    Anritsu Master series OTDR

    Presents:

    The next generation inOTDRs

    TheACCESS Master

    MT9083A / B / C

    TheNetwork Master

    MT9090A

    Slide 2

    Anritsu Master series OTDR

    OTDR 101

    Before we start, lets review key

    OTDR principles and terminology

  • 8/11/2019 Anritsu OTDRs

    2/45

    Slide 3

    Anritsu Master series OTDR

    OTDR Operation

    Bloc k Diagram

    Slide 4

    Anritsu Master series OTDR

    d = distance

    t = elapsed time

    c = speed of light in a vacuum

    n = Index of Refraction

    Test pulse fired

    Test pulse returned

    d

    d=t C

    2 n If n is incorrect, thenthe distance measuredwill also be wrong!!

    Must divide by

    two to get one

    way elapsed time.

    DistanceMeasurements

  • 8/11/2019 Anritsu OTDRs

    3/45

    Slide 5

    Anritsu Master series OTDR

    Fiber PhenomenaBackscattering

    Transmitted Light

    Random scattering

    Backscattering

    Fresnel Reflection

    Slide 6

    Anritsu Master series OTDR

    Fiber Type

    Multimode Fiber

    Wavelength: 850nm and 1300nm

    Short distance application such as in-building Ethernet

    connection Graded-Index support higher data speed than Step-index

    Singlemode Fiber

    Wavelength: 1310 and 1625nm

    High data speed over long distance application. Metro

    Network, Long-haul network.

  • 8/11/2019 Anritsu OTDRs

    4/45

    Slide 7

    Anritsu Master series OTDR

    Fiber Type

    Slide 8

    Anritsu Master series OTDR

    Fiber Parameter

    IOR (Index Of Refraction) obtain from fiber manufacturer, enter into OTDR

    Affecting the measured cable distance

    Backscatter coefficient obtain from fiber manufacturer, enter into OTDR

    Affecting the measured reflectance.

  • 8/11/2019 Anritsu OTDRs

    5/45

    Slide 9

    Anritsu Master series OTDR

    OTDR BasicsTypical OTDR trace with events

    Slide 10

    Anritsu Master series OTDR

    OTDR Basics

    Deadzones Why event deadzone matters

    Verification of two close reflections

    Provides ability to verify which

    connector is bad (high reflectance)

    With 80 cm deadzone, MT9083A can

    detect two connectors at 1 m apart.

    Why backscatter deadzone matters

    Verification of a second event after areflective event

    Provides ability to test vault splices

    Average MT9083A spec

    is 5 m (15 ft)

    EDZ

    ADZ

    Mechanical connectors

    Mechanical connectors

    Fusion

    splice

  • 8/11/2019 Anritsu OTDRs

    6/45

    Slide 11

    Anritsu Master series OTDR

    OTDR BasicsEvent Deadzone

    Connectors less than 80cm apart Connectors more than 1m apart

    Slide 12

    Anritsu Master series OTDR

    OTDR BasicsBackscatter Deadzone

    Connector Splice

    5m backscatter deadzone

    Undetectable splice

    Connector Splice

    5m backscatter deadzone

    Detectable splice

  • 8/11/2019 Anritsu OTDRs

    7/45

    Slide 13

    Anritsu Master series OTDR

    OTDR Basics

    10m (33)Actual Splice Location

    Measured Splice Location

    Actual End Location

    Measured End Location

    Low Density (wider Data Point Sampling)

    Resolution

    Slide 14

    Anritsu Master series OTDR

    OTDR Basics

    3m (10)Actual Splice Location

    Measured Splice Location

    Actual End Location

    Measured End Location

    High Density (narrow Data Point Sampling)

    Resolution

  • 8/11/2019 Anritsu OTDRs

    8/45

    Slide 15

    Anritsu Master series OTDR

    OTDR BasicsDynamic Range: How far can you measure.

    Dynamic Range OK to measure

    entire fiber.

    Insufficient Dynamic Range

    events near the end of the fiber

    cannot be measured.

    Slide 16

    Anritsu Master series OTDR

    OTDR BasicsHow far can our modules test?

    Converting Dynamic Range into Distance

    Can only be done on a straight piece of glass (no events)

    Need to know the db/Km loss for each wavelength

    Need to know the dynamic range at each wavelength.

