Top Banner

of 24

1-5 Units Study Materials

Aug 07, 2018

Download

Documents

lvrevathi
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/20/2019 1-5 Units Study Materials

    1/61

     

    Powered By www.technoscriptz.com

    DEPARTMENT OF ELECTRONICS AND INSTRUMENTATION

    ENGINEERING

    EI2251 INDUSTRIAL INSTRUMENTATION-1

    SEMESTER :IV

    BRANCH : EIE

    Lesson Planning Sheet

    Sub

    Code

    Name of the

    Subject

    Sem Department of

    Electronics &

    Instrumentation

    No. of

    Student

    s

    Time

    L T P

    EI2251 INDUSTRIALINSTRUMENTATION-1 

    IV Name of Faculty:

    M.THIRUMAGAL

    40 3 0 0

    L-Lecture, T-Tutorial, P-Practical

    Sl.

    No

    Lesson / Topic

    Covered

    (each lecture session

    wise)

    Methodology

      P  r  a  c  t  i  c  a  l   /

      G  u  e  s  t

      L  e  c  t  u  r  e

      I  n  d  u  s  t  r  i  a  l   V  i  s  i  t

    Theory Coverage Tutorial Support

    Black

    -

    Board

    OHPPower

    Point

    Exercis

    e

    Assignment

    s

  • 8/20/2019 1-5 Units Study Materials

    2/61

     

    Powered By www.technoscriptz.com

    1. 1. MEASUREMENT OFFORCE, TORQUE AND

    VELOCITY

    INTRODUCTION 

     

    2.  Electric balance – Different types of load

    cells 

     

    3. Hydraulic, pneumaticstrain gauge 

     

    4. Magneto elastic and Piezoelectric load cell 

     

    5. Different methods oftorquemeasurements:introduction 

     

    6. strain gauge-Relative angulartwist-   

    7. Speed measurement:- -Stroboscope.

     

    8. Capacitive tacho   

    9. Dragcup type tacho-D.C

    Ac tacho    

    Unit 2

    1 MEASUREMENT OFACCELERATION,VIBRATION AND

    DENSITY- introduction 

     

    2 Accelerometers:- LVDT,Piezo-electric,

    typeaccelerometer 

     

    3 Strain gauge and Variable

    reluctance type 

     

  • 8/20/2019 1-5 Units Study Materials

    3/61

     

    Powered By www.technoscriptz.com

    4 Mechanical type vibrationinstruments – 

       

    5 Seismic instruments asanaccelerometer – 

       

    6 Vibrometers : Calibrationof vibration pickups – 

     

    7 Units of density andspecific gravity – Baume

    scale, and API scale 

       

    8 Pressure head typedensitometers- Floattype

    densitometers 

     

    9 –  Ultrasonicdensitometer- Bridge

    type gas densitometer.

       

    Unit 3

    1 PRESSUREMEASUREMENTintroduction 

     

    2 Units of pressure-Manometers-

    introduction- -

     

    3 Different types –Elastictype pressure gauges:

    Bourdontube, bellows and

    diaphragms- 

     

     

    4 Elastic elements withLVDT and straingauges –

    Capacitive type pressure

    gauge –Piezo-resistive

    pressure sensor- 

     

  • 8/20/2019 1-5 Units Study Materials

    4/61

     

    Powered By www.technoscriptz.com

    5 Resonator pressuresensor 

     

    6 Measurement of vacuum:-McLeod gauge-Thermal

    conductivity gauges

     

     

    7 - Ionization gauges:– Coldcathode type and hot

    cathode 

     

    8 Testing and calibration ofpressure gauges 

     

    9 -

    Dead weight tester. 

     

    10 Electrical methods  

    Unit4

    1 Temperaturemeasurement-intro

     

    2Definitions and

    standards-Primary and

    secondary fixed points 

     

    3 Calibrationofthermometers 

     

    4 Different types of filledin system thermometer- 

     

    5Sources of errors in

    filledin systems and their

    compensation 

     

    6 – Electrical methods oftemperature

    measurement 

     

    7 Signal conditioning ofindustrial Bimetallic

    thermometers 

       

  • 8/20/2019 1-5 Units Study Materials

    5/61

     

    Powered By www.technoscriptz.com

    8 3 lead and 4 lead RTDs   

    9 Thermistors.   

    10Bimetallic thermometers

    and theircharacteristics-  

    Unit 5

    1THERMOCOUPLES AND

    RADIATION

    PYROMETERS- intro 

     

    2 Thermocouples-Laws ofthermocouple Response of

    thermocouple –––-– 

     

    3 Fabrication of industrialthermocouples – 

    Signalconditioning of

    thermocouple output- 

     

    4 –Isothermal blockreference junctions – 

    Commercial circuits for

    cold junction

    compensation- 

     

    5 Special techniquesformeasuring high

    temperature using

    thermocouples 

       

    6 Radiation fundamentals Radiation methods of

    temperature

    measurement 

       

    7 Total radiationpyrometers-Optical

    pyrometers- 

     

  • 8/20/2019 1-5 Units Study Materials

    6/61

     

    Powered By www.technoscriptz.com

    8 Fiber optic temperaturemeasurement 

     

    9 Two colour radiationpyrometers 

     

    UNIT 1

    MEASUREMENT OF FORCE, TORQUE AND VELOCITY

    AIM:

    Discussion of load cells, torque meter and various velocity pick-ups.

    KEY WORDS:

    Load cell- strain gauge- torque measurement- torque meters- speed measurement-

    Tachogenerators-stroboscope

  • 8/20/2019 1-5 Units Study Materials

    7/61

     

    Powered By www.technoscriptz.com

    UNIT-1

    MEASUREMENT OF FORCE , TORQUE AND VELOCITY

    Force may be defined as a cause that produces resistance or obstruction to any moving body, or

    changes the motion of a body, or tends to produce these effects. Force is given as F=MA.

    Force measurement is also done by electric means in which the force is first converted into

    displacement at an elastic element and the displacement is measured 

    A vector quantity has both magnitude and direction

    Units of force

    S.I unit= Newton (N)

  • 8/20/2019 1-5 Units Study Materials

    8/61

     

    Powered By www.technoscriptz.com

    ONE NEWTON:

    The force capable of giving a mass of one Kg an acceleration of one meter per second

    Types of forces

    Frictional force:

    Friction is a surface force that opposes relative motion

    Tensional force:

    Tension is the magnitude of the pulling force exerted by a string, cable, chain, or similar

    object on another object. Measured in newtons (or sometimes pounds-force)

    Compression force:

    Opposite of tension

    Elastic force:

    Elastic force is the physical property of a material that returns to its original shape after

    the stress

    Various methods of measuring force:

    1.  Balancing on standard mass, either directly or through levers

    2.  Measuring acceleration of the body if its mass is known on which the unknown force is

    applied

    3.  Balancing against a magnetic force of a current-carrying coil and a magnet

    4.  Transducing the force to fluid pressure and then measuring the pressure.

    5. 

    Force to elastic member and measuring the resulting deflection

    6.  Measuring the change in precession of a gyroscope caused by an applied torque due to

    applied force

    7. 

    Measuring the change in natural frequency of a wire tensioned by the force.

    Measurement methods

    1.  Direct method

    2. 

    Indirect method

    Measuring devices: Different types of electrical type force transducers are given below

    Force gauge-Load cell etc.,

    http://en.wikipedia.org/wiki/Physical_propertyhttp://en.wikipedia.org/wiki/Stress_(mechanics)http://en.wikipedia.org/wiki/Stress_(mechanics)http://en.wikipedia.org/wiki/Physical_property

  • 8/20/2019 1-5 Units Study Materials

    9/61

     

    Powered By www.technoscriptz.com

    Spring scale

    Strain gauge

    Load cell

    A load cell is a transducer that is used to convert a force into electrical signal

    This conversion is indirect and happens in two stages

      Stage-1: The force being sensed deforms a strain gauge.

      Stage-2:The strain gauge converts the deformation (strain) to electrical signals

    Types of Load cell:

    1.  Hydrostatic load cell

    2. 

    Pneumatic load cell

    3.  Magneto elastic load cell

    4.  Piezo electric load cell

    Hydrostatic Load cell:

    The force is made to exerted on the load platform which is connected to the diaphragm

    The diaphragm seals the chamber filled with fluid connected to bourdon gauge.

    During the measurement process the applied force pressurizes the oil which in turn

    activate the burdon gauge and the needle connected to it indicates the magnitude of the pressure

    exerted

    Full load deflection:;0.05 mm

    Measurement range:0-20 Tonnes

    Tare compensation of 0.2MPa is done

    Used in static measurement

    Also called as plunger

  • 8/20/2019 1-5 Units Study Materials

    10/61

     

    Powered By www.technoscriptz.com

    Pneumatic load cell:

    The force is applied to the platform connected to the sealing diaphragm of air chamber

    The applied force is measured by means of flapper nozzle principle described as follows

    The chamber is equipped with constant air supply

    The platform acts like a flapper and creates the backpressure on through the nozzle in thechamber according to the force applied on it

    This counter balance the platform and equilibrium is attained the value of pressure inside

    the chamber indicated on meter gives the value of the force applied

    If the

    Applied mass = W

    Output pressure= p

    Diaphragm stiffness=k s

    Flapper-nozzle gain=k f

    Area of diaphragm=α 

    p = W / (k s/k f  + α)

    The force is made to exert on the load platform which in turn compresses the fluid closed

     by the diaphragm resulting in deflection of meter pointer

    Used in static measurement

    Magneto elastic load cell:

  • 8/20/2019 1-5 Units Study Materials

    11/61

     

    Powered By www.technoscriptz.com

    Also called as pressductor load cell

    Principle:

    Stress on ferromagnetic material alters the magnetic moments resulting in thechange in permeability of the material.

