Top Banner

of 41

AL 100user Manual Extract

Jul 06, 2018

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/17/2019 AL 100user Manual Extract

    1/41

    Be sure to read this Operation Manual 

    throughout before using the AL-100 for your 

    correct and safe use.

    If you have any questions or comments on the

    operation of this Instrument, please contact 

    your local representative.

    Do not use procedures other than speci- 

    fied in this Operator Manual.

    Place this Operator Manual in a place of 

    access to the Operator while in opera- 

    tion.

    If you lose this Operator Manual, ask 

    your local representative for a new copy.

  • 8/17/2019 AL 100user Manual Extract

    2/41

    I O L

    M E  AS U R E 

    E D I T  

    2 COMPONENTS 

    2.1 Front and right side of AL-100

    2-1

  • 8/17/2019 AL 100user Manual Extract

    3/41

    2-2 

    (1) Screen(touch panel)

    Measurement data and other information are displayed on the screen. The instrument is operated by

    touching the keys.

    (2) Probe holder

    When the probe is not in use, place them in the holder.

    (3) Terminal of Biometry Probe

    The Biometry probe is connected here.

    (4) Power indicator

    Lit when power is on.

    (5) MEASURE Button

    Switch display to Measurement mode

    (6) IOL Calculation ButtonSwitch display to IOL Calculation Mode

    (7) EDIT button

    Switch display to EDIT Mode

    (8) PC card slot

    Memory card (Optional) should be inserted to store acquired data.

    (9) Printer

    Prints measured data and calculated results.

     (10) Biometry probe

    Use for axial length measurement.

  • 8/17/2019 AL 100user Manual Extract

    4/41

    2-3 

    2.2 Back and left side of AL-100

  • 8/17/2019 AL 100user Manual Extract

    5/41

    (1) RS-232C terminal

    The RS-232C cable is connected here.

    (2) Power supply terminal

    The power cord is connected here.

    (3) Fuse holders

    Fuses are in use in these holders.

    (4) Power switch

    Press the “ I ” side of the switch to turn on the power. Press the “ O ” side to turn it off.

    (5) Contrast adjuster

    Controls the contrast of the screen

    (6) Power cord with 3-prong plug

    Connection of this cord to a 3-prong power receptacle provides power to the AL-2000.

    (7) Meintenance Switch

    Switch for Manufacturer’s use only. Do not change its position.

    (8) Footswitch Terminal

    The footswitch is connected here.

    (9) Power supply terminal for chin rest fixation light

    The power plug for the fixation light for the chin rest.

    2-4

  • 8/17/2019 AL 100user Manual Extract

    6/41

    3-1

    3 SETUP 

    3.1 Safety precautions

     ALWAYS cleans the probe tip before taking a measurement on a human

    eye.

     NEVER use the probe if there is any visible damage to its tip. Such use may

    cause an incorrect maeasurement and/or damage to the cornea.

    This instrument is designed exclusively for ophthalmic use. DO NOT use

    the instrument for any purpose other than ophthalmic use.

     DO NOT use any cables or memory cards other than those specified in this

    manual. Such use may result in damage to the instrument.

  • 8/17/2019 AL 100user Manual Extract

    7/41

    3-2 

    3.2 Preparing the instrument for use

    3.2.1 Connection of accessaries

    a) Connecting the biometry probe

    Connector should be plugged into the terminal with a

     proper direction.

    Plug the biometry probe connector (1) into the terminal (2)

    labeled BIO on the front side of the instrument. When it is

     properly insetrted, you will hear it click.

    b) Connecting the power cord

     Power cord should be plugged into the terminal with a

     proper direction.

    Plug the power cord connector (1) into the terminal (2) on

    the back of the AL-100.

     

    O

     

    O

  • 8/17/2019 AL 100user Manual Extract

    8/41

    3-3

    c) Connecting the foot switch

    Connector should be plugged into terminal with a proper 

    direction.

    1) Plug the connector (1) for the foot switch to the terminal

    (2) labeled FOOT SW on the back of the AL-100. Line

    up the slot of the connector with the tab on the terminal.

    2) To source the connector, turn the locking ring (3) until

    you hear it click into place.

    d) Inserting the probe holder

    The holder should be installed onto the unit with a proper 

    direction.