    Distance = Dynamic Range = 38dB/0.20dB/km = 190km

    dB/km

    Wavelength Typical Loss

    850nm multimode 3.0dB/km

    1300nm multimode 1.0 dB/km

    1310nm single mode 0.35 dB/km

    1550nm single mode 0.20 dB/km

    1625nm single mode 0.25dB/km

  • 8/11/2019 Anritsu OTDRs

    9/45

    Slide 17

    Anritsu Master series OTDR

    OTDR BasicsPulse Width

    The single most important parameter Use the pulsewidth which still provides sufficient dynamic range

    Too short

    (short deadzonesbut insufficient DR)

    Too long

    (good DR

    but large

    deadzones)Just right

    (good DR, short

    deadzones)

    Slide 18

    Anritsu Master series OTDR

    Pulse Width (PW), what does it effect, Smaller PW = Smaller Dead Zones,

    Able to see close events,Not able to see longer distances. OTDRs specification deadzone is tested at the smallest PW.

    Larger PW = Larger Dynamic Range,Able so large distances,Not able to see close events. OTDRs specification Dynamic Range is tested at the largest PW.

    Its not possible to test close events with a large PW or largedistances with a small PW. Using the Autofunction on your OTDR sets the PWto the cable

    length.Pulse width Dead Zone Dynamic Range

    10ns 2 - 3m 6 - 9 dB

    100ns 10 - 15m 11 - 14dB

    500ns 50m 1821dB

    1us 100m 2225dB

    4us 400m 29 - 32 dB

    OTDR BasicsPulse Width

  • 8/11/2019 Anritsu OTDRs

    10/45

    Slide 19

    Anritsu Master series OTDR

    Presents:

    The next generation inOTDRs

    TheACCESS Master

    MT9083A / B / C

    Slide 20

    Anritsu Master series OTDR

    Historical Fiber Optic SystemBasic point to point deployment

    OP-AMP/RegenerationStations

    City A

    City B

    City A

    City B

    Splice points

    Yesterdays

    networks werevery simple and

    most test

    equipment was

    adequate.

  • 8/11/2019 Anritsu OTDRs

    11/45

    Slide 21

    Anritsu Master series OTDR

    But networks are evolvingCurrent Situat ion with Op tical Netwo rks

    Increased complexity

    Shift from long-haul to access installations

    Increase in Metro and FTTx installations

    Triple play (including sensitive video applications)

    And

    Budget reductions have forced service providersto get the job done with less people, equipment,

    knowledge and time

    Shouldnt your test equipment evolve too?

    Slide 22

    Anritsu Master series OTDR

    Introduce the all- in-one concept

    Power Meter

    and Light Source

    High Performance OTDR

    MT9083 Series

    ACCESS Master

    Visible LD

    IP Test function

    Less equipment to carry and maintain!

  • 8/11/2019 Anritsu OTDRs

    12/45

    Slide 23

    Anritsu Master series OTDR

    Introducing the MT9083 seriesdesignedwith the features that matter most.

    Dual mode design (HR/ER mode) Compact

    Lightweight

    Fool-proof ease of use

    All day battery operation

    Quick boot-up

    High resolution, but also goodrange

    Quick testing time

    Reliable

    Rugged Complete testing tool

    Inexpensive

    Slide 24

    Anritsu Master series OTDR

    MT9083B/C: Dual Circuit Design Standard, high resolution (HR)

    Current operation

    All pulsewidths

    Use for best deadzone

    Enhanced Range (ER) Optimized for PON

    Optimized for range

    50ns-2us

    User selectable

  • 8/11/2019 Anritsu OTDRs

    13/45

    Slide 25

    Anritsu Master series OTDR

    MT9083B/C: FTTx/PON Measurement Using ER

    In ER mode, MT9083B/C allows

    user to measure entire span Up to a 64/128-branched splitter!

    Section A Section CSection

    B

    MT9083B/C can detect

    fault location beyond

    splitter very accurately.

    Slide 26

    Anritsu Master series OTDR

    Best Technical specs where it counts Small Dead Zone to allow

    measurement of near endconnectors. Within CO In riser cables Premise

    5cm Resolution Event plotted accurately Up to 150,001 sampling

    points

    dB Range also allows for over100km measurement range. Up to 45dB (1310/1550nm) Most Telco long haul links

    are from 70km to 100km.

  • 8/11/2019 Anritsu OTDRs

    14/45

    Slide 27

    Anritsu Master series OTDR

    Best Technical specs where it counts One touch Fault Locate mode,

    Automatically sets all the Opticalparameters after a single touchthe user has the results.

    Will offer a result with theminimum required averages.

    When in Standard OTDR mode, By pushing the Start (Green)

    button all optical parameters willbe configured.

    Lost Test Mode, Allow for simple interface to

    Power Meter and Light Sourcemeasurements.

    High power meter option of+30dBm

    Slide 28

    Anritsu Master series OTDR

    Other Technical specs Battery life

    8hours of use battery life.

    Boot-up time Ready to use in 15sec

    Engineers dont waste time on thejob.

    Easy file transfer Simple drag and drop file transfer via

    USB.

