Alarm Management

IMPLEMENTING EVIDENCE

BASED PRACTICE TO REDUCE THE FREQUENCY OF PHYSIOLOGICAL

ALARMS Julie Zimmerman, RN, MS, CCRN, CNSAlbert Lobato, Telemetry TechnicianMark Miller, RN, USD CNS Graduate

Student

The Joint Commission 2014 National Patient Safety Goal 6

REDUCE THE HARM ASSOCIATED WITH CLINICAL ALARM SYSTEMS

Clinical alarm systems are intended to alert caregivers of potential patient problems, but if they are not properly managed, they can compromise patient safety.

CONCERNS: ▪ Numerous Alarms & Alarm devices ▪ Increased exposure to alarms ▪ Staff desensitization▪ Delayed responses & missed alarm signals▪ Improper modification/disabling of alarms ▪ Non-Actionable alarm settings/limits

Alarms – The Problem

Palomar Health Medical Device Alarm Safety Committee, 2013.

American Association of Critical Care Nurses Alarm Management

Recommendations

Expected Practice▪ Provide proper skin preparation for ECG electrodes. (Level B) ▪ Change ECG electrodes daily. (Level E) ▪ Customize alarm parameters and levels on ECG monitors. (Level E) ▪ Customize delay and threshold settings on oxygen saturation via

pulse oximetry (SpO2) monitors. (Level E) ▪ Provide initial and ongoing education about devices with alarms.

(Level E) ▪ Establish interprofessional teams to address issues related to

alarms, such as the development of policies and procedures. (Level E)

▪ Monitor only those patients with clinical indications for monitoring. (Level C)

AACN Best Practice Alert, May 2013.

HOW ARE WE DOING? ALARM FREQUENCY

Data collection on 10 patients admitted for at least 24 hours on: 2 East, 10 East CCU, and TICU

2 10

CCU

TICU

0

100

200

300

400

500

600

700

800

LOW ALERT YELLOW HIGH ALERT REDTOTAL IN 24HRS

SaO² L

OW

RR LOW

HR L

OW

0

100

200

300

400

500

2 East

10 East

CCU

TICU

Staff Survey Results from 2 East, 10 East CCU, and TICU

HOW ARE WE DOING? CURRENT ECG PRACTICE

2 EAST PILOT PRACTICE CHANGES & NURSING

ACTIONS▪ Provide proper skin preparation for ECG electrodes.▪ Properly apply and replace ECG electrodes daily.▪ Individualize alarm parameters▪ Ensure patient/patient family alarm education▪ Promote Alarm Safety- Audibility, Practices, &

Communication

Palomar Health Medical Device Alarm Safety Committee, 2013.

Proper Skin Preparation

▪ Clip excess hair at ECG electrode sites

▪Wash the isolated electrode area with soap and water.

▪Use a rough washcloth or gauze to roughen a small area of the skin.

▪ Ensure electrode sites have properly dried.

▪Do not use alcohol for skin preparation.

Palomar Health Medical Device Alarm Safety Committee, 2013.

FIVE-LEAD SYSTEM

RA – white electrode – Place where the Right arm joins the torso. (below R clavicle, 2nd ICS, R MCL)

RL – green electrode – At the level of the lowest rib. (R lower ribcage, 8th ICS, R MCL)

LA – black electrode – Place where the Left arm joins the torso. (below L clavicle, 2nd ICS, L MCL)

LL – red electrode –At the level of the lowest rib. (L lower ribcage, 8th ICS, L MCL)

V or Chest – brown electrode – placed in V1 (4th ICS, R SB) or V3 position (midway btwn 4th &5th ICS)

R= right, L= left, ICS = intercostal space, MCL – midclavicular line,

SB= sternal border

Correct 5 – Lead ECG Electrode Placement

V1V3

Correct ECG Electrode Placement

5 Lead ECG Site Selection:

▪ Avoid bony prominences

▪ Avoid fatty areas

▪ Avoid major muscles

Palomar Health Medical Device Alarm Safety Committee, 2013.

BEFORE AND AFTER: Proper ECG prep & placement

Palomar Health Medical Device Alarm Safety Committee, 2013.

Individualize Alarm Parameters

o Personalize alarms to meet the needs of individual patients.o Consider patient’s medical condition, status changes, and

activity levels.o Adjust alarms within 1 hour of assuming care.

▪ Individualization of physiological alarm parameters, conducted on a PCU/IMU resulted in a 43% reduction in critical monitor alarms. Additionally, clinical evidence demonstrates personalizing patient alarms reduces the number of false alarms.

MCP 572.1 Management of Clinical Alarms in Patient Care Settings

UCSD Medical Center Practice Guidelines for Assessment and Reassessment, Clinical Alarms

Promote Alarm Safety:

o Confirm ALL clinical alarms are audible from outside patient rooms.

***EVEN when patient doors are closed***

AUDIBILITY

PRACTICESDisconnecting patients from physiological monitors:o Constitutes a change in a patient’s level of ordered careo Can only occur with a required physician/provider order OR,o Only when essential for direct patient care or safety.o Is performed solely by qualified healthcare personnel.o Must NOT be performed by patients/patient family members.

Promote Alarm Safety:

COMMUNICATIONConduct proper communication with the telemetry

technicians.

o Guidance on when to call/notify telemetry technicians: 1) When disconnecting patients from the physiological monitors.

2) Before administration of intravenous cardiac medications/drips.

3) Prior to patient transportation/transfer off of a unit.

4) Upon returning to the unit.

5) Preceding pacemaker interrogation.

6) Prior to admission or discharge.