Top Banner
VAPORIZERS Tec 1 to Tec 5 MODERATOR: DR SARIKA SPEAKER: DR PREETI
59

anaesthesia Vaporizers tec1 to tec5

Aug 10, 2015

Download

Health & Medicine

Preeti Loona
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: anaesthesia Vaporizers tec1 to tec5

VAPORIZERS Tec 1 to Tec 5

MODERATOR: DR SARIKA

SPEAKER: DR PREETI

Page 2: anaesthesia Vaporizers tec1 to tec5

INTRODUCTION

Inhalational anaesthetic agents are usually liquids at room temperature and barometric pressure and need to be converted to vapour before being used and this conversion is effected using a vaporizer.Modern vaporizers are flow and temperature compensated, concentration calibrated, direct reading, dial controlled and are unaffected by positive pressure ‑ventilation.

Page 3: anaesthesia Vaporizers tec1 to tec5

The concentration of vapour is continuously monitored and adjusted by altering fresh gas flow through the vaporizerSafety features include an anti spill and a specific ‑vaporizer filling device.

Page 4: anaesthesia Vaporizers tec1 to tec5

Vaporizers and Standards

The ASTM anesthesia workstation standard contains the following provisions regarding vaporizers:

The effects of variations in ambient temperature and pressure, tilting, back pressure, and input flow rate and gas mixture composition on vaporizer performance must be stated in the accompanying documents.

The average delivered concentration from the vaporizer shall not deviate from the set value by more than ±20% or ±5% of the maximum setting, whichever is greater, without back pressure.

Page 5: anaesthesia Vaporizers tec1 to tec5

The average delivered concentration from the vaporizer shall not deviate from the set value by more than +30% or -20% or by more than +7.5% or -5% of the maximum setting, whichever is greater, with pressure fluctuations at the common gas outlet of 2 kPa with a total gas flow of 2 L/minute or 5 kPa with a total gas flow of 8 L/minute.

A system that prevents gas from passing through the vaporizing chamber or reservoir of one vaporizer and then through that of another must be provided.

The output of the vaporizer shall be less than 0.05% in the “OFF” or “zero” position if the “zero” position is also the “OFF” position.

Page 6: anaesthesia Vaporizers tec1 to tec5

All vaporizer control knobs must open counterclockwise.

Either the maximum and minimum filling levels or the actual usable volume and capacity shall be displayed.

The vaporizer must be designed so that it cannot be overfilled when in the normal operating position.

Page 7: anaesthesia Vaporizers tec1 to tec5

Vaporizers outside the breathing system must have noninterchangeable proprietary or 23-mm fittings and the inlet of the vaporizer must be male and the outlet female.

Vaporizers suitable for use in the breathing system must have standard 22-mm fittings or screw-threaded, weight-bearing fittings with the inlet female and the outlet male.

The direction of gas flow must be indicated by arrows and the vaporizer marked “for use in the breathing system.”

Page 8: anaesthesia Vaporizers tec1 to tec5

In order to give clinically useful concentrations of the agent, the anaesthetic vapor has to be diluted with fresh gas in one of the two ways:

1)By splitting the fresh gas flow so that only a portion passes through the vaporizing chamber and the rest bypasses it – variable bypass vaporizers.

2)By injecting the vapor directly to the total fresh gas flow without any split – measured flow vaporizers.

Page 9: anaesthesia Vaporizers tec1 to tec5

A. Method of regulating output concentration 1. Concentration calibrated (variable-bypass) 2. Measured flow (copper Kettle)

CLASSIFICATION

B. Method of vaporization 1. Flow over

2.Bubble Through 3.Injection

Page 10: anaesthesia Vaporizers tec1 to tec5

C. Temperature compensation 1. Thermocompensation

2. Supplied heat

E. Specificity 1. Agent specific

2. Multiple agent

D. Resistance 1. Plenum

2. Low resistance(draw over)

Page 11: anaesthesia Vaporizers tec1 to tec5

F.POSITION

1.VAPORIZER INSIDE CIRCUIT

2.VAPORIZER OUTSIDE CIRCUIT

Page 12: anaesthesia Vaporizers tec1 to tec5

TEC 1

introduced by Cyprane company in 1956 as FLUOTEC 1 for Halothane. Also called Mark 1. Problem of proportioning valve sticking there is a risk of overdose.

Page 13: anaesthesia Vaporizers tec1 to tec5

•variable bypass,

•flow over with wick

•out of system

•temperature compensated

•agent specific

TEC 2 VAPORIZER

Page 14: anaesthesia Vaporizers tec1 to tec5

Construction:

Vaporizing chamber- round with concentric wicks. Bimetallic strip- outlet of chamber. Dial in front.

