“SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT JASON WENDEROTH PRINCE OF WALES AND LIVERPOOL HOSPITALS SYDNEY [email protected]
33

“SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

Aug 18, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

“SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT

JASON WENDEROTH

PRINCE OF WALES AND LIVERPOOL HOSPITALS

SYDNEY

[email protected]

Page 2: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

TWO MAJOR CHALLENGES IN STENT DEPLOYMENT

• Stent torsion

• Stent prolapse into aneurysm

Page 3: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

THE TORSION PROBLEM

• Tortuous anatomy

• ”Round-the-world” access

• Long stents

Page 4: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

THE TORSION PROBLEM

• Torque on catheters transmitted to FDS

• Mid/proximal stent twists

• Non-opening

• Angioplasty-resistant

Page 5: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

THE PROLAPSE PROBLEM

• Broad-necked/giant aneurysms

• Aneurysms occupying >50% circumference of vessel

• Insufficient proximal, distal or lateral support during “push” phase of stent deployment

Page 6: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

FDS TORSION PROBLEM

• Alternate unsheathing of stent delivery catheter and stent

• Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent

• Intermediate catheter can be used to “nudge” proximal end of stent for optimal opening

Page 7: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

65M ex-smoker

Page 8: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

FIRST PROCEDURE

• Aneurysm crossed “round the world”, microcatheter in M2

• Intermediate catheter proximal to aneurysm

• Failed FDS deployment:

• PED 4.75 x 30mm

• PED 4.5 x 30mm

• FRED 4.5 x 26mm

• Acute stroke presented – procedure abandoned; rebooked

Page 9: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter
Page 10: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter
Page 11: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

THE ”SHEPHERD” TECHNIQUE

• Initial access beyond aneurysm via microcatheter

• Direct or “round the world”

• Standard or balloon microcatheter

• Torque accrues on primary access catheter

• Intermediate catheter advanced distally over primary catheter

• Stent delivery catheter (no torque) via intermediate catheter +/- buddy wire alongside if concerned re stent prolapse

Page 12: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

Intermediate catheter

• Distal balloon access

• Inflate balloon for stability

• Torque stored in loop

Page 13: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Tension on balloon catheter to remove loop

• Torque transmitted to balloon shaft

Page 14: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Balloon shaft straight, tension stored

• Use balloon shaft as rail

Page 15: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Advance intermediate catheter over balloon shaft

Page 16: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Place intermediate catheter well distal to aneurysm

• Can tension balloon shaft to cross aneurysm “lip”

Page 17: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Final position of intermediate catheter

Page 18: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Deflate balloon and remove

• May leave “buddy” wire in position if very difficult/tortuous access

Page 19: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

Buddy wire

Page 20: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Remove balloon

• Prepare stent-delivery catheter

Page 21: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Advance stent delivery catheter as distally as is safe

• Maintain intermediate catheter position +/-buddy wire

Page 22: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Commence unsheathing of stent lead wire

• Maintain intermediate catheter position

Page 23: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Withdraw intermediate catheter

• Expose stent delivery catheter in preparation for stent delivery

Page 24: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Gradual retraction of delivery catheter until stent opens distally and apposes wall

Page 25: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Slow unsheathing of delivery catheter to permit unsheathing of next stent segment

• Buddy wire can be very helpful here…

Page 26: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Next segment of stent delivered

• Manipulation by freely-moving delivery catheter with intermediate catheter support

Page 27: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Gradual retraction of delivery catheter and intermediate catheter until stent opens distally and apposes wall

Page 28: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

• Next segment of stent delivered

• Manipulation by freely-moving delivery catheter with intermediate catheter support

Page 29: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

65M ex-smoker

Page 30: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

65M ex-smoker

Page 31: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

65M ex-smoker

Page 32: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter

SUMMARY

• You’re probably doing this already

• If not, try it – it works!

• We find it useful with difficult anatomy to get our stents open