    The change in permeability is directly proportional to the applied force/stress.

    Working:

    Primary and secondary windings are wounded at right angles on diagonally drilled

    hole pairs on the transducer body enclosed with laminated sheets of ferromagnetic material

    Secondary windings remain undisturbed under no load condition. On load condition the

    angle between the primary and secondary changes the resulting flux linkage is given by

    Ф=αB cosθ 

    α = C.S.A of material 

    Ф = Total flux linkage 

    B= Magnetic flux density

    Cosθ = change in angle 

    Piezo Electric Load Cell:

    and induces a voltage proportional to the force applied

    es= -n dФ /dt 

    n = turn ratio (n2/n1)

    Unlike

    strain gages that can measurestatic forces, piezoelectric

    force sensors are mostly usedfor dynamic- force

    measurementssuch as oscillation,

    impact, or high speed

    compression or tension. Any

    force applied to the

     piezoelectric sensing element produces a separation of

    charges within the atomic

  • 8/20/2019 1-5 Units Study Materials

    12/61

     

    Powered By www.technoscriptz.com

    structure of the material, generating an electrostatic output voltage. The polarity of the voltage

    generated depends on the atomic structure of the material and the direction in which the force is

    applied.

    However, any leakage path lets electrons redistribute across the material, dropping the

    voltage output back to zero. Internal leakage paths are formed by impurities within the crystal

    while external paths are created by the electronics used to measure the voltage generated. Allleakages must be considered to determine the discharge time constant (DTC). The DTC typically

    follows an exponential curve similar to an RC time constant and is used to determine the sensor’s

    lowest frequency response.

    In a typical quartz-based force sensor, a charge-collection electrode is sandwiched

     between two quartz-crystal elements. The quartz elements are oriented to supply the same polarity voltage to the electrode when compressed, while the opposite polarity is applied to the

    sensor housing. This assembly resides between two mounting disks held together by an elastic, beryllium-copper stud and then weld-sealed within the enclosure to prevent contamination. The

    stud preloads the quartz elements to assure all parts are in intimate contact and to provide goodlinearity and tensile-force measurements.

    When a force is applied to the impact cap, the quartz elements generate an output voltage

    which can be routed directly to a charge amplifier or converted to a low-impedance signal within

    the sensor. The use of the direct sensor output demands that any connector, cable, and charge

    amplifier input must maintain a high insulation resistance on the order of >10≠″ Ω.

    Low-impedance quartz sensors have an internal MOSFET amplifier. Its output is a low-impedance voltage signal that uses standard cabling. However, force sensors with internal

    amplifiers do require external power to operate the amp.

    TORQUE MEASUREMENT

    The force which tends to change the linear motion or rotation of a body.

    It is also defined as the turning or twisting moment of a force about an axisT= FX D

    T=Torque

    F=Force

    D=perpendicular distance from the axis of rotation of the line of action of the force

    Methods of measurement:

    1. 

    Inline rotating sensor based torque measurement

    2.  Inline stationary sensor based torque measurement

    In line rotating sensor based torque measurement:

  • 8/20/2019 1-5 Units Study Materials

    13/61

     

    Powered By www.technoscriptz.com

    Strain gauge based measurement:

    A strain gage can be installed directly on a shaft. Because the shaft is rotating, the torque sensorcan be connected to its power source and signal conditioning electronics via a slip ring. The strain

    gage also can be connected via a transformer, eliminating the need for high maintenance sliprings. The excitation voltage for the strain gage is inductively coupled, and the strain gage output

    is converted to a modulated pulse frequency (Figure 6-5). Maximum speed of such anarrangement is 15,000 rpm.

    Strain gages also can be mounted on stationary support members or on the housing itself. These

    "reaction" sensors measure the torque that is transferred by the shaft to the restraining elements.

    The resultant reading is not completely accurate, as it disregards the inertia of the motor.

    Strain gages used for torque measurements include foil, diffused semiconductor, and thin filmtypes. These can be attached directly to the shaft by soldering or adhesives. If the centrifugal

    forces are not large--and an out-of-balance load can be tolerated--the associated electronics,

    including battery, amplifier, and radio frequency transmitter all can be strapped to the shaft.

    Torque measurement by relative angular twist method by proximity probe type

    Proximity and displacement sensors also can detect torque by measuring the angular

    displacement between a shaft's two ends. By fixing two identical toothed wheels to the shaft at

  • 8/20/2019 1-5 Units Study Materials

    14/61

     

    Powered By www.technoscriptz.com

    some distance apart, the angular displacement caused by the torque can be measured. Proximity

    sensors or photocells located at each toothed wheel produce output voltages whose phase

    difference increases as the torque twists the shaft.

    Torque measurement by relative angular twist method by optical type

    Another approach is to aim a single photocell through both sets of toothed wheels. As torquerises and causes one wheel to overlap the other, the amount of light reaching the photocell is

    reduced.

    SPEED MEASUREMENT:

    Speed is defined as rate of change of position of an object with respect to time.

    Units of speed

     

    Meters per second (symbol m s−1 or m/s), the SI derived unit; 

      Kilometers per hour (symbol km/h);

      Miles per hour (symbol mph);

      Knots (nautical miles per hour, symbol kn or kt);

      Feet per second (symbol fps or ft/s);

      Mach number, speed divided by the speed of sound; 

      The speed of light in vacuum (symbol c) is one of the natural units: 

      Revolution per minute (rpm)

    Measuring methods:

    Generally speed is calculated using tachometers which calculates the angular speed in

    revolution per minute(rpm) of the object and converted into the form required

    Types of Tachometer:

    1.  Mechanical tachometer: Associated only with mechanical units to measure speed

    2. 

    Electrical tachometer: Associated with transducer for converting rotational speed to

    electrical quantity.

    Capacitive tacho:

    http://en.wikipedia.org/wiki/Meters_per_secondhttp://en.wikipedia.org/wiki/SI_derived_unithttp://en.wikipedia.org/wiki/Kilometers_per_hourhttp://en.wikipedia.org/wiki/Miles_per_hourhttp://en.wikipedia.org/wiki/Knot_(nautical)http://en.wikipedia.org/wiki/Nautical_milehttp://en.wikipedia.org/wiki/Foot_per_secondhttp://en.wikipedia.org/wiki/Mach_numberhttp://en.wikipedia.org/wiki/Speed_of_soundhttp://en.wikipedia.org/wiki/Speed_of_lighthttp://en.wikipedia.org/wiki/Natural_unitshttp://en.wikipedia.org/wiki/Natural_unitshttp://en.wikipedia.org/wiki/Speed_of_lighthttp://en.wikipedia.org/wiki/Speed_of_soundhttp://en.wikipedia.org/wiki/Mach_numberhttp://en.wikipedia.org/wiki/Foot_per_secondhttp://en.wikipedia.org/wiki/Nautical_milehttp://en.wikipedia.org/wiki/Knot_(nautical)http://en.wikipedia.org/wiki/Miles_per_hourhttp://en.wikipedia.org/wiki/Kilometers_per_hourhttp://en.wikipedia.org/wiki/SI_derived_unithttp://en.wikipedia.org/wiki/Meters_per_second

  • 8/20/2019 1-5 Units Study Materials

    15/61

     

    Powered By www.technoscriptz.com

    It uses the principle of charging the capapcitor and discharging through a meter alternately . If

    the charging and discharging is controlled by the speed of the equipment, the average discharge

    current would be ppl to the speed, if (ω)omega is the speed of rotation, I=C R (ω) 

    Drag cup tachometer:

    This type is very common in rotational speed measurement. The source angular speed

    rotates a permanent magnet. An aluminum disc or cup is held close to the rotating magnet

    restrained by a control spring. When the magnet rotates eddy current is set up in the drag cup or

    disc and a torque is produced which tries to oppose the field produced by the eddy current. The

    cup is thus dragged or rotated in the direction of the rotating magnet. Due to the restraining action

    of the spring and angular rotation is indicated by the pointer which is proportional to speed. 

    D.C Tachogenerators

    The transducer that converts speed of rotation directly into electrical signal is aninduction pickup such a tachometer is more commonly used for speed cpntrol rotating equipments

  • 8/20/2019 1-5 Units Study Materials

    16/61

     

    Powered By www.technoscriptz.com

    Here the dc generator with the output voltage from the commutator is directly proportional to the

    speed measured

    A.C Tacho generators:

    Here the operation is similar to the dc tachometer but the magnet rotates in the

    stationary coil proportional to the speed to be measured in a stationary coil and generates a a.c

    voltage which is signal conditioned and displayed in the units of speed

    Stroboscopic method

    Stroboscopes are used to measure the speed of rotation or frequency of vibration of a

    mechanical part or system. They have the advantage over other instruments of not loading ordisturbing the equipment under test. Mechanical equipment may be observed under actual

    operating conditions with the aid of stroboscopes. Parasitic oscillations, flaws, and unwanteddistortion at high speeds are readily detected. The flashing-light stroboscopes employ gas

    discharge tubes to provide a brilliant light source of very short duration.

    http://www.answers.com/topic/vibrationhttp://www.answers.com/topic/unwantedhttp://www.answers.com/topic/unwantedhttp://www.answers.com/topic/vibration

  • 8/20/2019 1-5 Units Study Materials

    17/61

     

    Powered By www.technoscriptz.com

    UNIT 2

    MEASUREMENT OF ACCELERATION, VIBRATION AND DENSITY 

    AIM:

    . Exposure to various accelerometer pick-ups, vibrometers, density andviscosity pick-ups 

    KEY WORDS:

    Accelerometers:- LVDT - Vibrometers - density and specific gravity  –   densitometers- Ultrasonic densitometer-Bridge type gas densitometer  

  • 8/20/2019 1-5 Units Study Materials

    18/61

     

    Powered By www.technoscriptz.com

    UNIT-II 

    MEASUREMENT OF ACCELERATION, VIBRATION AND DENSITY

    MEASUREMENT OF ACCELERATION 

    Accelerometers:

    An accelerometer is a device that measures the vibration, or acceleration of motion of a structure. The

    force caused by vibration or a change in motion (acceleration) causes the mass to "squeeze" the

     piezoelectric material which produces an electrical charge that is proportional to the force exerted upon

    it. Since the charge is proportional to the force, and the mass is a constant, then the charge is also

     proportional to the acceleration.

