TUGAS AKHIR ANALISA KINERJA GAS CHROMATOGRAPHY (GC-FID 2010) PADA PENGARUH TEMPERATURE COLUMN TERHADAP NILAI RETENTION TIME DAN AREA OF DETECTION PEAK DARI BHYPENILE IN N-HEXANE DI PT. DITEK JAYA (SHIMADZU Analytical and Scientific Corp.) WENI YULIATI 2410.105.012 Pembimbing : Ir. Mochamad Ilyas Hs. Katherin Indirawati ST. MT. 1
24

Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

Aug 26, 2019

Download

Documents

lytuong
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

TUGAS AKHIR

ANALISA KINERJA GAS CHROMATOGRAPHY (GC-FID 2010) PADA PENGARUH TEMPERATURE COLUMN TERHADAP NILAI RETENTION TIME DAN AREA OF DETECTION PEAK

DARI BHYPENILE IN N-HEXANE DI PT. DITEK JAYA (SHIMADZU Analytical and Scientific Corp.)

WENI YULIATI2410.105.012

Pembimbing : Ir. Mochamad Ilyas Hs.Katherin Indirawati ST. MT.

1

Page 2: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

2

LATAR BELAKANG

HASIL PRODUKSI ANALISA GAS CHROMATOGRAPHY QUALITY CONTROL

Page 3: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

PERUMUSAN MASALAH

Sering kali para pengguna Gas Chromatography mendapatkan hasil analisa yang tidaksesuai dengan yang diharapkan, misalkan hasil chromatogram yang tidak sempurnayaitu seperrti bertumpukan atau tailling bahkan sering juga tidak munculnyachromatogram.

3

Page 4: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

TUJUAN

Menganalisa kinerja Gas Chromatography pada pengaruh temperature column terhadapnilai retention time dan area of detection peak dari bhypenile in n-hexane

4

Page 5: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

BATASAN MASALAH

� Standard Sample yang digunakan untuk menganalisa kinerja Gas Chromatographyadalah Bhypenile in n-Hexane dengan konsentrasi 4,6 ppm.

� Variasi temperature column yang digunakan adalah 140oC ; 150oC ; 160oC ; 170oC ;180oC ; 190oC ; 200oC

� Analisa dilakukan menggunakan metode repeatability dengan autosampler 5 kali injeksiStandard Sample per variasi temperature column dengan syarat nilai %RSD < 0.5 %.

� Carrier Gas yang digunakan adalah He (Helium) dengan tingkat kemurnian UHP (UltraHigh Purity) dengan presentase 99.999% dan menggunakan tekanan 100 kPa.

5

Page 6: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI

1

Start

Identifikasi

Permasalahan

Study Literatur

Study Lapangan +

Pengambilan Data

Analisa Data dan

Pembahasan

Kesimpulan dan

Saran

Page 7: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

TINJAUAN PUSTAKAChromatography

• Chromatography adalah metode pemisahan fisik dimana komponen yang dipisahkan

didistribusikan antara dua tahap

• Salah satunya adalah fase diam (stationary phase) dan lainnya adalah fase gerak (mobile

phase) yang bergerak dalam arah yang pasti.

• Pada fase diam ini akan terjadi pemisahan komponen campuran sampel

• Dengan teknik ini memungkinkan untuk menetukan keberadaan suatu senyawa hingga

menentukan konsentrasi sampelnya.

Sejarah Chromatography

• Sejarah kromatografi dimulai pada pertengahan abad ke-19. Pada awalnya, kromatografi

digunakan untuk pemisahan pigmen yang terdapat dalam tanaman seperti klorofil.

• Kromatografi jenis baru yang dikembangkan pada tahun 1940-an dan 1930-an

menjadikan teknik ini lebih berguna untuk berbagai jenis proses pemisahan.

Kromatografi dikembangkan oleh Archer John Porter Martin dan Richard Laurence

Millington Synge selama tahun 1940-an dan 1950-an.

2

Page 8: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

TINJAUAN PUSTAKAGas ChromatographyMain Component

3

Page 9: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

Injector Port

Injektor adalah tempat untuk memasukkan sampel yang akan dianalisa agar bisa

terbawa oleh sebuah fase gerak masuk ke dalam kolom.

TINJAUAN PUSTAKA

Page 10: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

Column Oven

Kolom adalah tempat dimana proses pemisahan terjadi, yaitu dengan menggunakan

suhu tertentu yang sesuai dengan karakteristik dari sampel.

TINJAUAN PUSTAKA

Page 11: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

Detector

Detector berfungsi sebagai pendeteksi kuantitas sampel yang sudah dipisahkan oleh

column dalam Gas Chromatography, sehingga detector akan membaca sinyal sesuai

dengan jumlah senyawa yang di injeksikan.

TINJAUAN PUSTAKA

Page 12: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIAN

2

Gas Chromatography

Page 13: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIAN

3

Flowchart Carrier Gas Flowline Gas Chromatography

Page 14: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIAN

Spesifikasi SHIMADZU GC FID – 2010

Injection Port

• Range Suhu : Suhu Ruang +4°C sampai 450°C dan -50 °C to 450 °C

• Akurasi Suhu : ±1 % (K) (kalibrasi pada 0.01 °C)

• Injection Unit : Split/Splitless Injection, Direct Injection

• Kecepatan Cooling : ± 6 menit dari 450 °C ke 50 °C

• Overheat Protection : Programmable > 470 °C

Advanced Flow Controller (AFC)

4

Page 15: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIANSpesifikasi SHIMADZU GC FID – 2010

Column Oven

• Dimensi (mm) : 280 (p) × 280 (l) × 175 (t)

