  • Pitot Static Tester ADSE650

    ADSE650User Manual

    PITOT STATIC TESTER

    November 2014 User Manual ED 01 Page 1 of 82

    Reference D2-0174Edition 01Date 11-24-2014Author F. PLATEL

    Approved by O. PLATEAU

    USERTypewriterhttp://www.avionteq.com/ATEQ-Omicron-ADSE-650-Pitot-Static-Tester.aspx

    USERTypewriter

    USERTypewriter

    USERTypewriterwww.avionteq.com

    USERTypewriter

    LyrahTypewriterTo buy, sell, rent or trade-in this product please click on the link below:

    http://www.avionteq.com/ATEQ-Omicron-ADSE-650-Pitot-Static-Tester.aspxhttp://www.avionteq.com/

  • Pitot Static Tester ADSE650

    Revisions

    Date Edition

    Creation November 2014 01

    November 2014 User Manual ED 01 Page 2 of 82

  • Pitot Static Tester ADSE650

    Table of contents1. PRESENTATION..................................................................................................................7

    1.1 General........................................................................................................................................7

    1.2 Applications.................................................................................................................................8

    1.3 Manufacturer’s part numbers.................................................................................................10

    1.4 Characteristics..........................................................................................................................11

    1.5 Units, Range and Precision from 10°C to 40°C......................................................................121.5.1 Static pressure - Ps.................................................................................................................................12 .......................................................................................................................................................................121.5.2 Total and differential pressures - Pt - ΔP...............................................................................................121.5.3 Ps dynamic variations.............................................................................................................................131.5.4 Pt dynamic variations.............................................................................................................................131.5.5 Rates of intrinsic leak.............................................................................................................................13

    1.6 Technical description................................................................................................................141.6.1 Electro-pneumatic Unit..........................................................................................................................15

    1.6.1.1 Front Panel......................................................................................................................................151.6.1.2 Electro-pneumatic block..................................................................................................................151.6.1.3 Pressure transducers........................................................................................................................151.6.1.4 Proportionnal Electrovalves............................................................................................................15

    1.6.2 Remote Control Unit..............................................................................................................................161.6.3 Software..................................................................................................................................................17

    1.6.3.1 Language.........................................................................................................................................171.6.3.2 Pressure control...............................................................................................................................171.6.3.3 Automatic Test programs................................................................................................................171.6.3.4 Flight Domain..................................................................................................................................181.6.3.5 Self test............................................................................................................................................18

    2. USER ACCESS....................................................................................................................192.1 ADSE650 RCU front panel......................................................................................................19

    2.2 ADSE650 EPU front panel ......................................................................................................20

    3. USE.......................................................................................................................................213.1 Initial Procedure.......................................................................................................................21

    3.1.1 ADSE650 alternative power feeding......................................................................................................213.1.2 Remote control link................................................................................................................................213.1.3 Pneumatic link........................................................................................................................................213.1.4 Preparing for use....................................................................................................................................223.1.5 Powering up............................................................................................................................................223.1.6 Interface software...................................................................................................................................22

    3.2 Using the operator interface....................................................................................................233.2.1 Starting software.....................................................................................................................................23

    3.2.1.1 Connection established....................................................................................................................243.2.1.2 Connection not established..............................................................................................................25

    3.2.2 Access to the flight domains...................................................................................................................26

    November 2014 User Manual ED 01 Page 3 of 82

  • Pitot Static Tester ADSE650

    3.2.2.1 Select a flight domain......................................................................................................................273.2.2.2 Creating a flight domain..................................................................................................................293.2.2.3 Editing, Renaming and deleting a flight domain.............................................................................30

    3.2.3 Access to the Measure/Manual Mode ...................................................................................................313.2.3.1 Changing the unit on the Ps and Pt..................................................................................................333.2.3.2 Activating a channel .......................................................................................................................353.2.3.3 Generating on the Ps or Pt channel.................................................................................................353.2.3.4 Leak test..........................................................................................................................................373.2.3.5 Return to the ground........................................................................................................................393.2.3.6 Displaying the current flight domain...............................................................................................39

    3.2.4 Automatic program mode.......................................................................................................................413.2.4.1 Writing a new program....................................................................................................................423.2.4.2 Editing, Renaming and deleting a program file...............................................................................473.2.4.3 Executing a program.......................................................................................................................48

    3.2.4.3.1 Operating Mode.......................................................................................................................553.2.4.3.2 File format for the data recorded by the semi-automatic program...........................................57

    3.2.5 Calibration mode....................................................................................................................................593.2.5.1 Calibrating the sensors....................................................................................................................603.2.5.2 Operating mode...............................................................................................................................60

    3.2.6 General Information...............................................................................................................................633.2.6.1 Parameters ......................................................................................................................................643.2.6.2 Password..........................................................................................................................................653.2.6.3 Enter a password.............................................................................................................................683.2.6.4 Languages .......................................................................................................................................69

    3.2.7 Help on line............................................................................................................................................703.2.8 Exiting the software................................................................................................................................703.2.9 Keyboard configuration..........................................................................................................................71

    3.2.9.1 Digital keyboard..............................................................................................................................713.2.9.2 Alpha-Numeric Keyboard...............................................................................................................72

    4. MAINTENANCE.................................................................................................................734.1 Passwords..................................................................................................................................73

    4.2 Internal......................................................................................................................................73

    4.3 Low Level Maintenance...........................................................................................................74

    4.4 Contacts.....................................................................................................................................754.4.1 Mailing address......................................................................................................................................754.4.2 Internet...................................................................................................................................................754.4.3 Telephone numbers................................................................................................................................75

    4.5 Maintenance..............................................................................................................................764.5.1 Autotest..................................................................................................................................................764.5.2 Maintenance Plan...................................................................................................................................82

    5. ANNEXE : Safety information ..........................................................................................835.1 Operating environment............................................................................................................83

    5.2 Safety for Lithium Polymer battery use..................................................................................83

    5.3 Important instructions (for service personnel only)..............................................................84

    5.4 Recycling....................................................................................................................................84

    November 2014 User Manual ED 01 Page 4 of 82

  • Pitot Static Tester ADSE650

    Introduction

    The purpose of this manual is to provide the user with the information required to

    implement and operate the ADSE650 Pitot Static Tester.

    Exhibit 1:General view of ADSE650

    November 2014 User Manual ED 01 Page 5 of 82

  • Pitot Static Tester ADSE650

    1. PRESENTATION

    1.1 General

    The ADSE650 pressure controller is housed in a very robust housing box, able to be

    moved directly to the aircraft on the tarmac or in the hangar to proceed "in situ" to the

    tests of equipment fitted to the aircraft.

    The two parts of ADSE650:

    ● The Electro Pneumatic Unit (EPU) includes the computer, the pressure

    measurement and regulation devices, the security systems as well as

    the pressure and vacuum pumps necessary for independent operation

    in generation mode,

    ● The Remote Control Unit (RCU), located in the cover of the housing

    box, includes the Human Machine Interface. It consists of a Tablet PC

    with Windows 8 Pro fitted with a watertight touch sensitive plate

    displaying the static (Ps) and total (Pt) pressure parameters

    (functions, measurement units, commands, rate of change, nominal

    values, program number, flight limit number).

    RCU and EPU are connected through the Wifi. (In case of problem with the wifi, a

    cable can be connected between RCU and EPU)

    The ergonomics of the ADSE650 caters for:

    ● Excellent readability of the information,

    ● Easy programming of the configuration (interactive menus, withexplanations and indication of potential handling errors),

    ● Direct calibration by the user.

    November 2014 User Manual ED 01 Page 6 of 82

  • Pitot Static Tester ADSE650

    1.2 Applications

    The ADSE650 pressure generator/controller has been specially designed to simplify

    pressure tests on equipment fitted on the aircraft (altimeter, air speed indicator,

    rate of climb indicator, air data computers ...) would it be in the workshop, in the

    hangar or directly on the tarmac.