    Insert the protruding rectangular tabs (1) on the probe

    holder into the holes (2) on the right ide of the instrument.

    DO NOT insert the holder upside down.

    Push inward on the center of the probe holder (3) and pull

    it downward to lock it in place.

  • 8/17/2019 AL 100user Manual Extract

    9/41

    3-4

    3.2.2 Connection of the optional parts

    a) Inserting and removing the memory card

     Before using memory card for the first time, set the bat-

    tery in accordance with the instruction Manual for 

     Memory Card (located in the box that contains the catd).

    Follow the instructions for BN-HSR series.

     Since the memory card pops quickly out of the slot when

    removed, hold the edge of the card when pressing the

    button for card removal.

    “Inserting the car”

    Hold the front of the memory card (1) toward you and in-

    sert it into the slot (2) until the card is aligned with the

    removal button.

    “Removing the car”

    Remove the memory card (1) by pressing the button (2).

     DO NOT INSERT or REMOVE the memory card, when

    data is loaded from or saved into the memory card.

  • 8/17/2019 AL 100user Manual Extract

    10/41

    3-5 

    b) Chin rest installation (AL-1100)

    An optional chin rest with a fixation light is available for use

    when measuring axial length.

    “ Installing the chin rest” 

     Make sure that the slider moves.

    1) From the operator’s side, carefully insert the biometry

     probe into the slider until it locks into place. (Be careful

    not to damage the eye contact area of the probe.)

    2) Coil the probe cord twice (as shown) and secure it with

    the cord hook (2). The cord length (3) from the probe to

    the hook should be approximately 20 cm to prevent

    tension on the cord when the slider is moved forward

    during the measurement.

    Figure 1 Figure 2

  • 8/17/2019 AL 100user Manual Extract

    11/41

    3-6 

    “ Connecting the fixation lamp power plug” 

    Insert the power plug (1) for the chinrest fixation lamp into

    the terminal (2) labeled “FIX LIGHT” on the back of the AL-

    100.

  • 8/17/2019 AL 100user Manual Extract

    12/41

    3-7 

    3.3 Biometry

    3.3.1 Biometry Mode Setup

    a) Turning the power on and initial adjustments

     ALWAYS clean the probe tip before taking a measurement on a human eye.

     NEVER use the probe if there is any visible damage to its tip. Such use may

    cause an incorrect measurement and/or damage to the cornea.

     As with any ultrasound instrument, it is recommended that we exposure be

    kept as low as reasonably achievable. (ALARA)

     Before turning on the power 

    *Make sure that the power plug is properly connected to

    the receptacle.

    *Make sure that the biometry probe is properly con-

    nected.

    *If the power turns off, wait for over one mimute to turn

    it on again.

    1) Turning on the power switch (1) at the upper left side of the

    instrument initiates self-checking of the probe. If the

     biometry probe is not detected (i.e., no probe or wrong probe

    connected), the message “Probe error!” will be presented.

    2) Adjust the contrast of the screen with the contrast adjuster 

    (2).

  • 8/17/2019 AL 100user Manual Extract

    13/41

    3-8 

    3) Tilt the screen forward by pulling out the “legs” (3) on

    the bottom of the instrument, if desired.

    b) Start up screen

    1) The probe is automatically calibrated, when the screen is

    shown.

    2) The catalog screen (Ready to measure screen) is shown,

    when the calibration is completed without any problem.

  • 8/17/2019 AL 100user Manual Extract

    14/41

    3-9

    3) When “Probe error!” is shown on the area (1), “OK” key

    will be shown on the area (2). Press OK to go to the

    catalog screen. The measurement cannot be done even

    on the catalog screen, if “Probe error!” is displayed. See

    Troubleshooting if “Probe error!” shows many times.

    Contact your TOMEY Representative, if the problem

    remains.

  • 8/17/2019 AL 100user Manual Extract

    15/41

    3-10 

    3.3.2 Guide to use Front buttons

    1) MEASURE

    The catalog screen (Ready to measure) screen is

    shown, when pushing this button. The settings with In-

    dex data input, selecting the eye to be measured, Con-

    tact/Immersion and selection of Measurement methods

    are also done with this button.

    2) IOL

    The IOL calculation screen appears, when pushing this

     button. IOL power calculation, Personal Correspon-

    dence value, IOL registration can be done with this

    function. The results can be printed out with the built-in

     printer.

