A

Agilent ChemStore C/S

Concepts Guide

Page 2: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

2 Agilent ChemStore C/S Concepts Guide

Notices© Agilent Technologies, Inc. 2002, 2004

No part of this manual may be reproduced in any form or by any means (including elec-tronic storage and retrieval or translation into a foreign language) without prior agree-ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Microsoft ® - Microsoft is a U.S. registered trademark of Microsoft Corporation.

Manual Part NumberG2181-90010

Edition03/04

Printed in Germany

Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn, Germany

Software RevisionThis handbook is for a module of the Agilent ChemStation Plus software. In the section “In This Guide…" on page 3 you will find details about the module and the revision of the module that this book is intended to be used for.

WarrantyThe material contained in this docu-ment is provided “as is,” and is sub-ject to being changed, without notice, in future editions. Further, to the max-imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par-ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec-tion with the furnishing, use, or per-formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep-arate agreement shall control.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor-dance with the terms of such license.

Restricted Rights LegendIf software is for use in the performance of a U.S. Government prime contract or subcon-tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer soft-ware” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclo-sure of Software is subject to Agilent Tech-nologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June

1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.

Safety Notices

CAUTION

A CAUTION notice denotes a haz-ard. It calls attention to an operat-ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi-tions are fully understood and met.

Page 3: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 3

In This Guide…This book introduces you to the main concepts of Agilent ChemStore C/S revision B.03.01. It highlights the main features of the product and gives you some guidance for setting up your studies and managing your data. It is organized as follows:

1 Introducing ChemStore C/S

This chapter gives an overview of the Agilent ChemStore C/S functionality to help you to get started.

Information for Operators If you are an Agilent ChemStore C/S operator, and want to learn about the operation of Agilent ChemStore C/S in detail, read Chapter 2, Chapter 3 and Chapter 4.

2 ChemStore C/S Concepts

This chapter describes the main benefits and concepts of using the ChemStore C/S database in the analytical laboratory.

3 Using the Custom Calculator

This chapter gives you an overview on creating and using calculation templates.

4 Using the Report Template Editor

This chapter gives you an overview on creating and using report templates.

Information for Administrators If you are an Agilent ChemStore C/S administrator, and want to learn about setting up Agilent ChemStore C/S to suit the needs of different laboratories and users, read Chapter 5 and Chapter 6.

5 Data Security

This chapter describes the facilities that are available in ChemStore C/S for ensuring the security of the data in the database.

Page 4: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

4 Agilent ChemStore C/S Concepts Guide

6 Data Management

This chapter describes how to manage the data inside your database using back-up, restore and archival.

7 ChemStore C/S Computations

Agilent ChemStore C/S includes a comprehensive suite of statistical calculations. You will find full details of how these calculations are performed in this chapter. It describes statistical calculations, custom calculator commands and error codes, calculations using custom fields and the time computations done by ChemStore C/S.

Page 5: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 5

Contents

1 Introducing ChemStore C/S

ChemStations and ChemStore C/S 12

A Brief ChemStore C/S Tour 13

2 ChemStore C/S Concepts

ChemStore C/S Workflow 18

ChemStore C/S Data Flow 19

Organizing Results 20

Organizing Results into a Study 20Using Custom Fields 21

Transferring Data from the ChemStation 23

Retrieving Data from the Database 24

Setting Up a Query 24Filtering the Dataset 27

Reviewing and Approving Results 29

Run Approval Status 29

The Graphical User Interface 33

Sample View 36Compound View 41Configuring Tables 44Intra-sample Calculations 45Statistics 46Generating Control Charts 47User Interface Settings 48

Transferring Data to the ChemStation 50

Versioning 51

Deleting Runs from the Database 53

Page 6: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

6 Agilent ChemStore C/S Concepts Guide

Contents

3 Using the Custom Calculator

What is the Custom Calculator? 56

Preparing to Use the Custom Calculator Script Editor 57

Retrieving the calculation data set 57Understanding the Custom Calculator Script Editor Views 57

Custom Calculator Script Editor 60

Understanding the Custom Calculator Script Wizards 60Example: Creation of a report calculation script 68Using the Custom Calculator Script Editor 79

4 Using the Report Template Editor

What is the Report Template Editor? 88

Preparing to Use the Report Template Editor 90

Retrieving the report data set 90Understanding the Report Template Editor Tree Views 91

Using the Report Template Editor 93

Using Dialog Boxes to Create Report Sections 93Using Tree Views for Tables and Charts 95

Understanding Report Template Components 96

Page Headers and Footers 96Data Sections 97Section Headers 100Section Elements 100

Understanding Tables 106

Grouping and Summaries for Tables 106Understanding Table Types 108Generating a Report Using a Template 121Using Automatic Reporting 121

Page 7: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Contents

Agilent ChemStore C/S Concepts Guide 7

5 Data Security

Introduction 124

Security 125

Setting Up and Managing Users 127

Setting Up Organizational Information 131

Setting Up and Managing Studies 131Setting Up and Managing Custom Fields 132

Setting Up and Managing ChemStore C/S Elements 134

Archiving and Dearchiving 135

Automated Archiving 135Archiving Interactively 139Archive units 141Deleting Runs 142Reopening Runs 142Dearchiving Runs 143Generic XML Archive Interface 143Understanding Archive Status 145Reviewing Archive Status and Reports 147Taking a Closer Look at the Archive/Dearchive Process 148

Managing Archiving and Dearchiving on Admin Client 151

Overview of the Archive/Delete Process 152Performing Archive/Dearchive Tasks on Admin Client 153

Audit Trail 155

E-Mail Notification 157

Database Logbook 160

Validation and Security of Printed Files 161

File Validation in Operation 161File Validation Functions 162

Page 8: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

8 Agilent ChemStore C/S Concepts Guide

Contents

Replacing the Built-in Security Scheme with a Customized Solution 162

6 Data Management

Backing Up Your Database 166

Back Up or Archive? 167

Backing Up the Stand-Alone Version 168

Windows 2000/XP Backup 169Automating and Scheduling Windows 2000/XP Backup 169Tape Drives 169CD-ROM 170

Backing Up the Client/Server Version 171

Oracle Back-up Types 171

Maintaining the ChemStation Computer 172

Cleaning up Left-Over Temporary Files 172Maintaining the PC’s File System 172Continuous operation 173

Disaster Recovery Plan 175

Disk drive Failure 175Power Failure 175Database is corrupted 175

7 ChemStore C/S Computations

Statistical Calculations 178

Single Value Calculations 178Regression Calculations 179The Linear Regression Model 180Matrix Solution of the Regression 180Statistical and Related Values 181Modifications for Weighted Linear Regression 182Modifications for Non-Linear Functions 183

Page 9: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Contents

Agilent ChemStore C/S Concepts Guide 9

Custom Calculator Commands 185

Error codes and descriptions 194

Calculations using custom fields 198

Original response factor 198List of built in functions 199

Time computation 201

Time synchronization 201Time stamps 201Time zones 202Setting Time zones on the PC 203

Index 205

Page 10: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

10 Agilent ChemStore C/S Concepts Guide

Contents

Page 11: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

11

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

1Introducing ChemStore C/S

ChemStations and ChemStore C/S 12

A Brief ChemStore C/S Tour 13

Page 12: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

12 Agilent ChemStore C/S Concepts Guide

1 Introducing ChemStore C/SChemStations and ChemStore C/S

ChemStations and ChemStore C/S

ChemStore C/S is a data evaluation and storage module for the ChemStation for LC, LC/MS, GC, CE, CE/MS or A/D systems. Based on standard database functionalities, ChemStore C/S adds data storage in and retrieval from a relational database, as well as cross-sample and cross-instrument review calculation and reporting facilities.

After successful installation, and selection of the Connection to Database checkbox, ChemStore C/S adds functionality to the ChemStation. It prompts for a specific user log-in to the ChemStation and adds funtionality to the user interface to handle the transfer of analysis results into the database, reviewing data, transferring it back to the ChemStation for batch review and/or reprocessing and other functions.

The ChemStore C/S client software user interface is installed as a separate application. It allows you to extract information from an ChemStore C/S database using queries, and to create cross-sample calculations, reports and charts.

ChemStore C/S is available as a stand-alone product, operating on a single ChemStation, or as a client/server product. The stand-alone version uses an MS Access database installed on a single ChemStation. The client/server version uses an Oracle database installed on an MS Windows or Unix server. Both versions provide centralized storage of raw data and results along with the metadata (method and sequence files) used to create the results.

In either case, the ChemStore C/S client software and the database communicate via a standard ODBC driver. The ChemStore C/S client software is the same for both configurations; the client/server version has additional facilities for archiving and de-archiving. Both versions track reprocessed runs, maintaining the current version and all previous versions in strict chronological order, and contain facilities for maintaining data integrity by linking the raw data to method and sequence information.

Page 13: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 13

Introducing ChemStore C/S 1A Brief ChemStore C/S Tour

A Brief ChemStore C/S Tour

ChemStore C/S is an add-on database module for your ChemStations for LC, LC/MS, GC, CE, CE/MS and A/D systems.

It provides the following functionality:

Organizing Data

ChemStore C/S is designed to help you to organize your results flexibly and efficiently in a database. You can set up Studies in which you can collect all runs associated with a specific project. Studies can be restricted to certain users to facilitate data security and ease of use. In addition, ChemStore C/S gives you the ability to specify custom fields into which you can incorporate additional data and organizational information associated with the run and/or study.

Transferring Results

The ChemStore C/S software is tightly integrated with your ChemStation software. After you have analyzed your results, you can easily transfer information for a single run, sequence or batch automatically or interactively to ChemStore C/S. In addition to the results, the information transferred can include the chromatograms and spectra, actual raw data, method information and sequence information.

Retrieving Data Sets

Once you have collected data inside the database, you can retrieve the set of data you are interested in by using the Query Builder to easily specify search criteria. The Query Builder allows you to set up queries easily using a graphical interface rather than a complicated SQL (Structured Query Language) statement.

To facilitate your work, you can also store queries for later re-use.

Page 14: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

14 Agilent ChemStore C/S Concepts Guide

1 Introducing ChemStore C/SA Brief ChemStore C/S Tour

Once you have specified the subset you are interested in, (the “query”), you can work on it in all further tasks: review and approval, cross-sample calculations, control charting, transfer to a ChemStation for additional batch processing, and reporting. You can also export the data to other applications (e.g. MS Excel) or a LIMSystem.

Reviewing and Approving Data

The on-line review of data retrieved using a query allows you to inspect a set of analysis results in context. Based on this condensed information, you can use the integrated approval and rejection functions or mark the runs for resubmitting to the ChemStation batch review for more detailed inspection and reprocessing. The approval function allows up to three electronic signatures per result version.

Tracking the Changes

An Audit Trail, maintained independently of the run data, keeps track of all the changes made to the run; you can display or print the Audit Trail and detailed information about the changes made for any run.

Generating Control Charts

To visualize trends in your selected result data, you can use the included control charts. You can view any combination of results in context, with overlaid charts if desired. You can also customize the charts to your own requirements.

Execute customized calculations

A built-in custom calculator offers you the possibility to define your own cross-sample calculations and statistics. The calculation is based on a calculation script that can be easily developed with the help of calculation script wizards.

Versioning is used for storage of those scripts. They can be re-used for all similar data sets. Calculation results and graphics can be included into reports.

Page 15: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 15

Introducing ChemStore C/S 1A Brief ChemStore C/S Tour

Generating Reports

Once you have reviewed and approved your data, you will probably want to create a report. A button on the main toolbar allows you to quickly create a report using one of the standard report templates. Reports can be printed, or saved to a file in one of several formats.

Customizing Reports

If the standard report templates included with ChemStore C/S don’t fully fit your requirements, you can easily modify them using the report template editor. You can also use the report template editor to create your own template designs. Essentially all data that are visible in ChemStore can be included in the report, as well as additional data defined by user-specified calculations.

Exporting to Other Applications

To process your data further, ChemStore C/S allows you to export information to MS Excel, or to print reports to a file in html, xml or csv format. You can also export the data to the Windows Clipboard for use in other Windows applications.

Archiving and Dearchiving Runs

The client/server version also allows you to copy data to an archive and, if appropriate, delete the archived runs from the database. Archived runs that are not deleted from the database are locked, and changes cannot be made to them unless they are reopened. Of course, archived runs that are deleted can also be dearchived and reopened.

Archiving allows long-term storage of data separate from the server database. Limiting the number of active runs in the database is crucial for the database size and performance.

You can archive interactively, on an ad hoc basis, or you can set up to archive automatically at regular intervals. You can use either the internal ChemStore archive utility or a third-party archiving tool based on XML technology. A generic XML archiving utility is available on the installation CD. For further details contact your Agilent Technologies representative.

Page 16: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

16 Agilent ChemStore C/S Concepts Guide

1 Introducing ChemStore C/SA Brief ChemStore C/S Tour

Data Protection and Integrity

To prevent unauthorized access to the database or tampering with results, each ChemStore C/S user is given an access password and a set of permissions that define which facilities are available for that user. Only the Administrator, or other user with the requisite permissions, can change a user’s permissions. If unauthorized access leads to the locking of an account, ChemStore can be set up to notify all affected users by e-mail.

Associated with each run in the database is an Audit Trail, which contains a log of all changes that are made to the run. In addition, a built-in versioning facility ensures that injections that are reprocessed multiple times are maintained in ChemStore C/S as a set and can be retrieved in strict chronological order. You can choose to retrieve all versions (that is, the complete set of reprocessed results) or only the latest; if you retrieve all versions, you can choose to restrict the display to only the latest. The existence of more than one version of the run in the database is denoted by an asterisk in the run table.

In addition, all database interactions that affect security are logged in a database logbook, which, if you are permitted the relevant access level, you can display.

Scalability

The database grows as your requirements grow. You can start with an MS Access based stand-alone system and upgrade to the Oracle based client/server version as you add further systems. A suite of utilities enables you to manage the migration of your stand-alone databases to the client/server version, which gives you the additional advantages of archiving and de-archiving your data, as well as allowing for much larger database sizes. The system grows with your requirements for the organization and reporting of your results, while you preserve your investment in data and training during the process, because the basic user interface does not change.

Page 17: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

17

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

2ChemStore C/S Concepts

ChemStore C/S Workflow 18

ChemStore C/S Data Flow 19

Organizing Results 20

Transferring Data from the ChemStation 23

Retrieving Data from the Database 24

Reviewing and Approving Results 29

The Graphical User Interface 33

Transferring Data to the ChemStation 50

Versioning 51

Deleting Runs from the Database 53

Page 18: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

18 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsChemStore C/S Workflow

ChemStore C/S Workflow

Figure 1 Typical workflow for the operation of ChemStore C/S

Create a Study Set up Sequence

Specify Custom Fields

Assign Study and Custom Field

Set up QueryRun data transferred

Reprocess RunsSample View Compound

viewStatistics

Chemstore C/S ChemStation

Review

Data

Batchtransferred

Reject Runs Accept Runs Mark Runs for Batch

Load ChemStore Batch

Define Report Define Chart

Print Report Print Chart

Page 19: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 19

ChemStore C/S Concepts 2ChemStore C/S Data Flow

ChemStore C/S Data Flow

Figure 2 Data flow in ChemStore C/S

Runs for batch reprocessingSpooler

Run data from the ChemStation and custom field values for the injection

Database

Query

Current data set

Filter

View

Approval Status

ChemStore C/S

ChemStation

Page 20: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

20 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsOrganizing Results

Organizing Results

The central organizational entity in ChemStore C/S is the run. A run is defined as the set of results that is obtained when a data analysis method is applied to the raw data resulting from a single injection. When a run is inserted into the ChemStore C/S database, all the available associated information about the injection, the sample etc. is also entered. This is the only method available for inserting data into the database.

Organizing Results into a Study

Within the ChemStore C/S database, runs can be logically grouped for any desired purpose. For example, it may be appropriate to collect together all the analytical results related to a clinical trial or a stability study. Such a group of runs is referred to as a Study; a user with the requisite permissions (see “Security” on page 125), may define as many studies as desired, but each study must be identified with a unique name. When results are transferred from the ChemStation to ChemStore C/S, the individual runs (or groups of runs) are assigned to a study; these runs, together with any previous runs assigned to the same study, are then available as a logical unit.

You can elect to store additional data along with the result data, for example the chromatogram and/or spectra. You can also optionally store the raw, unprocessed data, the method file and the sequence file, thus ensuring the maintenance of complete data integrity.

To help in the organization and reporting of results, you can associate a set of custom fields (see “Using Custom Fields” on page 21) with the study. These custom fields may be either specific to the study, or they may be used by other studies. Any new custom fields that you specify when you set up the study are automatically available for use in other studies.

When you set up a study that uses a custom field, you must specify how the data for the custom field will be entered. You can choose to enter the data automatically (by a ChemStation macro) or manually. If you choose to enter the data manually, you can specify whether input of a value is always required; if not, you can specify a default value. If you specify that a value is

Page 21: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 21

ChemStore C/S Concepts 2Organizing Results

always required, a valid value must be entered in the field before the transfer of the analysis results to ChemStore C/S is allowed. In addition, you can specify certain characteristics for some field types (maximum and minimum values for numeric fields, maximum string lengths for text fields). The data entry characteristics are specific to the study; if the custom field is used in other studies, its data entry characteristics may be different.

Studies are restricted to a list of assigned users. Only users who have been assigned to work on a study may add or access data for that study. This allows multiple teams to use the same database while being assured that their study data is secure from others who are not authorized to work with their data. Restricting study access also makes the product easier to use because the user only sees data from studies that they are assigned to access. Assignment of studies to users can only be performed by an administrator or other user with the proper ChemStore permission.

Using Custom Fields

In addition to the data normally associated with the analytical results, and provided by the ChemStation, external data can be appended to the database record. Such data are added in the form of Custom Fields, which are user-defined fields containing information of any kind. You can use such custom fields to organize your result data by defining the custom fields to match the kind of analyses you normally perform.

For a preclinical study for example, you may want to set up fields to add information to each set of results about the

• pH value

• Patient’s Name

• Patient’s Age

• Non-Smoker

• Date of Testing

• Patient’s Gender

• Dosage Applied

• etc.

You can then use these data to retrieve a specific set of analysis results, for example all results for a certain patient.

Page 22: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

22 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsOrganizing Results

Custom fields are of six types:

Organizing information in a database is helpful not just for retaining analysis results, with centralized storage, but also for flexible creation of cross-sample and/or cross-instrument reports, calculations and trend analyses on a set of results.

Following the example structure outlined above, you could for example compare results for a certain doses and generate a control chart that displayed the concentration of the active component found.

Custom fields are available on a database-wide basis so that any defined custom field can be appended to a study. Custom fields can have the status active or inactive; setting the status of a custom field to inactive does not affect studies that are configured to use that custom field, but makes the field unavailable for inclusion in any additional studies.

Custom fields used in a study may be configured as always required. You must enter valid values in such fields before the transfer of the analysis results to ChemStore C/S is allowed.

Custom fields are protected by user permissions and passwords (see “Security” on page 125). Only users with the requisite permission may create or modify custom fields, and changes need to be confirmed by electronic signature.

True/False The field can have the values true or false only, for example Non-Smoker would have the values True or False.

Selection list The value is restricted to a specific value from a user-defined list, for Patient’s Gender would offer the values Male or Female.

Integer The field can have an integer value only, for example Patient’s Age would generally be given as an integer.

Real The field can contain any numeric value, for example Dosage Applied might be a non-integer amount.

Text The field can contain text only, for example the Patient’s Name.

Date/Time The field can contain date and time information only, for example the Date of Testing.

Page 23: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 23

ChemStore C/S Concepts 2Transferring Data from the ChemStation

Transferring Data from the ChemStation

Analysis results can be transferred from the ChemStation to ChemStore C/S either interactively or automatically as part of a sequence, in a batch or as single samples. After the run, the data are transferred through a spooler into the database; the spooler operates as a background process so that the ChemStation can continue to run. This process ensures that the ChemStation is not slowed by the transfer process, and guards against data loss if the insertion of the data into the database fails.

Interactive data transfer is done from the Data Analysis view of the ChemStation, either in single data file mode, by reprocessing a sequence or in Batch Review. You can transfer the current data file into any active study which is assigned to you. You provide the necessary custom field values at the time of transfer. You may also change the default values of those custom fields that were given default values at setup.

Automatic data transfer is done either from a sequence or from a batch. Each run in the sequence or batch is assigned to a study (you may use multiple studies and assign each run to a different study), and you provide the values for the custom fields associated with the study during sequence set-up or batch processing.

Any data that has been selected to be saved along with the results (chromatograms, spectra or raw data, method and sequence files) are transferred along with the analysis results.

NOTE For either interactive or automatic data transfer, you must provide values for those custom fields specified as always required before the transfer of data can take place. These fields’ headers are marked with an asterisk.

Page 24: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

24 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsRetrieving Data from the Database

Retrieving Data from the Database

Runs are retrieved from the ChemStore C/S database using a Query (see “Setting Up a Query” on page 24). A query is a request to retrieve a set of runs from the many thousands of runs that comprise your database that matches certain criteria, for example by asking the question “Which samples contain Procaine?” The more exactly you can specify the criteria, the smaller will be the set of runs retrieved. Prior to retrieving the set of runs, ChemStore C/S confirms the number of runs to be retrieved using the current query criteria.

The data that comply with the criteria defined in the query are gathered together and copied to the local ChemStore C/S as a dataset — a snapshot of the data that you want to review (see Figure 2 on page 19).

The way that the runs within the database are organized, and the information that is stored with each run, can greatly increase the efficiency of retrieval of data. Runs assigned to studies designed to match the pattern of work of the laboratory can help to optimize retrieval of analysis results. This is aided by defining a well-structured set of custom fields that best reflect your data search requirements.

For example, if you normally group results by test participant, you can define Test Participant as custom field. The custom fields can then be searched along with the standard search criteria, such as operator name and instrument name, which are by default stored in the database.

Setting Up a Query

ChemStore C/S provides two levels of query: Simple and Advanced. Both levels are set up using a graphical constructor that helps you to define your search criteria easily and without the requirement to learn SQL (structured query language), the standardized language used to specify search criteria in relational databases. The graphical editor allows you to select the data categories you are interested in and either choose pre-existing data items or enter values manually. A study selection is offered to limit the number of existing data items presented in the dialog. The simplest query uses a single item from one category of information, for example “Find the samples from

Page 25: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 25

ChemStore C/S Concepts 2Retrieving Data from the Database

the Study called Beta1”, where “Study” is the data category and “Beta1” is the item. If the simplest query does not produce the runs you require, you can extend the query criteria by using either multiple items or multiple data categories. If you select multiple items within one data category, they are combined in the Simple query with an OR (for example “Find the samples from Study Beta1 OR Beta2”). If you select multiple data categories, they are combined in the Simple query with an AND (for example “Find the samples which contain Procaine AND which are control samples”). The Advanced query extends the query specification with the use of other logical operators (OR, NOT) and the specification of ranges of values (for example “Find the samples which contain Procaine AND which are NOT control samples”).

ChemStore C/S offers two forms of query builder:

• The Simple form (shown in Figure 3) allows you to build up the query by selecting data items and values using the graphical interface. Multiple criteria are ANDed: runs are retrieved that match all of the conditions.

