nice NeOumltech Medical Systems Pvt Ltd

Presentation Flow

COMPANY OVERVIEW

CLINICAL ASPECTS OF nice 2010 SERIES WARMERS

PRODUCT OVERVIEW

TROUBLESHOOTING

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

    Presentation Flow

    COMPANY OVERVIEW

    CLINICAL ASPECTS OF nice 2010 SERIES WARMERS

    PRODUCT OVERVIEW

    TROUBLESHOOTING

    Company INFO

    Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

    for growth and delighted customers

    Timeline-nice Neotech

    2005bullFirst New Factory Unit 1 Opened

    bullIncrease in production Line

    2001bullRegistered as Pvt ltd

    bullProduction started for different models

    2013bullSecond New Factory Unit 2 Opened

    1999bull Neotech Logo

    Designed

    bull Designed first incubator amp open

    care prototype

    2003bull Export to

    Vietnam

    2009-2011

    bull CE certification for products

    2015bull 125+skilled

    employee strong firm

    bull Multiple range of NICU products

    1997bullStarted Neotech Medical Systems with 2 employees

    bullFirst Infant Radiant Warmer

    Why nice Neotech

    User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

    Why Infant Radiant Warmers are Used

    Assist Thermoregulation

    Body warming by providing Far IR radiant heat

    Ensure baby is stabilized after birth

    INFANT RADIANT WARMERS

    MODES OF HEAT LOSS

    Conduction Heat Loss Direct heat transfer from skin to object

    Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

    Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

    (i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

    Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

    Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

    Electromagnetic Spectrum

    Infrared Spectrum

    Effects of Far IR in Infant body

    Far Infrared Rays are invisible waves of energy

    They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

    Types of heaters Quartz

    A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

    Far IR Infrared

    Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

    Ceramic Heaters

    Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

    Tubular Heater

    A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

    Types of heaters(Contrsquodhelliphellip)

    Benefits of Heaterhellip

    QUARTZ

    bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

    CERAM

    IC

    bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

    TUBULAR

    bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

    Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

    A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

    Designed to perfection for your baby

    nice 2010 series Infant Radiant Warmers

    nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

    nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

    nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

    bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

    Features

    How to operate

    Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

    Control Panel Operation

    1 Skin Temperature Display 15

    Manual ModeIndicator

    2 Sensor Failure Indicator 16

    Skin Mode Indicator

    3 Heater Failure Indicator 17

    Pre Warm Mode Indicator

    4 System Failure Indicator 18

    Increase Key

    5 PowerFailure Indicator 19

    Mute Key

    6 Skin Temperature gt 39degC Indicator

    20

    Heater OutputIndicator

    7 Lamp Indicator 21

    Set TemperatureManualAPGAR Timer

    8 Lock Key 22

    APGAR Timer Indicator

    9 Lock Indicator 23

    Mute ON Indicator

    10

    Lamp Key 24

    Manual Timer Indicator

    11

    APGAR Kay 25

    Skin Temperature Low Indicator

    12

    Decrease Key 26

    Skin Temperature High Indicator

    13

    Mode Key 27

    Alarm Indicator

    14

    Safe Mode Indicator

    Operating Modes

    Manual Mode Operation

    1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

    1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

    1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

    1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

    Skin Mode Operation

    Note The skin temperature probe must be properly attached before choosing Skin mode operation

    1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

    1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

    APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

    Operating Modes(contrsquodhellip)

    Operating Modes(contrsquodhellip)

    Pre-Warm Mode

    bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

    Safe Mode

    bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

    Note All the above mentioned modes can only be accessible when the keypad lock is off condition

    AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

    Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

    Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

    Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

    Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

    Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

    Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

    Heater Module

    Parabolic design

    Even Distribution of heat over bed surface

    Extensive safety

    measures(6 Layer safety for thermal

    cut- off)

