Top Banner
Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D., R&D PeteStone, Ph.D., Marketing
14

A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Jul 20, 2018

Download

Documents

vuongdang
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Page 1

A New Era in High Speed Analysis with Dynamic MRM and UHPLC

Tim Schlabach, Marketing

Anabel Fandino, Ph.D., R&D

PeteStone, Ph.D., Marketing

Page 2: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

ASMS TrainingPage 2

Page 2

Agilent LC/MS, GC/MS, and ICP-MS Products

Page 3: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

ASMS TrainingPage 3

Page 3

Agilent LC/MS, GC/MS, and ICP-MS Products

Page 4: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Agilent 1290 Infinity LC Attributes for MS - Infinitely Better for LC/MS

Page 4

1200bar @ 2mL/min for highest resolutionReduced Ion & Matrix Suppression

HT-Solution for up to 2,000 samples/dayComplete System Integration MassHunter

Less than 100 uL delay volumesFirst peaks can elute in < 6 seconds

High Pressures

LowDelay Vol.

High Throughput

Page 5: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Agilent 6000 Series LC/MS: Speed for UHPLC

Page 5

Fast MRMsfor UHPLC

40 MS Scans/sec10 MS/MS scans/sec

BestData

QualityFast Scans for UHPLC

Agilent 6460

Agilent 6530

1 ms dwell for MRM30 ms Pos/Neg Switching> 200 MRMs/ sec

Page 6: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

1290 Infinity LC means fast peaks, < 3 seconds

Page 6

Page 7: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Fast MRMs with 6400 Series QQQ – metabolism polarity switching in 30 millisecs plus 6 transitions (2 ISTDs)

Avg W1/2 = 1.3 sec 35 points across RSD [%] = 1.8

Avg W1/2 = 0.71 sec 15 points across RSD [%] = 2.9

Avg W1/2 = 1.3 sec 35 points across RSD [%] = 2.7

Avg W1/2 = 0.52 sec 11 points across RSD [%] = 6.5

Verapamil, 0.75 min

Buspirone, 0.40 min

Dextromethorphan, 0.53 min

Diclofenac, 1.04 min

Page 8: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Peaks widths with UHPLC only 2 seconds or less > ten uniform measurements across peak REQUIRED

Page 8

W = 2.1 sec

Cycle time < 115 ms

Avg. W1/2 = 0.72 sec

Verapamil, peak RSD 5%

Page 9: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Page 9

Peak Capacities are very high with UHPLC 40 Extracted Ion Transitions in 1 minute

Page 10: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Page 10

Dynamic MRM for 6400 Series Triple Quads monitors ion transitions only when compounds elute

With DMRM the number of concurrent MRMs are fewer than with Time Segments

Cycle Time:Concurrent MRMsDwell Time:

Page 11: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Page 11

Agilent 1290 Infinity LC + 6460 Triple Quad

9 replicates analyzed at 200 pg levelMean area %RSD = 3.2

Analysis of 600 Pesticide DMRM Transitions in 15 min.

Page 12: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

17 min.

1290/6230 Accurate Mass TOF Pesticide Screening250 pesticides, 50 pg on column

3 min.

1.5 min.

Most peaks 1-2 seconds wide!

Page 13: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

The Need for Speed in UHPLC/MS/MS

Cycle Times (MRM): <150 milliseconds

Dwell Times (MRM or DMRM): <= 2 millliseconds

Polarity Switching (MRM): < 50 milliseconds

Spectral scans: at 5 Hz or faster

PPM Mass Accuracy: at 5 Hz

20 K Resolution: at 5 Hz

Page 13

Page 14: A New Era in High Speed Analysis with Dynamic MRM and … New... · Page 1 A New Era in High Speed Analysis with Dynamic MRM and UHPLC Tim Schlabach, Marketing Anabel Fandino, Ph.D.,

Page 14

Agilent LC/MS can meet your most demandingapplication needs.

1290 Infinity LC and 6460 Triple Quad MS

1290 Infinity LC and 6530 Accurate Mass Q-TOF MS

The ultimate Qual solution

The ultimate Quant solution