-
ABSTRACT Five methods of measuring glucose in
the horse were compared with plasma glucose reference values
generated with an Autoanalyzer3 (Seal Analytical Ltd., Hampshire,
UK). Plasma glucose was measured with a One Touch Ultra (LifeScan,
Milpitas, CA) hand-held glucose meter and a YSI 2300 Stat Plus
Glucose and Lactate Analyzer (YSI; YSI Inc., Yellow Springs, OH).
Glucose also was measured in whole blood using One Touch Ultra and
YSI. Finally, interstitial glucose measurements were obtained with
a SEVEN continuous glucose monitor-ing device (Dexcom, San Diego,
CA). Glucose measurements were obtained on 2 consecutive days in 6
American Quarter Horses maintained on native prairie hay and water.
Each morning, 2 baseline glucose measurements were obtained 30 min
apart. After a meal of sweet feed, glucose measurements were
obtained every 30 min for 4 h. Patterns of postprandial glucose
increases and subsequent decreases were similar across all methods
but were displaced vertically from one another. Of the 5 methods,
the YSI method with plasma appeared to have the best
reproducibility of Autoana-
lyzer3 glucose values based on a random coefficients model with
intercept not statistically different from zero (P = 0.08) and
slope not statistically different from one (P = 0.08) and by having
the highest Lins concordance coefficient (r = 0.77). The other
methods had biased random coefficients models with inter-cepts not
equal to zero (P > 0.05) and slopes not equal to one (P >
0.05) and Lins concordance coefficient values that ranged from 0.39
to 0.64. Technical diffi-culties with the SEVEN device limited its
utility as a reliable method for evaluating glucose in the
equine.
Key words: continuous glucose monitor , equine , glucometer ,
glucose
INTRODUCTION Several disease conditions in the
horse have been linked to insulin resistance: equine metabolic
syndrome (Johnson, 2002), laminitis (Coffman and Colles, 1983;
Treiber et al., 2006), pituitary adenoma (Garcia and Beech, 1986),
hyperlipidemia (Jeffcott and Field, 1985), and osteochondrosis
dis-secans (Ralston, 1996). Other condi-tions, such as
polysaccharide storage myopathy, render the horse unable to store
glycogen normally. To manage these conditions, afflicted horses
often are fed low-starch or low-glycemic-in-dex diets (Johnson,
2002; Kronfeld et
al., 2005). Low-starch feeds are being marketed as an aid in
management of disease conditions, but they are also being marketed
to the general public as safer alternatives to traditional equine
rations.
To develop and assess low-starch diets, there is ongoing
research that requires the measurement of equine blood glucose
concentrations over a period of time. Using laboratory equipment to
measure glucose can be cumbersome and labor intensive. Sev-eral
point-of-care glucose monitoring devices are convenient, relatively
inex-pensive, and readily available for use in humans, but most
have not been validated for use in the equine. One of the newer
technologies available to diabetic humans is the continuous glucose
monitoring (CGM) device (Wentholt et al., 2007). Typically, CGM
devices obtain glucose read-ings through a sensor probe inserted
subcutaneously. Sensors detect an electrical current generated by a
reac-tion between interstitial glucose and glucose oxidase. Sensors
transmit a signal to a receiver that records data at predetermined
intervals. Aside from its near-continuous collection of glucose
data, the CGM dramatically reduces the number of whole blood
samples required for effective glucose monitoring, and it collects
continu-ous glucose data for up to 7 d. This
The Professional Animal Scientist 27 ( 2011 ):204214
A comparison of methodologies for measuring glucose
concentrations in the horse1 T. L. Slough,*2 C. D. Gunkel,* L. W.
Murray, and J. S. Drouillard* *Department of Animal Sciences and
Industry, and Department of Statistics, Kansas State University,
Manhattan 66506
2011 American Registry of Professional Animal Scientists
1 This is contribution No. 11-088-J from the Kansas Agricultural
Experiment Station, Manhattan. 2 Corresponding author:
[email protected]
-
Methods of measuring glucose in horses 205
study was designed to assess whether glucose readings obtained
from clini-cally normal horses through the use of either a CGM
device or a hand-held glucose monitor (both designed for use in
human patients) accu-rately reproduce those obtained with
laboratory analyses using a reference technique.
MATERIALS AND METHODSAll animal procedures were ap-
proved by the Kansas State Universi-ty Institutional Animal Care
and Use Committee. Six American Quarter Horses were used for this
study, including 3 mares and 3 geldings. All horses were 4 yr of
age and weighed 488 20 kg. Horses were maintained on an ad libitum
diet of native prairie hay and water. None of the horses had a
history of metabolic disorders and all were considered to be
clini-cally normal.
Preliminary Experiments
Preliminary experiments were con-ducted to identify potential
problems with the use of a CGM in horses. In the initial
experiment, sensors for the SEVEN CGM device (Dexcom, San Diego,
CA) were applied in the adipose tissue overlying the latis-simus
dorsi muscle of the horse, just posterior to the scapula. The
receiver, which the manufacturer recommends remain within 1.8 m of
the sensor, was fastened to the halter of each horse. During this
study, the SEVEN system had a large number of errors. Many of the
receivers required reini-tialization, some failed and had to be
replaced, and numerous readings were missed. Customer service
representa-tives from Dexcom indicated that SEVEN system receivers
do not func-tion optimally in close proximity to metal, such as
belt buckles on human patients. Thus, the metal buckles on the
halters may have interfered with receiver function. In addition,
when the sensor was located in this posi-tion, many of the
interstitial probes (consisting of a 330-m filament) were bent to
almost 90 upon removal.
