Page 1
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Spine Publish Ahead of Print
DOI: 10.1097/BRS.0000000000000660
Analysis of retrieved growth guidance sliding LSZ-4D devices for early
onset scoliosis and investigation of the use of Nitinol rods for this system
Elena Lukina1,2
, MSc, Mikhail Kollerov2, PhD, Jay Meswania
3, PhD, David Wertheim
1, PhD,
Peter Mason1, PhD, Paul Wagstaff
1, MSc, Aleksandr Laka
4, PhD, Hilali Noordeen FRCS (Tr
& Orth)5, Wai Weng Yoon FRCS (Tr & Orth)
5, Gordon Blunn
3 PhD
1 – Kingston University London, UK
[email protected] and [email protected] 104 Roehampton Vale, SW15 3RX
2 – “MATI”-RSTU, Russia
3 – University College London, UK
4 – Russian University of Peoples’ Friendship, Moscow, Russia
5 – Royal National Orthopedics Hospital, UK
Corresponding author:
Elena Lukina
e-mail: [email protected] [email protected]
Tel. +44(0)7895545709 and +79162266225
Address for correspondence and reprints:
4 GREENWOOD ROAD
KT70DY
THAMES DITTON
UK
Page 2
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
The device(s)/drug(s) that is/are the subject of this manuscript is/are not FDA-approved for
this indication and is/are not commercially available in the United States. No funds were
received in support of this work. Relevant financial activities outside the submitted work:
board membership, consultancy, payment for development of educational presentations,
employment, grants, payment for lectures, stocks, travel/accommodations/meeting expenses.
Study Design. Analysis of volumetric wear loss of retrieved growth guidance sliding devices
LSZ-4D for treatment of early onset scoliosis and laboratory in-vitro wear test for
comparison of Nitinol, Ti and CoCr alloys wear resistance.
Objective. To evaluate quantitatively the amount of wear debris from the sliding LSZ-4D
device and to investigate the potential of using Nitinol for replacing Ti alloys in spinal
instrumentation. In order to do that, wear resistance of Nitinol, Ti and CoCr was compared.
Summary of Background Data. There is little data regarding the amount of wear debris
associated with growth guidance sliding devices for patients with early onset scoliosis and the
wear resistance of superelastic Nitinol compared with Ti and CoCr.
Methods. Volumetric wear loss was measured on LSZ-4D devices made from titanium alloy
Ti6Al4V and each consisted of 2 rectangular section (6x4 mm) rods and 40±8 fixture
elements (20±4 hooks and 20±4 clips) retrieved from 3 patients (implantation period 3.5-5.8
years). Images of wear scars were taken on Bruker interferometer microscope and
incorporated into MATLAB software. Wear resistance of Nitinol, Ti and CoCr was studied
using reciprocation pin-on-disk wear test in bovine serum at 37±1OC.
Results. The volume wear rate of LSZ-4D device was found to be 12.5 mm3 per year from
which 5mm3 per year is the rods wear debris and 7.5 mm
3 per year is the contribution of
fixtures. Wear resistance of Nitinol is 100 times higher compared to Ti and comparable to
that of CoCr.
Page 3
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Conclusions. Application of wear resistant coatings on Ti components in growth guidance
sliding devices for the treatment of early onset scoliosis will be useful. High wear resistance
of Nitinol combined with its superelastic and shape memory properties could make
application of Nitinol rods for spinal instrumentation beneficial.
Key words: early onset scoliosis, wear, retrievals, growth guidance system, Nitinol, Implant
Level of Evidence: 5
Volumetric wear loss of retrieved growth guidance sliding devices LSZ-4D for early onset
scoliosis treatment made from Ti6Al4V was measured to be 12.5 mm3 per year. A laboratory
study using pin-on-disk wear test revealed wear resistance of Nitinol to be 100 times superior
to that of Ti and comparable to CoCr.
