Top Banner
Analysis of Explosives Using a Core Enhanced Technology Accucore HPLC Column Stephen Aspey, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract This application note demonstrates the use of the Thermo Scientific Accucore C18 column for the fast analysis of explosives. Introduction Accucore™ HPLC columns use Core Enhanced Technology to facilitate fast and high efficiency separations. The 2.6 μm diameter particles are not totally porous, but rather have a solid core and a porous outer layer. The optimised phase bonding creates a series of high coverage, robust phases. The tightly controlled 2.6 μm diameter of Accucore particles results in much lower backpressures than typically seen with sub-2 μm materials. Explosives are environmental toxins, regulated in EPA 8330. Here we compare the separation of a sub-set of the compounds covered by the EPA method, using a C18 and a perfluoronated phenyl (PFP) HPLC column. Results The Accucore C18 column demonstrated a separation of 5 explosive compounds in less than 3 minutes. This is in comparison with an 8 minute separation using the same conditions, on an equivalent PFP column. The different selectivity exhibited by the PFP column is believed to be a result of pi-pi interactions, which do not occur with conventional alkyl chain HPLC packings, such as C18. Conclusions The Accucore C18 column provides a good separation of 5 explosive compounds. This is achieved in less than 3 minutes. The Accucore PFP provides different selectivity to the C18 chemistry and could lend itself as a confirmatory column for these compounds. Key Words Fused core • Superficially porous • Explosives Core Enhanced Technology • Accucore C18 / PFP Application Note: ANCCSCETEXPLO Sample Preparation 1,3,5-trinitrobenzene (1.0 mg), 1,3-dinitrobenzene (1.2 mg), nitrobenzene (2.1 mg), 2,4,6-trinitrotoluene (1.0 mg) and 2,4-dinitrotoluene (1.7 mg) were each dissolved in 1.0 mL methanol/water (50:50). Each of the explosive solutions (100 μL) were mixed and made up to 1000 μL with methanol/water (50:50). Thermo Scientific Column Part Number Accucore C18 2.6 μm 100 x 2.1 mm 17126-102130 Accucore PFP 2.6 μm 100 x 2.1 mm 17426-102130 Measured backpressure: 264 bar Thermo Scientific HPLC System Column temperature 40 °C Injection volume 1.0 μL Flow rate 0.4 mL/min UV detection 254 nm Mobile Phase Water/Methanol (60:40) Consumables Part Number Fisher Scientific HPLC grade water W/0106/17 Fisher Scientific HPLC grade methanol M/4056/17 NSC Mass Spec Certified 2 mL clear vial with blue MSCERT4000-34W bonded PTFE silicone cap
2

Application Note: ANCCSCETEXPLO Analysis of Explosives Using … · 2011. 9. 27. · Technology Accucore HPLC Column Stephen Aspey, Thermo Fisher Scientific, Runco rn, Cheshire, UK

Aug 28, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Application Note: ANCCSCETEXPLO Analysis of Explosives Using … · 2011. 9. 27. · Technology Accucore HPLC Column Stephen Aspey, Thermo Fisher Scientific, Runco rn, Cheshire, UK

Analysis of Explosives Using a Core EnhancedTechnology Accucore HPLC Column Stephen Aspey, Thermo Fisher Scientific, Runcorn, Cheshire, UK

Abstract

This application note demonstrates the use of the ThermoScientific Accucore C18 column for the fast analysis ofexplosives.

Introduction

Accucore™ HPLC columns use Core EnhancedTechnology to facilitate fast and high efficiencyseparations. The 2.6 µm diameter particles are not totallyporous, but rather have a solid core and a porous outerlayer. The optimised phase bonding creates a series of highcoverage, robust phases. The tightly controlled 2.6 µmdiameter of Accucore particles results in much lowerbackpressures than typically seen with sub-2 µm materials.

Explosives are environmental toxins, regulated inEPA 8330. Here we compare the separation of a sub-set ofthe compounds covered by the EPA method, using a C18and a perfluoronated phenyl (PFP) HPLC column.