Perhaps rapid contraction of the large muscle mass underlying
the sensor in this location resulted in physical bending of the
probes, thus rendering them nonfunctional.
Due to the large number of miss-ing data points, a follow-up
experi-ment was conducted to determine optimum placement of the
sensor and receiver. Sensors were placed adjacent to the poll of
the horse, posterior to the scapula, on the crest of the neck,
posterior to the olecranon, and adjacent to the tail head.
Receivers were then attached to braided mane and tail hair, as well
as to a harness device (without any metal buckles) around the girth
of the horse. Wied-meyer et al. (2003) reported a Min-iMed CGM
worked well when the sensor was placed over the masseter muscle and
the receiver was attached to a halter on the head of the horse, but
the SEVEN system produced the fewest errors when the sensor was
placed near the base of the tail of the horse and the receiver was
attached to braided tail hair (Figure 1). Thus this positioning of
the sensor was used in subsequent experiments.
Current Experiment
On the day before the experiment, each horse was equipped with
an indwelling 14-gauge 13-cm jugular catheter and a SEVEN CGM
device. The SEVEN sensor and receiver were initialized, installed,
and calibrated following the manufacturers instruc-tions. Hair was
closely shaved from a patch of skin overlying the adipose tissue
adjacent to the base of the tail. The shaved area was cleaned with
alcohol, and after insertion, the sensor was firmly attached to the
skin of the horse with an adhesive-backed pad provided by the
manufacturer, with additional cyanoacrylate adhesive applied to the
pad to secure it to the skin. The sensor measured glucose
concentration in interstitial tissue and transmitted a signal to a
wire-less receiver every 5 min. The receiver was attached to
braided tail hair. The SEVEN receiver was calibrated every 12 h
(0700 and 1900 h, preprandially)
with the One Touch Ultra (OTU; LifeScan, Milpitas, CA) hand-held
glucose meter. Glucose measurements were obtained with the
hand-held glu-cose monitor following manufacturer directions.
Briefly, approximately 10 mL of blood was drawn through the jugular
catheter into a syringe, one drop (approximately 10 L) was
extracted and applied to a test strip, and a glucose reading was
obtained. The glucose reading obtained with the OTU was then
manually entered into the SEVEN receiver to complete the
calibration.
On the morning of each experiment, 2 baseline glucose
measurements were obtained 30 min apart. After acquisi-tion of the
second baseline sample, horses were fed 1.125 kg of commer-cial
sweet feed concentrate (Sweet Stuff 12, Cargill Animal Nutrition,
Minneapolis, MN) and allowed to eat uninterrupted for 30 min. These
horses were unaccustomed to grain meals, so a modest quantity of
sweet
Figure 1. The SEVEN sensor (A; Dexcom, San Diego, CA) was
inserted into the adipose tissue adjacent to the tail head and the
receiver (B) was attached to braided tail hair. This maintained the
sensor and receiver in close proximity to each other and avoided
possible interference from metal buckles on the halters of the
horses.
-
Slough et al.206
feed was offered to elicit a postpran-dial glucose response
without causing digestive upset in the horses. After feeding,
glucose measurements were obtained at 30-min intervals for 4 h. At
each sampling, the most recent glucose reading of the CGM (which
would have occurred within 5 min preceding sampling) was recorded
and 10 mL of whole blood was obtained via the indwelling jugular
catheter and placed into heparinized vacuum tubes. A single drop
(approximately 10 L) of the venous whole blood was applied to a
test strip, which was inserted into the OTU test port, as
recommended by the manufacturer, to obtain a glucose reading.
Whole blood from each sample also was subjected to glucose
analysis by a YSI 2300 Stat Plus Glucose and Lactate Analyzer (YSI;
YSI Inc., Yellow Springs, OH). All whole blood glucose measurements
were obtained within 30 min of sampling and the YSI produced an
intraassay CV of 0.12%. Whole blood was centrifuged at 876 g for 10
min at approxi-mately 13C and plasma was removed. A drop of plasma
(10 L) was applied to a test strip and analyzed using the OTU to
measure plasma glucose concentration. Remaining plasma was frozen
at 20C until analysis by an Autoanalyzer3 (AA3; Seal Analytical
Ltd., Hampshire, UK) according to procedures described by Fingerhut
et al. (1965). An intraassay CV of 0.35% was generated with the
AA3, and when the YSI was used to measure plasma glucose, an
intraassay CV of 0.03% was recorded. The entire sampling process
was repeated on the following day, yielding a total of 120 discrete
blood samples.
Analysis of Plasma Constituents
Each time blood was collected, glucose was measured using 6
meth-ods for each horse. These included 1) interstitial glucose
concentration as measured by the SEVEN system, 2) whole blood
glucose concentration as measured by the OTU, 3) plasma glucose
concentration as measured by
the OTU, 4) whole blood glucose con-centration as measured by
the YSI, 5) plasma glucose concentration as measured by the YSI,
and 6) plasma glucose concentration as measured by the AA3, the
reference method. All glucose measurements were obtained according
to manufacturer directions for each instrument used. For the YSI
and AA3, standards were run with each set of blood samples to
ensure quality control. Quality control was also monitored in the
OTU hand-held glucose meter by testing the control solution
provided by the manufacturer each morning and each time a new lot
of test strips was used.