Key points:
The volume wear rate measured for the retrieved LSZ-4D growth guidance device
made of titanium alloy Ti6Al4V was found to be 12.5 mm3 per year from which 5mm
3 per
year is the rods wear debris and 7.5 mm3 per year is the contribution of fixtures;
In-vitro pin-on-disk wear test study have revealed that wear resistance of Nitinol
tested against titanium alloy Ti6Al4V in the simulated body environment is 100 times higher
compared to titanium alloy Ti6Al4V and comparable to that of Cobalt Chromium alloy;
Application of wear resistant coatings on Ti components in growth guidance sliding
devices for treatment of early onset scoliosis will be useful. High wear resistance of Nitinol
Page 4
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
combined with its superelastic and shape memory properties could make application of
Nitinol rods for spinal instrumentation beneficial.
INTRODUCTION
Surgery for early onset scoliosis (EOS) requires fusion-less instrumentation as early fusion
may result in complications such as trunk shortening, pulmonary dysfunction and overloading
of the spine [1]. There are number of fusion-less instrumentation for EOS treatment that is
commonly used now. These include vertical expandable prosthetic titanium rib (VEPTR)
instrumentation [2], non-invasive growth guidance devices including magnetically extendable
growing rods [3] or sliding systems Shilla or LSZ-4D instrumentation [4-5] and invasively
(mechanically) extendable growing rods [6]. However, due to the absence of fusion
movement of rods against fixtures is possible and may result in undesirable wear debris
formation. This process is likely to be more pronounced for growth guidance sliding devices
where rods are able to slide in unlocked fixtures as the child is growing.
While absolute amounts of wear debris produced by joint prostheses such as total hip, knee
and spinal disk replacements (THP, TKP, TDP), which have a high range of motion, are
extensively covered in literature [7-8], there is limited information on quantitative values of
volumetric wear of spinal implants used for EOS.
Titanium alloys (Ti) are biocompatible; however they have lower wear resistance compared
to cobalt chromium (CoCr) and stainless steel (SS) [9].
Wear debris generated during articulation of total metal-on-metal hip prosthesis made of
CoCr is one of the main reasons for the failure of these implants [10]. Release of CoCr wear
debris and ions leads to tissue necrosis, pseudotumors and hypersensivity [11-12]. Polymer-
on-metal hip implants which include CoCr and polyethylene components also suffer from
aseptic loosening and osteolysis as a result of polyethylene wear debris [13]. Attempt to
Page 5
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
introduce knee implants with titanium on polyethylene friction pair was made in late 1980s.
However, these prostheses were withdrawn from the market because of the problems
associated with excessive Ti wear [14].
Prominent wear scars on Ti and SS spinal scoliosis instrumentation were reported by
Villarraga et al. [15]. Wang et al. revealed high concentrations of Ti in tissues collected from
areas near pedicle screw-rod junction [16]. In addition, titanium wear debris produced by
spinal implants has been reported to cause inflammation and osteolysis [17-19]. In contrast,
Singh et al. [20] estimated the wear loss of the Shilla growth guidance device (Medtronic,
USA) made of SS is less than 1 mm3 per year.
Nitinol, an alloy with Shape Memory effect could be another promising material for rods in
scoliosis instrumentation [21-24]. It has an ability to return to it’s preconfigured shape at
body temperature and has been used in scoliosis correction. It has the potential benefit of a
more gradual and sustained correction [24]. However, wear resistance of Nitinol in biological
environment is poorly understood, especially in combination with Ti which is the main
material used for fixtures.
The purpose of current work was to measure volume wear loss of retrieved LSZ-4D devices
made of titanium alloy Ti6Al4V. The second purpose was to investigate the potential of using
Nitinol for replacing Ti alloys in spinal instrumentation. In order to do that, we used a pin-on-
disk test to compare wear resistance of Nitinol, Ti and CoCr.
MATERIALS AND METHODS
Description of retrieved growth guidance sliding LSZ instrumentation and patient’s
sample
Growth guidance sliding LSZ-4D device (Conmet, Russia) for EOS scoliosis treatment was
retrieved from 3 patients (1 male, 2 female) undergoing routine final definitive fusion surgery
following skeletal maturity. The LSZ-4D sliding device consists of two rectangular cross-
Page 6
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
section 6x4 mm rods and fixture units. Locked fixtures are normally located on one spinal
level while unlocked fixtures are used at distal and proximal ends of the device thus enabling
sliding while spine is growing (fig.1A). Each fixture unit contains a clip and a hook (fig.1B).
