Top Banner
ambr ® analysis module
4

ambr analysis module - Sartorius · sensor calibration and subsequent in-process re-calibration. The analysis module eliminates the operator time needed for manual pH sample transfer,

May 28, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: ambr analysis module - Sartorius · sensor calibration and subsequent in-process re-calibration. The analysis module eliminates the operator time needed for manual pH sample transfer,

ambr® analysis module

Page 2: ambr analysis module - Sartorius · sensor calibration and subsequent in-process re-calibration. The analysis module eliminates the operator time needed for manual pH sample transfer,

2

Add-On Capability for Calibration of pH in ambr 15 Micro Bioreactors

The analysis module provides add-on capability for measurement of pH in ambr 15 micro bioreactors, for the purpose of initial bioreactor vessel pH sensor calibration and subsequent in-process re-calibration. The analysis module eliminates the operator time needed for manual pH sample transfer, measurement, data transfer and bioreactor pH sensor recalibration, by fully automating this process.

Operation of the analysis module is supported by user replaceable sensors (electrodes) and reagent kits including calibration solutions, cleaning solutions and a waste container. analysis module is suitable for use with both cell culture and microbial media. analysis module is not suitable for use with particulates (for example microcarriers will block internal tubing) or highly viscous cultures or samples (e.g. filamentous microbes).

The benefits

− Improve pH control and culture performance with frequent, accurate pH sensor recalibration

− Minimal user interaction enables more frequent pH checks and frees up scientist time

− Small sample volume allows more frequent pH checks and/or more culture volume for other samples

− Improves accuracy and consistency through full automation of sample handling and assay read

− Heater sensor cell enables measurement temperature to match the culture temperature, improving measurement accuracy

− Reduced CO2 outgassing effects and errors compared to manual offline samples

− Reduce utilisation of offline pH analyser, increasing capacity for other assays

− Compact size increases the ambr 15 system width by less than 25 cm

Specifications

Optimal working range pH 6.0 - 8.0

Sample volume (ambr 15) 60 µL

Cycle time per read 90 s

Resolution 0.01 pH

Buffer accuracy ±0.01 pH

Page 3: ambr analysis module - Sartorius · sensor calibration and subsequent in-process re-calibration. The analysis module eliminates the operator time needed for manual pH sample transfer,

3

Performance

− Accurate, precise, linear response when reading a range of pH buffers

− Y-axis: ambr 15 analysis module reading

− X-axis: pH buffer reference value

− Accurate, precise measurement of CHO culture pH

− ambr 15 CHO R&D test culture at Sartorius Royston

− N = 24 bioreactors, ~ 106 cell/mL at day 6

− Reference measurement: Mettler electrode inserted into ambr 15 vessel

7.5

7.4

7.3

7.2

7.1

7.0

6.9

6.8

6.7

6.60 1 2 3 6

Day of CHO culture

pH

analysis module

electrode in vessel

10.0

9.0

8.0

7.0

6.0

5.0

4.04.0 5.0 6.0 7.0 8.0 9.0 10.0

y = 0.9998xR2 = 0.9999

7.5

7.4

7.3

7.2

7.1

7.0

6.9

6.8

6.7

6.60 1 2 3 6

Day of CHO culture

pH

analysis module

electrode in vessel

10.0

9.0

8.0

7.0

6.0

5.0

4.04.0 5.0 6.0 7.0 8.0 9.0 10.0

y = 0.9998xR2 = 0.9999

Page 4: ambr analysis module - Sartorius · sensor calibration and subsequent in-process re-calibration. The analysis module eliminates the operator time needed for manual pH sample transfer,

Consumables

Operation of the analysis module is supported by replaceable sensors (electrodes) and reagent kits including calibration solutions, cleaning solutions and a waste container. A reagent kit typically provides sufficient reagents for up to 3 months operation. Replacement of electrodes and reagent kit is recommended every 3 months for optimum measurement accuracy.

pH and reference electrodes Part no. 001-4B80

Reagent kit, pH Part no. 001-4B60

Specifications subject to change without notice. Publication No. : SBI1558-e160401 Order No. : 85037-554-79 Ver. 04 | 2016

Sartorius Stedim Biotech GmbHAugust-Spindler-Strasse 1137079 Goettingen, GermanyPhone +49.551.308.0Fax +49.551.308.3289www.sartorius-stedim.com

USA Toll-Free +1.800.368.7178UK +44.1372.737159France +33.442.845600Italy +39.055.63.40.41Spain +34.90.2110935Russian Federation +7.812.327.5.327Japan +81.3.4331.4300

Specifications subject to change without notice. Printed and copyrighted by Sartorius Stedim Biotech GmbH. | WPublication No.: Order No.: Ver. 01 | 2012