Results

The Accucore C18 column demonstrated a separation of5 explosive compounds in less than 3 minutes. This is incomparison with an 8 minute separation using the sameconditions, on an equivalent PFP column. The differentselectivity exhibited by the PFP column is believed to be aresult of pi-pi interactions, which do not occur withconventional alkyl chain HPLC packings, such as C18.

Conclusions

The Accucore C18 column provides a good separation of5 explosive compounds. This is achieved in less than3 minutes. The Accucore PFP provides different selectivityto the C18 chemistry and could lend itself as aconfirmatory column for these compounds.

Key Words

• Fused core

• Superficially porous

• Explosives

• Core Enhanced Technology

• AccucoreC18 / PFP

Application Note:ANCCSCETEXPLO

Sample Preparation

1,3,5-trinitrobenzene (1.0 mg), 1,3-dinitrobenzene (1.2 mg), nitrobenzene(2.1 mg), 2,4,6-trinitrotoluene (1.0 mg) and 2,4-dinitrotoluene (1.7 mg) wereeach dissolved in 1.0 mL methanol/water (50:50). Each of the explosivesolutions (100 µL) were mixed and made up to 1000 µL with methanol/water(50:50).

Thermo Scientific Column Part Number

Accucore C18 2.6 µm 100 x 2.1 mm 17126-102130Accucore PFP 2.6 µm 100 x 2.1 mm 17426-102130Measured backpressure: 264 bar

Thermo Scientific HPLC System

Column temperature 40 °C Injection volume 1.0 µLFlow rate 0.4 mL/minUV detection 254 nm

Mobile Phase

Water/Methanol (60:40)

Consumables Part NumberFisher Scientific HPLC grade water W/0106/17Fisher Scientific HPLC grade methanol M/4056/17NSC Mass Spec Certified 2 mL clear vial with blue MSCERT4000-34Wbonded PTFE silicone cap

Page 2: Application Note: ANCCSCETEXPLO Analysis of Explosives Using … · 2011. 9. 27. · Technology Accucore HPLC Column Stephen Aspey, Thermo Fisher Scientific, Runco rn, Cheshire, UK

Part of Thermo Fisher Scientific

In addition to these

offices, Thermo Fisher

Scientific maintains

a network of represen -

tative organizations

throughout the world.

North America USA and Canada+1 800 332 3331

EuropeFrance+33 (0)1 60 92 48 34

Germany+49 (0) 2423 9431 -20or -21

Switzerland+41 56 618 41 11

United Kingdom+44 1928 534110

AsiaJapan+81 3 5826 1615

China+86-21-68654588or +86-10-84193588800-810-5118

India1800 22 8374 (toll-free)+91 22 6716 2200

Thermo FisherScientific AustraliaPty Ltd 1300 735 292 (free calldomestic)

Thermo FisherScientific NewZealand Ltd 0800 933 966 (free calldomestic)

All Other Enquiries+44 (0) 1928 534 050

TechnicalSupportNorth America800 332 3331

Outside NorthAmerica+44 (0) 1928 534 440

www.thermoscientific.com/chromatography©2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

ANCCSCETEXPLO 0611

Figure 1: The separation of 5 explosives on Accucore C18 and PFP columns.Compounds: 1. 1,3,5-trinitrobenzene 2. nitrobenzene 3. 1,3-dinitrobenzene 4. 2,4,6-trinitrotoluene 5. 2,4-dinitrotoluene

Table 1: Method precision - using retention time, asymmetry and theefficiency of peak 5 and derived from 6 replicate injections

Time (Minutes)0 1 2 3 4 5 6 7 8 9 10

mAU

0

20

40

60

80

100

120

1

2

3

4

5

1

2

3

45

Accucore C18

Accucore PFP

tr/min As N

Mean 7.86 1.11 14351

SD 0.03 0.10 349.9

%CV 0.35 8.6 2.4