• The Advanced form uses the same graphical interface but adds the facility to extend the criteria using absolute and relative Where clauses, the OR operator and negation. The Where clauses available depend on the data type, as shown in Figure 1 on page 18.

Figure 3 The Query Builder

Possible values of the selected data item

Query specification built up using graphical constructor

Restricts the view of data items below

Select all or latest version of data

Graphical constructor allows you to select data categories to search

Page 26: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

26 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsRetrieving Data from the Database

The NOT operator negates the condition: is not equal, is not greater than, is not less than, is not between and does not contain.

Once the query has been specified, the set of data that is retrieved is available for all subsequent tasks:

• on-screen review and approval

• customized calculations

• generation of control charts

• generation of reports

Because it is likely that you will want to use queries multiple times to retrieve the same dataset, they can be saved and re-run at a later date.

Queries form part of the ChemStore C/S Settings that can be shared between nominated users. Once a query has been specified and saved, it can be assigned to other users in addition to the owner. Queries are also protected from misuse by user permissions (see “Security” on page 125); only users with the necessary permissions can save queries or assign saved queries to other users.

Table 1 Where Clauses in the Advanced Query Builder

Numeric Date/Time Text Selection List

is equal Yes Yes Yes Yes

is greater than Yes Yes

is less than Yes Yes

is between Yes Yes

contains Yes

relative clause not older than currently logged on user,current computer,

Page 27: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 27

ChemStore C/S Concepts 2Retrieving Data from the Database

Filtering the Dataset

Often, the dataset that has been retrieved by a query contains runs that are not required to produce a specific control chart or report. In such cases, a filter can be used to filter out only those runs that are required, and hide information that is not required. For example, a query may return 50 samples that were run by two different operators; you can then apply a filter that displays the results from only one operator at a time. The filter operates only on the current dataset, not on the complete database, but it does not change the contents of the dataset. The application of a filter also speeds up the work time.

Two filters are available:

• a custom filter can be specified using a graphical constructor in a similar way to the query, and can be used to filter the dataset using query-type search criteria (see Figure 4). The search criteria available are the configured columns of the results table.

• the Standard filter contains a set of predefined filters that have been selected to reflect the most frequently-used criteria. The criteria available are approval status, archive status and run version, see Figure 5 on page 28. The run version filter is useful when you have chosen to retrieve all versions of a run, but want to review only the latest version.

If both the standard filter and a custom filter are applied at the same time, the criteria are combined with an AND operator.

Figure 4 The Custom Filter Graphical Constructor

Available data categories from which to select

Filter condition

List of values in the selected data category

Page 28: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

28 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsRetrieving Data from the Database

The filters are available in two states:

• The Filter state displays only those runs that match the filter conditions; runs that do not match the filter conditions are hidden.

• The Complement Filter state displays only those runs that do not match the filter conditions; runs that match the filter conditions are hidden.

Both filter states are set up in exactly the same way, and can be switched on and off. You can toggle between the Filter and Complement Filter to show the two complementary sets of runs that make up the dataset.

Like queries, custom filters form part of the ChemStore C/S Settings that can be shared between nominated users. Once a custom filter has been specified and saved, it can be assigned to other users in addition to the owner. Custom filters are also protected from misuse by user permissions (see “Security” on page 125); only users with the necessary permissions can save a custom filter specification or assign a custom filter to other users.

Figure 5 The Standard Filter Selection Items

Page 29: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 29

ChemStore C/S Concepts 2Reviewing and Approving Results

Reviewing and Approving Results

ChemStore C/S is designed for reviewing analysis results in context. The batch review function of the ChemStation is an ideal vehicle for review and transfer of data from the ChemStation to the ChemStore C/S database.

Once the results are in the database, you can consolidate and review them in context. Having retrieved the data you are interested in by using a query, and, if necessary, filtered out a sub-set of the retrieved dataset, you can display the remaining results organized by either sample (sample review) or compound (compound review).

Run Approval Status

When a run is transferred from the ChemStation to ChemStore C/S, it is given the approval status Approval Pending. Depending on the study configuration, all runs of this study might electronically approved and signed by one or multiple 1st level reviewers and one or two 2nd level reviewers. The approval can also be denied by the "Reject" option. A rejection submitted by one user overrules the approval of other reviewers.

User permissions for the 1st and 2nd approval level (see “Security” on page 125) can be assigned individually. The two 2nd level approvals are equivalent in respect to the execution order. A 2nd level approval can also be applied for runs with no 1st level approval.

A user with the requisite permissions can either approve or reject individual runs or complete sets of runs. The process of approval or rejection is divided in two steps, as outlined in Figure 6 on page 30.

Page 30: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

30 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsReviewing and Approving Results

Figure 6 Mark runs for approval

Step 1 is marking the runs for approval or rejection during the result review.

Step 2 is actually applying the electronic signature and approval/rejection comment to individual or all runs marked in step 1. To harmonize the approval comments, administrators can pre-define a set of standard comments in the Approval configuration menu (Figure 7 on page 31). However, the user can still add a new comment or modify or extend the pre-defined comments during the approval process.

Page 31: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 31

ChemStore C/S Concepts 2Reviewing and Approving Results

Each action of step 2 is documented in the audit trail (see Figure “Audit Trail” on page 156) with the time of the action, the status of the run, the operator name and a comment, supplied or selected by the operator. The change of the approval status is confirmed as '1st level approved', '2nd level approved' or 'Rejected'.

An approved run can be later rejected, and vice versa, and each of these actions, as for approval and rejection, invokes a request for a reason for the change in status. The change in status also generates an entry in the audit trail.

The runs can also be automatically locked when a specified approval status is reached. Locking a run prevents further batch reprocessing of this run. To undo the locking, the approval needs to be rejected.

Figure 7 Approval Comments configuration dialog

Page 32: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

32 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsReviewing and Approving Results

If modifications are made to the custom field values of an approved run, its approval status is reset to Approval Pending; in the client/server version, a run that is reopened is automatically given the Approval Pending approval status.

Figure 8 Run Approval Status Transitions

Approved

Approval Pending

Rejected

useruser

change or

reopen

user

user

change or reopen

Page 33: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 33

ChemStore C/S Concepts 2The Graphical User Interface

The Graphical User Interface

The ChemStore C/S window is divided into four areas (see Figure 9 on page 36 and Figure 10 on page 41):

• the main toolbar

• the second toolbar

• the left panel of the display

• the right panel of the display

The user interface settings, including the settings of all user-configurable elements, can be saved and later recalled (see “User Interface Settings” on page 48).

The Main Toolbar

The main toolbar is situated below the menu bar across the top of the ChemStore C/S window. It contains the tools that allow you to set up queries and filters, to switch views and layouts and to set up and print reports. All activities supported by the tools in all toolbars are also available from menu items.

The following tools are available in all modes:

Switches to the Sample View or Compound View, which you use to review, approve or reject, and mark runs for batch processing.

Switches to the Delete View in the stand-alone version or the Archive/Delete View in the client/server version, which you use to mark runs for archiving and/or deletion. This button is absent if the current user does not have archive/delete permission.

This tool appears twice. In the custom calculator context as well as in the report context. It displays a pop-up menu from which you can select to create a new calculation or report template, edit an existing calculation or report template or manage your calculation or report templates.

Page 34: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

34 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

Displays a preview of the current calculation results using the currently selected calculation template.

Displays a preview of the current report using the currently-selected report template.

Prints a report of the current results using the currently-selected report template on the currently-selected printer.

Displays a pop-up menu from which you can save the current user interface settings for later recall, manage previously-saved user interface settings or select a previously-saved user interface setting.

The two tabs, Sample and Compound, switch between Sample View and Compound View. Each layout includes a second toolbar, the format of which is dependent on the layout selected (see “Sample View” on page 36 and “Compound View” on page 41).

The following tools are available in the Sample View:

Switches to the Sample Review layout

Switches to the Sample Table layout

Switches to the Sample Chart layout

When Regression Statistics are selected in the Sample View, two additional tools are available:

Displays the Regression Results table and calculated statistics

Displays the Regression Residuals chart

Page 35: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 35

ChemStore C/S Concepts 2The Graphical User Interface

The following tools are available in the Compound View

Switches to the Compound Review layout

Switches to the Compound Table layout

Switches to the Compound Chart layout

When Regression Statistics are selected in the Compound View, two additional tools are available:

Displays the Regression Results table and calculated statistics

Displays the Regression Residuals chart

Page 36: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

36 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

Sample View

Sample View displays the retrieved runs based on sample information. It offers three layouts, which give different information on the current dataset:

• The Sample Review layout is a three-panel layout with the Run List in the left panel, the Chromatogram window in the upper part of the right panel and the Summary Sample table in the lower part of the right panel (see Figure 9 on page 36).

• The Sample Table layout is a two-panel layout with the Run List in the left panel and the Sample table in the right panel.

• The Sample Chart layout is a two-panel layout with the Run List in the left panel and the Sample chart in the right panel.

In addition, Sample View can show two different sets of statistical results:

• Summary Statistics splits the right panel of the Sample Table layout and shows the Sample table in the upper part of the panel and the Summary Statistics results in the lower part of the panel.

• Regression statistics makes two layouts available:

Figure 9 Sample Review Layout

Second toolbar

Run list

Main toolbar

Chromatogram of selected run

Summary sample table

Page 37: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 37

ChemStore C/S Concepts 2The Graphical User Interface

The Regression Results layout is a three-panel layout with the Run List in the left panel, the Regression Results table in the upper part of the right panel and the calculated regression statistics in the lower part of the right panel.

The regression Residuals layout is a two-panel layout with the Run List in the left panel and the Residuals chart in the right panel.

The Second Toolbar in Sample View

The tools that are available in the second toolbar in Sample View depend on the layout selected.

The following tools are available in all Sample View layouts:

Allows you to set the approval status of all runs marked for approval to Approved.

Allows you to set the approval status of all runs marked for rejection to Rejected.

Allows you to set up one or more batches for the runs marked for batch reprocessing.

Displays the sample-related information about the current run, including the sample-related processing parameters.

Displays the run- and data analysis-related information about the current run.

Displays the Audit Trail for all versions of the current injection.

Displays the ChemStation method parameters for the selected run. It is only active for run versions transferred to ChemStore revision B.02.02 or higher.

Displays the table column selection dialog box to allow you to select the columns to include in the Sample table.

Page 38: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

38 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

Displays a menu that allows you to switch between the normal custom filter and the complement filter.

Toggles the state of the current normal custom filter, on or off.

Toggles the state of the current complement filter, on or off.

Enables you to export data either to the Windows Clipboard or to a file in Microsoft Excel format.

Enables you to print one or more sections of the current view to the currently-selected printer.

Sample Review layout adds one tool to the general tool set:

Starts an automatic stepping through the runs. Choose the Stop button, which replaces the Start button to stop the automatic stepping through. You can configure the time interval for the automatic stepping through.

Sample Table layout adds two tools to the general tool set:

Sorts the results in alphanumeric or date/time order of the selected column of the Sample Table.

Sorts the results in reverse alphanumeric or date/time order of the selected column of the Sample Table.

Sample Chart layout adds one tool to the general tool set:

Enables you to specify the chart options.

Page 39: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 39

ChemStore C/S Concepts 2The Graphical User Interface

Regression Statistics adds one tool to the above tool sets:

Enables you to set up the parameters for the calculation of regression statistics and residuals.

Run List

The Run List (see Figure 9 on page 36) is a table containing information about all the runs in the current dataset. The table contains five columns:

• Run shows the number of the run in the dataset

• Mark run for shows how you have marked the run:

• P indicates that the run has been marked for approval

• R indicates that the run has been marked for rejection

• B indicates that the run has been marked for inclusion in a batch to be reprocessed in the ChemStation

• EXC indicates that the run has been excluded from reports and statistical calculations

• Sample name shows the sample name

• Audit shows if a run has already been reprocessed:

• If only the latest version of a run has been retrieved, an asterisk denotes that there are multiple versions.

• If all versions have been retrieved, they are numbered in sequence, 1 being the earliest version; + denotes the latest version.

• Displays the ChemStation method parameters for the selected run. It is only active for run versions transferred to ChemStore revision B.02.02 or higher.

• Status shows the current status of the run (approval status in data review, archive status in the archive/delete view).

Use the checkboxes at the bottom of the panel to mark runs for approval, rejection or batch processing, or use the right mouse button and select from the pop-up menu. Marking the runs does not change the approval status of the runs (see “Run Approval Status” on page 29), or transfer the runs to a batch. The approval status of the runs is changed only when the appropriate tool is used (the Approve tool sets the approval status of the runs marked for approval to Approved; the Reject tool sets the approval status of the runs marked for rejection to Rejected) and the batch is set up only when the Batch

Page 40: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

40 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

Processing tool is used. If you approve or reject a run, you must supply a reason for the change in status; this information, together with the change in status, is entered in the Audit Trail for the run, together with the change of status.

Summary Sample Table

The Summary Sample Table contains information about each peak in the run selected in the run list. You can configure the Summary Sample Table and select the information that is displayed (see “Configuring Tables” on page 44). You can use the Next and Previous buttons below the table to manually step through the runs, The Exclude button marks the run as Excluded and removes the run from the right panel of the sample review; you can remove the exclusion mark using the Include button.

Sample Table

The Sample Table of the Table Layout view contains information about each run in the run list. You can configure the Sample Table and select the information that is displayed (see “Configuring Tables” on page 44). Typically that view is used to display the equivalent to the sequence table in a sequence oriented workflow.

Page 41: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 41

ChemStore C/S Concepts 2The Graphical User Interface

Compound View

Compound View displays the retrieved runs based on compound information. Like Sample View, it offers three layouts, which give different information on the current dataset:

• Compound Review layout is a four-panel layout with the Compound List in the left panel, the Chromatogram window at the left of the upper part of the right panel, the Spectrum window at the right of the upper part of the right panel and the Summary Results table in the lower part of the right panel (see Figure 10 on page 41).

• Compound Table layout is a two-panel layout with the Compound List in the left panel and the Results table in the right panel.

• Compound Chart layout is a two-panel layout with the Compound List in the left panel and the Compound chart in the right panel.

In addition, Compound View can show two different sets of statistical results:

• Summary Statistics splits the right panel of the Compound Table layout and shows the Results table in the upper part of the panel and the Summary Statistics results in the lower part of the panel.

Figure 10 Compound Review Layout

Main toolbar

Spectrum of selected compound

Summary results table

Second toolbar

Compound list

Chromatogram of selected run

Page 42: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

42 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

• Regression statistics makes two layouts available:

• Regression Results layout is a three-panel layout with the Compound List in the left panel, the Regression Results table in the upper part of the right panel and the calculated regression statistics in the lower part of the right panel.

• Regression Residuals layout is a two-panel layout with the Compound List in the left panel and the Residuals chart in the right panel.

The Second Toolbar in Compound View

The tools that are available in the second toolbar in Compound View depend on the layout selected.

The following tools are available in all Compound View layouts:

Displays the table column selection dialog box to allow you to select the columns to include in the Results table.

Displays a menu that allows you to switch between the normal custom filter and the complement filter.

Toggles the state of the current custom filter on or off.

Toggles the state of the current complement filter, on or off.

Enables you to export data either to the Windows Clipboard or to a file in Microsoft Excel format.

Enables you to print one or more sections of the current view to the currently-selected printer.

Compound Review layout adds five tools to the general tool set:

NOTE The data that are available for the filter are only based on sample related data.

Page 43: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 43

ChemStore C/S Concepts 2The Graphical User Interface

Allows you to set the approval status of all runs marked for approval to Approved.

Allows you to set the approval status of all runs marked for rejection to Rejected.

Allows you to set up one or more batches for the runs marked for batch reprocessing.

Displays the peak-related information about the current compound.

Starts an automatic stepping through the runs. Choose the Stop button, which replaces the Start button to stop the automatic stepping through. You can configure the time interval for the automatic stepping through.

Compound Table layout adds two tools to the general tool set:

Sorts the results in alphanumeric or time/date order of the selected column of the Results Table.

Sorts the results in reverse alphanumeric or time/date order of the selected column of the Results Table.

Compound Chart layout adds one tool to the general tool set:

Enables you to specify the chart options.

Regression Statistics adds one tool to the above tool sets:

Enables you to set up the parameters for the calculation of regression statistics and residuals.

Page 44: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

44 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

Compound List

The Compound List (see Figure 10 on page 41) contains a list of all compounds (analytes) in the current dataset. Calibrated compounds are reported with their name, non-calibrated compounds are reported as Uncalibrated Compounds. You select the compound to display from the list. In Compound Review layout, you can select only one compound, but in Compound Table layout, an additional column allows you to select multiple compounds to display. In Compound Chart layout, you can chart multiple parameters of a single compound or single parameters of multiple compounds. If you chart multiple parameters of a single compound, the plots are normalized to the largest value in the dataset.

Summary Results Table

The Summary Results Table shows the information for the selected compound in each run. You can configure the Summary Results Table and select the information that is displayed (see “Configuring Tables” on page 44). Use the checkboxes below the table to mark runs for approval, rejection or batch processing. Marking the runs does not change the approval status of the runs (see “Run Approval Status” on page 29), or transfer the runs to a batch. The approval status of the runs is changed only when the appropriate tool is used (the Approve tool sets the approval status of the runs marked for approval to Approved; the Reject tool sets the approval status of the runs marked for rejection to Rejected) and the batch is set up only when the Batch Processing tool is used.

Results Table

The Results Table shows the information for all the selected compounds in each run. You can configure the Results Table and select the information that is displayed (see “Configuring Tables” on page 44). You can use the Results table to select the X- and Y-axis enumerators for your charts (see Figure 11 on page 45).

Configuring Tables

You can configure all tables that are displayed in the right panel of the ChemStore C/S window. Using a dialog box, you can select the columns that constitute each table, and their sequence in the table. For numeric columns

Page 45: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 45

ChemStore C/S Concepts 2The Graphical User Interface

and rows of the statistic results, you can also select the number format and precision; for date columns, you can also select the display style. The configuration dialog box contains only those columns that are valid for the table you are configuring.

During column configuration, you can also select the enumerators for the X- and Y-axes of your charts using the checkboxes in the column configuration dialog box. X- and Y-axis enumerators may be selected only from columns that are included in the table. You may choose only one X-axis enumerator, but as many Y-axis enumerators as you wish.

In addition to the standard results that are available from the ChemStation, you can also define columns that are the results of arithmetical calculations using the data available in the configured table.

Intra-sample Calculations

In your results tables, you can configure a column that contains a mathematical expression to perform a simple intra-sample calculation, for example a sum or a ratio, on the results for each run or compound.

These intra-sample relate to the run or peak in the row of the table. For calculations covering a cross-sample and enhanced intra-sample (relating to a column in the table) see “Statistics” on page 46 and “Preparing to Use the Custom Calculator Script Editor” on page 57

Figure 11 Column configuration dialog

Page 46: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

46 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

In your custom calculations, you can use any of the data items from the table; the mathematical expression can include arithmetical operators and mathematical functions as well as numerical constants.

Statistics

ChemStore C/S provides three levels of statistics:

• Summary Statistics are displayed below the Sample Table in Sample Review or the Results Table in Compound Review.

• Regression Statistics are displayed in two separate layouts: Regression Results Table layout and Residuals Chart layout.

• If statistic results needs to be used for further calculations the summary and regression statistics can also be calculated by the custom calculator. See “Example: Creation of a report calculation script” on page 68.

Summary Statistics

Summary statistics are calculated and displayed for each column in the Sample Table or Results Table. The statistics that are displayed are:

For details of the calculations used in Summary Statistics, see “Statistical Calculations” on page 178.

Regression Statistics

For details of the calculations used in Regression Statistics, see “Regression Calculations” on page 179.

Count Number of values in the data set

Max Maximum value in the data set

Mean Mean value of the data set

Min Minimum value in the data set

RSD Relative standard deviation of the values in the data set

Std. Dev Standard deviation of the values in the data set

Sum Sum of the values in the data set

Variance Variance of the values in the data set

Page 47: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 47

ChemStore C/S Concepts 2The Graphical User Interface

Generating Control Charts

You can generate charts from the results you choose to include in the Sample Table (Sample Review) or the Results Table (Compound View). The X- and Y-axis enumerators that you chose when you set up the table are transferred to the chart. However, you can use the buttons on the X-and Y-axes to change the selections, for example to simplify the chart.

Using the Chart Options dialog box, available either from the options menu or from the toolbar, you can format the color, style and shape of the lines and points for each of the lines on your chart. You can also format the style, size and position of the title, axis labels and chart legend.

By default, charts are displayed with automatic scaling of the axes: the X-axis is scaled from the minimum to maximum value of the X-axis enumerator range, the Y-axis is scaled from the minimum to maximum value of the set of Y-axis enumerators. You may change the scaling by setting maximum and minimum values, and you may also specify the precision of the axis scaling values. You may also choose to display the data on logarithmic axes, in which case the precision settings have no effect. You cannot display time enumerators in logarithmic form. Data points less than or equal to zero are not shown on logarithmic plots.

Adding Limit Lines

In addition to plotting your results, you can add up to five limit lines to your charts: center line, upper and lower warning limits and upper and lower critical limits. To set the positions of these lines on the Y-axis, you can use either calculated values (mean and standard deviation) or specified values:

Center Line Can be based on either the mean value of one of the Y-axis enumerators, or set to a specified value.

Limit Lines Can be specified values, or set to a specified difference from the center line. The specified difference can be an absolute value or based on the standard deviation of the selected Y-axis enumerator.

Page 48: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

48 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsThe Graphical User Interface

User Interface Settings

When you have reviewed your data, and set up tables, control charts and statistics options, you can save these user interface settings and recall them for later reuse (see Figure 12). The information that is saved includes not only the configuration and format of configurable tables and charts, but also the widths of columns in non-configurable tables (for example the run list and compound list) and the sizes of the panels in each of the layouts. The user interface settings are saved with a name, and are available for you to use again at a later date; a person with the user permission to manage user interface settings, can also allocate the saved settings to other users for their use.

Page 49: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 49

ChemStore C/S Concepts 2The Graphical User Interface

Figure 12 Configurable Settings of the User Interface

Column widths of all tables in all layouts

Line and point style and color, axis labelling, legend and title on all charts

Number format and precision in results tables

Panel sizes in all layouts

Page 50: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

50 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsTransferring Data to the ChemStation

Transferring Data to the ChemStation

Transfer of data from ChemStore C/S to the ChemStation takes place in two stages.

The first stage occurs during the review process in ChemStore C/S, when you select those runs to be transferred back to the ChemStation for reprocessing or re-analysis. The addition of a run to a batch request generates an entry in the audit trail for that run. When you have selected all relevant runs, they are transferred into one or more batches, together with all associated information (study name, custom field values and attributes). Before the runs are transferred, the system looks for identical files in the original location; if the files already exist, and they are identical, no transfer is made. If the files do not exist in the original location, and they are available in the database, they are transferred to the ChemStation in a separate, temporary storage area.

The second stage occurs in the ChemStation when you choose to load a ChemStore C/S batch into the ChemStation’s batch review function. If more than one batch has been set up in ChemStore C/S, you select the batch to load from a list of pending batches. Only those batches that have been assigned to the current ChemStation user, or to all users, are shown in the list.