    The computer manages the security systems, the measurement and regulation

    devices from the remote control unit with following operating functions:

    ● In "Measure" or "Manual" mode, independently or simultaneously on both

    static (Ps) and total (Pt) pressure channels:

    ● Measurement or generation of static (Ps) pressures (altitude),

    ● Measurement or generation of total (Pt) pressures (air speed

    /MACH number),

    ● Chronometric measurement of leak rate.

    ● In "Program" mode, independently or simultaneously on both static (Ps) and

    total (Pt) pressure channels:

    ● "Flight domains" writing,

    ● "Automatic test" programs writing,

    ● Execution of the written programs. In this mode, the operator can

    choose to enter at each program step the actual data read on the

    equipment under test, or to let the program run and to note

    manually the values read on the equipment under test.

    ● Security:

    ● Full time survey of all the flight domain limits parameters,

    ● Re-connection of the pressure circuits to the atmospheric pressure

    through manual flow adjustable valves in case of power supply

    network cut-off,

    ● Self test of all the ADSE650 internal functions each time the

    equipment gets switched on.

    November 2014 User Manual ED 01 Page 7 of 82

  • Pitot Static Tester ADSE650

    ● Calibration:

    ● The calibration of the ADSE650 pressure transducers can be

    fulfilled easily by the maintenance manager.

    ● Baro-setting:

    ● The baro-setting (Ref) adjustment is performed by the operator

    himself and remains displayed permanently.

    ● Altitude difference (between the bench position and the equipment under

    test):

    ● The operator can introduce in the measurement process the

    altitude difference between the bench on the ground and the

    equipment fitted in the aircraft cockpit.

    November 2014 User Manual ED 01 Page 8 of 82

  • Pitot Static Tester ADSE650

    1.3 Manufacturer’s part numbers

    Commercial Name Designation of materials ReferenceADSE650 Ps and Pt Air Data Test Set (0.01% FS) 650-0001

    HOSE-PS-10 Red Air Hose Ps (L=10 m) P6-0126HOSE-PT-10 Blue Air Hose Pt (L=10m) P6-0127CABLE-A-10 Alternative Current Cable (L=10m) T1-0025

    LINK-SERIAL-5 Serial Link Cable (5m) T1-0110

    November 2014 User Manual ED 01 Page 9 of 82

  • Pitot Static Tester ADSE650

    1.4 Characteristics

    Power supply 90 to 240 V AC from 50 to 400 Hz

    Normal operating temperature range -5°C to + 50°C

    Storage temperature -10°C to + 55°C

    Overall dimensions of the equipment Length 339 mm / 13.3 inch

    Depth 295 mm / 11.6 inch

    Height 152 mm / 5.9 inch

    Outside dimensions of the remote controlunit

    Width 230 mm / 9 inch

    Height 140 mm / 5.5 inch

    Depth 18 mm / 0.7 inch

    Total weight (without cables and tubes) 6 kg / 13.2 lbs

    Weight of the RCU 0.5 kg / 1.1 lbs

    Power consumption Less than 120 VA ~

    November 2014 User Manual ED 01 Page 10 of 82

  • Pitot Static Tester ADSE650

    1.5 Units, Range and Precision from 10°C to 40°C

    1.5.1 Static pressure - Ps

    Channel Unit Range Precision Resolution

    Ps hPa 70 to 1,200 hPa ± 0.1 hPa 0.01 hPa

    Ps InHg 1 to 35 inHg ± 0.0003 inHg 0.001 inHg

    Altitude

    Zp

    ft -2,300 to 60,000 ft ± 3 ft at 0 ft

    ± 8 ft at 30,000 ft

    ± 32 ft at 60,000 ft

    1 ft

    m -700 to 18,000 m ± 1 m at 0 m

    ± 3 m at 10,000 m

    ± 10 m at 18,000 m

    1 m

    1.5.2 Total and differential pressures - Pt - ΔP

    Channel Unit Range Precision Resolution

    Pt hPa 70 to 2000 hPa ± 0.15 hPa 0.01 hPa

    ΔP hPa 0 to 1000hPa ± 0.10 hPa 0.01 hPa

    Speed

    Vc

    kts 30 to 650 kts ± 2 kt at 50 kts

    ± 0.14 kt at 500 kts 0.1 kt

    km/h 35 to 1,480 ± 3 km/h at 100 km/h

    ± 0.26 km/h at 900 km/h

    0.1 km/h

    Mach M 0.1 to 4.0 M ± 0.0006 M at 0.8 M at 25,000 ft

    ± 0.0003 M at 1.7 M at 30,000 ft

    0.001 M

    November 2014 User Manual ED 01 Page 11 of 82

  • Pitot Static Tester ADSE650

    1.5.3 Ps dynamic variations

    Ps channel Unit Range Precision

    ΔPs / Δt hPa / min 10 to 150 hPa / min ±1hPa or ±1 % (1)

    ΔZp / Δt (VZ) ft / min 100 to 6,000 ft / min ±10 ft or ± 1 % (1)

    m / min 30 to 1,800 m / min ±3 m or ± 1 % (1)

    1.5.4 Pt dynamic variations

    Pt channel Unit Range Precision

    ΔPt / Δt hPa / min 10 to 150 hPa / min ±2 hPa or ± 1 % (1)

    Δ Δp / Δt Δ hPa / min 10 to 150 hPa / min ±2 hPa or ± 1 % (1)

    ΔVc / Δt Kts / min 10 to 200 kts / min ±2 kts or ± 1 % (1)

    ΔVc / Δt Kmh / min 10 to 370 kmh / min ±3.7 km/h or ± 1 % (1)

    ΔM / Δt M / min 0.1 to 1 M / min ±0.01M or ± 1 % (1)

    ● (1) on an integration time greater than 3 s,

    ● These performances are valid on a volume of 0 to 1000cc,

    ● The user can limit all values with "Flight limits".

    1.5.5 Rates of intrinsic leakThe intrinsic leak of the ADSE650 on the two channels is maximum 0.7 hPa/min at

    500 hPa without outside volume (test time: 15 minutes).

    November 2014 User Manual ED 01 Page 12 of 82

  • Pitot Static Tester ADSE650

    1.6 Technical description

    The ADSE650 includes two parts:.

    ● The Electro-pneumatic Unit,

    ● The Remote Control Unit.

    November 2014 User Manual ED 01 Page 13 of 82

  • Pitot Static Tester ADSE650

    1.6.1 Electro-pneumatic Unit

    1.6.1.1 Front Panel

    The front panel of the main box is equipped with:

    ● One static pressure output,

    ● One total pressure output,

    ● One power supply connector,

    ● Balancing Tap,

    ● One connector for the remote control unit link,

    ● Power on visualization and Power on,

    ● One Wifi Antenna,

    ● One TabletPC.

    1.6.1.2 Electro-pneumatic block

    Secured on a chassis below the upper plate, is located the metrological part of the

    system. The bloc includes the pressure transducers, several electro-valves, the

    servo-loop gates, the power supply unit and the computer.

    1.6.1.3 Pressure transducers

    Pressure transducers are of the newest technology ensuring best accuracy.

    The pressure sensors result from the last technological developments ensuring a

    reduction of the costs related to a manufacturing simplification. These sensors easily

    resist shocks such as overpressure.

    The sensors installed in the ADSE650 are guaranteed for uses of operation between

    10°C and 40°C with an accuracy of 0.01% of the full scale (with the

    pressure/temperature software correction).

    1.6.1.4 Proportionnal Electrovalves

    Two Proportionnal Electrovalves are used on each channel to regulate Pressure and

    vaccuum

    November 2014 User Manual ED 01 Page 14 of 82

  • Pitot Static Tester ADSE650

    1.6.2 Remote Control Unit

    The remote control Unit (RCU) is a commercial Tablet PC located in the cover of the

    housing with Windows 8 Pro.

    The link between RCU and EPU happens through the wifi network.

    One cable link can be used to connect the EPU and the RCU with RS232 protocol

    ( Serial Cable Link T1-0110).

    November 2014 User Manual ED 01 Page 15 of 82

  • Pitot Static Tester ADSE650

    1.6.3 Software

    1.6.3.1 Language

    The EPU computer’s Software is written in C langage .