Statistical Analysis
Perfect reproducibility between 2 methods occurs if the measured
glu-cose is identical for the methods, that is, when the
measurements are plotted in a bivariate graph, the points fall
exactly on the 45 line through the origin (Lin, 1989). Statistical
repro-ducibility occurs when the plotted line is not statistically
different from the 45 line through the origin.
Three statistical techniques were used to assess whether glucose
mea-surements obtained by methods 1 through 5 described in the
previous section statistically reproduce those obtained by the AA3,
our reference technique. These 3 techniques were used because they
each assess repro-ducibility slightly differently. All
cal-culations were performed using SAS version 9.2.1 software (SAS
Institute Inc., 2007).
First, a random coefficients re-gression model was fitted
(Littell et al., 1996) with AA3 values as the response (Y) and one
of the 5 other methods as predictor (X) in a simple linear
regression. In random coef-ficients regression models, an overall
fixed regression model (i.e., the aver-age relationship over all
horses on all days) is fitted, which (unlike ordinary regression
models) takes into account the inherent random variability due to
different horses and different days (i.e., the horseday
combination). For each model, the following parameters
were estimated: overall (fixed) inter-cept and slope; variance
components for random intercept, random slope, and residual; and
covariance compo-nent between random intercept and random slope. In
addition, an au-toregressive parameter was fitted to account for
serial correlation within each horseday combination. Models were
fitted using the SAS MIXED procedure with REML estimation.
Normality of residuals was checked using the SAS UNIVARIATE
proce-dure.
To assess reproducibility, t-tests were performed to check that
the intercept was not statistically differ-ent from zero and that
the slope was not statistically different from one for each model.
The t-test results, with 95% CI on the regression parameters, are
reported. For succinctness, in this paper we will refer to a model
with both intercept not different from zero and slope not different
from one as an unbiased model.
For an overall comparison of model fit among the 5 predictor
methods, Akaikes information criterion (AIC) was used. Akaikes
information crite-rion is used to compare models with complex
variancecovariance struc-tures (Littell et al., 1996; Burnham and
Anderson, 1998). For all methods having an unbiased regression
model, the predictor method with the model having the smallest AIC
value is considered the best for modeling AA3 values. Any
predictors with AIC val-ues within 2 of the minimum AIC are
considered equally good at modeling AA3 values, and AIC differences
of 7 or more indicate that the predictors are different in ability
to model AA3, with the higher AIC being less desir-able (Burnham
and Anderson, 1998).
The second statistical technique was described by Bland and
Altman (1986). BlandAltman calculations reported here are the mean
difference between measurements of 2 methods, the SD of
differences, and the number and percent of observed differences
outside the 95% prediction interval based on the t-distribution.
Assum-ing that differences are approximately normally distributed,
then only 5% of
-
Methods of measuring glucose in horses 207
values should be outside the predic-tion interval. If data are
not ap-proximately normal, BlandAltman results should be treated
with cau-tion. Therefore, normality of differ-ences was evaluated
using the SAS UNIVARIATE procedure. The mean difference reported is
the AA3 glucose mean minus the mean glucose of a predictor method,
so positive mean differences indicate that a predictor method has
lesser glucose measure-ments, on average, than the AA3, whereas
negative mean differences indicate the opposite. For these data,
BlandAltman calculations were per-formed on both original
observations and on residuals plus estimated means from the SAS
MIXED procedure to account for variability due to horseday
combinations. Results between the 2 were not different because the
horseday variance component was typically very small relative to
resid-ual variability. Therefore, BlandAlt-man results based on the
original data are reported. The authors note that, as Bland and
Altman (1986) point out, when possible this statistical analysis
should be used in conjunc-tion with knowledge of the clinical (as
opposed to statistical) significance of mean differences.
The third technique was Lins concordance coefficient (Lin,
1989),
which is a modification of the Pearson correlation coefficient
that not only evaluates the strength of the linear as-sociation
between 2 variables (X and Y) like Pearsons, but also evaluates the
degree of departure from the 45 line through the origin. Lins
concor-dance coefficient is reported with a 95% CI for each
predictor. If there is reproducibility between 2 methods, the 95%
CI should contain one. As with the BlandAltman analysis,
con-cordance coefficient calculations were performed on both the
raw data and the residuals plus estimated means to account for the
horseday variability. Unlike BlandAltman results, there were some
differences between concor-dance coefficients based on raw data or
based on residuals, so concordance coefficients based on residuals
are reported here.
The SEVEN system (X) also was assessed for its reproducibility
of values obtained by OTU used on both plasma (Y1) and whole blood
(Y2), because a hand-held glucose meter is generally used in
calibrating CGM devices. The comparison was done using all 3 of the
previously described statistical analyses.
There were some missing measure-ments in the current study, but
these had no effect on validity of the sta-tistical analyses other
than the issue
of smaller sample size with several predictor methods, as
discussed in the results. Data points were excluded only if there
was a mechanical or technical issue with the measurement
device.
