Line of ProductsLine of ProductsLine of ProductsLine of Products ●●●●ElectroElectroElectroElectro----dynamic Vibration Test Systemsdynamic Vibration Test Systemsdynamic Vibration Test Systemsdynamic Vibration Test Systems ●●●●Vibration ControllersVibration ControllersVibration ControllersVibration Controllers ●●●●Vibration MetersVibration MetersVibration MetersVibration Meters
e-mail for inquiry: inquiry@shinken-ltd.co.jp
HEAD OFFICE : KOJIMACHI SANKO BLDG. 3F,
6-4 KOJIMACHI CHIYODA-KU,
TOKYO, 102-0083, JAPAN
TEL:03(3261)3211 FAX:03(3264)2930
SHOCK & Vibration Test System G-5N Series
Model:G22Model:G22Model:G22Model:G22----121212125N5N5N5N Model:G22Model:G22Model:G22Model:G22----250NS250NS250NS250NS
500G Shock
GR-63 Seismic Wave Simulation 《Japanese Patent No. 3035280》
《《《《Sine ControlSine ControlSine ControlSine Control》》》》 《《《《PSD Random ControlPSD Random ControlPSD Random ControlPSD Random Control》》》》 《《《《ClassicClassicClassicClassical Shockal Shockal Shockal Shock》》》》
Challenging to Any Applications Related to Vibration・・・
Newly Patented
Technology Used
Simple-Function Digital Sine Controller
Model : G01-035
Fixed-Frequency Sine Tests & Swept Sine Tests of
Displacement Constant, Acceleration Constant and
Displacement-Acceleration Constant Easily Done
Up-to-date Digital Vibration Controller D-58 Series
Sine Control
PSD Random Control
Classical Shock Control
Road Simulation
Shock Response Spectrum
Sine on Random
Random on Random
profile(t)
high-abort(t)
low-abort(t)
control(t)
2-1 -0.9 -0.6 -0.3 0 0.3 0.6 0.9 1.2 1.5 1.8
680.0000
-320.0000-300.0000
-200.0000
-100.0000
0
100.0000
200.0000
300.0000
400.0000
500.0000
600.0000
Time (Milliseconds)
gn
The shock perfomance curve above iscalculated for the left half- sine shockwave in the case of ideal cont rol.
T : Shock Durat ion (msec)
T
A
1/ 5A A : Shock Accelerat ion(m/ s <G>)2
5 10 50 1001
1960(200)
980(100)
490(50)
98(10)
2
2
Max. Acc. 4900m/s <500G>(with 0.1kg load)4900(500)
Acceleration(Acc.)
m/s (G)
Max. Velocity
3m/ s
Durat ion(msec)
2
Max. Acc. 1649m/s <168G>(with no load)
Max. Acc. 2726m/s <278G>(with no load)2
Max. Acc. 2182m/s <222G>(with no load)2
Max. D
isplac
em
ent 1
00m
mp-p
Max. Acc. 2691m/s <274G>(with no load)2
Max. Acc. 1914m/s <195G>(with no load)2
Max. Acc. 1856m/s <189G>(with no load)2
Max. D
isplacem
ent 1
50m
mp-
pM
ax. Displac
ement 2
60m
mp-
p
Max. D
isplacem
ent 2
60m
mp-
p
Max. Displacement 150mmp- p
Max. D
isplacem
ent 20
0m
mp-
p
Max. Displacement 200mmp- p
<G-5220N>
<G-5250NS>
<G-5230NS>
<G-5230N>
<G-5212N>
<G-5125N>
<G-5160N>
● Distributor: EMS Test Measurement Sdn Bhd (Malaysia)
PT EMS Technology (Indonesia)
sales@ems-info.com.my
G-5N Series Shock & Vibration Test Systems can be used for:
Vibration Tests, Shock Tests and Seismic Simulation Tests!!
Has Developed Vibration Generators with New Structure!!
Generally speaking, with the Electro-dynamic Vibration Generators, periodical replacement of Bellows (Air Springs) and Roll Springs
for upholding static mass (specimens) is required under severe vibration tests, and so is of mechanical parts (Roller Bearings,
Rocking Arms, U-shaped or Loop Springs etc) for guiding Vibration Tables with Driving Coils.
has succeeded in developing
the new-structure Vibration Generators which
require no such replacements!!