    The use of this language ensures the separation of the processing and the

    permanent control of the various software modules in order to obtain a more stable

    software base.

    The following modules of the software are:

    ● Sensors management,

    ● Pressure control management,

    ● Security management,

    ● Wifi / RS232 management.

    The RCU Software is written in Labview and works under Windows 8.

    The following modules of the software are:

    ● User Interface management,

    ● Wifi / RS232 management.

    1.6.3.2 Pressure control

    When the execution of an instruction is required, the starting of the pressure and

    vacuum pump is activated.

    When the current pressure approaches the final set point, control pressure slows

    down its evolution until the complete stop. This threshold of deceleration is

    calculated according to the evolution speed and the pressure.

    1.6.3.3 Automatic Test programs

    All programs are listed in a table accessible through the "Program" menu.

    Writing or editing a program: each program step is presented on a separate screen,

    where the user chooses or modifies the instructions for this step (altitude, speed,

    evolution to these target values, leak test...)

    November 2014 User Manual ED 01 Page 16 of 82

  • Pitot Static Tester ADSE650

    1.6.3.4 Flight Domain

    All information requested to write or edit a flight domain (altitude, rate of climb,

    speed and acceleration) is presented in one screen, ensuring that keying in the

    requested data is made very easy.

    A table recapitulates all the written domains.

    1.6.3.5 Self test

    When the equipment gets turned on, an automatic test of the entire bench is

    performed. If an error is detected, a message is written on the screen and the bench

    is unusable until the problem is fixed.

    November 2014 User Manual ED 01 Page 17 of 82

  • Pitot Static Tester ADSE650

    2. USER ACCESS

    2.1 ADSE650 RCU front panel

    1. Charger Cable / Remote control link cable.

    2. ON/OFF Tablet Button.

    Exhibit 2: ADSE650 remote control front panel

    November 2014 User Manual ED 01 Page 18 of 82

    1 2

  • Pitot Static Tester ADSE650

    2.2 ADSE650 EPU front panel 1. Remote Control Connector,

    2. Wifi Antenna,

    3. Charger “USB” for TabletPC

    4. Power supply connector,

    5. Led charger,

    6. "Ps" channel ,

    7. Balance Tap (Ps to Pt),

    8. "Pt" channel,

    9. Vent Tap (Pt to Atmosphere)

    10. ON/OFF Button

    Exhibit 3: ADSE650 main box front panel

    November 2014 User Manual ED 01 Page 19 of 82

    1 2 3 4 5

    6

    10

    7

    8

    9

  • Pitot Static Tester ADSE650

    3. USE

    3.1 Initial Procedure

    3.1.1 ADSE650 alternative power feeding

    For alternative current (90 to 240 V from 50 to 400 Hz), use the cable T1-0025.

    Connect the plug to the connector 4 .

    3.1.2 Remote control link

    In case of Wifi connexion problem, connect the remote control unit with a tool kit

    cable. Connect one of the ends in the plug on the right side of the remote control

    unit and the other in the plug marked "RCU" of the front panel of the bench.

    (Connector 1)

    3.1.3 Pneumatic link

    Take the hoses with the pneumatic plugs, Red colour for Ps (P4-0126) and Blue

    colour for Pt (P4-0127). Connect them on the front panel.

    Then connect the hoses to the instruments to be tested.

    Warning:

    - In case of replacement of hoses, use the same kind with the same length.

    - During all procedures detailed in the next pages, observe the appropriate

    safety instructions and procedures stated in local orders and the aircraft

    and instruments maintenance manuals.

    November 2014 User Manual ED 01 Page 20 of 82

  • Pitot Static Tester ADSE650

    3.1.4 Preparing for use

    Check the following:

    ● Voltage source,

    ● Electrical connection(s).

    3.1.5 Powering up

    ● Set the momentary On/Off button during 2s on the EPU. The Blue Led

    blinks during the boot and remains lit.

    ● Switch on the TabletPC by pressing the ON/OFF button until the Logo

    “Dell” appears.

    3.1.6 Interface software

    On the power up, the TabletPC boot on Windows 8, run the program “ADSE” by

    clicking on the shortcut on the desk.

    3.1.7 Shutting off

    ● Set the momentary On/Off button on the EPU during 2s.

    ● Switch off the TabletPC by pressing the ON/OFF button during 1 to 5s and

    follow the instruction on the Tablet to shut down the PC.

    NB: To avoid to discharge the TabletPC quickly, shut down the PC as

    describe and avoid to stay in sleep mode.

    November 2014 User Manual ED 01 Page 21 of 82

  • Pitot Static Tester ADSE650

    3.2 Using the operator interface

    3.2.1 Starting software

    When the software gets started, the above screen is displayed as long as the

    computer is running its self test and the connection between the servo control

    computer and the user interface computer is not established.

    By pressing "Cancel Connection", the operator can interrupt the connection process.

    The computer will then go to "Disconnected Mode" and will not try to connect again.

    Exhibit 4

    November 2014 User Manual ED 01 Page 22 of 82

  • Pitot Static Tester ADSE650

    3.2.1.1 Connection established

    As soon as the connection is established, the information "Connected Mode"

    appears. All keys on the screen become accessible to the user.

    If the self test has been completed without problem, no message appears in the

    information area situated at the bottom of the screen.

    Exhibit 5

    Range limits: To select, create or modify domains,

    Measure / Manual Mode: To use ASDE in manual mode,

    Automatic Program: To select, create, modify and execute automatic

    program,

    Calibration : To calibrate transducer Ps and Pt,

    Exit: To Exit from application and come back to Windows.

    Note: The date of the last calibration is indicate on the bottom field text.

    November 2014 User Manual ED 01 Page 23 of 82

  • Pitot Static Tester ADSE650

    3.2.1.2 Connection not established

    If the computer could not complete its self test, the connection with the user

    interface computer does not happen. The message "Mode not connected" appears.

    To retry the connection:

    ● Check, if at the power up of windows, in the taskbar of windows 8, the

    icone “Wifi limited” appears.

    ● Check if the Wifi network is available and correct on the Tablet,

    (SSID=OMI650XXXX, XXXX= serial number of ADSE)

    ● Check the parameters of your Tablet (§3.2.6 General Information )

    ● Exit from the application

    ● restart the Pneumatic unit and waiting for the wifi connection on the

    Tablet

    ● and re-launch “ADSE” on the Tablet.

    If the problem persists, contact the after sales service of ATEQ.

    November 2014 User Manual ED 01 Page 24 of 82

  • Pitot Static Tester ADSE650

    3.2.2 Access to the flight domains

    Press "Range limits" to access this menu. It gives access to following functions:

    ● Select an existing flight domain,

    ● Create a new flight domain,

    ● Edit an existing flight domain.

    Exhibit 6

    November 2014 User Manual ED 01 Page 25 of 82

  • Pitot Static Tester ADSE650

    3.2.2.1 Select a flight domain.

    The selection of a flight domain happens through a window presenting all existing

    flight domains.

    ● Press "Cancel" to come back to the "Range limits" main menu,

    ● Select the desired Domains,

    ● Press "Validate" to visualize the selected flight domain.

    Exhibit 7

    November 2014 User Manual ED 01 Page 26 of 82

  • Pitot Static Tester ADSE650

    Exhibit 8

    All parameters of this flight domain appear, the user can choose between:

    ● "Return" to return to the list of flight domains,

    ● "Validate" to select this flight domain immediately (but this selection will be

    lost at the next start of the software),

    ● "By Default" to select this flight domain immediately and keep it as domain

    by default at the start of the system.

    Note: The flight domain called "MAX" is a reserved flight domain and can

    neither be modified nor deleted.

    November 2014 User Manual ED 01 Page 27 of 82

  • Pitot Static Tester ADSE650

    3.2.2.2 Creating a flight domain

    The range limits creating mode is accessible through a password.

    Upon creation of a flight domain, the value fields are pre-programmed with the

    maximum values acceptable by your ADSE650.