Gas flows: Controlled by spindle- rightward movement.

OFF position-flow only through bypass chamber. ON position-more gas through vaporizing chamber.

Page 15: anaesthesia Vaporizers tec1 to tec5
Page 16: anaesthesia Vaporizers tec1 to tec5

In a bimetallic strip, two metals with very different coefficients of thermal expansion are fixed together.

Page 17: anaesthesia Vaporizers tec1 to tec5

When the temp. of the vaporizing chamber drops, the bimetallic strip bends and moves away. This reduces the resistance to flow and thus more flow occurs into the vaporizing chamber

Page 18: anaesthesia Vaporizers tec1 to tec5

TEC 2 Evaluation and Use

Not accurate below 4Liters/min. With flow<2L/min, at low dial setting <2%, output is less than setting and more than setting with more dial setting.

With N2O gives lower output at higher setting and higher at lower settings.

With dial setting bet 0% and OFF, some output can occur varying with gas flow.

Prone to pumping effect at low flows and pressuring effect at high flows.

Small leak in the off position.

Page 19: anaesthesia Vaporizers tec1 to tec5

Disadvantages Halothane preservative Thymol gets deposited in vaporizer causing operating spindle to stick. Back pressure forcing saturated Vapour back into bypass chamber. High concentration at low flows.

Page 20: anaesthesia Vaporizers tec1 to tec5

TEC 3Models Fluotec Mark 3 Pentec Mark 2 Enfluratec Fortec

Page 21: anaesthesia Vaporizers tec1 to tec5

TEC 3 - Parts The VC has two concentric wick skirts, which enclose nickel plated copper helix in between.

This assembly forms a long spiral channel through which carrier gas flows before entering the VC preventing back pressure problems.

The bimetallic strip within the bypass chamber increases flow through the bypass chamber when temperature increase.

Page 22: anaesthesia Vaporizers tec1 to tec5

TEC 3 VAPORIZER

variable bypass

flow over with wick

automatic thermocompensation

agent specific

Construction: Long inlet of vaporizing chamber Concentration dial at the top, with locking lever- controls size of control channel. Filling & draining- bottom

Page 23: anaesthesia Vaporizers tec1 to tec5
Page 24: anaesthesia Vaporizers tec1 to tec5

TEC 3 Evaluation and Use

All are accurate at lower dial settings. At high dial settings, higher than expected concentration at low flow rates and vice versa.

Negligible effect of- O2 Flush, back pressure, increase or decrease of carrier gas.

N2O has little effect. Between OFF and 0.5%, output was less effected by FGF than in TEC 2.

Small leaks in bypass in OFF position. Tipping upto 90 degree does not affect output.

Page 25: anaesthesia Vaporizers tec1 to tec5

TEC 3 Disadvantages

Dial rotation problems-

rotation beyond OFF position.

rotated by 180 degree.

leak from dial setting because of damage or compression of gasket.

Page 26: anaesthesia Vaporizers tec1 to tec5

Tec 4Models- Fluotec Mark 4 Enfluratec 4 Fortec 4

Page 27: anaesthesia Vaporizers tec1 to tec5
Page 28: anaesthesia Vaporizers tec1 to tec5

Tec 4 A vapouriser designed for ‘out-of-circuit’ use in continuous flow techniques of inhalation anaesthesia with built in temperature-compensated and pressure-compensated capabilities.

The Tec 4 was introduced for BOC Model 2000 anaesthetic machine in 1983.

Page 29: anaesthesia Vaporizers tec1 to tec5

It was a remodelled Tec 3.

To overcome the problems of Tec 3 it incorporated internal baffle system to reduce the danger of liquid agent entering the bypass chamber on tilting.

Another interesting modification, to ensure only a single vapouriser operation at any time, was the safety interlock system.

This vapouriser is available for different specific agents, i.e. enflurane (dial setting range 0-7%), halothane (0-5%), isoflurane (0-5%)

Page 30: anaesthesia Vaporizers tec1 to tec5

TEC 4 Parts On top- Control Dial that is turn clockwise to increase the concentration.

A release button on the left of concentration dial must b depressed before the vaporizer is turned on.

