Transcript
Complete range of
Electro Surgical Solutions
Add Hi-Tech Precision to your Surgical Expertise...
H i g h P e r f o r m a n c e
With al l Reusable Attachments
Argon Plasma Coagulator
OUTPUT FLOW RANGE• Cut and Coag Flow rates:0.1-12 Standard Ltr/Min. with resolution of - 0.1 Ltr/Min for flowrates between 0.1-5 lpm - 0.2 Ltr/Min for flowrates between 5-10 lpm - 0.5 Ltr/Min for flowrates between 10-12 lpm
MODES• Cut : Open, Laparo
Coagulation : Open, Laparo, Gastro, Pulse• Introduced new Pulse mode with various pulse effects.•
GAS FILTRATION SYSTEM0.5 Micron external filter for Gas input0.1 Micron external filter for Gas output
INDICATOR
• Cylinder Gas level indicator : 0 to 100%
• Visual indicator for system warnings & errors.
• Beep tones for warning and errors.
• Over pressure indicator
DESIGN STANDARDClassification according to the EC- Directive 93/ 42/ EEC: Class IIbGeneral Safety Standards: IEC 60601-1-1, IEC 60601-2-2 EMI/EMC Compatibility Standard: IEC 60601-1-2IS Safety Standard: IS 7583
INPUT POWER (other ratings available on request)Operating Voltage : 165-265 Volt AC, 50-60 Hz, Fuse : 2 AFrequency : 50-60 Hz Current : 2 Amp
DIMENSIONSL x W x H in cm - 38.0 x 31.0 x 12.0, Weight in kg < 8.0
• Unit with 5" LCD Touch Panel.
• Fully microprocessor controlled : To achieve high degree of clinical precision.
• Argon Gas Level indicator with built in high pressure regulator.
• Separate flow setting for cut & coagulation.
• Device can switch between Electro-surgery Cut/Coag and Argon assist cut/coag.
• User set able flow rate range : 0.1 to 12 Ltr/Min.
• Internal purging facility which fill the accessory with non-ionized gas before surgery.
• Facility to detect over pressure from body and give an alarm in such condition.
• Internal leakage detection for Argon gas.
• Accessory selection option provided for Gastroscopy, Open and Laparoscopy surgery.
• Ten user settable programs with procedure, surgeon name, flow & modes can be saved in the system.
• Single hose pipe connection between the system and the cylinder.
• Decreases smoke while performing cut and coagulation.
• Its is based non-contact coagulation technique, it performs coagulation with less carbonization.
• Decreases blood loss, rebleeding, minimal tissue damage.
• Argon is inert, non-combustible and easily ionizable gas.
• The RF energy ionizes argon gas and produces plasma which helps in non-contact coagulation &
improves the quality of coagulation.
• Heavier than air, displacing Nitrogen & Oxygen have avoid tissue burning & smoke.
• Produces flexible eschar.
• Gastroscopy • Colonoscopy • Bronchoscopy • Laparoscopy Surgery • Onco Surgery, Open and General Surgical applications
XM
T/M
K/A
P1 r
ev
: 04 J
UL/1
9 C
opyr
ight by
XM
TP
L
INPUT GAS PRESSURE• Maximum input pressure of Argon Gas 150 Bar
OUTPUT GAS PRESSURE• 2.5 ± 0.5 bar
top related