07/05/22 The GC-in-a-PC 1 Pneumatic Focusing Gas Chromatography PFGC Robert O’Brien, Portland State Univ. & VOC Technologies The GC-in-a-PC Development funded by USEPA SBIR Phase 1 & 2 Grants
Jan 06, 2018
05/03/23 The GC-in-a-PC 1
Pneumatic Focusing Gas Chromatography PFGC
Robert O’Brien, Portland State Univ. & VOC Technologies
The GC-in-a-PC Development funded by USEPA SBIR Phase 1 & 2 Grants
05/03/23 The GC-in-a-PC 2
The PFGC technology was developed with SBIR (Small Business Innovative Research) from the US Government: NIH (National Institutes of Health) &
USEPA (Environmental Protection Agency.
The technology has 5 US patents.
PFGC Pneumatic Focusing Gas Chromatography Automated, Internet Monitoring of VOCs
05/03/23 The GC-in-a-PC 3
Pneumatic Focusing Gas ChromatographyPrinciples and Applications
Take a large volume air sample (e.g. 300 cc) and compress it to high pressure (e.g. 600 psi).
This concentrates the sample 30x & condenses the water, which remains in the sample loop.
Inject directly on-column, maintained at 600 psi. Build the whole thing inside a commercial PC. Connect it to the internet. Run it anywhere in the world from your office.
05/03/23 The GC-in-a-PC 4
PFGC Basic Principle
Separate VOCs by Pneumatic Focusing GC
Retention time (seconds)>
05/03/23 The GC-in-a-PC 5
Pneumatic Focusing Gas Chromatography: Features & Benefits
PFGC Completely enclosed - standard PC case
MS Windows w/ sophisticated VOCTEC OS
Data taken remotely, analyzed, transmitted--anywhere in the world via the Internet.
Easy set up & calibration. Two levels of peak identification
and integration, semi-manual & fully automated
On-line training, debugging, control and performance monitoring.
US Patent 6,865,926Below is the 1st PFGC built in a tower PC in 2001
05/03/23 The GC-in-a-PC 6
Portland Air ToxicsA section of 30 or so successive chromatograms.
isoPentane and Toluene have been integrated so far.
Hourly chromatograms in Portland air 1500-3600 seconds
05/03/23 The GC-in-a-PC 7
Pneumatic Focusing Gas Chromatography: Features & Benefits
PFGC Completely enclosed - standard PC case
MS Windows w/ sophisticated VOCTEC software
Data taken remotely, analyzed, transmitted anywhere in the world via the Internet.
Easy set up & self-calibrating. Very low cost per sample. 24/7 automated analysis & alert.
US Patent 6,865,926
Benzene from autos always present PCE rarely present
05/03/23 The GC-in-a-PC 8
Continuous 1-MonthToluene Record, NW Portland, Oregon USALow concentration period due to strong Pacific Ocean winds
Hourly samples, 24 per day
05/03/23 The GC-in-a-PC 9
Benzene/Toluene ratio for one month, showing mostly automobile sources with occasional non-vehicular source. Each point an hourly sample
Hourly samples, 24 per day
05/03/23 The GC-in-a-PC 10
05/03/23 The GC-in-a-PC 11
Comparison with Commercial VOC GCMS laboratory Benzene
05/03/23 The GC-in-a-PC 12
05/03/23 The GC-in-a-PC 13
Correlation of BTEX (Benzene, Ethylbenzene & Xylenes) with Toluene in Portland, Oregon US. Most comes from autos in Portland Oregon which has stationary source
controls and all BTEX have strong correlation.
Toluene (ppb)
05/03/23 The GC-in-a-PC 14
Correlation of BTEX (Benzene, Ethylbenzene & Xylenes) with Toluene in Boise, Idaho USA. Boise does not have stationary source controls & BTEX correlation is weaker.
BTEX = Benzene Toluene Ethylbenzene Xylenes(3)
Toluene (ppb)
05/03/23 The GC-in-a-PC 15
Same Portland data shown on same scale as Boise in previous slide. Note much lower concentrations and good correlation with automobile source when stationary VOC
sources are controlled in Portland but not Boise.
BTEX = Benzene Toluene Ethylbenzene Xylenes(3)
Toluene (ppb)
05/03/23 The GC-in-a-PC 16
Rack-Mounted (Pneumatic-Focusing Gas Chromatograph (PFGC) for VOC Measurement in Monitoring Station or
industrial plant.
05/03/23 The GC-in-a-PC 17
Interior of Pneumatic-Focusing Gas Chromatograph PFGC
Interior of the GC-in-a-PC
PC Power Supply
Sample Valve
Flame Ionization Detector
PC Mother Board
PFGC Oven
Hard Drive & Main Control Board
Sample Pump
Carrier & FID Gases
05/03/23 The GC-in-a-PC 18
Hourly daily chromatograms. Chromatograms start at midnight. A pollution episode starts about 2AM. A blank chromatogram (green) is done and then two calibrations (blue). After this the air becomes very clean again (yellow). The contamination in the blank run was due to a dirty sample line installed by the station technician. It was removed when the sample line was replace.
05/03/23 The GC-in-a-PC 19
Dual detector PFGC. One GC column, two detectors: FID (Flame Ionization Detector) & Agilent SCD (Sulfur Chemiluminescence Detector) To accommodate the SCD the case must be larger.
PFGC SCD Calibration Gas Computer Monitor
The computer allows the technician to operate the PFGC. The monitor is not necessary to run the PFGC over the Internet with the same screen in your office or lab.
05/03/23 The GC-in-a-PC 20
One day PFGC chromatograms. Alberta OilSands
1st chromatogram after midnight is calibration. Multiple VOCs at 10 ppb.
Automatic daily calibration, or on-demand over the internet.
05/03/23 The GC-in-a-PC 2105/03/23 The GC-in-a-PC 21
Specification & Applications
Concentration Range: 50ppt to 5000ppm depending on the configuration.
Chemical volatility Range: C1 to C10 (Higher C available)
Speciated VOCs or just Methane & Nonmethane
Questions?