostra first coats the surface with polypeptides and then with a low-molecularweight heparin, Liquemin. The coating is available in two types, one for CBP and one for long-term use in extracorporeal membrane oxygenation. Other more recent coatings include AOThel by Artificial Organ Technology, Corline by Corline Systems AB, and the Trilium Biopassive Surface by Avecor. Many studies have been performed on the efficacy of the heparin coatings and the required level of systemic heparin with the coatings. Aldea et al.[39] compared noncoated circuits and an ACT of 480 seconds with coated circuits and an ACT of 280 seconds. The heparin coating resulted in a 34% decrease in the need for blood products, 13.8% less bleeding, 43.6% shorter intubation time, 41.7% less time in the intensive care unit, and 17.8% less time in the hospital compared with noncoated circuits.

Page 38: Artificial Lung2

Extracorporeal Membrane Oxygenation (ECMO)

Uses blood oxygenators in pump-driven external circuits to provide respiratory support and lung rest and recovery for prolonged periods of time (1-30 days).

Used in patients with severe lung failure who fail traditional mechanical ventilation.

ECMO circuit contains a pump, a heat exchanger, an oxygenator (no venous reservoir and suctioning equipment).

Page 39: Artificial Lung2

Extracorporeal membrane oxygenation (ECMO) uses blood oxygenators in pump-driven external circuits to provide respiratory support and lung rest and recovery for prolonged periods of time (1–30 days).[40] ECMO is used in patients with severe lung failure who fail traditional echanical ventilation. Similar to CPB circuits,

the ECMO circuit contains a pump, a heat exchanger, and an oxygenator, but unlike CPB circuits, a venous reservoir and suctioning equipment is not used. In ECMO, the patient is continuously anticoagulated with heparin to achieve an ACT of 160–240 seconds, much less than that found in CPB.[41] The required blood flow in ECMO is 120 ml/kg/min for neonates, 75 ml/kg/min for pediatric, and 50 ml/kg/min for adults.[33] Extracorporeal membrane oxygenation is used to treat neonatal, pediatric, and adult patients with lung failure, and the effectiveness of ECMO differs in each of these groups. ECMO is most commonly used in neonates with a survival rate of 80%.[42] Indications for neonatal ECMO include meconium aspiration syndrome, respiratory distress syndrome, persistent fetal circlation, persistent pulmonary hypertension, and hyaline membrane disease.[43] Pediatric and adult patients have lower survival rates of 53% and 41%, respectively.[42] Indications for ECMO in pediatric or adult patients are viral, bacterial, or aspiration pneumonia and acute respiratory distress syndrome (ARDS), which can be caused by trauma, pneumonia, or sepsis.[44]

Page 40: Artificial Lung2

Cannulation techniques

Cannulation

techniques

Venovenous

Venoarterial

Page 41: Artificial Lung2

Venovenous (VV) ECMO

Drains and returns to the venous system.Most commonly used cannulation techniques.Sites for cannulation: internal jugular, saphenous,

femoral veins or the right atrium.Provides no cardiac support and is not used in

patients with cardiac arrest, arrhythmias, or myocardial failure.

The cannula must be designed to reduce recirculation of returned blood directly back into the ECMO circuit. Cardiac output, pump flow rate, cannula position, and right

atrium size are all factors that can affect recirculation.

Page 42: Artificial Lung2

Venovenous ECMO drains and returns to the venous system, venoarterial ECMO drains from the venous system and returns into the arterial system, and arteriovenous ECMO is the opposite. Venovenous (VV) ECMO was established in the 1960s and 1970s and is now

the most commonly used cannulation technique.[33,45] VV ECMO has several different sites for cannulation including the internal jugular, saphenous, or femoral veins or the

right atrium. In neonates, VV ECMO can use a single dual-lumen cannula or two cannulae. The single doublelumen cannula is used in the jugular vein, and the septum

offset produces a larger channel for venous inflow into the ECMO circuit. The cannula must be designed to reduce recirculation of returned blood directly back into the

ECMO circuit. Cardiac output, pump flow rate, cannula position, and right atrium size are all factors that canaffect recirculation. Single double-lumen cannulation

cannot be used in pediatric and adult ECMO due to inadequate venous inflow into the circuit and also high levels of hemolysis, recirculation, and pressure with flow

rates greater than 600 ml/min.

