Is Low Dose Really Low Dose? A Clinical Audit of Low Radiation Dose CT KUB Studies for Suspected Urinary Tract Calculi Chahal, B.S., Kwan, A.L.C, Lambert R.G.W., Nielsen, M.M., Gauvreau, D., Winters, S.D, Olubaniyi, B.O.
20

AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Apr 28, 2018

Download

Documents

vancong
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Is Low Dose Really Low Dose?

A Clinical Audit of Low Radiation Dose CT KUB Studies for Suspected Urinary Tract

Calculi

Chahal, B.S., Kwan, A.L.C, Lambert R.G.W., Nielsen, M.M., Gauvreau, D., Winters, S.D, Olubaniyi, B.O.

Page 2: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

•  Chahal, B.S. - None

•  Kwan, A.L.C. - None

•  Lambert, R.G.W. – Consultant for Bioclinica

•  Nielsen, M.M. - None

•  Gauvreau, D. - None

•  Winters, S.D. - Member of the Canadian Armed Forces, Department of National Defence

•  Olubaniyi, B.O - None

Declaration of Conflict of Interest

Page 3: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

The Audit Cycle 1. Identify

the problem

2. Set standards

3. Collect Data

4. Compare results with standards

5. Implement change

6. Conclude or re-audit

Page 4: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

•  Average mean effective dose of 11.2 mSv in 49, 903 scans1

•  “…reduced-dose renal protocol CT is used infrequently…”1

Identify the Problem

1. Lukasiewicz et al. Radiology, 2014 May; 271(2):445-451.

Page 5: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

•  CT KUB gold standard for investigating renal colic2

•  High prevalence requiring repeated studies3

•  Cumulative radiation dose can be high3

Background

Page 6: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Audit Team Fellow

MD Student

Department Head/Radiologist

Senior CT Technologist

CT Technologist

Body Imaging Radiologist

Quality Assurance Coordinator/Radiologist

Physicist

Page 7: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

TARGET: 80% at ≤ 3 mSv4

We considered:

A. Two targets

B.  3 protocols based on BMI

Not realistic

Set Standards

Page 8: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Audit Results (1st cycle) 40 studies (from one machine) 3 mSv or less achieved in 6/40 15% [Target: 3 mSv or less in 80%] Target not met

Page 9: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

1. Change Noise Index from 33 to 40

2.  Increase iterative reconstruction blend from 40% to 50% (ASIR)

Implement Change

Page 10: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Re-Audit Results (2nd cycle)

39 studies 3 mSv or less achieved in 7/39 18% [First cycle was 6/40 15%] [Target: 3 mSv or less in 80%]

Target STILL not met

Page 11: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

1. Change Noise Index: 40 to 50

2. mA: 10 minimum - 300 maximum

3. Scan length: •  “Above the liver to lesser trochanter” •  “1cm above the kidneys to 1 cm below the

pubic symphysis”

Implement Change…Again

Page 12: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Re-Audit Results (3rd cycle) 37 Studies 3 mSv or less achieved in 15/37 41% [Second cycle: 7/39 18%] [First cycle: 6/40 15%] [Target: 3 mSv or less in 80%] Target STILL not met

Page 13: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Comparing Cycles

0 2 4 6 8

10 12 14 16 18

First Cycle Second Cycle Third Cycle

Effe

ctiv

e D

ose

(mSv

)

Max Mean Min

Page 14: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Implementation Issues

1.  Over scanning 2.  Larger patients

Page 15: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Behavior Change STRATEGY EFFECTIVENESS

Emails *

ý

Word of mouth þ ý

Staff Meeting þ Employee Engagement7,8 þþþ

Page 16: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Audit Conclusion •  Target not met at first cycle

•  Target not met after first intervention

•  Target not met after second intervention

•  Third intervention performed. Await fourth cycle. •  Further changes?

Page 17: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Future Direction Kaye Edmonton Clinic

UAH Scanners

Edmonton Hospitals

Page 18: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

1.  Lukasiewicz, AMS., Bhargavan-Chatfield, M., Coombs, L., Ghita, M., Weinreb, J., Gunabushanam, G., Moore, CL. Radiation dose index of renal colic protocol CT studies in the United States: A report from the American College of Radiology National Radiology Data Registry. Radiology, 2014 May; 271(2):445-451.

2.  Yen P, Baerlocher, MO. Five things to know about…Imaging in urolithiasis. CMAJ. 2011 Dec 13;183(18):2133. doi:10.1503/cmaj.101638.

3.  Katz SI, Saluja A, Brink JA, Forman HP. Radiation dose associated with unenhanced CT for suspected renal colic: impact of repetitive studies. AJR Am J Roentgenol. 2006 Apr; 186(4):1120-4.

4.  Niemann T, Kollman T, Bongartz G. Diagnostic performance of low-dose CT for the detection of urolithiasis: a meta-analysis. AJR Am J Roentgenol. 2008 Aug; 191(2):396=401. Doi: 10.2214/AJR.07.3414.

5.  AAPM Report, No. 96. Retrieved from http://www.aapm.org/pubs/reports/RPT_96.pdf

6.  Gervaise A, Naulet O, Beuret F, Henry C, Pernin M,Portron Y, Lapierre-Combes M. Low-dose CT with automatic tube current modulation, adaptive statisical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: impact of body mass index. AJR Am J Roentgenol. 2014 Mar;202(3):553-60. doi: 10.2214/AJR.13.11350.

7.  Scott, TIM., Mannion, R., Davies, HT., Marshall, MN. Implementing culture change in health care: theory and practice. International Journal for Quality in Health Care, 2003; 15(2):111-118.

8.  Davies, H., Powell, A., Rushmer, R. Healthcare professionals’ views on clinician engagement I quality improvement. A literature review. 2007 April. Retrieved from: http://www.health.org.uk/public/cms/75/76/313/534/Healthcare%20professionals%20views%20on%20clinician%20engagement%20in%20quality%20improvement.pdf?realName=W5EHsp.pdf

References

Page 19: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

Acknowledgements: •  Sheara Wilson •  Linda O’Harris, Manager,

Specialty Imaging •  All CT Technologists at UAH

Page 20: AP008 Is Low Dose Really Low Dose Chahal - CAR Lifelong Learning/Meetings/ASM2015...Is Low Dose Really Low Dose? ... 39 studies 3 mSv or less achieved in 7/39 18% ... AAPM Report,

1.  Image quality

2.  Dissemination of recommendations

3.  Educating future technologists about local practice

4.  Preventing overcorrections

Challenges/Limitations