    Latest OTDR file format SR-4731 (issue 2).

    PC Emulation software available, Multiple version of PC software

    available included reporting softwaredownloadable from the web.

    System upgrade software available fromthe web. Download the latest software from the

    Anritsu web site.Your OTDR is never out of date.

  • 8/11/2019 Anritsu OTDRs

    15/45

    Slide 29

    Anritsu Master series OTDR

    High Performance, Field Tested Hardware

    Dedicated keys

    SetupsSave/recall

    Contrast

    Number pad

    Rotary encoder

    Precise marker placement

    Easy menu navigation

    START button

    True one-button fault

    location

    All parameters selected

    Perfect for novice users

    Dedicated softkeys

    Quick, easy operation

    ParametersWavelength

    Display modes

    Rugged Protector OptionSide bumpers

    Rigid display cover

    Shoulder strap

    Easy to read, TFT color

    displayStandard display or enhanced

    indoor/outdoor display

    LRUD keypad(left,

    right, up, down) for

    easy operation

    User Friendly

    Rugged Small

    Lightweight

    Easy to use

    Impressive Battery

    operation

    8 hour typical

    6 hour continuous

    LiIon no memory

    Slide 30

    Anritsu Master series OTDR

    Dedicated Testing Interfaces Clearly labeled Integrated dust caps A true all-in-one for optical fiber

    construction and maintenance

    OTDR ports

    Low cost, interchangeable

    adapters

    Built-in light source (std)

    Integrated power meter (std)

    SM/MM

    Optional power meters

    Interchangeable adapters

    For loop-back testing

    +3 to -70 dBm (SM/MM)

    +23 to -50 dBm (SM)

    +30 to 43dBm (SM)

    Visible laser diode

    option

    ~0 dBm output

    2.5mm universal

    ~5 km (3 mile) range

    CW or modulated

    Dual USB ports

    Plug & play

    mass storage

    IP testing Option

    RJ45 electrical

    10/100MB

    1000MB upgrade

    Ping, download test,

    throughput, frame loss

  • 8/11/2019 Anritsu OTDRs

    16/45

    Slide 31

    Anritsu Master series OTDR

    MT9083B/C: Waveform Comparison Using ER Mode

    MT9083B/C can capture

    waveform accurately using

    short pulse for entire fiberpath through PON splitter.

    Electrical noise

    Bad

    Bad

    Bad

    Good!

    Competitor

    A

    Competitor

    B

    Competitor

    C

    Slide 32

    Anritsu Master series OTDR

    MT9083B/C: Waveform Comparison Using ER Mode Competitor may be able to

    see waveform beyond

    splitter by using long

    pulse. However

    Zoom in

    Very sharp waveform even

    between close splitters

    OK

    ?

    Undefined waveformdue to poor dead zone

    performance

    Not

    OK

    OK! CompetitorA

    Competitor

    B

  • 8/11/2019 Anritsu OTDRs

    17/45

    Slide 33

    Anritsu Master series OTDR

    MT9083B/C: Longest Distance Using ER Mode

    MT9083B/C providessuperior range (100km) at100ns with excellent deadzone performance.

    100km

    100km

    100km

    100km

    (with 100ns

    Pulse)

    Competitor

    A

    Competitor

    B

    Competitor

    C

    Slide 34

    Anritsu Master series OTDR

    MT9083B/C: Best Dead Zone Using HR Mode

    There are faults our competitors can not see.

    MT9083B quickly identifies location of faults at the installationand reduces risk of future trouble.

    ANRITSU

    15m

    25m30m

    MT9083B/C is the only

    OTDR that can detect a

    fault in this zone.

    Competitor

    B

    Competitor

    C

  • 8/11/2019 Anritsu OTDRs

    18/45

    Slide 35

    Anritsu Master series OTDR

    Access Master MT9083[ ] Lineup

    Opt Wavelength MT9083A MT9083B MT9083C

    053 1310/1550 42/41 45/45

    055 1310/1550/1650 37.5/36/33.5 42/41/35

    056 1310/1490/1550 36/34.5/34.5 40/39/39

    057 1310/1550/1625 36/34.5/31.5 40/39/38 45/45/43

    058 1310/1490/1550/1625 38/37/37/36

    059 1310/1383/1550/1625 38/37/37/36

    063 M850/1300/S1310/1550 28/27/38/36.5 29/28/42/41

    064 MMF850/1300 28/27

    073 1310/1550 38/36.5

    A = 35dB class

    B = 40dB class

    C = 45dB class

    Slide 36

    Anritsu Master series OTDR

    Easy to use, task-oriented top menu

    Dedicated testing

    modes simplify

    operation

    1. Select the testing

    to be performed

    2. The GUI is then

    optimized to

    perform it

    3. Saves time,

    increases

    productivity

    Upon boot-up or whenthe Top Menu key ispushed, the Top Menuwill appear. A testingfunction to be performedcan be chosenimmediately.