• Volume (L) : 13.7

• Range Suhu : Suhu Ruang +4°C sampai 450°C dan -50 °C to 450 °C

• Akurasi Suhu : ±1 % (K) (kalibrasi pada 0.01 °C)

• Deviasi Suhu : ± 2°C

• Stabilitas Suhu : ± 0.05 °C

• Koefisien Suhu : 0.01 °C / °C

• Kecepatan Cooling : ± 6 menit dari 450 °C ke 50 °C

• Overheat Protection : Programmable > 470 °C

5

Page 16: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIANSpesifikasi SHIMADZU GC FID – 2010

Detector

• Detektor Tipe : Flame Ionization Detector (FID)

• Range Suhu : Suhu Ruang +4°C sampai 450°C dan -50 °C to 450 °C

• Akurasi Suhu : ±1 % (K) (kalibrasi pada 0.01 °C)

• Overheat Protection : Programmable > 470 °C

• Jet Material : Fused Quartz

• Waktu Konstant : 4 ms to 2 s selectable

Advanced Pressure Controller

1

Page 17: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIANGC - Solution

2

Page 18: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIANReal Time Analysis

3

Page 19: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

METODOLOGI PENILITIANGC – Solution (Setting INJECTOR)

4

Page 20: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

PENGUJIAN DAN ANALISA DATAHasil Analisa Pada Temperature Column 200oC

1

Page 21: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

PEMBAHASANDari analisa data diatas dapat dilihat bahwa hasil injeksi standard sample sudah

memenuhi persyaratan yaitu variasi perubahan nilai retention time dan area of detection

peak serta nilai height pada setiap variasi temperature dengan nilai rata – rata %RSD

dibawah 0.5%. Untuk memperjelas digunakan tabel untuk representatif dari hasil

perhitungan nilai %RSD sebagai berikut :

2

Page 22: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

PEMBAHASAN

3

RETENTION TIMETemperature Column

(°C)n-Hexane Bhypenile

140 2.085 4.035150 1.855 3.858160 1.587 3.267170 1.425 3.022180 1.223 2.923190 1.028 2.682200 0.892 2.357

0.000

0.500

1.000

1.500

2.000

2.500

3.000

3.500

4.000

4.500

140 150 160 170 180 190 200

min

Temperature Column (deg.C)

Retention Time

(n-Hexane)

Bhypenile

AREATemperature Column

(°C)n-Hexane Bhypenile

140 779304940 37920037150 778735257 37765976160 773612140 37573903170 775838641 37775757180 773183254 37132913190 771813981 37137270200 772469280 37135200

0

100000000

200000000

300000000

400000000

500000000

600000000

700000000

800000000

900000000

140 150 160 170 180 190 200

uV/m

in

Temperature Column (deg.C)

Area

(n-Hexane)

Bhypenile

Page 23: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

PEMBAHASAN• Hasil perhitungan nilai %RSD diatas menunjukkan bahwa tingkat kestabilan dari

instrument GC (Gas Chromatography) disini adalah sesuai dengan persyaratan dari

SHIMADZU Performance Qualification yaitu dibawah 0.5% untuk nilai %RSD dengan

teknis injeksi menggunakan autosampler.

• Dari hasil analisa tersebut dapat dianalisa mengenai kinerja GC (Gas Chromatography)

yaitu pengaruh temperature column terhadap retention time dan area of detection peak

adalah bahwa temperature column dapat mempengaruhi nilai retention time dari

detection peak yaitu semakin besar nilai temperature column maka nilai retention

time semakin kecil begitu pula sebaliknya semakin kecil nilai temperature column

maka nilai retention time semakin besar atau dengan kata lain waktu analisa bisa

diperkecil dengan cara setting temperature column.

• Untuk nilai dari area of detection peak cenderung stabil atau tidak berubah secara

signifikan. Hanya saja bentuk chromatogram dari hasil analisa yang berbeda yaitu

semakin besar nilai temperature column maka bentuk dari chromatogram semakin

meruncing atau bisa dikatakan nilai height semakin besar begitu pula sebaliknya

semakin kecil nilai temperature column maka bentuk dari chromatogram semakin

melebar atau bisa dikatakan nilai height semakin kecil.

4

Page 24: Analisa kinerja gas chromatography tipe Shimadzu GC-FID ... · Analisa kinerja gas chromatography tipe Shimadzu GC-FID 2010 pada pemgaruh perubahan temperature column terhadap nilai

KESIMPULAN dan SARAN

Kesimpulan

• Telah dilakukan analisa data dengan parameter methode yang telah ditentukan dan didapatkan hasil

sesuai dengan persyaratan yaitu variasi perubahan nilai retention time dan area of detection peak serta

nilai height dan concentraion dengan nilai rata – rata %RSD dibawah 0.5%.

• Kinerja SHIMADZU GC-2010 mengenai pegaruh temperature column terhadap nilai retention time adalah

semakin besar nilai temperature column maka nilai retention time semakin kecil begitu pula sebaliknya

semakin kecil nilai temperature column maka nilai retention time semakin besar atau dalam arti waktu

analisa bisa diperkecil dengan cara setting temperature column.

• Kinerja SHIMADZU GC-2010 mengenai pegaruh temperature column terhadap nilai area of detection peak

adalah pada bentuk chromatogram dengan ketentuan semakin besar nilai temperature column maka

bentuk dari chromatogram semakin meruncing begitu pula sebaliknya semakin kecil nilai temperature

column maka bentuk dari chromatogram semakin melebar tetapi tidak berpengaruh pada nilai area of

detection peak itu sendiri.

Saran

• Untuk penelitian selanjutnya agar dapat dikembangkan dari segi karakteristik Gas Chromatography yang

lain seperti setting nilai variable proses yang lain seperti pressure dan flow.