The Scientific World JournalVolume 2012, Article ID 439875, 7
pagesdoi:10.1100/2012/439875
The cientificWorldJOURNAL
Research Article
Stabilizing Agents for Calibration in theDetermination of
Mercury Using Solid SamplingElectrothermal Atomic Absorption
Spectrometry
Hana Zelinkova, Rostislav Cervenka, and Josef Komarek
Department of Chemistry, Faculty of Science, Masaryk University,
Kotlarska 2, 61137 Brno, Czech Republic
Correspondence should be addressed to Josef Komarek,
[email protected]
Received 15 November 2011; Accepted 9 January 2012
Academic Editors: K. Kannan, R. G. Wuilloud, and M. C.
Yebra-Biurrun
Copyright 2012 Hana Zelinkova et al. This is an open access
article distributed under the Creative Commons AttributionLicense,
which permits unrestricted use, distribution, and reproduction in
any medium, provided the original work is properlycited.
Tetramethylene dithiocarbamate (TMDTC), diethyldithiocarbamate
(DEDTC), and thiourea were investigated as stabilizing agentsfor
calibration purposes in the determination of mercury using solid
sampling electrothermal atomic absorption spectrometry(SS-ETAAS).
These agents were used for complexation of mercury in calibration
solutions and its thermal stabilization ina solid sampling
platform. The calibration solutions had the form of methyl isobutyl
ketone (MIBK) extracts or MIBK-methanolsolutions with the TMDTC and
DEDTC chelates and aqueous solutions with thiourea complexes. The
best results were obtainedfor MIBK-methanol solutions in the
presence of 2.5 gL1 TMDTC. The surface of graphite platforms for
solid sampling wasmodified with palladium or rhenium by using
electrodeposition from a drop of solutions. The Re modifier is
preferable due toa higher lifetime of platform coating. A new
SS-ETAAS procedure using the direct sampling of solid samples into
a platform withan Re modified graphite surface and the calibration
against MIBK-methanol solutions in the presence of TMDTC is
proposed forthe determination of mercury content in solid
environmental samples, such as soil and plants.
1. Introduction
Mercury and its compounds belong among the most
toxiccontaminants and have the ability to bioaccumulate. Themain
sources of mercury are volcanic activity, combustionof coal, and
other human activities, through which mercuryis released into
water, soil, and sediments, whereby it entersinto the food chain
and causes health damages. Hence, thestudy of mercury content in
environmental samples is veryimportant [13].
For direct analysis of solid samples over the past years,solid
sampling electrothermal atomic absorption spectro-metry (SS-ETAAS)
has been used. The solid samples areweighed on a graphite platform,
which is inserted intoa graphite tube. The advantages of this
method are theuse of a very small amount of sample and little
samplepretreatment. The precision and accuracy of the results
depend on the weighing process, distribution of particlesin the
sample, and its homogeneity. The disadvantages areincreases in
interferences and calibration technique [1, 2, 410].
In SS-ETAAS, the interferences, kinetic of atomization,shape of
the signal, and sensitivity depend on the amountof the sample, the
form of the analyte, and the matrixcomposition. If the properties
of the sample and standardsfor calibration are dierent, an error
can occur. Therefore,the right calibration technique is very
important. The firstmethod is the application of solid standards,
such as certifiedreference materials with properties similar to the
analyzedsample. The reference material is weighed on the
graphiteplatform in various amounts, and every point of the
cali-bration curve corresponds to one weight and measurement[11].
Vale et al. found that a higher amount of the sample hasa
depressive influence on the signal and distorts the
-
2 The Scientific World Journal
calibrationcurve [5]. Maia et al. used five dierent
referencematerials and constructed the calibration curve as
depen-dence of the normalized absorbance on the certified
mercurycontent [11]. The disadvantage of calibration with
solidstandards is their low availability, high cost, limited
con-centration range, and limited possibility to prepare
artificialsamples [5, 6, 11]. If the matrix components interfere,
thestandard additionmethodmay be used. Thismethod is basedon the
assumption that a change in response for the sampleand the sample
with an addition of standard correspondsonly to the change of the
concentration. For solid samplestwo techniques can be used: the
addition of an aqueous stan-dard or the addition of reference
material to the solid sample.A disadvantage of this method is that
it is impossible toensure the constant sample mass [7].