  • 8/11/2019 Anritsu OTDRs

    19/45

    Slide 37

    Anritsu Master series OTDR

    Fault Locate Mode for novice users

    True one-button fault location

    Quick test for cable damage

    Perfect for CO fault locationReduces MTTR

    Ideal for inexperienced users

    Launch condition verified

    Test parameters automatically selected

    Quick data averageTypically less than 10 seconds

    Easy to read results for even the most

    inexperienced user

    Includes break/end location, span loss Full trace view if desired

    Slide 38

    Anritsu Master series OTDR

    Perform a Fault Locate test (1/2)1. Clean and connect fiber

    2. Highlight fault locate using the

    arrow keys & press enter

    3. Connection check immediately

    verifies connection

    1. Press continue to begin test if

    status bar is black

    2. Clean & reconnect if red

    3. User can not start test i f f iber is

    conn ected incorrect ly nochance of incorrect resul ts

    4. Automatically test sequence begins

    1. longest wavelength is selected

    2. testing parameters selected

    3. Fiber test begins

    1

    SM OTDRMM OTDR

    2

  • 8/11/2019 Anritsu OTDRs

    20/45

    Slide 39

    Anritsu Master series OTDR

    Perform a Fault Locate test (2/2)

    1. Results displayed in seconds1. End/fault location2. Event/loss3. Total span loss

    4. Full event table with end fault

    highlighted

    2. Select view trace for more

    information (if desired)1. Full event table2. Full trace

    3. Or connect next fiber and

    press

    Slide 40

    Anritsu Master series OTDR

    Standard OTDR Modefor any task

    Flexible operation for any skill level or task

    Customize the unit Units of measure File save format File save location Reflectance calculation

    Complete event table control Break/end threshold Loss specification Reflectance specification End to end loss

    Full functionality OTDR testing Multiple trace comparison Automated or manual parameter

    selectionyour choice Selectable display modes - from A, B,

    Origin or Anywhere (zoom box)

  • 8/11/2019 Anritsu OTDRs

    21/45

    Slide 41

    Anritsu Master series OTDR

    Run a quick test Press f1 and

    All parameters are automatically selected

    +

    =

    Complete analysislocation, type,

    loss, pass/fail in seconds!

    Slide 42

    Anritsu Master series OTDR

    Customize Test parameter(1/2) From the top menu, select

    Press f4 Select 1km (3.3kft), low density, 3 ns

    Press

    Press Averaging Time, enter 10 sec, then

  • 8/11/2019 Anritsu OTDRs

    22/45

    Slide 43

    Anritsu Master series OTDR

    Customize Test parameter (2/2) Press , then f1

    Slide 44

    Anritsu Master series OTDR

    Real Time Test Select 1550 or 1625 nm

    Select 2.5 km, low density, 10 ns & press

    Contract as shown using the arrow keys Comment on the screen update rate (0.15 sec)

    While still in real time, bend the fiber near the OTDR The trace will drop out

    Release the bend

    Trace will restore instantly

  • 8/11/2019 Anritsu OTDRs

    23/45

    Slide 45

    Anritsu Master series OTDR

    All fiber information is on a single screennoneed to flip through tabs Battery status (%

    and time remaining)Current

    testing mode

    Axis scaling

    A and B

    measurement

    cursors

    Softkeys

    Loss mode

    type and

    results

    Complete event

    table:

    Event #

    Location

    Type

    Loss

    Reflectance

    Segment dB/km

    Cumulative loss

    Key testing parameters:

    Wavelength

    Distance range

    PulsewidthIndex of refraction

    Resolution

    Number or averages

    Slide 46

    Anritsu Master series OTDR

    Interpreting Event Table

    1

    2

    3

  • 8/11/2019 Anritsu OTDRs

    24/45

    Slide 47

    Anritsu Master series OTDR

    Construction OTDR Mode for cable instal lat ion

    Automated testing of multiple fibers at multiple wavelength

    Enter the information oncesaves times

    Each fiber tested, named and saved automatically

    Once the single page set-up is complete, the software becomes

    the project manager

    Identifies which fiber to connect

    Verifies connection

    Ensures consistency in filename and parameters

    Eliminates user errors such as

    missing files

    Inconsistent naming Inconsistent parameter selection

    Slide 48

    Anritsu Master series OTDR

    Perform a Construction Mode Test (1/2)

    1. Select file storage location

    2. Determine base file order

    (example shown in filename

    field)

    3. Enter test site locations

    4. Use other field for cable

    ID, project, customer, etc.