You can also choose to reload any or all files saved with a run or a complete sequence interactively from the run information interface (raw data, method or sequence files). In this case, if the files already exist in the ChemStation, you are prompted change the destination path or cancel the transfer.

Reprocessed runs that are transferred to ChemStore C/S are stored in the database as new runs; they do not overwrite the original results. The original results are retained in the database, and are optionally retrieved (if they fit the search criteria). You can use the standard filter to display all versions (the latest version plus all previous versions), or to display only the latest version.

Page 51: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 51

ChemStore C/S Concepts 2Versioning

Versioning

ChemStore C/S retains all previous versions of files that are reprocessed by the ChemStation. If you choose to store raw data, method and sequence files with the data, only those files that have been changed are stored with the new version, i.e. the raw data file is stored only once, the method file is stored only if the method has been modified and the sequence file is stored only if the run is transferred in a sequence, and not if the run is transferred in a batch or manually.

When you set up a query, you can choose to retrieve either all versions of the run or only the latest version. If you choose to retrieve all versions of the run, you can apply a standard filter to display only the latest version.

When all versions of the run are displayed, the Audit column of the run list shows the version number of the run: 1 is the oldest version, 2 the next oldest; the run marked with + is the latest version.

If only the latest version of the run is displayed, a * in the Audit column of the run list denotes that multiple versions of the run exist in the database.

An <A> in the Audit column generally denotes that there is more detailed information for this run besides the information that the version has been added to the database (event: NEW). The following events triggers the <A> sign:

• run added to a new batch

• run approval status changed or approval rejected

• run reopened

• custom field value changed

• sample name changed

Page 52: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

52 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsVersioning

Figure 13 Retrieval and display of versions

Only latest version retrieved

All versions retrieved

Denotes that multiple versions exist in the database

Denotes an entry in the audit trail for the latest version

Denotes the latest version

Previous versionsDenotes an extended entry in the audit trail for this version

Page 53: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 53

ChemStore C/S Concepts 2Deleting Runs from the Database

Deleting Runs from the Database

You can delete runs from the ChemStore C/S database using the Delete view of the stand-alone version or the Archive/Delete view of the client/server version. In the client/server version, you can choose to archive the runs before deleting, or to delete them without archiving. To archive the data in the stand-alone version, copy the *.mdb database files to your archive medium.

You select the runs for deleting from the run list in the left panel of the Delete view (stand alone version) or Archive/Delete view (client/server version). When you select the Delete button from the toolbar, the runs are deleted.

Deleting is protected by user permission and password. The Archive/Delete and Delete views are available only to users with the requisite permissions, and you must enter your password before deletion of the runs will begin.

CAUTION Runs that you delete are removed from the database and cannot be retrieved. Be sure that you only delete runs that you have either archived, copied to another medium, or no longer need.

Page 54: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

54 Agilent ChemStore C/S Concepts Guide

2 ChemStore C/S ConceptsDeleting Runs from the Database

Page 55: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

55

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

3Using the Custom Calculator

What is the Custom Calculator? 56

Preparing to Use the Custom Calculator Script Editor 57

Custom Calculator Script Editor 60

This chapter describes the usage of the custom calculator. It shows how to create customized calculations on data sets of the ChemStore database and how to integrate the results into a ChemStore report.

Page 56: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

56 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorWhat is the Custom Calculator?

What is the Custom Calculator?

The Custom Calculator is an integrated ChemStore component that provides enhanced calculation capabilities.

Calculations are described by calculation templates (scripts) using a simple, SQL-like programming language. A number of wizards help the development of syntactically correct calculation templates. These wizards guides the user through the creation of each calculation step. For experienced users, a script editor allows the direct programming of calculation scripts.

The programming language contains constructs that range from simple mathematical operations and functions to complex summary calculations. A set of built-in calculation templates is designed for common calculations; these may serve as a starting point for the development of custom calculation templates.

When a calculation template is written, it can be tested and debugged using powerful debugging tools such as step-by-step execution and breakpoints. Calculation templates can be stored and assigned to specific users for execution only. When modifying a calculation template, the user can store it either as a separate version or as a new calculation template. The ChemStore C/S logbook keeps track of such operations.

The user can preview the results of a calculation template, format and include them in reports. Calculation templates included in reports are automatically executed each time a report is generated. The results of these calculations can be printed in a tabular form as well as in charts.

Access to these calculation capabilities is regulated by a special user privilege, assignment of calculation templates to users, and management of calculation template ownership.

Page 57: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 57

Using the Custom Calculator 3Preparing to Use the Custom Calculator Script Editor

Preparing to Use the Custom Calculator Script Editor

Retrieving the calculation data set

The calculator template works on the current dataset downloaded by a database query. It’s the dataset that can be seen in the review table. However, the review tables doesn’t need to display all columns that are supposed to be used for calculations. The Table Wizard of the Custom Calculator allows to specify all columns of the current dataset, independent of the user interface settings.

When starting the development of a new calculation template it is important to query a dataset that is representative for the data where the template should be used for.

Understanding the Custom Calculator Script Editor Views

Windows selection tree

The window navigator is a tree view that helps the user to navigate through the custom calculation windows represented as leafs. Double clicking on a leaf node of this tree view finds, shows, and brings to the top the selected window that is partially or completely overlapped by other windows or even invisible at the time. The top two leaf nodes (Calculation, Errors) are always present. These nodes bring to the top the Calculation editor and Calculation errors windows. Below the Tables node all the currently available calculation table windows are listed. These nodes bring to the top the selected custom calculation table.

Page 58: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

58 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorPreparing to Use the Custom Calculator Script Editor

Calculation window

This window is used for editing and debugging of calculation scripts. The title bar contains the name and version of the loaded calculation template, if available. The left border contains current execution line and breakpoint markers. The bottom border contains the cursor position (line and column) and an execution progress bar. The client area of the window is used for editing of the calculation script. The calculation script lines are written in black unless an error is detected or a warning is triggered. Error lines are written in red, while warning lines appear in blue.

Variables list

The list of variables is a list containing names and values of currently available variables of a calculation Double clicking on a variable in the list of variables pops up a dialog containing a read-only copy of the variable name and value. This dialog can be used for viewing long variable names and values. It is also possible to copy these values to clipboard.

Figure 14 Windows selection tree

Figure 15 Calculation window

Page 59: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 59

Using the Custom Calculator 3Preparing to Use the Custom Calculator Script Editor

Errors and warnings window

The errors and warnings window lists detailed information on all errors and warnings resulting from the execution of a calculation script. A reference to the script line and a visual indication of the position of the malfunction helps to resolve the error.

Figure 16 Variables list

Page 60: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

60 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Custom Calculator Script Editor

The custom calculator script editor is a yellow colored calculation window where the scripting of the custom calculator template is done. This can either be done by using the custom calculator wizards to create the appropriate script code or by free-hand programming, like writing a macro.

Understanding the Custom Calculator Script Wizards

Wizards are used to support the user in generating a script code for the desired calculation. For each individual calculation operation task a specific wizard is used to create this particular part of the script.

The wizards can be invoked either from the toolbar or by clicking on a line in the template (right mouse click, select Edit from the upcoming popup menu). According to the content of this line the correct wizard is displayed holding the content of this line in its appropriate fields.

The wizards can be used to create new lines or edit given lines in the template.

Common wizard characteristics

Figure 17 shows an example of a simple table wizard. At the left of each wizard dialog there is a symbol that visualizes the calculation statement’s goal. A table wizard is e.g. used to put data from the current dataset into a table.

In the middle there are syntactical components that define the parts of the command line to be generated. The order of these components is the same as they appear later on in the calculation statement. Each syntactical component contains a terminal name (large blue letters), description (small black letters), and controls for selecting and/or entering statement specific information.

Page 61: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 61

Using the Custom Calculator 3Custom Calculator Script Editor

Common sub-Wizards

Selecting columns

Figure 18 shows a dialog used for selecting table columns, e.g. when browsing the SELECT function of the table wizard. A tree view at the left contains all available tables and their columns. By moving items to the right, the new table gets defined. A list of selected columns is placed at the right of the dialog. The icons before selected tree nodes are green while those before unselected tree nodes are white.

Figure 17 Table Wizard

Page 62: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

62 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Expressions

The Expressions dialog shown in Figure 19 is the heart of the custom calculator. It allows the creation and edition of expressions and conditions. The edit box in the topmost frame contains the expression composed so far. It is possible to compose the contents by selecting the items from Data Items and Functions frame and clicking on the ‘Add to Expression’ buttons. The contents of the edit box can also be typed in manually. A check icon at the top right corner and an additional information line indicate if the currently built expression is syntactically correct. These two items are updated continuously if the check box for Continuous check is checked.

If the currently built expression is not syntactically correct, the check icon changes to a question mark. The check information gives some hints about the problem an the offending part of the expression is colored red in the edit box.

Figure 18 Column selection screen (Custom Calculator)

Page 63: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 63

Using the Custom Calculator 3Custom Calculator Script Editor

Table Wizard

The table wizard displayed in Figure 17 offers the possibility to select a set of columns from the current dataset and store it into a new calculation table. This is the initial step to create a new calculation template. It is recommended to add all columns used for sample identification and used for the desired calculation steps to one overview table as a starting point.

The wizard writes the result of the user’s selection into the script window as given in the following example:

SELECT Study,’StudyName’,Sample.'Sample Name', etc.

FOR "Antipyrine",”Phenacteine”

WHERE (Phenacetine.PeakArea>0

INTO "1_overview"

Figure 19 Set condition dialog

Page 64: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

64 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

In this example the Study name column was selected from the study table and the Sample column was selected from the sample table. The selection is based on the compounds Antipyrine and Phenacetine and limited to all samples where a Phenacetine peak was found. The result of this section will be written to the new table called 1_overview.

Column Wizard

The Column wizard allows the extension of a given table by a new calculation column. This column contains values based on an expression defined using the expression wizard similar to Figure 19.

The example shown in Figure 20 adds a new column ‘ResRatio’ to the table ‘1_overview’. The contents of the column is defined by the area ratio of the compounds ‘Ethanol’ and ‘Methanol’.

Figure 20 Column Wizard

Page 65: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 65

Using the Custom Calculator 3Custom Calculator Script Editor

Variable Wizard

The Variable Wizard is similar to the Column wizard. The user can apply a mathematical expression of values from one or more columns and stores the result in a variable. Usually this is the case when the result comprises the evaluation of a series of values.

If Wizard

The previous examples showed how to calculate new values based on given values in columns and assign them either to new values in a column or to a single variable.

In cases where this assignment should be dependent on a certain condition, the IF wizard can be utilized.

Format Wizard

The custom calculator displays numbers as raw figures using all available decimals. To achieve a better readability or display significant digits only, the Format wizard formats single values (variables), a column of a specific table or even a set of columns from multiple tables. The precision of calculation is not affected by the number format.

Besides numbers it also possible to format strings, dates and times and even empty cells. The on-line help of the Format Wizard contains examples of these extended formatting options.

Transpose Wizard

The Transpose Wizard allows to transpose an existing table by turning columns into rows and rows into columns. This can be useful when calculating statistics over values which are given in a row. Those might result from a previous calculation. However, statistics are only possible along columns and a transpose step needs to be done prior to statistics calculation.

The Transpose wizard dialog (Figure 21 on page 66) contains the following syntactical components:

• In the TRANSPOSE combo box the user can specify the name of the table to be transposed. Note: The selection contains only those tables that are available at the given calculation line - all previously generated tables.

Page 66: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

66 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

• In the BY combo box the user needs to specify which table column should be used as the header row for the transposed table. Typically this is the sample name or run ID, as those items defined the rows of the original table.

• In the INTO edit box the user can specify a name for the transposed table.

Group Wizard

Grouping a table is useful e.g. in cases of calculating statistics over a certain set of samples, rather than overall statistics. The same results can be achieved by defining multiple sub-tables, but the Group wizard combines this to one step.

The dialog shown in Equation 22 on page 67 contains the following syntactical components:

• In the GROUP combo box the user specifies the name of the table considered to be grouped.

Figure 21 Transpose Wizard

Page 67: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 67

Using the Custom Calculator 3Custom Calculator Script Editor

• In the BY combo box the user specifies the identifier of the group. E.g. grouping by ‘Sample calibration level’ allows to combine samples and calibration in separate groups. Calibration runs will also be grouped by the calibration level.

• The DO component defines a (statistical) function being applied to a given table column resulting into a new named column.

The INTO box needs to be filled with a name for the resulting table.

Figure 22 Group Wizard

Page 68: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

68 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Example: Creation of a report calculation script

In the following chapter it is shown how a simple example template can be developed. The example is based on the data you receive when executing the query ‘SSR for quality control’ on the ChemStoredemo.mdb database installed with ChemStore C/S. It is recommended to follow the steps in this chapter to get familiar with the usage and the functionality of the calculator wizards. This example does not represent a recommendation on how to evaluate acceptance criteria for a data set.

Assuming you have been given the following task:

Develop a template where the relative standard deviation of the areas of a given compound gets calculated for calibration runs and sample runs. The comparison should mark sample runs as ‘pass’ if the relative standard deviation is less in comparison to the relative standard deviation for calibration runs. The results should be integrated in a report.

To achieve a solution for this task you have to go through four phases:

1 Set up a plan including the steps which are needed to achieve a solution.

2 Develop the template according to this plan.This includes testing and correcting of each development step.

3 Refine the template

4 Include template into report.

Set up a plan

To set up the plan you first have to fully understand the task. Divide the task into steps and write down what you want to achieve in each step:

NOTE In case the described script cannot get achieved, a copy of this example script is located on the ChemStation Plus installation CD. Browse the support folder for simple_example.cct. To import the example script to the demo database, use the import function of the custom calculator.

Page 69: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 69

Using the Custom Calculator 3Custom Calculator Script Editor

1 Create a table which consists of all parameters which you need for your calculation and all sample parameters which need for sample identification and reporting. For calculation you need:

• Peak area

• Sample type to distinguish calibration runs from sample runs

For sample identification you want to see:

• Sample name

• Raw data file name

2 Calculate the relative standard deviation of peak areas per calibration level and store it into a new table.

3 Calculate the average of the relative standard deviations and store it in a variable.

4 Calculate the mean areas for all sample runs and the absolute and relative deviation from this mean amount. This should be done on a per run basis.

5 Check whether the relative deviation of each sample is less than the average relative standard deviation of the calibration runs and mark it appropriately.

After each step you should test the intermediate result to see whether you are on the right track. If the intermediate result deviates from your expectation you should undertake corrective actions.

Developing the template

Create an overview table

To create the starting table you will use the table wizard as shown in Figure 23 on page 70. Select the columns Sample Name, Sample Calibration level, Sample type, Acq. Sequence name, Raw data file name and Peak area for compound ‘Antipyrine’ and store it in the table '1_overview'. It's a good practice to add a leading number to the tables created by the script in order to reflect the script structure in the results.

Page 70: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

70 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Figure 23 Table Wizard

Figure 24 Overview table

Page 71: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 71

Using the Custom Calculator 3Custom Calculator Script Editor

For testing this step you execute the script and you should get a table where sample names, calibration levels, sample types, acq. sequence names, raw data file names and peak areas for the compound 'Antipyrine' are listed for all samples retrieved by the query 'SSR for quality control’ (see Figure 24 on page 70).

Create sub-tables

The next step is to create a sub-table to separate calibration runs from sample runs. This can be achieved by using a sub-table wizard. The key to separate the sample types is the WHERE condition, that has to match exactly example shown in Figure 25 on page 71. The condition "Calibration" has to be typed manually in the set condition dialog and it needs to be enclosed in double quotes.

Executing the script will result in a simple table that lists all calibration samples with the calibration level and the peak areas of compound 'Antipyrine' (see Figure 26 on page 72).

Figure 25 sub-table wizard for separating calibration runs

Page 72: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

72 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Calculate the relative standard deviation of peak areas for each calibration level

To calculate the relative standard deviation per calibration level you can utilize the Group wizard. Grouping the table '2_calib_runs' on calibration

levels generates one result per level and stores it in the new table '3_calibstatistics'. The calculation to be executed per group gets defined in the DO/AS section of the group wizard (Figure 27 on page 72). The group wizard

Figure 26 Calibration run sub-table

Figure 27 Group Wizard

Page 73: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 73

Using the Custom Calculator 3Custom Calculator Script Editor

allows to define multiple calculations for multiple compounds at the same time, but the current example is restricted to one, for the sake of simplification.

Executing the script results in a small table containing the relative standard deviation for each calibration level.

To average the relative standard deviations, the variables wizard should be used with the following definition statement:

Avg(RSTD_Antipyrine)=mean(‘3_Calibstatistics’.’RSTD_Antip_area’)

For testing this step you execute the calculation and you will find your variable with the appropriate result in the variable list. The calculation on the calibration runs is finalized now and the next step is to perform calculations on the sample run results and finally perform the desired comparison.

Calculate the relative area deviation for sample runs

To list the sample runs in a separate table generate a sub-table as shown in Figure 25 on page 71, but replace the sample type ’Calibration’ in the WHERE condition by ‘Sample’ and the destination table name to '4_sample_runs'.

The next two steps are adding two additional columns to the sample table - one for calculating the average response over all samples (see Figure 29) and another one for calculating the absolute deviation of the individual responses from the average. Use the 'add columns' wizard twice to achieve this result. The 'set condition' dialog for calculating the absolute deviation is shown in Figure 30.

Figure 28 Relative standard deviation of Antipyrine peak areas

Page 74: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

74 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Figure 29 Averaging Antipyrine responses for sample runs

Figure 30 Absolute deviation of Antipyrine area from mean value

Page 75: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 75

Using the Custom Calculator 3Custom Calculator Script Editor

Executing the script results in the following sample table:

Check and mark

Now, all results are available to define a final result table where the comparison will be executed. Use a sub-table wizard to generate an extract of the table '4_sample_runs' and store it in the final result table

Figure 31 Results of deviation calculations

Figure 32 Calculating relative area deviation

Page 76: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

76 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

'5_Antipyrine_results'. The example uses the columns sample name, raw data file name, peak area, delta_ant. If other columns for sample identification should appear in the final report, they need to be added in addition to the results table.

Before comparing sample runs and calibration runs, the absolute area deviation of the sample runs needs to be divided by the area to get the relative deviation. The add column wizard for this calculation is shown in Figure 32 on page 75.

If the relative area deviation is less than the average value for the calibration runs, the runs should be marked with PASS, otherwise with FAIL. This can be achieved with the If wizard. The IF condition in Figure 33 on page 76 shows this comparison. The THEN and ELSE statements are defined by column wizards, accessible through the browse buttons of the IF wizard. This step, shown in Figure 34 on page 77, will add a 'status' column to the result table.

Figure 33 If wizard for result evaluation

Page 77: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 77

Using the Custom Calculator 3Custom Calculator Script Editor

For testing this final step you execute the script and a new column is added to the result table having PASS or FAIL next to the relative deviation column (Figure 35 on page 77)

Figure 34 Defining a THEN statement with the help of a column wizard

Figure 35 Final result table

Page 78: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

78 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Refining the template

Formatting columns

Last but least you might want to format some of the columns. Use the format wizard to format e.g. the delta_Ant column of table’ 4_sample_runs’ and in table ‘5_Antipyrine_results’, the ‘rel_dev’ column and the Antipyrine.Peak.Area column as shown in Figure 36.

All table columns and all variables that should appear with the same number format can be defined by one table wizard.

Embedding template into report

In section “Calculations” on page 104 it is described how to embed a calculation result table into a report.

Figure 36 Formatting the final result table

Page 79: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 79

Using the Custom Calculator 3Custom Calculator Script Editor

Using the Custom Calculator Script Editor

The script editor is like a text editor for custom calculator scripts. Experienced users can write scripts without the help of wizard. This also allows to create script constructs which cannot be created with a wizard. Such script constructs can make the script code much shorter and clearer to overview.

Commands

In the following, the most important commands used for scripting the template will be briefly described. For a detailed description of all commands, refer to the reference section “Custom Calculator Commands” on page 185.

Selection of columns from database

To select columns from the database, optionally using a condition for shrinking the resulting dataset, you use the SELECT command:

SELECT column(s) FOR compound(s) WHERE aCondition INTO aTable

column(s) can be one or more columns, specified by name, from any available table in the database.

compound(s) can be one or more compounds, specified by name, for which the columns should be selected. Wildcards are allowed here. See the reference part of “Custom Calculator Commands” on page 185 for more details.

aCondition is optional. It specifies a condition for the selection.

aTable is the name of a table to which the selection will be written.

Selection of columns from a table

To select columns from a custom calculator table, you use the FROM command:

FROM aTable SELECT column(s) FOR compound(s) WHERE aCondition INTO anotherTable

This command does the same as the SELECT command, except that it retrieves the columns only from a table which you have received from a previous SELECT or FOR command.

Page 80: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

80 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Making decisions

Decisions can be made by using conditional statements. The IF THEN ELSE command is used to perform this operation:

IF aCondition THEN doThis ELSE doThat

If the condition is fulfilled, the THEN part will be evaluated; if the condition is not fulfilled, the ELSE part will be evaluated. The ELSE part is optional and can be omitted.

Formatting results in table(s)

The resulting values in the custom calculator tables or from variables can be formatted using the FORMAT command:

FORMAT column(s)/variable(s) USING aFormat

aFormat is a string (enclosed with quotes) containing the format specification. See the reference part of “Custom Calculator Commands” on page 185 for more details.

Transposing table(s)

Calculations are performed on selected columns for each row. Results are written into new columns. Certain (statistical) calculations need to be done on one row for each column. This is not possible directly: this row has to be turned onto a column. To do so, you use the TRANSPOSE command:

TRANSPOSE aTable BY aColumn INTO destinationTable

A column is used as key for the transpose operation. The entries of this column will be used as column headers for the destination table. Since the column names must be unique, the destination table does not contain any column with a name that appears more than once.

Grouping table(s)

Grouping tables is a way to better organize the contents of a table to make some statistics possible. You group tables using the GROUP command:

NOTE The conditional statement can be assigned only to operations on variables or table columns. It cannot be used for conditional evaluation of your template script.

Page 81: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 81

Using the Custom Calculator 3Custom Calculator Script Editor

GROUP aTable BY aColumn DO operation(s) INTO destinationTable

Grouping of a atable is defined by the values of a column. Each group of rows contains rows with equal values in the column. The operations to be performed on each group of rows are defined by the operation(s).

Adding comments

The #-symbol is used to start a comment. Text written after this #-symbol will NOT be evaluated:

# This is a comment!

In spite of a future wizard which will handle the declaration part of your script the comments are added after the variable definition in case they are related to this definition:

strCompund = “Barbital”# This is the definition for the compound

The declaration part of the script starts with the

# D E C L A R A T I O N

comment and ends with the

# I M P L E N T A T I O N

comment.

Syntax

The detailed description for the syntax of all commands and script parts you will find in the reference section of “Custom Calculator Commands” on page 185.

The following part describes some specials which should be considered to avoid unnecessary errors in the script.

Using quotes

Quotes must always be used to enclose strings. Strings are names of compounds, samples etc. and names of columns. Furthermore, as you will see in the next section, strings also can contain commands.

If the string also contains quotes, these quotes must be turned into double quotes.

Page 82: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

82 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Apostrophes around column names must remain when enclosed by quotes.