Another technique is calibration against aqueous stan-dards. The
main problem with the determination of mer-cury in a solution by
ETAAS is the high volatility ofthe element and its compounds.
Therefore, additions ofthermal stabilizing agents are applied to
avoid losses ofmercury. Because inorganic mercury compounds are
lessvolatile than the element itself, various oxidizing agentssuch
as hydrogen peroxide, permanganate, or dichromatewere used to
prevent their reduction [1, 3, 8, 12, 13].Reagents containing
sulphur as dithizone, diethyldithio-carbamate (DEDTC), or
tetramethylene dithiocarbamate(TMDTC) stabilize mercury by the
formation of complexand subsequently mercury sulphide [1, 1315]. A
successfulapproach used to stabilize mercury is the application
ofmodifiers to the graphite atomizer surface. Gold,
platinum,palladium, rhodium, and iridium or their mixtures
wereinvestigated. Palladium is applied most frequently.
Themodifiers can be deposited onto an atomizer surface bythe
thermal or electrochemical method [1, 1013]. In SS-ETAAS,
calibration against aqueous standards was applied,utilizing
oxidizing agents and modifiers of the graphiteatomizer surface [1,
10]. In [1], a loss-free determinationof mercury in aqueous
calibration solutions was reachedonly through the addition of
potassium permanganate andby using Pd, thermally deposited on the
SS platform. Thisprocedure was satisfactory for mercury
determination in ash,sludge, and sediment reference materials. In
our previouswork [10], permanganate was used together with a
Pdmodifier, electrochemically deposited on the SS platform.However,
the use of permanganate has some disadvantages.For technical
reasons the dosing of only 3 L of KMnO4solution onto the SS
platform with concentration >10 g L1
is possible. By the injection of a volume >3 L, a drop
ofsolution with great viscosity is superimposed on the innerspace
of the SS platform and the insertion of the SS platforminto to the
graphite tube without any spills, using tweezers,is impossible. For
the optimal total amount of KMnO4(0.3mg), a concentration of 100 g
L1 is required for 3 L ofthe solution. Moreover the preparation of
such a solution ofpermanganate is dicult [10].
Therefore, the aim of this work was to select anothersuitable
stabilizing agent for calibration solutions. For thispurpose, the
influence of thiourea, tetramethylene dithio-carbamate, and
diethyldithiocarbamate was studied. In our
Table 1: Temperature program for the determination of
mercury.
Stage Temperature (C) Ramp (C s1) Hold (s)
Drying 90a, 120b 30 15
Pyrolysis 200 30 40
AZc 200 0 6
Atomizationd 1100 1500 10
Cleaning 1700 200 4aDrying temperature for aqueous solutions,
bdrying temperature for MIBKcalibration solutions in the presence
of TMDTC and DEDTC, cauto zero,dgas stop.
previous work [10], the electrodeposition from a drop ofa
modifier solution proved to be suitable method of prepa-ring the Pd
surface for the determination of mercury bySS-ETAAS. Because
palladium has a relatively low boilingpoint, another metal for
coating the SS platform was tested.On the basis of our previous
results by the determinationof gold with the complete
electrochemical coating of thegraphite tube surface, rhenium was
chosen [16]. Palladium,and newly, rhenium were used as modifiers of
the graphiteplatform surface for the determination of mercury in
solidenvironmental samples as soil and plant.
2. Experimental
2.1. Instrumentation. A ZEEnit 650 atomic absorption
spec-trometer (Analytik Jena, Germany) with a transverselyheated
graphite tube and a solid sampling system SSA61Z was used for all
measurements. The spectrometer wasequipped with a Zeeman-based and
deuterium backgroundcorrector. The magnetic field of an
electromagnet wasapplied to the graphite atomizer by the
2-fieldmode. Zeemancorrections were used throughout the work, and a
deuteriumdevice was used only in special cases. A mercury
hollowcathode lamp at current 4.5mA was used as the
radiationsource. Measurements were performed in the peak areamode
at 253.7 nm using a spectral bandwidth of 0.5 nm.Calibration
solutions were applied manually onto an SSgraphite platform
(Analytik Jena, Part no. 407-152.023) andintroduced into the
graphite tubes without a dosing hole(Analytik Jena, Part no.