Page 43: Artificial Lung2

Venoarterial (VA) ECMO

Drains from the venous system and returns into the arterial system.

Original cannulation technique used in ECMO.

Indicated when cardiac support is required in addition to respiratory support.

Disadvantages of VA ECMO: Cannulation of a major artery. Lack of pulmonary perfusion. Decreased cardiac output due to a

higher afterload. Increased risk of neurological events.

Page 44: Artificial Lung2

Venoarterial (VA) ECMO was the original cannulation technique used in ECMO and is indicated when cardiac support is required in addition to respiratory support. In contrast, VV ECMO provides no cardiac support and is not used in patients with cardiac arrest, arrhythmias, or myocardial failure.[45] The disadvantages of VA ECMO include cannulation of a major artery, lack of pulmonary perfusion, decreased cardiac output due to a higher afterload, and increased risk of neurological events. VV ECMO has several advantages over VA ECMO including preserving pulsatility and avoiding the cannulation of a major artery. Neurological events can also be reduced since thromboemboli from the circuit travel to the lungs instead of the brain. VV ECMO also prevents ischemic injury to the lungs since the lungs remain perfused with blood, but blood flow must be carefully regulated in order to prevent an imbalance in the central venous system.[45]Given its advantages compared to VA ECMO, several institutions are now using VV ECMO and comparing results with VA ECMO. Knight et al.[46] found an increased survival rate of 91% with VV ECMO compared to 80% with VA ECMO in neonates. Zahraa et al.[47] performed a retrospective study from 1986–1997 comparing VV and VA ECMO in pediatric patients and found a trend for improved survival with VV ECMO with survival rates of 60% and 56%, respectively.

Page 45: Artificial Lung2

Advantages of VV ECMO over VA ECMO

Preserving pulsatility.Avoiding the cannulation of a major artery. Neurological events can also be reduced

since thromboemboli from the circuit travel to the lungs instead of the brain.

Prevents ischemic injury to the lungs since the lungs remain perfused with blood, but blood flow must be carefully regulated in order to prevent an imbalance in the central venous system.

Page 46: Artificial Lung2

Given its advantages compared to VA ECMO, severalinstitutions are now using VV ECMO and comparingresults with VA ECMO. Knight et al.[46] found anincreased survival rate of 91% with VV ECMO

comparedto 80% with VA ECMO in neonates. Zahraa et al.[47]performed a retrospective study from 1986–1997

comparingVV and VA ECMO in pediatric patients and founda trend for improved survival with VV ECMO withsurvival rates of 60% and 56%, respectively.

Page 47: Artificial Lung2

Complications with ECMO

Plasma wetting of hollow fiber membranes from the longer-term exposure of the ECMO oxygenator to blood. Plasma wetting decreases gas exchange, can

occur quickly and unpredictably, and requires replacement of the oxygenator.

Microporous hollow fiber membranes can be coated with thin siloxane layers to prevent plasma wetting and increase the biocompatibility.

Page 48: Artificial Lung2

The inflammatory and thrombogenic complications associated with cardiopulmonary bypass are exacerbated in ECMO due to the longer blood exposure to the extracorporeal circuit. As for CPB circuits, ECMO circuits and oxygenators are heparin-coated to help minimize systemic heparinization, decrease inflammatory responses, and prevent thrombosis. One complication of ECMO not seen in CPB is plasma wetting of hollow fiber membranes from the longer-term exposure of the ECMO

oxygenator to blood. Plasma wetting decreases gas exchange, can occur quickly and unpredictably, and requires replacement of the oxygenator. Microporous hollow fiber membranes can be coated with thin siloxane layers to prevent plasma wetting and increase the biocompatibility.[ 48–52] New polymer coatings are also being developed to resist plasma leakage while attenuating the inflammatory

response. Saito et al.[52] coated CBP circuits with poly(2-methoxyethylacrylate) (PMEA) and compared the inflammatory response with that caused by uncoated circuits in swine. Protein adsorption was significantly less on the PMEA circuits compared with control (0.3±0.03

mg/cm2 versus 3.42±0.04 mg/cm2). Peek et al.[53] performed an initial clinical trial with the Medos Hilite 7000LT oxygenator, which uses a polymethyl pentene (PMP) asymmetric hollow fiber membrane, which was also coated with heparin. Additional studies are needed to fully evaluate the effectiveness of these new coated fiber oxygenators.

