Top Banner
Anger Camera for neutrons F. A. F. Fraga, L. M. S. Margato ,S. T. G. Fetal, M. M. F. R. Fraga, A. Morozov, L. Pereira, R. Ferreira Marques, P. Mendes, J. P. Rodrigues LIP Coimbra, Universidade de Coimbra
32

Anger Camera for neutrons

Mar 18, 2016

Download

Documents

Anger Camera for neutrons. F. A. F. Fraga, L. M. S. Margato ,S. T. G. Fetal, M. M. F. R. Fraga, A. Morozov, L. Pereira, R. Ferreira Marques, P. Mendes, J. P. Rodrigues LIP Coimbra, Universidade de Coimbra. FP6 NMI3 JRA2 RII3-CT-2003-505925 ended 2008. - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Anger Camera for neutrons

Anger Camera for neutrons

F. A. F. Fraga, L. M. S. Margato ,S. T. G. Fetal, M. M. F. R. Fraga, A. Morozov, L. Pereira, R. Ferreira Marques, P. Mendes, J. P. Rodrigues

LIP Coimbra, Universidade de Coimbra

Page 2: Anger Camera for neutrons

FP6 NMI3 JRA2 RII3-CT-2003-505925 ended 2008MILAND – development of thermal neutron detectors•Would increase knowledge about neutron detectors, but should have a deliverable working detector prototype with a firm and clear specification

• Partners and Observers• P1 CCLRC N. Rhodes (UK) T3, T6, T7• P2 GKSS R. Kampmann (Germany) T1, T4, T5, T6, T7• P3 BNC L. Rosta (Hungary) T1, T5, T6, T7• P4 ILL B. Guerard (France) T1, T2, T3, T4, T5, T6, T7 coordinator• P5 LLB C. Fermon (France) T2, T4, T6, T7• P6 FRM-II K. Zeitelhack (Germany) T1, T2, T4, T6, T7• P7 LIP F. Fraga (Portugal) T2, T3, T6, T7

ObserversEFO1 BNL G. Smith (USA)

• EFO2 SNSR. Cooper (USA) • EFO3 TU H. Takahashi (Japan) • EFO4 RAL J. Mir (UK)• EFO5 BNL R. Krueger TU Delft (NL)

•32 x 32 cm2 detector•1 mm resolution (FWHM) res/length = 3x10-4

•1 MHz global count rate•100 kHz / mm2 local (peak) count rate•Gamma background rejection 10-8

•50% efficient at 1.8Å•parallax free•Cost is critical

Integrated Infrastructure Initiative for Neutron Scattering and Muon Spectroscopy

Page 3: Anger Camera for neutrons

Neutron Detectors nuclear reactions to convert neutrons

• n + 3He 3H + 1H + 0.764 MeV Gaseous detectors (CF4, prop.)

• n + 6Li 4He + 3H + 4.79 MeV Scintillators• n + 10B 7Li* + 4He7Li + 4He + 0.48 MeV +2.3 MeV

(93%) 7Li + 4He +2.8 MeV ( 7%)

• n + 155Gd Gd* -ray spectrum conversion electron spectrum • n + 157Gd Gd* -ray spectrum conversion electron spectrum • n + 235U fission fragments + ~160 MeV • n + 239Pu fission fragments + ~160 MeV

Page 4: Anger Camera for neutrons
Page 5: Anger Camera for neutrons

Scintillators for Neutron Detectors

CWE van Eijk, NIM A529(2004)260-267

Page 6: Anger Camera for neutrons

Gaseous detectors

n + 3He 3H + 1H + 0.764 MeV

CF4 should be added to control the range -3 bar for 1mm FWHM

• CF4 is a good scintillator, but only a few primary photons• Secondary scintillation ~.3 photon per secondary electron

200 300 400 500 600 700 800 9000

200

400

600

800

I norm (a

.u.)

(nm)

CF3

*

CF4+*, CF3

+*

Page 7: Anger Camera for neutrons

PMT 2

10mm

35m

m

Boron Mask

Neutron Beam

Quartz Window

anodes

PMT 3

AQUISISITION CARDS: Acqiris, CC103

PMT 1 PMT 4

PC

trigger (from anodes or cathodes)

Drift electrode

n

T

p

Page 8: Anger Camera for neutrons

Scintillation Anger camera

PMT

Transparent window

MSGC

PMT base

gasNeutron window

PMT

PMT base

Page 9: Anger Camera for neutrons

New MSGC Batch

NMI3 Meeting, CORSICA, 25-28th June 2008

Remarks• Cathodes width of the new set of microstrips is more or less half of the

older ones (181m vs. 398 m) and the anodes seems to be larger.