407-152.316) in the same way as thesolid samples. The calculated
integrated absorbance per mgof the sample is introduced as the
normalized absorbance.The temperature program for the determination
of mercuryis presented in Table 1.
For comparison purposes, the mercury content in envi-ronmental
materials was also determined using the AMA254 analyzer (Altec,
Czech Republic). The measurement inthis single-purpose atomic
absorption spectrometer is basedon the combustion of a sample in a
flow of oxygen andthe subsequent capture of mercury by a gold
amalgamator.After thermal release from amalgamator, the mercury
vapouris measured. This pyrolysis approach in AAS is frequentlyused
in analysis of environmental and biological materials,for example,
marine sediments, soil, citrus and tomatoleaves [17]. Each time,
40100mg of a sample was weighed
-
The Scientific World Journal 3
mA
V
1
2
3 4
+
Figure 1: Schematic diagram for electrodeposition from a drop.
(1)Power supply, (2) SS platform: cathode, (3) Pt wire: anode, (4)
dropof modifier solution.
or 10200 L of solution was dosed in nickel boats. Thesolid
samples were dried at 120C for 60 s and decomposedat 650C for 150
s. The AMA 254 analyzer was regularlycalibrated using standard
solutions of 11000 gL1 ofmercury for the first (06 ng Hg) and
second (0200 ngHg) calibration intervals. The calibration solutions
wereprepared by diluting the stock standard solution with
0.05%(m/v) K2Cr2O7 and 0.6% HNO3 to improve their stability.The
accuracy of the results was controlled by analysis ofthe standard
reference material GBW 07405. The relativestandard deviation (RSD)
was 3.2% (at 0.29mgkg1 Hg,n = 10).
2.2. Chemicals and Solutions. Hg(II) solutions were preparedfrom
the stock standard solution for mercury (1.000 0.002 g L1 Hg,
Analytika, Czech Republic) in 2% HNO3by dilution with 5% (v/v)
HNO3. Thiourea p.a. (SigmaAldrich), ammonium tetramethylene
dithiocarbamate p.a.(TMDTC, Sigma Aldrich), acetic acid p.a.
(Fluka), sodiumacetate p.a. (Fluka), sodium diethyldithiocarbamate
p.a.,(DEDTC, Lachema), acetylacetone p.a. and methyl isobutylketone
p.a. (MIBK, Lachema) were used for the preparationof calibration
solutions. A stock solution of 110 g L1
KMnO4 (Merck) was prepared as in [10] with the supportof an
ultrasonic bath and added to the calibration solutionsfor a final
concentration of 100 g L1. PdCl2 (Merck, Darm-stadt, Germany) and
NH4ReO4 (Analytika, Czech Republic)standard solutions containing 10
g L1 Pd or Re were used tomodify the graphite platform surface.
2.3. Samples and Their Treatment. A Certified reference
ma-terial (CRM) soil GBW 07405 (National Centre for
StandardMaterials, Beijing, China) and the environmental samplesof
soil II and plant Scirpus from the Hg-polluted area wereused. The
environmental samples were ground in a millFritsch Pulverisette 7
with balls from Si3N4 and passedthrough a nylon sieve for a
particle size of 56 m. The
Table 2: Temperature program (according to study [1])
fortreatment of platform after the electrodeposition of modifiers
froma drop of solutions.
Stage Temperature (C) Ramp (C s1) Hold (s)
Drying 90 30 15
Pyrolysis 250 20 35
AZa 250 0 6
Atomization 1000 1000 10
Cleaning 2000 200 5aAuto zero.
aliquots of environmental samples between 0.1 and 0.5mgor CRM
GBW 07405 210mg were weighed directly ontothe SS platforms and
inserted into a graphite tube. Beforeeach weighing on the SS
platform, these ground sampleswere carefully stirred. The residues
of solid samples afteratomization were easily removed from the
platform.