Using brackets

Brackets must always be used when several operations in an expression have to be ordered hierarchically. Expressions in brackets will be evaluated first. Therefore each part of expression which must be evaluated before another part has to be enclosed with brackets.

Case sensitivity

Usually the scripts (commands, column names, expressions) are not case-sensitive. There is just two exceptions:

• When a comparison is made on strings to check the identity of strings (string1=string2), this operation is case sensitive.

• The names of variables must be used in the script as they were defined in the beginning

Usage of string variables

The major feature of the script editor is to create string variables. In the following we will prefix string variable names with str. A string variable is a piece of text enclosed with quotes. The content of these string variables can be a simple name e.g. a compound name or the definition of the sample name column:

strCompound = “Barbital”

strSmplName = “Sample.’Sample Name’“

but also a piece of script code itself e.g. the instruction to select columns from the database:

strSelect = “SELECT Sample.’Sample Name’ FOR Barbital INTO aTable”

Consider that column names using a blank:

Sample.’Sample Name’

or a special characters in their names (like ‘1_overview’)

have to be enclosed by’ (apostrophe).

The purpose of using string variables is to define such variables once and reuse them in the rest of the script even multiple times.

Page 83: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 83

Using the Custom Calculator 3Custom Calculator Script Editor

To reuse these string variables you must prefix them with a $-sign:

$strCompound

$strSelect

The $-sign will instruct the script interpreter to interpret the content of this variable. Instead of writing

SELECT Sample.SampleName FOR Barbital INTO aTable

you also can write:

SELECT Sample.SampleName FOR $strCompound INTO aTable

The interpreter will substitute $strCompound by Barbital before evaluating the SELECT command and doing the database selection.

This has the advantage that you define a compound name once in the beginning of the script and then you can refer to it in the rest of the script. When you have to change the compound name you just do it once in this variable at the beginning of the script.

As shown in the upper example you can also define string variables for script coding:

strSelect = “SELECT Sample.SampleName FOR Barbital INTO aTable”

or considering the name substitution:

strSelect = “SELECT $strSmplName FOR $strCompound INTO aTable”

To use this type of construct you just write in the script editor the string variable with the $-sign as prefix:

$strSelect

The script interpreter first will substitute the sample and compound name string variables by their contents and then do the selection itself which means in this case to select all sample names for Barbital.

As you have seen, string variables can also contain string variables (with $-sign as prefix). The string variables are substituted by their contents at the moment when the script interpreter is evaluating and executing the command either from a string variable as shown in the last example:

$strSelect

or from the entered command itself as shown in the previous example:

Page 84: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

84 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

SELECT $strSmplName FOR $strCompound INTO aTable

There is another very special sign for substitution. Peak parameters such as area, height but also tailing factor, resolution etc. are dependent on the chosen compound. You usually select these parameters for specific compounds. The script editor allows you to use the ! (exclamation mark) for selecting these peak parameters referring to the currently used compound:

SELECT !Peak.Area FOR $strCompound INTO aTable

Executing this command will select only peak areas of the compound defined in strCompound (e.g. Barbital). The ! stands for Barbital or whatever compound name was defined in the strCompound string variable.

You could have also written:

SELECT Barbital.Peak.Area FOR $strCompound INTO aTable

but then you are fixed to Barbital or

SELECT $strCompound.Peak.Area FOR $strCompound INTO aTable

but using ! is simpler.

Example: Creation of a report calculation script

Remember the example which you have developed in the previous section. In this chapter this example is used to demonstrate the advantages of using string variables. This adds more flexibility when adapting the script to other data sets.

First you define some string variables. This will become the defining section in your script. Then replace the appropriate part of the script with these variables. This will become the implementation part of the script. Finally you add some comments to the script.

NOTE In case the described script cannot get achieved, a copy of this example script is located on the ChemStation Plus installation CD. Browse the support folder for refined_example.cct. To import the example script to the demo database, use the import function of the custom calculator.

Page 85: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 85

Using the Custom Calculator 3Custom Calculator Script Editor

Defining string variables

Define a string variables for the compound name. Then it will be easy to replace the compound name in case you need to use this script for another compound:

strCompound = “Antipyrine”

Define a string variable for the response column name. Then it becomes easier to switch between area and height as response indicator:

strResponse = “Peak.Area”

For better readability define a string variable for the selection. This makes it easier for modifying the complete selection. You can even use the upper defined string variables as part of this one with the $-sign as prefix:

strSelection="Sample.'Sample Name', Sample.'Sample Calibration Level', Sample.'Sample Type', 'Acq Sequence'.'Acq. Sequence Name', 'Raw Data'.'Raw Data File Name', !$strResponse"

Applying string variables

The first select statement will turn into:

SELECT $strSelection FOR $strCompound INTO '1_overview'

and the second one will be reduced to:

FROM '1_overview' SELECT $strSelection FOR $strCompound WHERE '1_overview'.Sample.'Sample Type'="Calibration" INTO '2_calib_runs'

The three calculation statements will simplify to

GROUP '2_calib_runs' BY 'Sample.Sample Calibration Level' DO rstdev ($strCompound.$strResponse) AS 'RSTD_Antip_area' INTO '3_Calibstatistics'

'5_compound_results'.'rel_dev'='5_compound_results'.'delta_Ant'/'5_compound_results'.$strCompound.$strResponse

Commenting script

The last step is adding meaningful comments to separate the definition section from the implementation section and to simplify the customization of the definition section. See also the refined_example.cct template on the installation CD-ROM.

Page 86: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

86 Agilent ChemStore C/S Concepts Guide

3 Using the Custom CalculatorCustom Calculator Script Editor

Page 87: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

87

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

4Using the Report Template Editor

What is the Report Template Editor? 88

Preparing to Use the Report Template Editor 90

Using the Report Template Editor 93

Understanding Report Template Components 96

Understanding Tables 106

Generating a Report Using a Template 121

Using Automatic Reporting 121

Page 88: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

88 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorWhat is the Report Template Editor?

What is the Report Template Editor?

The Report Template Editor is an application that is available by clicking the left icon under the name of the report (Compound Amounts [(built-in)] in the figure below) or choosing a menu item from the Report menu in the main dialog box in ChemStore C/S.

The Report Template Editor is used to create and edit report templates, which are then used to generate reports. Because you may want to use the same report many times, but with slightly different specifications and with different data, the Report Template Editor was designed to allow you to create easily modifiable report templates.

Figure 37 Report template editor

Figure 38 Report Template Editor

Page 89: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 89

Using the Report Template Editor 4What is the Report Template Editor?

When you create a report template, you specify such things as the contents and placement of a header or footer, tables or charts to include and the fonts used in each area of the report. In effect, you create a specific design for the layout and contents of a report and save the design to use again the next time you want to generate that report. In addition, you can change the design of any report template and save it under a new name.

Page 90: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

90 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorPreparing to Use the Report Template Editor

Preparing to Use the Report Template Editor

Retrieving the report data set

In ChemStore C/S, you query the database based on data fields that you choose and specify values for. The group of data records that result from the query is called the current data set (also known as a snapshot). Then, you select a report template and use it to generate a report. This report prints information contained in the current data set according to the structure of the template. You can even restrict the data that appears on the report further, but if the data is not in the current data set, it does not appear on a report that was generated using the current data set.

Page 91: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 91

Using the Report Template Editor 4Preparing to Use the Report Template Editor

Understanding the Report Template Editor Tree Views

The Report Template Editor uses tree views to display various objects.

The first object that is displayed in a tree view, when you open a template, is the report template itself (on the left-hand side of the main dialog). When you edit an existing report template, the Report Template Editor main dialog box contains a tree view of the structure of the template. The report template name is the top of the tree. Directly under the template name, you might see report template components:

• page headers and footers

• data sections

• section headers

• section elements, such as tables, charts, calculations, chromatograms and spectra

Figure 39 Tree View of Report Template in Report Template Editor

Page 92: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

92 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorPreparing to Use the Report Template Editor

Anything in a tree view that can be expanded (like a directory structure in Windows Explorer) is adjacent to a plus sign in a box. Page headers never have a plus sign next to them because they are never expanded. But data sections contain tables and charts; so if you click the plus sign next to a data section name, you can view the elements that make up the data section.

See “Understanding Report Template Components” on page 96 to learn more about the sections and elements that make up a report template.

Page 93: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 93

Using the Report Template Editor 4Using the Report Template Editor

Using the Report Template Editor

Using Dialog Boxes to Create Report Sections

You use the Report Template Editor main dialog box to create the sections and elements that you want to appear on the report by using menu items or by right-clicking the mouse.

The parameters associated with the section or element that you are editing appear in the right-hand side of the dialog box. For example, if you are adding a page header, the right-hand side of the dialog box contains parameters such as those that specify whether or not to print the header on every page or whether or not to print a border around the header, etc.

Figure 40 Menu for Creating Elements of Data Sections

Page 94: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

94 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUsing the Report Template Editor

Then, on the bottom of the dialog box, there are one or more buttons that you click to reach the dialog boxes used to edit the parameters. When editing a page header, there is only one button on the Report Template Editor main dialog box and it is labeled “Edit”. When you click it, you can edit the many different parameters associated with a header such as the number of columns and rows in the header and what text or data items to print in the columns/rows.

If you are editing a table, there are many more buttons at the bottom of the Report Template Editor main dialog box because there are many more parameters that can be configured for a table than for a header.

Figure 41 Page Header Dialog Box on Report Template Editor

Figure 42 Table Dialog Box on Report Template Editor

Page 95: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 95

Using the Report Template Editor 4Using the Report Template Editor

Using Tree Views for Tables and Charts

Another area where a tree view is used is during data field selection, such as for table columns or chart axes. It is convenient to use a tree view when selecting database fields because fields are members of records that are themselves members of a data set. For example, if you are selecting the columns for a sample table, you would expand the view of the data set that contains the information about sample organization so that you could view all of the fields associated with samples.

Figure 43 Tree View for Designing a Table

Page 96: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

96 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Report Template Components

Understanding Report Template Components

A report template comprises one or more of the following objects: a header, a footer, page breaks and data sections. Data sections contain report elements such as tables, calculations, chromatograms, spectra and charts. Any one of these sections or elements can be excluded from a generated report without deleting it from the report template simply by using a check box. Each of the components of a report template is called an object because you can edit them on an individual basis and they do not depend on the other components that you have selected for the template.

Page Headers and Footers

You can set up a page header or footer for the report template to indicate Report Name, Report Comment, User, Time, Page, Report Modified Date and Report Modified By. Each of these items can be placed in one of three columns in the header and formatted with any font and size that your computer supports. In addition, each of these items can be labeled and justified within the column. Graphics, such as your company logo, can also be placed in the header or footer.

Page 97: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 97

Using the Report Template Editor 4Understanding Report Template Components

You edit headers and footers using a dialog that allows you to see where each item is placed, in relation to the header and in relation to other items in the header. This dialog is set up so that you can see what the header/footer looks like, as you edit it.

Data Sections

Data sections contain one or more report elements, such as tables, calculations, charts, chromatograms and spectra. There are few parameters directly associated with a data section, as you can see if you click on a data section name in the tree view of a report template. There is a description field (which appears in the tree view) and there is a check box for whether or not to include this data section in a report. The one button, “Repeat”, that appears on the dialog box is used to select fields for repeating. Thus, the repeat function has a major influence on the internal order of the report you generate. It’s the key to create all different type of reports.

Figure 44 Editing Headers and Footers

Page 98: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

98 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Report Template Components

Repeating a Section

You can repeat a data section for every unique combination of the fields that you select. For example, suppose you want to create a summary report for sequences, including statistical calculations. However, you don't like to calculate one overall statistic over all calibration levels and sample runs. To report calibration levels and sample types separately, the items 'Sample calibration level' and 'Sample type' needs to be added to the repetition section, as shown in Figure 45.

Figure 45 Data Sections

Page 99: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 99

Using the Report Template Editor 4Understanding Report Template Components

Figure 46 Repetition

In our example, we also want to avoid cross-sequence and cross-study statistics, so these items are also added as separators.

Finally, we like to report the sequence results for each compound in an individual table, including statistics. There for, the item 'Compound name' is added.

Another important criteria is the order of the field items in the data section. The lowest item in the list gets iterated first. In our example, the pages of the report would show the following order:

Compound A (Cal.level 1),Compound B (Cal.level 1), Compound A (Cal.level 2), Compound B (Cal.level 2), Compound A (samples), Compound B (samples)

To get all reports for Compound A in one series of report pages, the item 'Compound Name' needs to get moved above the item 'sample type'. This would result in the following printing order:

Compound A (Cal.level 1),Compound A (Cal.level 2), Compound A (samples), Compound B (Cal.level 1), Compound B (Cal. level 2), Compound B (samples).

Page 100: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

100 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Report Template Components

Calculated Fields

You can create calculated fields using any of the fields in the current data set. Once created, calculated fields can be used to determine the repeating section as well as columns in a table or axes in a chart. Calculated fields are selected in the same way as other data fields except that they all appear together under the heading “Calculated Fields” in the tree view of the data set.

Section Headers

Section headers are used to differentiate report sections. In addition to the items listed for page headers/footers (Report Name, Report Comment, Page and so on), section headers can contain any data that you have selected as data items to repeat on. For example, if you chose to repeat the section based on the unique combinations of Sample Type and Study Name, then those two data items appear in the list of data items that can be placed in the section header. If you do not select repeating data items, then the only items you can place in a section header are the ones that are available to the standard page header.

Section Elements

Tables

You can create eight different types of tables; the type of table that you select determines the data set fields that you can use. The choices are injection, run, signal, compound, peak, column configuration, instrument components and audit trail. See “Understanding Tables” in this chapter to learn more about how these tables work.

For each table, you specify the following:

• the data fields to use as columns in the table

• the data fields to use to group data in the table

• the columns to sort on (up to 3)

• any summaries to generate for each group or for the overall report

• the font and size of the information that is printed

Page 101: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 101

Using the Report Template Editor 4Understanding Report Template Components

Whenever a table uses columns containing only names or strings and no numerical entries, rows containing the same entries will collapse into one row.

Whenever a table uses columns containing numerical entries, the table rows will remain separate, even though the entries in the rows appear identical, as in rows one and three in the Peak Table below:

Figure 47 Run Table

Page 102: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

102 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Report Template Components

Charts

You can create two types of chart for your report: compound or sample. For both types of chart, you specify the following:

• the data fields to be used to create the X and Y axes

• restrictions on the actual data used to create the chart and for compound charts, the compounds to include

• the data fields to use to group charts

• additional chart options including curve type, weighting, treatment of the origin, limits, titles and scaling

Table 2 Peak Table

Study Name Sample Name Inj Vol Compound Name Expected RT

Procaine Procaine decay 3.0 Amino-Benzoic Acid 1.734

Procaine Control Procaine 3.0 Amino-Benzoic Acid 1.734

Procaine Procaine decay 3.0 Amino-Benzoic Acid 1.734

Page 103: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 103

Using the Report Template Editor 4Understanding Report Template Components

When creating a sample chart, the values for the Y-axis can be chosen from per-run fields. For each Y-axis field that is selected, there is a single point per run, with the values for the various runs spread out along the X-axis of the chart.

When creating a compound chart, the values for the Y-axis can be chosen from per-compound fields. If the section has data for multiple compounds, that data may be overlaid on a single chart (with one curve per compound) or on multiple charts.

You can also group data in charts. If the data is grouped, then only data from a single group is included in a single chart. Therefore, there is one chart printed for each different group of data.

Figure 48 Compound Chart

Page 104: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

104 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Report Template Components

Calculations

You can include the results generated by a custom calculator script in your report.

The results to display are either a calculation table, a list of variables used for calculation or a calculation chart visualizing the calculation results.

In addition, the calculation script itself and the list of calculation errors and warnings can be added to the report.

For each items, you specify the following:

• the name of the calculation template

• the version of the calculation template

• the font and size of the information that is printed

Figure 49 Calculation Table

Page 105: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 105

Using the Report Template Editor 4Understanding Report Template Components

• For calculation tables and charts you need to select in addition the name of the result table generated by the calculation template. Once the calculation template and version are chosen, the calculation tables are selectable. The 'data series' dialog then allows to select the columns to be displayed or plotted.

Chromatograms

Chromatograms can be added to the report if they were stored in the database. Storing chromatograms is an option in the study transfer settings. When chromatograms are printed in the report, they are presented simply as graphical metafile representations of the chromatogram plots generated in the ChemStation software. They may not be further customized within the ChemStore reports.

If your ChemStation system used a color or greyscale printer driver, then the chromatograms stored in ChemStore will be in color. Otherwise, they will be black and white.

The size, position and grouping of chromatograms can all be specified from the report template designer.

Spectrum plots

If spectra were stored with the data, they can also be included on the report. Storage of spectra is also controlled through the study transfer settings. As with tables and certain charts, additional restrictions such as printing only spectra for a particular compound or group of compounds can be set up.

If your ChemStation system used a color or greyscale printer driver, then the spectrum plots stored in ChemStore will be in color. Otherwise, they will be black and white.

The size, position and grouping of spectrum plots can all be specified from the report template designer.

CAUTION In order to successfully create a report template including custom calculations it is important to have a calculation template that runs smoothly on the current data set.

Report and calculation templates are not hard-linked when exporting those to a file system. Both templates needs to be exported and imported individually.

Page 106: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

106 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

Understanding Tables

The concepts behind the process of constructing tables with the ChemStore Report Template Editor are the most complex within the software. In this section these concepts are presented so that you can understand the tables that appear from the choices you make.

Grouping and Summaries for Tables

Grouping

Data in tables can be grouped, meaning that all rows with the same value of a field are placed together. The groups are sorted by the field that controls grouping while the rows in each group are sorted according to the sort order you select. If you choose to group rows in a table, then you can save a column in the table by not including the grouping item as a column. In fact, if you have set up the columns in your table and then you set up grouping, using a field that is already represented by a column, a message box appears letting you know of the duplication.

For example, you can group the rows of a table on Study Name so that all data for the same study appears together. The grouping data item, Study Name, appears at the top of each group and then all the data for that group is sorted according to any sort you set up. Another way to do this is to include Study Name as a column and then assign Study Name as the first sort item. By grouping on Study Name instead, though, you can create space in your table to include another column.

Page 107: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 107

Using the Report Template Editor 4Understanding Tables

Statistics

Statistics can be printed in table summaries. You can request an overall summary and/or group summaries (if you have specified a data item for grouping). Summaries consist of up to seven lines of statistical information such as Mean, Std Deviation, Minimum, Maximum, etc. You can specify whether or not a column is included in group or overall summaries.

Figure 50 Compound Table

Page 108: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

108 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

Understanding Table Types

Relationships between Table Elements

The relationships between table elements, even within a small database, can be quite complex. The figure below shows an example of how some table elements within a small database are related to one another.

Within the figure, injections, runs, compounds, signals and peaks are related to one another in the following ways:

• The database holds data from two different injections. Each of these has been analyzed twice, so there are a total of four runs.

• During the first injection, signals (chromatograms) I and II were acquired (e.g., signals from two different detectors, or two different channels of a multi-channel detector). During the second injection, three signals (I, II and III) were analyzed.

• In run 1 only the compound X was detected, but run 2 found both compounds X and Y. Runs 3 and 4 both found compounds Y and Z.

Figure 51 Relationships between Table Elements

?

Injections Runs Compounds Peaks

1

2

1 (1A)

2 (1B)

3 (2A)

4 (2B)

X (1)X (2)Y (2)Y (3)Y (4)

Z (3)Z (4)

Signals

I

II

III

1, I, 1.0 main1, II, 1.0 qual2, I, 1.0 main2, II, 1.0 qual2, I, 2.0 qual2, II, 2.0 main2, I, 3.0 unk2, II, 3.5 unk3, I, 2.0 main3, II, 2.0 3, III, 2.0 3, I, 4.5 maincontinued

Page 109: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 109

Using the Report Template Editor 4Understanding Tables

• The results for signal I from run 2 include peaks at 1.0, 2.0 and 3.0 minutes. Signal II for that run has peaks at 1.0, 2.0 and 3.5 minutes.

• Signal I was used to quantitate Compound X, so the peak (2, I, 1.0) is identified as a “main” peak. The peak at the same time from signal II is still associated with Compound X, but not as the main peak.

• There are also two peaks associated with Compound Y for run 2; in this case Signal II is designated the main peak.

Note that the peaks for Run 2 at 3.0 and 3.5 minutes are not associated with ANY compound.

This example is now used to help explain the Injection, Run, Signal, Compound and Peak Tables.

Injection Table

In an Injection Table, the only fields (columns) that you can include are those with one value per injection (for example, injection date/time, name of the instrument used, vial number, data file name). Effectively, the data model shown in Figure 51 on page 108 is simplified to that shown in Figure 52.

Generally speaking, the table in the report will have no more than one row for every injection in the snapshot database (in this case, two rows), see Table 3 on page 110.

Figure 52 Data Model for an Injection Table

1

2

Injections

Page 110: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

110 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

There are several reasons there might be fewer rows than this:

• If the table is in a section that is repeated, each repetition of the table will only have the rows that are associated with the values of the repeat fields for that repetition. For instance, if the section is repeated on compound name, then there will be three repetitions (one for X, Y and Z). The injection table in the first repetition would have only injection I (because compound X is only detected in runs based on the first injection). The “Y” repetition would have injections I and II, and the “Z” repetition would have only a row for injection II.

• If there are any data restrictions specified for the table element, then any row which doesn’t meet the criteria will be eliminated.

• If the table does not contain any numeric or date/time columns, then only unique rows are shown in the table. In the above example, if only the “instrument” and “operator” columns were included then there would be only a single row, rather than two rows with identical contents.

Table 3 Example of Table in a Report

Injection Date Instrument Operator Vial Number

1-Apr-1977 08:00 LC System A Fred 10

1-Apr-1977 08:20 LC System A Fred 33

Page 111: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 111

Using the Report Template Editor 4Understanding Tables

Run Table

The only fields that you can include in a Run Table are those with one value per injection or one value per run (for example, study name, custom field values, processing method name). The data model from Figure 51 on page 108 is reduced to that shown in Figure 53.

Generally speaking, the table in the report will have no more than one row for every run in the snapshot database (in this case, four rows), see Table 4.

Figure 53 Fields in a Run Table

Table 4 Example of Table in a Report

Run No. Injection Date Vial Run Version Approval Patient Code

1 1-Apr-1977 08:00 10 2 Pending PKH9653

2 1-Apr-1977 08:00 10 1 Approved PKH9653

3 1-Apr-1977 08:20 33 2 Pending PKH9004

4 1-Apr-1977 08:20 33 1 Approved PKH9004

1

2

1 (1A)

4 (2B)

3 (2A)

2 (1B)

Injections Runs

Page 112: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

112 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

As in the case of Injection Tables, there may be fewer rows for the following reasons:

• The table is in a repeated section.

• There are data restrictions specified for the table element.

• No numeric or date/time columns are present and some rows would otherwise be identical. For instance, if you include only the Approval OR Patient Code columns, there would be only two rows. If you include both Approval AND Patient Code, there are still four rows, as there are four unique combinations for these columns.

Signal Table

The only fields that you can include in a Signal Table are those with one value per injection, run or signal (for example, signal description, start time, end time, type). The accessible data model looks like Figure 54.

