ARPANSA
1. Structure of organization
2. Personnel
3. Facilities and conditions
4. Equipment
5. QA/QC programme
6. Number of service/ samples
7. Future research/ new topics
2
In Melbourne, Australia
Page 3
Structure
3
Office of the CEO
Radiation Health
Services
Corporate Office
Medical Radiation Services
Regulatory services
PSDL
ACDS
Medical Imaging
ARPANSA CEO
Primary Standard Dosimetry Laboratory
Australian Clinical Dosimetry Services
~ 140 staff in totalConcerned with radiation protection (UV, ionising, mobile phones, EM, nuclear safety…)
Personal Monitoring
UV
CTBT
…
Page 4
ARPANSA: PSDL Staff
4
Page 5
PSDL staff (Nov 2018)
Viliami TakauAnna HaytonTracy Bailey (0.4)Chris OliverMax Hanlon (0.6)Clare SmithPeter HartyCallum WatsonDuncan Butler (on secondment for 1 year)
5
Page 6
ACDS : Australian Clinical Dosimetry Service
• Checks radiotherapy doses in hospitals
6
By mail: Al2SO4 OSLDsOn-site
Page 7
ACDS
7Director, Dr Jessica Lye
Page 8
Personal Radiation Monitoring Service
8
Page 9
PSDL• Maintain primary standards for radiation dosimetry
• Provide personal monitor irradiations
• Calibrate equipment
- Our focus is on medical radiation
Research/projects:
• Monte Carlo modelling of ion
chambers
• Small field dosimetry
• IAEA Hp(10) comparisons
• Adopting ICRU90 data
• Spatial response of radiation detectors9
Page 10
Laboratory facilities
10
60Co Linac (2009) New linac (2019)
40 – 320 kVp X-rays 10 – 100 kVp X-rays Am/Be neutrons
137Cs – not shown
Page 11
Protection-level irradiations
• Have put irradiators into storage (!)
• Can no longer perform low dose rate calibrations
but
• Personal monitor irradiations can be performed with the high dose rate sources (1 mSv/min)
• Attenuate beam for many protection calibrations
11
Page 12
New linac at ARPANSA
12
Pouring concrete for shielding
Page 14
14
Official opening March 2019
Page 15
15
Official opening March 2019
Page 16
Equipment – Primary Standards
16
Cavity Chamber for air kerma in 60Co
Free air chambers for air kerma in X-ray beams
10-100 kVp 40-320 kVp
Graphite calorimeter for absorbed dose in 60Co and
linac beams
Page 17
Graphite calorimeter in use…
17Measuring absorbed dose to graphite in linac beam
Page 18
Example: Irradiation using an ISO Narrow Series X-ray beam
18
Radiation source
Primary standard(ARPANSA free air chamber)
Page 19
Example: Irradiation using an ISO Narrow Series X-ray beam
19
Radiation source
Primary standard(ARPANSA free air chamber)
Air kerma here
Uncertainty in kerma is 0.4% (k=1)
Page 20
Example: Irradiation using an ISO Narrow Series X-ray beam
20Uncertainty in Hp(10) > uncertainty in K
Radiation source
Page 21
Equipment – Electrometers
21
Use several PTW UNIDOS Webline electrometers- used for protection-level and therapy-level measurements
Many radiotherapy ionization chambers and solid state detectors
Page 22
Equipment – Other
22
Exradin A4 and A5 ionization chambers
• Temperature (thermistors and PRT)
• Pressure (Druck 141)• Humidity
• Voltage calibrator • Standard resistors
All traceable to national standards
Page 23
Personal monitor irradiations
• 60Co…1250 keV
• 137Cs…662 keV
• kV X-rays…40 – 320 kVp
ISO Narrow Series X-rays
ISO Wide Series X-rays
ARPANSA therapy beams
• Am/Be neutrons
23
Page 24
QA program for a PSDL
1. International comparisons of primary standards “Key comparisons” – in Asia Pacific APMP and with BIPM in Paris.
2. Supplementary and bi-lateral comparisons
3. Internal checks: records of constancy
Accreditation of services to ISO 17025(17025 = standard for
calibration laboratories)
On-site peer reviews
24
Page 25
1. Example: key comparison air kerma in kV X-rays
25
1%
Laboratory
Page 26
26
2. Example: Bi-lateral comparison Japan-Australia
Page 27
History (QC charts)
27
Example: Co-60 air kerma measurements
Page 28
QA/QC for personal monitoring
ARPANSA PSDL provides exposures for:
Commissioning
• Energy response (from kV X-rays, 137Cs, 60Co)
• Linearity (0.01 mSv to 100 mSv)
(also establishes stdev versus dose)
• Fading
Ongoing QA
• Monthly 1 mSv control monitors (20)28
Page 29
IAEA-ARPANSA Comparison of Personal Dosimeters for the Asia
Pacific Region 2018
29
Page 30
Dosimetry course for medical physicists
30
Page 31
Ion chamber spatial response
31
Published in Nuclear Inst. and Methods in Physics Research A 914 (2019) 15–24
Response maps in B-fields – in preparation
Page 32
Measuring small fields: application of TRS-483
• Used protocols reference conditions:
– Water phantom (PTW Blue Phantom II), Z= 10 gcm-2, SSD = 100 cm.
• Elekta Synergy linac (6 MV WFF X-ray beam)
32
Page 33
Comparison of detector types
33
0.6
0.7
0.8
0.9
1.0
5 15 25 35 45
Ou
tpu
t fa
cto
r (u
nco
rre
cte
d)
Cone size (mm)
Uncorrected output factor for cones
MD 1 MD 2 MD 3 MD 4
e- diode EFD p diode PFD
SRS PP 31014 PP 31016 CC04
0.6
0.7
0.8
0.9
1.0
5 15 25 35 45
Ou
tpu
t fa
cto
r (c
orr
ect
ed
)
Cone size (mm)
Corrected output factor for cones
MD 1 MD 2 MD 3 MD 4
e- diode EFD p diode PFD
SRS PP 31014 PP 31016 CC04
Page 34
Monte Carlo calculations of kQ
34
Figure 1: Visualisation of interactions inside a NE2571 ionisation
chamber with photon source incident from the top. The yellow marks
represent photons, and the red marks electrons. Cross section
enhancement is used to increase the production of electrons.
Page 35