2.4. Electrodeposition of Palladium and Rhenium from a Dropof
Solutions. The surface modifiers were applied to thegraphite
platform using 7 injections of 20 L of solutionof 2 g L1 Pd or Re.
The graphite platform with a dropof modifier solution served as the
cathode and a Pt wirewas used as the anode (Figure 1).
Electrodeposition of everydrop proceeded by the current 10mA for
5min. After eachdeposition, the surface of the SS platform was
rinsed withwater, dried, the SS platform was inserted into the
graphitetube, and the temperature program started according toTable
2. The amount of Pd or Re electrodeposited ontothe SS platform was
calculated from the dierence of itscontent in the solution before
and after electrolysis. Duringelectrodeposition 250 g Pd or Re was
deposited.
2.5. Preparation of Calibration Solutions of Hg(II) in
thePresence of TMDTC and DEDTC. Calibration solutions ofHg(II) in
the presence of TMDTC and DEDTC wereprepared using two methods:
(i) The extraction of mercury with chelating agents intoMIBK or
acetylacetone.1mL of mercury(II) solution, 2mL of 2.5 g L1
TMDTC or DEDTC aqueous solution and 1mL ofacetate buer (pH 5)
were pipetted into the extrac-tion tube. After shaking, 4mL
acetylacetone or MIBKwas added. The chelates were extracted into
theorganic phase on the shaker at a speed of 300 RPMfor 1 h. The
extraction eciency was checked bymeasuring the absorbance of the
aqueous phase.
(ii) The preparation ofMIBK-methanol solution from anaqueous
methanol solution refilled by MIBK.The calibration solutions of
Hg(II) in the presence ofTMDTC were prepared from 0.5mL of Hg(II)
aque-ous solution, 1mL of 25 g L1 TMDTC in methanoland 1mL of 1mol
L1 sodium acetate in methanol.After shaking, the solution was
diluted with MIBK to10mL to formation of single phase.
-
4 The Scientific World Journal
Table 3: Maximum pyrolysis temperatures for the determination
ofmercury in calibration solutions with the stabilizing agents.
Temperature/C
Agents Pd Re
Potassium permanganate 250a 280
Thiourea 280
TMDTC 200 200
DEDTC 200 230a[10].
3. Results and Discussion
3.1. Stability of Surface Modifiers. The modifier of the
graph-ite platform surface has a limited lifetime. To its
investiga-tion, the solution of a constant concentration of mercury
wasalways applied after the 10 atomization cycles, and
mercuryabsorbance was measured. In our previous work [10],
thelifetime for the Pd modifier was found to be 100120atomization
cycles. In case of the Re, a sensitivity decreaseof 10% was
observed after 200 cycles. The Re modifier ismore stable due to a
higher boiling point than Pd. Thesurface of the platforms was
always recoated with optimalmass of 250 g Pd or Re after 100 or 200
cycles. By usingthe less mass of modifier, lower sensitivity was
observed.The electrodeposition from a drop proved to be a
suitableway for graphite surface modification with rhenium as
well.This technique does not require a special cell, the
electrolysisspans a short time (35min), and the electrochemical
coatingof the SS platform is ensured.
3.2. Stabilizing Agents for Hg(II) in Solution. The
pyrolysiscurves (Figures 2 and 3) and the influence of the amounts
ofstabilizing agents on mercury absorbance were investigatedfor
both surface modifiers and all stabilizing agents. Thesolutions
were injected in a volume 20 L. For comparisonthe results obtained
for solutions of mercury(II) only indiluted nitric acid without a
stabilizing agent are shown.The integrated absorbance for mercury
is low and indicatesthat part of mercury was lost, probably already
during thedrying stage. The investigated graphite surfacemodifiers
thushave little stabilizing eect for mercury in a diluted
nitricacid solution during the drying stage. Therefore, the
additionof a stabilizing agent into calibration solutions is
necessary.Maximum usable pyrolysis temperatures for solutions
ofmercury(II) with stabilizing agents are shown in Table 3. Inthis
table, the data for potassium permanganate with Pdmodifier [10] and
newly measured with Re modifier arementioned.