    Far IR heating ensures

    stimulation of cells for growth

    The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

    Bed Platform Operation

    bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

    bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

    Side Panel Operation

    bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

    X-ray Procedure

    bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

    bullThe heater housing rotates to the left or right for X-ray procedures

    bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

    bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

    Comparison of nice 2010 Infant Radiant Warmers

    nice 2010 BC nice 2010 B nice 2010

    Skin Mode Y Y Y

    Manual Mode Y Y Y

    Safe Mode Y Y Y

    Pre-warm Mode Y Y Y

    X Ray Facility Y Y N

    Bed Y Y N

    Pivoting Warmer ModuleY Y Y

    Apgar Timer Y Y Y

    Height Adjustment N N Y

    Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

    How nice Infant Radiant Warmers fare over competitors

    nice 2010 series warmers

    User-friendly easy to access all parameters

    Long life quartz heater has a long life compared to other heaters

    6 Layer safety for the protection of infant from burning

    6 Layer Safety Features

    Alarms if infant temperature greater or less than Set Temperature

    Thermal cut off if infant temperature exceeds 39 degC

    Heater output reduces to 30 after 15 mins in Manual mode

    Heater output reduces to 30 if electronics senses no variable feedback

    Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

    Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

    TROUBLESHOOTING

    AND MAINTENANCE

    Cleaning Procedure

    1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

    2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

    3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

    Cleaning Solutions for Mattress

    Generic Formulations Maximum concentration level

    Hydrogen Peroxide 6

    Quaternary Compound 05 aqueous solution

    Quaternary Ammonium 02

    Lodophors 50mglitre

    Preventive Maintenance

    1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

    Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

    1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

    TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

    Sl No Problem Cause Remedy

    1 Power failure LED is ON with audio alarm

    Power FailureCheck the unit is Plugged in to Main supply

    Check the mains are switched ON

    Else Contact nice Neotech

    2System Failure LED is ON with audio alarm

    System Failure Contact nice Neotech

    General System Failure

    System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

    1 No Display on the Control PanelCheck AC Supply Voltage

    Check if Power Cord is disconnected

    Check Fuse in the Rear side of Pillar

    Contact nice Neotech

    2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

    Set Temperature may be less than Actual Temperature

    Check Heater Connection

    Contact nice Neotech

    3 Equipment Working But Observation Lamp amp Heater output not working

    Check Link cord connected between Control Module toHeater Module

    Check Fuse in the Rear side of Heater Module

    Check Transformer Voltage - 12V AC

    Replace Lamp

    4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

    Check Mute ON

    Contact nice Neotech

    5 Patient Temperature Display shows ldquoS-Ordquo

    with Continuous Alarm(S-O Sensor Open)

    Check if skin Probe is Disconnected from theEquipment

    If skin probe is defective replace new probe

    Contact nice Neotech

    7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

    Check Heater Connection

    Contact nice Neotech

    8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

    Check Environment Temperature above 39deg C

    Check Infant Fever

    9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

    Temperature

    Check Environment Temperature is 1degC above SetTemperature

    Contact nice Neotech

    10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

    Temperature

    Probe may be wet

    Contact nice Neotech

    13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

    Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

    Low Supply Voltage

    14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

    Percentage Heater Output is Set at High in ManualMode

    15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

    Check for Strong Air conditioning draft

    16Temperature Display does not display InfantsTemperature accurately

    Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

    17Baby skin Temperature not correlate with display Temperature

    Check sensitive portion of the Probe is fixed with babyskin properly

    Check Probe is fixed to the lower abdomen of baby

    18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

    19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

    Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

    System Fault Infant Radiant Warmer(Contrsquodhellip)

    Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

    Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

    components Do not autoclave or gas sterilize the skin temperature probe Do not

    immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

    Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

    should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

    Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

    and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

    equipment performance

    Certifications and Approvals

    Electrical Safety IEC 60601-1 IEC 60601-1-4

    Particular product IEC 60601-2-21

    EMC Safety IEC 60601-1-2

    Contact Us

    nice Neotech Medical Systems Private Limited

    No 85 Krishna Industrial Estate

    Mettukuppam Vannagaram

    Chennai ndash 600095 Tamil Nadu

    INDIA

    Tel +91-44-2476259424764608

    Fax +91-44-24766920

    Web wwwniceneotechcom

    E-mail infoniceneotechcom marketingniceneotechcom

    Overseas Enquiry overseasniceneotechcom

    Toll Free +91-1800-425-2594

    • nice NeOumltech Medical Systems Pvt Ltd
    • Presentation Flow
    • Company INFO
    • Timeline-nice Neotech
    • Why nice Neotech
    • Why Infant Radiant Warmers are Used
    • MODES OF HEAT LOSS
    • Slide 8
    • Effects of Far IR in Infant body
    • Types of heaters
    • Slide 11
    • Benefits of Heaterhellip
    • Slide 13
    • Slide 14
    • Slide 15
    • How to operate
    • Control Panel Operation
    • Operating Modes
    • Slide 19
    • Slide 20
    • Alarms
    • Heater Module
    • Bed Platform Operation
    • Side Panel Operation
    • X-ray Procedure
    • Comparison of nice 2010 Infant Radiant Warmers
    • How nice Infant Radiant Warmers fare over competitors
    • Slide 28
    • Slide 29
    • Cleaning Procedure
    • Preventive Maintenance
    • Troubleshooting
    • System Fault Infant Radiant Warmer
    • Slide 34
    • Slide 35
    • Certifications and Approvals
    • Contact Us