    5. Select wavelengths

    6. Select testing options

    7. Enter number of fibers to

    test

    1 2 3 4

    5

    67

  • 8/11/2019 Anritsu OTDRs

    25/45

    Slide 49

    Anritsu Master series OTDR

    Perform a Construction Mode Test (2/2)

    Once set-up is complete, connect thefiber as instructed and press

    continue

    Connection level is verified

    Optimize and press continue if

    bar is black

    Clean and reconnect if red

    Sit back and watch up to four

    wavelengths tested, analyzed and

    savedwithout pressing a button

    Connect next fiber and presscontinue

    Slide 50

    Anritsu Master series OTDR

    Software Overview Set-ups always available from front panel 5

    Neatly arranged into categories

    Notes explain each setting

    Simply one-button

    software upgrades via

    USB!

  • 8/11/2019 Anritsu OTDRs

    26/45

    Slide 51

    Anritsu Master series OTDR

    Loss Test Set Modefor opt ical bud get veri f icat ion

    Stabilized light source standard More cost effective than buying stand-alone handheld

    sources

    CW, 1kHz or 2kHz output modes

    Universal connector easily adapts to all connector

    styles same port as the OTDR

    High (-8 dBm) output provides exceptional dynamic

    range for testing even the longest fibers (-8 dBm output) - (-65 dB sensitivity) = 57 dB of range

    Integrated power meter standard

    -5 to -50 dBm

    High power CATV range (+20 to55 dBm) optional

    Data table for easy storage of results

    Slide 52

    Anritsu Master series OTDR

    Perform Optical Loss Test Set (OLTS) function

    Select same wavelength for

    Power Meter and Light source

    Set Reference Connect patch cord between power

    meter and light source through an

    adapter.

    Turn on the light source and record

    the power in the Reference field

    Connect patch cord to cableunder test

    Measure loss

  • 8/11/2019 Anritsu OTDRs

    27/45

    Slide 53

    Anritsu Master series OTDR

    Reference Procedure for Loss Measurement

    Use two short patch cable

    for referencing

    Connect the short patch

    cable to both end of the fiber

    under test.

    Light Source Access Master in

    Power Meter Mode

    Slide 54

    Anritsu Master series OTDR

    USB connection Can be used on PCs with USB

    port

    Fiberscope for x200/400

    magnification

    Adaptors for 1.25 & 2.5mm

    patch cord.

    Universe connectors including

    APC

    Image saving function

    Video Inspection Probe (VIP)

  • 8/11/2019 Anritsu OTDRs

    28/45

    Slide 55

    Anritsu Master series OTDR

    Visible Fault Locator Option

    Integrated visible redlight fortroubleshooting within the OTDR dead

    zone

    Provides easy identification of a

    particular fiber

    Select output mode: CW or 2 Hz

    modulated

    Universal 2.5 mm connector fits all

    standard size connectors

    0dBm (1mW) max output

    ~5km (3 mile) range

    Slide 56

    Anritsu Master series OTDR

    Dedicated field test solution for installation and

    maintenance of optical links

    OTDR Tester.

  • 8/11/2019 Anritsu OTDRs

    29/45

    Slide 57

    Anritsu Master series OTDR

    Introducing the MT9090A with

    MU909014x/15x OTDR Module

    A truly revolutionary OTDR!

    Pocket-sized

    Easy Operation

    Excellent OTDR Performance

    Unique Battery Operation

    Priced Right

    Slide 58

    Anritsu Master series OTDR

    Overview. Product overview,

    Operation, Performance and Options,

    Specs,

    Comparison to MT9083x,

  • 8/11/2019 Anritsu OTDRs

    30/45

    Slide 59

    Anritsu Master series OTDR

    Value without compromise.

    OCA

    (Optical Channel

    Analyzer for CWDM)

    DCFL

    (Drop Cable Fault Locator)

    GiGE

    (Gigabit Ethernet test)

    Part of an ever-expanding platform!???

    Modular platform ensures

    maximum return on investment!

    OTDR

    Slide 60

    Anritsu Master series OTDR

    For Highly Portable Field Use. Easily held with one hand,

    Palm-size (190mm x 96mm x 48mm), Approximately 700g (1.54 lbs).

    Battery operation, Up to 8 hour operation

    (with STD Battery pack), 4 hour recharge

    (with STD Battery pack), NiMH battery pack or AA Alkaline Dry Battery

    operation.

    Rugged/sealed design, No vents or fans, Wrist strap to prevent accidental drops.

    Color indoor/outdoor display standard, 4.3 high resolution, Landscape format naturally fits OTDR trace.

  • 8/11/2019 Anritsu OTDRs

    31/45

    Slide 61

    Anritsu Master series OTDR

    The Foolproof testing device.