The aqueous calibration solutions in the presence of1 g L1
thiourea were prepared at pH 1.5.With Zeeman back-ground correction
for both Pd and Re modifiers, an over-correction of the signal was
observed and the absorbancerecord was made impossible. A change of
pH in range1.55, similarly to the concentration of thiourea in
range0.130 g L1, did not eliminate this eect. With
deuteriumbackground correction and Pd modifier, a dual-split
peak
100 150 200 250 300 350 4000
0.2
0.4
0.6
0.8
1
1.2
1.4
0
0.02
0.04
0.06
0.08
0.1
0.12
Pyrolysis temperature (C)
Inte
grat
ed a
bsor
ban
ce (
s)
Inte
grat
ed a
bsor
ban
ce (
s)
Figure 2: Pyrolytic curves for Hg(II) solutions in the presence
ofstabilizing agents and Pd surface modifier. left axis: DEDTC,
TMDTC; right axis: without stabilizing agent.
100 150 200 250 300 350 400 4500
0.2
0.4
0.6
0.8
1
1.2
0
0.02
0.04
0.06
0.08
0.1
0.12
Pyrolysis temperature (C)
Inte
grat
ed a
bsor
ban
ce (
s)
Inte
grat
ed a
bsor
ban
ce (
s)
Figure 3: Pyrolytic curves for Hg(II) solutions in the presence
ofstabilizing agents and Re surface modifier. left axis:
thiourea(deuterium background correction), DEDTC, TMDTC; rightaxis:
without stabilizing agent.
was observed. By using the graphite platform, modified withRe
and with deuterium background correction, the determi-nation of
mercury was possible with RSD = 3.54.1% for 510 ng Hg (n = 5). The
calibration curve was linear to 10 ngHg (R2 = 0.9955).
TMDTC forms with Hg(II) stable chelate, which maybe extracted in
an organic solvent. Acetylacetone and MIBKwere selected for this
purpose. The use of an acetylacetoneas a solvent was not
appropriate, because overcorrectionof the signal was observed.
Preevaporation of acetylacetoneunder an infralamp did not eliminate
this eect. By usingMIBK as a solvent, mercury was stabilized to
200C for bothsurface modifiers. However, the overcorrection of the
signalwas observed with the use of a Pd modifier and
Zeemanbackground correction at pyrolysis temperatures 260340C.This
eect was not observed by using deuterium backgroundcorrection. The
presence of 0.014mg TMDTC in 20 L ofcalibration solution had no
influence on the mercury signal,and the amount of 0.05mg TMDTC was
chosen as optimal.Both methods of preparation of the calibration
solutions,with the use of extract in MIBK and with aqueous
methanolsolution refilled by MIBK, yielded the same results. TheRSD
values for the measurements with extracts by pyrolysistemperature
200C were 1.32.4% for 1015 ng Hg (n = 5)
-
The Scientific World Journal 5
Table 4: Results obtained for mercury content in environmental
materials using SS-ETAAS with modification of platform surface
andcalibration against calibration solutions.
Material
Obtained value SD/mg kg1 (n = 5) Certified anddetermined
value SD/mg kg1TMDTC/MIBK TMDTC/methanol/MIBK KMnO4
a
Pd Re Pd Re Pda Re
Soil II 37.2 1.8 37.7 1.9 38.9 1.3 38.3 1.4 38.8 1.6 36.6 2.1
38.1 0.7Plant 35.7 1.7 37.5 1.8 37.4 1.6 37.7 1.4 37.7 2.0 36.5 1.8
37.6 0.5GBW 07405 0.25 0.02 0.27
0.020.30 0.02
0.29 0.02
0.30 0.01
0.27 0.02
0.29 0.03aAccording to procedure in [10].