    3) EDIT

    The EDIT screen appears, when pushing this button.

    Waveform, measured data, Caliper, Utility, Data Save/

    Load and print out can be done with this function.

    BIO METER AL-100

    MEASURE IOL EDIT

    MEASUREButton

    IOLButton

    EDITButton

  • 8/17/2019 AL 100user Manual Extract

    16/41

    3-11

    3.3.3 Measurement

    Press MEASURE button to go into measurementmode.

    Widow will be switched from Data re-

    view to measurement automatically,

    when it is taking data.

    a)Data Review Window / Measurement Window

    a-1 Data Review Window

    a-2 Measurement Window

  • 8/17/2019 AL 100user Manual Extract

    17/41

    3-12 

    3.3.4 Setting of the measurement conditions

     Press “New” key and delete data of both eyes, before tak-

    ing new patient. Otherwise two patients’f data may be

    mixed up and cause serious probelm.

    a) Setting the eye to be measured

    Touch Eye key to select the eye you wish to measure (right

    or left).

    Axial length data and IOL power calculation result of each

    eye are aquired with different data storage.

     Please make sure the display is showing right eye, when

    measuring right eye, and left eye for left.

     Figure 1

    1) The right eye (OD) or the left eye (OS) is selected alternately each time the Eye key is touched.

  • 8/17/2019 AL 100user Manual Extract

    18/41

    3-13 

    c) ID / Patient’s name, sex / Physician’s name input

    Index Data (ID / Patient’s name, sex / Physician’s name)

    should be input on the screen.

    b) New Patient

    Press the button (1) to measure new patient.

    Figure 1

    1) Press (1) button (Figure 1) until you hear “beep” sound.

    2) New screen is ready to take new patient data.

    Former ID / Patient’s name, sex / Physician’s name are

    deleted. K1 and K2 for IOL calculation are also

    deleted.

    Eye type will be set as “Normal”, and Eye to be measured 

    will be set as “Right” eye.

    Figure 1

  • 8/17/2019 AL 100user Manual Extract

    19/41

    3-14

    1) Touch the Index key (1) on the measurement window

    (Fig.1 )to display the Name/ID entry window. (fig.2)

    2) Touch the number keys and the alphabetic character 

    keys (2) to enter the patient name, ID number and 

     physician’s name. If the physician’s name is already

    registered in the physician list, it can be selected by

     pressing the corresponding number. “.” and ”space”

    cannot be used as ID number.

    3) To switch between letters and numbers, use the switch-over key (3). In the event of an erroneous entry, touch

    the Delete key.

    4) Male/Female will be selected with pressing icon buttons

    on the key board.

    5) To set the entered value and advance to the next entry,

    touch the Return key (5). The cursor location indicates

    the category of the next entry.

    6) Touch “Exit” key to return to the measurement window.

    Setting the values for meausrement

    Eye type / Gain / Contact Immersion mode / Measurement

    mode can be selected as settings for measurement.

    Setting the eye type and converted velocity

    Figure2

  • 8/17/2019 AL 100user Manual Extract

    20/41

    3-15 

    d) Physician’s list Entry

    Physician’s name can be stored in advance as a list.

    Figure1

    1) Touch Entry key to reverse the color of “Entry” icon.

    [Figure1]

    2) Touch PhyList, which you wish to change and it reverses

    the color.

    3) Type the name with the alphabetic character keys.

    4) Touch the “Enter” key again to save change.

    5) Touch the “Phy. List” to put the name as the physician.

  • 8/17/2019 AL 100user Manual Extract

    21/41

    3-16 

    e) Setting items for measurement

    (Figure 1)

    1 Touch Eye type/ Gain key(1) to go into its setting

    window. “Figure2”

    (Figure 2)2 Select the Key (2) for measured eye which suits the

    eye to be measured. (See the next page.).

    3 In case of changing the sound velocity, select the

    sound velocity setting key to display the number keys

    (5) (as highlighted). In (Fig. 3)

    4 Select the key, then change the entry item and set the

    sound velocity.. (As for the entry range, see the next

  • 8/17/2019 AL 100user Manual Extract

    22/41

    3-17 

     page.)

    (Figure 3)

    5 When (3) is again selected, the highlighted display is

    released and the screen is returned to that in (Fig. 2).