All components are made of titanium alloy Ti6Al4V.
In total six rods and sixty fixture units (60 hooks and 60 clips) were analyzed from 3 patients.
The components were implanted for periods for between 3.5 and 5.8 years (average: 4.3
years). None of these patients had any clinical complications. Clinical information for all
retrieved devices is given in Table 1.
Volume wear loss of LSZ-4D sliding device components was measured by taking images of
wear scars using a Bruker interferometer microscope. The interferometry data was then
incorporated into MATLAB software (The MathWorks Inc., Natick, MA, USA). The system
then calculates the displaced volume in parallel planes with 1 pixel spacing lines along the
length of the groove. The median height over 20 pixels is calculated at the end of each line.
Using linear interpolation, the volumes under the lines are calculated and summed over the
length of the groove. In case one edge of the groove was worn out, the system also allowed
the volume to be calculated with respect to the mean of a region; in this mode the user could
select the top surface and the displaced volume beneath is computed taking into account the
tilting of the image. Following processing with MATLAB, 3D visualizations were performed
using Amira v5.4 (Visualization Sciences Group, Germany).
Photos of wear grooves on rods (fig. 2A), clips (fig. 2C) and hooks (fig. 2E) and their
corresponding interferometry images reconstructed in MATLAB followed by 3D
visualizations (fig. 2B, D and F respectively) are presented in figure 2.
This volumetric wear loss measurement was validated in the in-vitro test where the wear of
test components were measured volumetrically using the method described above and related
Page 7
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
to the loss of material measured gravimetrically. The difference in measurements was less
than 5%.
Scanning electron microscopy (SEM) was used to examine the wear scars on these devices
(EVO 50, Carl ZEISS).
Wear particles were extracted from tissues adjacent to the implanted LSZ-4D device in the
lumbar part of the spine during routine final definitive fusion surgery. Particles were digested
in papain and proteinase K according to ISO 17853:11 [26]. The particles size and shape was
determined by analyzing SEM images taken at 30,000 magnification using ImagePro3
software (Nexis, Russia). Chemical composition of particles was analysed using energy
dispersive x-ray analysis (EDAX).
In-vitro pin-on-disk wear test
In-vitro wear tests were conducted using a pin-on-disk set up (fig.3A and B) according to
ASTM G99-05(2010) [27] at 37±1OC with 1 Hz frequency of reciprocating sliding
movement. Diluted bovine serum was used as a lubricant according ISO 18192-1:2011 [28].
The load on the pins was 10N which provided contact stresses of approximately 3 MPa which
was the calculated stress level based on literature data on bending moments encountered by
spinal rods during various functional activities [29]. The amplitude of reciprocation
movement was 6 mm.
Pins were made of titanium alloy Ti-5.8wt.% Al-3.9wt.%V (Ti6Al4V); Ti-55.8 wt.%Ni
(Nitinol), and Co-27wt.%Cr-5wt.%Mo, low carbon (CoCr alloy). All disks were made of
titanium alloy Ti6Al4V.
The pins were 5.5 mm in diameter, and were rounded to 20 mm radius. The discs measured
40 mm in diameter and 5mm in height. The temperature of shape recovery of Nitinol
measured according ASTM 2082-06 [30] was Af=37±1OC. Surfaces of the pins and discs
were ground and polished to the roughness of Ra 0.04±0.01µm. Volume wear loss (∆V) of
Page 8
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
both counter-parts separately was determined after 0.05, 0.2, 0.5 and 0.85 million cycles as
∆V=∆m/ρ, where ∆m – weight loss gravimetrically measured, g and ρ – density of the
material, g/mm3. Number of tests for each condition n=3. Wear scars were analyzed using
SEM. Wear particles were isolated from the bovine serum according to ISO 17853:11 [26]
and analysed using SEM and EDAX.
The Mann-Whitney U test was used to determine if there was statistically significant
difference in the size of wear particles digested from tissues adjacent to the implanted LSZ-
4D device and those digested from the bovine serum after in-vitro wear tests. A p value of
less than 0.05 was defined as statistically significant. This test was run after Kolmogorov-
Smirnov test revealed that the wear particle’s size significantly deviated from a normal
distribution. Statistical analysis was performed with SPSS 22.0 software (IBM Corp., USA).