90 130 150 180 200 230 250 280 300 3300
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
0
0.002
0.004
0.006
0.008
0.01
0.012
0.014
Nor
mal
ized
inte
grat
ed a
bsor
ban
ce
Nor
mal
ized
inte
grat
ed a
bsor
ban
ce
Pyrolysis temperature (C)
Figure 4: Pyrolytic curves for solid samples in presence Pd
surfacemodifier. left axis: plant, soil II; right axis: CRMCBW
07405.
with Pd modifier and 1.9-2.0% for 1015 ng Hg (n = 5) withRe
modifier. The RSD of mercury determination in MIBK-methanol
solution was 1.92.4 for 1015 ng Hg (n = 5)with Pd modifier and
2.02.5% for 1015 ng Hg (n = 5)with Re modifier. Calibration curves
were linear to 15 ng Hg(R2 = 0.9975 or R2 = 0.9994 for extracts
with Pd or Remodifier and R2 = 0.9976 or R2 = 0.9994 for
MIBK-methanol solution with Pd or Re modifier). By using the
Pdmodifier, the detection limit, 128 pg Hg, acquired through6
repetitive firings of the platform with TMDTC enables
thedetermination of 0.43mg kg1 Hg for an optimum samplemass of
0.3mg. By using the Re modifier, the detection limitwas 120 pg Hg
and 0.40mg kg1 Hg.
DEDTC as the chelating agent was not suitable for sta-bilization
of mercury in calibration solutions, because dual-split peaks were
obtained for both types of modifiers andbackground corrections. The
change of the concentrationof DEDTC in a range of 0.014mg DEDTC in
20 L ofsolution or modification of the temperature program had
noinfluence on the shape of peak.
The use of 100 g L1 of permanganate in combinationwith the Re
modifier of the graphite platform surfaceprovided better results
than those in combination with Pdmodifier. The disadvantage of
permanganate is the necessityof dosing only 3 L of KMnO4 solution
and diculty inpreparing the stock solution of permanganate. The RSD
ofmercury determination was 4.04.9% for 1015 ng Hg (n =5) with the
Pd modifier [10] and 3.55.0% for 1015 ng
130 150 170 200 230 250 2800
0.3
0.6
0.9
1.2
1.5
1.8
2.1
0
0.005
0.01
0.015
0.02
0.025
Nor
mal
ized
inte
grat
ed a
bsor
ban
ce
Nor
mal
ized
inte
grat
ed a
bsor
ban
ce
Pyrolysis temperature (C)
Figure 5: Pyrolytic curves for solid samples in presence Re
surfacemodifier. left axis: plant, soil II; right axis: CRMCBW
07405.
Hg (n = 5) with the Re modifier. Calibration curves werelinear
to 15 ng Hg (R2 = 0.9991 for the Pd modifier andR2 = 0.9998 for Re
modifier). By using the Pd modifier, thedetection limit, 120 pg Hg,
was acquired from 10 repetitivefirings of platform with KMnO4 and
enabled determinationof 0.40mg kg1 Hg for an optimum sample mass of
0.3mg.By using the Re modifier, the detection limit was 110 pg
Hgand 0.37mg kg1 Hg.
3.3. Analytical Results. The pyrolytic curves (Figures 4 and
5)were investigated for solid samples by using both
surfacemodifiers and from their shape maximum usable
pyrolysistemperature results for soil II 280C with Pd or 230C
withRe, for plant 200C with Pd or 230C with Re, and for CRMGBW
07405 200C with both surface modifiers. The resultsobtained for
mercury content in soil II, plant, and CRMGBW 07405 using SS
platforms modified with palladiumor rhenium electrolytic from a
drop of modifier solutionsand calibration against MIBK extracts or
MIBK-methanolsolutions in presence of TMDTC are given in Table 4.
Forcomparison the results obtained for permanganate with thePd
modifier presented in [10] or with the Re modifier newlymeasured in
this paper are also mentioned. In all cases,the results are in good
agreement with the certified valueand with those obtained by
measurement on the AMA 254analyzer. The RSD values are dependent on
the numberof platform firings and consequently on the state of
the
-
6 The Scientific World Journal
platform surface. According to the results, the switch fromthe
calibration solutions to the samples on the modifiedsurface can
also play a certain role. It can be connected withincreased
precision of mercury determination using calibra-tion solutions in
the presence of TMDTC.