    The user must:

    ● Enter a domain name by clicking on the field, which gives access to an

    alphanumeric keyboard,

    ● Modify the values by using the digital keyboard which appears by clicking

    on the associated field.

    Once all parameters are entered, press "Save" to validate and record the data or

    press "Return" to abandon the flight domain creation and come back to the "Range

    limits" main menu.

    Exhibit 9

    November 2014 User Manual ED 01 Page 28 of 82

  • Pitot Static Tester ADSE650

    3.2.2.3 Editing, Renaming and deleting a flight domain

    The range limits editing mode is accessible through a password.

    Editing a flight domain happens through a window presenting all existing flight

    domains (except "MAX").

    ● Rename domain by clicking on the field, which gives access to an

    alphanumeric keyboard,

    ● "Return" to come back to the "Range limits" main menu,

    ● "Validate" to edit the selected flight domain.

    All parameters of this flight domain appear; the user can modify the values

    by using the digital keyboard which appears by clicking on the associated

    field.

    After entering the data, press

    ● "Return" to come back to the "Range limits" main menu,

    ● "Save" to record the selected flight domain.

    Note: You can erase a domain file when you edit it. For this, you can use the "Delete"

    button (you must confirm your choice).

    Exhibit 10

    November 2014 User Manual ED 01 Page 29 of 82

  • Pitot Static Tester ADSE650

    3.2.3 Access to the Measure/Manual Mode

    Exhibit 11

    November 2014 User Manual ED 01 Page 30 of 82

    State of ADSE

    Activating Channel

    Back To main Menu

    To Stop Generation

    To go back at Ground

    To make leak testTo visualize

    Domain

    Select UnitPt

    Reference for AltitudeCurrent Unit

    Gap Altitude between ADSE and pitot tube

    To Execute a set point

    Step by step

    increment Or Alticoder Information

    Select unitPs

    To visualize The User Manual

    Don't stop generation if leak or big volume

  • Pitot Static Tester ADSE650

    In this Mode, the operator can access to:

    ● Manual Mode: Access to the unit, pressure information, generation

    mode,

    ● Leak test: To run Leak test on Ps and Pt,

    ● Domain: To view the current domain,

    ● Help: To view the Help in line.

    On each channel (Ps and Pt), the user can:

    ● Change the pressure unit with the scrolling menu,

    ● Enabling or disabling the channel, in order to access the instruction

    fields for a generation, a leak test...,

    ● Enter the generation instructions,

    ● Launch a generation,

    ● Increment the value step by step in the selected unit,

    ● Move to the Leak test function,

    ● Come back to the ground,

    ● Stop a generation.

    Note: Mode Big Capacity: This mode is used when you have a big volume and the rate of

    climb is not an important parameter. In this case, the ADSE regulate with the

    desired rate of climb if the volume is too big, or if there is an important leak, the

    tester will keep generating at a lower rate of climb. In the standard case, the ADSE

    stop generation if one parameter can not be respect.

    In this mode, you have not messages on the ADSE limits (example with Ps channel :

    "Ps : Limits Exceeded").

    November 2014 User Manual ED 01 Page 31 of 82

  • Pitot Static Tester ADSE650

    3.2.3.1 Changing the unit on the Ps and Pt

    To change the unit, press the "Alt." button (for the Ps channel) or "Speed" button

    (for the Pt channel) and select the unit from the scrolling list.

    The change is active when the requested unit appears in the field "Actual unit". The

    pressure evolution is then displayed in the new selected unit.

    Important notice: a unit change does not result in an update of the target and rates

    values, due to the non linearity between pressure units on one side and altitude and

    speed units on the other side.

    Unit Ps: hPa, Psi, mmHg, inHg, ft, m...

    Exhibit 12

    November 2014 User Manual ED 01 Page 32 of 82

  • Pitot Static Tester ADSE650

    ● Unit Pt: hPa, Psi, mmHg, inHg, DhPa, DPsi, DmmHg, DinHg, Kts, km/h, MPH,Mach...

    Exhibit 13

    Note: If the focus stays on the Units button more of five seconds (see: softwareparameter), an Unit Window opens. With it, you can select units more easily.

    Exhibit 14

    November 2014 User Manual ED 01 Page 33 of 82

  • Pitot Static Tester ADSE650

    3.2.3.2 Activating a channel Enabled or disabled a channel happens by clicking on the "Ps" or "Pt" button. A

    green or red light and a message indicate the status of the channel:

    ● If the channel is active, the light is green and the message displayed is

    "Channel ON".

    ● If the channel is inactive, the light is red and the message displayed is

    "Channel OFF".

    Activating a channel gives access to the instruction fields and to the "Execute",

    "Leak" and "Ground" functions.

    When a channel is inactive and will not be used during a generation, it is

    recommended to leave the corresponding outlet open, in order to keep the internal

    pressure of this channel equal to the ambient pressure.

    Warning: The function of generating on Ps and leaving the Pt channel inactive

    should be used with major care, especially when the static circuit of the aircraft is

    pneumatically linked to the airspeed indicator: generating on Ps will cause the

    airspeed indicator to climb, potentially outside its range limits.

    3.2.3.3 Generating on the Ps or Pt channel

    To perform a generation,

    ● Active the channel,

    ● Select the desired unit

    ● Enter the altitude (for Ps) or speed (for Pt) value and the climb rate (Ps) or

    acceleration (Pt):

    ● Select the instruction field. This field will flicker in red.

    ● Enter the desired value with the keyboard. Repeat this operation with the

    "Rate" field.

    If one of the values entered is outside the flight domain, a message "Domain

    overflow" will appear at the bottom of the screen and the value will be forced to the

    November 2014 User Manual ED 01 Page 34 of 82

  • Pitot Static Tester ADSE650

    closest value within the flight domain (either min or max value within the flight

    domain, depending whether the entered value was below or over the allowable

    values).

    Executing the instruction:

    In order to execute the instruction, click on "Execute". The data entered in the

    instruction fields of the active channels are taken into account and the ADSE650

    starts generating.

    Different messages appear during the different phases of the generation.

    ● Before the generation starts, the message indicates "Measure": the ADSE

    is in slave mode and displays the pressure without controlling it.

    ● When the generation starts, the ADSE will manage the evolution of the

    pressure. During the first phase, the tester will start the pumps (message

    displayed: "Starting"), then balance the pressure internally to avoid a

    pressure surge (Message displayed: "Balancing").

    ● During the next phase, the pressure will evolve at the set rate to reach the

    set target. The message displayed indicates "Run".

    ● During the last phase, the pressure is stabilized at the set target. The

    message displayed indicates "Ready".

    Note #1: In execution mode, the automatic increment of the target is accessible. By

    using this function, the operator can increment or decrement the target pressure by

    the value indicated between the "+" and "-" buttons. By clicking the "+" or "-"

    button, the operator increases or decreases the target value. The ADSE650 updates

    the display of the instructions accordingly and starts generating the pressure

    according to this new target. To modify the increment step, click on the

    corresponding field and enter the requested value with the keyboard.

    Note #2: If you change an instruction during the generation, click on "Execute" to

    request the generator to take this new value into account.

    November 2014 User Manual ED 01 Page 35 of 82

  • Pitot Static Tester ADSE650

    Note #3: When both channels are active, clicking on "Execute" requests the

    generation on both channels simultaneously. In this case, the generation starts first

    on the Pt channel. When the rate is stabilized on Pt, the generation starts on Ps.

    Note #4: If the operator clicks "Execute" without having entered a value in the

    "Rate" field, the system will display the message "Domain overflow" and force the

    rate to the minimum value.

    3.2.3.4 Leak test

    When the operator switches to the Leak test mode, the ADSE moves automatically

    to the "Measure" mode on both channels. The unit for the leak test is the unit

    selected in the main menu of the manual mode.

    To access this screen, click on the "LEAK" button of the manual mode menu.

    Exhibit 15

    To configure the test, enter a time. Click on the minutes and seconds field and enter

    the values with the keyboard.

