Effects of paper and plastic bags on the performance of Thermo Scientific portable XRF analyzers
Ap
plica
tion
No
te
IntroductionField Portable X-Ray Fluorescence (FPXRF) instruments, including more than 3,000 Thermo ScientificTM portable XRF analyzers, are used extensively in various stages of mining activity from grass root exploration to exploitation, ore grade control, and even environmental investigations. With or without sample preparation, FPXRF analyzers deliver fast, accurate and repeatable results on samples containing a broad range of elements from magnesium (Mg) to uranium (U).
Handheld Thermo ScientificTM NitonTM XL3t and Niton XL2 Series analyzers and the Niton FXL field x-ray lab analyzers are making a critical difference in mining industry operations worldwide, providing geologists with new capabilities to prequalify and reduce laboratory samples, with exceptional portability and ease-of-use.
ApplicationFPXRF is a surface analysis technique that is capable of analyzing more than 30 elements in seconds in any type of sample, from soil to rock chips and drill cores. In some cases, depending on the sample type and required accuracy, clients may want to assay their samples in a paper or plastic bag. Depending on the thickness of bag and its composition, the number of x-rays reaching the detector will be lower than with direct assay. In such cases, calibration of the instrument using “cal-factors” is necessary. However, the overall time savings gained by eliminating various sample preparation steps can be significant in some projects.
MethodPulverized samples from a sedimentary-hosted lead-zinc deposit were analyzed in both plastic (0.06 mm thick or 2-2.5 mils, 38 samples) and paper (0.11 mm thick or 4-5 mils; 18 samples) bags (Figure 1) and the results were compared to their lab assays (Figure 2). Samples were crushed and pulverized in mill steel to 95% passing 105 micron mesh. The laboratory assay methods used were a combination of instrumental neutron activation (INAA) with a 30g aliquot and inductively coupled plasma optical emission spectrometry (ICP-OES), with a four acid, near- total sample digestion of 0.25 g. The portable analyzer used was an XL3t-500 series with PIN detector using a filter time of 30 seconds per filter with analysis through the sample bags returned from the lab.
Thermo Scientific XRF analyzers use Mining Mode and Soil Mode. Mining Mode is based on a Fundamental Parameters Calibration and relies on the detector’s response to pure element spectra, whereas Soil Mode uses Compton Normalization where scatter (Compton Peak) in spectrum is related to sample matrix. Mining Mode provides suitable assay data on samples tested in a paper or plastic bag. Preliminary investigation of the samples and their assay data indicate that Soil Mode is not suitable due to high concentration of elements such as Fe, Pb and Zn. The summary of results is shown in Table 1.
Figure 1. Plastic and paper bags used in this study.
2
ResultsAs Figure 2 shows, there are good correlations between lab assay data and portable XRF results. Mining Mode is a preferred method for elements in the ranges of concentrations shown in Table 1. If an application does not need very low levels of detection, the analyzer can be calibrated to the sample through various media.
Figure 2. Comparison of lab data with those from portable XRF in Mining Mode used on samples in paper and plastic bags.
Element Bag ConcentrationLOD
Plastic Bag Paper Bag
Ag Plastic <60 ppm 16 15
As Plastic <70 ppm 88 78
Ca Plastic <11.1% 560 882
Cd Plastic/paper <1700 ppm 16 16
Cu Plastic/paper <1200 ppm 94 90
Fe Plastic/paper <40% 13245 11088
K Plastic <1.7% 1032 900
Mn Plastic/paper <2% 716 642
Ni Plastic/paper <650 ppm 198 176
Pb Plastic/paper <9% 546 446
S Plastic <35% 20954 19154
Sr Plastic <300 ppm 44 42
Ti Plastic <4500 ppm 213 171
Zn Plastic/paper <50% 252 217
Table 1: Recommended bag type based on concentration and composition of samples used in this study. The concentration column shows the maximum concentration in the samples used in this study. Limits of Detection (LOD) are shown for each bag type using Mining Mode.