    There are two types of piezoelectric accelerometers (vibration sensors). The first type is a "high

    impedance" charge output accelerometer. In this type of accelerometer the piezoelectric crystal produces

    an electrical charge which is connected directly to the measurement instruments. The charge output

    requires special accommodations and instrumentation most commonly found in research facilities. This

    type of accelerometer is also used in high temperature applications (>120C) where low impedance

    models cannot be used.

    The second type of accelerometer is a low impedance output accelerometer. A low impedance

    accelerometer has a charge accelerometer as its front end but has a tiny built-in micro-circuit and FET

    transistor that converts that charge into a low impedance voltage that can easily interface with standard

    instrumentation. This type of accelerometer is commonly used in industry. An accelerometer power

    supply like the ACC-PS1, provides the proper power to the microcircuit 18 to 24 V @ 2 mA constantcurrent and removes the DC bias level, they typically produces a zero based output signal up to +/- 5V

    depending upon the mV/g rating of the accelerometer. All OMEGA(R) accelerometers are this lowimpedance type.

    LVDT accelerometer

  • 8/20/2019 1-5 Units Study Materials

    19/61

     

    Powered By www.technoscriptz.com

    A second type of accelerometer takes advantage of the natural linear displacement

    measurement of the LVDT to measure mass displacement. In these instruments, the LVDT coreitself is the seismic mass. Displacements of the core are converted directly into a linearly

     proportional ac voltage. These accelerometers generally have a natural frequency less than 80 Hzand are commonly used for steady-state and low-frequency vibration. Figure shows the basic

    structure of such an accelerometer.

    Piezoelectric Accelerometer:

    The piezoelectric accelerometer is based on a property exhibited by certain crystals where

    a voltage is generated across the crystal when stressed. This property is also the basis for suchfamiliar sensors as crystal phonograph cartridges and crystal microphones. For accelerometers,the principle is shown in Figure 5.28. Here, a piezoelectric crystal is spring-loaded with a test

    mass in contact with the crystal. When exposed to an acceleration, the test mass stresses thecrystal by a force (F  = ma), resulting in a voltage generated across the crystal. A measure of this

    voltage is then a measure of the acceleration. The crystal per se is a very high-impedance source,

    and thus requires a high-input impedance, low-noise detector. Output levels are typically in the

    millivolt range. The natural frequency of these devices may exceed 5 kHz, so that they can be

    used for vibration and shock measurements.

    Variable Reluctance

  • 8/20/2019 1-5 Units Study Materials

    20/61

     

    Powered By www.technoscriptz.com

    This accelerometer type falls in the same general category as the LVDT in that an

    inductive principle is employed. Here, the test mass is usually a permanent magnet. The

    measurement is made from the voltage induced in a surrounding coil as the magnetic mass moves

    under the influence of acceleration. This accelerometer is used in vibration and shock studies

    only, because it has an output only when the mass is in motion. Its natural frequency is typically

    less than 100 Hz. This type of accelerometer often is used in oil exploration to pick up vibrationsreflected from underground rock strata. In this form, it is commonly referred to as a geophone. 

    Seismic instruments as accelerometer

    The mass is connected through the parallel spring and damper arrangement to the housing

    frame. This frame is then connected to the vibration source whose characteristics are to bemeasured. The mass tends to remain fixed in its spatial position, so that the vibration motion is

    registered as a relative displacement between the mass and the housing frame. The displacement

    is then sensed and indicated by an appropriate transducer. The seismic instrument may be used

    for either displacement or acceleration measurement by proper selection of mass, spring and

    damper combinations.

    Vibration instruments

    Calibration of vibration pick ups:

      Constant Acceleration method: Constant acceleration methods, which are suitably

    only for calibrating accelerometers include the tilting- support method and the centrifuge.

    The tilting-support method utilises the accelerometer’s inherent sensitivity to gravity..

    Static acceleration over the range +-1g may be accurately applied by fastening the

    accelerometer to a tilting support whose tilt support whose tilt angle from vertical is

    accurately measured. This method requires that the accelerometer respond to staticaccelerations; therefore piezoelectric devices cannot be calibrated in this way. 

      This method consists of a modified electro dynamic vibration shaker which has

     been carefully designed to provide uniaxial pure sinusoidal motion which is equipped with

    an accurately calibrated moving coil velocity pick up to measure its table motion. If a

    motion is purely sinusoidal, knowledge of its velocity pickup to measure its table motion.If Motion is known to be purely sinusoidal, knowledge of its velocity and frequency

    enables accurate calculation of the displacement and displacement. The motion frequencyis easily obtained with high accuracy by electronic counters. This technique is thus useful

    for displacement , velocity or acceleration pickups.

    LASER DOPPLER VIBROMETERS:

  • 8/20/2019 1-5 Units Study Materials

    21/61

     

    Powered By www.technoscriptz.com

    During the last years the growing importance of the correct determination of the state of

    conservation of artworks has been stated by all personalities in care of Cultural Heritage. There

    exist many analytical methodologies and techniques to individuate the physical and chemical

    characteristics of artworks, but at present their structural diagnostics mainly rely on the expertise

    of the restorer/technician and the typical diagnostic process is accomplished mainly through

    manual and visual inspection of the structure. For this reason, many innovative optical techniques

    have been tried and applied to this issue and in these pages we will show some examplesregarding the use of the laser Doppler vibrometer (LDV); The basic idea behind the employment

    of LDV is to substitute human senses and contact sensors with measurement systems capable ofremote acquisition and, if necessary, of remote structural excitation: surfaces are very slightly

    vibrated by mechanical and acoustical actuators, while a laser Doppler vibrometer means theobjects measuring surface velocity and producing 2D or 3D maps. Think, for example, of a fresco

    with delaminated areas: where these defects occur, velocity is higher than neighbouring areas sodefects can be easily spotted by a LDV. Laser vibrometers also identify structural resonance

    frequencies thus leading to a complete characterization of these defects, and this holds true also

    for massive structures, like towers, buildings, churches.

    Laser Doppler Vibrometers, or better Scanning Laser Doppler Vibrometers (SLDV), have beenapplied to different types of movable or decorative artworks, like frescoes, icons, mosaics,

    ceramics, inlaid wood and easel painting, with different degrees of success, but always showingan impressive list of important advantages:

       —  no remarkable intrusivity,

       —  remote measurements,

       —  ample frequency response,

       —  high sensibility,

       —  portability.

    Moreover all existing systems are completely PC controlled and this allows digital data storage

    and easy data transfer to other applications like software packages for structural and modalanalysis, and to spreadsheets applications like Excel or Matlab.

    The application to historical buildings is more recent [2] and still limited but looks promising and

    will be the subject of much research in the immediate future. Of course there still exist a lot of

    difficulties, mainly related to the non-optically collaborative surfaces of tested structures and the

    necessity of working at great distances to get data that can be considered representative of the

    examined object. These two factors work one against the other, and this makes the application of

    SLDV mainly a ―prototype‖ application yet, but already exist situations where this is not  the case

    anymore [3].

    Also we must not forget other problems, like instrument isolation from ground vibrations and the

    realization of special excitation techniques but it has been already demonstrated the capability of

    the LDV to acquire non-intrusively vibrational data on not-treated surfaces up to 10-15 meters, a

    real asset when dealing with large structures. Regular monitoring of important parameters related

    to the state of conservation of these huge objects, like frequencies of resonance, is thus possible

    with no external intervention on the structure and may be performed quickly and with a high

    degree of accuracy.

    Optical Sensors for Vibration Measurements

    When we say ―optical sensors‖, we mean an immense variety of instruments, devices andsystems. Just think of such different instruments like infra red thermal cameras or a Bragg grating

    strain sensor. Even to measure vibrations we may have such different solutions like laser based,

  • 8/20/2019 1-5 Units Study Materials

    22/61

     

    Powered By www.technoscriptz.com

    LED based, fibre based sensors; not to mention the physical scale of such sensors, going from the

    micro scale, e.g. optical MEMS accelerometers, to relatively ―immense‖ laser Doppler

    vibrometers. If we confine ourselves to laser based instrumentation, we may mention full field

    techniques (holography, shearography, ESPI, for example) or focused beam ones (laser Doppler

    vibrometers).

    The advantages of optical sensors are outstanding and their use is spreading more and more, day

    after day. Think for example to optical fibres sensors: their solid state design is resistant to

    vibration, unaffected by electromagnetic interference (nor do they create additional EMI), and,

     because the light source can be located far away from explosive materials, do not run the risk of

    sparking an explosion. They also offer superior multiplexing capabilities, thanks to the possibility

    of having multiple sensors in a single fibre line.

    Also, fibre optic sensors fall into a variety of sensor types: chemical, temperature, strain,

     biomedical, electrical and magnetic, rotation, vibration, displacement, pressure, and flow. Manyof these categories were developed by military organizations during the nineties. These sensors

    are extremely effective at creating "smarter" structures, widely used nowadays for chemicalsensing (especially in the petrochemical industry), transportation, building and structural

    monitoring, and biomedical.