      Company INFO

      Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

      for growth and delighted customers

      Timeline-nice Neotech

      2005bullFirst New Factory Unit 1 Opened

      bullIncrease in production Line

      2001bullRegistered as Pvt ltd

      bullProduction started for different models

      2013bullSecond New Factory Unit 2 Opened

      1999bull Neotech Logo

      Designed

      bull Designed first incubator amp open

      care prototype

      2003bull Export to

      Vietnam

      2009-2011

      bull CE certification for products

      2015bull 125+skilled

      employee strong firm

      bull Multiple range of NICU products

      1997bullStarted Neotech Medical Systems with 2 employees

      bullFirst Infant Radiant Warmer

      Why nice Neotech

      User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

      Why Infant Radiant Warmers are Used

      Assist Thermoregulation

      Body warming by providing Far IR radiant heat

      Ensure baby is stabilized after birth

      INFANT RADIANT WARMERS

      MODES OF HEAT LOSS

      Conduction Heat Loss Direct heat transfer from skin to object

      Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

      Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

      (i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

      Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

      Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

      Electromagnetic Spectrum

      Infrared Spectrum

      Effects of Far IR in Infant body

      Far Infrared Rays are invisible waves of energy

      They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

      Types of heaters Quartz

      A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

      Far IR Infrared

      Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

      Ceramic Heaters

      Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

      Tubular Heater

      A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

      Types of heaters(Contrsquodhelliphellip)

      Benefits of Heaterhellip

      QUARTZ

      bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

      CERAM

      IC

      bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

      TUBULAR

      bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

      Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

      A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

      Designed to perfection for your baby

      nice 2010 series Infant Radiant Warmers

      nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

      nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

      nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

      bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

      Features

      How to operate

      Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

      Control Panel Operation

      1 Skin Temperature Display 15

      Manual ModeIndicator

      2 Sensor Failure Indicator 16

      Skin Mode Indicator

      3 Heater Failure Indicator 17

      Pre Warm Mode Indicator

      4 System Failure Indicator 18

      Increase Key

      5 PowerFailure Indicator 19

      Mute Key

      6 Skin Temperature gt 39degC Indicator

      20

      Heater OutputIndicator

      7 Lamp Indicator 21

      Set TemperatureManualAPGAR Timer

      8 Lock Key 22

      APGAR Timer Indicator

      9 Lock Indicator 23

      Mute ON Indicator

      10

      Lamp Key 24

      Manual Timer Indicator

      11

      APGAR Kay 25

      Skin Temperature Low Indicator

      12

      Decrease Key 26

      Skin Temperature High Indicator

      13

      Mode Key 27

      Alarm Indicator

      14

      Safe Mode Indicator

      Operating Modes

      Manual Mode Operation

      1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

      1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

      1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

      1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

      Skin Mode Operation

      Note The skin temperature probe must be properly attached before choosing Skin mode operation

      1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

      1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

      APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

      Operating Modes(contrsquodhellip)

      Operating Modes(contrsquodhellip)

      Pre-Warm Mode

      bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

      Safe Mode

      bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

      Note All the above mentioned modes can only be accessible when the keypad lock is off condition

      AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

      Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

      Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

      Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

      Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

      Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

      Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

      Heater Module

      Parabolic design

      Even Distribution of heat over bed surface

      Extensive safety

      measures(6 Layer safety for thermal

      cut- off)

