Journal of Automatic Chemistry of Clinical Laboratory Automation, Vol. 7, No. 2 (Apr-Jun 1985), pp. 99-101 Evaluation of an automated haemolytic method for the determination of anti-streptolysin O antibodies F. Parri and E. de Majo Laboratorio di Batteriologia Virologia, Ospedale Careggi U.S.L. 10D, Via della Quiete 8, Firenze, Italy 1. Introduction Streptolysin O (SO) is an exotoxin produced by the majority of group A, C, and G beta-haemolytic strepto- cocci, Groups C and G, however, rarely cause diseases in man, so finding high anti-Streptolysln O (ASO) titer is an index of group A beta-haemolytic streptococcal infection. This is of great clinical help because serious complica- tions may follow this kind of infection. ASO titer determination is a currently used laboratory procedure for the diagnosis of streptococcal infections; many authorities agree that high ASO titers will be found in about 80% of streptococcal infections. The ASO titer is generally carried out by the haemolytic method of Rantz and Randall or modifications of this method, which requires a series of steps (sample inactiva- tion, preparation of a series of sample dilutions, prepara- tion of a 5% rabbit erythrocyte suspension) which limit its usefulness. Further, it has a critical point due to the poor stability of reduced SO in the presence of oxygen with consequent loss of lytic activity. Recently, more useful methods have been devised. Some kits on the market make use of the agglutination of latex particles or formol-treated erythrocytes as support for SO. Other kits (Aso-Quantum, Sclavo, I 53100 Siena, Italy, and Taso-tec/Taso-matic Diesse, Diesse, I 53035 Monteriggioni, Italy) are based on the haemolytic method and use oxidized SO, which is without lytic activity and hence capable of binding the specific antibodies. The antigen-antibody reaction is detected after the addition of a reducing agent, which makes the unbound SO once again capable of lysing the erythrocytes [2, 3 and 4]. Aso-Quantum is a manual method which may be used with whole blood since it employs the patient’s own erythrocytes as detectors of the antigen-antibody reac- tion. This means, however, that the determination must be performed within 24 h of the blood sample being taken. Since laboratories are not always in a position to do this, modifications of the original method have been developed which allow the use of the patient’s serum together with human (O group and Rh negative) or rabbit erythrocytes [5 and 6]. Taso-tec/Taso-matic is an automated method [7], which can be used both on undiluted whole blood or serum; in the latter case, fresh human erythrocytes are used as detectors by adding them directly to oxidized SO. The rate of haemolysis is followed photometrically on a Taso-matic instrument in which a control containing only SO, erythrocytes and reducing reagent allows more accurate results and greater standardization to be obtained. In the presence of known levels of antibodies a series ofsigmoid curves are obtained with slopes inversely proportional to the concentration of antibodies (see figure 1). It is therefore possible to establish a correlation between the time required to reach a 10% reduction of the initial absorbance and the level of antibodies in the sample [7]. Taso-matic equipment takes a first reading at to (the time after the reducing agent is automatically added) and then measures until tx is reached (the time at which the absorbance becomes 90% of the initial value). According to the time between to and t,, the instrument calculates the ASO titer of each sample and prints them directly in International Units (IU). The possibility of introducing a control serum at each analysis cycle allows a true quality-control programme to be carried out. The Taso-matic can take up to 17 serum sample cuvettes and performs automatic stirring, thermoregulation, addition of the reducing agent to each cuvette, data storage, 3.6" 3.2. 2.8. 2.4. 0"40-i0 to 14 16 20 tTt2 ta t, ts t6 Time (min) Figure 1. Haemolysis curves for samples as known ASO titer. Where to time when the reducing reagent was added and l, 12, 14, 15, 16 time at which a 10% decrease in the initial absorbance value due to haemolysis was obtained, for samples having 200, 300, 400, 500, 600 and 700 IU of ASO tiler. 99
4

an method for the of anti-streptolysin O antibodiesdownloads.hindawi.com/journals/jamc/1985/874231.pdf · F. ParriandE.deMajoAnautomatedhaemolyticmethodfor anti-streptolysin 0antibodies,,,oo