    November 2014 User Manual ED 01 Page 36 of 82

  • Pitot Static Tester ADSE650

    To launch the leak test, click on "Start". The time starts being counted down until 0

    minute and 0 second. The leak is displayed in the "Leak" field. In "Leak/min", you

    find the leak for one minute.

    During the test, the fields "Value" and "Rate" evolve. The operator can stop the leak

    test at any time by clicking on the "Stop" button.

    To come back to the main screen of the manual mode, click on the "MANUAL"

    button. The ADSE650 returns to the generation mode.

    Important notice: if a unit change has been performed, the set targets might be

    inconsistent with the units. Recheck carefully the set values before starting a

    generation.

    Note:

    ● For pressure unit (hPa, Psi ...):

    Result=Final Pressure – Initial Pressure

    ● For altitude and airspeed unit (ft, m, kts, Mach...):

    Result=Initial value – Final Value

    November 2014 User Manual ED 01 Page 37 of 82

  • Pitot Static Tester ADSE650

    3.2.3.5 Return to the ground

    When the operator clicks on "Ground", the ADSE650 compares the ambient pressure

    with the pressures in the Ps and Pt circuits. If on one of the channels, the pressure

    is different, the ADSE650 turns to the generator mode and returns to the

    atmospheric pressure in the selected unit and using the last rate entered. The

    message "Back to Ground" is displayed at the bottom of the screen

    Once the internal pressure has reached the ambient pressure, the ADSE turns to the

    "Measure" mode and opens the internal circuit to the ambient pressure. The

    message "Atmosphere" is displayed at the bottom of the screen. The equipment in

    test can be securely disconnected from the ADSE650.

    Note: When the system is back to ground ("Atmosphere" displayed on both

    channels) the operator has access to:

    ● The correction of Qfe: Choice between 1013.25 hPa and atmospheric

    (local) pressure as reference for altitude simulation.

    ● The correction for altitude gap: Parameter in feet or meter equal to

    the difference between the measurement of the pressure at the level

    of the test set (Ps outlet) and the altimeter situated in the aircraft.

    3.2.3.6 Displaying the current flight domainExhibit 16

    At any time, the operator can click on the "Domain" button to display the flight

    domain currently in use.

    To return to the main screen, click on the "Back" button.

    November 2014 User Manual ED 01 Page 38 of 82

  • Pitot Static Tester ADSE650

    3.2.4 Automatic program mode

    The program mode allows the user to create, visualize, edit and start an automatic

    test program.

    Exhibit 17

    Note : See "Autotest" section in "Maintenance" chapter to use the "Autotest" button.

    November 2014 User Manual ED 01 Page 39 of 82

  • Pitot Static Tester ADSE650

    3.2.4.1 Writing a new program

    The program writing mode is accessible through a password.

    Steps for writing a flight program:

    ● Click on "New Program". This gives access to the screen shown,

    Exhibit 18

    ● Enter a file name by clicking on the field, in order to have access to the

    alphanumerical keyboard. The list of existing files is given for

    information. When the "Validate" button is clicked, a consistency check is

    performed.

    ● If the field is empty, the application returns to the main menu

    ● If the program name already exists, a connection box is displayed,

    indicating the problem. After validation, the user is requested to enter

    a new name.

    November 2014 User Manual ED 01 Page 40 of 82

  • Pitot Static Tester ADSE650

    ● Write a program with a maximum of 40 steps

    ● To save the data, click on "Save",

    ● To delete a current program step, click on "Delete (Step)",

    ● To insert a step before the step displayed, click on "Insert (Step)",

    ● To Add a step click on " ",→

    ● To cancel the creation of the program, click on "Main Menu".

    ● "Measure" Mode: The ADSE650 switches to the pressure measure mode and

    stops regulating the pressure. In this mode, the measurement unit can be chosen,

    in which the information shall be displayed, as well as the level time before the

    program goes to the next step.

    Exhibit 19

    November 2014 User Manual ED 01 Page 41 of 82

  • Pitot Static Tester ADSE650

    ● "Generation" Mode: In this mode, the altitude or speed value, the climb or

    acceleration rate and the measurement unit used for the generation need to be

    entered. Once the generation level has been reached, the level time entered by the

    user is counted down before the program goes to the next step.

    Exhibit 20

    November 2014 User Manual ED 01 Page 42 of 82

  • Pitot Static Tester ADSE650

    ● "Leak" Mode: In this mode, the ADSE650 switches into measure mode, waits until

    the pressure stabilizes in the circuit during a time corresponding to the level entered by

    the user, then starts the leak test entered in the "Time" field and in the entered

    measurement unit. Once the time has been counted down, the user is requested to

    validate, so that the program continues to the next step.

    Exhibit 21

    November 2014 User Manual ED 01 Page 43 of 82

  • Pitot Static Tester ADSE650

    ● "Ground" Mode: The ADSE650 automatically comes back to the ground, using

    the last measurement unit and the last rate of climb validated, and then opens the

    electro-valve putting the circuit at atmospheric pressure. If the generator is already

    at atmospheric pressure (+/- 5 hPa), only the electro-valve opening to the

    atmospheric pressure is activated. Once the ADSE650 has reached the atmospheric

    pressure, the level time entered by the user is counted down before the program

    goes to the next step.

    Exhibit 22

    Note: In the case of a request to return to ground, when no generation has been

    performed before but the generator is at a pressure different from the atmospheric

    pressure, the equipment comes back to the ground using the hPa as measurement

    unit and a climb rate equal to 100 hPa/min.

    November 2014 User Manual ED 01 Page 44 of 82

  • Pitot Static Tester ADSE650

    3.2.4.2 Editing, Renaming and deleting a program file

    ● Click on "Read/Edit Program" on the Automatic Program menu (Exhibit

    17),

    ● Select the desired program,

    Exhibit 23

    ● The screens displayed are identical to the screens of the creation mode.

    The user can edit the program, delete or modify existing steps or add

    new steps.

    ● To save the modifications, click on "Save", and enter the right password.

    ● To quit without saving, click on "Menu".

    Note: You can erase a program file when you edit it. For this, you can use the

    "Delete" button. You must confirm your choice.

    November 2014 User Manual ED 01 Page 45 of 82

    file:///L:/RX_suivi/Suvi%20D%C3%A9veloppement/R%20&%20D%20ADSE/ADSE%20650/enterapassword

  • Pitot Static Tester ADSE650

    3.2.4.3 Executing a program

    Important Notice: Before starting a program, it is recommended to make

    sure that the test set is at ground level:

    ● Select "Measure/Manual Mode" in the main menu:

    ● Return the system back to the ground ("Atmosphere" displayed on both

    channels),

    Exhibit 24

    November 2014 User Manual ED 01 Page 46 of 82

  • Pitot Static Tester ADSE650

    Exhibit 25

    Exhibit 26

    November 2014 User Manual ED 01 Page 47 of 82

  • Pitot Static Tester ADSE650

    Note: You can select "Big Capacity" mode before to start program.

    Exhibit 27

    Exhibit 28

    November 2014 User Manual ED 01 Page 48 of 82

  • Pitot Static Tester ADSE650

    Exhibit 29

    Exhibit 30

    November 2014 User Manual ED 01 Page 49 of 82

  • Pitot Static Tester ADSE650

    Exhibit 31

    Exhibit 32

    November 2014 User Manual ED 01 Page 50 of 82

  • Pitot Static Tester ADSE650

    Exhibit 33

    Exhibit 34

    November 2014 User Manual ED 01 Page 51 of 82

  • Pitot Static Tester ADSE650

    Exhibit 35

    Exhibit 36

    November 2014 User Manual ED 01 Page 52 of 82

  • Pitot Static Tester ADSE650

    3.2.4.3.1 Operating Mode

    ● Click on "Start Program" (Exhibit 17) to access screen shown on Exhibit 24

    ● Select a test program in the list of existing programs and "Validate",

    ● A configuration window (Exhibit 25) appears to:

    ● Enter a number of loops (number of times you want the program to

    repeat).

    ● Select the pressure reference (Choice between 1013.25 hPa and local

    (Qfe) pressure).