Generally speaking, the report will have no more than one row for each combination of run and signal. In this case, there are two signals associated with runs 1 and 2, and three signals associated with runs 3 and 4, so there may be as many as ten rows, see Table 5 on page 113.

Figure 54 Data Model of a Signal Table

1

2

1

4

3

2

Injections Runs

I

II

Signals

III

Page 113: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 113

Using the Report Template Editor 4Understanding Tables

As in the case of Injection and Run Tables, there may be fewer rows for the following reasons:

• The table is in a repeated section.

• There are data restrictions specified for the table element.

• No numeric or date/time columns are present and some rows would otherwise be identical. For example, if only the channel column were included, the table would have only three rows (one for each of the three signals).

Table 5 Example of Table in a Report

Run No. Version Vial Channel

1 1 10 DAD1 A, Sig=282,8 Ref=550,100

1 1 10 DAD1 B, Sig=295,8 Ref=550,100

2 2 10 DAD1 A, Sig=282,8 Ref=550,100

2 2 10 DAD1 B, Sig=295,8 Ref=550,100

3 1 33 DAD1 A, Sig=282,8 Ref=550,100

3 1 33 DAD1 B, Sig=295,8 Ref=550,100

3 1 33 DAD1 C, Sig=320,8 Ref=550,100

4 2 33 DAD1 A, Sig=282,8 Ref=550,100

4 2 33 DAD1 B, Sig=295,8 Ref=550,100

4 2 33 DAD1 C, Sig=320,8 Ref=550,100

Page 114: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

114 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

Compound Table

The Compound Table may include fields with one value per injection, run, or compound. In addition, it may include peak and signal fields. However, the only peak values that will be included are those for the main peak for each compound result. This means that there will be only one value for, say, peak area associated with compound X in run 1. Similarly, the only signal information included in the Compound Table is from the signal associated with the main peak for the compound.

The data model in this case looks like that shown in Figure 55. Note that only main peaks are shown, and they have a 1:1 relationship with both compound result and signal entries.

In this example, there are three different compounds, but there are seven sets of compound results for the four runs (because some compounds are detected in more than one run), so the table may have up to seven rows, see Table 6 on page 115.

Figure 55 Data Model for Compound Table

Injections Runs Compounds Peaks Signals

1

2

1

2

3

4

X (1)

X (2)

Y (2)

Y (3)

Y (4)

Z (3)

Z (4)

I (1)

I (2)

II (2)

I (3)

I (4)

II (3)

II (4)

1,I,1.0 main

2,I,1.0 main

2,II,2.0 main

3,I,2.0 main

4,I,2.0 main

3,II,4.5 main

4,II,4.5 main

continued

Page 115: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 115

Using the Report Template Editor 4Understanding Tables

As in the case of other tables, there may be fewer rows for the following reasons:

• The table is in a repeated section.

• There are data restrictions specified for the table element.

• No numeric or date/time columns are present and some rows would otherwise be identical. For example, if you include only the Cpd Name column, the table will have only three rows (one for each of the three compounds).

Peak Table

Like the Compound Table, Peak Tables may include fields with one value per injection, run, compound, signal or peak. Unlike peaks in the Compound Table, peaks in the Peak Table are not restricted to the main peak for each compound. That is, the table may contain information for all peaks, even those not associated with any compound. The data model in this case looks like that shown in Figure 56 on page 116.

Table 6 Example of Compound Result Table

Run No. Vial Cpd Name Cpd Amt RT Channel

1 10 X n.nnn 1.000 DAD1 A, ...

2 10 X n.nnn 1.000 DAD1 A, ...

2 10 Y n.nnn 2.000 DAD1 B, ...

3 33 Y n.nnn 2.000 DAD1 A, ...

3 33 Z n.nnn 4.500 DAD1 B, ...

4 33 Y n.nnn 2.000 DAD1A, ...

4 33 Z n.nnn 4.500 DAD1 B, ...

Page 116: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

116 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

In this example, there are eight peak entries just for the first two runs.

If you were to create a Peak Table with exactly the same columns as the previous Compound Table example, the Peak Table would have many more rows, as shown below. Note that peaks not associated with a target compound have no amounts and have compound names like “Unknown n”.

Figure 56 Data Model for Peak Table

1

2

Injections

Runs Compounds

1

4

3

2

X (2)

X (1)

Y (2)

Y (3)

Y (4)

Z (3)

Z (4)

2, I, 3.0 unk

2, II, 3.5 unk

3, I, 2.0 main

1, I, 1.0 main

2, II, 2.0 main

2, I, 2.0 qual

2, II, 1.0 qual

2, I, 1.0 main

1, II, 1.0 qual

3, II, 2.0 qual

3, I, 4.5 main

3, III, 2.0 qual

Peaks

?

?

continued

I

II

Signals

III

Page 117: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 117

Using the Report Template Editor 4Understanding Tables

As in the case of other tables, there may be fewer rows for the following reasons:

• The table is in a repeated section.

• There are data restrictions specified for the table element.

• No numeric or date/time columns are present and some rows would otherwise be identical. For example, if you included only the Cpd Name column, the table would have five rows (for the compounds X, Y, Z, Unknown 1 and Unknown 2).

Table 7 Example of Peak Table in a Report

Run No. Vial Cpd Name Cpd Amt RT Channel

1 10 X n.nnn 1.000 DAD1 A, ...

1 10 X n.nnn 1.000 DAD1 B, ...

2 10 X n.nnn 1.000 DAD1 A, ...

2 10 X n.nnn 1.000 DAD1 B, ...

2 10 Y n.nnn 2.000 DAD1 A, ...

2 10 Y n.nnn 2.000 DAD1 B, ...

2 10 <Unknown 1> 3.000 DAD1 A, ...

2 10 <Unknown 2> 3.500 DAD1 B, ...

3 33 Y n.nnn 2.000 DAD1 A, ...

3 33 Y n.nnn 2.000 DAD1 B, ...

3 33 Y n.nnn 2.000 DAD1 C, ...

CAUTION When multiple channels of data are collected, the peak table can have more entries than any other table in the database and can consume significant amounts of resources (computer time and printer paper!).

Page 118: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

118 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

Instrument Components Table

In an Instrument Components Table, you can include columns with one value per run, injection or instrument components. Each run is associated with some single instrument, which is associated with possibly several components.

If the above diagram represents the data we are dealing with, an Instrument Component Table would have a maximum of 14 rows (the four components of Instrument 1 are used for Injection 1 and runs 1 and 2; the three components of Instrument 2 are used for runs 3 and 4), as represented in Table 8 on page 119.

Figure 57 Data Model for Instrument Components Table

1Instrument I

2Instrument II

1

4

3

2

VWD 2

Pump 1

DAD 2

Pump 2

DAD 1

FLD 1

Oven 1

Injections RunsComponents

Page 119: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 119

Using the Report Template Editor 4Understanding Tables

The number of rows may be reduced by any of the three factors that apply to other tables mentioned earlier.

Column Configuration Table

This table works exactly like the Instrument Components Table, with column configuration information replacing the instrument components information (each injection may be associated with multiple column configuration entries). So the only columns you can include are injection, run, or column config fields.

Table 8 Example of Instrument Components Table

Component Name Injection No. Run No.

Pump 1 1 1

Oven 1 1 1

FLD 1 1 1

DAD1 1 1

Pump 1 1 2

Oven 1 1 2

FLD 1 1 2

DAD 1 1 2

Pump 2 2 3

DAD 2 2 3

VWD 2 2 3

Pump 2 2 4

DAD 2 2 4

VWD 2 2 4

Page 120: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

120 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

Audit Trail Table

In this table, you may include columns with one value per injection, run or audit trail entry. Each run entry is associated with one or more audit trail entries (one for each entry when you change the approval status or custom field values).

Figure 58 Data Model for Audit Trail Table

Table 9 Example of an Audit Trail Table

Inj. No. Run No. Approval Status Audit Time

1 1 Pending 01-Apr-97 08:10

1 2 Pending 02-Apr-98 10:15

1 2 Approved 02-Apr-98 14:30

2 3 Pending 01-Apr-97 08:30

2 4 Pending 02-Apr-98 10:20

2 4 Approved 02-Apr-98 14:30

1

2

1

4

3

2

Injections Runs

Pending

Approved

Pending

Pending

Approved

Pending

AuditEntries

Page 121: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 121

Using the Report Template Editor 4Understanding Tables

Generating a Report Using a Template

When you have finished preliminary editing of a report template, you can preview the report that will be created using the template. Then, you can make any changes you need to make in the template without ever having left the Report Template Editor application. When you are satisfied with the template, then you can print the report.

From within the ChemStore C/S application, you select the set of analysis results you want to report on by specifying a query, and then select the report template you want to use.

Using Automatic Reporting

ChemStore C/S allows automatic reporting after a sequence. After a sequence a sequence summary report can be generated based on the results of the current sequence. The reports are initiated via a DDE (Microsoft dynamic data exchange) link to ChemStore C/S. The Preferences dialog box of the ChemStore menu in the ChemStation Method and Run Control View allows you to set up the name and type of report for post-sequence. For the Post-Sequence report you can choose between any of the available ChemStore C/S reports (standard templates and your own customized templates).

Page 122: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

122 Agilent ChemStore C/S Concepts Guide

4 Using the Report Template EditorUnderstanding Tables

Page 123: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

123

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

5Data Security

Introduction 124

Setting Up and Managing Users 127

Setting Up Organizational Information 131

Setting Up and Managing ChemStore C/S Elements 134

Managing Archiving and Dearchiving on Admin Client 151

Audit Trail 155

E-Mail Notification 157

Validation and Security of Printed Files 161

Page 124: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

124 Agilent ChemStore C/S Concepts Guide

5 Data SecurityIntroduction

Introduction

Data security and integrity is a major concern when an analytical laboratory is working under regulatory control, such as Good Laboratory Practice (GLP) and current Good Manufacturing Practice (cGMP) regulations.

ChemStore C/S provides the following capabilities to fulfill these needs:

• Logging on to ChemStore C/S is password-protected; users must supply a valid password to start ChemStore C/S. In addition, many other activities that make changes to the database require the user to supply the user ID and password.

• ChemStore C/S uses a system of user permissions to prevent unauthorized users from executing management functions such as defining report layouts or deleting runs from the database. The user profile specifies the user’s permissions; you can use one of the built-in user profiles or customize user profiles for each individual user.

• ChemStore C/S does not allow stored analysis results to be modified; they are protected. The user can change only the custom fields or the approval status of a record. All changes are documented in an audit trail.

• The client/server version of ChemStore C/S provides archive functions which allow you to archive your data when your project is completed and to restore them if you need to review them at a later date.

• In order to trace the transfer of the data, ChemStore C/S provides an audit trail that records the following events:

• additions to the database (analysis runs),

• changes to the custom fields,

• changes to the approval status of a run,

• adding to or deleting from a batch

• changes to the archive status of a run.

NOTE If you require the software to fulfill explicitly the GLP/GMP regulations, add the Security Pack G2183AA, available from your local Agilent Technologies supplier, to your ChemStore C/S installation.

Page 125: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 125

Data Security 5Introduction

• Both the ChemStation and ChemStore can be locked while they are running, to prevent tampering by an unauthorized person. The lock-out may be Private, when either the current user’s or an Administrator’s user ID and password must be supplied to unlock the session, or Non-Private, when any valid user ID/password combination is sufficient to unlock the session.

Security

ChemStore C/S provides two levels of security to prevent unauthorized access or use of the database. The first level of security is the password protection, which prevents unauthorized access; the second level of security is through user permissions, which restrict the available facilities to authorized users. Passwords and user permissions are not global; they are configured individually for each ChemStore C/S database.

Password Protection

ChemStore C/S is password-protected; in order to start ChemStore C/S, the user must supply a valid password. Several other activities within ChemStore C/S also need password verification in addition to user permission, for example:

• changing to a different user

• changing passwords

• editing custom fields

• reopening a previously-archived run

• deleting a run

• archiving and dearchiving runs

• approving and rejecting runs

NOTE For reasons of security, we recommend that you change the default user passwords after installation of ChemStore C/S. See “Server—Change default Passwords” in the ChemStore C/S Installation Guide.

Page 126: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

126 Agilent ChemStore C/S Concepts Guide

5 Data SecurityIntroduction

Password Settings

A number of conditions apply to the validity and longevity of passwords. The administrator can specify the values for these conditions.

Minimum length This is the minimum acceptable length (in characters) of a password. Passwords shorter than the minimum length are invalid and are rejected by ChemStore C/S. The default minimum password length is 8 characters.

Password validity This is the length of time (in days) over which the password remains valid. The password expires after the specified validity, and a new password must be provided. The default validity is 90 days.

Minimum password recycle

This is the minimum number of new, unique passwords that a user must use before a password can be re-used. The default value is 12, which means that a user must change the password at least 12 times before re-using the original password.

Retries before rejection This is the permitted number of consecutive unsuccessful attempts that a user can enter before ChemStore C/S rejects the user. The default value is 3. If the maximum number of retries is exceeded the current password is invalidated.

Page 127: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 127

Data Security 5Setting Up and Managing Users

Setting Up and Managing Users

Many of the facilities of ChemStore C/S are safeguarded from modification by unauthorized personnel by applying a set of user permissions. Each ChemStore C/S user is granted permission to perform certain tasks. Individual users can be granted different sets of permissions depending on each laboratory’s business needs and each user’s job requirements.

A ChemStore C/S user is someone who has been registered in ChemStore C/S with a user log-on name, a user display name, which is shown in audit trails and other reports, a password and a set of permissions. The set of permissions identifies the ChemStore C/S facilities that are available to the user.

Only an Administrator, or a user with the relevant permissions, can set up a new ChemStore C/S user. To assist in configuring user permissions for new users, four standard user profiles are provided:

Operator The Operator is the basic user. The Operator can retrieve data from the database, but has no permission to make changes to the ChemStore C/S configuration.Permissions: Create batch only.

Chemist The Chemist can change the approval status of retrieved data, and can change the values of custom fields. Chemists can also save their user interface settings and, in the client/server version, can archive/dearchive and retrieve files.Permissions: Archive/dearchive runs, First level approval, Create batch, Create/edit queries, advanced filters, calculation and report templates. Define columns/expressions, Edit/save settings.

Page 128: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

128 Agilent ChemStore C/S Concepts Guide

5 Data SecuritySetting Up and Managing Users

Users can be set up using one of the four standard user profiles, but individual permissions may be granted or refused, to give each user a custom set of permissions. The permissions contained in the templates provided are shown in Table 10.

Lab Manager The Lab Manager has all the permissions of the Chemist and, in addition, can create and modify studies and custom fields. The manager is permitted to approve runs on a 2nd level and to change the data reload configuration. The lab manager also has the ability to allocate ChemStore C/S settings to other users, and to delete runs.Permissions: All except Admin client test mode (a special diagnostic and maintenance mode), Administer DB server, Administer users (see Table 10 on page 128), Assign studies to users, define the approval configuration, perform database migration or compact a stand-alone database.

Administrator The Administrator has the same permissions as the Lab manager, but in addition has the ability do administer users, Assign studies to users and, in the client/server version, to administer the ChemStore C/S server and perform database migration.Permissions: All (see Table 10 on page 128).

NOTE In case Security Pack is installed, permissions for tasks like pausing the spooler, deleting spool jobs or pre-selecting a database are handled on the operating system level. The standard operation of the software should be performed by a member of the Windows user group users. See also ChemStation Plus Security Pack User’s Guide.

Table 10 User Permissions Templates

Permission Operator Chemist Lab. Manager

Administrator

Admin client test mode Yes

Administer DB server Yes

Administer custom fields Yes Yes

Administer users Yes

Administer user settings Yes Yes

Assign studies to users Yes

Archive/de-archive runs Yes Yes Yes

Page 129: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 129

Data Security 5Setting Up and Managing Users

No permission is needed for the following tasks:

• Storing results in the database.

• Retrieving and reviewing data using saved queries to which the user has been granted access.

• Accessing any user setting (query, filter, report template or user interface setting) that is not controlled by security permissions or that the user owns or has been granted access to.

• Reloading raw data files, methods files or sequences files to the ChemStation requires the Create batch permission.

1st level approval Yes Yes Yes

2nd level approval Yes Yes

Create/edit queries Yes Yes Yes

Create/edit advanced filters Yes Yes Yes

Create/edit report templates Yes Yes Yes

Create/edit calculator templates

Yes Yes Yes

Create/modify study Yes Yes

Define columns/expressions Yes Yes Yes

Delete runs Yes Yes

Edit custom field values Yes Yes Yes

Reopen runs Yes Yes

Create batch Yes Yes Yes Yes

Compact database (standalone only)

Yes

Edit/save settings Yes Yes Yes

Migrate database Yes

Table 10 User Permissions Templates (continued)

Permission Operator Chemist Lab. Manager

Administrator

Page 130: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

130 Agilent ChemStore C/S Concepts Guide

5 Data SecuritySetting Up and Managing Users

Deactivating Users

An Administrator (or other user with the permission to administer users) can deactivate any registered ChemStore C/S user. Deactivated users cannot log on to ChemStore C/S.

Page 131: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 131

Data Security 5Setting Up Organizational Information

Setting Up Organizational Information

Only users with the requisite permissions can set up and manage Studies and Custom Fields. For standard user profiles, only Lab Managers and Administrators have the requisite permissions.

Setting Up and Managing Studies

A study may only be created or modified by an Administrator, Lab Manager or other user with the requisite permission. A new study can be created from scratch, or can be based on an existing study. During the creation of the study, the creator identifies the items (chromatograms, spectra, raw data, method and sequence information) that will be stored by ChemStore C/S along with the results, and specifies the custom fields to be associated with the study. All custom fields that have been created for other studies and have not been deactivated are available for inclusion in the new study. Any new custom fields for inclusion in the study can also be specified at this time (see “Setting Up and Managing Custom Fields” on page 132). Irrespective of whether the custom fields included in the study are newly-created or already existing, the values in the custom field are unique to the runs in the study.

Once the study has been set up, users are either granted or denied access to that study depending on the study assignments (see “Assigning Studies” on page 133). Users who have been assigned to the study may add data into the study and retrieve runs using the study name as a search criterion.

When creating a new study, a user with sufficient privileges also has the option to change the approval mode, run lock mode and LIMS notification mode.

Studies can be given the status Active or Inactive by a user with the requisite permission. An active study is available for all activities of the ChemStation and ChemStore C/S. An inactive study is no longer available to the ChemStation for adding data into. It is still available in ChemStore C/S for retrieving data, but is not available for use as a template for a new study or for modification.

Page 132: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

132 Agilent ChemStore C/S Concepts Guide

5 Data SecuritySetting Up Organizational Information

Setting Up and Managing Custom Fields

Only an Administrator, Lab Manager or other user with the requisite permission may create or modify custom fields. The creation of a custom field requires the provision of a name and the specification of the data type (see “Using Custom Fields” on page 21); for some data types, additional information (units, a list of possible values) is also necessary. A newly-created custom field is available for inclusion in any study.

Data Entry for Custom Fields

Part of the specification of the custom field is the method of data entry. Custom field values can be entered automatically (by a ChemStation macro) or manually. Manual data entry can be specified as always required, when a valid value must be entered in the field before the transfer of the analysis results to ChemStore C/S is allowed. If a value is not always required, a default value may be specified; this default value can be overwritten prior to data transfer. In addition, data masks can be specified for some field types: maximum and minimum values for numeric fields, maximum string lengths for text fields. These data masks are specific to the study; if the custom field is used in other studies, its data masks can have different specifications.

If the data is to be entered into the custom field automatically, using a ChemStation macro, you must ensure that

• the macro returns the correct data type (in accordance with the custom field specifications)

• the macro is loaded in the ChemStation

For more details on macros, refer to the ChemStation on-line help on Commands.

Custom fields can be given the status Active or Inactive by a user with the requisite permission. If the status of a custom field is changed to inactive, it becomes unavailable to all ChemStation and ChemStore C/S activities except use as a query criterion.

Page 133: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 133

Data Security 5Setting Up Organizational Information

Assigning Studies

Only users who are assigned to a study will have access to that study. The assignment of studies to users can be done either during study creation and modification or through a separate interface where all study assignments can be edited.

Users will only be able to access studies that are assigned to them within the ChemStore and ChemStation software interfaces. This protects studies from being modified, accessed or have inappropriate data added by unauthorized users. Assigning a limited number of studies to each user also makes the software easier to use, as fewer studies and options are shown to the user.

A specific user permission “Assign Studies to Users” is required to make study assignments.

Page 134: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

134 Agilent ChemStore C/S Concepts Guide

5 Data SecuritySetting Up and Managing ChemStore C/S Elements

Setting Up and Managing ChemStore C/S Elements

Five major elements of the ChemStore C/S configuration can be saved and later reopened; these elements can also be specified for use by other ChemStore C/S users, even when they do not have the requisite permissions to create or save the elements. This setup limits data access to predefined data sub-sets. The creator of the element is assigned as the owner, but ownership can be transferred to another user by a user with the requisite permission.

In addition to allocating ChemStore C/S elements for use by other users and transferring ownership to other users, a user with the requisite permission can rename a saved element. Elements can also be modified by a user with the requisite permission and resaved either with a new name or with the same name, in which case the original element is overwritten.

Queries Queries can be saved with a name for later re-use; saved queries can be allocated for other users to use.

Filters Custom filters can be saved with a name for later re-use; saved custom filters can be allocated for other users to use.

Report templates Report templates can be saved with a name for later re-use; saved report templates can be allocated for other users to use.

User Interface Settings

User interface settings include all current configurations of all windows, tables and charts. these settings can be saved with a name for later re-use; saved user interface settings can be allocated for other users to use.

Calculation templates

Calculator templates can be saved with a name and version number for later re-use; saved calculator templates can be allocated for other users to use. The allocation of a calculator templates affects all versions with the same name.

Page 135: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 135

Data Security 5Archiving and Dearchiving

Archiving and Dearchiving

The client/server version of ChemStore allows you to archive (copy and save) runs present in the database to a separate file on disk or tape. After the runs have been archived, you can automatically delete the runs from the ChemStore C/S database, to free up space in the database.

Before you can archive, dearchive or reopen a file, your administrator must give you permission to do so. Your administrator also sets up and manages certain archive functions through the ChemStore C/S Admin Client. See the section “Managing Archiving and Dearchiving on Admin Client” on page 151.

If you have not deleted runs after archiving, you can still access the runs on the database, although you cannot modify (e.g., approve) them unless you reopen them first. Before you can reopen a deleted run you must dearchive and restore it to the database.

Archiving, deleting, dearchiving and reopening act on all versions of a run. If multiple versions of a run exist, all versions are archived - even if they are distributed over different studies. Similarly, if you specify to automatically delete the runs after archiving, all versions of the run are deleted.

ChemStore C/S provides two modes of archiving: automated and interactive.

Automated Archiving

A user with administrator permissions can set up to archive specified runs automatically at regular intervals. The runs to be archived are selected by a query. To set up new archive queries the user has to select Automated Archiver from the ‘administration’ menu of the Review Client.

The archive query section allows to enter information similar to the Create Archive Unit-dialog used for manual archiving. A name for the archive query has to be specified (mandatory and unique) as well as the name of the archive unit and the path. The archive unit name is extended with creation date, i.e. ‘<filename>-yyyy-mm-dd’.