Jul 07, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: an method for the of anti-streptolysin O antibodiesdownloads.hindawi.com/journals/jamc/1985/874231.pdf · F. ParriandE.deMajoAnautomatedhaemolyticmethodfor anti-streptolysin 0antibodies,,,oo

Journal of Automatic Chemistry of Clinical Laboratory Automation, Vol. 7, No. 2 (Apr-Jun 1985), pp. 99-101

Evaluation of an automated haemolyticmethod for the determination ofanti-streptolysin O antibodies

F. Parri and E. de MajoLaboratorio di Batteriologia Virologia, Ospedale Careggi U.S.L. 10D, Via dellaQuiete 8, Firenze, Italy

1. Introduction

Streptolysin O (SO) is an exotoxin produced by themajority of group A, C, and G beta-haemolytic strepto-cocci, Groups C and G, however, rarely cause diseases inman, so finding high anti-Streptolysln O (ASO) titer is anindex ofgroup A beta-haemolytic streptococcal infection.This is of great clinical help because serious complica-tions may follow this kind of infection.

ASO titer determination is a currently used laboratoryprocedure for the diagnosis of streptococcal infections;many authorities agree that high ASO titers will be foundin about 80% of streptococcal infections.

The ASO titer is generally carried out by the haemolyticmethod of Rantz and Randall or modifications of thismethod, which requires a series ofsteps (sample inactiva-tion, preparation of a series of sample dilutions, prepara-tion of a 5% rabbit erythrocyte suspension) which limitits usefulness. Further, it has a critical point due to thepoor stability of reduced SO in the presence of oxygenwith consequent loss of lytic activity.

Recently, more useful methods have been devised. Somekits on the market make use of the agglutination of latexparticles or formol-treated erythrocytes as support forSO. Other kits (Aso-Quantum, Sclavo, I 53100 Siena,Italy, and Taso-tec/Taso-matic Diesse, Diesse, I 53035Monteriggioni, Italy) are based on the haemolytic methodand use oxidized SO, which is without lytic activity andhence capable of binding the specific antibodies. Theantigen-antibody reaction is detected after the addition ofa reducing agent, which makes the unbound SO once

again capable of lysing the erythrocytes [2, 3 and 4].

Aso-Quantum is a manual method which may be usedwith whole blood since it employs the patient’s own

erythrocytes as detectors of the antigen-antibody reac-tion. This means, however, that the determination must

be performed within 24 h ofthe blood sample being taken.Since laboratories are not always in a position to do this,modifications of the original method have been developedwhich allow the use of the patient’s serum together withhuman (O group and Rh negative) or rabbit erythrocytes[5 and 6].

Taso-tec/Taso-matic is an automated method [7], whichcan be used both on undiluted whole blood or serum; in

the latter case, fresh human erythrocytes are used asdetectors by adding them directly to oxidized SO. Therate of haemolysis is followed photometrically on a

Taso-matic instrument in which a control containingonly SO, erythrocytes and reducing reagent allows moreaccurate results and greater standardization to beobtained. In the presence of known levels of antibodies a

series ofsigmoid curves are obtained with slopes inverselyproportional to the concentration ofantibodies (see figure1). It is therefore possible to establish a correlationbetween the time required to reach a 10% reduction ofthe initial absorbance and the level of antibodies in thesample [7]. Taso-matic equipment takes a first reading at

to (the time after the reducing agent is automaticallyadded) and then measures until tx is reached (the time at

which the absorbance becomes 90% of the initial value).According to the time between to and t,, the instrumentcalculates the ASO titer of each sample and prints themdirectly in International Units (IU). The possibility ofintroducing a control serum at each analysis cycle allowsa true quality-control programme to be carried out. TheTaso-matic can take up to 17 serum sample cuvettes andperforms automatic stirring, thermoregulation, additionof the reducing agent to each cuvette, data storage,

3.6"

3.2.

