Transcript

1

Automated Hematology:

An Overview

2

Session Outline• Principles of Automation• Initial Setup• Calibration• Quality Control• Flagging• Troubleshooting• Case Studies

3

Objectives• Describe the electrical impedance and light

scatter principle for performing cell counts• Explain the underlying causes of invalid

automated hematology results. • Utilize quality control procedures to determine

if patient results are acceptable• Recognize the significance of flagging and

take appropriate actions

4

ABX Micos 60

Range of Technologies in Zambia

5

ABX Pentra 60 C+

6

Hematology Automation• Two General Principles

Electronic resistance ( impedance) Light scattering

7

Cell counting• Coulter Principle

Dilution Vacuum and

pressure Electrical impedance Reagent systems

8

Electronic Resistance (Impedance)• Utilizes non-conductive properties of blood cells

as blood cell passes through orifice of aperture it displaces its own volume

increased resistance between electrodes results in an electrical pulse

RBCs and Platelets counted together, separated by pulse heights

hydrodynamic focusing forces cells to pass single file through sensing zone

9

Light scattering• Cells counted as passed through focused beam of light( LASER) • Sum of diffraction(bending around corners), refraction (bending

due to change in speed) and reflection (light rays turned back by obstruction).

• Multi angle polarized scatter separation (M.A.P.S.S) 0° : indicator of cell size 10° : indicator of cell structure and complexity 90° polarized: indicates nuclear lobularity 90° depolarized: differentiate eosinophils

10

TechnologiesABX

11

12

Initial Setup

13

Out of the BoxFor the service Engineer

• Check instrument for visual damage• Check for any loose parts or connections• Make sure all computer boards are properly sealed• Check the socket to verify proper voltage outlet• Plug instrument power cord into (voltage stabilizer)

electrical supply• Confirm the correct voltage on instrument

Main power supply Photometric voltage Any other voltage supply that is pertinent to instrument

functions

14

Out of the BoxFor the service Engineer

• Permit instrument to stabilize/equilibrate Let all components reach proper

temperature• Set in any parameters that may be required

Ranges Temps

15

When to Calibrate

You should calibrate your instrument:• At installation• After the replacement of any component that

involves dilution characteristics or the primary measurements (such as the apertures)

• When advised to do so by your service representative

16

Calibration• Calibration fine tunes your hematology analyzer and

provides the most accurate results possible.• In the normal process of tracking data for an extended

period of time, your laboratory can make a specific decision to recalibrate a given parameter. Never adjust to a specific value for an individual sample.

• For best performance, calibrate all the CBC parameters. The WBC differential is calibrated at the factory. They do not require calibration in the laboratory.

17

Performing Reproducibility Check – CBC (N=10)

Sample Requirements - For reproducibility studies, ensure the patient for the sample that is being tested:

• Is receiving no medication• Has normal hematologic parameters, with a WBC count of 5.0 ± 1.0.• Has normal erythrocyte, leukocyte, and platelet morphology and, if

you want to check the Diff parameters, with Diff values Neutrophils 40 to 72% Lymphocytes 17 to 45% Monocytes 4 to 12% Eosinophils 0 to 10% Basophils 0 to 1%Ensure you have enough normal whole blood from a single donor for 11 cycles.

18

Daily startup procedures(Daily Maintenance)

• Daily cleaning• Background counts• Electronic checks• Compare open and closed mode sampling (use a

normal patient sample)• Run controls

Must be within specified limits

19

Ensure the Instrument is Functioning Properly

1. Check the reagent containers for: Sufficient quantity Not beyond expiration date No precipitates, turbidity, particulate

matter, or unusual color Proper connections between the

instrument and the reagent containers

20

2. Check the waste container for: Sufficient capacity Proper connections

3. Perform daily startup4. In addition to verifying daily startup results, verify

acceptable: Reproducibility Carryover Control Results

Ensure the Instrument is Functioning Properly

21

Quality Control

22

Quality Control• Purpose of QC

Assure proper functionality of instrumentation Means of assuring accuracy of unknowns Monitoring the Integrity of the Calibration

