21/04/23 1

First FRCR Examination in Clinical Radiology

Diagnostic Radiology & Radionuclide Radiology

(4b) Patient Dosimetry

John SaundersonRadiation Protection Adviser

Page 2: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 2

RCR Syllabus

•Methods

•Diagnostic reference levels (including high dose techniques)

•Magnitude and measurements .

Page 3: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 3

Methods

• General radiology

• Fluoroscopy

• Computed Tomography

• Nuclear Medicine .

Page 4: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 4

General radiology

• ESD– Thermoluminescent dosemeter (TLD)– exposure factors

• Dose-Area Product (DAP)

• Effective dose .

Page 5: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 5

T.L.D.

• Crystals, e.g. lithium fluoride• Radiation causes electrons to be caught in

“traps”• At lab. TLDs heated to 240oC• Electrons released, light emitted• Amount of light emitted proportional to dose .

Page 6: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 6

T.L.D.s +/-

+ Small+ Tissue equivalent+ Read 1 Gy+ Easy for radiographer

- No direct readout- Sensitive to heat, UV,

dirt- Tricky to calibrate- Easy to loose- Special ones needed for

low dose (e.g. chest) .

Page 7: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 7

Using exposure factors• kV, mAs, field size, FSD• If D = dose at 100 cm for 80 kV, 100 mAs

then• ESD = D

• x kV2/802 • x mAs/100 • x 1002/FSD2 • x BSF

• BSF = backscatter factor, depends on HVL and field size.

Page 8: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 8

E.g.• What is ESD for 100 kV, 50mAs, 15 x 15 cm field, 75 cm

FSD for Room 3?• From annual survey

– @ 80 kV, D = 9.1 mGy/100mAs @ 1 m– HVL for 100kV was 3 mmAl

• From tables– BSF for 3mmAl, 15 x 15 field = 1.33

• Therefore, entrance surface dose= 9.1mGy x 1002/802 x 50/100 x 1002/752 x 1.33

= 0.17 mGy .

Page 9: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 9

Using exposure factors +/-

+ No fiddly TLD for radiographers!

+ No fiddly TLDs for physicist!

+ Radiographers can calculate ESDs

+ No minimum dose

- No direct readout- Exposure parameters

must be recorded- Assumes tube output not

changed from last survey.

Page 10: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 10

Dose Area Product

• Because dose falls with 1/d2 and area increases with d2, DAP is independent of distance.

Page 11: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 11

DAP +/-

+ No fiddly TLD for radiographers!

+ Only one number to record

+ Instant answer

- Doesn’t take into account backscatter

- Initial cost (several thousand £)

- Units sometimes cause confusion (cGy.cm2, or

Gy.cm2, Gy.m2) .

Page 12: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 12

Effective dose

• Complicated to calculate from ESD or DAP

• Can use computer models which make assumptions on– field size– patient size– field position

Page 13: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 13

Page 14: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 14

Page 15: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 15

Page 16: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 16

Page 17: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 17

Page 18: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 18

Fluoroscopy

• Dose-Area Product (DAP)

• Exposure factors

• Effective dose .

Page 19: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 19

DAP +/- for fluoroscopy

+ Instant answer, etc.+ DAP moves with the

tube+ Gives good indication of

relative risks of inducing cancer

- Not directly linked to erythema risk.

Page 20: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 20

Exposure factors

• Based on assumed FSDs, etc.

Page 21: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 21

Effective dose

• Can be “fudged” using radiograph software.

Page 22: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 22

GAFCHROMIC film

Page 23: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 23

Page 24: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 24

Page 25: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

25

GAFCHROMIC film

• optical density is proportional to the absorbed dose• 0.01 Gy to 50 Gy • Energy independent from 30 keV to 30 MeV• £20 per 14” x 17” sheet

Page 26: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 26

Computed Tomography

• CT Dose Index (CTDI)

• Dose-Width Product (DWP)

• Effective dose

Page 27: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 27

CT Dose Index (CTDI)

• Applies to a single slice

• Can be used to compare– different slice widths– different physical filter– different scanners– etc.

dzTN

zDCTDI

mm

mm

50

50

100

)(

Page 28: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 28

Dose-Length Product (DLP)

• DWP = CTDI x n x T

• Gives an idea of relative dose for a whole scan

• Can be used to compare effect of pitch, etc.

Page 29: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 29

Effective dose

• NRPB program similar to radiography one

Page 30: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 30

Page 31: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 31

CT dose/risk calculation• e.g. wrong patient referred for CT scan

(laryngectomy)

Page 32: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 32

Nuclear Medicine

• MIRD

• Add up dose to each organ from the dose irradiating from each organ.