    However, fibre sensors must be placed in contact or closed to the object to be measured, and so

    they maybe not used in many occasions, where the objects cannot be reached or are impossible tomodify, e.g. a fresco in a church.

    For cases like these, instrumentation based on a laser beam used as a probe is much more suited,

    and we will deal exclusively with these devices in the following of this publication. Major

    advantages of such instruments rely not only in this absence of invasivity, but also in their high

    sensibility and in their capacity of acquiring detailed data in the terms of space, time, and

    frequency. Many of these systems are still quite expensive, but their contribution to solve design,

     production process, or quality control problems is invaluable.

    More specifically we will deal with focused laser beam instruments, laser Doppler vibrometers.We will avoid detailed mathematical description of involved theory, preferring a more intuitive

    approach to make this matter more palatable to a wider range of learners.

    The scanning version of the LDV may automatically and accurately measure point-by-point

    surface velocities using interferometric techniques and a couple of galvanometric driven mirrorssteering the laser beam. In this way it is possible to scan a grid of acquisition points acquiring

    response spectra and time histories of the velocity of each point; these data are then processed and presented as 2D or 3D colour maps. Modern SLDVs may scan 100 points/second for a total

    number of more than 100.000 points working with a maximum frequency in the range of sometens of MHz, and with a lower limit of less than a Hertz. Full-scale highest range is typically 10

    m/s with lower ranges in the order of 1 mm/s, corresponding to a displacement of some tens ofnanometres.

    These features make the SLDV an ideal instrument in applications where it is impossible or very

    difficult to use standard vibration measuring devices, such as accelerometers. Accelerometers will

    load the examined structures and may even damage the delicate surface of precious objects.Moreover, to perform an accurate vibrational analysis it would require to employ many

    transducers or to move one all around the tested piece and in both cases time and cost would riseconsiderably

    MEASUREMENT OF DENSITY:

    Density: 

    http://alpha1.infim.ro/cost/pagini/handbook/chapters/ldv.htmhttp://alpha1.infim.ro/cost/pagini/handbook/chapters/shea.htmhttp://alpha1.infim.ro/cost/pagini/handbook/chapters/shea.htmhttp://alpha1.infim.ro/cost/pagini/handbook/chapters/ldv.htm

  • 8/20/2019 1-5 Units Study Materials

    23/61

     

    Powered By www.technoscriptz.com

    Density is defined as an objects mass per unit volume. Mass is a property. 

    Mass and Weight - the Difference! - What is weight and what is mass? An explanation of

    the difference between weight and mass.

    The density can be expressed as

     ρ = m / V  = 1 / v g (1) 

    where 

     ρ = density (kg/m3 ) 

    m = mass (kg) 

    V = volume (m3 ) 

    v g  = specific volume (m3 /kg) 

    The SI units for density are kg/m3. The imperial (BG) units are lb/ft3 (slugs/ft3). While peopleoften use pounds per cubic foot as a measure of density in the U.S., pounds are really a measure

    of force, not mass. Slugs are the correct measure of mass. You can multiply slugs by 32.2 for a

    rough value in pounds.

      Unit converter for other units

    The higher the density, the tighter the particles are packed inside the substance. Density is a

     physical property constant at a given temperature and density can help to identify a substance.

      Densities and material properties for common materials

    Relative Density (Specific Gravity)

    Relative density of a substance is the ratio of the substance to the density of water, i.e.

    Specific Weight

    Specific Weight is defined as weight per unit volume. Weight is a force.

      Mass and Weight - the difference! - What is weight and what is mass? An explanation ofthe difference between weight and mass.

    Specific Weight can be expressed as

    γ = ρ g   (2) 

    where 

    γ = specific weight (N/m3 ) 

     ρ = density (kg/m3 ) 

     g = acceleration of gravity (m/s2 ) 

    The SI-units of specific weight are N/m3. The imperial units are lb/ft3. The local acceleration g  is

    under normal conditions 9.807 m/s2 in SI-units and 32.174 ft/s

    2 in imperial units.

    http://www.engineeringtoolbox.com/mass-weight-d_589.htmlhttp://www.engineeringtoolbox.com/unit-converter-d_185.htmlhttp://www.engineeringtoolbox.com/material-properties-t_24.htmlhttp://www.engineeringtoolbox.com/mass-weight-d_589.htmlhttp://www.engineeringtoolbox.com/accelaration-gravity-d_340.htmlhttp://www.engineeringtoolbox.com/accelaration-gravity-d_340.htmlhttp://www.engineeringtoolbox.com/mass-weight-d_589.htmlhttp://www.engineeringtoolbox.com/material-properties-t_24.htmlhttp://www.engineeringtoolbox.com/unit-converter-d_185.htmlhttp://www.engineeringtoolbox.com/mass-weight-d_589.html

  • 8/20/2019 1-5 Units Study Materials

    24/61

     

    Powered By www.technoscriptz.com

    Ultrasonic densitometer

    Twin folks are inserted into liqu9d or gas media whose density needs to be

    measured since the natural frequency of the forks is a function of density of the media, small

    changes in the natural frequency must be monitored accurately

    Pressure head type densitometers:

    The pressure at the bottom of the tank of the constant liquid column is proportional to

    density and the weight of the given volume of the fluid is proportional to density.

    It compares hydrostatic pressures due to the height of the liquids in two tanks. one is the

    reference tank, consisting of a liquid of constant height and density. The other tank maintains the

    height constant by overflow, so that the manometer can be directly in terms of density

    measurement.

    Float type densitometer:

    The plumet is located entirely under the liquid surface the effective weight of the chain on

    the plumet varies depending on the position of the plumet which in turn the function of density of

    the liquid

    Bridge type gas densitometer

    It consist of four arm of pipe connections like wheat stone bridge for a balanced flow the

    detector elements are equally cooled and when they are connected in the wheat stone bridge and

    they indicates null balance. The detector bridge unbalance will therefore will beameasure of gas

    density 

  • 8/20/2019 1-5 Units Study Materials

    25/61

     

    Powered By www.technoscriptz.com

    UNIT 3

    PRESSURE MEASUREMENT 

    AIM:

    To have an adequate knowledge about pressure transducers 

    KEY WORDS:

    Manometers- pressure gauges- Elastic elements- bellows- diaphragms- LVDT and straingauges- Piezo-resistive

     pressure sensor- Ionization gauges- Dead weight tester. 

  • 8/20/2019 1-5 Units Study Materials

    26/61

     

    Powered By www.technoscriptz.com

    UNIT-III

    PRESSURE MEASUREMENT 

    Pressure :

    Pressure is defined as force per unit area. It is usually more convenient to use pressure

    rather than force to describe the influences upon fluid behavior. The standard unit for pressure is

    the Pascal, which is a Newton per square meter.

    For an object sitting on a surface, the force pressing on the surface is the weight of the

    object, but in different orientations it might have a different area in contact with the surface and

    therefore exert a different pressure

    Units of pressure

    http://hyperphysics.phy-astr.gsu.edu/hbase/mass.html#wgthttp://hyperphysics.phy-astr.gsu.edu/hbase/mass.html#wgt

  • 8/20/2019 1-5 Units Study Materials

    27/61

     

    Powered By www.technoscriptz.com

    Pressure measurements:

    Manometer methods:

    1. 

    U-tube manometer:

    When there is a pressure difference between two ends of the tube the liquid

    goes down on one side of the tube and up on other side the difference in liquid levels from one

    side to other indicates the difference in pressure

    Well type manometer:

    The well type manometer is

    widely used because the

    reading a single leg is required in it consist of very large diameter vessel connected on one side to

    a very small size tube thus the zero level moves very little when pressure is applied

    Inclined manometer:

    Inclined manometer is used to measure very small pressure differences the manometer is tipped

    so that the liquid moves a longer distance through the tube as it rises

    Elastic type pressure gauges:

  • 8/20/2019 1-5 Units Study Materials

    28/61

     

    Powered By www.technoscriptz.com

    Bourdon-tube designs

    Since the invention of the Bourdon-tube gauge more than a century ago, pressure gauge

    manufacturers have been developing different types of gauges to meet specific needs without ever

    changing the basic principle of the Bourdon tube's operation. Bourdon-tube gauges, Figure 1, are

    now commonly available to measure a wide range of gauge, absolute, sealed, and differential

     pressures, plus vacuum.

    They are manufactured to an accuracy as high as 0.1% of span and in dial diameters from 1-1/2 to

    16 in. A variety of accessories can extend their performance and usefulness. For example,

    snubbers and gauge isolators can be installed to protect the sensitive internal workings of the

    gauge from pressure spikes. The availability of Bourdon-tube pressure gauges to meet specific

    needs, coupled with their inherent ruggedness, simplicity, and low cost has resulted in their wide

    use in many applications.

    Gauges using C-shaped Bourdon tubes as theelastic chamber - the type shown in Figure 1 -

    are by far the most common. Pressurized fluidenters the stem at the bottom (which is

    sometimes center-back-mounted instead) and

     passes into the Bourdon tube. The tube has a

    flattened cross section and is sealed at its tip.

    Any pressure in the tube in excess of the

    external pressure (usually atmospheric) causes

    the Bourdon tube to elastically change its shape

    to a more circular cross section.

    This change in shape of the cross section tends to straighten the C-shape of the Bourdon tube.With the bottom stem end fixed, the straightening causes the tip at the opposite end to move ashort distance - 1/16 to 1/2 in., depending on the size of the tube. A mechanical movement then

    transmits this tip motion to a gear train that rotates an indicating pointer over a graduated scale to

    display the applied pressure. Often, a movement is incorporated to provide mechanical advantage

    to multiply the relatively short movement of the tube tip.

