Top Banner
Albumin Analysis by Potentiometric Method
19

Albumin Analysis by Potentiometric Method. Results.

Dec 13, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Albumin Analysis by Potentiometric Method. Results.

Albumin Analysis by Potentiometric Method

Page 2: Albumin Analysis by Potentiometric Method. Results.

Results

Page 3: Albumin Analysis by Potentiometric Method. Results.

0

5

10

15

20

25

0 2 4 6 8Conc.(ppm)

E(m

V)

WE: Cu (sputter)

Page 4: Albumin Analysis by Potentiometric Method. Results.

-4-3.5

-3-2.5

-2-1.5

-1-0.5

0

0 5000 10000 15000 20000T(sec)

E(m

V)WE: graphite (sputter)

Page 5: Albumin Analysis by Potentiometric Method. Results.

-4.5

-4

-3.5

-3

-2.5

-2

-1.5

-1

-0.5

0

0 2000 4000 6000 8000T(sec)

E(m

V)

WE:graphite + polymer film(DMAEMMC/TEGDMA:1/19, adhesive:epoxy resin, initiator:benzoin)

Page 6: Albumin Analysis by Potentiometric Method. Results.

-600

-550

-500

-450

-400

0 2000 4000 6000 8000 10000 12000T(sec)

E(m

V)

WE:氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/25,adhesive:TEGDMA, initiator:benzoin)

Page 7: Albumin Analysis by Potentiometric Method. Results.

-470

-460

-450

-440

-430

-420

0 5000 10000 15000 20000

T(sec)

E(m

V)

WE: 氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/12, adhesive:epoxy resin, initiator:CHP+ TMEDA)

Page 8: Albumin Analysis by Potentiometric Method. Results.

0

5

10

15

20

25

0 1 2 3 4 5Conc.(ppm)

E(m

V)

WE:氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/12,adhesive:epoxy resin, initiator:CHP + TMEDA)

Page 9: Albumin Analysis by Potentiometric Method. Results.

-470

-450

-430

-410

-390

-370

0 5000 10000 15000 20000 25000T(sec)

E(m

V)

WE:氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/35, adhesive:epoxy resin, initiator:CHP+ TMEDA)

Page 10: Albumin Analysis by Potentiometric Method. Results.

-6

-4

-2

0

2

4

6

8

0 2 4 6 8 10 12 14Conc.(ppm)

E(m

V)

WE:氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/35, adhesive:epoxy resin, initiator:CHP + TMEDA)

Page 11: Albumin Analysis by Potentiometric Method. Results.

-420

-410

-400

-390

-380

-370

-360

0 5000 10000 15000 20000T(sec)

E(m

V)

WE:氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/25, adhesive:epoxy resin, initiator:CHP + TMEDA)

Page 12: Albumin Analysis by Potentiometric Method. Results.

0

1

2

3

4

5

6

0 1 2 3 4 5 6Conc.(ppm)

E(m

V)

WE:氧化鋁板 +polymer film(DMAEMMC/TEGDMA:1/25, adhesive:epoxy resin, initiator:CHP + TMEDA)

Page 13: Albumin Analysis by Potentiometric Method. Results.

Adhesion Testing

Page 14: Albumin Analysis by Potentiometric Method. Results.

WE:氧化鋁板 + polymer film(DMAEMMC/TEGDMA: 1/35, adhesive:epoxy resin, initiator:CHP + TMEDA)

Page 15: Albumin Analysis by Potentiometric Method. Results.

WE:氧化鋁板 + polymer film(DMAEMMC/TEGDMA:1/35, adhesive:epoxy resin, initiator:CHP + TMEDA)

Page 16: Albumin Analysis by Potentiometric Method. Results.
Page 17: Albumin Analysis by Potentiometric Method. Results.

Conclusions

Page 18: Albumin Analysis by Potentiometric Method. Results.

1. 就 crosslinker而言 , 親油性的 (TEGDMA)優於親水性 (PEG400DMA).

2. 白蛋白的濃度改變與電位變化之間具有線性關係 .

3. Epoxy resin 所得實驗結果優於 TEGDMA.

4. Polymer film 與氧化鋁板間會有局部膨脹與脫落現象 ,影響實驗結果 .

Page 19: Albumin Analysis by Potentiometric Method. Results.

END