    ● Enter the altitude gap between the test set (Ps outlet) and the altimeter

    situated in the aircraft. For example, if the altimeter is 10 feet above the

    test set, enter "10". If it is 10 feet below the test set, enter "-10".

    ● Select automatic (Exhibit 25) or semi automatic (Exhibit 26) program. In

    semi automatic mode, the test set pauses after each program step. The

    operator can enter the values read on the instruments as well as

    comments, separately for the Ps and the Pt channels. These data (value

    and comments) is saved in a spreadsheet (see next paragraph).

    ● Validate to start program.

    ● The step #1/x is displayed with the program Loop #1/y and the name of the

    selected program (Exhibit 27).

    ● Click on "Execute" to start the program.

    ● The step #1 of the loop #1 is displayed and executed and so forth until step

    #x/x and Loop #y/y for the automatic mode.

    November 2014 User Manual ED 01 Page 53 of 82

  • Pitot Static Tester ADSE650

    ● Note #1: In semi automatic mode, at the end of each step, two buttons "Edit

    Ps" and "Edit Pt" appear (Exhibit 28) to enter the value shown on the

    altimeter and/or airspeed indicator and comments (Exhibit 29). To resume

    the program, press "Execute".

    ● Note #2: At any time, the user can switch to the standby mode by clicking on

    "Pause". If the generator is counting down a level time, the "Pause"

    instruction will be taken into account at the end of the countdown. To resume

    the execution of the program, click on "Execute" (Exhibit 28).

    ● Note #3: At the end of a program, the user can exit the program by clicking

    on "Menu" or re-launch the program by clicking on "Exec".

    ● Note #4: At any time, the currently selected flight domain can be visualized

    by clicking on "Domain".

    ● Note #5: The mode "Big Capacity" is only visible. You can not enabled or

    disabled it when the program is started.

    ● Note #6: At the end of a program, if the channels are not at the atmosphere

    pressure, you could go back to the ground pressure with the button

    "Ground".

    November 2014 User Manual ED 01 Page 54 of 82

  • Pitot Static Tester ADSE650

    3.2.4.3.2 File format for the data recorded by the semi-automaticprogram

    At the end of the semi automatic mode, a text file, formatted to be opened with

    Excel, is created in the directory "C:\ADSE 743\Results\" with the format "XXXX

    YYYY ZZZZ.txt" where XXXX is the Name of program, YYYY corresponds to the date

    and ZZZZ corresponds to the time.

    This example shows the "TOTAL 23_09_2010 05_49_34_PM.txt" file program

    opened with Excel.

    This file is accessible by users allowed to access the underlying Windows XP

    operating system (see 3.3.11: Exiting the software). From there, the remote control

    unit is usable as a normal Windows PC.

    Exhibit 37

    Spreadsheet cells content

    ● "A1": Program name

    ● "A4": Date of test

    ● "A5": Time of test beginning

    ● "A8": Field Name

    ● "A11": Field Comments

    November 2014 User Manual ED 01 Page 55 of 82

  • Pitot Static Tester ADSE650

    ● "A16": Pressure reference configuration

    ● "A19": Altitude gap between the test set and the aircraft altimeter

    ● Line "25 to xx": Information step by step:

    ● Column A: "Ps Mode" ->Measure, Manual, Leak, Ground

    ● Column B: "Ps Value" ->Set point for altitude

    ● Column C: "Ps Rate" ->Set point for rate of climb

    ● Column D: "Ps Unit" ->Unit used for the altitude

    ● Column E: "Ps Read" ->Value entered by the operator (or

    automatically in Leak mode in the unit used)

    ● Column F: "Ps Comments" ->Comments entered by the operator

    ● Column H: "Pt Mode" ->Measure, Manual, Leak, Ground

    ● Column I: "Pt Value" ->Set point for airspeed

    ● Column J: "Pt Rate" ->Set point for acceleration

    ● Column K: "Pt Unit" ->Unit used for speed

    ● Column L: "Pt Read" ->Value entered by the operator (or

    automatically in Leak mode in the unit used)

    ● Column M: "Pt Comments" ->Comments entered by the operator

    ● Column O: "Time Leak Minute" ->Indicates the leak test duration

    (minutes)

    ● Column P: "Time Leak Second" ->Indicates the leak test duration

    (seconds)

    November 2014 User Manual ED 01 Page 56 of 82

  • Pitot Static Tester ADSE650

    3.2.5 Calibration mode

    The calibration mode is accessible through a password.

    The first step is to validate the date or enter the right date.

    Exhibit 38

    Choose the channel to be calibrated (Ps or Pt),

    Exhibit 39

    November 2014 User Manual ED 01 Page 57 of 82

  • Pitot Static Tester ADSE650

    3.2.5.1 Calibrating the sensors

    Calibrating the sensors of the ADSE650 requires a pressure standard in a climate

    controlled room with a temperature regulated between 20 and 25°C.

    ● Place the ADSE 743 in the climate controlled room, next to the pressure

    standard,

    ● Turn it on,

    ● Leave the equipment for at least 30 min to stabilize in temperature.

    3.2.5.2 Operating mode

    Choose the channel to be calibrated (Ps or Pt),

    Exhibit 40

    Different information are displayed on the screen:

    ● The serial number of the sensor,

    ● The uncorrected pressure measured by the sensor in the "Measure" field.

    November 2014 User Manual ED 01 Page 58 of 82

  • Pitot Static Tester ADSE650

    Three buttons are accessible as well as a data entry field "Standard".

    The calibration procedure consists of:

    ● Generating controlled pressures with a pressure/vacuum generator connected

    on one of the Ps or Pt pneumatic connectors and to a pressure standard,

    ● Reading the value of the pressure standard and entering it in the "Standard"

    field with the digital keyboard which appears by clicking in this field.

    ● Click on "Validate Measure" to Validate the value entered.

    ● Validating the calibration once all points (pressure values) have been entered,

    by clicking on "Validate calibration".

    When the calibration is validated:

    ● The data is saved in a backup file,

    ● The data is sent to the computer which takes into account these new

    parameters.

    To obtain an optimal calibration of the equipment, ATEQ-OMICRON recommends to

    perform the calibration by measuring following pressure values:

    On the Ps channel: 65, 100, 200, 300, 400, 500, 600, 700, 800, 900, 950,

    1000, 1050, 1100, 1150 hPa.

    On the Pt channel: 0, 5, 10, 25, 50, 100, 200, 300, 400, 500, 600, 700, 800,

    900, 1000 hPa.

    Note #1: The user can enter up to 40 pressure values per channel on the whole

    scale. These data must be entered starting from the minimum pressure upwards to

    the maximal pressure.

    When the data has been successfully transferred to the computer, a message

    "Calibration OK" appears in the information window. If the transfer did not happen

    properly, a message "Calibration KO" appears. The data needs to be loaded again.

    Click on the button "Load data". To validate the data transfer, click on "Validate"

    otherwise click on "Cancel".

    November 2014 User Manual ED 01 Page 59 of 82

  • Pitot Static Tester ADSE650

    The data transfer procedure resumes.

    Note #2: If a calibration validation is requested but less than two calibration points

    have been validated, a control prevents this validation and an information window is

    displayed.

    November 2014 User Manual ED 01 Page 60 of 82

  • Pitot Static Tester ADSE650

    3.2.6 General Information

    Exhibit 41

    By clicking on the "Config" button of the main menu, following information isdisplayed:

    ● "Main Menu" brings back to the main menu,

    ● "Help" opens an Help file for the user,

    ● "Parameters" gives access to the modification of:

    ● the password requested to exit the remote control unit software(access to the underlying operating system), to access the calibrationmode, to write and to modify a program and a flight domain,

    ● the RS232 parameters (if needed),

    ● the wifi parameter,

    ● "Languages" allows to change the language of the software.

    Note 1: On the bottom right, you can find the serial number of the ADSE650on wich the TabletPC must communicate, the revision software of the EPU Boot and Application.

    Note 2: On the top left, you have informations about TabletPC Battery.