RESULTS
Analysis of retrievals
Wear
The whole assemble of LSZ-4D sliding device analyzed from each of three patients included
2 rods and 40±8 fixture elements (20±4 hooks and 20±4 clips).
Analysis of wear regions on the retrieved rods revealed larger scars on distal and proximal
parts of rods and minor wear in the central region near the spine level with locked screw. The
average width of scars was 10±2 mm.
Wear of the hooks and clips was also observed to be uneven. Only 17 of 60 examined clips
and 18 of 60 examined hooks were severely damaged with deep grooves while the rest had
minor scratches with no measurable volume wear. Severely damaged fixture elements were
located on distal or proximal part of rods. None of the patient’s devices was seen to be any
Page 9
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
more or less worn than the others. The examples of most heavily damaged clip, hook and
distal part of rods are shown on Figure 2.
The average volume wear rate for the whole LSZ-4D sliding device assuming a linear wear
rate was 12.5±1.5mm3 per year with individual contributions from rods, hooks and clips
being 5±1.2mm3, 3±1mm
3 and
4.5±1.5mm
3 per year respectively (fig.4).
Analysis of wear scars and particles
SEM analysis of wear scars on retrieved rods, hooks and clips of LSZ-4D device revealed
deep abrasive grooves and adhesion deposits of titanium (fig. 5A) indicating abrasive and
adhesion mechanisms of Ti6Al4V alloy wear damage.
The average size of wear particles (fig. 5B) digested from tissues surrounding implanted
LSZ-4D device was 0.46µm (range from 0.09 to 1.59µm). Average aspect ratio was 1.2-1.5
and they were classified as round to oval according to ISO 17853:2011 [26]. The size
distribution of particles is given in Figure 5C where it can be seen that approximately 50% of
particles have a size less than 0.4 µm.
EDAX analysis of wear particles chemical composition demonstrated typical profiles of
Ti6Al4V alloy with titanium and aluminum (fig. 5D). Oxygen peaks make it possible to
assume oxidation of titanium wear particles within the biological environment. Gold and
palladium peaks appear from the nanosize coating deposited on particles for EDAX analysis.
In-Vitro pin-on-disk wear test
Wear
Figure 6A and B demonstrates volume wear loss of Ti6Al4V, Nitinol and CoCr pins tested
against titanium disks. Wear loss of Ti6Al4V pins was 33±2mm3 after 0.5 million cycles
which is at least 100 times more compared to Nitinol and CoCr pins which were measured to
Page 10
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
be 0.15±0.02 and 0.1±0.01 mm3 respectively after 0.5 million cycles (fig.6B). Nevertheless,
wear resistance of disk counter-parts (fig. 6C) made of Ti6Al4V did not demonstrate much
difference regardless of pin material and was measured to be 37±3, 25±2 and 20±2 mm3 after
0.5 million cycles when tested against Ti6Al4V, Nitinol and CoCr pins respectively.
Analysis of wear scars and particles
Ti6Al4V- Ti6Al4V friction combination
SEM analysis of wear scars on pins and disks made of Ti6Al4V alloy (fig.7A) revealed
abrasive and adhesion mechanisms of Ti6Al4V alloy wear damage similar to that observed
on the retrievals of LSZ-4D devices (fig.5A).
SEM micrographs of wear particles collected after 0.5 million cycles for Ti6Al4V-Ti6Al4V
combination, their size distribution and EDAX chemical composition are given in Figure
7B,C,D. The average particle size was 0.45µm (range from 0.13 to 1.6 µm). Their shape was
round to oval, similar to that collected from patient’s tissues (average aspect ratio was 1.3-
1.6). Approximately 70% of particles had size less than 0.4 µm. EDAX analysis of the
particles revealed the presence of titanium, aluminum and vanadium. There was no
statistically significant difference in the size of wear particles digested from tissues adjacent
to the implanted LSZ-4D device and those collected after in-vitro wear test (p=0.57).
Nitinol-Ti6Al4V friction combination
Mild abrasive grooves and deposits of the titanium counter-part were seen over the entire
surface of the Nitinol pins (fig.8A). The roughness of wear scars on Nitinol was 0.170.05
Ra.