    6 If the EXIT key (4) is selected, the screen will be

    returned to that shown in (Fig. 1).

  • 8/17/2019 AL 100user Manual Extract

    23/41

    3-18 

    [Setting the eye type and converted velocity]

    The AL-100 can be set to measure the following tyoes of 

    eyes:

     Normal

    Select when the lens nucleus of the patient’s eye is rather 

    soft, e.g., incipient cataract.

    Average velocity

    Average axial length velocity (Avg) : 1500 ~ 1600 m/s

    Lens velocity(LENS) : 1540 ~ 1740 m/s

    Anterier chamber velocity(ACD) : 1430 ~ 1630 m/s

    Dense Cataract

    Select when the lens nucleus of the patient’s eye is rather 

    hard e.g., hypermature cataract. Hard to measure with Nor-

    mal mode.

    Average axial length velocity (Avg) : 1500 ~ 1600 m/s

    Lens velocity(LENS) : 1540 ~ 1740 m/s

    Anterier chamber velocity(ACD) : 1430 ~ 1630 m/s

    Aphakic

    Select when the patient’s eye is aphakic.

    Average axial length velocity (Avg) : 1430 ~ 1630 m/s

    Pseudophakic

    Select when the patient’s eye is pseudophakic.

    IOL velocity (LENS) : 800 ~ 3000 m/s

    Vitreous velocity (Vit) : 800 ~ 2000 m/s

    Anterier chamber velocity (ACD) : 1430 ~ 1630 m/s

    IOL thickness (Thickness) : 0.10 ~ 4.00 mm

  • 8/17/2019 AL 100user Manual Extract

    24/41

    3-19

    [ACD] and [LENS] are not displayed with Aphakic, and 

    [LENS] is not displayed with Pseudophakic.

    Posterier lens waveform might not be recognized, due to mul-

    tiple echo with Dense Cataract mode.

    Measured data is displayed.

    Measured data may not be displayed.

    Measured data is not displayed.

    Setting of the gain (ultrasound emitter power)

    Set the gain level based on the waveform amplitude. There

    are 8 levels. The higher you set the gain, the larger the am-

     plitude will be.

    Figure 1

    1 Touch (1) keys to adjust the gain with the window

    (Fig. 1).

    2 New value is displayed on main window.

  • 8/17/2019 AL 100user Manual Extract

    25/41

    3-20 

    Setting of the Contact / Immersion

    Contact / Immersion mode can be changed with touching

    “Contact” or “Immersion”.

    Figure 1

    1 Select Contact with contact mode and Immersion with

    immersion mode.

    2 Selected key should be highlighted.

    [How to apply biometry probe for Contact mode]

    The tip of the probe should be applied to the center of the

    cornea verticaly.

  • 8/17/2019 AL 100user Manual Extract

    26/41

    3-21

    [Measurement of axial length with immersion mode]

    Measurement without using immersion attachment (cap)

     In case of using corneal protective gel, care must be

    taken not to put excessive amount of the agent, which

    may otherwise influence measurement data.

    Measurement with immersion attachment

    Care must be taken not to let air bubbles in the cup pro-

    vided at the tip of the Immersion Attachment, when plac-

    ing the ultrasound gel.

    1 Apply a small amount of the ultrasound gel to the tip of 

    the Biometry probe. (Fig. 1)

    2 Cover the b iometry probe wi th the immersion

    attachment.(Fig.2)

    3 Put the ultrasound gel in the cup provided at the top of 

    the immersion attachment.(Fig3)

    4 Apply the immersion attachment to the probe so that its

    tip is aligned to the probe axis and the visual axis, but it

    is positioned in such a place where the tip of the

    immersion attachment does not touch the cornea. (Fig 4)

    Hand, Chin Measurement mode

    1 Select the key, (1) key or (2) key, depending on the

    measurement method.

    2 The selected key is displayed as reverse highlighted.

    The converted velocity directly affects the measurement.

     Ascertain that you have set the desired converted veloc-

    ity.

     In immersion mode, the ultrasound gel must intervenebetween the eye contact area of the biometer probe and 

    the cornea so that distance between the probe and the

    cornea is approximately 1.8 to 3.2 mm.

     Excessive ultrasound gel may affect the accuracy of the

    measurement.