Page 136: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

136 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

The checkbox Active allows to enable an archive query. The archive query will be shown as 'active' in the list of automatic archive queries. If the checkmark is removed, the query will be disabled showing a status inactive.

Optionally runs can be deleted from the database after they have successfully been archived.

The query criteria include items such as injection time, instrument name, operator name, acquisition sequence name, method name, sample name, study name, custom fields, injection operator, approval status and so on. Operands depend on the data category and can be negated with NOT.

Page 137: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 137

Data Security 5Archiving and Dearchiving

• Text based data: is equal, contains (wildcards)

• Numeric data: is equal, is greater than, is lower than, is between

• Date : is older than X days/months

For optimum performance, the number of clauses that can be defined for automated archives is restricted to 10.

Figure 59 The Automated Archives dialog box

Page 138: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

138 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

An individual frequency for the archival can be set for each archive query. The time interval can be configured, e.g. daily, weekly, monthly or in conjunction with a counter (every x days, weeks etc.). A First starting date when the query should be executed on the server has to be entered. The user must authenticate and confirm the creation or editing of an archive query with UserID and password, similar to the create archive unit dialog for creating manual archives. The button Test Query provides information of how many runs the query currently returns with the given query condition.

When an archive query is enabled it will run automatically for the first time as defined through the 'first starting date. The following archives will be created according to the specified time interval.

Page 139: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 139

Data Security 5Archiving and Dearchiving

Archiving Interactively

The users have also the option to select runs for archiving from the Archive/Delete view or graphics layout of the main ChemStore panel.

Note that the Archive icon is depressed, rather than the Sample Review icon.

Figure 60 Archive/Delete View of the main ChemStore panel

Page 140: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

140 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

The Archive/Delete view contains a similar toolset to the Sample Review graphics layout with the exception of the following three icons:

Enables you to specify where and when your archive unit will be stored.

Deletes all runs marked in the run list for deletion. This button is also present in the Delete view of the stand-alone version.

Reopens all runs marked in the run list for reopening and sets their approval status to Approval Pending (see “Run Approval Status” on page 29).

See “Sample View” on page 36 for more information on the Sample Review graphics layout.

You select the runs for archiving from the run list in the left panel of the Archive/Delete view, just as you would search data for reviewing or reporting in the Review view. Runs marked for archiving are protected from any further modification of approval status or custom field values. Since archiving can be postponed to a suitable time (for example, during non-working hours), runs marked for archiving are stored in a dataset until archiving has been completed.

In principle, manual archiving jobs do not interfere with automatically triggered ones. It operates on a ‘first in - first served’ basis. A run that is scheduled for archiving in two separate ‘jobs’ will be archived by the first job in the queue.

Figure 61 Create Archive Unit Panel

Page 141: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 141

Data Security 5Archiving and Dearchiving

Archive units

Archived runs are stored as archive units; an archive unit contains all versions of the specified runs. The archive unit is created either automatically or manually, as described above. When you set up the archive unit, you specify a file name and a path, and you can also add a description of the archive unit. These items can be used later to search for a specific archive unit. When ChemStore C/S creates the archive unit, it attaches a system-generated archive ID and a catalog file in XML format describing the contents of the binary archive unit. The file name corresponds to the name of the archive unit with the extension XML. Figure 62 on page 141 shows an example of an XML catalog file.

For enhanced archive management, for example in external document management systems (DMS), this catalog file can be used for indexing.

Figure 62 An example of an XML catalog file

Page 142: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

142 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

Archiving is protected by user permission and password. The Archive/Delete view as well as the ‘Automated archives’ menu item are available only to users with the correct permissions. To submit an archiving job the user must enter the user ID and password.

Deleting Runs

If there is no Security Pack for ChemStore C/S installed, it is possible to delete runs without archiving first. In the Delete view of the stand-alone version, and in the Archive/Delete view of the client/server version, you can mark runs explicitly for deletion. Data security is ensured by allowing only users with the required permissions (“Setting Up and Managing Users” on page 127) to perform this task, and by requiring password-protected confirmation that the runs are to be deleted.

In the client/server version, it is also possible to automatically delete runs from the database after they have been archived.

Reopening Runs

Runs that have been archived but not deleted are still in the database, but they are locked and cannot be modified. You can reopen archived runs for access by selecting them for reopening from the run list in the left panel of the Archive/Delete view and then clicking the Reopen icon described on the previous page.

Like archiving, reopening is available only to users with the correct permission.

CAUTION Runs that you delete are removed from the database and cannot be retrieved. Be sure that you only delete runs that you have already archived or no longer need.

Page 143: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 143

Data Security 5Archiving and Dearchiving

Dearchiving Runs

Runs that have been archived and deleted are no longer in the database. In order to access archived, deleted runs, you must first dearchive them (copy the runs back into the database) and then reopen them.

You can dearchive a complete archive unit, or you can select one or more runs to dearchive from within the archive unit. To select an archive unit for dearchiving, you use the Archive History, which displays all archive units in chronological order of creation. Because you may have many archive units in the database, making it difficult to find the one containing the required run(s), a search facility is available to allow you to search the list of archive units for the required one.

You can specify that dearchiving, like archiving, begin immediately (within the wake-up interval of the server); alternatively you can postpone dearchiving until a later time. Dearchiving copies all versions of the selected run in the archive into the database.

Dearchiving is protected by user permission, user ID, and password. Only users with the required permission have access to the Archive/Delete view, and the user’s ID and password must be entered before dearchiving will begin.

Generic XML Archive Interface

In addition, a generic XML-based archive interface can be optionally added to the ChemStore C/S server. It allows to implement a linkage to external archive management systems (DMS, HSM etc.).The archive interface is installed from the directory \G1410a\archiver_interface on the ChemStation Plus CD-ROM. Please refer to the readme.txt for installation details.

In case it is installed, ChemStore C/S notifies the generic XML-based archive interface every time an archive unit is created. The name and location of the archive unit and the corresponding XML file is passed on to the default plug-in. Depending on the nature of the plug-in, different activities, such as the transfer of the archive unit to a DMS followed by local deletion, can be started. After successful processing of the archive unit a notification message with an external ID for the storage location can be reported back to the generic XML-based archive interface, which sends this information to ChemStore C/S.

Page 144: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

144 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

A prerequisite for using the generic XML-based archive interface is the development of a plug-in. On request Agilent supplies a developer's kit with the necessary information to program such a plug-in.

A simple example for a plug-in comes along with the installation of the XML Archive Interface and it needs no further configuration. The simple plug-in triggers a dialog box popping up on the server PC each time a new archive file has been created.

The message and function of the simple plug-in can be changed by modifying the file AddArchive.cmd in the simple plug-in folder ("C:\Program Files\Agilent Technologies\XML-based Archive Restore\Examples\Simple Plugin").

A schematic overview of the data flow with the simple plug-in and the potential extension with a customized plug-in is given in Figure 64.

Figure 63 Message generated by simple plug-in of generic XML-based archive interface

Page 145: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 145

Data Security 5Archiving and Dearchiving

Restoring has to be done manually first by putting the binary archive file back to its old location and then by means of ChemStore in the same way as in previous revisions of ChemStore.

Understanding Archive Status

When a run is transferred from the ChemStation to ChemStore C/S, it is given the archive status Never Archived. A run that has been selected for archiving is given the archive status Archive Pending and is locked against further modification. If deletion after archiving has been specified, the run is given the status Archive Delete Pending. This status prevents runs being retrieved by queries. The statuses Archive Pending and Archive Delete Pending apply until the archive has been completed, when the run’s archive status is changed

Figure 64 Generic XML-based archive interface

genericarchive interface

scheduled archiving

dearchiving

Review Client

genericarchive Interface

DMS or HMS

Oracledatabase

manualarchival

(e.g. on tapes/CDs/DVDs)

archive unit+ XML file

archive unit+ XML file

SimplePlug-In

DMS or HMSplug-in

DMS or HMSplug-in

Page 146: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

146 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

to Archived for a completed archive operation. For Archive Delete operations, the archiving and delete information is only available from the archive unit interface in the admin client application.

If the archive fails — perhaps because the storage medium filled before the archive was complete — the runs are given the status Archive Failed. In such a case, your administrator must use the Admin Client to solve the problem and reschedule the archive.

A run that has been archived but not deleted — that is, it remains in the database — is locked against any change, such as a change to its approval status or to a custom field value, until the run is explicitly reopened. Then it is given the status Reopened. Reopened runs may be archived again in the normal way.

Certain rules exist for determining the status of a run when you dearchive runs to the database.

If the run no longer exists in the database, then the following rules apply:

• If you attempt to dearchive the last archive created, the run is transferred to the database and marked Dearchived.

• If the run was archived multiple times and you attempt to dearchive an archive that is not the most recent run, the run is transferred to the database and marked Read-Only.

If the run exists in the database after an actual delete and dearchive process, then the following rules apply:

• If you attempt to dearchive the most recent archive the run will be transferred to the database and have the status dearchived

• If you attempt to dearchive an archive dated that is not the most recent one, the run will be dearchived and marked read-only.If you dearchive as a next step a more recent but not the latest run version, the older dearchived run will be overwritten and all runs get the status read-only.In case you try to dearchive older versions of the run than those that are already dearchived, the application will not allow to dearchive these runs or to overwrite runs in the database

NOTE Archive status is not automatically updated; you must refresh the status by conducting another query on the runs of interest to update archive status.

Page 147: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 147

Data Security 5Archiving and Dearchiving

If the run still exists in the database and has not been restored from an archive unit so far, the application will not allow to dearchive or overwrite an existing run in the database.

You can find a listing of the runs that were overwritten or not overwritten for the last dearchive operation of an archive unit in the Archive Reports/View of the admin client. This information is also written to the server log file.

The relationships between the archive statuses are shown in Figure 65. See also the topic “Taking a Closer Look at the Archive/Dearchive Process” in this section for a specific example of how the archiving and dearchiving process affects the status of the run.

Reviewing Archive Status and Reports

To see the schedule of pending archives and dearchives, look under the Review menu at the top of the main panel of the Archive/Delete view (see Figure 60 on page 139). You can review the history of an archive from the same menu and assess why an archive or dearchive may have failed.

Automated archives are not listed in the schedule of pending archives because automatic archive queries are directly executed at the specified date and time.

You can also view the archive schedule and history from the Admin Client panel. Any user can view these reports without needing a password to access them on the Admin Client. You can also view the following additional reports through the Admin Client:

• Archive Units

• Runs in an archive unit

• Runs to be deleted

Figure 65 Archive Status Relationships

Never Archive

Archive Pending

Archive

De-Archive

ReopenedSelect for Archive

Archive

Reopen

ArchiveReopen(if not deleted)

Dearchive

(if deleted)

Page 148: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

148 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

• Audit Trail

• Information for an object

• Base path, wake-up interval and time-out interval

• Active spool/download table

Taking a Closer Look at the Archive/Dearchive Process

An example may provide a deeper understanding of the archive/dearchive process.

Example The following example shows two figures: Figure 66 on page 149 represents the steps in the process of archiving a run to several archive units. Figure 67 on page 150 is an example of the dearchive process when the run is no longer in the database based on Figure 66.

Page 149: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 149

Data Security 5Archiving and Dearchiving

Figure 66 Example Archive Process

Archive Status for Run10

Never Archived

Archive Pending

ArchivedReopened(Changes Made

12/6)

Archived

Database no longer contains run10

Select for archive

Archive

Reopen

Archive at specified time

Archive Again/Delete

Database

Archive Unit ID

ID1(12/3)

ID2(12/7)

ID3(12/11)

Tape/Disk

Page 150: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

150 Agilent ChemStore C/S Concepts Guide

5 Data SecurityArchiving and Dearchiving

Figure 67 Example Dearchive Process—Run10 no longer on database

Archive Status for Run10

Read Only

Read Only

Dearchived

Database no longer contains run10

Dearchive run10

(12/12)

Dearchive run10

If you attempt to dearchive run10 in the archive units dated earlier than the most recent one, the runs are transferred to the database but given the status of Read Only.

Dearchive run10

Since ID3 is the most recent archive unit, run10 will be dearchived and available for reopening on the database.

Archive Unit ID

ID2(12/7)

ID1(12/3)

ID3(12/11)

Database Tape/Disk

Page 151: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 151

Data Security 5Managing Archiving and Dearchiving on Admin Client

Managing Archiving and Dearchiving on Admin Client

Administrators may perform certain setup tasks on ChemStore C/S Admin Client before users with the correct permissions archive or dearchive runs. ChemStore C/S Admin Client is a web-based application that you can use from any PC on the network to easily manage the archive and dearchive functions and other server processes. Admin Client communicates with the Archive Server running as an NT service on the server computer. You access Admin Client using a web browser. You can perform the following tasks through Admin Client:

• schedule archive, dearchive and delete operations

• modify scheduled operations

• view reports detailing pending operations

• review archive and dearchive history

• view information about archives, runs and other objects

• execute archive, dearchive and delete operations immediately

• set server operational parameters such as wakeup interval

• connect the Archive Server to Oracle

• add and modify database connection information

See the online help for Admin Client for more information on how to use these administrative functions.

See the section “Archiving and Dearchiving” on page 135 for details on the concepts behind archiving and dearchiving with the ChemStore C/S software.

Page 152: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

152 Agilent ChemStore C/S Concepts Guide

5 Data SecurityManaging Archiving and Dearchiving on Admin Client

Overview of the Archive/Delete Process

The Admin Client helps manage the archive process taking place on the archive server. Below are listed the steps involved in archiving and deleting a run.

1 Using a function inside of ChemStore C/S itself, a user specifies which runs in the database to archive and sets a time to archive them. These runs are placed in Archive Tables that are used exclusively by the archive server. The operation and the scheduled time are placed in the Schedule Table.

2 When the server wakes up next, it checks the Schedule Table to determine if there are any pending archive requests. If there are, the system checks the integrity of the requests; for example, runs can be archived only in certain status states.

3 If a run is determined valid to archive, then the data is written into a binary file (in the directory specified by the base path variable). In addition to writing the data from a run in the archive file, all data about related objects, such as the method and raw data, are also written to the binary file. If there is a failure during this part of the process, the archive is marked as “Failed” in the Archive Tables and deleted from the Schedule Table.

4 When the server is finished writing data to the binary file, it goes back to the database to mark every archived run for Delete, if the archive includes the AutoDel flag set to TRUE.

5 The delete operation runs automatically after a successful archive. Administrators can use ChemStore C/S Admin Client to schedule a run delete operation manually at any time.

6 When the delete operation runs, all runs marked for delete during archive or manually from ChemStore C/S are deleted.

Page 153: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 153

Data Security 5Managing Archiving and Dearchiving on Admin Client

Performing Archive/Dearchive Tasks on Admin Client

The tasks listed at the beginning of this section are divided into five groups. There are archive reports, archive tasks, server control, global configuration options and test functions. Administrators must log on to the Admin Client with a user name and a password to perform archive, server control or testing tasks. Any user with access to the network can view archive reports without logging on to the Admin Client.

You access the tasks under these groups from the main web page for Admin Client.

Figure 68 Main Web Page for Admin Client

Page 154: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

154 Agilent ChemStore C/S Concepts Guide

5 Data SecurityManaging Archiving and Dearchiving on Admin Client

Once you have accessed the main web page for Admin Client, you can access the task depending on your permissions. Most of the views and reports have the following format:

Figure 69 Archive Server History Report

Page 155: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 155

Data Security 5Audit Trail

Audit Trail

The Audit Trail tracks all changes that are made to a run. Each sample has an Audit Trail which is maintained separately from the sample data. The Audit Trail is not archived with the run, so that archiving and dearchiving activities can be added to the Audit Trail; the link to the Audit Trail is maintained, even when the run is archived. Activities that give rise to an entry in the Audit Trail are:

• changes to custom field values

• changes to approval status

• changes to the sample name

• changes to reprocessing status

• adding to or deleting from a batch

• changes to archive status

The Audit Trail for any run can be displayed; it contains a table showing a summary of all the changes made to the run, who made the change and when (see Figure 70 on page 156). Buttons are available to display additional information on changes to custom field values and archive status.

Page 156: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

156 Agilent ChemStore C/S Concepts Guide

5 Data SecurityAudit Trail

Figure 70 Audit Trail

Page 157: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 157

Data Security 5E-Mail Notification

E-Mail Notification

The following facility is available on Client/Server systems only . It sends an automatic e-mail notification on specific events and needs to be set up by the administrator (or other user with administrator permissions). The events to trigger an automated e-mail notification are:

• locking of an user account due to exceeded number of unsuccessful log-on attempts.

• submitting a batch for review

• clearing a password

• creation of a new user

• permission modification

With the exception of 'notification of batch submission', the e-mail notifications should only be sent to administrators to warn about security violations. The operation of e-mail notification requires a working e-mail server. The mail can be sent in two ways:

• By anonymous SMTP to a mail server.

In this case, the mail server must be configured to allow anonymous SMTP. For increased security, the mail server can add the IP address of the ChemStore database server to its trust list, and allow anonymous SMTP only from trusted servers.

• Using LDAP

For this purpose, the SMTP virtual server from the Windows 2000 server, which has LDAP routing capability, can be used. The e-mail notification is then sent via LDAP to the mail server. This requires an existing account on the mail server that is used to send the mail; this account is then the originator of the mail.

Page 158: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

158 Agilent ChemStore C/S Concepts Guide

5 Data SecurityE-Mail Notification

The E-Mail Notification Setup dialog box, shown in Figure 71, contains settings for the e-mail server connections, and the e-mail addresses of the recipients of the e-mail notification. A default subject e.g. Account locked can be edited, and additional text can be added.

The e-mail notification message contains the date and time of the intitiation of the message, the user ID of the affected user and the ChemStore client details, as well as the additional text specified during setup.

A test mail function allows to verify the successful setup of this function.

Figure 71 The E-Mail Notification dialog box

Page 159: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 159

Data Security 5E-Mail Notification

NOTE Valid e-mail address formats are:

a) [email protected] e.g.: [email protected]

b) some text<[email protected]> e.g.: Jim Sample<[email protected]>

c) “some text”<[email protected]> e.g.: “Jim Sample”<[email protected]>

Several e-mail addresses are separated by ’;’ (semicolon).

Page 160: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

160 Agilent ChemStore C/S Concepts Guide

5 Data SecurityDatabase Logbook

Database Logbook

All interactions that affect the security of the database and the application are entered into a logbook. The logbook entries can be displayed in a table; you use a query to display only the entries that match selected criteria. Actions that cause an entry in the logbook are:

• user setup (created new user, modified permissions, changed password, cleared password, user deactivated, user reactivated, user's display name changed)

• run status changed (approved runs, rejected runs, reopened runs, deleted from batch request,

• deleted spooler jobs, edit custom fields)

• archiving and DB maintenance (archived runs, dearchived runs, archived runs with delete, modified archive, modified archive with delete, deleted runs, migrated Data, updated database schema)

• calculations (created Calculation Template, created new version of calculation template)

The logbook table is displayed with the latest entry first and can be printed.

Page 161: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 161

Data Security 5Validation and Security of Printed Files

Validation and Security of Printed Files

To support the export of results to a file-based LIMS systems, ChemStore C/S allows you to print reports to a file in xml or csv format. These formats are simple text files, which could easily be tampered with. To address this problem, ChemStore C/S contains a validation hook that you can use to ensure that the report files have not been changed.

File Validation in Operation

As soon as the report generator has completed writing to the text file, it calls a function to secure the file. ChemStore C/S includes a “secure file” function that calculates a numeric signature for the file, and stores the signature with the file. At any time, the numeric signature of the file can be recalculated and checked against the stored value; a difference in the two signatures indicated that the file has been changed.

If you want to automatically verify the identity of the exported report files with the original data, you can use a validation batch file to calculate the checksum. The batch file c:\hpchem\chemstor\validatefile.bat call hpexfs.exe and displays an error message if the security codes do not match.

NOTE When hpexfs00.exe is installed on a computer which does not have ChemStore C/S installed, the files hpcsfs00.dll and msvbwm50.dll must be installed and registered manually. Hpcsfs00.dll and hpexfs00.exe run under Microsoft Windows operating systems; they do not run under UNIX.

Page 162: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

162 Agilent ChemStore C/S Concepts Guide

5 Data SecurityValidation and Security of Printed Files

File Validation Functions

A DLL (hpcsfs00.dll) implements three functions:

SecureFile(filepath) This function takes action to secure the specified file. It calculates a “signature” of the file, and appends it to the file. The signature is the hash value — a 24-character-long value based on the data RSA Data Security, Inc. MD5 Message Digest algorithm.

ValidateFile(filepath, key) This function re-computes the signature of the file and checks it against the signature recorded by SecureFile(). If the signatures do not match, or the file has no signature, the file has been changed and ValidateFile() returns an error code. Note that the key parameter is not used in the current implementation.

UnsecureFile(filepath, key, newpath) This function strips the signature from the file and writes a new version of the file to newpath. Note that the key parameter is not used in the current implementation.

Replacing the Built-in Security Scheme with a Customized Solution

If your security needs are higher than provided by ChemStore C/S, you can simply replace the hpcsfs00.dll with your own hpcsfs00.dll.

You have to ensure that your hpcsfs00.dll does provide the three functions as outlined in this section.

Public Function SecureFile (strFilePath As String, strKey As String) As Long

Returns zero if the operation is successful, otherwise returns an error code. The report generator calls this function after it has finished writing to the text file.

NOTE You have to unregister the Agilent hpcsfs00.dll first and than register your new hpcsfs00.dll using regsrv32.

Page 163: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 163

Data Security 5Validation and Security of Printed Files

Public Function ValidateFile (strFilePath As String, strKey As String) As Long

Checks whether the file at the specified path has been secured and has not been tampered with since it was secured.

Returns zero if the file was found to be secured and the signatures matched, otherwise returns an error code.

Note that, depending on the implementation, strKey may or may not be used. It is not used in the default implementation, which validates a file by checking that it has a signature which matches the contents of the file. strKey might, however, be used in an implementation that secured the file by encryption.

Public Function UnsecureFile (strFilePath As String, strKey As String) As Long

Unsecures the file at the specified path.

Returns zero if the operation is successful, otherwise returns an error code.

Note that, depending on the implementation, strKey may or may not be used. It is not used in the default implementation, which unsecures a file by removing the signature embedded in the file between the opening and closing signature tags (leaving the tags in place in the file). strKey might, however, be used in an implementation that secured the file by encryption.

Page 164: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

164 Agilent ChemStore C/S Concepts Guide

5 Data SecurityValidation and Security of Printed Files

Page 165: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

165

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

6Data Management

Backing Up Your Database 166

Backing Up the Stand-Alone Version 168

Backing Up the Client/Server Version 171

Maintaining the ChemStation Computer 172

Disaster Recovery Plan 175

Page 166: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

166 Agilent ChemStore C/S Concepts Guide

6 Data ManagementBacking Up Your Database

Backing Up Your Database

In order to protect your valuable data, an appropriate backup strategy is mandatory. You must back up your data files at regular intervals. If your data are no longer required to be available on line, you should also archive them and remove them from your hard disk. This will free up valuable storage space on your computer’s hard disk.

The back-up process copies your database files (and any other files you choose to back up) from your computer’s hard disk to another medium, typically a recordable CD-ROM or a magnetic tape. The back-up process is recommended for any important information stored on any computer system. Back-up intervals should be as frequent as required to protect your most important data; a regime that backs up the complete database on a weekly basis is common, but if you have a high throughput of samples, you may find it necessary or desirable to do a daily back-up.