    Bellows and diaphragms:

    Low-pressure applications do not generateenough force in the Bourdon tube to operate the

    multiplying mechanism; therefore, Bourdon-

    tube gauges are not generally used for pressure

    spans under 12 psi. For these ranges, some other

    form of elastic chamber must be used, a metallic

     bellows, Figure 4, for example. These bellows

    generally are made by forming thin-wall tubing.

    However, to obtain a reasonable fatigue life and

    motion that is more linear with pressure, a coil

    spring supplements the inherent spring rate of

    the bellows. These spring-loaded bellowsgauges generally are used in pressure ranges having spans to 100 psi and to 1 in. Hg.

    Simplified view of spiral Bourdon-tube

    pressure gage and movement.

    Cross-sectional view of spring-loaded

    bellows pressure gauge.

  • 8/20/2019 1-5 Units Study Materials

    29/61

     

    Powered By www.technoscriptz.com

    Metallic diaphragms also are used as the elastic chamber in low-pressure gauges. A diaphragm

     plate is formed from thin sheet metal into a shallow cup having concentric corrugations. To make

    an element with a low spring rate that generates substantial deflection from a small change in

     pressure, two plates can be soft soldered, brazed, or welded at their periphery to form a capsule,

    and additional capsules can be joined at their centers to form a stack, Figure 5.

    Generally, the measured pressure is applied to the interior of the element and no supplementalcoil springs are used. A 2-in. diameter capsule (two plates) will provide about 0.060 in. of motion

    without exceeding the elastic limit of the material. This is usually enough to operate a high-ratio

    multiplying movement because diaphragm deflection can transmit high force.

    Diaphragm elements often are used in gauges to indicate absolute pressure. In this form, the

    diaphragm element is evacuated. sealed, and mounted within a closed chamber. The pressure to be measured is admitted to the closed chamber and surrounds the diaphragm element. Changes in

    the measured pressure cause the element to deflect, but because atmospheric pressure is excludedand has no effect on the indication, the gauge may be calibrated in terms of absolute pressure. If

    the applied pressure is atmospheric pressure, the gauge is known as a barometer.

    Diaphragm elements also may be used in an opposing arrangement. By evacuating one side of the

    assembly, the gauge can indicate absolute pressure. If a pressure is applied to one side of the

    assembly, and a second pressure is applied to the other side, then the differential pressure will be

    indicated. The differential pressure is limited with respect to the static pressure that can be

    applied. That is, the gauge may be suitable to indicate between 10 psi and 12 psi, but not be

    suitable to indicate between 100 psi and 102 psi. Also, the consequence of inadvertently applying

    full pressure to one side of the element and no pressure to the other side of the element must be

    considered.

    Elastic element with LVDT based pressure measurement

    Any change in pressure will given to bellows which in turn actuate the core of the

    LVDT and produces output in the secondary the value of output is directly proportional to the

     pressure input to the bellows

    Capacitive type pressure transducer:

  • 8/20/2019 1-5 Units Study Materials

    30/61

     

    Powered By www.technoscriptz.com

    Any change in pressure causes the change in distance between diaphragm and fixed

     plate which the unbalance the bridge the bridge is proportional to pressure applied.

    Piezoresistive pressure sensor:

    Piezoresistive materials are materials that change resistance to the flow of current when they are

    compressed or strained. Metal is piezoresistive to some degree, but most pressure sensors use the

    semiconductor silicon. When force is put on the silicon, it becomes more resistant to a current

     pushing through. This resistance is usually very linear--twice as much pressure results in twice as

    large a change in resistance.

    A Piezoresistive Pressure Sensor contains several thin wafers of silicon embedded between

     protective surfaces. The surface is usually connected to a Wheatstone bridge, a device for

    detecting small differences in resistance. The Wheatstone bridge runs a small amount of current

    through the sensor. When the resistance changes, less current passes through the pressure sensor.

    The Wheatstone bridge detects this change and reports a change in pressure.

    Resonant Wire pressure sensor: 

  • 8/20/2019 1-5 Units Study Materials

    31/61

     

    Powered By www.technoscriptz.com

    The resonant-wire pressure transducer was introduced in the late 1970s. In this design , a wire is

    gripped by a static member at one end, and by the sensing diaphragm at the other. An oscillator

    circuit causes the wire to oscillate at its resonant frequency. A change in process pressure changes

    the wire tension, which in turn changes the resonant frequency of the wire. A digital counter

    circuit detects the shift. Because this change in frequency can be detected quite precisely, this

    type of transducer can be used for low differential pressure applications as well as to detectabsolute and gauge pressures.

    The most significant advantage of the resonant wire pressure transducer is that it generates aninherently digital signal, and therefore can be sent directly to a stable crystal clock in a

    microprocessor. Limitations include sensitivity to temperature variation, a nonlinear outputsignal, and some sensitivity to shock and vibration. These limitations typically are minimized by

    using a microprocessor to compensate for nonlinearities as well as ambient and processtemperature variations.

    Resonant wire transducers can detect absolute pressures from 10 mm Hg, differential pressures

    up to 750 in. water, and gauge pressures up to 6,000 psig (42 MPa). Typical accuracy is 0.1% of

    calibrated span, with six-month drift of 0.1% and a temperature effect of 0.2% per 1000¡ F.

    Measurement of vacuum:

    McLeod gauge:

    A McLeod gauge isolates a sample of gas and compresses it in a modified mercury manometer

    until the pressure is a fewmmHg. The gas must be well-behaved during its compression (it must

    not condense, for example). The technique is slow and unsuited to continual monitoring, but is

    capable of good accuracy.

    Useful range: above 10-4

     torr  [3]

     (roughly 10-2

     Pa) as high as 10−6 Torr (0.1 mPa),

    mPa is the lowest direct measurement of pressure that is possible with current technology. Other

    vacuum gauges can measure lower pressures, but only indirectly by measurement of other

     pressure-controlled properties.

    The McLeod gauge measures the pressure of gases by compressing a known volume with a fixed

     pressure. The new volume is then a measure of the initial absolute pressure.-- The McLeod gauge has been used until recently for calibrating other gauges.

    http://en.wikipedia.org/wiki/McLeod_gaugehttp://en.wikipedia.org/wiki/Torrhttp://en.wikipedia.org/wiki/Pressure_measurement#cite_note-THV_V-2http://en.wikipedia.org/wiki/Pressure_measurement#cite_note-THV_V-2http://en.wikipedia.org/wiki/Pressure_measurement#cite_note-THV_V-2http://en.wikipedia.org/wiki/Pressure_measurement#cite_note-THV_V-2http://en.wikipedia.org/wiki/Torrhttp://en.wikipedia.org/wiki/McLeod_gauge

  • 8/20/2019 1-5 Units Study Materials

    32/61

     

    Powered By www.technoscriptz.com

    - It covers the vacuum range between 1 and 10-6 torr.

    Thermal Designs:

    The thermal conductivity of a gas changes with its pressure in the vacuum range. If an element

    heated by a constant power source is placed in a gas, the resulting surface temperature of the

    element will be a function of the surrounding vacuum. Because the sensor is an electrically heated

    wire, thermal vacuum sensors are often called hot wire gauges. Typically, hot wire gauges can be

    used to measure down to 10-3 mm Hg.

    Pirani guage:

    In this design, a sensor wire is heated electrically and the pressure of the gas is

    determined by measuring the current needed to keep the wire at a constant temperature

    Ionisation gauges:

    Hot cathode vaccum guage

    The operating principles of this gauge are similar to the Penning gauge except that the electrons

    are produced by a hot filament and accelerated to a grid. The pressure range covered is either 1 to

    10-5 torr or 10-2 to 10-7 torr, depending on the electrode structure. Electrons emitted from the

    filament ionize residual gas molecules in the container being evacuated; the ion current arriving at

    the collector plates is directly proportional to the pressure and the ionization probability of the

    residual gas. This is a clean, accurate gauge that can be used down to about 10-6 torr; below this pressure its accuracy is reduced due to the soft X-rays produced by electrons striking the grid.

    These X-rays generate a current in the collector circuit independent of pressure.

  • 8/20/2019 1-5 Units Study Materials

    33/61

     

    Powered By www.technoscriptz.com

     Bayard-Alpert hot-filament ionization gauge. In this ionization gauge, the cross section of the

    collector is reduced to minimum to reduce the X-ray effect. This is achieved by inverting the

    gauge — that is, the collector (a fine wire) is surrounded by the grid. The pressure range covered is

    10-3 to 10-9 torr or down to 10-11 torr if a modulated instrument is used. Operating principles are

    the same as for the other ionization gauges

    Cold cathode vacuum guage:

    This gauge makes use of the fact that the rate of ion production by a stream of  electrons in a

    vacuum system is dependent on pressure and the ionization probability of the residual gas. Also

    called the Penning gauge, it consists of two cathodes opposite one another with an anode centrally

    spaced between them inside a metal or glass envelope. Outside the envelope a permanent magnet

     provides a magnetic field to lengthen the path travelled by the electron in going from cathode to

    anode, thus increasing the amount of ionization occurring within the gauge. Normally the anodeis operated at about 2 kV, giving rise to a direct current caused by the positive ions arriving at thecathode. The pressure is indicated directly by the magnitude of the direct current produced. The

     pressure range covered by this gauge is from as low as 10-7

     torr. It is widely used in industrialsystems because it is rugged and simple to use.

    Testing and calibration of pressure gauges-Dead weight tester.