RESULTS AND DISCUSSIONTechnical Issues with Devices
One of the issues of most concern was the number of missing
observa-tions, especially with the SEVEN system (Table 1). Although
the greatest number of missing observa-tions (33) occurred in whole
blood tested with the YSI, these errors were easily identified and
are simple to remedy. In those samples, air bubbles in the sample
tubes caused the YSI to generate obviously unrealistic glucose
values. Had the presence of the air bubbles been noticed before
separat-ing the plasma, the bubbles would have been removed and
these missing data points would have been avoided.
It is important to note that there were 12 missing data points
observed with the SEVEN system at the time of blood sampling, but
because the CGM records glucose measurements every 5 min, there
were far more missing data points that occurred throughout the
period. It just so
Table 1. A summary of glucose measurements collected1,2,3
Item
AA3 OTU YSI SEVEN
Plasma Blood Plasma Blood Plasma Interstitial fluid
Range (mM) Low 3.820 1.389 4.000 1.470 2.630 2.167 High 8.780
9.111 12.889 6.870 9.215 11.111Mean (mM) 5.824 5.803 8.018 4.147
5.920 4.887Missing data points4 0 0 0 33 0 12Percent of values
within 20% of AA3 N/A 70.8 21.7 21.7 84.2 37.51Methods of measuring
glucose included measurement of plasma glucose with an
Autoanalyzer3 (AA3; Seal Analytical Ltd., Hampshire, UK), glucose
in whole blood and plasma with a One Touch Ultra (OTU; LifeScan,
Milpitas, CA) hand-held glucometer, glucose in whole blood and
plasma with a YSI 2300 Stat Plus Glucose and Lactate Analyzer (YSI;
YSI Inc., Yellow Springs, OH), and interstitial glucose with a
Dexcom SEVEN continuous glucose monitor (SEVEN; Dexcom, San Diego,
CA).2For each method, the following values are given: maximum and
minimum glucose value, mean glucose concentration, number of
missing data points, and the percent of glucose values within 20%
of the values generated by the AA3.3Note that this is a summary of
raw data and ignores the autocorrelation among values for a given
horse over time.4There were 120 total possible observations with
each method tested.
-
Slough et al.208
happened that only 12 of those oc-curred at the time of blood
sampling. Although changing the position of the SEVEN system sensor
resulted in a considerable improvement com-pared with the 54
missing data points recorded during the preliminary experiment,
missing observations with the SEVEN system are not easily remedied
and appear to be due to mechanical errors within the system itself
when used in the horse.
Another technical difficulty arose during later experiments when
ambient temperatures and humidity increased. As the days became
more hot and humid, both the OTU and SEVEN collected condensation
inter-nal to the device. The receiver screen on the SEVEN and the
output screen on the OTU both became difficult to read, and in many
cases the devices did not display any output. These problems,
however, were short-lived and most devices returned to normal
function when weather conditions im-proved. However, 2 SEVEN
receivers never regained normal function and were replaced.
Reproducibility Results
A graph of raw data (Figure 2) re-vealed that, as expected, each
method of glucose measurement detected a postprandial increase,
followed by a decrease, in glucose concentra-tions. Overall,
glucose concentrations recorded using the different methods
followed the same trajectory, which was fairly consistent, but they
were displaced vertically from one another, with some methods
reading high and others reading low relative to the AA3 reference
method. Further informa-tion on the raw data is presented in
scatter plots (Figures 3 and 4), which illustrate the relationship
between measurements obtained by the various methodologies.
Descriptive statistics based on raw data are summarized in Table
1 for all methods. Note that these statistics do not account for
horseday variability, autocorrelation over time, or correla-tions
between method readings made on individual samples. Therefore,
means and ranges should be treated with caution in
nonstatistical compar-isons of methods. However, regarding
reproducibility of AA3 readings, the methods of the YSI with plasma
and the OTU with whole blood had, re-spectively, 84.2 and 70.8% of
samples with glucose readings within 20% of the AA3 readings,
whereas fewer than 50% of the glucose measurements obtained by the
OTU with plasma, the YSI with whole blood, and the SEVEN system
were within 20% of the numerical glucose values obtained by the
AA3. In 2003 the International Organization for Standardization
established criteria to evaluate the accuracy of glucose monitors:
If the glucose concentration is
-
Methods of measuring glucose in horses 209
system with respect to Lins concor-dance coefficient, but only
slightly better for the random coefficients models. Note that it
makes sense only to compare AIC among the unbiased random
coefficients models to pick a best model. Because there is only one
unbiased random coefficients model, AIC values are not compared
here.
Because the SEVEN system was calibrated with the OTU, random
coefficients regression models, BlandAltman calculations, and Lins
con-
cordance coefficient were also used to evaluate whether the
SEVEN system was able to reproduce the OTU using whole blood or
plasma. Neither of the random coefficients models were close to
having an unbiased relationship with (and hence reproducibility of)
OTU values. BlandAltman calcula-tions indicated that the difference
in means was closer to zero and the prediction interval was
narrower when whole blood was used rather than plasma, but both
methods generated
the same number of observations outside the prediction interval
(
-
Slough et al.210
It is important to note that the reference values obtained by
the AA3 in this experiment were obtained
with plasma samples; however, this is common practice in studies
designed to evaluate the accuracy of hand-held
glucometers in measuring glucose in whole blood (Hussain and
Sharief, 2000; Kozar et al., 2008; Kong et al., 2010). One might
expect glu-cose concentrations to be higher in plasma than in whole
blood, simply because of the dilution effect of the red blood cells
in whole blood, as well as the fact that blood cells metabo-lize
glucose at a rate of 5 to 10% per hour (Stockham, 1995).