The stand-alone version of the database is an MS Access file; you can use any commonly available back-up program to copy the contents of the hpchem/database folder to another medium.

The client/server database is an Oracle database; you must shut down the database instance in order to perform a full backup.

NOTE Since in both the stand-alone and client/server versions of the product you must not be accessing the database while a backup is being performed, it is important to plan for regular down-time to perform this critical maintenance operation.

Page 167: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 167

Data Management 6Backing Up Your Database

Back Up or Archive?

In contrast to archiving (see Chapter 5, “Data Security”), back-up always works on the complete data set. In the case of a system failure, you can restore the complete database to the system after it has been repaired.

Archiving allows you to transfer individual records from the database to an Archive Unit, which is separate from the database; the transferred records may optionally be deleted from the database after archiving. Note that archiving is available only with the client/server version of ChemStore C/S.

A back-up of the database allows you to restore the complete database (including archive units) if your computer system should fail. In contrast, you cannot restore your database from archive units. For this reason, it is essential that you back up your database regularly.

CAUTION It is important to realize that the archives created with ChemStore C/S can only be restored to the database from which they were created, so backing up the full database is still critical.

Page 168: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

168 Agilent ChemStore C/S Concepts Guide

6 Data ManagementBacking Up the Stand-Alone Version

Backing Up the Stand-Alone Version

The stand-alone version of the database is an MS Access file; you can use any commonly-available back-up program to back up the contents of the database folder to another medium. Back-up and restore software is often bundled with back-up devices such as writable CD-ROM drives or magnetic tape drives.

A utility is available to repair a damaged MS Access database. The utility is a stand-alone application that is run from the ChemStore C/S group in the Programs menu. The utility also contains facilities for compacting an MS Access database. You should compact your database on a regular basis, especially if you make changes to, or delete, runs in the database. You can also use the utility to create a new, empty database; the new database can be optionally created to include the settings (queries, filters, user settings and report templates) from an existing database.

If the database becomes irreparably corrupted, delete the corrupt files from your computer’s hard disk before restoring the database files from the most recent back-up set. We also recommend that you optimize your hard disk after deleting the corrupt files and before restoring the database files using, for example, the disk defragmentation tool provided with your operating system software. This guarantees that any physical damage on the hard disk surface is identified and excluded. It also ensures that your disk is optimized for storage space and speed.

CAUTION Since backing up requires exclusive access to a file, it is important that the ChemStore C/S client and ChemStation software are not running during the backup process.

NOTE When restoring a database from CD-ROM medium, you must also remember to remove the read-only file attribute that will be carried from the CD-ROM medium as you copy the file back to your disk. This can be done by right-clicking the file after restoring it and selecting ‘properties’ from the context menu, then de-selecting the read-only checkbox.

Page 169: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 169

Data Management 6Backing Up the Stand-Alone Version

Windows 2000/XP Backup

Windows 2000 and XP both include compatible backup programs that support many standard tape devices. The backup utility provided with operating system also allows for backing up to a single file, which can then be placed on an offline medium. Both backup utilities allow the Windows Registry to be backed up as well as data.

Automating and Scheduling Windows 2000/XP Backup

Windows allows automation and scheduling of backup jobs through the use of batch files and the Scheduler service.

To configure the Scheduler service to start automatically every time press Start and Run. At the command line type: ntbackup.

Use the backup wizard to define reoccurring backup jobs and the data to be backed up. Further details can be found in the on-line help of the Backup tool.

Tape Drives

Tape drives are the most common backup devices available today. The simplest drives read and write Travan tapes (older standards are QIC-3020, QIC-3010, QIC-80, TR-3 and TR-1 cartridges) and combine lowest initial hardware cost with acceptable performance (up to 70 MB/min.). Travan tapes can store more than 2 GB of data on a single tape and are thus sufficient to back up the MS Access database. Travan tape drives can be connected to the internal IDE port or the external parallel port and do not require additional hardware other than the driver and the tapes itself. However, due to the poor performance of Travan drives they are not recommended for server backups.

CAUTION Never perform a backup on a stand-alone system while the Review Client is open. The back up file will not be suitable to restore the database. Never perform a back up during data acquisition, as this will degrade the performance of the hard disk and the acquired data might not be written to the hard disk fast enough, causing a power fail or buffer overrun message in the instrument log book and possible loss of data.

Page 170: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

170 Agilent ChemStore C/S Concepts Guide

6 Data ManagementBacking Up the Stand-Alone Version

DAT tapes usually connect to a SCSI port and offer greater storage capacity and higher data transfer rates than Travan tapes. They are also more reliable if used heavily. They require the installation of a SCSI adapter and are usually available in both internal and external versions. DAT tapes offer a low-cost medium for backing up small server systems or large standalone databases.

DLT tape drives are also typically SCSI devices and offer from 2 to 4 times the backup speed and data storage than DAT tape drives. These drives are highly recommended for backup of the Oracle database in the Client/Server product.

CD-ROM

CD-R (CD-Recordable) systems allow you to record up to 700 MB of data on a single CD-R disk. The CD reader/recorder is usually connected to your PC’s IDE interface. The CD-R media have a very long life, making them ideally suited to archiving. The CD-R disks can be read by other computers fitted with CD-ROM drives and appropriate software.

If you are planning to use CD-Recordable to back up the standalone MS-Access databases, you should plan for each database to be kept under 700 MB in size.

The G2181A folder of the installation CD contains a setup of a NT service that can be installed for monitoring the database size automatically.

CAUTION Never perform a backup with a Windows-based backup utility while the ChemStation is acquiring data or the Review Client is open. This may result in inconsistencies in the file structure stored on the backup medium.

Page 171: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 171

Data Management 6Backing Up the Client/Server Version

Backing Up the Client/Server Version

Back-up and restoring the ChemStore C/S server database is done with Oracle tools. Please refer to the relevant Oracle documentation.

Oracle Back-up Types

There are generally two types of back-up performed on Oracle databases; Cold Back-up and Hot Back-up. Cold Back-ups are performed while the database instance is not running, which Hot Back-ups can be done while the database is in use. An example of a cold backup and restore process can be found in the ChemStore Installation Guide in the section “Administration Tasks and Reference Information”.

Cold Back-up

A cold backup is a simple file backup of all of the necessary files that make up the Oracle database. In order to perform a cold backup, the database instance must first be stopped and all user must be disconnected from the database so that the backup can lock the files for exclusive use. Failure to stop the database instance before backing up results in files missing from the backup set and/or an inconsistent database. Either of these conditions will result in a backup set that cannot be restored to produce a working database.

Hot Back-up

A hot back-up is performed while the database is in archive log mode, and allows the back-up to be performed without first stopping the database instance. However, since the ChemStore C/S application uses large binary data records, running the database in archive log mode is not recommend; as performance of the database is seriously degraded. Agilent generally recommend that only cold backups be performed unless the system has been optimized for continuous operation in archive log mode.

NOTE We recommend that you perform this task under the supervision of an Oracle-trained professional.

Page 172: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

172 Agilent ChemStore C/S Concepts Guide

6 Data ManagementMaintaining the ChemStation Computer

Maintaining the ChemStation Computer

On every system, regular maintenance is mandatory to assure correct operation.

This section describes the maintenance steps that you should perform on a regular basis. These steps include cleaning out left-over temporary files, checking the integrity of the logical and physical structure of the file system, virus-checking and regular back-ups.

Cleaning up Left-Over Temporary Files

Occasionally, temporary files may accumulate in the directory specified by the TEMP environment variable. These files generally are left over when Windows is abnormally terminated, for example, by powering your computer down without closing Windows first. The Windows temporary files are named ~XXXXXXX.TMP, where XXXXXXX are characters and numbers produced by the program that created the temporary file. To recover the temporary space, you should delete these files after closing all currently running applications.

To determine which directory is currently being used for temporary files, type SET at the command prompt. This will echo back the current settings for all user and system environmental variables.

Maintaining the PC’s File System

Scanning and Repairing NTFS Volumes

Use the Windows utility CHKDSK from the command prompt to scan and repair NTFS volumes. This program incorporates all of the functionality of the MS DOS-based Chkdsk and Scandisk utilities, including a surface scan. To do a surface scan, use CHKDSK /R. The disk check can also be performed by accessing a disk drive’s properties window and selecting the tools tab.

Page 173: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 173

Data Management 6Maintaining the ChemStation Computer

Windows 2000/XP perform an autocheck routine upon every boot. If this routine detects a consistency error on the volume, it automatically runs the CHKDSK /F command to fix it. If Windows is unable to run the CHKDSK /F command (for example, because you want to run it on the boot partition or because someone is accessing the partition from the network), this task is postponed until the next time the system reboots.

Using compression on NTFS partitions

Due to the performance decrease when using the Windows compression feature, Agilent does recommend this for ChemStation plus systems. The ChemStation data are compressed quite heavily in their native format, so disk compression does not usually offer much advantage in space.

Using encryption with Windows 2000 and XP

Due to the performance decrease and risk of losing access to data when using the Windows encryption feature, Agilent does recommend using encryption for ChemStation plus systems. The ChemStation data are typically quite large and are stored in checksum-protected binary files, so encryption is not necessary and reduces the system performance.

Continuous operation

If your environment requires continuous routine operation of the ChemStation plus system over days or even weeks without restarting your client system, the system performance may degrade over time due to memory and resource leaks in the system. To overcome this problem, we recommend installing current Service Packs as they become available and are supported by Agilent. Operating system Service Packs are available from the Microsoft home page.

In addition, we recommend that you defragment your hard disk at regular intervals, at least once a week.

Page 174: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

174 Agilent ChemStore C/S Concepts Guide

6 Data ManagementMaintaining the ChemStation Computer

Defragmenting an NTFS Volume

Compared to the FAT file system, fragmentation of files on an NTFS volume is greatly reduced. By design, fragmentation only occurs on an NTFS volume if a file increases in size after it is on the drive. This means that backing up a volume to tape and restoring it from tape will result in an unfragmented volume.

Windows 2000 Professional and XP does include a simple utility for this task. Third-party programs are available that can defragment both Windows 2000 and XP volumes. Agilent recommend Diskeeper from Executive Software (http://www.execsoft.com/).

For more information on system maintenance and administration, refer to the ChemStore Installation Guide in the section “Administrative Tasks and Reference Information”.

CAUTION Do not defragement your hard disk drive during data acquisition.

Page 175: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 175

Data Management 6Disaster Recovery Plan

Disaster Recovery Plan

You should also formulate a disaster recovery plan that includes copies of your entire database to protect you against a complete system failure such as a hard disk problem. The purpose of a disaster recovery plan is to plan for failure scenarios and then identify and implement procedures to recover in the even of such scenarios. The following short list of common failures is presented.

Disk drive Failure

If a disk drive fails, you will need to repair your system and then restore it to full functionality from a backup.

Agilent recommend that server systems use hardware RAID controllers with redundant disk configurations so that a single disk drive failure will not result in down-time or data loss.

Power Failure

A power failure on a database system may result in corruption of data unless the system can be shut down cleanly. This is more important for Oracle server systems, where uninterruptable power supplies are required.

Database is corrupted

The disaster recovery plan should include how often and how a regular backup should be done in order to be able to recover your data in case the database becomes corrupted. Then it can be done from the last backup.

Page 176: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

176 Agilent ChemStore C/S Concepts Guide

6 Data ManagementDisaster Recovery Plan

1 Delete all files and directories which are affected and which should be recovered from the last backup.

2 Optimize the disk using a qualified defragmenting utility.

3 Recover the data by retrieving your data from the backup.

Page 177: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

177

Agilent ChemStore C/SConcepts Guide

Agilent Technologies

7ChemStore C/S Computations

Statistical Calculations 178

Calculations using custom fields 198

Time computation 201

Page 178: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

178 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsStatistical Calculations

Statistical Calculations

Single Value Calculations

The following calculations do not require value pairs.

In all cases, the calculations are based on the current dataset that is displayed in the ChemStore C/S window or report (excluding runs marked as excluded).

Number Displays the number of values:

Sum Displays the sum of the values:

Minimum Displays the smallest value.

Maximum Displays the largest value.

Mean Displays the arithmetic mean (sum/number) of the all the values.

Variance Displays the square of the standard deviation.

Number n( )=

Sum Xi∑( )=

Mean x⟨ ⟩Xi∑( )

n-----------------= =

Variance s2x x⟨ ⟩–( )∑

2

n 1–( )-------------------------------------= =

Page 179: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 179

ChemStore C/S Computations 7Statistical Calculations

Regression Calculations

The calculations supported for value pairs are based on weighted linear regression. They require the specification of additional parameters.

The parameters need not be specified until a statistical calculation is requested. The parameters are:

Standard Deviation

Displays the standard deviation of the values.

Relative Standard Deviation

Displays the relative standard deviation, standard deviation divided by the mean. When displayed as a percent value, the RSD gets multiplied by 100.

Std. Dev. s Variance( )= =

RSD%

------------ sx⟨ ⟩

-------- 100=

Curve Type Determines the type of function to which the data is fitted. The following types are supported:Linear Types: Zero Order (mean), First order (linear), Second order (quadratic), Third order (cubic) Non-linear Types: Exponential, Logarithmic, Power

Origin Treatment Determines how the point at (0,0) is used. The following types are supported:Ignore: use the data set “as is”; no special treatment Include: add a point at (0,0) to the actual data set. Force: constrain the curve so that it must pass through (0,0).

Weighting Determines how values are weighted in the calculation, from None, 1 x⁄ 1 x2⁄ 1 y⁄ 1 y2⁄, , ,

Page 180: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

180 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsStatistical Calculations

The Linear Regression Model

The regression model used for zero, first, second and third order curve types is one of the following, depending on whether or not the curve is constrained to pass through the origin:

(I) unconstrained origin

(II) forced origin

where are the independent x-values

are the dependent (measured) y-values

n is the number of data points m is the order of regression

are the calculated regression coefficients

are the y-values calculated from the function using

Matrix Solution of the Regression

The same regression models can be expressed in matrix notation as

where

Column vector of calculated values

Column vector of coefficients

and F is one of the following rectangular matrices (which one depends on whether or not the curve is constrained to pass through the origin):

case (I), an matrix of independent values

y'i a0 a1x1 … amxim

+ + +=

y'i a1x1 … amxim

+ +=

xi

yi

ao…am

y'i ao…am

NOTE If , then case (I) must be used, and the resulting is just the mean of the values.

m 0= y'i yi

y' Fa=

y' y1'…yn'( )T=

a a1…am( )T=

F1 x1 x1

m

1 xn xnm

= n x m 1+( )

Page 181: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 181

ChemStore C/S Computations 7Statistical Calculations

case (II), an matrix of independent values

The values of the vector of coefficients, a, can be found via

wherey is the vector of dependent variables

is the inverse of F calculated by singular value

decomposition ( ), a robust method which helps minimize round-off errors and problems with nearly-singular equations.

Statistical and Related Values

Once the coefficients have been found, then the following statistics and related values can be calculated:

Residuals

One value for each measured :

Sample Standard Deviation

One value. This is also called the “residual standard deviation”:

where Number of degrees of freedom, case (I)

Number of degrees of freedom, case (II)

Standard Deviation of the Coefficients

One value for each of the (m+1) coefficients. Standard deviation of coefficient :

Fx1 x1

m

xn xnm

= n x m

a F#y=

F# VΛ 1– UT=

F UΛVT=

yi

ei y'i yi–=

s y'i yi–( )2∑( ) q⁄=

q n m– 1–=

q n m–=

ai

Page 182: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

182 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsStatistical Calculations

where are taken from the singular value decomposition.

Correlation Coefficient (expressed as coefficient of determination)

One value.

where for case (I)

for case (II)

Modifications for Weighted Linear Regression

In case of weighting, the weights are stored as diagonal elements of a matrix W:

Diagonal matrix of weights

weights (one of )

In case xi or yi is equal to zero, wi is calculated as the average of the remaining non-zero values.

Set

the solution to the regression equation becomes

where

denotes the inverse of calculated by the singular value decomposition.

The calculations for standard deviation of coefficients and correlation coefficient are modified as follows:

sais VΛ 2– VT( )ii=

V Λ,

R2 y'i yavg–( )2

yi yavg–( )2-----------------------------∑=

yavg Σyi n⁄=

yavg 0=

W wii( )=

wi 1 xi⁄( ) 1 xi2⁄( ) 1 yi⁄( ) 1 yi

2⁄( ),,,

Φ W12---–

diag w1( )… wn( )( )= =

a ΦF( )#Φy=

ΦF( )# VΛ 1– UT= ΦF

Page 183: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 183

ChemStore C/S Computations 7Statistical Calculations

Relative Fit Error

Standard deviation of coefficients

Correlation coefficient (expressed as coefficient of determination)

where case (I)

case (II)

Modifications for Non-Linear Functions

The same linear regression matrix equation is used to fit data to non-linear functions if those functions can be “linearized”.

Exponential

The “real” equation is

This can be expressed as a “fitting function” that is linear in the coefficients:

In practice, modified y-values are calculated as

and then the following equation is used in the regression:

sw wii y'i yi–( )2( ) q⁄∑=

saisw VΛ 2– VT( )ii=

R2 wii y'i yavg–( )2( )∑wii yi yavg–( )2( )∑

-------------------------------------------------=

yavg wiiyi( )∑ wii( )∑⁄=

yavg 0=

yi a bxi( )exp⋅=

yi( )ln a( )ln bxi+=

Yi yi( )ln=

Yi a0 a1xi+=

Page 184: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

184 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsStatistical Calculations

Power

the real equation

function linear in coefficients

modifications of x- and y-values

function used for regression

Logarithmic

the real equation (already linear in coefficients)

modification of x-values

function used for regression

yi axib

=

yi( )ln a( )ln b xi( )ln⋅+=

Yi yi( ) Xi xi( )ln=,ln=

Yi a0 a1Xi+=

yi a b xi( )ln⋅+=

Xi xi( )ln=

yi a0 a1Xi+=

Page 185: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 185

ChemStore C/S Computations 7Custom Calculator Commands

Custom Calculator Commands

SELECT

Syntax:

select <column list> for <compound list> where <condition> into <table>

Description:

This statement is used for retrieving data from selected rows of one ore more DB tables into a custom calculation table. After the reserved word select a list of columns <column list> of DB tables has to be specified. These columns can be either simple columns or sub columns.

Sub columns are those columns that are available for each compound specified by a for clause in the <compound list>. Sub columns must be preceded by an exclamation sign '!'. If at least one sub column is specified in the <column list> then the for clause is mandatory. The compounds of the for clause can be specified also by using a wildcard '*'. The wildcard alone means all compounds. Appending a wildcard to a compound prefix specifies all compounds that start with a given prefix. The following for clause specifies all compounds starting with "Bar" and those starting with "C":

For Bar*,C*

The optional where clause defines a subset of retrieved table rows using a <condition> expression. The resulting table <table> is specified by the into clause..

FROM

Syntax:

from <table> select <column list> for <compound list> where <condition> into <table>

Description:

Page 186: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

186 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCustom Calculator Commands

This statement is used for retrieving data from one custom calculation table (source table) into another (destination table). The source table is specified after the reserved word from. The rest of the statement is the same as for the select statement above except for the following: The <column list> and <compound

IF

Syntax:

if <condition> then <first assignment> else <second assignment>

Description:

This statement is used for conditional assignment. The else part of the statement is optional. If the condition evaluates to a single value and the <condition> expression is fulfilled then the <first assignment> is made else the <second assignment> is made. If the condition evaluates to a column but the assignment is not columnar, then the condition is further evaluated as a logical and of all column rows and the assignment is made. If both the condition and the assignment are columnar then a conditional assignment is made row by row depending on the value of the condition column row. If the length of the condition column (m) is greater then the length of the assignment column (n), then only the first n rows are conditionally assigned. The exceeding part of the conditional column is ignored. If the length of the conditional column (m) is smaller than the length of the assignment column (n) then only the first m rows are conditionally assigned. The rest of assignments is not made at all.

FORMAT

Syntax:

format <operand list> using <format specification>

Description:

This statement is used for formatting of column and variable data. The comma separated list of columns and variables is specified as <operand list> and the format specification is given by the <format specification> string. The formatting affects only the output (screen, paper) representation of data while the actual data remain unchanged.

Page 187: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 187

ChemStore C/S Computations 7Custom Calculator Commands

Format Specification

Format specification is a string containing instructions for operand formatting. These instructions are given by characters of the format specification that have special meaning described in the following subsections. Characters that do not have any special meaning are displayed literally. In order to display literally a character that has a special meaning, precede it with a backslash (\) or enclose it in double quotation marks (" "). The backslash itself and double quotes are not displayed. To display a backslash, use two backslashes (\\).

Formatting numeric operands

A format specification for numeric operands can have from one to four sections separated by semicolons.

The following format specification has two sections: the first specifies the format for positive values and zeros; the second section specifies the format for negative values.

"$#,##0;($#,##0)"

If you use The result is

One section only The format specification applies to all values.

Two sections The first section applies to positive values and zeros, the second to negative values.

Three sections The first section applies to positive values, the second to negative values, and the third to zeros.

Four sections The first section applies to positive values, the second to negative values, the third to zeros, and the fourth to null values.

Page 188: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

188 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCustom Calculator Commands

If you include semicolons with nothing between them, the missing section is printed using the format of the positive value. For example, the following format specification displays positive and negative values using the format in the first section and displays "Zero" if the value is zero.

"$#,##0;;\Z\e\r\o"

The following characters can be used for numeric format specifications:

Character Description

None Display the number with no formatting.

0 Digit placeholder. Display a digit or a zero. If the operand has a digit in the position where the 0 appears in the format specification, display it; otherwise, display a zero in that position. If the operand has fewer digits than there are zeros (on either side of the decimal) in the format specification, display leading or trailing zeros. If the operand has more digits to the right of the decimal separator than there are zeros to the right of the decimal separator in the format specification, round the number to as many decimal places as there are zeros. If the operand has more digits to the left of the decimal separator than there are zeros to the left of the decimal separator in the format specification, display the extra digits without modification.

# Digit placeholder. Display a digit or nothing. If the operand has a digit in the position where the # appears in the format specification, display it; otherwise, display nothing in that position. This symbol works like the 0 digit placeholder, except that leading and trailing zeros aren't displayed if the operand has the same or fewer digits than there are # characters on either side of the decimal separator in the format specification.

. Decimal placeholder. The decimal placeholder determines how many digits are displayed to the left and right of the decimal separator. If the format specification contains only number signs to the left of this symbol, numbers smaller than 1 begin with a decimal separator. To display a leading zero displayed with fractional numbers, use 0 as the first digit placeholder to the left of the decimal separator.

% Percentage placeholder. The operand is multiplied by 100. The percent character (%) is inserted in the position where it appears in the format string.

Page 189: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 189

ChemStore C/S Computations 7Custom Calculator Commands

Formatting date/time operands

The following characters can be used for date/time format specifications:

, Thousand separator. The thousand separator separates thousands from hundreds within a number that has four or more places to the left of the decimal separator. Standard use of the thousand separator is specified if the format specification contains a thousand separator surrounded by digit placeholders (0 or #). Two adjacent thousand separators or a thousand separator immediately to the left of the decimal separator (whether or not a decimal is specified) means "scale the operand by dividing it by 1000, rounding as needed." For example, you can use the format specification "##0,," to represent 100 million as 100. Operands smaller than 1 million are displayed as 0. Two adjacent thousand separators in any position other than immediately to the left of the decimal separator are treated simply as specifying the use of a thousand separator.