2.8.

2.4.

0"40-i0to

14 16 20

tTt2 ta t, ts t6Time (min)

Figure 1. Haemolysis curves for samples as known ASO titer.Where to time when the reducing reagent was added and l, 12,14, 15, 16 time at which a 10% decrease in the initial absorbancevalue due to haemolysis was obtained,for samples having 200, 300,400, 500, 600 and 700 IU ofASO tiler.

99

Page 2: an method for the of anti-streptolysin O antibodiesdownloads.hindawi.com/journals/jamc/1985/874231.pdf · F. ParriandE.deMajoAnautomatedhaemolyticmethodfor anti-streptolysin 0antibodies,,,oo

F. Parri and E. de Majo An automated haemolytic method for anti-streptolysin 0 antibodies

, ,,oo 1 ; : ,oo

Z

soo 0 soo’1

nnmIIII

300 300

1102

i:ll:iiml

m66

200

<200

mmm

292

mmmmmmmmmmmmmmmmmmmmmmmmmmmm mmmmmmmmmmmmmmmmmm

250

200

<200 200 300 500 800 1600

RANTZ RANDALLIU <200 200 300 500 800 1600 IU

TASO-TEC/TASO-MATIC

Table 1. Correlation between the Rantz and Randall method andTaso-tec/ Taso-matic automated method.

Table 3. Correlation between Taso-tec/Taso-matic and Aso-Quantum methods.

1600

800

500

300

20O

200

mmmmmmmmmmm m9o

mmmmmmmmmmm

mmmmmmmm

mmummmmmmmmmmmmmmmmmmmmnmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmm)

,,,.,mlll:.,.,,,,.mmmmmm

m250

<200 200 300 500 800 1600

RANTZ RANDALL

Table 2. Correlation between the Rantz and Randall andAso-Quantum methods.

processing, and printing-out of results in IUs within 30min. The Rantz and Randall method requires a serialdilution of each sample, while Aso-Quantum needs serialdilution of SO. Taso-tec/Taso-matic method performseach determination in one cuvette with a single dose ofSO.

Materials and methods

The evaluation was done on 600 serum samples sent tothe laboratory for ASO titer determination. The Rantz

and Randall method was used as a refirence and theTaso-tec/Taso-matic method using human erythrocytes,was compared to the reference and to the Aso-Quantummethods utilizing rabbit erythrocytes. All the methodswere performed according to the manufacturers’ tech-nical instructions.

Results

Table compares the results obtained with the Rantz andRandall and Taso-tec/Taso-matic methods. Statisticalanalysis of the results shows an excellent agreementbetween the two procedures (y 22"23 + 0"95x; r- 0-96),indeed the correlation coefficient r 0"96 does not differsignificantly from (p<0"05).

Table 2 shows the correlation between the Rantz andRandall and the Aso-Quantum methods. Statisticalanalysis of the results indicates less agreement betweenmethods (y 127"15 + 0"42x; r 0"73); the value of theintercept (127"15) and the value of the slope (0"42)indicate that the Aso-Quantum gives lower results thanthe reference method at high levels, and higher results atlow levels.

Table 3 compares the results obtained with the Taso-tec/Taso-matic and Aso-Quantum methods. Statisticalanalysis of the results shows a poor correlation betweenthe methods (y 121"76 + 0.43x; r 0.73) and indirectlyconfirms the good correlation between the reference andthe Taso-tec/Taso-matic methods.