    Dead weight tester

    Deadweights are usually used for pressure gauge calibration as they come with high accuarcy, So

    they can be used as primery standard (as mentioned before).there are many types of them

    depending on the application and they are operated with oil (hydrulic) or with air (penumatic). 

    Deadweight testers are the basic primary standard for accurate measurement of pressure.

    Deadweight testers are used to measure the pressure exerted by gas or liquid and can also

    generate a test pressure for the calibration of numerous pressure instruments.

    Hope this helps!

    Description

    http://www.britannica.com/EBchecked/topic/183692/electron-tubehttp://www.britannica.com/EBchecked/topic/357048/magnetic-fieldhttp://www.britannica.com/EBchecked/topic/164851/direct-currenthttp://www.britannica.com/EBchecked/topic/99895/cationhttp://www.britannica.com/EBchecked/topic/99895/cationhttp://www.britannica.com/EBchecked/topic/164851/direct-currenthttp://www.britannica.com/EBchecked/topic/357048/magnetic-fieldhttp://www.britannica.com/EBchecked/topic/183692/electron-tube

  • 8/20/2019 1-5 Units Study Materials

    34/61

     

    Powered By www.technoscriptz.com

    Known weights are placed on a rotating plate on top of a calibrated piston, connected by tubing to

    the pressure sensor being tested. This puts a known force (weights) on a known surface area

    (piston). The rotation eliminates any static friction that would affect the reading.

    Dead Weight Testers.

    1 - Handpump

    2 - Testing Pump3 - Pressure Gauge to be calibrated

    4 - Calibration Weight

    5 - Weight Support

    6 - Piston

    7 - Cylinder

    8 - Filling Connection 

    Dead weight testers are a piston-cylinder type measuring device. As primary standards, they are

    the most accurate instruments for the calibration of electronic or mechanical pressure measuring

    instruments.

    They work in accordance with the basic principle that P= F/A, where the pressure (P) acts on aknown area of a sealed piston (A), generating a force (F). The force of this piston is then

    compared with the force applied by calibrated weights. The use of high quality materials result insmall uncertainties of measurement and excellent long term stability.

    Dead weight testers can measure pressures of up to 10,000 bar, attaining accuracies of between

    0.005% and 0.1% although most applications lie within 1 - 2500 bar. The pistons are partly madeof tungsten carbide (used for its small temperature coefficient), and the cylinders must fit together

    with a clearance of no more than a couple of micrometers in order to create a minimum frictionthus limiting the measuring error. The piston is then rotated during measurements to further

    minimise friction.

    The testing pump (2) is connected to the instrument to be tested(3), to the actual measuring

    component and to the filling socket. A special hydraulic oil or gas such as compressed air or

    nitrogen is used as the pressure transfer medium. The measuring piston is then loaded with

    calibrated weights (4). The pressure is applied via an integrated pump (1) or, if an external

     pressure supply is available, via control valves in order to generate a pressure until the loadedmeasuring piston (6) rises and 'floats' on the fluid. This is the point where there is a balance between pressure and the mass load. The piston is rotated to reduce friction as far as possible.

  • 8/20/2019 1-5 Units Study Materials

    35/61

     

    Powered By www.technoscriptz.com

    Since the piston is spinning, it exerts a pressure that can be calculated by application of a

    derivative of the formula P = F/A.

    The accuracy of a pressure balance is characterised by the deviation span, which is the sum of the

    systematic error and the uncertainties of measurement.

    Today's dead weight testers are highly accurate and complex and can make sophisticated physicalcompensations. They can also come accompanied by an intelligent calibrator unit which canregister all critical ambient parameters and automatically correct them in real time making

    readings even more accurate.

    UNIT 4

    TEMPERATURE MEASUREMENT 

    AIM:

    To have an idea about the temperature standards, calibration and signalconditioning used in RTD’s 

    KEY WORDS:Thermometers- Filled in thermometers- Bimetallic thermometers- RTDS- thermistors 

  • 8/20/2019 1-5 Units Study Materials

    36/61

     

    Powered By www.technoscriptz.com

    UNIT-IV

    TEMPERATURE MEASUREMENT

    Temperature measurement: 

    Measurement of the hotness of a body relative to a standard scale. The fundamental scale of

    temperature is the thermodynamic scale, which can be derived from any equation expressing the

    second law of thermodynamics. Efforts to approximate the thermodynamic scale as closely as

     possible depend on relating measurements of temperature-dependent physical properties of

    systems to thermodynamic relations expressed by statistical thermodynamic equations, thus ingeneral linking temperature to the average kinetic energy of the measured system. Temperature-

    measuring devices, thermometers, are systems with properties that change with temperature in asimple, predictable, reproducible

    In the establishment of a useful standard scale, assigned temperature values of thermodynamic

    equilibrium fixed points are agreed upon by an international body (General Conference ofWeights and Measures), which updates the scale about once every 20 years. Thermometers for

    interpolating between fixed points and methods for realizing the fixed points are prescribed, providing a scheme for calibrating thermometers used in science and industry.

    The scale now in use is the International Temperature Scale of 1990 (ITS-90). Its unit is the

    kelvin, K, arbitrarily defined as 1/273.16 of the thermodynamic temperature T  of the triple point

    of water (where liquid, solid, and vapor coexist). For temperatures above 273.15 K, it is common

    to use International Celsius Temperatures, t 90 (rather than International KelvinTemperatures, T 90), having the unit degree Celsius, with symbol °C. The degree Celsius has thesame magnitude as the kelvin. Temperatures, t 90, are defined as t 90/°C = T 90/K - 273.15, that is, as

    differences from the ice-point temperature at 273.15 K. The ice point is the state in which theliquid and solid phases of water coexist at a pressure of 1 atm (101,325 pascals). [The Fahrenheit

    scale, with symbol °F, still in common use in the United States, is given by t F/°F = (t 90/°C × 1.8)+ 32, or t F/°F = (T 90/K × 1.8) - 459.67.] The ITS-90 is defined by 17 fixed points.

    Primary thermometers are devices which relate the thermodynamic temperature to statistical

    mechanical formulation. The fixed points of ITS-90 are all based on one or more types of gasthermometry or on spectral radiation pyrometry referenced to gas thermometry. Secondary

    thermometers are used as reference standards in the laboratory because primary thermometers areoften too cumbersome. It is necessary to establish standard secondary thermometers referenced to

    one or more fixed points for interpolation between fixed points. Lower-order thermometers are

  • 8/20/2019 1-5 Units Study Materials

    37/61

  • 8/20/2019 1-5 Units Study Materials

    38/61

     

    Powered By www.technoscriptz.com

    deviation from the instrument scale recorded.[16] For many modern devices calibration will be

    stating some value to be used in processing an electronic signal to convert it to a temperature.

    Precision, accuracy, and reproducibility:

    The "Boyce MotoMeter" radiator cap on a 1913 Car-Nation automobile, used to measure

    temperature of vapor in 1910s and 1920s cars.

    The precision or resolution of a thermometer is simply to what fraction of a degree it is possible

    to make a reading. For high temperature work it may only be possible to measure to the nearest

    10°C or more. Clinical thermometers and many electronic thermometers are usually readable to

    0.1°C. Special instruments can give readings to one thousandth of a degree. However, this

     precision does not mean the reading is true or accurate.

    Thermometers which are calibrated to known fixed points (e.g. 0 and 100°C) will

     be accurate (i.e. will give a true reading) at those points. Most thermometers are originally

    calibrated to a constant-volume gas thermometer.[citation needed]

     In between a process

    of  interpolation is used, generally a linear one.[16] This may give significant differences between

    different types of thermometer at points far away from the fixed points. For example the

    expansion of mercury in a glass thermometer is slightly different from the change in resistance of

    a platinum resistance of the thermometer, so these will disagree slightly at around 50°C.[17] There

    may be other causes due to imperfections in the instrument, e.g. in a liquid-in-glass thermometer

    if the capillary varies in diameter .[17] 

    For many purposes reproducibility is important. That is, does the same thermometer give thesame reading for the same temperature (or do replacement or multiple thermometers give the

    same reading)? Reproducible temperature measurement means that comparisons are valid in

    scientific experiments and industrial processes are consistent. Thus if the same type of

    thermometer is calibrated in the same way its readings will be valid even if it is slightly

    inaccurate compared to the absolute scale.

    An example of a reference thermometer used to check others to industrial standards would be a platinum resistance thermometer with a digital display to 0.1°C (its precision) which has been

    calibrated at 5 points against national standards (-18, 0, 40, 70, 100°C) and which is certified to

    an accuracy of ±0.2°C.[18] 

    According to a British Standard, correctly calibrated, used and maintained liquid-in-glass

    thermometers can achieve a measurement uncertainty of ±0.01°C in the range 0 to 100°C, and alarger uncertainty outside this range: ±0.05°C up to 200 or down to -40°C, ±0.2°C up to 450 or

    down to -80°C.[19] 

    Temperature Measurement: Filled-System Thermometers:

    Many physical properties change with temperature, such as the volume ofa liquid, the length of a

    metal rod, the electrical resistance of a wire, thepressure of a gas kept at constant volume, and the

    volume of a gas kept atconstant pressure. Filled-system thermometers use the phenomenon of

    thermal expansion of matter to measure temperature change.