SEM micrographs of wear particles collected after 0.5 million cycles for Nitinol-Ti6Al4V
combination and their size distribution are given in Figure 8B,C. The average size of the wear
particles was 0.3µm (range from 0.10 to 1.97 µm) with a round shape (average aspect ratio
Page 11
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
was 1.2-1.4). Approximately 70% of particles were less than 0.4 µm. EDAX analysis of
particles revealed no traces of Ni (fig. 8D).
Discussion
The volume wear rate measured for the retrieved components of LSZ-4D growth guidance
device made of titanium alloy Ti6Al4V was found to be 12.5 mm3 per year with 5mm
3 and
7.5 mm3 per year contributed from rods and fixtures. We measured wear rate on three devices
which had typical size and depth of wear scars normally observed during routine operations
to replace the sliding mechanism with a spinal fusion devices. The absolute values of wear
rate will depend on the design of the device, fixtures density and patient activity. Wear loss
values observed in our study are greater compared to the Shilla growth-guidance device made
of SS. The latter is reported to have 0.78 mm3 per year [20], which can be explained by the
higher wear resistance of SS compared with titanium and the difference in fixtures density.
However wear rate measured in the current work for LSZ-4D devices, is similar to values
(15-26 mm3 per year) reported for metal-on-metal THR made of CoCr alloys which are
known for their superior wear resistance [7,31-32]. In hip replacements this amount of metal
debris although composed of CoCr, often leads to adverse biological reactions, bone
resorption and implant failure. Taking into account the poor wear resistance of titanium
alloys and the size of the wear scars on the rods the range of motion in sliding growth
guidance devices may be much lower compared to THR articulations. Patients with
implanted LSZ-4D devices analysed in this study developed no clinical complications or
device loosening. However, in other cases seromas, fistulas and inflammation of the tissues
adjacent to the device was observed after LSZ-4D instrumentation implantation, which may
be caused by the metal wear debris. Work is now in progress to estimate the occurrence and
severity of these complications. Retrieved spinal instrumentation made of more wear resistant
SS has been reported to have corrosion damage in addition to wear scars [15]. Therefore,
Page 12
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
using of spinal instrumentation made of titanium with additional wear resistant coatings to
limit the release of metal debris may be beneficial.
This work has shown that over 50% of wear particles retrieved from tissues surrounding
LSZ-4D devices were less than 400 nm. This is similar to the size described for metal
particles from THR which is reported to be 0.02-0.8 μm for CoCr and from 0.04-0.9 μm for
Ti [33]. The larger size of wear particles observed for Shilla devices [20] might be possibly
explained by using laser scattering methodology for measuring particles size where the effect
of particle agglomeration is unclear.
Results of in-vitro pin-on-disk test in this work revealed that the wear resistance of Nitinol is
comparable to that of CoCr with a volume wear rate is 100 times less compared to Ti6Al4V.
Similar performance of Nitinol was observed by Li [34] and can be explained by superelastic
deformation of Nitinol asperities thus reducing contact stresses and abrasive wear [35-36].
Comparison of wear scars on retrieved LSZ-4D device components and titanium pins and
disks in in-vitro test reveals similar abrasive and adhesive wear damage mechanisms. The
size and shape of wear particles retrieved from tissues adjacent to the implanted LSZ-4D
device and those digested from bovine serum after the laboratory wear test are also similar,
which makes it possible to assume that in-vitro test adequately simulates reciprocating
translation movement of rods against fixtures.
The wear rate of titanium pins in Ti6Al4V-Ti6Al4V friction combination being 33±2mm3 per
0.5 million cycles corresponds approximately to wear debris produced by two rods of LSZ
device after 6.5 years of implantation. Volume wear loss of Nitinol pins after 0.5 million
cycles was measured to be 0.15 mm3, which is relatively low for 6 years of implantation. It
may be concluded that the high wear resistance of Nitinol rods combined with their
superelastic and shape memory properties could be beneficial for the application in spinal
instrumentation. Nevertheless, further in-vitro wear tests of assembled devices incorporating
Page 13
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Nitinol rods are needed in order to evaluate the volume of wear debris generated by specific
device which will take into account the device design, lubrication and edge effects.