Page 33: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 33

Page 34: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 34

Diagnostic Reference Levels

• Early 80’s survey

• DRLs today

• IRMER

Page 35: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 35

Mid-80’s surveyMethod

• Survey of twenty UK hospitals

• 13 most common views

• For each 10-20 patients (60-80kg) at

• DAP or ESD by TLD measured.

Page 36: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 36

Mid-80’s survey Results

• E.g. Chest PA– Median ESD = 0.18 mGy– Minimum ESD = 0.03 mGy– Maximum ESD = 1.43 mGy– Max / min = 48 !!.

Page 37: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 37

Entrance Surface Dose (mGy) Radiograph

Min. Median 3rd Quartile Max.

Lumbar spine AP 0.8 7.7 10 59

Lat 2.4 20 30 108

LSJ 7.4 35 40 131

Abdomen AP 0.7 6.7 10 62

Pelvis AP 0.9 5.7 10 32

Chest PA 0.03 0.2 0.3 1.4

Lat 0.14 1.0 1.5 11

Skull AP 0.7 4.0 5 14

PA 1.8 4.3 5 13

Lat. 0.4 2.2 3 9

Dose-area product (Gy.cm2) Radiograph

Min. Median 3rd Quartile Max.

Lumbar spine

(3.4 films)

2.0 12 15 93

Barium enema

(8.5 films, 224 s fluoro)

6.2 41 60 272

Barium meal

(7.8 films, 193 s fluoro)

0.5 17 25 163

IVU

(8.2 films)

3.3 29 40 251

Abdomen

(1.4 films)

0.7 5 8 30

Pelvis

(1.1 films)

0.49 4 5 19

Page 38: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 38

Mid-80’s surveyRecommendation

• Use 75th percentile as reference value

• i.e. carry out local surveys and take action if average dose is greater than ¾ of national survey doses

• e.g. for chest PA reference = 0.3mGy ESD

• Send results to NRPB to review national reference doses every 5 years.

Page 39: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 39

DRLs today• A DRL is essentially a guide to the

rather indistinct border between “good and normal practice” and “bad and abnormal practice”.

Page 40: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 40

IRMER

• National DRLs – set as 3rd quartile– average from survey of “standard patients”– should be below DRL

• Local DRLs– ?Average for Trust?– Most should be below average

Page 41: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 41

Page 42: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 42

Page 43: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 43

Page 44: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 44

Page 45: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 45

Page 46: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 46

Page 47: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 47

Staff and Environmental Monitoring

• Devices– TLD– OSLD– Film– Electronic

• Body

• Extremity

• Environment

Page 48: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 48

Relevant measurement techniques

Page 49: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 49

Measuring Dose

• Luxel dose badges

• TLD finger rings– Can be cold sterilized– Heat sensitive

Page 50: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 50

Luxel badges

• Wear underneath lead rubber apron

• Assume dose to badge = effective dose

• Can be worn for 2 weeks to 3 months (usually 1 month)

• Must be returned promptly.

Page 51: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 51

Electronic Dosemeters• Used by Radiation Physics staff to test x-ray unit

and measure environmental doses

Page 52: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 52

Page 53: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 53

Absent Badges

Participant Name, Number

Begin wear date

End wear date

Dosimeter Location

Body Dose (mSv)

Extremity Dose (mSv)

Body Dose YTD

Exremity Dose YTD

Notes Derived

FALSE ARCHER, MS G,00176 01/09/2013 30/09/2013 Chest 0.1   1.27     MEASURED

FALSE ARCHER, MS G,00176 01/09/2013 30/09/2013 Right Finger   0.2   1.2   MEASURED

FALSE ATKINSON,A,00308 01/09/2013 30/09/2013 Chest 0.03   0.18     MEASURED

FALSE AVERY, DR G,00298 01/09/2013 30/09/2013 Chest 0.04   0.24     MEASURED

FALSE AVERY, DR G,00298 01/09/2013 30/09/2013 Left Finger   0.2   9.6   MEASURED

FALSE AVERY, DR G,00298 01/09/2013 30/09/2013 Right Finger   0.2   23.7   MEASURED

FALSE AZOR, B,00587 01/09/2013 30/09/2013 Chest 0.02   0.1     MEASURED

FALSE BAJALAN, L,00325 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSEBARLOW,

NICHOLAS,00053 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSE BEADLE, MRS E,00609 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSE BEADLE, MRS E,00609 01/09/2013 30/09/2013 Right Finger   0.2   1.2   MEASURED