Researchers have confirmed that plasma glucose values often are
larger than values obtained from whole blood (Hussain and Sharief,
2000), as was the case in this experiment, but some attribute these
higher values to a hematocrit-dependent error where glucose
readings tend to be overestimated in samples with low hematocrit
(Louie et al., 2000). Others have cited a differ-ence in glucose
distribution in various species, with humans having 50% of glucose
distributed within erythro-cytes and 50% within plasma (Cold-man
and Good, 1967), whereas dogs have a glucose distribution of 12.5%
in erythrocytes and 87.5% in plasma (Coldman and Good, 1967).
Hollis et al. (2008) observed more accurate readings when their
Accu-Chek hand-held glucometer was used with equine plasma than
when it was used with whole blood. However, one must use caution in
comparing results obtained by differing brands of glucometers.
Forbes and Brayton (2008) reported higher glucose values obtained
by an OneTouch Ultra2 glucometer compared with the Accu-Chek when
both models were used in mice. These contrasting findings might be
explained by differences in the indicators used by the 2 systems.
The OTU uses a carbon electrode indicator and the Accu-Chek
Ad-vantage uses a palladium electrode. The OTU also uses a glucose
oxidase reaction (Hones et al., 2008), whereas the Accu-Chek
Advantage does not (Russell et al., 2007). As a result, the OTU is
more sensitive to differ-ences in oxygen tension in samples. Both
the Accu-Chek and the OTU glucometers use hexacyanoferrate
III/hexocyanoferrate II in the media-tor reaction. This system is
prone to
Figure 3. Scatter plots illustrating the agreement in raw
glucose concentrations obtained between measurement methods. For
each, the vertical axis (Y) represents glucose measurements
obtained in plasma using the Autoanalyzer3 (AA3; Seal Analytical
Ltd., Hampshire, UK). The horizontal axis represents corresponding
glucose measurements obtained in plasma with the One Touch Ultra
(OTU; LifeScan, Milpitas, CA; a), whole blood with the OTU (b),
plasma with the YSI 2300 State Plus Glucose and Lactate Analyzer
(YSI; YSI Inc., Yellow Springs, OH; c), whole blood with the YSI
(d), or interstitial fluid with the Dexcom SEVEN continuous glucose
monitor (SEVEN; Dexcom, San Diego, CA; e).
-
Methods of measuring glucose in horses 211
interferences by uric acid, bilirubin, acetaminophen, dopamine,
and ascor-
bic acid (Hones et al., 2008). In the normal, healthy,
nonfasting, relatively
sedentary horse, one would not expect to see large quantities of
these inter-fering substances (Newton and Davis, 1922; Stillions et
al., 1971; Snow and Frigg, 1989; Engelking, 1993).
The major advantage of the OTU is its convenience. Although it
may be a better predictor of reference values when plasma is used,
the data presented here do not unequivocally indicate that using
plasma with the OTU would increase accuracy of readings relative to
those of the AA3. One caveat to using plasma with the OTU is that
the process of separating plasma from whole blood reduces the
convenience associated with the hand-held glucose meter and it
introduces a potential source of error.
Hand-held glucose meters intended for use in humans are designed
for samples containing 30 to 55% hema-tocrit (LifeScan
informational package insert, 2009). Under normal circum-
Figure 4. Scatter plots illustrating the agreement in raw
glucose concentrations obtained between measurement methods. For
each, the vertical axis (Y) represents glucose measurements
obtained in interstitial fluid using the Dexcom SEVEN continuous
glucose monitor (SEVEN; Dexcom, San Diego, CA). The horizontal axis
represents corresponding glucose measurements obtained in whole
blood with the One Touch Ultra (OTU; LifeScan, Milpitas, CA; a) and
plasma with the OTU (b).
Table 2. Estimated regression coefficients comparing equine
glucose concentrations obtained by various methodologies to glucose
values obtained in plasma by the AA31,2
Item
AA33
OTU3
Blood Plasma
OTU4 YSI4 SEVEN4 SEVEN4
Blood Plasma Blood Plasma Interstitial fluid Interstitial
fluid
Intercept 3.71 2.83 4.35 1.41 4.60 3.97 5.70 SE 0.45 0.43 0.65
0.72 0.38 0.45 0.49 P-value (H0: 0 = 0)
-
Slough et al.212
stances, the relaxed equine will have a similar hematocrit
range. Hackett and McCue (2010) reported that a veteri-nary glucose
meter tested in horses was accurate when median packed cell volume
was 38%, but when packed
cell volume was 66%, glucose val-ues were only about 52% of
values generated by their reference method. However, Hollis et al.