Limitations of the study:
Only 3 retrieved implants have been investigated in our study. Nevertheless, it was shown
that the in-vitro wear tests are able to replicate wear damage mechanisms seen on retrievals
and the shape and size of wear particles isolated from tissues adjacent to LSZ-4D implants
was similar to those isolated from the in- vitro tests. This made it possible to compare the
wear performance of Nitinol, which might be a prospective material for spinal
instrumentation with metal materials which are currently used. However, further in vitro and
in-vivo testing of Nitinol is necessary for its safe application.
Conclusions
This paper investigates the wear of sliding titanium devices to treat early onset scoliosis in
adolescent patients. We showed the average volumetric wear rate for the retrieved LSZ-4D
growth guidance device made of titanium alloy (Ti6Al4V) was in total 12.5 mm3 per year. Of
this, the rods contributed 5mm3 per year with the fixtures contributing 7.5 mm
3 per year. We
used an in-vitro pin-on-disk wear test to investigate the wear resistance of Nitinol tested
against titanium alloy in a simulated body environment. We showed that Nitinol wear loss
was 100 times lower compared with titanium alloy Ti6Al4V and comparable to that of Cobalt
Chromium alloy. The wear mechanism and the size and shape of wear particles after in-vitro
tests was similar to wear patterns identified from retrieved titanium components and to the
wear debris isolated from tissues adjacent to these implants.
References:
Page 14
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
1. Karol LA, Johnston C, Mladenov K et al. Pulmonary function following early
thoracic fusion in non-neuromuscular scoliosis. The Journal of Bone & Joint Surgery
2008;90(6):1272-1281.
2. Campbell Jr, R. M. (2013) VEPTR: past experience and the future of VEPTR
principles. European Spine Journal 2013; 22(2): 106-117.
3. Akbarnia BA, Cheung K, Noordeen H et al. Next Generation of Growth-Sparing
Techniques: Preliminary Clinical Results of a Magnetically Controlled Growing Rod
in 14 Patients With Early-Onset Scoliosis. Spine 2013;38(8):665-670.
4. McCarthy RE, Luhmann S, Lenke L et al. The Shilla Growth Guidance Technique for
Early-Onset Spinal Deformities at 2-Year Follow-Up: A Preliminary Report. J.
Pediatr. Orthop. 2014;34:1-7.
5. Sampiev MT, Laka AA, Zagorodniy NV et al. The application of growth-guidance
LSZ device for infantile and adolescent scoliosis treatment. Russian Medical Journal
2013;5:24-28.
6. Thompson, G. H., Akbarnia, B. A., & Campbell Jr, R. M. (2007). Growing rod
techniques in early-onset scoliosis. Journal of Pediatric Orthopaedics 2007; 27(3):
354-361.
7. Lord JK, Langton DJ, Nargol AV, et al. Volumetric wear assessment of failed metal-
on-metal hip resurfacing prostheses. Wear 2011, 272:79-87.
8. Lee JL, Billi F, Sangiorgio SN, et al. Wear of an experimental metal-on-metal
artificial disc for the lumbar spine. Spine 2008; 3: 597-606.
Page 15
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
9. McKellop H, Rostlund R, Ebramzadeh E, et al. Wear of titanium 6-4 alloy in
laboratory tests and in retrieved human joint replacements. In: Brunette D, ed.
Titanium in medicine: material science, surface science, engineering.Berlin: Springer;
2001: 747-770.
10. Fabi, D., Levine, B., Paprosky, W., Della Valle, C., Sporer, S., Klein, G.&
Hartzband, M. Metal-on-metal total hip arthroplasty: causes and high incidence of
early failure. Orthopedics, 2012; 35(7): e1009-16.
11. Ollivere, B., Darrah, C., Barker, T., Nolan, J., & Porteous, M. J. Early clinical failure
of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and
soft-tissue necrosis. Journal of Bone & Joint Surgery, British Volume 2009; 91(8):
1025-1030.
12. Mahendra, G., Pandit, H., Kliskey, K., Murray, D., Gill, H. S., & Athanasou, N.
Necrotic and inflammatory changes in metal-on-metal resurfacing hip arthroplasties:
relation to implant failure and pseudotumor formation. Acta orthopaedica 2009;
80(6): 653-659.
