e u r o t r o n TEST & MEASUREMENTS EUROTRON Instruments S.p.A. Portable Temperature Calibrators Bulletin 09-06.3 E MicroCal T series Ranges from -50°C to 1100°C High stability and uniformity Fast cool-dow n design Microprocessor temperature controller RS232 seri al interface CalpMan 2000 documenting P C software Eurotron - One of the leaders in high t echnology equipment: high accuracy calibrators - non-contact IR thermometers - portable f lue gas analyzers. ns I trn uo mr t eo nr tu sE n s I t r u m t e n t s E ISO 9001 Qd ue t aln ita yr G
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Portable Dry-block and FluidTemperature Calibrators
GENERAL M icroCal T100 M icroCal T350
Designed for Testing and ca librating temp erature sensors Switch test input for therm ostat Compa ct, rugged, and easy to use
Microprocessor temp erature controller optional RS232 serial interface
TM and CalpMan 2000 Windows software
Compa ct, rugged a nd easy to use.
Built-in reference Pt100 with high ac curac y 0.01°C resolution indicator Auxiliary input s for external Pt100 and therm ostat switch-test.
Internal fa n and Peltier unit for fastcool-down.RS232 seria l interfa ce and optional CalpMan 2000
TM Windows software
The temperature parameter representthe most important factor to fulfil quality,operational safety and reliability ofindustrial processes. Thermocouples.resistance thermometers and any othertemperature sensors. When installed ina industrial process, they are subjectedto mechanical, thermal and chemicalstresses which accelerate their ageing.
Therefore it is a recommendedprocedure to inspect, check andcalibrate each sensor during thecommissioning phase and at regular andprogrammed time intervals.The temperature calibrators seriesMicroCal T include five different, ruggedand transportable units designed toobtain a controlled temperature with highstability, high uniformity in all calibratorsranges.
Internal block temperature stabilityA built-in microprocessor basedcontroller handle the man-machine
interface, the internal logic and assuresthe high stability of the internal metalblock through a sophisticatedtemperature algorithm.
A complete Automatic CalibrationSystemThe MicroCal T can be used togetherwith a Personal Computer, amultifunction high precision calibratorMicroCal 200/200+ (see bulletin 08-44)and a CalpMan 2000 Windows® software package to improve a trueautomatic temperature calibrationsystem.Using the RS232 interfaces of theinstruments it is possible to programtrough the PC the complete testprocedure for Your thermocouple orresistance thermometer to be certificate.The MicroCal 200 is able to read thetesting thermo-element with the desiredaccuracy and the software calculate theerrors and prepare the certificate.
Table A Options0 None1 Carrying case3 RS232 serial interface4 blank insert5 2 holes insert ( φ 6.5 - 12.5 mm)
Table B Power supply1 120 V USA plug2 230 V Schuko plug3 240 V UK plug4 230 V European plug
Optional 2 inputs for externa l 3/4 wires Pt100 and Tc J, K, N, R, S.Therm ostat switch-test input .RS232 interfac e a nd opt ional
TM CalpM an 2000 Windows software.Wate r (< 90°C), 47V20 Silicon oil (< 120°C) and 47V100 Silicon oil (< 200°C).
Operating range from Tam b to + 600.0 °C
Built-in reference Pt100 with high ac curac y 0.01°C resolution indicator.Auxiliary input for external Pt100.
Therm ostat switch-test input .Internal fan for rap id c ool-down.RS232 interface a nd optiona l
TM CalpM an 2000 Windows software
Tem pera ture sensors field and laboratory inspection and calibration.Optional 2 inputs for externa l 3/4 wires Pt100 and Tc J, K, N, R, S.Therm ostat switch-test input .
Internal fan for rapid cool-down.Opera ting rang e from + 200 °C to + 1100 °C RS232 interface a nd optiona l
TM CalpMan 2000 Windows software
Table A Options0 None1 Alluminium carrying case + EE460016
and EE460017 inserts + extration tool2 Maximum temperature +600°C4 Vinyl carrying case7 Special drilling for 50mm block (550°C)7X Special drilling for 65mm block (550°C)8 Special drilling for 50mm block (600°C)8X Special drilling for 65mm block (600°C)
Table B Power supply1 120 V USA plug2 230 V Schuko plug3 240 V UK plug4 230 V European plug5 100 V Japan/USA plug
Table C Report of calibration1 Eurotron Report
Table A Options0 None
2 n.2 Tc & Pt100 inputs9 Special φ44mm block
Table B Power supply1 120 V USA plug2 230 V Schuko plug3 240 V UK plug4 230 V European plug5 100 V Japan/USA plug
Table C Report of calibration1 Eurotron Report
Table A Options0 None
1 Vinyl carrying case2 n.2 Tc & Pt100 inputs
Table B Power supply1 120 V USA plug2 230 V Schuko plug3 240 V UK plug4 230 V European plug
Table C Report of calibration1 Eurotron Report
ORDERING CODE ORDERI NG CODE ORDERI NG CODE
Cat . 8174 - A - B - C Cat . 8176 - A - B - C Cat . 8152 - A - B - C
Metal inserts for T350 (nr. / holes -ID):cat. EE460037 undrilledcat. EE460038 (2 / 6.5-12.5 mm)
Custom inserts and blocks diameters are available on request.
Model MicroCal T350 MicroCal T100 MicroCal T200Fluid
MicroCal T500 MicroCal T1100
Temp. range °C amb. to 350 -50 from amb. to 150 amb. to 200 amb. to 550 (600) 200 to 1100*Stability °C ± 0,1 ± 0,03 ± 0,02 @150°C ± 0,05 @ 450 °C ± 0,3 @ 1000°CUniformity