A case of inappropriate ICD shock causing cardiac arrest

Dr Darragh Moran, Dr Niall Mahon

Page 2: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• 2003 ED MMUH• 16 year old male• Chest discomfort• Abnormal ECG• Normal coronaries

Page 3: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 4: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• IVDd of 22mm• No LVOT gradient noted • Satisfactory TMET• Rx betablockade• Normal 24 holter• Episodic left-sided chest discomfort persisted• No family history of SCD

Page 5: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• May 2007, aged 20• ED MMUH – collapse following game of soccer

Page 6: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 7: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 8: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 9: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• No arrhythmia seen during inpatient monitoring

• TMET – drop in blood pressure at peak exertion

Page 10: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

ACC/ESC expert consensus HCM

Maron BJ, McKenna WJ, Danielson GK, Kappenberger LJ, Kuhn HJ, Seidman CE, et al. American College of Cardiology/European Society of Cardiology clinical expert consensus document on hypertrophic cardiomyopathy. A report of the American College of Cardiology Foundation Task Force on clinical expert consensus documents and the European Society of Cardiology Committee for practice guidelines. J Am Coll Cardiol 2003;42:1687–713.

Page 11: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• 18 days later – ED MMUH• OHCA whilst cycling• Prolonged CPR, external DCCV for VF at scene• GCS 3, intubation and ventilation with inotropic

support• GCS 11 day 2, GCS 15 day 4• Aspiration pnemonia• Post-traumatic amnesia, poor concentration,

attention deficit

Page 12: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 13: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• RV lead repositioning• DFT testing – failed 25J and 35J shocks

requiring 200J external rescue shock

Page 14: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 15: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 16: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• Defibrillation testing was repeated 1 week later. No further T wave oversensing was detected during this interrogation. VF was appropriately sensed by the ICD, both a 20J and 25J shock were unsuccesful, with a 30J rescue shock being required to restore sinus rhythm. The patient was discharged home in the days that followed.

Page 17: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

2/12 later, patient returned to hospital as SC array had dislodged whilst swimming

Page 18: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 19: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• 2 years of uneventful follow-up• 2009 – threatened pocket erosion of device• Deep submuscular repositioning• Device check April 2010, high impedance

values, >200J, sensitivity of 0.6mV• Admitted for revision and testing of leads and

epicardial patch

Page 20: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 21: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• 3/7 later patient returned for DFT testing• 1st 35J shock was unsuccessful, 2nd successfully

restored SR• Long discussion with patient regarding risks

and benefits of further intervention• Discharged home, albeit with unsatisfactory

DFT

Page 22: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• Few months later…• Pain and swelling around device insertion site• Chronic pocket infection• Device explanted

Page 23: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.
Page 24: A case of inappropriate ICD shock causing cardiac arrest Dr Darragh Moran, Dr Niall Mahon.

• 7 years of device complications• No ventricular events in that time• High risk of SCD• SC implantable defibrillator offered• AEDs located at workplace and also at the

gym patient attends to reduce risk of SCD• No further symptoms/adverse events to date