• To achieve the same charge gain the new microstrip needed a higher anode voltage than the older one.

• For instance, at 1 bar of CF4, to reach a gain of G ~103 we should apply a voltage of Va~1900V, something like 840V more than in the older microstrip (Va=1060V).

• Taking off the drawbacks linked to a higher voltage, the light yield obtained with the new microstrip (for G ~103 ) is ~ 0.263 ph/sec-e, about 4.5 times higher than the light yield in the older one (~0.045 ph/sec-e).

• This MSGCs batch are unable to operate above 1 bar CF4 (discharges).

Page 10: Anger Camera for neutrons

Charge gain and light/charge versus anode voltage 15, 20 and 30µm anodes

CF4 (1bar)

CF 4 (1bar)

1.0E+01

1.0E+02

1.0E+03

1.0E+04

500 700 900 1100 1300 1500Va (V)

Gai

n

15 micron anodes20 micron anodes30 micron anodes

CF 4 (1bar)

0

5

10

15

20

500 700 900 1100 1300 1500Va (V)

S PM

T / Q

anod

e

15 micron anodes20 micron anodes30 micron anodes

Page 11: Anger Camera for neutrons

New MSGC Batch

NMI3 Meeting, CORSICA, 25-28th June 2008

Charge and light measurements

Charge gain versus anode voltage Number of emitted photons per

secondary electron

CF 4 (1bar); V d =-700V

1.0E+01

1.0E+02

1.0E+03

1.0E+04

800 1000 1200 1400 1600 1800 2000V anode (V)

Gai

n

15 micronNew Microstrip - April 2008

CF 4 (1bar)

0.00

0.10

0.20

0.30

0.40

0.50

1.0E+02 1.0E+03 1.0E+04Gain

Emitt

ed P

hoto

ns /

s.e.

15 micronNew Microstrip - April 2008

Page 12: Anger Camera for neutrons

MSGC ILL6C at High pressure

NMI3 Meeting, CORSICA, 25-28th June 2008

Measurements at 3 bar CF4

CF 4 (3bar)

0.00

0.10

0.20

0.30

0.40

0.50

1.0E+02 1.0E+03 1.0E+04Gain

Emitt

ed P

hoto

ns /

s.e

ILL6C - Nº1ILL6C - Nº2LLB - 15 micron

CF4 -3bar

1.0E+02

1.0E+03

1.0E+04

1400 1600 1800 2000V anode (V)

Gai

n

ILL6C - Nº2ILL6C - Nº3LLB - 15 micron

Charge gain versus anode voltage Number of emitted photons per

secondary electron

Page 13: Anger Camera for neutrons

Simulations F.Fraga, Hamburg meeting

Page 14: Anger Camera for neutrons

GSPC details

Page 15: Anger Camera for neutrons

GSPC in position at the ILL beamAquisition system by Acquiris and E. Shooneveld

Page 16: Anger Camera for neutrons

2D position images Vanode = 1360, 1550 1650 and 1850V

Page 17: Anger Camera for neutrons

Mask_1650_3.eve: Va=1650 V; VPMT= 800 V.

fwhm (yy)~1.63 mm

fwhm (xx)~1.68 mm

Page 18: Anger Camera for neutrons

Mask_1650_3.eve: Va=1650 V; VPMT= 800 V

FHWM center peak 1.68 mm

FHWM center peak 1.63 mm

Page 19: Anger Camera for neutrons

FWHM resolution versus Vanode

0.00

0.50

1.00

1.50

2.00

2.50

3.00

1300 1500 1700 1900

Vanode (V)

peak

FW

HM

(mm

)

(XX)(YY)

•For the best safe conditions (Vanode =1850 V; VPMT= 800 V) the intrinsic resolution is 1.1mm after we deconvolute the beam width

Page 20: Anger Camera for neutrons

Detector uniformity (1850V)

Page 21: Anger Camera for neutrons

Typical PMT signals

High amplitude – good discrimination

Very fast signals – risetime ~20ns

Page 22: Anger Camera for neutrons
Page 23: Anger Camera for neutrons

Transparent electrode MSGC

• Manufactured at the Tokyo University and supplied by Hiroyuki Takahashi

• 90% ITO electrode transparency

• The transparent window can be the active element

• Multigrid approach proposed –it has the advantage of a very high local count rate.