      Far IR heating ensures

      stimulation of cells for growth

      The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

      Bed Platform Operation

      bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

      bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

      Side Panel Operation

      bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

      X-ray Procedure

      bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

      bullThe heater housing rotates to the left or right for X-ray procedures

      bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

      bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

      Comparison of nice 2010 Infant Radiant Warmers

      nice 2010 BC nice 2010 B nice 2010

      Skin Mode Y Y Y

      Manual Mode Y Y Y

      Safe Mode Y Y Y

      Pre-warm Mode Y Y Y

      X Ray Facility Y Y N

      Bed Y Y N

      Pivoting Warmer ModuleY Y Y

      Apgar Timer Y Y Y

      Height Adjustment N N Y

      Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

      How nice Infant Radiant Warmers fare over competitors

      nice 2010 series warmers

      User-friendly easy to access all parameters

      Long life quartz heater has a long life compared to other heaters

      6 Layer safety for the protection of infant from burning

      6 Layer Safety Features

      Alarms if infant temperature greater or less than Set Temperature

      Thermal cut off if infant temperature exceeds 39 degC

      Heater output reduces to 30 after 15 mins in Manual mode

      Heater output reduces to 30 if electronics senses no variable feedback

      Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

      Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

      TROUBLESHOOTING

      AND MAINTENANCE

      Cleaning Procedure

      1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

      2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

      3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

      Cleaning Solutions for Mattress

      Generic Formulations Maximum concentration level

      Hydrogen Peroxide 6

      Quaternary Compound 05 aqueous solution

      Quaternary Ammonium 02

      Lodophors 50mglitre

      Preventive Maintenance

      1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

      Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

      1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

      TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

      Sl No Problem Cause Remedy

      1 Power failure LED is ON with audio alarm

      Power FailureCheck the unit is Plugged in to Main supply

      Check the mains are switched ON

      Else Contact nice Neotech

      2System Failure LED is ON with audio alarm

      System Failure Contact nice Neotech

      General System Failure

      System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

      1 No Display on the Control PanelCheck AC Supply Voltage

      Check if Power Cord is disconnected

      Check Fuse in the Rear side of Pillar

      Contact nice Neotech

      2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

      Set Temperature may be less than Actual Temperature

      Check Heater Connection

      Contact nice Neotech

      3 Equipment Working But Observation Lamp amp Heater output not working

      Check Link cord connected between Control Module toHeater Module

      Check Fuse in the Rear side of Heater Module

      Check Transformer Voltage - 12V AC

      Replace Lamp

      4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

      Check Mute ON

      Contact nice Neotech

      5 Patient Temperature Display shows ldquoS-Ordquo

      with Continuous Alarm(S-O Sensor Open)

      Check if skin Probe is Disconnected from theEquipment

      If skin probe is defective replace new probe

      Contact nice Neotech

      7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

      Check Heater Connection

      Contact nice Neotech

      8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

      Check Environment Temperature above 39deg C

      Check Infant Fever

      9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

      Temperature

      Check Environment Temperature is 1degC above SetTemperature

      Contact nice Neotech

      10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

      Temperature

      Probe may be wet

      Contact nice Neotech

      13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

      Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

      Low Supply Voltage

      14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

      Percentage Heater Output is Set at High in ManualMode

      15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

      Check for Strong Air conditioning draft

      16Temperature Display does not display InfantsTemperature accurately

      Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

      17Baby skin Temperature not correlate with display Temperature

      Check sensitive portion of the Probe is fixed with babyskin properly

      Check Probe is fixed to the lower abdomen of baby

      18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

      19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

      Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

      System Fault Infant Radiant Warmer(Contrsquodhellip)

      Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

      Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

      components Do not autoclave or gas sterilize the skin temperature probe Do not

      immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

      Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

      should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

      Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

      and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

      equipment performance

      Certifications and Approvals

      Electrical Safety IEC 60601-1 IEC 60601-1-4

      Particular product IEC 60601-2-21

      EMC Safety IEC 60601-1-2

      Contact Us

      nice Neotech Medical Systems Private Limited

      No 85 Krishna Industrial Estate

      Mettukuppam Vannagaram

      Chennai ndash 600095 Tamil Nadu

      INDIA

      Tel +91-44-2476259424764608

      Fax +91-44-24766920

      Web wwwniceneotechcom

      E-mail infoniceneotechcom marketingniceneotechcom

      Overseas Enquiry overseasniceneotechcom

      Toll Free +91-1800-425-2594

      • nice NeOumltech Medical Systems Pvt Ltd
      • Presentation Flow
      • Company INFO
      • Timeline-nice Neotech
      • Why nice Neotech
      • Why Infant Radiant Warmers are Used
      • MODES OF HEAT LOSS
      • Slide 8
      • Effects of Far IR in Infant body
      • Types of heaters
      • Slide 11
      • Benefits of Heaterhellip
      • Slide 13
      • Slide 14
      • Slide 15
      • How to operate
      • Control Panel Operation
      • Operating Modes
      • Slide 19
      • Slide 20
      • Alarms
      • Heater Module
      • Bed Platform Operation
      • Side Panel Operation
      • X-ray Procedure
      • Comparison of nice 2010 Infant Radiant Warmers
      • How nice Infant Radiant Warmers fare over competitors
      • Slide 28
      • Slide 29
      • Cleaning Procedure
      • Preventive Maintenance
      • Troubleshooting
      • System Fault Infant Radiant Warmer
      • Slide 34
      • Slide 35
      • Certifications and Approvals
      • Contact Us

        Timeline-nice Neotech

        2005bullFirst New Factory Unit 1 Opened

        bullIncrease in production Line

        2001bullRegistered as Pvt ltd

        bullProduction started for different models

        2013bullSecond New Factory Unit 2 Opened

        1999bull Neotech Logo

        Designed

        bull Designed first incubator amp open

        care prototype

        2003bull Export to

        Vietnam

        2009-2011

        bull CE certification for products

        2015bull 125+skilled

        employee strong firm

        bull Multiple range of NICU products

        1997bullStarted Neotech Medical Systems with 2 employees

        bullFirst Infant Radiant Warmer

        Why nice Neotech

        User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

        Why Infant Radiant Warmers are Used

        Assist Thermoregulation

        Body warming by providing Far IR radiant heat

        Ensure baby is stabilized after birth

        INFANT RADIANT WARMERS

        MODES OF HEAT LOSS

        Conduction Heat Loss Direct heat transfer from skin to object

        Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

        Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

        (i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

        Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

        Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

        Electromagnetic Spectrum

        Infrared Spectrum

        Effects of Far IR in Infant body

        Far Infrared Rays are invisible waves of energy

        They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

        Types of heaters Quartz

        A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

        Far IR Infrared

        Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

        Ceramic Heaters

        Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

        Tubular Heater

        A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

        Types of heaters(Contrsquodhelliphellip)

        Benefits of Heaterhellip

        QUARTZ

        bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

        CERAM

        IC

        bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

        TUBULAR

        bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

        Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

        A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

        Designed to perfection for your baby

        nice 2010 series Infant Radiant Warmers

        nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

        nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

        nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

        bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

        Features

        How to operate

        Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

        Control Panel Operation

        1 Skin Temperature Display 15

        Manual ModeIndicator

        2 Sensor Failure Indicator 16

        Skin Mode Indicator

        3 Heater Failure Indicator 17

        Pre Warm Mode Indicator

        4 System Failure Indicator 18

        Increase Key

        5 PowerFailure Indicator 19

        Mute Key

        6 Skin Temperature gt 39degC Indicator

        20

        Heater OutputIndicator

        7 Lamp Indicator 21

        Set TemperatureManualAPGAR Timer

        8 Lock Key 22

        APGAR Timer Indicator

        9 Lock Indicator 23

        Mute ON Indicator

        10

        Lamp Key 24

        Manual Timer Indicator

        11

        APGAR Kay 25

        Skin Temperature Low Indicator

        12

        Decrease Key 26

        Skin Temperature High Indicator

        13

        Mode Key 27

        Alarm Indicator

        14

        Safe Mode Indicator

        Operating Modes

        Manual Mode Operation

        1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

        1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

        1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

        1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

        Skin Mode Operation

        Note The skin temperature probe must be properly attached before choosing Skin mode operation

        1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

        1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

        APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

        Operating Modes(contrsquodhellip)

        Operating Modes(contrsquodhellip)

        Pre-Warm Mode

        bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

        Safe Mode

        bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

        Note All the above mentioned modes can only be accessible when the keypad lock is off condition

        AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

        Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

        Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

        Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

        Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

        Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

        Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

        Heater Module

        Parabolic design

        Even Distribution of heat over bed surface

        Extensive safety

        measures(6 Layer safety for thermal

        cut- off)