Locking lever-connected to control knob. When the vaporizer is turned on, 2 plungers within the vaporizers operate to open the valve into the Fresh gas stream. Also prevent the use of any adjacent vaporizers,

2 Filling mechanisms- Screw cap and Keyed Filling system

Page 31: anaesthesia Vaporizers tec1 to tec5

TEC 4 Gas Flow

OFF Position- incoming gas flows from inlet and into one channel across the top of the vaporizer , without coming in contact with vaporizing chamber or temperature compensating device and leave through the outflow

ON Position- •incoming gas split into 2 streams by a rotary valve attached to concentration dial. • 1 stream goes to VC and that enters one of the two chambers which surrounds the bypass chamber.•After passing it is directed over 2 concentric wicks that enclose a copper helix which converts this space into long spiral outlet channel.

Page 32: anaesthesia Vaporizers tec1 to tec5

TEC 4•Wicks assure maximum contact between gas and agent.

•This vapor laden gas leaves via the second chamber surrounding bypass chamber to outlet.

The remaining Fresh gas flows to the bypass chamber which contains temperature sensitive element.

No spillage after tilting or inversion.

Page 33: anaesthesia Vaporizers tec1 to tec5

TEC 4 Evaluation and Use

Fluctuation back pressure can increase the concentration.

Greater effects seen at-

low flow rates, low dial setting, large and frequent pressure fluctuations.

N2O- decreases output slightly.

Vaporizer is filled and used in upright position- deviation do not affect the output or safety.

Page 34: anaesthesia Vaporizers tec1 to tec5

TEC 4 Disadvantages

Difficulty in operation one handed.

Yearly service interval.

Page 35: anaesthesia Vaporizers tec1 to tec5

TEC 5 Vaporizer Variable Bypass

Flow over wick

Out of System

Temperature Compensated by automatic flow alteration

Concentration Calibrated

Agent Specific

Page 36: anaesthesia Vaporizers tec1 to tec5
Page 37: anaesthesia Vaporizers tec1 to tec5

TEC-5

Page 38: anaesthesia Vaporizers tec1 to tec5
Page 39: anaesthesia Vaporizers tec1 to tec5
Page 40: anaesthesia Vaporizers tec1 to tec5

Construction:

The control dial is at the top, the dial must be pushed in before the

vaporizer can be turned on.

At the rear is a locking lever that is connected to the control dial so

that the vaporizer cannot be turned on until it is locked on the

manifold.

At the bottom right front is a sight glass that shows the liquid agent

level in the vaporizing chamber.

Page 41: anaesthesia Vaporizers tec1 to tec5

When the dial is turned past zero, inflowing gas is split into

two streams by the rotary valve.

One stream is directed to the vaporizing chamber, the other

through the bypass.

Page 42: anaesthesia Vaporizers tec1 to tec5

The internal baffle system is designed to keep liquid from

reaching the outlet if the vaporizer is tipped or inverted.

When the concentration dial is in the zero position, all of the gas

from the flowmeters bypasses the vaporizer through the select a

tec bar.

Page 43: anaesthesia Vaporizers tec1 to tec5

Gas flowing through the bypass flows down one side of the

vaporizer and past the thermostat,which is bimetallic strip in the

base.

As the temperature in the vaporizer decreases,the thermostat

permits less gas flow through the bypass so that more gas passes

through the vaporizing chamber.

Page 44: anaesthesia Vaporizers tec1 to tec5

From the thermostat, gas flows up the other side of the

vaporizer and near the outlet joins the gas that has passed

through the vaporizing chamber.

Gas flowing to the vaporizing chamber first passes through

the central part of the rotary valve after which it is directed

through a helical channel then past a spiral wick that is in

contact with the wick skirt, which dips into the liquid agent.

Page 45: anaesthesia Vaporizers tec1 to tec5

Gas with vapour leaves the vaporizing chamber via a channel in the concentration dial rotary valve and flows to the outlet.

Page 46: anaesthesia Vaporizers tec1 to tec5

Benefits of TEC 5:

Helical IPPV assemble to minimize effects of positive pressure ventilation.

Volatile Agent capacity increased from 125ml to 300ml

One Handed Dial and more obvious off position

Internal Baffle system designed to prevent liquid from reaching the outlet if vaporizer is tipped or inverted.

Page 47: anaesthesia Vaporizers tec1 to tec5

Tec 5 vaporizers are available with either of two filling devices:

Keyed system

Funnel fill Funnel Fill System Vaporizer Components The vaporizer filling components include a funnel and cap. When the cap is removed, liquid can be poured into the vaporizing chamber through the funnel.

When the full level is reached, liquid will accumulate in the funnel, and no more liquid will enter the vaporizer.

A funnel-fill vaporizer can be converted to an agent-specific keyed filling system by the addition of an adaptor that screws into the vaporizer filler.

Page 48: anaesthesia Vaporizers tec1 to tec5

Bottle Component

A color-coded adaptor is available to aid the filling process.