    The filled thermal device consists of a primary element that takes the formof a reservoir or bulb, a

    flexible capillary tube, and a hollow Bourdon tubethat actuates a signal-transmitting device and/or

    a local indicating temperaturedial. A typical filled-system thermometer is shown in Figure .In this

    system, the filling fluid, either liquid or gas, expands as temperatureincreases. This causes the

    Bourdon tube to uncoil and indicate thetemperature on a calibrated dial.

    http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/car-nationhttp://www.answers.com/topic/gas-thermometer-1http://en.wikipedia.org/wiki/Wikipedia:Citation_neededhttp://www.answers.com/topic/interpolation-2http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Peak-17http://www.answers.com/topic/thermometer#cite_note-Peak-17http://www.answers.com/topic/thermometer#cite_note-Peak-17http://www.answers.com/topic/british-standardshttp://www.answers.com/topic/thermometer#cite_note-BS1041-18http://www.answers.com/topic/thermometer#cite_note-BS1041-18http://www.answers.com/topic/thermometer#cite_note-BS1041-18http://www.answers.com/topic/thermometer#cite_note-BS1041-18http://www.answers.com/topic/british-standardshttp://www.answers.com/topic/thermometer#cite_note-Peak-17http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Duncan-16http://www.answers.com/topic/thermometer#cite_note-Benedict11-15http://www.answers.com/topic/interpolation-2http://en.wikipedia.org/wiki/Wikipedia:Citation_neededhttp://www.answers.com/topic/gas-thermometer-1http://www.answers.com/topic/car-nationhttp://www.answers.com/topic/thermometer#cite_note-Benedict11-15

  • 8/20/2019 1-5 Units Study Materials

    39/61

     

    Powered By www.technoscriptz.com

    The filling or transmitting medium is a vapor, a gas, mercury, or another liquid. The liquid-filled

    system is the most common because it requires a bulb with the smallest volume or permits asmaller instrument to be used.The gas-filled system uses the perfect gas law, which states the

    following for an ideal gas:

    T = kPV

    where:

    T = temperature

    k = constant

    P = pressure

    V = volume

    If the volume of gas in the measuring instrument is kept constant, then the ratio of the gas

     pressure and temperature is constant, so that

    The only restrictions on Equation are that the temperature must be expressed in degrees Kelvin

    and the pressure must be in absolute units.

    Different types of Filled in thermometers:

    1.Gas filled thermometers

    2.Liquid filled thermometers

    3.Mercury filled thermometers

    4.Vapour pressure Thermometers

    Sources of errors in filled system:

    1.Ambient temperature effect:

  • 8/20/2019 1-5 Units Study Materials

    40/61

     

    Powered By www.technoscriptz.com

    The change of temperature causes volume changes in the capillary tube and the bourdon tube

    therby causing error in measurement.

    2.Head or elevation effect:

    If the thermometer bulb is placed at a different height with respect to the bourdon tube, elevation

    errors are produced

    3.Barometric effect:

    The effect due to change in the atmospheric pressure is known as the barometric effect.

    4.Immersion effect:

    If the bulb is not properly immersed or fully immersed and the head of the bulb is lost due to

    conduction through the not properly insulated heat from the bulb is lost due to conduction

    through the extension neck and thermal well.This causes what is known as immersion error.

    5.Radiation effect:

    radiation error occurs due to temperature difference between the bulb and other solid bodies

    around.

    Bimetallic Strip Thermometers

    Bulb thermometers are good for measuring temperature accurately, but they are harder to use

    when the goal is to control the temperature. The bimetallic strip thermometer, because it is made

    of metal, is good at controlling things.

    The principle behind a bimetallic strip thermometer relies on the fact that different metalsexpand at different rates as they warm up. By bonding two different metals together, you can

    make a simple electric controller that can withstand fairly high temperatures. This sort of

    controller is often found in ovens. Here is the general layout:

    Two metals make up the bimetallic strip (hence the name). In this diagram, the green metal would be chosen to expand faster than the blue metal if the device were being used in an oven. In a

    refrigerator, you would use the opposite setup, so that as the temperature rises the blue metalexpands faster than the green metal. This causes the strip to bend upward, making contact so that

    current can flow. By adjusting the size of the gap between the strip and the contact, you controlthe temperature.

    You will often find long bimetallic strips coiled into spirals. This is the typical layout of a

     backyard dial thermometer. By coiling a very long strip it becomes much more sensitive to small

    http://home.howstuffworks.com/refrigerator.htmhttp://home.howstuffworks.com/refrigerator.htm

  • 8/20/2019 1-5 Units Study Materials

    41/61

     

    Powered By www.technoscriptz.com

    temperature changes. In a furnace thermostat, the same technique is used and a mercury switch 

    is attached to the coil. The switch turns the furnace on and off. `

    Electrical methods Of Temperature Measurement:

    Thermistors.:

    A thermistor is a type of  resistor whose resistance varies significantly(more than in standard

    resistors) with temperature. The word is a portmanteau of  thermal  and resistor  . Thermistors are

    widely used as inrush current limiters, temperature sensors, self-resetting overcurrent protectors,

    and self-regulating heating elements. 

    Thermistors differ from resistance temperature detectors (RTD) in that the material used in a

    thermistor is generally a ceramic or polymer, while RTDs use pure metals. The temperatureresponse is also different; RTDs are useful over larger temperature ranges, while thermistors

    typically achieve a higher precision within a limited temperature range [usually −90 °C to 130°C].

    Thermistor symbol

    Assuming, as a first-order approximation, that the relationship between resistance andtemperature is linear, then:

    where

    = change in resistance

    = change in temperature

    = first-order temperature coefficient of resistance

    Thermistors can be classified into two types, depending on the sign of . If is positive, 

    the resistance increases with increasing temperature, and the device is called a positive

    temperature coefficient (PTC) thermistor, or posistor. If is negative, the resistancedecreases with increasing temperature, and the device is called a negative temperature

    http://wapedia.mobi/en/Resistorhttp://wapedia.mobi/en/Electrical_resistancehttp://wapedia.mobi/en/Temperaturehttp://wapedia.mobi/en/Portmanteauhttp://wapedia.mobi/en/Thermal_(disambiguation)http://wapedia.mobi/en/Thermal_(disambiguation)http://wapedia.mobi/en/Thermal_(disambiguation)http://wapedia.mobi/en/Resistorhttp://wapedia.mobi/en/Resistorhttp://wapedia.mobi/en/Resistorhttp://wapedia.mobi/en/Electric_currenthttp://wapedia.mobi/en/Sensorshttp://wapedia.mobi/en/Heating_elementhttp://wapedia.mobi/en/Resistance_temperature_detectorhttp://wapedia.mobi/en/Linearhttp://wapedia.mobi/en/Positive_numberhttp://wapedia.mobi/en/Temperature_coefficient#Positive_temperature_coefficient_of_resistancehttp://wapedia.mobi/en/Temperature_coefficient#Positive_temperature_coefficient_of_resistancehttp://wapedia.mobi/en/Temperature_coefficient#Negative_temperature_coefficienthttp://wapedia.mobi/en/Temperature_coefficient#Negative_temperature_coefficienthttp://wapedia.mobi/en/Temperature_coefficient#Negative_temperature_coefficienthttp://wapedia.mobi/en/Temperature_coefficient#Positive_temperature_coefficient_of_resistancehttp://wapedia.mobi/en/Temperature_coefficient#Positive_temperature_coefficient_of_resistancehttp://wapedia.mobi/en/Temperature_coefficient#Positive_temperature_coefficient_of_resistancehttp://wapedia.mobi/en/Positive_numberhttp://wapedia.mobi/en/Linearhttp://wapedia.mobi/en/Resistance_temperature_detectorhttp://wapedia.mobi/en/Heating_elementhttp://wapedia.mobi/en/Sensorshttp://wapedia.mobi/en/Electric_currenthttp://wapedia.mobi/en/Resistorhttp://wapedia.mobi/en/Thermal_(disambiguation)http://wapedia.mobi/en/Portmanteauhttp://wapedia.mobi/en/Temperaturehttp://wapedia.mobi/en/Electrical_resistancehttp://wapedia.mobi/en/Resistor

  • 8/20/2019 1-5 Units Study Materials

    42/61

     

    Powered By www.technoscriptz.com

    coefficient (NTC) thermistor. Resistors that are not thermistors are designed to have a as

    close to zero as possible(smallest possible k), so that their resistance remains nearly constantover a wide temperature range.

    Instead of the temperature coefficient k , sometimes the temperature coefficient of resistance 

    (alpha) or is used. It is defined as [1] 

    For example, for the common PT100 sensor, or 0.385 %/°C. This

    coefficient should not be confused with the parameter below.

    1. Steinhart-Hart equation

    2. B parameter equation

    3. Conduction model4. Self-heating effects

    5. Applications

    1. Steinhart-Hart equation

    In practice, the linear approximation (above) works only over a small temperature range.

    For accurate temperature measurements, the resistance/temperature curve of the device

    must be described in more detail. The Steinhart-Hart equation is a widely used third-order

    approximation:

    where a, b and c are called the Steinhart-Hart parameters, and must be specified for each

    device. T  is the temperature in kelvins and R is the resistance in ohms. To give resistanceas a function of temperature, the above can be rearranged into:

    where

    and

    The error in the Steinhart-Hart equation is generally less than 0.02 °C in themeasurement of temperature[citation needed ]. As an example, typical values for a thermistor

    with a resistance of 3000 Ω at room temperature (25 °C = 298.15 K) are: 

    http://wapedia.mobi/en/Thermistor#1.http://wapedia.mobi/en/Thermistor#2.http://wapedia.mobi/en/Thermistor#3.http://wapedia.mobi/en/Thermistor#4.http://wapedia.mobi/en/Thermistor#5.http://wapedia.mobi/en/Steinhart-Hart_equationhttp://wapedia.mobi/en/Kelvinhttp://wapedia.mobi/en/Ohm_(unit)http://wapedia.mobi/en/Wikipedia:Citation_neededhttp://wapedia.mobi/en/Wikipedia:Citation_neededhttp://wapedia.mobi/en/Wikipedia:Citation_neededhttp://wapedia.mobi/en/Wikipedia:Citation_neededhttp://wapedia.mobi/en/Ohm_(unit)http://wapedia.mobi/en/Kelvinhttp://wapedia.mobi/en/Steinhart-Hart_equationhttp://wapedia.mobi/en/Thermistor#5.http://wapedia.mobi/en/Thermistor#4.http://wapedia.mobi/en/Thermistor#3.http://wapedia.mobi/en/Thermistor#2.http://wapedia.mobi/en/Thermistor#1.