    November 2014 User Manual ED 01 Page 61 of 82

  • Pitot Static Tester ADSE650

    3.2.6.1 Parameters

    Exhibit 42

    User Settings:

    ● "Password": To modify password,

    ● “Autotest threshold...”: to configure the threshold for the leak during theAutotest

    ● "Display Unit Windows...": to configure the time in seconds to open the Unitwindows in the application,

    RS232:

    ● If you use the tool kit to communicate in RS232, the Com Port is indicate.

    WIFI:

    ● SN ADSE650 : Enter the serial number of the ADSE.

    Configuration of the wifi network:

    ● IP Address : By default: 201.197.57.1

    ● Subnet Mask : By default: 255.255.255.252

    ● Default Gateway : By default: 210.197.57.1

    ● IP Port: 2000

    November 2014 User Manual ED 01 Page 62 of 82

  • Pitot Static Tester ADSE650

    3.2.6.2 Password

    By default, when the equipment is delivered, the password is the serial number of

    the equipment.

    By clicking on the "Config" button of the main menu and "Parameters", following

    information is displayed:

    Exhibit 43

    Exhibit 44

    November 2014 User Manual ED 01 Page 63 of 82

  • Pitot Static Tester ADSE650

    Exhibit 45

    Exhibit 46

    November 2014 User Manual ED 01 Page 64 of 82

  • Pitot Static Tester ADSE650

    ● Select "Calibration" to change the password of calibration Mode or "By

    default" to change the other password (To write or modify program,

    domain, to exit).

    ● Enter the current password by clicking in the field and using a alpha

    numeric keyboard,

    ● If the password entered is wrong, a message appears: "Incorrect

    password",

    ● If the password is correct, a window appears, where the user is requested

    to enter twice the new password and validate (Exhibit 45 and 46),

    ● If both passwords are different, an error message "The passwords are

    different" is displayed,

    ● If the passwords are identical, a window appears, requesting to validate

    the change (Exhibit 47),

    ● If the user clicks on "Return" without validating, the change is not taken

    into account.

    November 2014 User Manual ED 01 Page 65 of 82

  • Pitot Static Tester ADSE650

    3.2.6.3 Enter a password

    Exhibit 47

    When a password is requested:

    ● click on the field,

    ● enter the current password by using the alphanumeric keyboard,

    ● Validate to continue or Cancel to come back to the main menu.

    November 2014 User Manual ED 01 Page 66 of 82

  • Pitot Static Tester ADSE650

    3.2.6.4 Languages

    Exhibit 48

    To change Languages:

    ● Select the desired Languages,

    ● Press "Validate" to change.

    November 2014 User Manual ED 01 Page 67 of 82

  • Pitot Static Tester ADSE650

    3.2.7 Help on lineThe "HELP" button on the main menu open a html windows with the user manual.

    3.2.8 Exiting the software• To exit to the software and come back to Windows 8, click on "Exit" in the

    main menu, click “shutdown Software” and enter a password.

    • To exit to the software and turn off the TabletPC, click on "Exit" in the main

    menu and click “shutdown Windows” and enter a password.

    November 2014 User Manual ED 01 Page 68 of 82

  • Pitot Static Tester ADSE650

    3.2.9 Keyboard configurationTo enter a numeric or an alpha-numeric value, 3 kinds of keyboards appear.

    3.2.9.1 Digital keyboard

    To enter a digital value, this screen is displayed:

    Exhibit 49

    ● Enter a value and validate by "Enter" or escape with hitting the "Abort"

    key.

    November 2014 User Manual ED 01 Page 69 of 82

  • Pitot Static Tester ADSE650

    3.2.9.2 Alpha-Numeric Keyboard

    To enter an alphanumeric value, this screen displayed:

    Exhibit 50

    ● Enter an alphanumeric value and validate by "Enter" or escape by"Escape"

    Note: For password, Text is hidden as follows:

    Exhibit 51

    November 2014 User Manual ED 01 Page 70 of 82

  • Pitot Static Tester ADSE650

    4. MAINTENANCE

    4.1 Passwords

    The initial passwords (upon delivery of the equipment) needed to access some

    secured menus are the serial number of the equipment.

    4.2 Internal

    Your ADSE 650 is engineered to give you full satisfaction for a large number of

    years. As it contains some components potentially subject to ageing (filters,

    membranes...) we recommend to follow the maintenance plan (in attachment), in

    order to ensure a perfect operation on the long run.

    This maintenance plan is valid for a use in normal conditions of temperature, dust,

    moisture and salt, with only occasional use in very harsh environment. In the case

    of a permanent use in harsh conditions (equipment embarked permanently on an

    aircraft carrier, on an airbase in the desert...) we recommend to divide by two the

    time between two maintenances.

    November 2014 User Manual ED 01 Page 71 of 82

  • Pitot Static Tester ADSE650

    4.3 Low Level Maintenance

    Housing: Clean every month with soft sponge (water and mild liquid detergent)

    Front panels and touch screen: Clean every week with damp lint-free cloth and

    mild liquid detergent.

    Note: NEVER USE AGRESSIVE DETERGENT OR SOLVENT

    Electric Cable: Check the cable and connector before each use.

    Pneumatic hose: Check hose and plug before each use.

    Internal Leak: Check every month the internal leak of the ADSE.

    November 2014 User Manual ED 01 Page 72 of 82

  • Pitot Static Tester ADSE650

    4.4 ContactsDo not hesitate to contact us, should you have any questions regarding the use and the maintenance of your equipment

    4.4.1Mailing address

    ATEQ15, rue des dames78340 Les Clayes-sous-BoisFrance

    4.4.2Internetwww.ateq.com

    4.4.3Telephone numbers

    Switchboard: +33 1 30 80 10 90

    Telefax +33 1 30 54 11 00

    November 2014 User Manual ED 01 Page 73 of 82

    http://www.ateq-omicron.com/

  • Pitot Static Tester ADSE650

    4.5 Maintenance

    4.5.1AutotestAfter shipping or in case of doubt on the bench, the "Autotest" function checks the

    ADSE (limits and leaks).

    This test, available in "program mode", runs automatically. At the end of the test,

    one window "Results" is displayed.

    At the end of the test, the box is green if the test passed, red otherwise.

    Two cases are possible:

    ● During the test, the ADSE has detected a problem with the generation. In this

    case, the test is stopped and a message corresponding to the problem is

    displayed at the bottom of the screen.

    ● One or more tests leaks are incorrect. Values implied appear in red in the

    results table.

    CAUTION:

    To run the Autotest:

    ● disconnect any device (risk of damage),

    ● connect tubes on Ps and Pt and put the plugs accessories to

    close them.

    November 2014 User Manual ED 01 Page 74 of 82

  • Pitot Static Tester ADSE650

    To Run Autotest :

    In the menu "Automatic Program", press the Autotest button at the bottom left ofthe screen.

    Exhibit 52

    A window appears, requesting confirmation of the launch and giving instructions for the ADSE.

    Exhibit 53

    November 2014 User Manual ED 01 Page 75 of 82

  • Pitot Static Tester ADSE650

    All steps appear and Autotest starts automatically.

    Exhibit 54

    The active step is flashing in green during the test. At any time, the user can:

    ● see the graphs of measure for all channels by clicking on the button

    "Windows".

    ● stop the Autotest by pressing the "STOP" button. A "Ground" button replaces

    it and allows the return to atmospheric pressure on all channels, if applicable.

    November 2014 User Manual ED 01 Page 76 of 82

  • Pitot Static Tester ADSE650

    Examples of measures :

    Exhibit 55

    Exhibit 56

    Note: To return to the Autotest menu press again the "Windows" button.

    November 2014 User Manual ED 01 Page 77 of 82

  • Pitot Static Tester ADSE650

    At the end of the Autotest, the result is given by a LED:● Green: The bench is functional,● Red: leak problem or malfunction.

    Note: If the test is successful (result: green) all channels are at atmosphericpressure at the end.

    Exhibit 57

    If it is a leak problem, the leak shows up in red.

