AMPTEC 601ES Explosive Safety DMM– Operator Training 2 W ire 4 W ire + HIG H RESEARCH AM PTEC 601ES EXPLOSIVE SAFETY DIGITAL M U LTIM ETER STRAY CURRENT PRESS TO MEASURE milliammeter 0 - 10 mA DC PRESS TO MEASURE CASE GROUND 2.0 K OHM 20 O HM 200 O HM 2.0 O HM R ESISTA NC E 20 K OHM LO W V I9999 >>> O R LO BAT V A .... PO W ER ON OFF STR AY C U R R EN T FU SE (5 A M PER E) 13 13 VLO VHI 50mV - 5mA SIG NA L SOURCE D IS P LAY BACKLIGHT C A LIB R ATED 601ES 601ES-503 0 5 A PR 2 0 0 5 0 5 A PR 2 0 0 6 I FUNCTIO N / RANG E 2 A MP A C/DC 600V R M S 200V AC/DC 20V AC/DC 2.0V AC/DC A C -D C V O LTA G E VRMS press /hold B attery C om partment 4 each "A A " A lkaline or N i-M H rechargeable 2 Wi re T erm inal s
19
Embed
AMPTEC 601ES Explosive Safety DMM– Operator Training
AMPTEC 601ES Explosive Safety DMM– Operator Training. AMPTEC 601ES Explosive Safety Safety DMM – Startup Check Out Procedure. The first item to examine is to check the calibration sticker due date for the AMPTEC 601ES Explosive Safety DMM. If the due date is still valid and not - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
AMPTEC 601ES Explosive Safety DMM– Operator Training
2 Wire
4 Wire
+ HIGH
RESEARCHAMPTEC
601ES EXPLOSIVE SAFETYDIGITAL MULTIMETER
STRAY CURRENTPRESS TO MEASURE
milliammeter 0 - 10 mA DC PRESS TO MEASURE
CASEGROUND
2.0 KOHM
20 OHM
200 OHM
2.0 OHM RE
SIS
TA
NC
E
20 KOHM LOW
V
I9999>>> OR
LO BATVA
. . . .
POWER ON
OFF
STRAY CURRENT FUSE ( 5 AMPERE)
13
13
VLO
VHI
50mV - 5mA
SIGNAL SOURCE
DISPLAYBACKLIGHT
CALIBRATED
601ES601ES- 503
05APR200505APR2006
I
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OL
TAG
E
V
RM
S
press/hold
Battery Compartment4 each "AA" Alkaline
or Ni-MH rechargeable
2 WireTerminals
AMPTEC 601ES Explosive Safety Safety DMM –Startup Check Out Procedure
The first item to examine is to check the calibration sticker due date for the AMPTEC 601ES Explosive Safety DMM. If the due date is still valid and not expired, it Is ready to begin the safety check procedure. If the calibration due date is expired then contact your calibration laboratory or AMPTEC RESEARCH. A tamper resistant calibration sticker is usually located around the perimeter areato the AMPTEC 601ES DMM covering one of the calibration access screws.
AMPTEC 601ES Explosive Safety Safety DMM –Startup Check Out Procedure
POWER ON
OFF
Hold downON
“ ”.
the “POW ER ON” switch ~ (2) secondsfor power The 601ES automatically turns after (30) minutes of idle use to conserve battery power.
“OFF”
Step 1: Procedure for turning on the 601ES DMM – Using main operating power switch.
Step 2: Observe the 601ES DMM display after unit is “powered up”. There should be no low battery (“LO BAT”) indicator present in the LCD display. ( see red circle area below)
The “LO BAT” display means it istime to replace the unit’s “AA” batterries. ( shown without backlight)
Since no “LO BAT” is displayed means,it means the unit’s batterries are OK. ( display shown with backlight lit mode )
The 601ES DMM is capable of indicating : “A” for amperes, “V” for voltage, , “k” for Kilo, “Ω ” for ohms. When the display indicates “>>> OR” or Overrange mode – it means possible “open circuit connection” or a higher range selection is required.
Step 3: Place the unit in “2 wire” measurement mode to begin the self check. (make sure that the 2 wire / 4 wire switch is in the the “up” position).
2 Wire
4 Wire
+ HIGH
LOW
VVLO
VHI
I
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
4 6
5 milliamperelimit point
810
2.0 KOHM
20 OHM
200 OHMR
ES
ISTA
NC
E
20 KOHMFUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- DC
VO
LTA
GE
VR
MS
2.0 OHM
2 sections of the AMPTEC 601ES DMM are used for this process.The FUNCTION/RANGE section and the on-board analog milliammeter.
