1 MRI Physics 1: Image Acquisition Chris Rorden & Paul Morgan – Magnetic Resonance – Radio frequency absorption – Relaxation: Radio frequency emission – Gradients – Increases signal/noise: antenna selection, field strength. Excellent references: – www.cis.rit.edu/htbooks/mri/ – www.easymeasure.co.uk/principlesmri.aspx
34
Embed
1 MRI Physics 1: Image Acquisition Chris Rorden & Paul Morgan –Magnetic Resonance –Radio frequency absorption –Relaxation: Radio frequency emission –Gradients.
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
1
MRI Physics 1: Image Acquisition
Chris Rorden & Paul Morgan– Magnetic Resonance– Radio frequency absorption – Relaxation: Radio frequency emission– Gradients– Increases signal/noise: antenna selection, field strength.
Tesla and Gauss are measures of magnetic field strength– Earth’s magnetic field ~0.5 Gauss.– 1Tesla = 10,000 Gauss.– Our fMRI system is 3T.
~x60,000 earth’s field strength
3
Anatomy of an atom
Atoms are the building blocks of our world. Composed of 3 components
– Electrons: tiny negatively charged particles.– Protons: heavy positively charged particles.– Neutrons: heavy particles without charge.
Electrons are often thought of like planets: tiny objects distantly orbiting a massive core.
Neutrons and protons form the dense nucleus of an atom.
4
Atomic Nuclei Atomic nuclei are composed of protons and neutrons. Protons determine the element, e.g. hydrogen has one proton, helium has two. Neutrons determines the isotope.
– Helium: Usually, two neutrons: 4He. Rarely, only have one neutron: 3He.
– Hydrogen Usually, no neutrons: 1H. Rarely, Deuterium (2H) is found in ‘heavy’ water Tritium (3H) is radioactive (spontaneous decay).
4He 3He 1H 2H
5
Nuclear Magnetic Resonance
Felix Block and Edward Purcell– 1946: the nuclei of some elements absorb and re-emit radio frequency energy
when in magnetic field– 1952: Nobel prize in physics
Atoms with odd number of protons/neutrons spin in a magnetic field– Nuclear: properties of nuclei of atoms– Magnetic: magnetic field required– Resonance: interaction between magnetic field and radio frequency
Bloch Purcell
6
How fast do hydrogen atoms spin?
Field Strength (Tesla)
180
401.0 4.0
The rate of rotation is determined by the strength of the magnetic field.
Larmor equationf = B0
For Hydrogen, = 42.58 MHz/T
At 1.5T, f = 63.76 MHzAt 3T, f = 127.7 MHz
Res
on
ance
Fre
qu
ency
(M
Hz)
7
Radiofrequency Pulses
A radiofrequency (RF) pulse at the Larmor frequency will be absorbed.
This higher energy state tips the spin, so it is no longer aligned to the field.
An RF pulse at any other frequency will not influence the nuclei.
This is resonance.
8
Source of MR signal
At rest, atoms align parallel to magnetic field.– Lowest energy state.
After RF absorption, net magnetization is orthogonal to magnetic field– RF emission at larmor frequency– RF emission decays with time
9
Absorption and Relaxation
Our RF transmission is absorbed by atoms at Larmor frequency.
After the RF pulse, atoms will begin to realign with the magnetic field: relaxation
During this period, an RF signal is emitted.This signal will be at the Larmor frequency.An antenna can measure this signal.
10
Hydrogen is the mainstay for MRI
We will focus on Hydrogen– Hydrogen abundant in body (63% of atoms).– Elements with even numbers of neutrons and protons have no
spin, so we can not image them (4He, 12C).– 23Na and 31P are relatively abundant, so can be imaged.
Therefore, by sending in a RF pulse at a specific frequency we can selectively energize hydrogen.
11
Electromagnetic Spectrum
MRI signals are in the same range as FM radio and TV (30-300MHz). MRI frequency is non-ionizing radiation, unlike X-rays. Absorbed RF will cause heating. Specific absorption rate (SAR): measure of the energy absorbed by tissue.
– Increases ~ with square of field strength.– Higher SAR = more energy = more signal = more heating– FDA limits SAR, and is a limiting factor for some protocols (3 W/kg averaged over 10
minutes).
Ioni
zing
Rad
iatio
nB
reak
s B
onds
Non
-Ion
izin
g R
adia
tion
Hea
ting
Excites Electrons
Excites Nuclei
12
Making a spatial image
To create spatial images, we need a way to cause different locations in the scanner to generate different signals.
To do this, we apply gradients. Gradients make the magnetic field slightly stronger
at one location compared to another. Lauterbur: first MRI: 2003 Nobel Prize.
Lauterbur
13
Slice Selection Gradient
Gradients make field stronger at one location compared to another.
Larmor frequency different along this dimension.
RF pulse only energizes slice where field strength matches Larmor frequency.
