Acceptance Testing And Commissioning Data Measurements For Photon Beam Of A Varian Linear Accelerator Performed By Md. Din Muhammad B.Sc.(Hons.) MPBME Gono Bishwabidyalay, Savar, Dhaka
17

Aceeptance testing and commisioning data measurement for photon beam

Apr 15, 2017

Download

Education

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Aceeptance testing and commisioning data measurement for photon beam

Acceptance Testing And Commissioning Data Measurements For Photon Beam Of A Varian

Linear Accelerator

Performed ByMd. Din Muhammad

B.Sc.(Hons.) MPBME

Gono Bishwabidyalay, Savar, Dhaka

Page 2: Aceeptance testing and commisioning data measurement for photon beam

1. Introduction 1.1 General introductionDuring accepting testing and commissioning medical

physicists are faced with many challenges, like a variety of scientific testing methods, setting for highest precision, data accuracy, etc.

Accepting testing procedure agreed on between the manufacturer’s representative and the facility physicist

Commissioning is a process to prepare machine for any clinical use and investigated, measured, and recorded. Commissioning of a treatment unit should only start after a thorough acceptance process has established its mechanical and radiation consistency.

Page 3: Aceeptance testing and commisioning data measurement for photon beam

1.2 Objectives of the studyThe objectives of this study are-To prepare necessary documents for acceptance

testing and commissioning.To perform acceptance testing and safety checksOutput Calibration of the machine for each

photon energy.To measure all relative data required to

configure AAA algorithm.To commission the machine for all designated

clinical usesTo establish treatment procedures and train

personnel

Page 4: Aceeptance testing and commisioning data measurement for photon beam

1.3 Aims & ScopesThe main aims of acceptance tests and

commissioning are: To assure all specifications are fulfilled. To calculate how much area is free of radiation. To protect the staffs from irradiation. To know how much accurate beam is produced To prepare the machine for clinical treatment To ensure the better quality of treatment

Page 5: Aceeptance testing and commisioning data measurement for photon beam

2.Materials and Methods2.1 Materials:There are some equipments

that used in my project work . These are:

Linear Accelerator(Varian 2300 C/D)

Mini Water Phantom (WP34)Cylindrical Ionization ChamberMonitoring DeviceGeiger Muller (GM) CountersElectrometer(IBA-DOSE-1)Plane-parallel ChamberFilm (X-Omat V)

Page 6: Aceeptance testing and commisioning data measurement for photon beam

2.2.1 Safety ChecksWarning lightsDoor Interlock. Patient monitoring equipmentEmergency ButtonsInterlocks

2.2 Acceptance TestsAcceptance tests procedures are divided into three groups:

Safety ChecksMechanical checksDosimetry measurements

Page 7: Aceeptance testing and commisioning data measurement for photon beam

2.2.2Radiation Survey:This test is followed by IEC 61217 protocol Materials: Geiger Muller (GM) Counters

2.2.3 Collimator Axis of RotationThis test is followed by IEC 61217 protocolMaterials: Front pointer Graph paper Microspore tape Pinpricks, etc.Method of Test:The graph paper sets under the front pointer tip rotate the collimator from 90° to 270° while observing the pointer run-out .Verify that the worst case run-out ≤1.0 mm.

Fig-Collimator Axis of Rotation

Page 8: Aceeptance testing and commisioning data measurement for photon beam

2.2.4 Gantry Axis of RotationThis test is followed by IEC 61217 and

IAEA TECDOC_1540 protocol Materials: Front pointerMicrospore tape Pinpricks, etc.Method of Test: Rotate the gantry through 360° angle and visually

checking the front pointer run-out after the rotation of every 90°.

Fig-Gantry Axis of Rotation

2.2 Acceptance Tests

Page 9: Aceeptance testing and commisioning data measurement for photon beam

2.2 Acceptance Tests2.2.5 Couch Axis of RotationThis test is followed by IEC 61217 and IAEA TECDOC_1540 protocol Materials: Front pointerGraph paperMicrospore tape Pinpricks, etc.

Method of Test:Slowly rotate the Couch from 90° to 270° while observing the

front pointer run-out every 45°Verify the worse case run-out is ≤1.0 mm radius for

counterweight systems and ≤2.0 mm for Beam-stopper systems.

Fig-Couch Axis of Rotation

Page 10: Aceeptance testing and commisioning data measurement for photon beam

2.2.6 Light Field and X-Ray FieldMethods:Set up a piece of film on the Couch with the light field visible on one half of the film sheet. Mark the edges of the light field and expose the film. Shift the film to expose the other half of the film using the same condition for X-ray field. After developing the film compare 50% intensity regions of X-ray edges to the marks representing the light field edges. The difference between the light field marks and the x-ray field edges should coincide within ±2.0 mm.

Fig: Light Field and X-Ray Field

2.2.7 Gantry Rotation Spoke ShotMethod of Test:Rotate gantry successfully 90 t0 185 for exposed filmBisect 1.0cm exposureMeasure the longest line of the trapezoid

Page 11: Aceeptance testing and commisioning data measurement for photon beam

2.2.9 ODI CalibrationMethod of Test:Turn on the Field Light and project the

crosshairs and ODI rangefinder display on a piece of white paper. Verify the ODI meets specification for the 100 cm position.

Drive the Couch 20 cm above and then 30 cm below the 100 cm reference position. Verify the ODI display meets specification.

Page 12: Aceeptance testing and commisioning data measurement for photon beam

2.2.10 Photon Energy Depth of IonizationMethod of Test: Beam on and scan the central axis depth of ionization for

both.Analyze the scans and verify each energy meets the

specifications.

2.2.11 Photon Field Flatness and SymmetryMethod of Test: Scan through the X-ray field central axis at a depth if 10 cm in the radial plane for each energy

at field sizes of 40×40 cm and 10×10 cm. Radial scans should always be performed prior to transverse scans in cause any steering adjustments are required.

Setup the Welhofer scanning system for CROSSPLANE scanning and repeat the scans in the transverse plane. For Buddle-ship tanks rotate the couch to 90° or 270°.

Page 13: Aceeptance testing and commisioning data measurement for photon beam

3 Result:PDD:

Beam Profile

Page 14: Aceeptance testing and commisioning data measurement for photon beam

Wedge Profile:

Page 15: Aceeptance testing and commisioning data measurement for photon beam

4. Discussion: The study helps to understand the important of

Acceptance tests and commissioning for linear accelerator.

The study gives an ideas to improve the work accurately.

The study helps the staffs to update their knowledge in this procedures

The study helps to measure the errors.

Page 16: Aceeptance testing and commisioning data measurement for photon beam

5. Conclusion:All measurements are done properly and

carefully.All data readings are include carefully.During Absolute and Reference Dosimetry all

equipments are setup accurately according to the protocol TRS-398.

I could not include some data like beam profiles for both photon energy at all depth ,wedge ,PDD etc.

All data are measured before treating the patients.