Low Water Levels in Various Solvents: On-line Monitoring by a Miniature
MIR Spectrometer
Joseph P Sung and Jim Tatum, Rohm and Haas Company, Huston, TX, USA
Yael Barshad and Yoav Barshad, Applied Analytics, Inc., Chestnut Hill, MA, USA
Applied Analytics, Inc.
MicroSpec
In-line Infrared Sensors Covering the Mid Infrared from 2 to 14µm (5000 to 700 cm-1 )
Applied Analytics, Inc.
Direct in Stream Measurements
IR light source
A single, dual or up to 6 elements array detector with a filter that isolates a specific IR wavelength
A thermistor to measure temperature
ATR element.
Transmission element
Applied Analytics, Inc.
Applied Analytics, Inc.
Applied Analytics, Inc.
Applied Analytics, Inc.
Applied Analytics, Inc.
Applied Analytics, Inc.
Applied Analytics, Inc.
Applied Analytics, Inc.
MicroSpec Specifications
Detection: MIR Transducer
Pressure: Maximum 150psig (10Bar)
Temperature: 1000C (2120F)
Drift: 0.5% of reading/ 24 hours
Noise: ±0.2% full scale
Electrical: 24 V DC 100 mA
Output: 4 – 20 mA 700 W load galvanically isolated.
Connection: The body is threaded for insertion into the process pipe 1-2-UNFF
Material of construction stainless steel body (316L)
Applied Analytics, Inc.
Applied Analytics, Inc.
Features
•Makes in line quantitative measurements at selected wavelengths
•Contains no moving parts and functions in process stream environments
•Replaces spectrometers in the process plant and eliminates the need for optical fibers
Applied Analytics, Inc.
Applied Analytics, Inc.
0
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.01
1300 1350 1400 1450 1500 1550 1600 1650 1700
nm
AU
Absorbance spectra of water in DMF (100 PPM), DMF used as blank (1cm cell)
Applied Analytics, Inc.
Calibration of Microspec on Sample X on 12/4/02Karl Fischer Data (PPM) Microspec Data180 0189 0.143537 11.83772 23.21216 35.41305 38.33
Applied Analytics, Inc.
Microspec Calibration Curve on Sample X
y = 0.0344x - 5.9332
R2 = 0.9944
0
5
10
15
20
25
30
35
40
45
0 200 400 600 800 1000 1200 1400
Karl Fischer Data (PPM)
Mic
ros
pe
c D
ata
Applied Analytics, Inc.
Microspec vs. Karl Fischer (PPM), Sample X
y = 0.9276x - 71.0565
R2 = 0.9713
0
200
400
600
800
1000
1200
1400
0 200 400 600 800 1000 1200 1400
Karl Fischer Data (PPM)
Mic
ros
pe
c D
ata
Applied Analytics, Inc.
Applied Analytics, Inc.
Microspec Calibration (Sample Y)
y = -2.1454x + 272.29
R2 = 0.9806
228
230
232
234
236
238
240
15 16 17 18 19 20 21
Karl Fischer Data %
Mic
ros
pe
c D
ata