13. Hallan, G., Lie, S. A., & Havelin, L. I. High wear rates and extensive osteolysis in 3
types of uncemented total hip arthroplasty: a review of the PCA, the Harris Galante
and the Profile/Tri-Lock Plus arthroplasties with a minimum of 12 years median
follow-up in 96 hips. Acta orthopaedica 2006; 77(4): 575-584.
14. Schroeder, H. P., Smith, D. C., Gross, A. E., Pilliar, R. M., Kandel, R. A.,
Chernecky, R., & Lugowski, S. J. Titanemia from total knee arthroplasty: a case
resulting from a failed patellar component. The Journal of arthroplasty 1996;
11(5):620-625.
15. Villarraga ML, Cripton PA, TetiSD, et al Wear and corrosion in retrieved
thoracolumbar posterior fixation. Spine 2006;31:2454-2462.
Page 16
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
16. Wang J, Warren D, Harvinder S, et al Metal debris from titanium spinal implants.
Spine 1999;24:899-903.
17. Hallab NJ, Cunningham BW, Jacobs JJ. Spinal implant debris-induced osteolysis.
Spine 2003;28(suppl):125–38.
18. Cunningham BW, Orbegoso CM, Dmitriev AE, et al. The effect of spinal
instrumentation particulate wear debris: an in vivo rabbit model and applied clinical
study of retrieved instrumentation cases. Spine J 2003;3:19–32.
19. Mochida Y, Bauer TW, Nitto H, et al. Influence of stability and mechanical properties
of a spinal fixation device on production of wear debris particles in vivo. J Biomed
Mater Res 2000;53:193–8.
20. Singh V, Simpson J, Rawlinson J, et al Growth Guidance system for early-onset
scoliosis. Spine 2013;38:1546-1553.
21. Yoshihara H. Rods in spinal surgery: a review of the literature Spine J 2013: In press.
22. Kollerov, M., Lukina, E., Gusev, D., Mason, P., & Wagstaff, P. (2013). Impact of
material structure on the fatigue behaviour of NiTi leading to a modified Coffin–
Manson equation. Materials Science and Engineering: A 2013; 585: 356-362.
23. Otsuka C, Wayman M. Shape Memory Materials, Cambridge: Cambridge University
Press, 1998.
24. Cheung K, Kuong E, Samartzis D, et al. Assessing the Safety and Efficacy of a Novel
Superelastic Rod in Comparison to Conventional Titanium Rod for Scoliosis Curve
Correction. Paper presented at: 18th
International Meeting on advanced spinal
techniques. 2011; Copenhagen.
Page 17
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
25. Lenke, L. G., Betz, R. R., Harms, J., Bridwell, K. H., Clements, D. H., Lowe, T. G., &
Blanke, K. Adolescent idiopathic scoliosis a new classification to determine extent of
spinal arthrodesis. The Journal of Bone & Joint Surgery 2001; 83(8): 1169-1181.
26. ISO 17853:2011 Wear of implant materials – Polymer and metal wear particles –
Isolation and characterization 11.040.40: Implants for surgery, Prosthetics and
Orthotics. Geneva, Switzerland: International Organization for Standardization, 2008.
27. ASTM G99-05(2010) Standard Test Method for Wear Testing with a Pin-on-Disk
Apparatus.Section Three – Corrosion of metals; Wear and Erosion. West
Conshohocken, PA:ASTM International, 2010.
28. ISO 18192-1:2011 Implants for surgery -- Wear of total intervertebral spinal disc
prostheses -- Part 1: Loading and displacement parameters for wear testing and
corresponding environmental conditions for test. Implants for surgery, Prosthetics
and Orthotics. Geneva, Switzerland: International Organization for Standardization,
2011.
29. Rohlmann A, Graichen F, Bergmann G. Loads on an Internal spinal fixation device
during physical therapy. J Bone Joint Surg Am 2002;82:44-52.
30. ASTM 2082-06 Standard Test Method for Determination of Transformation
Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free
Recovery.Section Thirteen - Medical and Surgical Materials and Devices. West
Conshohocken, PA:ASTM International, 2006.