The inner structure of the Vibration Generator
is prominent with the employment of the Air
Bearing for guiding the Vibration Table with
the Driving Coil and the unique axial suspension
by storing in the Air Room the compressed air
used for the Air Bearing, with air quantity in
the Room controlled proportional to the weight of
a specimen, keeping the vibration table at center.
Enhanced Reliability permitted by Air Bearing No Worry About Wear & Tear of Bellows
Guide due to no mechanical friction involved and due to seal-less axial suspension structure by
reduction in No. of parts, especially consumables! re-using the compressed air for air bearing!
Long Stroke of up to 260mmp-p Easy Check by opening the Upper Cover only
despite of small systems, suited to Shock and <No need for disassembling them deep inside
Seismic Simulation Tests!! as used to>
● System Specifications
The line-up of the G-5N Series Vibration Test Systems are as follows:
System Model
G-5125N G-5160N G-5212N G-5220N G-5230N G-5230NS G-5250NS
Sub-system Model
1 Controller G01-035/D-58 Series
2
Power Amplifier G14-007 G14-04 G14-021 G14-028 G14-042 G14-042 G14-063
Field Power Supply G31-002 G31-003 G31-005 G31-007 G31-009 G31-009 G31-012
3 Vibration Generator G22-125N G22-160N G22-212N G22-220N G22-230N G22-230NS G22-250NS
4 Cooling Blower 0.75kW 1.5kW 7.5kW 11kW 15kW
● System Specifications
System Model G-5125N G-5160N G-5212N G-5220N G-5230N G-5230NS G-5250NS
Sho
ck
Max. Shock Force KN (Kgf) 7.5(765) 15(1530) 24(2450) 35(3570) 61.2(6250) 61.2(6250) 85.7(8570)
Max. Acceleration m/s² (G) 4900(500)
100g load
2726(278)
No load
2182(222)
No load
2691(274)
No load
1914(195)
No load
1856(189)
No load
1649(168)
No load
Max. Velocity m/s 3.0 <may be changed to 4m/s by reducing Shock Force by 25% >
Pulse Width msec 1~70 1~100 1~70 1~50 1~100 2~100 2~100
Vib
rati
on
Max. Force KN (Kgf) 2.5(255) 6(612) 12(1224) 20(2040) 30(3061) 30(3061) 49(5000)
Max. Acceleration m/s² (G) 1000(102)
1.1kg load
1000(102)
0.5kg load
1000(102)
1kg load
1537(156)
no load
937(95)
no load
857(87)
no load
942(96)
no load
Max. Velocity m/s 2.0
Frequency Range Hz 3 (0.4*)~2000 0.4~1400 0.4~1000
Co
mm
on
Max. Displacement mmp-p★ 150 200 150 100 200 260 260
Movable Weight Kg 1.4 5.5 11 13 32 35 52
Features High Shock Acc. Long-stroke - High Vib. Acc. Long-stroke GR-63 GR-63
Power Required 3φKVA 10 18 28 30 50 50 90
Max. Payload Kg 20 50 150 150 370 370 370
★ Relevant accelerometer and charge amplifier are needed for low frequency tests from 0.4Hz. ★ Between Electric Limits (for sine tests, effective displacement is around 80% of the max. displacement .with pneumatic isolators ON)
● Max. Performance Curve ⅠⅠⅠⅠ <SINE>
Guide Roller Stopping
Table from Rotating
Air Bearing
Vibration Table
Compressed Air In
Lead Plate
Air Out<Adjusted with Valve>
Driving Coil
Field Coil
Air Out Air Out
Air Room
5 10 50 1K100 500 2KFrequency(Hz)
980(100)
490(50)
98(10)
49(5)
1960(200)
10.5
9.8(1)
4.9(0.5)
Max
. Dis
plac
ement
210m
mp-
p
Max
. Veloc
ity 2
m/s
Max. Acc. 1537m/ s (156G)<with no load>2
Max. Displacement 160mmp- p
Max
. D
ispl
acem
ent
80m
mp-
p
Max
. D
ispl
acem
ent
210
mm
p-p
Max
. Dis
plac
ement
120m
mp-
p
Max. Acc. 937m/ s (95G)<with no load>
Max. Acc. 942m/ s (96G)<with no load>
Max. Acc. 857m/ s (87G)<with no load>2
2
2
Max. Acc. 1000m/ s (102G)2
<with 1kg load><with 0.5kg load><with 1.1kg load>
<G-5220N>
<G-5250NS>
<G-5230NS>
<G-5230N>
<G-5212N>
<G-5125N>
<G-5160N>
Acceleration(Acc.)
m/s (G)2