    Easy to use,

    True one-button fault location,

    Automatic testing parameters selection in

    Full-automatic mode (Full-manual

    operation also available),

    Quick fiber evaluation,

    Event analysis table and full Trace view.

    Quick boot-up,

    Ready to test in

  • 8/11/2019 Anritsu OTDRs

    32/45

    Slide 63

    Anritsu Master series OTDR

    Complete Optical Fiber Testing. Excellent OTDR Performance,

    Capable of,Full parameter selection (5ns-20us),125,000 data points,Up to 37dB Dynamic range,High resolution accuracy,Dual wave 1310/1550nm config,Complete testing from Access to Core.

    Complete PON testing tool,High resolution and extremely shortdead zones ensure entire PON (1X32, 1X64 branches)network evaluation,ITU-T L.41 compliant - 1625nm/1650nmwavelength for in-service maintenanceof PONs without external filters.

    Slide 64

    Anritsu Master series OTDR

    1:32 Splitter

    2km Fiber4km Fiber

    2.2km Fiber

    Complete PON Measurement for I&M.

  • 8/11/2019 Anritsu OTDRs

    33/45

    Slide 65

    Anritsu Master series OTDR

    Long Battery Life,

    Up to 8 hours / 4 hour recharge,

    with Standard NiMH Battery pack (Supplied with the original

    unit),

    Alkaline battery (AA)operation also available.

    Batteries.

    Available 4 x AA Type

    NiMH

    Or

    Alkaline Dry Batteries

    Slide 66

    Anritsu Master series OTDR

    Additional Highlights.

    Built-in 10m (30ft) launch fiber,

    Measure front-end connection

    without additional patchcords,

    Can be enabled or disabled.

    Complete data management,

    Up to 1,000 traces internal/10,000+ with USB,

    Easy drag and drop transfers to PC,

    Reporting and printing with NETWORKS or

    TraceView PC software.

  • 8/11/2019 Anritsu OTDRs

    34/45

    Slide 67

    Anritsu Master series OTDR

    Optional Features. Visual Fault Locator Option,

    Integrated 650nm laser,

    Can also be used to identify a

    particular fiber (in the splicing tray),

    Can be used simultaneously with power meter or OTDR,

    Up to 5km (3 mile) range.

    Connector Inspection Option,

    400x image displayed,

    All software pre-loaded,

    Eye-safe (Often an OH&S requirement),

    Image save/recall,

    Hardware includes probe/tips.

    Slide 68

    Anritsu Master series OTDR

    Different Applications. FTTx,

    Full network testing: ONT to OLT (incl. splitter(s)), In service maintenance (filters installed).

    General fiberinstallation/certification, Premise/CO, Access, Metro, Core/Long Haul.

    Backhaul, FTTc (fiber to the cell site), DAS, RRH.

    Optical Fiber(when used Ethernet)

    BaseTerminalStation :NodeB

    Antenna

    Optical

    Fiber

    BTS

    RRH

    RemoteRadioHead

    Optical Drop CableFault Locator,OTDR

    Optical Fiber Test

  • 8/11/2019 Anritsu OTDRs

    35/45

    Slide 69

    Anritsu Master series OTDR

    MU909015B/B1-056/066 MU909014B/B1-056/066 MU909014A/A1-053/063 MU909014A/A1-054/064

    1310/1550 20 nm * 1310/1550 20 nm * 1625 15 nm 1650 15 nm

    PW=20sec 37/36dB 32.5/31dB 32.5dB*(1490/1550 cut filter included)

    32.5dB*(1490/1550 cut filter included)

    PW=500nsec 28/26dB 24.5/23dB 24.5dB*(1490/1550 cut filter included)

    24dB*(1490/1550 cut filter included)

    Modulation: CW, 1Hz

    * Please see detailed on Product brochure

    Fiber Type 10m/125m SMF (ITU-T G.652)

    Loss Measurement mode 2 point loss, Splice loss, dB/km Loss LSA, ORL, Event

    Environment Operating temperature and humidity: -5 to +40C*10,

  • 8/11/2019 Anritsu OTDRs

    36/45

    Slide 71

    Anritsu Master series OTDR

    MU909011A Drop Cable Fault Locator Module

    Slide 72

    Anritsu Master series OTDR

    PURPOSE-BUILT FOR TESTING SHORT FIBERS

    Uniqueonly true short-fiber tester available

    Perfect solution for FTTX drop cables, CO,

    MDU risers, wireless towers & premise

    Integrated 10m launch cord standard

    Dynamic range of >7dB (10dB typ)

    For testing up to 10km (~6 miles)

    Unit will display transmission power levels,

    then fault locate with the press of a single

    button

    Testing parameters are fixed

    Operation simplified

    Ensures proper settings and valid results

    658 feet!