Step 4: The next set of steps envolves checking (by viewing the analog milliammeter) each range and function to see that less than 0.005 A ( or 5 mA) is ever output from any FUNCTION or selected range, into the unit’s onboard analog milliammeter.
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
4 6
5 milliamperelimit point
810
2.0 KOHM
20 OHM
200 OHMR
ES
ISTA
NC
E
20 KOHMFUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- DC
VO
LTA
GE
VR
MS
2.0 OHM
Step 5: Select the 600 VRMS range (rotary switch in far counterclock position). Then press and momentarily hold down the milliammeter “measure” switch. You should see virtually no movement in the analog milliammeter. (it’s scale is 0 to 10 mA fullscale).
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
4 6
5 milliamperelimit point
810
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
Step 6: Select the 200 VRMS range (rotary switch in 8 O’clock position). Then press and hold the milliammeter “measure” switch. Again , you should see virtually no movement in the analog milliammeter. (it’s scale is 0 to 10 mA fullscale).
The yellow line on range switch is shown for emphasis of position purposes.
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
4 6
5 milliamperelimit point
810
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
Step 6: Select the 20 VRMS range (rotary switch in 10 O’clock position). Then press and hold the milliammeter “measure” switch. Again, you should see virtually no movement in the analog milliammeter. (it’s scale is 0 to 10 mA fullscale).
The yellow line on range switch is shown for emphasis of position purposes.
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
4 6
5 milliamperelimit point
810
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
Step 7: Select the 2.0 VRMS range (rotary switch in ~11 O’clock position). Then press and hold the milliammeter “measure” switch. Again, you should see virtually no movement in the analog milliammeter.
The yellow line on range switch is shown for emphasis of position purposes.
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
4 6
5 milliamperelimit point
810
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OL
TAG
E
V
RM
S
2.0 OHM
Step 8: Select the 2.0 AMP AC/DC range (rotary switch in 12 O’clock position). This step requires simultaneously pressing and holding down 2 buttons. *Press and hold the 2.0 AMP button ( see red arrow below) while you also press and hold the milliammeter “measure” switch. Again, you should see virtually no movement in the analog milliammeter.
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES2
0
4 6
5 milliamperenominal limit point
8
10
Step 9: Select the 2.0 OHM range (rotary switch in the 1 O’clock position). Press and momentarily hold down the milliammeter “measure” switch. You should see the analog milliammeter needle deflect to to midscale or the 5 mA (verticle needle position is “OK” – the milliammeter scale is 0 to 10 mA fullscale). You should not see 5 ½ milliamperes, avoid paralax viewing error of the milliammeter by taking reading straight over the milliammeter - use one eye to view. ( 8 mA is the failsafe current level).
2.0 KOHM
20 OHM
200 OHMR
ES
ISTA
NC
E
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
milliammeter 0 - 10 mA DC PRESS TO MEASURE
MILLAMPERES
Step 10: Select the 20 OHM range (rotary switch in the 2 O’clock position). Press and momentarily hold down the milliammeter “measure” switch. You should see the analog milliammeter needle deflect to to midscale or the 5 mA level. It should be the same as the 2.0 ohm current source level. A verticle needle position is “OK” – the milliammeter scale is 0 to 10 mA fullscale. Again, you should not see 5 ½ milliamperes, avoid paralax viewing error of the milliammeter by taking reading straight over the milliammeter - use one eye to view.
MILLAMPERES2
0
4 6
5 milliamperenominal limit point
8
10
milliammeter 0 - 10 mA DC PRESS TO MEASURE
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
Step 11: Select the 200 OHM range (rotary switch in the ~3 O’clock position). Press and momentarily hold down the milliammeter “measure” switch. You should see the analog milliammeter needle deflect to barely ~1 mA level. You should not see any level even close to 5 ½ milliamperes.
2.0 KOHM
20 OHM
200 OHM
RE
SIS
TAN
CE
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
MILLAMPERES2
0
4 6
5 milliamperenominal limit point
8
10
milliammeter 0 - 10 mA DC PRESS TO MEASURE
Step 12: Select the 2.0 KOHM range (rotary switch in the ~ 4 O’clock position). Press and momentarily hold down the milliammeter “measure” switch. You should see the analog milliammeter needle deflect to barely any current or ~<1 mA level. You should not see any level even close to 5 ½ milliamperes.