4. Conclusion
Agents such as TMDTC, DEDTC, and thiourea were in-vestigated for
complexation of mercury in calibration solu-tions and its thermal
stabilization in a solid samplingplatform in the determination of
mercury using SS-ETAAS.The calibration solutions were used in the
form of MIBKextracts or MIBK-methanol solutions for the TMDTC
andDEDTC chelates and aqueous solutions for the thioureacomplexes.
Only calibration with TMDTC was successful.MIBK-methanol solutions
in the presence of TMDTC areeasier to prepare than MIBK extracts.
Therefore, calibrationwith MIBK-methanol solutions in the presence
of TMDTCwas preferred. Higher precision in calibration and
easiermanipulation with the solutions makes the calibration
withMIBK-methanol solutions preferable to the heretofore
usedpotassium permanganate [1, 10]. The use of standardsolutions
for calibration also provides the best precision andlowest
uncertainty prior to the use of reference materials.The surface of
the graphite platforms for solid sampling wasmodified with
palladium or rhenium using electrodepositionfrom a drop of
solutions. This process of electrochemicalcoating of the SS
platform surface is promising for the prepa-ration of graphite
surface modifiers in SS-ETAAS. The Remodifier is preferable due to
the higher lifetime of platformcoating. The use of the SS-ETAAS
method with a modifiedsurface of the SS platform and calibration
against stabilizedcalibration solutions reduces the time of
analysis comparedwith the mercury determination after the sample
digestion.Sample preparation requires only routine grinding
andhomogenization. The new SS-ETAAS procedure using directsampling
solid samples into a platform with an Re modifiedgraphite surface,
and the calibration against MIBK-methanolsolutions in the presence
of TMDTC is proposed for thedetermination of mercury content in
solid environmentalsamples, such as soil and plants. With
calibration againstMIBK-methanol solutions in the presence of
TMDTC, thedetection limit was 120 pg and with a sample mass of
0.3mgit was 0.4mg kg1 Hg.
Acknowledgments
This work was supported by the project MSM 0021622412 ofthe
Ministry of Education, Youth and Sports of the CzechRepublic and
the project MUNI/A/0992/2009 of MasarykUniversity in Brno.
References
[1] A. F. da Silva, B. Welz, and A. J. Curtius, Noble metals
aspermanent chemical modifiers for the determination of mer-cury in
environmental reference materials using solid sam-pling graphite
furnace atomic absorption spectrometry and
calibration against aqueous standards, Spectrochimica ActaPart
B, vol. 57, no. 12, pp. 20312045, 2002.
[2] K. H. Grobecker and A. Detcheva, Validation of
mercurydetermination by solid sampling Zeeman atomic
absorptionspectrometry and a specially designed furnace, Talanta,
vol.70, no. 5, pp. 962965, 2006.
[3] I. Lopez-Garca, M. Sanchez-Merlos, and M. Hernandez-Cordoba,
Determination of mercury in soils and sedimentsby graphite furnace
atomic absorption spectrometry withslurry sampling, Spectrochimica
ActaPart B, vol. 52, no. 14,pp. 20852092, 1997.
[4] B. Welz, Atomic absorption spectrometrypregnant againafter
45 years, Spectrochimica acta, Part B, vol. 54, no. 14,
pp.20812094, 1999.
[5] M. G. R. Vale, M. M. Silva, B. Welz, and E. C. Lima,
De-termination of cadmium, copper and lead in mineral coalusing
solid sampling graphite furnace atomic absorptionspectrometry,
Spectrochimica ActaPart B, vol. 56, no. 10,pp. 18591873, 2001.
[6] N. Coskun and S. Akman, Direct determination of man-ganese
in vitamin-mineral tablets using solid sampling elec-trothermal
atomic absorption spectrometry, Talanta, vol. 64,no. 2, pp. 496500,
2004.
[7] M. Resano, E. Garca-Ruiz, M. Aramenda, and M. A. Be-larra,
Solid sampling-graphite furnace atomic absorptionspectrometry for
Hg monitoring in soils. Performance asa quantitative and as a
screeningmethod, Journal of AnalyticalAtomic Spectrometry, vol. 20,
no. 12, pp. 13741380, 2005.
[8] A. F. da Silva, F. G. Lepri, D. L. G. Borges, B.Welz, A. J.