  • 8/20/2019 1-5 Units Study Materials

    43/61

     

    Powered By www.technoscriptz.com

    2. B  parameter equation

     NTC thermistors can also be characterised with the B parameter equation, which is

    essentially the Steinhart Hart equation with ,

    and ,

    where the temperatures are in kelvins and R0 is the resistance at temperature T 0 

    (usually 25 °C = 298.15 K). Solving for R yields:

    or, alternatively,

    where . This can be solved for the temperature:

    The B-parameter equation can also be written as . This can be used to

    convert the function of resistance vs. temperature of a thermistor into a linear function of

    vs. . The average slope of this function will then yield an estimate of the value of the

     B parameter.

    3. Conduction model

    Many NTC thermistors are made from a pressed disc or  cast chip of a semiconductor such as asintered metal oxide. They work because raising the temperature of a semiconductor increases the

    number of  electrons able to move about and carry charge - it promotes them into the conduction

    band . The more charge carriers that are available, the more current a material can conduct. This is

    described in the formula:

    = electric current (amperes) = density of charge carriers (count/m³)

    = cross-sectional area of the material (m²)

    = velocity of charge carriers (m/s)

    = charge of an electron ( coulomb)

    http://wapedia.mobi/en/Kelvinhttp://wapedia.mobi/en/Castinghttp://wapedia.mobi/en/Semiconductorhttp://wapedia.mobi/en/Sinteringhttp://wapedia.mobi/en/Oxidehttp://wapedia.mobi/en/Electronhttp://wapedia.mobi/en/Electric_chargehttp://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Current_(electricity)http://wapedia.mobi/en/Current_(electricity)http://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Conduction_bandhttp://wapedia.mobi/en/Electric_chargehttp://wapedia.mobi/en/Electronhttp://wapedia.mobi/en/Oxidehttp://wapedia.mobi/en/Sinteringhttp://wapedia.mobi/en/Semiconductorhttp://wapedia.mobi/en/Castinghttp://wapedia.mobi/en/Kelvin

  • 8/20/2019 1-5 Units Study Materials

    44/61

     

    Powered By www.technoscriptz.com

    The current is measured using an ammeter. Over large changes in temperature, calibration is

    necessary. Over small changes in temperature, if the right semiconductor is used, the resistance of

    the material is linearly proportional to the temperature. There are many different semiconducting

    thermistors with a range from about 0.01 kelvin to 2,000 kelvins (−273.14 °C to 1,700 °C). 

    Most PTC thermistors are of the "switching" type, which means that their resistance rises

    suddenly at a certain critical temperature. The devices are made of a doped polycrystallineceramic containing barium titanate (BaTiO3) and other compounds. The dielectric constant of this

    ferroelectric material varies with temperature. Below the Curie point temperature, the high

    dielectric constant prevents the formation of potential barriers between the crystal grains, leading

    to a low resistance. In this region the device has a small negative temperature coefficient. At theCurie point temperature, the dielectric constant drops sufficiently to allow the formation of

     potential barriers at the grain boundaries, and the resistance increases sharply. At even highertemperatures, the material reverts to NTC behaviour. The equations used for modeling this

     behaviour were derived by W. Heywang and G. H. Jonker in the 1960s.

    Another type of PTC thermistor is the polymer PTC, which is sold under brand names such as

    "Polyswitch" "Semifuse", and "Multifuse". This consists of a slice of plastic with carbon grainsembedded in it. When the plastic is cool, the carbon grains are all in contact with each other,

    forming a conductive path through the device. When the plastic heats up, it expands, forcing thecarbon grains apart, and causing the resistance of the device to rise rapidly. Like the BaTiO3 

    thermistor, this device has a highly nonlinear resistance/temperature response and is used forswitching, not for proportional temperature measurement.

    Yet another type of thermistor is a silistor, a thermally sensitive silicon resistor. Silistors are

    similarly constructed and operate on the same principles as other thermistors, but employ siliconas the semiconductive component material.

    4. Self-heating effects

    When a current flows through a thermistor, it will generate heat which will raise the temperature

    of the thermistor above that of its environment. If the thermistor is being used to measure the

    temperature of the environment, this electrical heating may introduce a significant error if a

    correction is not made. Alternatively, this effect itself can be exploited. It can, for example, make

    a sensitive air-flow device employed in a sailplane rate-of-climb instrument, the electronic

    variometer, or serve as a timer for a relay as was formerly done in telephone exchanges. 

    The electrical power input to the thermistor is just:

    where I  is current and V  is the voltage drop across the thermistor. This power is converted to heat,

    and this heat energy is transferred to the surrounding environment. The rate of transfer is well

    described by Newton's law of cooling: 

    where T(R) is the temperature of the thermistor as a function of its resistance  R, is the

    temperature of the surroundings, and K  is the dissipation constant, usually expressed in units ofmilliwatts per degree Celsius. At equilibrium, the two rates must be equal.

    http://wapedia.mobi/en/Ammeterhttp://wapedia.mobi/en/Kelvinhttp://wapedia.mobi/en/Ceramichttp://wapedia.mobi/en/Barium_titanatehttp://wapedia.mobi/en/Dielectric_constanthttp://wapedia.mobi/en/Ferroelectrichttp://wapedia.mobi/en/Curie_pointhttp://wapedia.mobi/en/Dielectric_constanthttp://wapedia.mobi/en/Polymerhttp://wapedia.mobi/en/Polyswitchhttp://wapedia.mobi/en/Carbonhttp://wapedia.mobi/en/Plastichttp://wapedia.mobi/en/Electrical_conductorhttp://wapedia.mobi/en/Sailplanehttp://wapedia.mobi/en/Variometerhttp://wapedia.mobi/en/Timerhttp://wapedia.mobi/en/Relayhttp://wapedia.mobi/en/Telephone_exchangehttp://wapedia.mobi/en/Newton%27s_law_of_coolinghttp://wapedia.mobi/en/Newton%27s_law_of_coolinghttp://wapedia.mobi/en/Telephone_exchangehttp://wapedia.mobi/en/Relayhttp://wapedia.mobi/en/Timerhttp://wapedia.mobi/en/Variometerhttp://wapedia.mobi/en/Sailplanehttp://wapedia.mobi/en/Electrical_conductorhttp://wapedia.mobi/en/Plastichttp://wapedia.mobi/en/Carbonhttp://wapedia.mobi/en/Polyswitchhttp://wapedia.mobi/en/Polymerhttp://wapedia.mobi/en/Dielectric_constanthttp://wapedia.mobi/en/Curie_pointhttp://wapedia.mobi/en/Ferroelectrichttp://wapedia.mobi/en/Dielectric_constanthttp://wapedia.mobi/en/Barium_titanatehttp://wapedia.mobi/en/Ceramichttp://wapedia.mobi/en/Kelvinhttp://wapedia.mobi/en/Ammeter

  • 8/20/2019 1-5 Units Study Materials

    45/61

     

    Powered By www.technoscriptz.com

    The current and voltage across the thermistor will depend on the particular circuit configuration.As a simple example, if the voltage across the thermistor is held fixed, then by Ohm's Law we

    have and the equilibrium equation can be solved for the ambient temperature as a

    function of the measured resistance of the thermistor:

    The dissipation constant is a measure of the thermal connection of the thermistor to its

    surroundings. It is generally given for the thermistor in still air, and in well-stirred oil. Typical

    values for a small glass bead thermistor are 1.5 mW/°C in still air and 6.0 mW/°C in stirred oil. If

    the temperature of the environment is known beforehand, then a thermistor may be used to

    measure the value of the dissipation constant. For example, the thermistor may be used as a flow

    rate sensor, since the dissipation constant increases with the rate of flow of a fluid past thethermistor.

    5. Applications

    PTC thermistors can be used as current-limiting devices for circuit protection, as replacements for

    fuses. Current through the device causes a small amount of resistive heating. If the current is large

    enough to generate more heat than the device can lose to its surroundings, the device heats up,

    causing its resistance to increase, and therefore causing even more heating. This creates a self-

    reinforcing effect that drives the resistance upwards, reducing the current and voltage available tothe device.

    PTC thermistors are used as timers in the degaussing coil circuit of CRT displays and televisions.

    When the unit is initially switched on, current flows through the thermistor and degauss coil. The

    coil and thermistor are intentionally sized so that the current flow will heat the thermistor to the

     point that the degauss coil shuts off in under a second.

     NTC thermistors are used as resistance thermometers in low-temperature measurements of the

    order of 10 K.

     NTC thermistors can be used as inrush-current limiting devices in power supply circuits. They

     present a higher resistance initially which prevents large currents from flowing at turn-on, and

    then heat up and become much lower resistance to allow higher current flow during normal

    operation. These thermistors are usually much larger than measuring type thermistors, and are

     purposely designed for this application.

     NTC thermistors are regularly used in automotive applications. For example, they monitor things

    like coolant temperature and/or oil temperature inside the engine and provide data to the ECU

    and, indirectly, to the dashboard. They can be also used to monitor temperature of an incubator.

    Thermistors are also commonly used in