Fie
ld S
tren
gth
Z Position
128
Mh
z
127
Mh
z
126
Mh
z127 MhzRF pulse
LarmorFreq
14
Slice Selection Gradient
Gradual slice selection gradients will select thick slices, while steep gradients select thinner slices.
– The strength of your scanner’s gradients can limit minimum slice thickness.– FDA limits speed of gradient shift (dB/dt) and some of our protocols can
elicit slight tingling sensation or brief muscle twitches. Position of gradient determines which 2D slice is selected.
Fie
ld S
tren
gth
Z Position
Fie
ld S
tren
gth
Z Position
Fie
ld S
tren
gth
Z Position
Fie
ld S
tren
gth
Z Position
15
Phase encoding gradient
Orthogonal gradient applied between RF pulse and readout This adjusts the phase along this dimension. Analogy: Phase encoding is like timezones. Clocks in
different zones will have different phases.Y
Pos
ition
Field Strength
16
Frequency encoding gradient
Apply final orthogonal gradient when we wish to acquire image.
Slice will emit signal at Lamour frequency, e.g. lines at higher fields will have higher frequency signals.
Aka ‘Readout gradient’.
X P
ositi
on
Field Strength
17Source: Traveler’s Guide to K-space (C.A. Mistretta)
Raw MRI image: K-Space
Raw MRI is 2D Frequency Map.Apply Fourier Transform to reconstruct image.
18
Reconstruction
Medical scanners automatically reconstruct your data. You can manually reconstruct data using Tokarczuk’s
Slices can be either collected sequentially (e.g. ascending axial slices) or in interleaved order (e.g. collect all odd slices, then all even slices).
……
Sequential Interleaved
Time Time
26
Sequential vs Interleaved Volumes
Interleaving reduces interference between slices. Useful if thin gap between slices.
Unfortunately, any head movements will cause worse ‘spin history’ effects for interleaved slices than sequential. (TR will be shorter for regions that were previously in another slice).
You must know slice order if you want to ‘slice time correct’ data (temporal processing lecture).
Slice 4
Slice 3
Slice 2
Slice 1
Slice 4
Slice 3
Slice 2
Slice 1
Slice 4
Slice 3
Slice 2
Slice 1
27
Signal to Noise
Signal To noise is given by the formulaSignal = VN
Where V is the volume and N is the number of samples averaged (referred to as ‘Nex’, as in ‘number of excitations’).
For example, to get the same SNR as a single 3x3x3mm scan (27mm3) we would need to collect 12 2x2x2mm (8mm3) scans or 768 1x1x1mm (1mm3) scans.
28
Signal to Noise: Antennas
Volume coil Surface coilHead coil
www.fmrib.ox.ac.uk/~karla/
Surface coil
The MRI antenna is called a coil.We use different coils for different body parts.For brains, the most common antenna is the head
coil, which is a volume coil: it shows the whole brain.We can also use a surface coil: it gives great signal
for a small field of view.
29
Parallel Imaging (SENSE, iPat)
Parallel imaging uses multiple surface coils to generate a volume image.
Dramatically reduces spatial distortion and increases signal.
Optionally, you can acquire images more rapidly by only collecting a portion of k-space.– SENSE R=2 collects half of the lines.– SENSE R=3 collects one third– Reduces spatial distortion and increases speed
of acquisition. Some loss in signal.
8-channel array
30
Parallel Imaging (SENSE, iPat)
Increasing SENSE reduction factor decreases acquisition time and spatial distortion, but high values lead to reduced signal.
Eff
ect
s of
SEN
SE f
act
or
(R)
on E
PI
R=1 R=2 R=3
31
Outside magnetic field •Spins randomly oriented
In magnetic field:•Spins tend to align parallel or anti-parallel to magnetic field.•At room temperature, ~4 parts per million more protons per Tesla align with versus against field.•As field strength increases, there is a bigger energy difference between parallel and anti-parallel alignment (faster rotation = more energy).
•A larger proportion will align parallel to field.•More energy will be released as nuclei align.•Therefore, MR signal increases with square of field strength.
Signal and Field Strength
32
Signal and Field Strength
Most clinical MRI: 1.5T Our fMRI systems: 3.0T Maximum for NbTi MRI ~11.7T World record for superconductive MRI ~22T (Nb3Sn) Field strength influences:
– Faster Larmor frequency : Bigger energy difference between parallel and anti-parallel alignment
• Larger ratio of nuclei aligned = more signal• More signal as nuclei realign.
Reduced TR and TE: less time to take images (next week).
33
Signal and Field Strength In theory:
– Signal increases with square of field strength– Noise increases linearly with field strength– A 3T scanner should have twice SNR of 1.5T scanner; 7T should have ~4.7
times SNR of 1.5T. Unfortunately, physiological artifacts also increase, so advantage is less in