(2008) reported that packed cell volume had no effect on the
accuracy of the Accu-Chek
hand-held glucose meter in measur-ing glucose in equine whole
blood. As Hollis et al. (2008) and Russell et al. (2007) pointed
out, equine whole blood may impair the ability of test strips to
effectively filter erythrocytes
Table 3. BlandAltman comparisons of equine glucose
concentrations obtained by various methodologies to values obtained
by the AA31,2
Item
AA33
OTU3
Blood Plasma
OTU4 YSI4 SEVEN4 SEVEN4
Blood Plasma Blood Plasma Interstitial fluid Interstitial
fluid
Observations (n) 120 120 87 120 108 108 108diffmean 0.0206
2.1937 1.6681 0.0962 0.9570 0.9339 3.2515SD 1.18 1.31 1.19 0.73
1.60 1.63 1.64No. outside diffmean 2 SD
5 4 4 4 2 4 4
% outside diffmean 2 SD
4.2 3.3 4.5 3.3 1.9 3.7 3.7
1Methods of measuring glucose included measurement of plasma
glucose with an Autoanalyzer3 (AA3; Seal Analytical Ltd.,
Hampshire, UK), glucose in whole blood and plasma with a One Touch
Ultra (OTU; LifeScan, Milpitas, CA) hand-held glucometer, glucose
in whole blood and plasma with a YSI 2300 Stat Plus Glucose and
Lactate Analyzer (YSI; YSI Inc., Yellow Springs, OH), and
interstitial glucose with a Dexcom SEVEN continuous glucose monitor
(SEVEN; Dexcom, San Diego, CA).2BlandAltman calculations performed
on mean of Y minus mean of X (diffmean) with SD of differences.3Y
variable.4X variable.
Table 4. Lins concordance coefficient comparisons of equine
glucose concentrations obtained by various methodologies to values
obtained by the AA31,2
Item
AA33
OTU3
Blood Plasma
OTU4 YSI4 SEVEN4 SEVEN4
Blood Plasma Blood Plasma Interstitial fluid Interstitial
fluid
Observations (n) 120 120 87 120 108 108 108Lower CL 0.46 0.55
0.22 0.69 0.23 0.32 0.45Concordance coefficient 0.58 0.64 0.41 0.77
0.39 0.47 0.59Upper CL 0.68 0.72 0.57 0.84 0.52 0.60 0.701Methods
of measuring glucose included measurement of plasma glucose with an
Autoanalyzer3 (AA3; Seal Analytical Ltd., Hampshire, UK), glucose
in whole blood and plasma with a One Touch Ultra (OTU; LifeScan,
Milpitas, CA) hand-held glucometer, glucose in whole blood and
plasma with a YSI 2300 Stat Plus Glucose and Lactate Analyzer (YSI;
YSI Inc., Yellow Springs, OH), and interstitial glucose with a
Dexcom SEVEN continuous glucose monitor (SEVEN; Dexcom, San Diego,
CA).2Lins concordance coefficient calculations performed on mean +
MIXED residuals accounting for horseday variability, which was
usually not significant. Upper and lower confidence limits (CL) are
shown.3Y variable.4X variable.
-
Methods of measuring glucose in horses 213
because of the tendency of rouleaux formation in equine red
blood cells, or other factors such as microclot forma-tion,
hemolysis, viscosity, or inflam-matory markers.
There is considerable variability in the literature regarding
how accu-rately glucose values obtained from interstitial fluid
reflect blood glucose values. Whereas some have reported a 5-min
delay between blood and interstitial glucose concentrations
(Re-brin et al., 1999; Kulcu et al., 2003), others have reported
that interstitial glucose concentrations are approxi-mately 60%
lower in human adipose tissue compared with arterialized plasma
concentrations (Regittnig et al., 2003). Several published reports,
however, indicate that interstitial glucose in adipose tissue tends
to be 85 to 100% of arterialized plasma concentrations (Lonnroth et
al., 1987; Jansson et al., 1988; Bolinder et al., 1989; Moberg et
al., 1997; Rosdahl et al., 1998; Niklasson et al., 2000). Others
point out that a hypoglycemic environment may develop in
inter-stitial tissue preceding a decline in blood glucose values
(Pickup et al., 1999). Some speculate that differences between
interstitial glucose readings and blood glucose measurement may be
due to factors other than sensor sensitivity, such as insulin
sensitivity of the subject and vascularization of the sensor site
(Kulcu et al., 2003).
The potential to use a CGM to measure glucose concentration
every 5 min is extremely appealing. Horses in the current
experiment exhibited only mild and transient signs of discomfort
during sensor placement, and symp-toms of irritation around the
insertion site were minimal. Although there were horses that rubbed
their sen-sors on the fence, those occurrences were relatively
rare. If the CGM obtained accurate measurements and performed
dependably, it would be immensely useful as a research or clinical
tool. Although conclusions drawn in this experiment regarding the
ability of the SEVEN system to estimate AA3 values might vary from
one statistical analysis to another, the reproducibility of
measurements ob-
tained was inferior to that of the YSI when plasma was used.
Others have reported accurate readings obtained with CGM devices in
horses, although previous researchers validated CGM readings in
only 36 samples and used a different CGM device (Wiedmeyer et al.,
2003).
The ability to gather near-continu-ous glucose data in the
equine could be invaluable to researchers, but the technical
problems and frequent system failures experienced in this research
were an impediment to usage of the CGM in horses. Sensors are
advertised to last >5 d in humans, and although some sensors
functioned for longer than this in individual horses, many of them
ceased normal function by d 3. This required the insertion and
initialization of new sen-sors. The devices are expensive, and each
initialization takes several hours. Consequently, some of the
logistical issues in implementing these devices may limit their
usefulness to equine researchers or clinicians.