- When controls begin to show evidence of unusual trends- When controls exceed the manufacturer’s defined acceptable

limits

23

Quality Control Methods• Assayed or Unassayed stabilized material

(Commercial)• Previously analyzed patient samples

Easily obtained Cost effective Results and samples readily available

24

QC Methods: Assayed or Unassayed Stabilized Material

Commercially available Known values (Assayed only) Analyze low, normal and high control Results stored in the instrument computer

(Pentra only) Monitored with Levy-Jennings charts

- Easily illustrates trends and shifts

25

Quality Control• OUT OF CONTROL!!!

Repeat the assay ( One time occurrence ) Check for trends (from Levy Jennings) Check integrity of material Troubleshoot Verify instrumentation

26

Remedial Action to Take When a Control is Outside its Expected Range

1. Ensure the control• Material was mixed properly. If not, mix it according to the

package insert.• Identification information was entered correctly. If using

the Numeric Keypad, ensure you typed the correct information.

• Setup information (assigned values and expected ranges) matches the control package insert for the current lot number being used. If they do not match, change the control’s information to match the package insert.

27

Remedial Action to Take When a Control is Outside its Expected Range

2. If any of the problems still exist, rerun the control; otherwise, proceed to the next step

3. Rerun the control to ensure the problem was not a statistical outlier.

4. Ensure the control material was not contaminated by running another vial or level of control.

Automated Hematology

Problem Solving – Troubleshooting

Specimen – Related OR Instrument

29

Data Review• A review of instrument data, such as

background, control, and blood sample results, is helpful in detecting problems.

• Sometimes a questionable blood sample result is the only symptom of subtle reagent or pneumatic problems.

30

• An instrument problem is differentiated from a specimen-related problem by running a control.

• If the control results are acceptable, the problem is probably specimen-related. Check for: clots hemolysis lipemia

Specimen-Related Problems

31

Instrument Problems• If the control shows similar problems, it indicates an

instrument problem. Electronic? Pneumatic / Hydraulic? Reagent?

• Because it is easiest to detect a problem in the electronic subsystem and hardest to detect a problem in the reagent subsystem, the subsystems are usually checked in the following order: electronic, pneumatic / hydraulic, reagent.

32

Electronic Troubleshooting• Detecting a problem in the electronic subsystem –

or eliminating the electronic subsystem as the source of the problem – is simplified by indicators and electronic tests.

• Correcting Electronic Problems: minor problems,such as loose cables most electronic problems require the

assistance of your instrument service representative.

33

Pneumatic / Hydraulic Troubleshooting• Most pneumatic / hydraulic problems are detected by

observing the Diluter section in operation.• When you identify a symptom of a malfunction, try to

isolate the malfunction to the specific part of the cycle, for example, during preparation, counting, or cleanup.

• Then, try to isolate the malfunction to the specific components and tubing.

• Next, look for one of four possible problems – pinched tubing, plugs, leaks, or defective components.

34

Correcting Pneumatic/Hydraulic Problems

You can correct most pneumatic / hydraulic problems, including defective components. Tubing, (Need other examples) Follow manufacturer’s instructions Sticking float in moisture chamber

35

Reagent Troubleshooting• A reagent problem can be as obvious as precipitate

in the reagent tubing.• In the less obvious cases, the most effective way of

detecting a problem is by keeping a log of the lot numbers with the opening and expiration dates of the reagents in use, and knowing how each reagent affects the data.

• Refer to the labeling information with your reagents for details.

36

Correcting Reagent Problems

• You can correct most reagent problems by: changing the container of reagent priming the instrument with the new

reagent.