Ag
(ppm
)C
a (p
pm)
As
(ppm
)C
d (p
pm)
Niton and Pulps Comparison
7/11/2013 page 1 of 1
0
10
20
30
40
50
60
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Ag
(ppm
)
Lab
Plastic Bag
Paper Bag
0
10
20
30
40
50
60
70
80
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
As
(ppm
)
Lab Plastic Bag Paper Bag
0
20000
40000
60000
80000
100000
120000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Ca (p
pm)
Lab Plastic Bag Paper Bag
0
100
200
300
400
500
600
700
800
900
1000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Cd (p
pm)
Lab Plastic Bag Paper Bag
3
Cont. Figure 2.
0
200
400
600
800
1000
1200
1400
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Cu (p
pm)
Lab Plastic Bag Paper Bag
0
50000
100000
150000
200000
250000
300000
350000
400000
450000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Fe (p
pm)
Lab Plastic Bag Paper Bag
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
K (p
pm)
Lab Plastic Bag Paper Bag
0
5000
10000
15000
20000
25000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Mn
(ppm
)
Lab Plastic Bag Paper Bag
0
200
400
600
800
1000
1200
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Ni (
ppm
)
Lab Plastic Bag Paper Bag
0
10000
20000
30000
40000
50000
60000
70000
80000
90000
100000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Pb (p
pm)
Lab Plastic Bag Paper Bag
Cu
(ppm
)
Fe (p
pm)
K (p
pm)
Ni (
ppm
)
Mn
(ppm
)P
b (p
pm)
0
500
1000
1500
2000
2500
3000
3500
4000
4500
5000
A00
001
A00
002
A00
003
A00
004
A00
005
A00
006
A00
007
A00
008
A00
009
A00
010
A00
011
A00
012
A00
013
A00
014
A00
015
A00
016
A00
017
A00
018
A00
019
A00
020
A00
021
A00
022
A00
023
A00
024
A00
025
A00
026
A00
027
A00
028
A00
029
A00
030
A00
031
A00
032
A00
033
A00
034
A00
035
A00
036
A00
037
A00
038
Ti (p
pm)
Lab Plastic Bag Paper Bag
0
100000
200000
300000
400000
500000
600000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Zn (p
pm)
Lab Plastic Bag Paper Bag
0
200
400
600
800
1000
1200
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
Sr (p
pm)
Lab Plastic Bag Paper Bag
0
50000
100000
150000
200000
250000
300000
350000
400000
A000
01
A000
02
A000
03
A000
04
A000
05
A000
06
A000
07
A000
08
A000
09
A000
10
A000
11
A000
12
A000
13
A000
14
A000
15
A000
16
A000
17
A000
18
A000
19
A000
20
A000
21
A000
22
A000
23
A000
24
A000
25
A000
26
A000
27
A000
28
A000
29
A000
30
A000
31
A000
32
A000
33
A000
34
A000
35
A000
36
A000
37
A000
38
S (p
pm)
S Plastic Mining Pb Mining Zn Mining
Cont. Figure 2.
S (p
pm)
Sr (
ppm
)
Ti (p
pm)
Zn (p
pm)
ConclusionThis case study shows that for many applications, portable XRF analysis of samples through a plastic bag or paper bag yields valuable and reliable data that can be used to make timely decisions in the field. Furthermore, the study documents that Mining Mode (for most sample matrices) is preferred if the user wishes to analyze samples in a bag.
To discuss your particular applications and performance requirements, or to schedule an on-site demonstration, please contact your local Thermo Scientific portable analyzer representative or contact us directly by email at niton@thermofisher.com, or visit our website at www.thermoscientific.com/niton.
We would like to thank private consultant Cris Carman for his technical and geological support.
5-332 07/2013
thermoscientific.com/niton © 2013 Thermo Fisher Scientific Inc. All rights reserved. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
AmericasBoston, MA USA+1 978 670 7460niton@thermofisher.com
Europe, Middle East, Africa & South AsiaMunich, Germany+49 89 3681 380niton.eur@thermofisher.com
Asia PacificNew Territories, Hong Kong+852 2885 4613niton.asia@thermofisher.com
In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world.
top related