#17017 Biopsy Channel Adapter (WHITE TUBING) Adapting to Pentax Small Caliber Endoscopes Groups A and B type instruments ONLY as described below Use Pentax Suction Channel Cleaning Cap (Pentax OF-B155) #17017 Balloon Channel Adapter Male luer to Pentax OE-B5 tube (WHITE TUBING) Group A Group B Refer to Scope Adaptation Charts for the specific endoscope model and recommended adapters: Group A = Has a permanent Suction Connector & a luer Biopsy Channel port. Uses a Pentax (Suction Valve Cylinder) adapter cap (Ex. EB-30/70/70K-series, FB-V/X/P-series, Fl-P/BS-series, FNL/VNL) CU adapters = 17002 (Suction Connector) and 17017 (Biopsy port) Group B = This scope also has a water injection/balloon channel w/unique connector with threads to secure a Pentax OE-B5 injection tube. (Ex. EBUS – EB-1970UK) CU adapters = 17002, 17017 #17017 Biopsy Channel Adapter (WHITE TUBING) Use Pentax Suction Channel Cleaning Cap (Pentax OF-B155) The Pentax Connection Diagram is for reference purposes only – the illustration shown may not represent a specific model instrument. CAUTION: Remove all valves (suction) and biopsy caps, selector mechanism, etc.) consult scope manufacturer manual for proper removal and cleaning procedures of components. Install certain reprocessing parts supplied by Pentax and all adapters supplied at time of In-Service. To confirm proper hook up refer to In-Service Program Report Form and/or Scope Adaptation For All Model Endoscopes Chart. If you have any questions on adapting to your scopes or if you acquire any new scopes not listed on your In-Service Program/Report Form please contact Custom Ultrasonics. #17002 Suction Channel Adapter (GREEN TUBING) OE-B5 #17002 Suction Channel Adapter (GREEN TUBING) customultrasonics.com (215) 364-1477 #17017 Pentax(4), Rev 01/16/2015
1

Adapting to Pentax Small Caliber Endoscopes Group A · Adapting to Pentax Small Caliber Endoscopes Groups A and B type instruments ONLY as described below ... Male luer to Pentax

Apr 20, 2018

Download

Documents

vutu
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Adapting to Pentax Small Caliber Endoscopes Group A · Adapting to Pentax Small Caliber Endoscopes Groups A and B type instruments ONLY as described below ... Male luer to Pentax

#17017Biopsy Channel Adapter

(WHITE TUBING)

Adapting to Pentax Small Caliber EndoscopesGroups A and B type instruments ONLY as described below

Use Pentax Suction Channel Cleaning Cap

(Pentax OF-B155)

#17017 Balloon Channel Adapter

Male luer to Pentax OE-B5 tube(WHITE TUBING)

Group A

Group B

Refer to Scope Adaptation Charts for the specific endoscope model andrecommended adapters:

Group A = Has a permanent Suction Connector & a luer Biopsy Channel port. Uses a Pentax (Suction Valve Cylinder) adapter cap (Ex. EB-30/70/70K-series, FB-V/X/P-series, Fl-P/BS-series, FNL/VNL) CU adapters = 17002 (Suction Connector) and 17017 (Biopsy port)

Group B = This scope also has a water injection/balloon channel w/unique connector with threads to secure a Pentax OE-B5 injection tube. (Ex. EBUS – EB-1970UK) CU adapters = 17002, 17017

#17017 Biopsy Channel Adapter(WHITE TUBING)Use Pentax Suction Channel

Cleaning Cap(Pentax OF-B155)

The Pentax Connection Diagram is for reference purposes only – the illustration shown may not represent a specific model instrument.

CAUTION: Remove all valves (suction) and biopsy caps, selector mechanism, etc.) consult scope manufacturer manual for proper removal and cleaning procedures of components. Install certain reprocessing parts supplied by Pentax and all adapters supplied at time of In-Service. To confirm proper hook up refer to In-Service Program Report Form and/or Scope Adaptation For All Model Endoscopes Chart. If you have any questions on adapting to your scopes or if you acquire any new scopes not listed on your In-Service Program/Report Form please contact Custom Ultrasonics.

#17002 Suction Channel Adapter

(GREEN TUBING)

OE-B5

#17002 Suction Channel Adapter

(GREEN TUBING)

customultrasonics.com(215) 364-1477

#17017

Pentax(4), Rev 01/16/2015