31. Morlock MM, Bishop N, Zustin J et al. Modes of implant failure after hip resurfacing:
morphological and wear analysis of 267 retrieval specimens. J. Bone Joint Surg. Am.
2008;90 (Suppl. 3):89–95.
Page 18
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
32. Witzleb W, Hanisch U, Ziegler J et al. In vivo wear rate of the Birmingham hip
resurfacing arthroplasty: a review of 10 retrieved components. J.Arthroplasty
2009;24(6):951–956.
33. Doorn PF, CampbellPA, Worrall J. Metal wear particle characterization from metal
on metal total hip replacements: transmission electron microscopy study of
periprosthetic tissues and isolated particles. J Biomed Mat Res 1998:103-111.
34. Li DY. A new type of wear-resistant material: pseudo-elastic TiNi alloy. Wear
1998;221:116-123.
35. Qian L, Zhou Z, Sun Q. The role of phase transition in the fretting behavior of NiTi
shape memory alloy. Wear 2005;259:309-318.
36. Yan W. Theoretical investigation of wear-resistance mechanism of superelastic shape
memory alloy NiTi. Materials Science and Engineering: A. 2006;427: 348-355.
Figure 1. (A) Illustration of LSZ-4D sliding growth guidance device. Locked fixture is used
on one spinal level. Unlocked fixtures are used at distal and proximal end of the device
enabling sliding and continued spinal growth; (B) Assembled fixture unit consisting of clip
and hook.
Page 19
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Figure 2. Photos of wear grooves on LSZ-4D sliding device components (rods (A), clips (C)
and hooks (E)) and their corresponding images (B, D and F) taken using Bruker
interferometry microscope and reconstructed in MATLAB followed by 3D visualisations
performed using Amira software.
Figure 3. Photo (A) and scheme (B) of pin-on-disk wear test for the in-vitro evaluation of
wear resistance of metal materials used in the manufacture of spinal implants.
Page 20
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Figure 4. Average volume wear for retrieved components of LSZ-4D device made of titanium
alloy Ti6Al4V. The whole device analyzed from each patient included 2 rods and 40±8
fixture elements (20±4 hooks and 20±4 clips).
Page 21
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Figure 5. (A) SEM micrograph of the wear scar on LSZ-4D device retrieved hook revealing
abrasive and adhesion mechanisms of wear damage; (B) SEM micrograph of wear particles
enzymatically digested from retrieved tissues. (C) Particle size distribution calculated using
SEM micrographs (D) The chemical composition of typical wear particles measured by
EDAX analysis.
Page 22
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Figure 6. (A) Volume wear loss of Ti6Al4V, Nitinol and CoCr pins tested against Ti6Al4V
disks in-vitro pin-on-disk; (B) detailed comparison revealed that volume wear loss of Nitinol
pins is comparable to that of CoCr; (C) volume wear loss of titanium disks counter-parts
remains high regardless of the pin material.
Page 23
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Figure 7. (A) SEM micrograph of Ti6Al4V pin wear scar after pin-on-disk test of Ti6Al4V-
Ti6Al4V combination revealing abrasive/adhesion mechanisms of wear damage similar to
that seen on retrievals; (B) SEM micrograph of wear particles collected after 0.5 million
cycles; (C) Particle size distribution (D) The chemical composition of wear particles.
Figure 8. (A) SEM micrograph of Nitinol pin wear scar after pin-on-disk test of Nitinol-
Ti6Al4V combination revealing mild abrasive wear and adhesion of titanium counter-part;
(B) SEM micrograph of wear particles; (C) Particle size distribution (D) The chemical
composition of wear particles revealed no traces of Ni.
Page 24
Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Table 1 Clinical information for all retrieved LSZ-4D sliding devices
Implant Age at
implantation
Gender Scoliosis
type* and
degree
Level of
device
implantation
Implan
tation
time
(years)
Number of device
elements
rods fixtures
clips hooks
LSZ-4D(1) 11 Female IBN,
76
T2/L4 5.8 2 24 24
LSZ-4D(2) 12 Female IIAN,
63
T4/L2 3.5 2 20 20
LSZ-4D(3) 10 Male IAN,
66
T2/T12 3.5 2 16 16
*Lenke classification [25]