  • 8/11/2019 Anritsu OTDRs

    37/45

    Slide 73

    Anritsu Master series OTDR

    Additional Features Integrated power meter option for FTTx

    maintenance For measuring power to ONT Connect to MT9090A/MU909011A Verify power level and classify

    PASSfiber OK, ONT faultFAILfiber damaged, launch fault locate

    Power meter and fault locate from same port

    Complete breakdown as to failure and responsibilityONT, drop cable or feed fiber

    General purpose power meter -5 to -45dB range

    Slide 74

    Anritsu Master series OTDR

    74

    MT9090A SpecificationsMT9090A CONTROLLER SPECIFICATIONS

    Display 4.3 inches TFT Color LCD (480 x 272 pixels, transmissive)

    Interface USB (v1.1)

    Battery type NiMH

    Battery life time 3.5 hours (typ)

    Battery charging time

  • 8/11/2019 Anritsu OTDRs

    38/45

    Slide 75

    Anritsu Master series OTDR

    CMA5000aOTDRProduct Introduction

    Slide 76

    Anritsu Master series OTDR

    CMA5000a The all-in-one, portable CMA5000a supports network

    transmission and optical physical layer tests forinstallation and maintenance. Transmission

    SDH/SONETEthernet

    Optical physical layerFiber end testLoss test

    OTDRChromatic dispersionPolarization mode dispersionSpectrum analysis

  • 8/11/2019 Anritsu OTDRs

    39/45

    Slide 77

    Anritsu Master series OTDR

    Modular Structure One mainframe supports full range of versatile

    measurements by changing modules

    Transmission systems (modules)SDH/SONET (UTA, XTA)

    Ethernet (UTA, GigE II)

    Optical physical layer (modules)Fiber end face check (VIP)

    Loss test (OTDR)

    Optical Time Domain

    Reflectometer (OTDR)

    Chromatic dispersion

    (CD-OTDR)Polarization mode dispersion (PMD)

    Spectrum analyzer (OSA)

    Slide 78

    Anritsu Master series OTDR

    Mainframe Types and Modules Two mainframes models allowing greater flexibility

    SBA: Two bays up for up two modules

    MBA: Four bays for up to four modules

    Multiple measurement modules can be

    set and operated simultaneously in

    one mainframe.

    1 bay used

    GigE II

    OTDR

    CD-OTDR

    SBA

    2 bays used

    UTA

    XTA

    Bays emptyUSB Connection

    VIP

    MBA

    4 bays used

    OSA

    PMD

  • 8/11/2019 Anritsu OTDRs

    40/45

    Slide 79

    Anritsu Master series OTDR

    CMA 5000a OTDR Avai lable Modu les

    Model

    Number

    Wavelengths Dynamic Range

    5261 QUAD SM/MM:

    850/1300nm (62.5um)/1310/1550nm

    24/26dB

    35/35dB

    5262 QUAD SM/MM:

    850/1300nm (62.5um)/1310/1550nm

    24/26dB

    40/40dB

    5266 850/1300nm62.5um 24/26dB

    5269 850/1300nm50um 24/26dB

    5283 1310/1383/1550/1625nm 35dB

    5491 1310/1490/1550nm 40/40/40dB

    5493 1310/1490/1550/1625nm 37dB

    Model

    Number

    Wavelengths Dynamic Range

    5225 1310/1550nm 37/36dB

    5235 1310/1550nm 40/40dB

    5236 1310/1550/1625nm 40/40/40dB

    5245 1310/1550nm 43/45dB

    5246 1310/1550/1625nm 43/45/43dB

    5254 1550nm 50dB

    5260 QUAD SM/MM:850/1300nm (50um)

    /1310/1550nm

    24/26dB35/35dB

    Slide 80

    Anritsu Master series OTDR

    Key Features and B enefi ts

    Unparalleled data sampling - Up to

    256,000 data points

    Twice as many as the nearest

    competitor!

    Unparalleled waveform creation

    Ultra-fast real time screen update - 0.1

    sec

    Great for optimizing

    splices/connectors

    Gives perception of fast testing

    Fully configurable to your customers

    needs

    Over 20 different optics choices

    single mode or multimode, 25 to

    50dB, dual, tri- or quad wavelengths

    State of the art platform

    Powerful PC, all common I/O

    Active productcontinued

    development

    Designed for simplicity and ease of use

    Dedicated testing modesno macros

    One button testing (AUTO-TEST)