Page 24: Anger Camera for neutrons

ITO M - MSGC

NMI3 Meeting, CORSICA, 25-28th June 2008

Photographs

Page 25: Anger Camera for neutrons

ITO M - MSGC

NMI3 Meeting, CORSICA, 25-28th June 2008

Experimental Setup

Drift length = 10 mm

Distance between microstrip and PMT = 34.8mm

Page 26: Anger Camera for neutrons

ITO M - MSGC

NMI3 Meeting, CORSICA, 25-28th June 2008

Ar+5%CF4 (1atm, 100cc/min.) and CF4 (1atm, 100cc/min.)

ITO M - MSGC

1.0E+02

1.0E+03

1.0E+04

1.0E+05

500 700 900 1100V anode (V)

Gai

n

Ar+5%CF4 (Vg1=400V)

100%CF4 (Vg1=600V)

ITO - Microstrip

0.00

0.10

0.20

0.30

0.40

1.0E+02 1.0E+03 1.0E+04Gain

Nº o

f em

itted

pho

tons

/ s.e

Ar+5%CF4 (Vg1=400V)

100%CF4 (Vg1=600V)

Charge gain versus anode voltage Number of emitted photons per

secondary electron

Page 27: Anger Camera for neutrons

ITO M - MSGC

NMI3 Meeting, CORSICA, 25-28th June 2008

Effect of GRID potential (Vg) on Gain and light yield

ITO M - MSGC: 100%CF 4 , Vd=-500V

1.0E+02

1.0E+03

1.0E+04

1.0E+05

600 800 1000 1200V anode (V)

Gai

n

Vg1=200VVg1=400VVg1=600Vvg1=800V

ITO - Microstrip: 100%CF 4 , Vd=-500V

0.00

0.10

0.20

0.30

0.40

1.0E+02 1.0E+03 1.0E+04 1.0E+05Gain

Nº o

f em

itted

pho

tons

/ s.e Vg1=200V

Vg1=400VVg1=600VVg1=800V

Charge gain versus anode voltage for several GRID voltages

Number of emitted photons per secondary electron

Page 28: Anger Camera for neutrons

ITO M - MSGC

NMI3 Meeting, CORSICA, 25-28th June 2008

Example of typical signal taken directly from the PMT Hamamatsu R1387 (RL=50 ): 55Fe (5.9keV) X-ray source

•100%CF4 (1atm)

•Vdrift= -500V, Vcathode=0

•Vanode=990V Vg1= 600V

•VPMT=-1000V

60ns

100 mV

Page 29: Anger Camera for neutrons

MSGC ILL6C at High pressure

NMI3 Meeting, CORSICA, 25-28th June 2008

Measurements at 3, 4 and 5 bar CF4

CF4 - 3, 4 & 5 bar

1.0E+02

1.0E+03

1.0E+04

1400 1900 2400 2900V anode (V)

Gai

n

LLB - 15mic.-3 bar ILL6C - 3 barILL6C - 4 barILL6C - 5 bar

CF 4 - 3, 4 & 5 bar

0.00

0.10

0.20

0.30

0.40

0.50

1.0E+02 1.0E+03 1.0E+04Gain

Emitt

ed P

hoto

ns /

s.e

LLB - 15mic. - 3barILL6C - 3 barILL6C - 4 barILL6C - 5 bar

Charge gain versus anode voltage Number of emitted photons per

secondary electron

Page 30: Anger Camera for neutrons

2009-2011 FP7 WP22CP-CSA_INFRA-2008-1.1.1 Number 226507-NMI3

• Development of new thermal neutron detector technologies based on Gaseous Scintillation Proportional Counters (GSPC) • Explore their potential to overcome the limitations in light output and rate capability of existing scintillation detectors • Investigate their potential as high resolution detectors for Reflectometry or time resolved SANS. • Built and study small scalable prototypes

http://detectors.neutron-eu.net/Detectors

Page 31: Anger Camera for neutrons

The new family members, born in Munich two months ago!

We will go on studying the scintillation and the PMT readout

Page 32: Anger Camera for neutrons

The good not so good new

• European Spallation Source (ESS) went to Lund in Sweden

The bad new

• 3He crisis