Because of the need to obtain imme-diate glucose readings for
calibration, readings obtained by the OTU using whole blood were
used for calibra-tion of the SEVEN system. Using a more accurate
method of measuring glucose concentrations for the cali-bration of
the SEVEN system may further decrease the bias of the read-ings
obtained by the CGM relative to AA3 readings, as might further
study into the effect of timing of calibration (Buckingham et al.,
2006). Calibrat-ing the SEVEN system with plasma glucose measured
by the OTU would be considerably less convenient than using whole
blood, and there is not overwhelming evidence, based on this
experiment, that there would be any advantage in doing so.
IMPLICATIONSIn this work, none of the methods
tested met the 2003 criteria estab-lished by the International
Organiza-tion for Standardization (ISO 15197). Consequently, if
absolute glucose values or statistically significant data are
required, as in clinical emergency
cases, a standard laboratory refer-ence technique will provide
the most accurate glucose measurements. However, all methods tested
detected a postprandial increase and then decrease in glucose
concentrations. If the objective is simply to determine trends or
make comparisons, then the hand-held glucometer or CGM may be
sufficient. If the SEVEN system is to be used, technical issues
will have to be addressed.
ACKNOWLEDGMENTSThe authors would like to thank
the staff at the K-State Beef Cattle Research Center and the
K-State Horse Unit (Kansas State University, Manhattan) for their
assistance with horses and facilities used in the proj-ect. As
well, students enrolled in ASI 661 Special Problems at Kansas State
University were tremendously helpful in the collection of glucose
samples.
LITERATURE CITEDBland, J., and D. Altman. 1986. Statistical
methods for assessing agreement between two methods of clinical
measurement. Lancet 1:307.
Bolinder, J., E. Hagstrom, U. Ungerstedt, and P. Arner. 1989.
Microdialysis of subcu-taneous adipose tissue in vivo for
continuous glucose monitoring in man. Scand. J. Clin. Lab. Invest.
49:465.
Buckingham, B. A., C. Kollman, R. W. Beck, A. Kalajian, R.
Fiallo-Scharer, M. J. Tansey, L. A. Fox, D. M. Wilson, S. A.
Weinzimer, K. J. Ruedy, W. V. Tamborlane, and D. S. Grp. 2006.
Evaluation of factors affecting CGMS calibration. Diabetes Technol.
Ther. 8:318.
Burnham, K. P., and D. R. Anderson. 1998. Model Selection and
Inference. Springer-Verlag, New York, NY.
Coffman, J. R., and C. M. Colles. 1983. Insulin tolerance in
laminitic ponies. Can. J. Comp. Med. 47:347.
Coldman, M. F., and W. Good. 1967. The distribution of sodium,
potassium and glucose in the blood of some mammals. Comp. Bio-chem.
Physiol. 21:201.
Engelking, L. R. 1993. Equine fasting hyper-bilirubinemia. Adv.
Vet. Sci. Comp. Med. 37:115.
Fingerhut, B., R. Ferzola, W. H. Marsh, and A. B. Miller. 1965.
Automated methods for blood glucose and urea. Clin. Chem.
11:803.
-
Slough et al.214
Forbes, N. A., and C. F. Brayton. 2008. Glu-cometer validity in
mouse research. J. Am. Assoc. Lab. Anim. Sci. 47:167.
Garcia, M. C., and J. Beech. 1986. Equine in-travenous
glucose-tolerance testGlucose and insulin responses of healthy
horses fed grain or hay and of horses with pituitary-adenoma. Am.
J. Vet. Res. 47:570.
Grosenbaugh, D. A., J. E. Gadawski, and W. W. Muir. 1998.
Evaluation of a portable clini-cal analyzer in a veterinary
hospital setting. J. Am. Vet. Med. Assoc. 213:691.
Gunkel, C. D., J. S. Drouillard, and T. L. Slough. 2009. The use
of a handheld glucom-eter for measuring glucose concentrations from
whole blood collected from the horse. J. Anim. Sci. 87(E-Suppl.
2):772.
Hackett, E. S., and P. M. McCue. 2010. Evaluation of a
veterinary glucometer for use in horses. J. Vet. Intern. Med.
24:617.
Hollis, A. R., B. L. Dallap Schaer, R. C. Boston, and P. A.
Wilkins. 2008. Compari-son of the Accu-Chek Aviva point-of-care
glucometer with blood gas and laboratory methods of analysis of
glucose measurement in equine emergency patients. J. Vet. Intern.
Med. 22:1189.
Hones, J., P. Muller, and N. Surridge. 2008. The technology
behind glucose meters: Test strips. Diabetes Technol. Ther.
10:S10.
Hussain, K., and N. Sharief. 2000. The inac-curacy of venous and
capillary blood glucose measurement using reagent strips in the
newborn period and the effect of haematocrit. Early Hum. Dev.
57:111.
ISO (International Organization for Stan-dardization). 2003. In
vitro diagnostic test systemsRequirements for blood-glucose
monitoring systems for self-testing in managing diabetes mellitus.
ISO 15197.
http://www.techstreet.com/standards/ISO/15197_2003?product_id=1095349.
Jansson, P. A., J. Fowelin, U. Smith, and P. Lonnroth. 1988.
Characterization by microdialysis of intracellular glucose level in
subcutaneous tissue in humans. Am. J. Physiol. 255:E218.
Jeffcott, L. B., and J. R. Field. 1985. Cur-rent concepts of
hyperlipemia in horses and ponies. Vet. Rec. 116:461.
Johnson, P. J. 2002. The equine metabolic syndrome peripheral
Cushings syndrome. Vet. Clin. North Am. Equine Pract. 18:271.
