“A study of the electrical bio-impedance of tumors”, Morimoto (1993) Measurement System: Three-electrode method, frequency range 0-200kHz, a thin coaxial needle electrode, large plate (80mm x 160mm) Puls-response method Calculated: • extracellular resistance • intracellular fluid resistance • cell membrane capacitance Electrical impedance of varius tumors, in vivo. 54 patiens with breast disease. 57 patients with pulmonary disease. Needle Electrode
5
Embed
“A study of the electrical bio-impedance of tumors”, Morimoto (1993)
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
“A study of the electrical bio-impedance of tumors”, Morimoto (1993)
Measurement System:
Three-electrode method,
frequency range 0-200kHz,
a thin coaxial needle electrode,
large plate (80mm x 160mm)
Puls-response method
Calculated:• extracellular resistance
• intracellular fluid resistance
• cell membrane capacitance
Electrical impedance of varius tumors, in vivo. 54 patiens with breast disease. 57 patients with pulmonary disease.
Needle Electrode
“In vivo electrical conductivity of hepatic tumours”, Haemmerich, Staelin, Tsai, Tungjitkusolmun,Mahvi and J. G.Webster (2003)
Classical four-electrode plunge probe
In vivo hepatic tumour and normal tissue conductivity at seven frequencies from 10 Hz to 1 MHz, in rats.
Ag-AgCl electrodes
The proximal 2 mm of the probe was insulated to reduce errors caused by fluid buildup on the probe surface.
“Design considerations for optimum impedance probes with planar electrodes for bioimpedance measurements.”, Ivorra, A., Aguilo´ , J., Milla´n, J., (2001)
Planar electrodes- design electrodes as large as possible - use the same geometry for the inner electrodes- electrode-electrolyte interface impedance as low as possible.
Pt electrodes can be coated with a porous layer of “black platinum”
The electrode ring system on the tip of the non-invasive probe, The outermost electrode (1) is approximately 10 mm indiameter. The two outer electrodes [(1) and (2)] are sourceelectrodes. The innermost electrode (4) is a current sink.The electrode (3) surrounding the innermost electrode is a guard electrode, which reduces surface currents.
Microinvasive probe, voltage is applied at the beams marked (a) and (b), and the current is detected in beam (c)
Area of the electrode is 5 x 5 mm2.
”Micromachined electrodes for biopotential measurements” Griss,Enoksson,Tolvanen-Laakso,Meriläinen,Ollmar (2001)
Concentric ring electrode, non-invasive probe
Electrode array
“Electrical impedance scanning: a new approach to skin cancer diagnosis”, Glickman (2003)
GN= average of the normalized conductivity
Detecting probe: matrix of 8x8 pointed, gold-coated electrodes (4mm x 4 mm)
Voltage source: a hand held cylinder (metallic clip)
Trans-impedance measurement technique: while the evoked current under a particular electrode was recorded, the other electrodes were kept at ground potenttial.
Results show that Electrical impedance spectroscopy (EIS) is able to repeatedly detect small tumors (<3 mm) and tumorassociated changes, whereas CT and US were not routinely capable of detecting pathological developments on this scale.