FALSEBOTTOMLY, JAYNE,00694 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

1BOWERING, SUE,00279 01/09/2013 30/09/2013 Chest           ABSENT

FALSE BRAMALL, M,00588 01/09/2013 30/09/2013 Chest 0.01   0.03     MEASURED

FALSE BROWN, SUSAN,00326 01/09/2013 30/09/2013 Chest 0.03   0.07     MEASURED

FALSE BYASS, DR O,00384 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSE BYASS, DR O,00384 01/09/2013 30/09/2013 Left Finger   0.2   13.3   MEASURED

FALSE BYASS, DR O,00384 01/09/2013 30/09/2013 Right Finger   0.2   14.9   MEASURED

FALSE CAST, DR J,00376 01/09/2013 30/09/2013 Chest 0.01   0.99     MEASURED

FALSE CAST, DR J,00376 01/09/2013 30/09/2013 Waist 0.48   4.13   overapron MEASURED

FALSE CAST, DR J,00376 01/09/2013 30/09/2013 Right Finger   1.1   14.8   MEASURED

YTD = year to date

Luxel badge results(Sept 2013, CHH Radiology)

Page 54: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

21/04/23 54

Absent Badges

Participant Name, Number

Begin wear date

End wear date

Dosimeter Location

Body Dose (mSv)

Extremity Dose (mSv)

Body Dose YTD

Exremity Dose YTD

Notes Derived

FALSE ARCHER, MS G,00176 01/09/2013 30/09/2013 Chest 0.1   1.27     MEASURED

FALSE ARCHER, MS G,00176 01/09/2013 30/09/2013 Right Finger   0.2   1.2   MEASURED

FALSE ATKINSON,A,00308 01/09/2013 30/09/2013 Chest 0.03   0.18     MEASURED

FALSE AVERY, DR G,00298 01/09/2013 30/09/2013 Chest 0.04   0.24     MEASURED

FALSE AVERY, DR G,00298 01/09/2013 30/09/2013 Left Finger   0.2   9.6   MEASURED

FALSE AVERY, DR G,00298 01/09/2013 30/09/2013 Right Finger   0.2   23.7   MEASURED

FALSE AZOR, B,00587 01/09/2013 30/09/2013 Chest 0.02   0.1     MEASURED

FALSE BAJALAN, L,00325 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSEBARLOW,

NICHOLAS,00053 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSE BEADLE, MRS E,00609 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSE BEADLE, MRS E,00609 01/09/2013 30/09/2013 Right Finger   0.2   1.2   MEASURED

FALSEBOTTOMLY, JAYNE,00694 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

1BOWERING, SUE,00279

01/09/2013

30/09/2013 Chest          

ABSENT

FALSE BRAMALL, M,00588 01/09/2013 30/09/2013 Chest 0.01   0.03     MEASURED

FALSE BROWN, SUSAN,00326 01/09/2013 30/09/2013 Chest 0.03   0.07     MEASURED

FALSE BYASS, DR O,00384 01/09/2013 30/09/2013 Chest         Component Note: H-Unused MEASURED

FALSE BYASS, DR O,00384 01/09/2013 30/09/2013 Left Finger   0.2   13.3   MEASURED

FALSE BYASS, DR O,00384 01/09/2013 30/09/2013 Right Finger   0.2   14.9   MEASURED

FALSE CAST, DR J,00376 01/09/2013 30/09/2013 Chest 0.01   0.99     MEASURED

FALSE CAST, DR J,00376 01/09/2013 30/09/2013 Waist 0.48   4.13   overapron MEASURED

FALSE CAST, DR J,00376 01/09/2013 30/09/2013 Right Finger   1.1   14.8   MEASURED

Luxel badge results

Page 55: 28/10/20151 First FRCR Examination in Clinical Radiology Diagnostic Radiology & Radionuclide Radiology (4b) Patient Dosimetry John Saunderson Radiation.

Absent Badges

Participant Name, Number

Begin wear date

End wear date

Dosimeter Location

Body Dose (mSv)

Extremity Dose (mSv)

Body Dose YTD

Extremity Dose YTD

Notes Derived

FALSEC##T, DR J,00376 01/09/2013 30/09/2013 Chest 0.01   0.99     MEASURED

FALSEC##T, DR J,00376 01/09/2013 30/09/2013 Waist 0.48   4.13   overapron MEASURED

FALSEC##T, DR J,00376 01/09/2013 30/09/2013

Right Finger   1.1   14.8   MEASURED

Luxel badge results

Effective dose assumed to be underapron dose. 0.99mSv to end Sept, so predicted annual dose = 12/9 x 0.99 = 1.32mSv for year

Finger dose, 4.13 to end Sept, so predicted annual dose = 12/9 x 14.8 = 19.73 mSv for year