Curtius,and U. Heitmann, Determination of mercury in
biologicalsamples using solid sampling highresolution
continuumsource electro thermal atomizaion atomic absorption
spec-trometry with calibration againts aqueous standards, Journalof
Analytical Atomic Spectrometry, vol. 21, pp. 13211326,2006.
[9] A. Detcheva and K. H. Grobecker, Determination of Hg, Cd,Mn,
Pb and Sn in seafood by solid sampling Zeeman atomicabsorption
spectrometry, Spectrochimica ActaPart B, vol.61, no. 4, pp. 454459,
2006.
[10] R. Cervenka, H. Zelinkova, M. Konecna, and J.
Komarek,Electrochemical modification of a graphite platform fora
solid sampling electrothermal atomic absorption spectrome-try of
mercury, Analytical Sciences, vol. 26, no. 9, pp. 989993,2010.
[11] S. M. Maia, B. Welz, E. Ganzarolli, and A. J. Curtius,
Fea-sibility of eliminating interferences in graphite furnace
atomicabsorption spectrometry using analyte transfer to the
perma-nently modified graphite tube surface, Spectrochimica
ActaPart B, vol. 57, no. 3, pp. 473484, 2002.
[12] D. Baralkiewicz, H. Gramowska, M. Kozka, and A.
Kanecka,Determination of mercury in sewage sludge by direct
slurrysampling graphite furnace atomic absorption
spectrometry,Spectrochimica ActaPart B, vol. 60, no. 3, pp. 409413,
2005.
[13] E. Bulska, W. Kandler, and A. Hulanicki, Noble metals
aspermanent modifiers for the determination of mercury
byelectrothermal atomic absorption spectrometry, Spectrochi-mica
ActaPart B, vol. 51, no. 9-10, pp. 12631270, 1996.
[14] I. Kunert, J. Komarek, and L. Sommer, Determination
ofmercury by atomic absorption spectrometry with cold vapourand
electrothermal techniques, Analytica Chimica Acta, vol.106, no. 2,
pp. 285297, 1979.
[15] I. Grigic, V. Hudnik, and S. Gomiscek, Behaviour of
mercurycomplexes in a graphite tube furnace for atomic
absorption
-
The Scientific World Journal 7
spectrometry, Analytica Chimica Acta, vol. 226, no. 2,
pp.203211, 1989.
[16] M. Konecna and J. Komarek, Utilization of
electrodepositionfor electrothermal atomic absorption spectrometry
determi-nation of gold, Spectrochimica ActaPart B, vol. 62, no.
3,pp. 283287, 2007.
[17] C. T. Costley, K. F. Mossop, J. R. Dean, L. M. Garden,
J.Marshall, and J. Carroll, Determination of mercury
inenvironmental and biological samples using pyrolysis
atomicabsorption spectrometry with gold amalgamation,
AnalyticaChimica Acta, vol. 405, no. 1-2, pp. 179183, 2000.
-
Submit your manuscripts athttp://www.hindawi.com
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Inorganic ChemistryInternational Journal of
Hindawi Publishing Corporation http://www.hindawi.com Volume
2014
International Journal ofPhotoenergy
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Carbohydrate Chemistry
International Journal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Journal of
Chemistry
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Advances in
Physical Chemistry
Hindawi Publishing Corporationhttp://www.hindawi.com
Analytical Methods in Chemistry
Journal of
Volume 2014
Bioinorganic Chemistry and ApplicationsHindawi Publishing
Corporationhttp://www.hindawi.com Volume 2014
SpectroscopyInternational Journal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
The Scientific World JournalHindawi Publishing Corporation
http://www.hindawi.com Volume 2014
Medicinal ChemistryInternational Journal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Chromatography Research International
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Applied ChemistryJournal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Theoretical ChemistryJournal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Journal of
Spectroscopy
Analytical ChemistryInternational Journal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Journal of
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Quantum Chemistry
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
Organic Chemistry International
ElectrochemistryInternational Journal of
Hindawi Publishing Corporation http://www.hindawi.com Volume
2014
Hindawi Publishing Corporationhttp://www.hindawi.com Volume
2014
CatalystsJournal of