     Auto measurement (Hand mode) assists the examiner to

    take measurement, and the function itself does not diag-

    nose the data.

     Measurement accuracy might be over 0.1mm, due to the

    measurement condition or the difference among pa-

    tients.

    Figure 1

    Figure 2

    Figure 3

    Figure 4

  • 8/17/2019 AL 100user Manual Extract

    27/41

    3-22 

    Figure 1

    Setting of the measurement mode

    The Hand, chin measurement (automatic measurement) or 

    the manual measurement can be selected.

  • 8/17/2019 AL 100user Manual Extract

    28/41

    3-23 

    3.3.5 Actual measurement

    a) Checking the performance

    The biometry test piece is used only for checking the op-

    erating performance of the instrument. It cannot be used  for determining the precision of the instrument or for 

    calibration of the instrment.

    Check the performance of the AL-100 by using the biometry test piece(found in the box containing the biometry probe).

    1 Select the following settings:

      Eye type/converted velocity: Aphakic/1,532 m/s

      Gain: 8  Measurement Mode: Hand-Held 

    2 Apply a drop of water to the upper and to the lower test piece surfaces. Apply the biometer perpendicular to theupper surface.

    b) Preparation for measurement

     If the patient is relaxed and cooperative, the measure-

    ment will be easier and more successful. Prior to taking 

    measurements, explain the purpose and method of mea-

     surement to the patient to reduce his/her anxiety.

    1 Confirm that the measurement conditions have been set.

    2 Anesthetize the eye with an appropriate topical

    anesthetic.

    3 When taking measurements using the chin rest, instruct

    the patient to sit in front of the chin rest. Adjust the

    height of the chair, the lift table and/or the chin rest to

    maximize patient comfort.

    When taking measurements holding the probe by hand,

    instruct the patient to be seated or to lie supine.

    4 If the fixation light is used to guide the direction of 

    gaze(when using the chin rest), set it at an appropriate

    height for the patient to fixate.

  • 8/17/2019 AL 100user Manual Extract

    29/41

    3-24

    c) Measurement methods

    There are two automatic modes, hand-held and chin rest,

    and one manual mode.

    Hand (Hand-held)

    1 Touch “New” key(1) to measure new patient’s eye.

    Apply biometery probe perpendicular to the cornea.

    2 The measurement window will be automatically shown,when certain waveform conditions are satisfied.

    Figure 1

  • 8/17/2019 AL 100user Manual Extract

    30/41

    3-25 

    3 When a sa tisfactory measurement is taken, the

    instrument makes a “beep” sound.

    4 The instrument aquires the data automatically, when the

    data becomes stable.

    The instrument makes a “beep” sound, when a data is

    aquired.

    5 The instrument shows the waveform, average axial

    length and maximum(as L) / minimum(as S)

    measurement data.

    6 When data acquisition is complete, a higher-pitched 

     beep sounds and the waveform for the measurement

    closest to the average value is displayed.

    Using the hand-held method, up to 15 measurements in therange of +/- 0.2 mm from the average value are taken. If 

    measurement data are largely scattered, “ERROR” is dis-

     played.

    When “ERROR” is displayed, the probe may not be ap-

     plied properly. Touch “Retake” key for two seconds and 

    take the measurement again.

    Figure 2

  • 8/17/2019 AL 100user Manual Extract

    31/41

    3-26 

    Chin (Chin measurement)

    Select “Chin” for taking measurement using the Chin Rest.

    The operation and display of the Measure, Setting screen

    and the Measure screen are same as those done for Hand 

    (hand measurement). (See Figs. 1 and 2.)

    1) Measurement is performed by using the Chin Rest.

    2) Assemble the Biometry Probe in accordance with “3.3.1

    c) Assembling of the Biometry Probe”. Give a sufficient

    reach of the Probe cord between the Probe and the cord 

    hook, which may otherwise apply an undue pressure

    against the cornea.

    3) When taking measurement, apply the Probe in such a

    manner that the planes of the slider moving plane and the

    fixed part are aligned. Insufficient application of the

    Probe may cause to measure a slightly longer axial length

    due to the corneal protective gel and tears. Care must also

     be given not to apply an excessive force against the cornea

    with the Probe.

    4) Measurement must be continued until all the ten data

    come within the range of +/-0.1mm against the average

    value.