Kong, G. W. S., W. H. Tam, M. H. M. Chan, W. Y. So, C. W. K.
Lam, I. P. C. Yiu, K. M. Loo, and C. Y. Ki. 2010. Comparison in the
performance of glucose meters in blood glu-cose monitoring during
pregnancy. Gynecol. Obstet. Invest. 69:264.
Kozar, J., A. M. Simundic, N. Nikolac, M. Zirovic, and E. Topic.
2008. Accu Chek Compact Plus blood glucometer evaluation. Biochem.
Med. 18:361.
Kronfeld, D. S., K. H. Trieber, T. M. Hess, and R. C. Boston.
2005. Insulin resistance in the horse: Definition, detection, and
dietetics. J. Anim. Sci. 88:E22.
Kulcu, E., J. A. Tamada, G. Reach, R. O. Potts, and M. J. Lesho.
2003. Physiological differences between interstitial glucose and
blood glucose measured in human subjects. Diabetes Care
26:2405.
Lin, L. I.-K. 1989. A concordance coeffi-cient to evaluate
reproducibility. Biometrics 45:255.
Littell, R., G. Milliken, W. Stroup, and R. Wolfinger. 1996. SAS
for Mixed Models. SAS Inst. Inc., Cary, NC.
Lonnroth, P., P. A. Jansson, and U. Smith. 1987. A microdialysis
method allowing char-acterization of intercellular water space in
humans. Am. J. Physiol. 253:E228.
Louie, R. F., Z. P. Tang, D. V. Sutton, J. H. Lee, and G. J.
Kost. 2000. Point-of-care glu-cose testingEffects of critical care
variables, influence of reference instruments, and a modular
glucose meter design. Arch. Pathol. Lab. Med. 124:257.
Moberg, E., E. Hagstrom-Toft, P. Arner, and J. Bolinder. 1997.
Protracted glucose fall in subcutaneous adipose tissue and skeletal
muscle compared with blood during insulin-induced hypoglycaemia.
Diabetologia 40:1320.
Newton, E. B., and A. R. Davis. 1922. Combined uric acid in
human, horse, sheep, pig, dog, and chicken blood. J. Biol. Chem.
54:603.
Niklasson, M., A. Holmang, M. Sjostrand, L. Strindberg, and P.
Lonnroth. 2000. Muscle glucose uptake is effectively activated by
ischemia in type 2 diabetic subjects. Diabe-tes 49:1178.
Pickup, J., L. McCartney, O. Rolinski, and D. Birch. 1999. In
vivo glucose sensing for diabetes management: Progress towards
non-invasive monitoring. BMJ 319:1289.
Ralston, S. L. 1996. Hyperglycemia/hyper-insulinemia after
feeding a meal of grain to young horses with osteochondritis
dissecans (OCD) lesions. Pferdeheilkunde 12:320.
Rebrin, K., G. M. Steil, W. P. van Antwerp, and J. J.
Mastrototaro. 1999. Subcutaneous glucose predicts plasma glucose
independent of insulin: Implications for continuous moni-toring.
Am. J. Physiol. 277:E561.
Regittnig, W., M. Ellmerer, G. Fauler, G. Sendlhofer, Z.
Trajanoski, H. J. Leis, L. Schaupp, P. Wach, and T. R. Pieber.
2003. Assessment of transcapillary glucose ex-change in human
skeletal muscle and adipose tissue. Am. J. Physiol. Endocrinol.
Metab. 285:E241.
Rosdahl, H., K. Hamrin, U. Ungerstedt, and J. Henriksson. 1998.
Metabolite levels in human skeletal muscle and adipose tissue
studied with microdialysis at low perfusion flow. Am. J. Physiol.
274:E936.
Russell, C., J. E. Palmer, R. C. Boston, and P. A. Wilkins.
2007. Agreement between point-of-care glucometry, blood gas and
laboratory-based measurement of glucose in an equine neonatal
intensive care unit. J. Vet. Emerg. Crit. Care 17:236.
SAS Institute Inc. 2007. SAS OnlineDoc 9.1.2. SAS Inst. Inc.,
Cary, NC. http:// support.sas.com/onlinedoc/912/
docMainpage.jsp.
Snow, D. H., and M. Frigg. 1989. Oral-admin-istration of
different formulations of ascorbic-acid to the horse. J. Equine
Vet. Sci. 9:30.
Stillions, M. C., S. M. Teeter, and W. E. Nelson. 1971. Ascorbic
acid requirement of mature horses. J. Anim. Sci. 32:249.
Stockham, S. L. 1995. Interpretation of equine serum biochemical
profile results. Vet. Clin. North Am. Equine Pract. 11:391.
Treiber, K. H., D. S. Kronfeld, and R. J. Geor. 2006. Insulin
resistance in equids: Pos-sible role in laminitis. J. Nutr.
136:2094S.
Wentholt, I. M., J. B. Hoekstra, and J. H. Devries. 2007.
Continuous glucose monitors: The long-awaited watch dogs? Diabetes
Technol. Ther. 9:399.
Wiedmeyer, C. E., P. J. Johnson, L. A. Cohn, and R. L. Meadows.
2003. Evaluation of a continuous glucose monitoring system for use
in dogs, cats, and horses. J. Am. Vet. Med. Assoc. 223:987.