2.0 KOHM
20 OHM
200 OHMR
ES
ISTA
NC
E
20 KOHM
FUNCTION / RANGE
2 AMP AC/DC
600VRMS
200V AC/DC
20V AC/DC
2.0V AC/DC
AC
- D
C V
OLT
AG
E
V
RM
S
2.0 OHM
Step 13: Select the 20 KOHM range (rotary switch in the 6 O’clock position). Press and momentarily hold down the milliammeter “measure” switch. You should see the analog milliammeter needle indicate barely any current at all or ~<1 mA level. You should not see any level even close to 5 ½ milliamperes.
MILLAMPERES2
0
4 6
5 milliamperenominal limit point
8
10
milliammeter 0 - 10 mA DC PRESS TO MEASURE
If the unit has successfully completed all the steps indicated it is ready to use.
This concludes the AMPTEC 601ES Explosive Safety DMM check out procedure.
1. Calibration/Maintenance access screws
2. Purge Valve
3. Handles
4 On/Off
-
ONProtection Fuse
6 LCD Display AV
k >>>ORif
6A. LCD Display "LO BAT"
7. Display Backlight
8. Meter Test Lead Input Jacks
V VV V all
9. Zero.
sealed with tamper resistance calibration sticker (see red arrow).
- Automatic Air Pressure Equalization
- See item #1, .
. Power “ ” switch, silicon rubber boot covered, operates on 4 “AA” batteries (alka-line or NiMH). Automat ically turns “OFF” after (30) min. of in-activity. Located near shoulder of handle to avoid accidental power turn .
(5.0 amp fast blow type) protects the 2.0 ampere Stray Current Range (measurement). The type of fuse used is a Little Fuse 3AG or equivalent. . 20,000 count with ( ) amps,
( ) voltage,( ) Ohms, ( ) kilo symbols plus ( ) overrange indicator a range’s magnitude is exceeded.
The will indicatewhen it is time to replace the batteries.
Replace (4) ea. “AA” batteries with a heavy duty 1.5 V Alkaline i.e. Duracell Ultra, orrechargeable Energizer 2500 mAHr 1.2V Nickel Metal Hydride rechargeable or equiva-lent.
is push button activated (dim ambient light areas). Note: Use sparingly to conserve battery power.
- For two wire test lead connections, you should use the far left input jacks both
and , shown as a dotted box. Use these two terminals and for volt-age, current and simple ohms measurements. The “4 Wire” switch position and terminals are only for 4 Wire Kelvin resistance measure-ments.
'Short' the test lead ends prior to adjustment of the Zero knob only for resis-tance measurements. Readjust of the Zeroknob is only needed if connections are changed. This is a precision 10 turn pot with delicate stops at both ends (i.e. used with 2 Wire 2 ohm, 20ohm and 200 ohm ranges).
only lift for calibra-tion access
LOLO
5.
HIHI
10. Case Chassis Ground Jack
11. RMS AC/DC
12. Stray Current
13 0 to 10 mA DC Analog MilliammeterPress to
Measure”
Press to Measure14
or is connected to the meter’s internal Faraday Cage, metal top plate and the unit’s bottom plate (an accessible chassis to earth ground connection).
: 600.0V,200.0V, 20.0 V & 2.0 VRMS AC/DC (
4 ½ digit resolution).
. measures DC current output levels with the "
button. To test the AMPTEC 601ES resistance function's test current level select the resistance range ( rotary knob) and press the milliammeter's " " button.
. AMPTEC 601ES Safety DMM Resistance Ranges include a 2.0 Ohm range with 0.1 milliohm resolution, 20.0 Ohm, 200 Ohm, 2.0 KOhm and 20 KOhm ranges.
Voltage Measurement i.e. 0.0500
VRMS measurement on the 2.0 VRMS range -
Press To Measure - To measure AC/DC current, connect input and output wires from the UUT to the far left 2 terminals ( / LO inputs). Connect to the “UUT”. Depress and Hold the Stray Current button. A 5.0 ampere fuse protects the meter from current than 5 amps.
The function is used to verify DVM voltage measurement (i.e. the range) and stray current measurement . Select the 2.0 V Range. Depress and hold the signal source button. The display should indicate approx. 0.0500 V around 50 mV . Next, the range.
Depress and hold the button the button. The display should
indicate approx. 0.0050 amperes ( ~ 5 mA ). Place palm on the lid corner. Press
while flipping up the latch to open the AMPTEC601ES DMM.