Discussion

The well correlated results with the reference method andthe easy-to-use equipment, mean that the Taso-tec/Taso-

100

Page 3: an method for the of anti-streptolysin O antibodiesdownloads.hindawi.com/journals/jamc/1985/874231.pdf · F. ParriandE.deMajoAnautomatedhaemolyticmethodfor anti-streptolysin 0antibodies,,,oo

F. Parri and E. de Majo An automated haemolytic method for anti-streptolysin 0 antibodies

matic system is useful for clinical laboratories as a validalternative to the traditional, manual method for deter-mining ASO titer. Important assets are its ease ofuse andspeed and the possibility ofperforming a complete qualitycontrol program.References1. RANTZ, L. A. and RANDALL, E., Proceedings of the Society of

Experimental Medicine, 59 (1945), 22.2. ALOUF, J. E. and RAYNAtD, E. M., Annales de l’Institut Pasteur,

114 (1968), 812.

3. BEINn.XtdER, A. W., Biochemica & Biophysica Acta, 344(1974), 27.

4. B.xtqHM., A. W. and AwA), L. S., Infection and Immunity,1 (1970), 509.

5. RlccI, A., BERTI, B., MOAURO, C., PORRO, M., NERI, P., andTAL, P., Journal of Clinical Microbiology, 8 (1978), 263.

6. CUALBJ DF.z, G., Quaderni Sclavo di Diagnostica, 16 (1980),312.

7. RlccI, A. Hemolytic method for the kinetic determination ofantistreptolysin 0 antibodies in blood or serum samples, using oxidizedstreptolysin 0 (U.S. Patent No. 4, 379,850).

35th CANADIAN CHEMICAL ENGINEERING CONFERENCE

Calgary, Alberta, 6-9 October 1985

Organized. by the Canadian Society for Chemical Engineering/Socit Canadienne du G(nie Chimique, theconference will be held at the Calgary Convention Centre, and will consist of seven concurrent technical andgeneral-interest sessions. The papers will cover a wide range of topics from fundamentals to industrialapplications of chemical engineering. There will also be sessions relevant to the chemical, process, and energyindustries..Several sessions, including one on government relations, will include invited speakers. The economicand Business Management Division (EBM) of the Chemical Institute of Canada is co-sponsoring andorganizing.several sessions on forecasts, forecasting and planning, petrochemicals, and the business side oflargeprojects.

Technical sessions at the conference are planned on the following subjects:

BiotechnologyBusiness side of large projects (EBM)Chemical engineering fundamentals with applicationsChemical processingCoal, oil and tar sandsCogenerationComputer aided designComputer controlEntrepreneurs in chemical engineeringEnvironmental opportunitiesEnvironmental regulationsForecasts, forecasting and planning (EBM)Government relationsPetrochemical outlook (EBM)Plastics and materialsThe gas plant industryUse of PCs in chemical engineeringUtilization of methane.

Further information from Roger M. Butler, Department of Chemical and Petroleum Engineering, University of Calgary,Calgary, Alberta T2N IN4, Canada. Tel.: 403 284 7133.

101

Page 4: an method for the of anti-streptolysin O antibodiesdownloads.hindawi.com/journals/jamc/1985/874231.pdf · F. ParriandE.deMajoAnautomatedhaemolyticmethodfor anti-streptolysin 0antibodies,,,oo

Submit your manuscripts athttp://www.hindawi.com

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Inorganic ChemistryInternational Journal of

Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

International Journal ofPhotoenergy

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Carbohydrate Chemistry

International Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Chemistry

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Advances in

Physical Chemistry

Hindawi Publishing Corporationhttp://www.hindawi.com

Analytical Methods in Chemistry

Journal of

Volume 2014

Bioinorganic Chemistry and ApplicationsHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

SpectroscopyInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

The Scientific World JournalHindawi Publishing Corporation http://www.hindawi.com Volume 2014

Medicinal ChemistryInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Chromatography Research International

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Applied ChemistryJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Theoretical ChemistryJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Spectroscopy

Analytical ChemistryInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Quantum Chemistry

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Organic Chemistry International

ElectrochemistryInternational Journal of

Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

CatalystsJournal of