ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on Building Asia-Pacific Individual Monitoring Service (IMS) Capabilities toward Regional Sustainable Network. At Pullman Bangkok King Power Hotel, Bangkok Thailand 18-20 March 2019 1
35

ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Aug 18, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

ARPANSA - Primary Standards Dosimetry Laboratory (PSDL)

Report

Duncan Butler

Workshop on Building Asia-Pacific Individual Monitoring Service (IMS) Capabilities toward Regional Sustainable Network.

At Pullman Bangkok King Power Hotel, Bangkok Thailand 18-20 March 2019

1

Page 2: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

ARPANSA

1. Structure of organization

2. Personnel

3. Facilities and conditions

4. Equipment

5. QA/QC programme

6. Number of service/ samples

7. Future research/ new topics

2

In Melbourne, Australia

Page 3: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Structure

3

Office of the CEO

Radiation Health

Services

Corporate Office

Medical Radiation Services

Regulatory services

PSDL

ACDS

Medical Imaging

ARPANSA CEO

Primary Standard Dosimetry Laboratory

Australian Clinical Dosimetry Services

~ 140 staff in totalConcerned with radiation protection (UV, ionising, mobile phones, EM, nuclear safety…)

Personal Monitoring

UV

CTBT

Page 4: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

ARPANSA: PSDL Staff

4

Page 5: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

PSDL staff (Nov 2018)

Viliami TakauAnna HaytonTracy Bailey (0.4)Chris OliverMax Hanlon (0.6)Clare SmithPeter HartyCallum WatsonDuncan Butler (on secondment for 1 year)

5

Page 6: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

ACDS : Australian Clinical Dosimetry Service

• Checks radiotherapy doses in hospitals

6

By mail: Al2SO4 OSLDsOn-site

Page 7: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

ACDS

7Director, Dr Jessica Lye

Page 8: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Personal Radiation Monitoring Service

8

Page 9: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

PSDL• Maintain primary standards for radiation dosimetry

• Provide personal monitor irradiations

• Calibrate equipment

- Our focus is on medical radiation

Research/projects:

• Monte Carlo modelling of ion

chambers

• Small field dosimetry

• IAEA Hp(10) comparisons

• Adopting ICRU90 data

• Spatial response of radiation detectors9

Page 10: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Laboratory facilities

10

60Co Linac (2009) New linac (2019)

40 – 320 kVp X-rays 10 – 100 kVp X-rays Am/Be neutrons

137Cs – not shown

Page 11: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Protection-level irradiations

• Have put irradiators into storage (!)

• Can no longer perform low dose rate calibrations

but

• Personal monitor irradiations can be performed with the high dose rate sources (1 mSv/min)

• Attenuate beam for many protection calibrations

11

Page 12: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

New linac at ARPANSA

12

Pouring concrete for shielding

Page 13: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

13

Page 14: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

14

Official opening March 2019

Page 15: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

15

Official opening March 2019

Page 16: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Equipment – Primary Standards

16

Cavity Chamber for air kerma in 60Co

Free air chambers for air kerma in X-ray beams

10-100 kVp 40-320 kVp

Graphite calorimeter for absorbed dose in 60Co and

linac beams

Page 17: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Graphite calorimeter in use…

17Measuring absorbed dose to graphite in linac beam

Page 18: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Example: Irradiation using an ISO Narrow Series X-ray beam

18

Radiation source

Primary standard(ARPANSA free air chamber)

Page 19: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Example: Irradiation using an ISO Narrow Series X-ray beam

19

Radiation source

Primary standard(ARPANSA free air chamber)

Air kerma here

Uncertainty in kerma is 0.4% (k=1)

Page 20: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Example: Irradiation using an ISO Narrow Series X-ray beam

20Uncertainty in Hp(10) > uncertainty in K

Radiation source

Page 21: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Equipment – Electrometers

21

Use several PTW UNIDOS Webline electrometers- used for protection-level and therapy-level measurements

Many radiotherapy ionization chambers and solid state detectors

Page 22: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Equipment – Other

22

Exradin A4 and A5 ionization chambers

• Temperature (thermistors and PRT)

• Pressure (Druck 141)• Humidity

• Voltage calibrator • Standard resistors

All traceable to national standards

Page 23: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Personal monitor irradiations

• 60Co…1250 keV

• 137Cs…662 keV

• kV X-rays…40 – 320 kVp

ISO Narrow Series X-rays

ISO Wide Series X-rays

ARPANSA therapy beams

• Am/Be neutrons

23

Page 24: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

QA program for a PSDL

1. International comparisons of primary standards “Key comparisons” – in Asia Pacific APMP and with BIPM in Paris.

2. Supplementary and bi-lateral comparisons

3. Internal checks: records of constancy

Accreditation of services to ISO 17025(17025 = standard for

calibration laboratories)

On-site peer reviews

24

Page 25: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

1. Example: key comparison air kerma in kV X-rays

25

1%

Laboratory

Page 26: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

26

2. Example: Bi-lateral comparison Japan-Australia

Page 27: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

History (QC charts)

27

Example: Co-60 air kerma measurements

Page 28: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

QA/QC for personal monitoring

ARPANSA PSDL provides exposures for:

Commissioning

• Energy response (from kV X-rays, 137Cs, 60Co)

• Linearity (0.01 mSv to 100 mSv)

(also establishes stdev versus dose)

• Fading

Ongoing QA

• Monthly 1 mSv control monitors (20)28

Page 29: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

IAEA-ARPANSA Comparison of Personal Dosimeters for the Asia

Pacific Region 2018

29

Page 30: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Dosimetry course for medical physicists

30

Page 31: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Ion chamber spatial response

31

Published in Nuclear Inst. and Methods in Physics Research A 914 (2019) 15–24

Response maps in B-fields – in preparation

Page 32: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Measuring small fields: application of TRS-483

• Used protocols reference conditions:

– Water phantom (PTW Blue Phantom II), Z= 10 gcm-2, SSD = 100 cm.

• Elekta Synergy linac (6 MV WFF X-ray beam)

32

Page 33: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Comparison of detector types

33

0.6

0.7

0.8

0.9

1.0

5 15 25 35 45

Ou

tpu

t fa

cto

r (u

nco

rre

cte

d)

Cone size (mm)

Uncorrected output factor for cones

MD 1 MD 2 MD 3 MD 4

e- diode EFD p diode PFD

SRS PP 31014 PP 31016 CC04

0.6

0.7

0.8

0.9

1.0

5 15 25 35 45

Ou

tpu

t fa

cto

r (c

orr

ect

ed

)

Cone size (mm)

Corrected output factor for cones

MD 1 MD 2 MD 3 MD 4

e- diode EFD p diode PFD

SRS PP 31014 PP 31016 CC04

Page 34: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on

Monte Carlo calculations of kQ

34

Figure 1: Visualisation of interactions inside a NE2571 ionisation

chamber with photon source incident from the top. The yellow marks

represent photons, and the red marks electrons. Cross section

enhancement is used to increase the production of electrons.

Page 35: ARPANSA - Primary Standards Dosimetry Laboratory (PSDL) Reportssdl.oap.go.th/ssdlonline2/wp-content/uploads/2019... · Dosimetry Laboratory (PSDL) Report Duncan Butler Workshop on