    Measurement does not have to be completed, but can be

    stopped on its way.

    Manual

    1) When taking measurement from a new Patient in the

    screen shown in Fig. 1, select the New key (1) and apply

    the Probe to the cornea at a right angle.

    2) When the Instrument has been fulfilled with waveform

    conditions to a certain extent, the screen is changed to that

    for measurement.

    3) When the Instrument has been fulfilled with the taking-

    in conditions, a monitor sound of “pee” is given.

    4) For manual measurement, data taking is given by using

    the footswitch in (Fig. 5)

    5) If the footswitch is again pressed, the operation will

     proceed in the successive measurement.

    6) Measurement will be followed in the same manner as

    above. When ten measured data have been taken in, mea-

    surement is finished with “beep” sound.

    Figure 4

    Figure 3

    Figure 5

  • 8/17/2019 AL 100user Manual Extract

    32/41

    3-27 

    d) Deleteing all data for one eye

    1 To delete all data currently displayed for either the right

    or the left eye, press the Retake key (1) until you hear a

     beep.

     Note: This deletes currently acquired data from the instrument memory; it does not delete data stored on

    the memory card.

    e ) Selecting the location of the retina waveform

    In the event that a spike occurs between the back of the lens

    and the retina (e.g., due to a vitreous hemorrhage), manu-

    ally reposition the retinal cursor (2) by pressing the retinal

    cursor shift keys (1).

    The waveform to the right of this cursor location is taken as

    the retinal waveform.

    1 Move cursor by pressing move keys to locate it on the

    left side of the actual retina.

    Figure 1

  • 8/17/2019 AL 100user Manual Extract

    33/41

    3-28 

    2 The waveform to the right of this cursor(2) is taken as

    the retinal.(Fig2)

    3.3.6 Acceptable waveforms

    a) Contact mode

    In the automatic measurement modes, waveforms are evaluated 

    and acquired only if the following criteria are met:

    1. The following crests rise above the level cursor:

     Normal:

    The spikes of the back and front of the lens and of theretina.

    Dense Cataract:

    The spike of the front of the lens and of the retina.

    Aphakic:

    The spike of the retina.

    Pseudophakic1 ~ 3:The spikes of the front of the IOL and of the retina.

    2. The retinal waveform stands upright.

    3. The variability among measurements is low.

    Figure 2

    Retina Select Cursor

    Retina

    Uprightwaveform

    Sclera

    A valley between Retina and Sclera

  • 8/17/2019 AL 100user Manual Extract

    34/41

    3-29

    The following items are used for checking to see if ultra-

    sound can correctly catch the axial length and, therefore,

    reasonable waveforms are obtainable, provided that these

    items are not good conditions for taking-in of measure-

    ment data.

    i The retinal echoes rise high.: implies that the probe is

    applied perpendicular to the cornea.

    ii The echo rises high in the front and at the rear of the

    crystalline lens: implies that the axial length was caught

    with the probe.

    iii The retinal and sclera echoes are distinctive.: implies

    that the probe is applied at a right angle. In canse of high

    gain, the drop (choroid) between two echoes cannot be

    identified, which is not always necessary to be detected.

    iv No tail waves followed after corneal echos.: implies that

    the probe directly touches the cornea. If there is any tear 

    or corneal protective gel left between the cornea and the

     probe, the corneal echoes are followed by tail waves. If 

    so, the measurement of axial length may not be stable or 

    may be longer than its actual length.

    b) Immersion mode

    For immersion mode, the following conditions are added 

    to those of Contact mode.

    The cornea front echoes must be within the range of 1.8

    and 3.2 mm (which is shown as the range of dotted line in

    the left figure.)

    The following items are not to indicate the conditions of 

    data acquisition, but to confirm if acquired waveform is

    acceptable or not.

    Also confirm “a Contact mode i - iii ”.

    i No unnecessary echoes arise if there is air voids

    included in the ultrasound gel used in the inside of the tip

    of the immersion attachment or between the probe and 

    the cornea.

    Initial Echo

    Lens front surface echo

    Retina echo

    Lens rear surface echo

    Corneal echo

  • 8/17/2019 AL 100user Manual Extract

    35/41

    3-30 

    3.3.7 EDIT

    Press EDIT button on the front, and show EDIT window.