Page 49: Artificial Lung2

Artificial Lung vs. Natural LungNatural lung Hollow fiber oxygenator

PerformanceMust meet demand at rest and during exercise, fever, etc.Constant temperature process around 37oC.O2 and CO2 transfer are matched to achieve the respiratory quotient imposed by foodstuff metabolism (around 0.8).Continuous over a life time

Must meet demand at rest and under anesthesia.Can be coupled with heat exchanger to lower body temperature.O2 and CO2 transfer are largely independent of each other and must be controlled by the operator.Usually limited to few hours.

Exchange surface areaWide transfer are (~70m2).Highly permeable alveolar-capillary membrane

Short diffusion distances (1-2 μm)Hydrophilic membraneNo hemocompatibility problemsSelf-cleaning membrane

Limited transfer area (1-3m2).Diffusion barriers in synthetic membrane and blood oxygenation boundary layerRelatively thick membranes (50-100 μm).Hydrophobic polymersHemocompatibility problemsProtein-build-up on membrane

Gas sideTo-and-fro ventilationOperates with airMembrane structure sensitive to high oxygen partial pressurePressure below that in blood phase to avoid capillary collapseOperates under water vapor saturation conditionsVentilation linked to perfusion by built-in control mechanisms

Can be used for gaseous anesthesia

Steady cross flow gas supplyOperates with oxygen-rich mixtureMembrane insensitive to high oxygen partial pressurePressure below that on blood side to avoid bubble formationCan be clogged by water vapor condensationVentilation dissociated from perfusion, with risks of hyper- or hypoventilationCan be used for gaseous anesthesia

Blood sideShort capillaries (0.5-1 mm)Narrow diameter (3-7 μm)Short exposure time (0.7 s)Low resistance to blood flowSophisticated branchingMinimal priming volumeCapillary recruitment capabilityNo recirculationLimited venous admixtureNo on-site blood mixingOperates with normal hemoglobin concentrationDoes not require anticoagulation

Long blood path (10-15cm)Thick blood film (150-250 μm)Long exposure time (5-15 s)Moderate to high resistance to blood flowCrude manifolding of entry and exit portsModerate to large priming volumeFixed geometry of blood pathPossibility of recirculationRisk of uneven perfusion of parallel bedsPossibility of blood stirring and mixingHemodilution is common Requires anticoagulation.

Page 50: Artificial Lung2

In the natural lungs, the factors underlying exchange across the alveolo-capillary barrier and transport by the blood can be grouped into four classes:

1) The ventilation of the lungs (the volume flow rate of gas) and the composition of the gas mixture to which mixed venous (pulmonary artery) blood will be expored.

2) The permeability of the materials which separate the gas phase from the blood phase in the pulmonary alveoli.3) The pattern of pressure and flow through the airways and through the pulmonary vascular bed and the distribution

of inspired air and circulating blood among the various zones of the exchange system.4) The gas carrying capacity of the blood as regards oxygen and carbon dioxide (and secondarily nitrogen and

anesthetic gases).

In an artificial lung, replacing the gas transfer function of the natural organ implies that blood circulation can be sustained by mechanical pumps for extended periods of time to achieve a continuous, rather than a batch process, and that venous blood can be arterialized in that device by exposure to a gas mixture of appropriate composition. The external gas supply to an artificial lung does not pose particular problems, since pressurized gas mixtures are readily available. Similarly, the components of blood which provide its gas-carrying capacity are well identified and can be adapted to the task at hand. In clinical practice, it is important to minimize the amount of donor blood needed to fill the extracorporeal circuit, or priming volume. Therefore a heart-lung machine is generally filled with an electrolyte or plasma expander solution (with or without donor blood), resulting in hemodilution upon mixing of the contents of the extracorporeal and intracorporeal blood circuits. The critical aspects for the operating of an artificial lung are blood distribution to the exchanger, diffusion resistances in the blood mass transfer boundary layer, and stability of the gas exchange process.

Artificial lungs are expected to perform within acceptable limits of safety and effectiveness. The most common clinical situation in which an artificial lung is needed is typically of short duration, with resting or basal metabolism in anesthetized patients. Table 66.3 compares the structures and operating conditions of the natural lung and standard hollow fiber artificial membrane lungs with internal blood flow.

Page 66-7