E- E+ e- e+ Scientific format. If the format specification contains at least one digit placeholder (0 or #) to the right of E-, E+, e-, or e+, the operand is displayed in scientific format and E or e is inserted between the number and its exponent. The number of digit placeholders to the right determines the number of digits in the exponent. Use E- or e- to place a minus sign next to negative exponents. Use E+ or e+ to place a minus sign next to negative exponents and a plus sign next to positive exponents.

Character Description

: Time separator. The time separator separates hours, minutes, and seconds when time values are formatted.

/ Date separator. The date separator separates the day, month, and year when date values are formatted.

C Display the date as ddddd and display the time as ttttt, in that order. Display only date information if there is no fractional part to the date serial number; display only time information if there is no integer portion.

D Display the day as a number without a leading zero (1 31).

Dd Display the day as a number with a leading zero (01 31).

Ddd Display the day as an abbreviation (Sun Sat).

Dddd Display the day as a full name (Sunday Saturday).

Page 190: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

190 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCustom Calculator Commands

Ddddd Display the date as a complete date (including day, month, and year). The default short date format is m/d/yy.

Dddddd Display a date serial number as a complete date (including day, month, and year). The default long date format is mmmm dd, yyyy.

W Display the day of the week as a number (1 for Sunday through 7 for Saturday).

Ww Display the week of the year as a number (1 54).

M Display the month as a number without a leading zero (1 12). If m immediately follows h or hh, the minute rather than the month is displayed.

Mm Display the month as a number with a leading zero (01 12). If m immediately follows h or hh, the minute rather than the month is displayed.

Mmm Display the month as an abbreviation (Jan Dec).

Mmmm Display the month as a full month name (January December).

Q Display the quarter of the year as a number (1 4).

Y Display the day of the year as a number (1 366).

Yy Display the year as a 2-digit number (00 99).

Yyyy Display the year as a 4-digit number (100 9999).

H Display the hour as a number without leading zeros (0 23).

Hh Display the hour as a number with leading zeros (00 23).

N Display the minute as a number without leading zeros (0 59).

Nn Display the minute as a number with leading zeros (00 59).

S Display the second as a number without leading zeros (0 59).

Ss Display the second as a number with leading zeros (00 59).

t t t t t Display a time as a complete time (including hour, minute, and second). A leading zero is displayed if the leading zero option is selected and the time is before 10:00 A.M. or P.M. The default time format is h:mm:ss.

AM/PM Use the 12-hour clock and display an uppercase AM with any hour before noon; display an uppercase PM with any hour between noon and 11:59 P.M.

Page 191: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 191

ChemStore C/S Computations 7Custom Calculator Commands

Formatting string operands

A format specification for strings can have one section or two sections separated by a semicolon (;).

The following characters can be used for string format specification:

am/pm Use the 12-hour clock and display a lowercase AM with any hour before noon; display a lowercase PM with any hour between noon and 11:59 P.M.

A/P Use the 12-hour clock and display an uppercase A with any hour before noon; display an uppercase P with any hour between noon and 11:59 P.M.

a/p Use the 12-hour clock and display a lowercase A with any hour before noon; display a lowercase P with any hour between noon and 11:59 P.M.

AMPM Use the 12-hour clock and display the AM string literal with any hour before noon; display the PM string literal with any hour between noon and 11:59 P.M.

If you use The result is

One section only

The format applies to all string data.

Two sections The first section applies to string data, the second to null values and zero-length strings ("").

Character Description

@ Character placeholder. Display a character or a space. If the string has a character in the position where the at symbol (@) appears in the format string, display it; otherwise, display a space in that position. Placeholders are filled from right to left unless there is an exclamation point character (!) in the format string.

& Character placeholder. Display a character or nothing. If the string has a character in the position where the ampersand (&) appears, display it; otherwise, display nothing. Placeholders are filled from right to left unless there is an exclamation point character (!) in the format string.

< Force lowercase. Display all characters in lowercase format.

Page 192: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

192 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCustom Calculator Commands

TRANSPOSE

Syntax:

Transpose <source table> by <column> into <destination table>

Description:

This statement is used for transposing of a custom calculation table. The table to be transposed is defined by <source table> and the result is put into <destination table>. The by <column> specifies which column of the source table will be used as column names (header) of the destination table. Since the column names must be unique the destination table does not contain any column with a name that appears more than once in the by <column>. The resulting table <destination table> is specified by the into clause

GROUP

Syntax:

Group <source table> by <column> do <operation list> into <destination table>

Description:

This statement is used for performing a list of operations on groups of rows. The source table for this statement is given by <source table> and the result is written into the <destination table>. Grouping of the <source table> rows is defined by the values of the by column. Each group of rows contains rows with equal values in the by column. The operations to be performed on each group of rows are defined in a comma separated <operation list>. Each element of this list has the following syntax:

<aggregate> ( <column> ) as <alias>

> Force uppercase. Display all characters in uppercase format.

! Force left to right fill of placeholders. The default is to fill placeholders from right to left.

Page 193: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 193

ChemStore C/S Computations 7Custom Calculator Commands

The aggregate function <aggregate> is applied to each group of rows of the column <column>. The results of these operations form a column <alias> of the destination table. The resulting table <destination table> is specified by the into clause.

Page 194: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

194 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCustom Calculator Commands

Error codes and descriptions

This section describes all errors detected by the custom calculator. These errors are either syntax or runtime errors. Syntax errors are detected during syntax check of the calculation statement. If a statement contains a syntax error then it cannot be understood by the custom calculator and thus cannot be evaluated. Custom calculator indicates the position of a syntax error inside the calculation statement either by red color of the text and by error position marker (^). The error position marker points to the unexpected syntax element

Only syntactically correct calculation statements are evaluated. Even if a statement is syntactically correct it can still produce runtime errors due to unavailability of data or due to unexpected data type or value. Custom calculator does not track the error position for runtime errors.

Syntax errors

These errors are detected during syntax check. All syntax errors immediately terminate syntax check and an adequate error message is always displayed. If a calculation statement contains more than one syntax error only the first one is detected and displayed.

Code Description Cause

2 'by' expected By is missing (transpose and group statements).

4 'do' expected Do is missing (group statement).

5 End of line expected Statement ending with one or more syntactical elements that are not needed and thus unexpected.

6 '=' expected Equal sign is missing (first word of a statement misspelled and treated as a variable name by custom calculator).

7 Expression expected Expression is missing (right part of an expression containing relational operator is missing).

8 Factor expected A factor is missing (in an expression after a *, /, or and operator).

9 For expected For is missing in a statement that contains at least one sub column. (select and from statements).

11 'into' expected Into is missing (select, from, transpose, and group statements).

Page 195: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 195

ChemStore C/S Computations 7Custom Calculator Commands

12 '(' expected Left parenthesis is missing (in a function invocation).

14 Name expected Incorrect name (name does not start with a letter)

15 Number expected Incorrect number (digits followed by a decimal point followed by letters)

17 ')' expected Right parenthesis is missing (the numbers of left and right parentheses are not balanced; in a function invocation).

18 'select' expected Select is missing (from statement).

19 string expected String is missing (format specification in a format statement)

20 Term expected A term is missing (in an expression after a +, -, or operator).

21 'then' expected Then is missing (if statement).

22 'using' expected Using is missing (format statement).

24 Invalid column type A column that is not a sub column is specified as sub column or a column that is a sub column is not specified as a sub column (select and from statements).

25 DB column expected Specified column is not a column of the specified DB table (select statement).

26 DB table expected Specified table does not exist in the DB (select statement).

27 Compound expected Specified compound does not exist in the DB (select and from statements).

28 Variable not found Specified variable does not exist (format statement).

29 Table not found Specified table does not exist (format statement).

30 Column not found Specified column does not exist in the specified table (format statement).

32 Table already exists Specifying an existing table in an INTO clause (select, from, transpose, and group statements).

41 Duplicate column name A column is specified more than once (select and from statements).

42 Column is read only Assignment of an expression to a table column directly or indirectly obtained from the DB (column assignment statements).

Page 196: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

196 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCustom Calculator Commands

Calculation errors

There might be cases where you want to include error messages of your calculation in the final report. At this stage of report generation this is the only way to see the error messages of a calculation because calculation is done right at the moment when the report is generated and the result of the calculation is embedded into the report.:

Runtime errors

These errors are detected during runtime. If a runtime error occurs more than once in a single calculation line then it is reported only once.

Bad argument

This runtime error occurs if an arithmetical function is called with an illegal argument value. The following operations produce this error:

• Arithmetical functions (log, ln) of zero or negative argument. Argument may be a constant, a variable or a column.

• Arithmetical function (sqrt) of a negative argument. Argument may be a constant, a variable or a column.

Overflow

This runtime error occurs if a numerical result of an operation is too large or too small. The following operations produce this error:

• Addition (+), subtraction (-), multiplication (*) of a large numerical operands. Both operands may be a constant, a variable or a column.

• Division (/) of a zero or close to zero operands. Both operands may be a constant, a variable or a column.

Code Description Cause

37 Bad argument A function was called with an illegal argument value.

38 Overflow Result of operation is too large.

39 Division by zero Division by zero was attempted.

40 Type mismatch The type of the operand(s) conflicts with attempted operation.

Page 197: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 197

ChemStore C/S Computations 7Custom Calculator Commands

• Arithmetical functions (exp, sqr) and aggregate functions (mean, rstdev, stdev, sum, var) applied to large operands. Argument may be a constant, a variable or a column.

• Negation (not) of numerical values larger than 2147483646.

Division by zero

This runtime error occurs if a nonzero value is divided by zero. The following operations produce this error:

• Division (/) of a nonzero operand by a zero operand. Both operands may be a constant, a variable or a column.

Type mismatch

This runtime error occurs if a function argument or operator’s operand(s) are not of the appropriate type. The following operations produce this error:

• Addition (+), subtraction (-), multiplication (*), division (/), conjunction (and), and disjunction (or) of a string and a numeric operand. Both operands may be a constant, a variable or a column. It is not important which of the two arguments is of a string, and which is of a numeric type.

• Negation (not) and changing a sign (-) of a string operand. Operands may be a constant, a variable or a column.

• Subtraction (-), multiplication (*), division (/), conjunction (and), and disjunction (or) of two string operands. Both operands may be a constant, a variable or a column.

• Arithmetical functions (abs, exp, ln, log, sqr, sqrt) and aggregate functions (mean, rstdev, stdev, sum, var) applied to string arguments. Argument may be a constant, a variable or a column.

Page 198: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

198 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCalculations using custom fields

Calculations using custom fields

As described in the section “Using Custom Fields” on page 21 you can use custom fields to transfer sample related data to the ChemStore C/S database that is not included in the standard database model. This data can be used to calculate compound related data using the custom calculation as described in the section “Understanding the Custom Calculator Script Wizards” on page 60.

In this section we describe how to calculate commonly required data using built in macro functions.

Original response factor

The original response factor is the response factor calculated by dividing the amount from the calibration table with the peak area for a specific compound. The response factor field in the database only stores the calculated response factor after averaging as it appears in the ChemStation calibration table. Use the function ChemStoreCompoundVal to obtain the original response factor for each compound in the calibration runs of the specific compound.

1 Create custom fields for each compound you need the original response factor for. Use a naming convention that includes the compound name for easier identification

2 Add the custom fields to a study and select the field details button.

3 In the data entry section select by ChemStation function and type ChemStoreCompoundVal ("AmtPerRespOrg", "CompoundName") as the function. Replace CompoundName with the appropriate name of the compound.

Page 199: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 199

ChemStore C/S Computations 7Calculations using custom fields

List of built in functions

ChemStoreAreaSum(signal$)

The function ChemStoreAreaSum calculates the sum of all peak areas in a chromatogram for a specified signal or for all signals. If a signal is specified, only peaks in that signal are summed. If the parameter is omitted all peaks in the chromatogram are summed.

Parameters:

[signal$]

You can either specify a signal numerically, by the full signal description or the first n characters of the signal description. The parameter has to be in double quotes “, as it must be a string.

Example ChemStoreAreaSum (“2”)

Sums all peaks of second signal in ChromRes[1].Signal

ChemStoreAreaSum (1)

error - use ChemStoreAreaSum ("1") instead

ChemStoreAreaSum ("")

Sums all peaks of all loaded signals

ChemStoreAreaSum ("DAD1 B, Sig=305,190 Ref=550,100")

Sums all peaks of signal "DAD1 B, Sig=305,190 Ref=550,100"

ChemStoreAreaSum ("DAD1 B")

Sums all peaks of DAD1 channel B

Page 200: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

200 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsCalculations using custom fields

ChemStoreCompoundVal(expression$,compound$)

The function ChemStoreCompoundVal returns a per compound value denoted by expression$ for the compound compound$.

Parameters:

[expression$]: Specifies the information to be returned

[compound$]: The compound name exactly as it appears in the ChemStation calibration table. The compound name is not case sensitive.

Example ChemStoreCompoundVal ("FirstPeak~MeasRetTime", "Biphenyl")

Returns the retention time of the peak identified as the component "Biphenyl".

ChemStoreCompoundVal ("AmtPerRespOrg", "Biphenyl")

Returns the original response factor of the component "Biphenyl"

ChemStoreCompoundText(expression$,compound$)

The function ChemStoreCompoundText returns a per compound text denoted by expression$ for the compound compound$.

Parameters:

[expression$]: Specifies the information to be returned

[compound$]: The compound name exactly as it appears in the ChemStation calibration table. The compound name is not case sensitive.

Example ChemStoreCompoundText$ ("FirstPeak~IntPeakType", "Biphenyl")

Returns the peak type as of the component "Biphenyl", like "BV". See the ChemStation online help for a description of baseline codes.

Page 201: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 201

ChemStore C/S Computations 7Time computation

Time computation

Time synchronization

As a ChemStation plus system consist of several components, the individual clocks need to be synchronized to show a consistent time throughout the system.

The client PC's and the servers clock should be synchronized using standard network tools that synchronizes times in a network.

The analytical instruments that feature a real time clock that is used for generating time stamps, like the Agilent 1100 LC system, are synchronized during start-up of the ChemStation.

The individual Agilent 1100 series LC modules synchronize within the system with each injection to a new system time.

Time stamps

Time stamps in a ChemStation plus system are generated each time an event happens. Events can be a sample injection, a method saved to disk, a data transfer to the database is finished, a run is approved, etc. There are generally two possibilities to store time stamps:

• Time stamps using the time zone settings

• Time stamps ignoring the time zone settings

All time stamps in the ChemStore C/S database (Oracle or Access) are stored independent of the current time zone setting of the server or the review client PC. Because of this, changing the time zone setting on either computer does not change any time stamp reported by ChemStore C/S.

The time stamps in ChemStation log files are also independent of the time zone settings of the computer. These are for example method and sequence start and stop times, error messages and warnings.

Page 202: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

202 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsTime computation

Time stamps associated with ChemStation data files and methods however do respect the time zone settings on the PC they are created on. These are for example injection times, calibration times, method save times etc.

This dependency on the time zone settings is only notable, if the data files are send across time zones, or if the time zone on the reporting PC was changed after data acquisition.

Time zones

To be able to compare events that happen in different time zones, the ChemStation stores the time zone dependent time stamps always in Coordinated Universal Time (UTC). UTC is basically the same as Greenwich Main time (GMT) and the local time in all other time zones is calculated from this.

The storage of the time zone dependent time stamps like the injection time is performed by the ChemStation in the following way:

1 The event occurs.

2 The local time is read from the system.

3 The system variable TZ is read, indicating the time zone the computer is located.

4 UTC is calculated using TZ and the local time.

5 UTC is stored as seconds since the 01.01.1970 00:00:00.

The reporting of the time zone dependent time stamps like the injection time is performed by the ChemStation in the following way:

1 The system variable TZ is read, indicating the time zone the computer is located.

2 UTC is read from the files as seconds since the 01.01.1970 00:00:00.

3 The system variable TZ is used to calculate the local time for this PC when the event happened.

4 The local event time in seconds since the 01.01.1970 00:00:00 is converted according to the selected formatting.

As the storage and the reporting of these times can happen on computers in different time zones, the time zone dependent time stamps vary with the setting of the system variable TZ.

Page 203: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 203

ChemStore C/S Computations 7Time computation

Setting Time zones on the PC

In Windows 2000 and Windows XP the time zone your PC is located in can be set by using the Time Zone tab in the Control Panel applet Date/Time. Unfortunately the ChemStation can not read this setting and requires the system variable TZ to be set to the correct value.

The ChemStore C/S client installation program sets TZ system variable in the windows registry. If your PC Time zone setting is changed after the installation, you must manually update the TZ system variable from the properties panel under ‘My Computer’.

Refer to the ChemStore Installation Manual in the “Reference Information” section for instructions on modifying the TZ variable.

NOTE Please note that when your time zone in the Control Panel is set to a positive value, like (GMT +01:00) Brussels, Berlin, Bern, Rome, Stockholm, Vienna the TZ variable needs to be negative, like WES-01WED01.

Page 204: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

204 Agilent ChemStore C/S Concepts Guide

7 ChemStore C/S ComputationsTime computation

Page 205: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 205

Index

Aadministrator, 128analytes, 44approval status, 27, 39, 44approving runs, 14, 39, 44, 50archiving, 15, 135, 167

automated, 135interactive, 139

audit trail, 14, 30, 124automated archiving, 135axes

chart, 40, 44

Bbacking up, 166backup

strategy, 166batch, 39, 44, 50batch review, 29

Ccalculated field, 100calculations, 96

single value, 178center line, 47cGMP, 124chart, 96, 102

control, 47chart axes, 40, 44, 95chemist, 127ChemStation, 12, 23, 50ChemStoreAreaSum, 199ChemStoreCompoundText, 200ChemStoreCompoundVal, 200CHKDSK, 172

chromatogram, 20chromatograms, 96client/server version, 12coefficient

correlation, 182, 183standard deviation, 181, 183

column, 95complement filter, 28compound list, 44compound review, 35configuring tables, 44control chart, 14, 22, 47correlation coefficient, 182, 183corrupted database, 168, 175count, 46criteria

search, 24critical limits, 47cross-instrument report, 22cross-sample report, 22current Good Manufacturing Practice, 124curve type, 179custom field, 32custom field status, 22custom fields, 20, 21, 24customized report, 15

Ddata

exporting, 15external, 21organizing, 13retrieval, 13reviewing, 14transfer, 23

data entry, 132

data flow, 19data integrity, 124Data section, 97data section, 91, 96data security, 124data type, 132data types, 22database

corrupted, 168dataset, 90date display style, 45date/time, 22dearchiving, 15, 135dearchiving, runs, 143deleting runs, 53disaster recovery plan, 175

Eelement, 92, 96e-mail notification, 157excluding runs, 40exponential, 183exporting data, 15external data, 21

Ffile validation, 161files

corrupted, 168reloading, 50

filter, 27, 134complement, 28

footer, 96format

number, 45

Page 206: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

206 Agilent ChemStore C/S Concepts Guide

Index

GGLP, 124GMT, 202Good Laboratory Practice, 124grouping, 106

Hhash value, 162header, 96

Iinformation retrieval, 24installation, 12integer, 22integrity

data, 124interactive archiving, 139interface settings, 48

Llab manager, 128limit lines, 47linear regression, 179lines

limit, 47locking ChemStore and ChemStation, 125logarithmic, 184logbook, 160

Mmacro, 132main toolbar, 33manager, 128marking runs, 39, 44, 50maximum, 178maximum value, 46mean, 178mean value, 46menu, 33method, 20method files, 23

minimum, 178minimum value, 46

Nnotification

e-mail, 157NT service pack, 173number, 178number format, 45number precision, 45

OODBC, 12operator, 127optimizing disk, 168organizing data, 13origin, 179

Ppage break, 96page footer, 96page header, 91, 96paragraph tags, 56password, 124, 125permissions, 124post-sequence report, 121power, 184precision

number, 45preferences dialog box, 121preview, 121

Qquery, 24, 134

builder, 25

Rraw data, 20, 50raw data files, 23real, 22regression statistics, 34, 35, 46, 179

rejecting runs, 39, 44, 50relative fit error, 183relative standard deviation, 46, 179reload files, 50report

cross-instrument, 22cross-sample, 22customized, 15

report template, 134report template editor, 88reports, 15, 88reprocessed run, 50reprocessing, 50residual, 181residual standard deviation, 181results

transferring, 13results table, 44retrieving data, 13review

compound, 35sample, 34, 36

reviewing data, 14RSD, 46, 179run, 20

reprocessed, 50run list, 39runs, 142

approving, 14, 39, 44, 50marking, 39, 44, 50rejecting, 39, 44, 50

Ssample review, 34, 36sample standard deviation, 181sample table, 40scalability, 16search criteria, 24second toolbar, 37, 42section, 93section header, 100security, 125

data, 124

Page 207: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent ChemStore C/S Concepts Guide 207

Index

selection list, 22sequence, 20, 121sequence files, 23sequence summary report, 121service pack, 173session locking, 125settings

user interface, 48single value calculations, 178spectra, 20, 96SQL, 24stand-alone version, 12standard deviation, 46, 179

coefficients, 181, 183relative, 46residual, 181sample, 181

statistics, 36, 41, 46, 107regression, 34, 35, 179summary, 46, 178

statusapproval, 27, 39, 44custom field, 22

study, 20, 24, 131sum, 46, 178summary results table, 44summary sample table, 40summary statistics, 46, 178

Ttable, 94, 96, 100

results, 44sample, 40summary results, 44summary sample, 40

tablesconfiguring, 44

tagsparagraphs, 56

text, 22Time stamps, 201Time zones, 202toolbar

main, 33second, 37, 42

transaction log, 124transferring results, 13tree view, 91trend analysis, 22trend chart, 14true/false, 22TZ, 202

Uupgrading, 16user interface settings, 48, 134user permissions, 124, 127user profile, 124, 128UTC, 202

Vvalidation, file, 161variance, 46, 178version, 27

number, 51versions, 51

Wwarning limits, 47weighting, 179where clause, 25workflow, 18

XXML, 141

Page 208: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

208 Agilent ChemStore C/S Concepts Guide

Index

Page 209: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...
Page 210: Agilent ChemStore C/S · Validation and Security of Printed Files 161 File Validation in Operation 161 ... Agilent ChemStore C/S Concepts Guide Agilent Technologies ...

Agilent Technologies 2002, 2004

Printed in Germany 03/04

*G2181-90010**G2181-90010*G2181-90010

www.agilent.com

Agilent Technologies

In This Book

This book introduces you to the main concepts of Agilent ChemStore C/S. It highlights the main features of the product and gives you some guidance for setting up your studies and managing your data.

It includes

• A brief overview of the features and capabilities of Agilent ChemStore C/S,

• Details on the operation of Agilent ChemStore C/S

• Details on setting up Agilent ChemStore C/S to suit the needs of different laboratories and users,

• Full details of how Agilent ChemStore C/S’s comprehensive suite of statistical calculation calculations are performed