    Exhibit 58

    November 2014 User Manual ED 01 Page 78 of 82

  • Pitot Static Tester ADSE650

    If a generation problem is concerned, a message indicating the problem shows up atthe bottom of the screen.

    Exhibit 59

    Note: If the test is stopped (result: red). The operator can, with the "Ground"button, go to atmospheric pressure for all channels.

    November 2014 User Manual ED 01 Page 79 of 82

  • Pitot Static Tester ADSE650

    4.5.2Maintenance Plan

    November 2014 User Manual ED 01 Page 80 of 82

    X XX XX XX XX XX XX XX XX XX XX XT TX XX XP PX XB BB BM M

    Notes: X: pour tous les ADSE / f or all ADSE equipment

    B: pour les ADSE av ec option pack batterie/ f or ADSE with battery pack option only

    M: pour les ADSE av ec option multi-v oies / f or ADSE with multiple outlet option only

    P: pour ADSE av ec pompes internes / f or ADSE with internal pumps only

    T: pour ADSE avec écrans tactiles / f or ADSE with touch screen

    CalibrationCalibrationChangement des f iltres de v entilationFan f ilters exchangeChangement f iltres internesInternal f ilters exchangeVérif ication alimentation interneInternal v oltage checkVérif ication pneumatique interneInternal pneumatic checkChangement des pièces en caoutchoucRubber part exchangeTest des électrov annesElectrov alv es testTest cartes électroniquesElectronic board testTest connecteurs pneumatiquesPneumatic plug testTest connecteurs électriquesElectric plug testRéglage v anne modulanteRegulation v alv e adjustmentRemplacement dalle tactileTouch screen exchangeRénov ation des électrov annesElectrov alv es renov ationRemplacement connecteurs pneumatiquesPneumatic plug exchangeTest groupe de pompagePump group testChangement v entilateursFan exchangeTest des batteriesBattery testChangement des batteriesBattery pack exchangeRénov ation des électrov annes 3Ps 3PtElectrov alv es renov ation 3Ps 3Pt

  • Pitot Static Tester ADSE650

    5. ANNEXE : Safety information CAUTION: This unit contains an internal Lithium Polymer battery. The battery can burst or explode, releasing hazardous chemicals. To reduce the risk of fire or burns, do not disassemble, crush, pierce or dispose of the battery or the instrument in fire or water, do not short circuit or short the contacts with a metalobject.

    5.1 Operating environment

    • Remember to follow any special current regulations any area, and always switch off your device when its use is prohibited or when it may cause interference or danger.

    • Allow ventilation around the device when using it. Do not cover it that will reduce cooling.

    • Connect the charger to a proper power source. The voltage requirements are found on the product case and/or packaging.

    • Unplug it before attempting to service or clean it.

    • NEVER charge a battery pack immediately after use and while still hot. Leaveit cool down to ambient temperature.

    • If the battery is fully discharge, a full charge must be done between 1 and 5 Hours. In other cases thanks to call the service after sales.

    • If the battery is fully charge, the time of using in normal operation must be between 1 and 3 hours. In other cases thanks to call the service after sales.

    5.2 Safety for Lithium Polymer battery use

    If you see some smoke or some liquid out of the device, stop it immediately. Disconnect the device and place it in an isolated area for at least 15 minutes. DO NOT USE THE DEVICE AGAIN.

    In the unlikely event that the Lithium Polymer battery will ignite, DO NOT use waterto extinguish the fire, take some sand or fire extinguisher described above.

    Do not pierce, cut or NEVER compress the jacket of a Lithium Polymer battery. If the battery is swelling up or if the cover is damaged, DO NOT USE AGAIN. Replacewith a new one.

    Do not dispose of Lithium Polymer batteries to the dustbin.

    November 2014 User Manual ED 01 Page 81 of 82

  • Pitot Static Tester ADSE650

    A Lithium Polymer battery should be stored with a minimum charge of 30%. If you store completely discharged, it will quickly become unusable.

    The ATEQ Company disclaims any responsibility for damage sustained in case of non compliance with these safety instructions.

    The ATEQ Company couldn't control the proper use of the battery for each customer (charge, discharge, storage etc.), It can not be held responsible for damage to persons and property.

    5.3 Important instructions (for service personnel only)

    CAUTION: Risk of explosion if battery is replaced by an incorrect type. Dispose of used batteries according to the instructions.

    Replace only with the same or equivalent type recommended by the manufacturer ATEQ-OMICRON.

    Charge the Lithium Polymer battery ONLY with the charger provided.

    NEVER use a Ni-MH (Nickel Metal Hydride) type battery charger to charge a LithiumPolymer battery.

    The battery must be recycled or disposed of properly.

    5.4 Recycling

    Do not dispose of the rechargeable Lithium-Ion battery or the device tothe dustbin.

    These components must be collected and recycled.

    The crossed-out wheeled dustbin means that within the EU the product must be taken to separate collection at the product end-of life. This applies to your tool but also to any enhancements marked with this symbol. Do not dispose of these products as unsorted municipal waste. For further information, please contact ATEQ.

    November 2014 User Manual ED 01 Page 82 of 82

    1. PRESENTATION1.1 General1.2 Applications1.3 Manufacturer’s part numbers1.4 Characteristics1.5 Units, Range and Precision from 10°C to 40°C1.5.1 Static pressure - Ps1.5.2 Total and differential pressures - Pt - ΔP1.5.3 Ps dynamic variations1.5.4 Pt dynamic variations1.5.5 Rates of intrinsic leak

    1.6 Technical description1.6.1 Electro-pneumatic Unit1.6.1.1 Front Panel1.6.1.2 Electro-pneumatic block1.6.1.3 Pressure transducers1.6.1.4 Proportionnal Electrovalves

    1.6.2 Remote Control Unit1.6.3 Software1.6.3.1 Language1.6.3.2 Pressure control1.6.3.3 Automatic Test programs1.6.3.4 Flight Domain1.6.3.5 Self test

    2. USER ACCESS2.1 ADSE650 RCU front panel2.2 ADSE650 EPU front panel

    3. USE3.1 Initial Procedure3.1.1 ADSE650 alternative power feeding3.1.2 Remote control link3.1.3 Pneumatic link3.1.4 Preparing for use3.1.5 Powering up3.1.6 Interface software3.1.7 Shutting off

    3.2 Using the operator interface3.2.1 Starting software3.2.1.1 Connection established3.2.1.2 Connection not established

    3.2.2 Access to the flight domains3.2.2.1 Select a flight domain.3.2.2.2 Creating a flight domain3.2.2.3 Editing, Renaming and deleting a flight domain

    3.2.3 Access to the Measure/Manual Mode3.2.3.1 Changing the unit on the Ps and Pt3.2.3.2 Activating a channel3.2.3.3 Generating on the Ps or Pt channel3.2.3.4 Leak test3.2.3.5 Return to the ground3.2.3.6 Displaying the current flight domain

    3.2.4 Automatic program mode3.2.4.1 Writing a new program3.2.4.2 Editing, Renaming and deleting a program file3.2.4.3 Executing a program3.2.4.3.1 Operating Mode3.2.4.3.2 File format for the data recorded by the semi-automatic program

    3.2.5 Calibration mode3.2.5.1 Calibrating the sensors3.2.5.2 Operating mode

    3.2.6 General Information3.2.6.1 Parameters3.2.6.2 Password3.2.6.3 Enter a password3.2.6.4 Languages

    3.2.7 Help on line3.2.8 Exiting the software3.2.9 Keyboard configuration3.2.9.1 Digital keyboard3.2.9.2 Alpha-Numeric Keyboard

    4. MAINTENANCE4.1 Passwords4.2 Internal4.3 Low Level Maintenance4.4 Contacts4.4.1 Mailing address4.4.2 Internet4.4.3 Telephone numbers

    4.5 Maintenance4.5.1 Autotest4.5.2 Maintenance Plan

    5. ANNEXE : Safety information5.1 Operating environment5.2 Safety for Lithium Polymer battery use5.3 Important instructions (for service personnel only)5.4 Recycling