  • 8/17/2019 AL 100user Manual Extract

    36/41

    3-31

    a) Data display and delete

    The waveforms for each eye of the current patient mea-

    surements can be retrieved and reviewed at any time dur-

    ing the measuring process.

    “Displayng a waveform”

    1 If you wish to review the waveform for a particular 

    measurement, select the measurement by pressing the up

    and down arrow keys. The selected data will be

    highlighted and the corresponding waveform will bedisplayed. (Fig1)

    2 Press Select key (2) to select the data.

    3 Press Echo key (3) to show Echo data (Fig2).

    When pressing Echo without selecting any data, the data

    closest tothe average will be displayed. Average data will 

    also be displayed as Ave. AXIAL.

    Figure 1

    Figure 2

  • 8/17/2019 AL 100user Manual Extract

    37/41

    3-32 

    [Deleting individual measurements]

    1 Touch the up and down arrow keys (2) to move to the

    measurement you wish to delete.

    2 Touch Select key to select the data you wish to delete.

    Touch Delete/Recall key to delete the selected data.

    At this time, touch Delete/Recall key again to recover the

    data.

    When switching to the other window, deleted data can

    not be recovered by touching Delete/Recall key again.

    Figure1

  • 8/17/2019 AL 100user Manual Extract

    38/41

    3-33 

    b) Caliper Function

    The distance between two points can be measured on the

    waveform.

     A reading displayed in the caliper mode is an

    appoximation and may be different from the actual read-

    ing.

    1 After measurement, touch the Caliper key (1) to switch

    to Echo window. (Fig2)

    2 Press the right or left arrow key (2) to set the point from

    which to start measuring, which will be indicated by avertical dashed line. Touch the Fix key (3) to set the

    location.

    Figure 1

    Figure 2

  • 8/17/2019 AL 100user Manual Extract

    39/41

    3-34

    3 Touch the right or left arrow key again to set the

    second measurement location, which will be indicated 

     by a second vertical dashed line.

    4 The distance between the two points will be displayed 

    in real time. (Fig3)

    Figure 3

  • 8/17/2019 AL 100user Manual Extract

    40/41

    8-1

    8 SPECIFICATIONS 

    8.1 Specifications for the Instrument

    8.1.1 Biometry Function and IOL Calculation Function

    Measurement range

     Axial length: 15:00 - 40.00 mm

     Anterior depth: 1.80 - 7.00mm

    Crystalline lens thickness: 2.00 - 6.00mm

    Instrument accuracy

    Measuring accuracy: ± 0.1mm

    Resolution: 0.01mm

    Converted sound velocity set when shipped

    Normal

     Average axial length sound velocity: 1,550m/s

    Crystalline lens sound velocity: 1,641m/s

     Anterior depth sound velocity: 1,532m/s

    Dence

     Average axial length sound velocity: 1,548m/s

    Crystalline lens sound velocity: 1,629m/s

     Anterior depth sound velocity: 1,532m/s

     Aphakic

     Average axial length sound velocity: 1,532m/s

    Pseudophakic

    IOL sound velocity (Pseudo1): 2,718m/s

    IOL sound velocity (Pseudo2): 1,049m/s

    IOL sound velocity (Pseudo3): 2,200m/s

    Vitreous body sound velocity: 1,532m/s

     Anterior depth sound velocity: 1,532m/s

    IOL thickness (Pseudo1): 0.8mm

    IOL thickness (Pseudo2): 1.0mm

    IOL thickness (Pseudo3): 0.8mm

  • 8/17/2019 AL 100user Manual Extract

    41/41

    IOL calculations

    SRK ll equation

    SRK/T equationHOLLADAY equation

    HOFFER Q equation

    HAIGIS optimized equation

    HAIGIS standard equation

    Biometry Probe

    Type: Solid type

    Fixation lamp: Built in the Probe (red LED)

    Oscillator frequency: 10MHz

    Tip diameter: 5.3mmφ  (Concaved)

    Dimensions and weight: 8mmφ × 97mm/30g

    8.1.2 Main unit

    STN liquid crystal: 5.7 inch, color  

    Dimensions and weight: W220 × D222 × H275mm/4kg

    8.1.3 Power source

    Voltage: 100VAC

    Frequency 50/60Hz

    Consumption power: less than 35VA + 15%

    * The specifications and outer appearance may be changed for improved performances.