37

Flagging• Condition flags

• Describes cell populationnormalabnormal

• WBC Suspect flags Blasts Imm Grans/Bands 1 Imm Grans/Bands 2 Variant lymphs Review Slide

Check Slide

38

Check Slide

More Flagging• RBC Suspect flags

NRBCs Macrocytic RBCs Dimorphic RBC population Micro RBCs/RBC fragments RBC agglutination

• Definitive Flagging Based on predetermined lab limits Provide information for review

39

Review ofHistograms

as aQualityControl

Test

40

Histograms• RBC, PLT, and WBC

plotted on histogram• X-Axis

Cell size in femtoliters (fL)

• Y-Axis # of cells

41

Histogram WBC: Distribution with three individual peaks and valleys at specific regions representing the lymphocytes, monocytes, and granulocytes. All curves normally start and end at baseline

42

RBC Histogram As A Quality Control Tool

ABN / INDICATOR PROBABLE CAUSE COMMENTLeft of curve does not touch baseline

Schistocytes and extremely small red cells

Review smear CBC and Platelet histogram

Bimodal peak Transfused cells, therapeutic response

Review Smear

Right portion of curve extended

Red cell autoagglutination

Review CBC & Smear

Left shift of curve Microcytes Review smear & CBC

Right shift of curve Macrocytes Review smear & CBC

43

WBC Histogram As A Quality Control Tool

ABN / INDICATOR PROBABLE CAUSE COMMENTTrail extending downward at extreme left, or lymph peak not starting at baseline

NRBC, Plt clumping, unlysed RBC, cryoproteins, parasites

Review smear and correct WBC for NRBC

Peak to the left of lymph peak or widening of lymph peak towards left

NRBC Review smear & correct WBC for NRBC

Widening of lymph peak to right

Atypical lymphs, blasts, plasma cells, hairy cells, eosinophilia, basophilia

Review smear

Wider mono peak Monocytosis, plasma cells, eosinophilia, basophilia, blasts

Review smear

44

WBC Histogram As A Quality Control Tool

ABN / INDICATOR PROBABLE CAUSE COMMENT

WBC histogram (lymph peak) does not start at baseline

Giant platelets, NRBC, Plt clumping

Review smear, correct WBC for NRBC

Elevation of left portion of granulocyte

Left Shift Review smear

Elevation of right portion of granulocyte peak

Neutrophilia Review smear

45

Platelet Histogram As A Quality Control Tool

ABN / INDICATOR PROBABLE CAUSE COMMENTPeak or spike at left end of histogram (2-8 fl)

Cytoplasmic fragments

Review smear

Spike towards right end of histogram

Schistocytes, microcytes, giant platelets

Review smear + CBC( MCV & RDW)( MPV & PDW)

Bimodal peak Cytoplasmic fragments

Review smear

46

Interpreting Test Results

Misleading results can occur if the specimen was not properly collected, stored, or transported. The following situations can also yield misleading results for the parameters listed:

47

Interfering substances• Cold Agglutinins• Cryoglobulins• Lipemia• Platelet Clumps• RBC fragments• NRBCs

48

Interferences Cont. All Parameters:

WBC:

Clotted specimen. Inspect every specimen for clots

Unusual RBC abnormalities that resist lysing

malarial parasites giant platelets, platelet clumps, NRBCs fragmented white cells, agglutinated white

cells, lyse-resistant red cells, cryoglobulin, some extremely elevate

proteins, unlysed particles greater than 35 fL in size

49

Interferences

• RBC: Very high WBC count, high concentration of very large platelets, auto-agglutination

• Hgb: Very high WBC count, severe lipemia, heparin, certain unusual RBC abnormalities that resist lysing

• MCV: Very high WBC count, high concentration of very large platelets, auto-agglutination

• RDW: Very high WBC count, high concentration of very large platelets, auto-agglutination

• Plt: Giant platelets, platelet clumps, white cell fragments, electronic noise, very small red cells, red cell fragments

• Hct: Known references related to RBC and MCV

50

Interferences

• MCH: Known interferences related to Hgb and RBC

• MCHC: Known interferences related to Hgb, RBC, and MCV

Troubleshooting Flagged Results:

• Refer to the Instrument Manual to troubleshoot codes, flags, and messages displayed with patient results

top related