    Minimal set-up time, maximum testing time

    A strength for all skill levels

    Superior trace linearity0.04 dB/dB

    Provides greater measurement accuracy

    and consistency

    Overlay up to 8 traces simultaneous

    Compare between wavelengths to seestresses effecting the cable

    Compare between fibers in a cable to see

    inconsistencies

  • 8/11/2019 Anritsu OTDRs

    41/45

    Slide 81

    Anritsu Master series OTDR

    Test ing Modes

    Fault Locate One Button

    Automated Testing Standard

    Traditional OTDRTesting

    Construction Automated Multi-

    Wavelength Multi-Fiber Testing

    Loss Test Set Integrated Power

    Meter and LightSource

    Visual Fault Locator Visible Red Light

    Video Inspection Probe Connector

    inspectionmicroscope

    Slide 82

    Anritsu Master series OTDR

    Fault Locate Mode One Button Au tomated Test ing

    One-button break/end locator

    Quick test for cable damage

    Ideal testing mode forinexperienced users Launch condition verified Test parameters automatically

    selected Data averaged just long enough

    to produce valid results Easy to read results for even the

    most inexperienced user Includes break/end location,

    span loss and distance pastprior event

  • 8/11/2019 Anritsu OTDRs

    42/45

    Slide 83

    Anritsu Master series OTDR

    Standard OTDR Mode Tradi t ional OTDR Testing

    Full functionality OTDR testing Multiple trace comparison Automated or manual parameter

    selectionyour choice Selectable display modes - from A, B,

    Origin or Anywhere (zoom box)

    Customize the unit Units of measure File save format File save location Reflectance calculation

    Complete event table control Break/end threshold Loss specification Reflectance specification

    End to end loss

    Slide 84

    Anritsu Master series OTDR

    Construct OTDR ModeAu tom ated Mult i -Wavelength Testing

    Automated testing of multiple fiber

    cables

    Test multiple fibers at multiple

    wavelength quickly and easily

    Enter the information once and

    test/save each fiber with a single key

    press

    Ensures consistency in filename andparameters

    Combines the best of CMA 4000

    Construction and Multiple

    Wavelength modes

  • 8/11/2019 Anritsu OTDRs

    43/45

    Slide 85

    Anritsu Master series OTDR

    Loss Test Set ModeOptional Power Meter and Light Source

    Adds a stabilized mode to the OTDRlaser for point to point testing More cost effective than buying

    stand-alone handheld sources

    CW, 1kHz or 2kHz output modes

    Universal connector easily adapts toall connector styles Uses the same UC-130-XX

    adapters as the OTDR

    High (-8 dBm) output providesexceptional dynamic range fortesting even the longest fibers (8 dBm output) (65 dB

    sensitivity) = 57 dB of range

    High power CATV range +20 to 55 dBm Data table for easy storage of results

    Slide 86

    Anritsu Master series OTDR

    Visual Fault Locator ModeOptional Red Light Test

    Visible redlight fortroubleshooting withinthe OTDR dead zone

    Provides easyidentification of aparticular fiber

    Select output mode: CW

    or 2 KHz modulated

    Universal 2.5 mmconnector fits allstandard sizeconnectors

    0dBm (1mW) max output ~5km (3 mile) range

  • 8/11/2019 Anritsu OTDRs

    44/45

    Slide 87

    Anritsu Master series OTDR

    NetWorks Data Processing UtilityThe ult imate data proc essing to ol

    Integrated software utility for viewing,printing and analyzing OTDR data

    Analyze data on any PC

    Complete report generating tool

    Uni- and Bi-directional splice loss

    Exception

    Fiber Acceptance

    Trace Summary

    Batch processing for high count fiber

    cables Full analysis and Bellcore/Telcordia file

    format processing

    Anritsu Master series OTDR

    Slide 88

    Anritsu Master series OTDR

    The New AnritsuWhy Choose Us?

    Now includes Nettest

    The portable OTDR market share leader

    Over 30 years in the optical industry

    Over 25 years OTDR experience

    Stable and debt-free: your partner for thelong haul

    Focused on core competencynot ajack of all trades

    Solutions and support that enable you tobe more competitive and profitable

    Superior sales and technical support:anytime, anywhere

    Anritsu Master series OTDR

  • 8/11/2019 Anritsu OTDRs

    45/45

    Slide 89

    Anritsu Master series OTDR

    The right productplus the support yo u deserve

    Easy and always FREE software upgrades Simply, one time registration

    https://www1.anritsu.co.jp/Download/MService/Login.asp

    Dedicated, knowledgeable technical support Via our worldwide network.Email : [email protected]

    Local service centers Via our worldwide network.Email : [email protected]

    Anritsu Master series OTDR

    Anritsu Master series OTDRAnritsu Master series OTDR

    https://www1.anritsu.co.jp/Download/MService/Login.aspmailto:[email protected]:[email protected]:[email protected]:[email protected]://www1.anritsu.co.jp/Download/MService/Login.asp