At one end is a connector with a screw thread to match the thread on the bottle and a skirt that extends beyond the screw threads.

It has slots that match the projections on the bottle collar. The adaptor for a different agent than the adaptor is intended for will not screw on either because of different threads or bottle opening size or because the projections will not line up with the slots on the adaptor.

Page 49: anaesthesia Vaporizers tec1 to tec5

Filling Procedure

The filler cap is removed by turning it counterclockwise. Agent is poured into the filling port until the level reaches the full mark.

The level may fall slightly as the wicks absorb the agent. The filler cap is then securely replaced.

Draining Procedure

To drain the vaporizer, a container is placed under the drain. Removing the filling cap will usually speed drainage.

The mechanism for opening the drain varies with the vaporizer. After draining is complete, the drain plug should be replaced and tightened to minimize leaks.

Page 50: anaesthesia Vaporizers tec1 to tec5

Keyed Filling System The keyed filling system has been used by many manufacturers for a variety of anesthetic agent

Vaporizer Component The vaporizer filler receptacle (filler socket or block, vaporizer filler unit) permits only the intended bottle adaptor to be inserted.

The receptacle should be sealed when the bottle adaptor is not inserted.

A metal filler block (plug) may be used. There may be a single port for both filling and draining or two ports, the upper one for filling and the lower one for draining .

Page 51: anaesthesia Vaporizers tec1 to tec5

Bottle Component Each bottle of liquid anesthetic has a color-coded collar attached securely at the neck .

The colors are red for halothane, orange for enflurane, purple for isoflurane, and yellow for sevoflurane.

These colors are also used on the bottle labels. Bottle adaptors are also color coded. At one end, the adaptor has a connector with a screw thread to match the thread on the bottle and a skirt that extends beyond the screw threads and has slots that match the projections on the bottle collar.

At the other end is the male connector that fits into the vaporizer filler receptacle.

Page 52: anaesthesia Vaporizers tec1 to tec5
Page 53: anaesthesia Vaporizers tec1 to tec5

Filling To fill a vaporizer, the cap from the bottle is removed, and the bottle adaptor is screwed to the bottle until tight.

If the connection is not tight, the vaporizer may be overfilled or a leak may occur.

The vaporizer should be turned OFF before proceeding further. The plug, if present, is removed. The bottle component is then inserted with the groove matching that on the vaporizer receptacle.

During insertion, the bottle should be below the inlet level. After the filler block is inserted, the retaining device is tightened, and the fill valve (vent) is opened.

The bottle is then held higher than the filler receptacle so that liquid enters the vaporizer

Gentle up-and-down motion may help to clear air bubbles and facilitate filling

Page 54: anaesthesia Vaporizers tec1 to tec5

Draining

To drain the vaporizer, the bottle adaptor is attached to an appropriate bottle.

In the dual-port filler, the bottom socket is used.

The filler plug is removed, the male adaptor inserted, and the retaining device tightened.

The bottle is held below the receptacle, and the drain (spool) valve is opened.

After the vaporizer is drained, the drain valve is closed, the retaining device loosened, and the bottle adaptor removed. The filler plug should be reinserted and the retaining device tightened

Page 55: anaesthesia Vaporizers tec1 to tec5

Evaluation of Tec 5 Vaporisers

Accuracy is maximum at a flow of 5lit/min with a dial setting <3%.

Greatest accuracy is between 15˚ and 35˚c

<15˚C output will be less

>35˚C output will be high

Prone to Increase in output due to pumping effect

Carrier Gas Composition effects the output

At low flows output is less if air or nitrous oxide is used as compared to when oxygen is used

Page 56: anaesthesia Vaporizers tec1 to tec5

Maintenance:

The exterior of the vaporizer may be wiped with a damp cloth.

The vaporizer should be drained every 2 weeks or when the level is low.

The vaporizer should be returned to a service center every 3 years.

Page 57: anaesthesia Vaporizers tec1 to tec5

Hazards:

If the locking lever is partially loose or the filling port open, there will be a gas leak.

Tilting the vaporizer can result in overfilling

Reversed flow through the vaporizer results in increased output.

Page 58: anaesthesia Vaporizers tec1 to tec5

References:

1.Dorsch and Dorsch: Understanding Anaesthesia equipment;5th edition.

2.Wards

3.Barash Clinical Anesthesia;7th edition.

4.Chakravarti S, Basu S. Modern anaesthesia vaporizers. Indian J Anaesth 2013;57:464-71.

5.Internet references.

Page 59: anaesthesia Vaporizers tec1 to tec5

THANK YOU