        Far IR heating ensures

        stimulation of cells for growth

        The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

        Bed Platform Operation

        bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

        bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

        Side Panel Operation

        bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

        X-ray Procedure

        bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

        bullThe heater housing rotates to the left or right for X-ray procedures

        bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

        bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

        Comparison of nice 2010 Infant Radiant Warmers

        nice 2010 BC nice 2010 B nice 2010

        Skin Mode Y Y Y

        Manual Mode Y Y Y

        Safe Mode Y Y Y

        Pre-warm Mode Y Y Y

        X Ray Facility Y Y N

        Bed Y Y N

        Pivoting Warmer ModuleY Y Y

        Apgar Timer Y Y Y

        Height Adjustment N N Y

        Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

        How nice Infant Radiant Warmers fare over competitors

        nice 2010 series warmers

        User-friendly easy to access all parameters

        Long life quartz heater has a long life compared to other heaters

        6 Layer safety for the protection of infant from burning

        6 Layer Safety Features

        Alarms if infant temperature greater or less than Set Temperature

        Thermal cut off if infant temperature exceeds 39 degC

        Heater output reduces to 30 after 15 mins in Manual mode

        Heater output reduces to 30 if electronics senses no variable feedback

        Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

        Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

        TROUBLESHOOTING

        AND MAINTENANCE

        Cleaning Procedure

        1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

        2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

        3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

        Cleaning Solutions for Mattress

        Generic Formulations Maximum concentration level

        Hydrogen Peroxide 6

        Quaternary Compound 05 aqueous solution

        Quaternary Ammonium 02

        Lodophors 50mglitre

        Preventive Maintenance

        1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

        Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

        1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

        TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

        Sl No Problem Cause Remedy

        1 Power failure LED is ON with audio alarm

        Power FailureCheck the unit is Plugged in to Main supply

        Check the mains are switched ON

        Else Contact nice Neotech

        2System Failure LED is ON with audio alarm

        System Failure Contact nice Neotech

        General System Failure

        System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

        1 No Display on the Control PanelCheck AC Supply Voltage

        Check if Power Cord is disconnected

        Check Fuse in the Rear side of Pillar

        Contact nice Neotech

        2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

        Set Temperature may be less than Actual Temperature

        Check Heater Connection

        Contact nice Neotech

        3 Equipment Working But Observation Lamp amp Heater output not working

        Check Link cord connected between Control Module toHeater Module

        Check Fuse in the Rear side of Heater Module

        Check Transformer Voltage - 12V AC

        Replace Lamp

        4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

        Check Mute ON

        Contact nice Neotech

        5 Patient Temperature Display shows ldquoS-Ordquo

        with Continuous Alarm(S-O Sensor Open)

        Check if skin Probe is Disconnected from theEquipment

        If skin probe is defective replace new probe

        Contact nice Neotech

        7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

        Check Heater Connection

        Contact nice Neotech

        8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

        Check Environment Temperature above 39deg C

        Check Infant Fever

        9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

        Temperature

        Check Environment Temperature is 1degC above SetTemperature

        Contact nice Neotech

        10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

        Temperature

        Probe may be wet

        Contact nice Neotech

        13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

        Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

        Low Supply Voltage

        14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

        Percentage Heater Output is Set at High in ManualMode

        15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

        Check for Strong Air conditioning draft

        16Temperature Display does not display InfantsTemperature accurately

        Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

        17Baby skin Temperature not correlate with display Temperature

        Check sensitive portion of the Probe is fixed with babyskin properly

        Check Probe is fixed to the lower abdomen of baby

        18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

        19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

        Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

        System Fault Infant Radiant Warmer(Contrsquodhellip)

        Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

        Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

        components Do not autoclave or gas sterilize the skin temperature probe Do not

        immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

        Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

        should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

        Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

        and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

        equipment performance

        Certifications and Approvals

        Electrical Safety IEC 60601-1 IEC 60601-1-4

        Particular product IEC 60601-2-21

        EMC Safety IEC 60601-1-2

        Contact Us

        nice Neotech Medical Systems Private Limited

        No 85 Krishna Industrial Estate

        Mettukuppam Vannagaram

        Chennai ndash 600095 Tamil Nadu

        INDIA

        Tel +91-44-2476259424764608

        Fax +91-44-24766920

        Web wwwniceneotechcom

        E-mail infoniceneotechcom marketingniceneotechcom

        Overseas Enquiry overseasniceneotechcom

        Toll Free +91-1800-425-2594

        • nice NeOumltech Medical Systems Pvt Ltd
        • Presentation Flow
        • Company INFO
        • Timeline-nice Neotech
        • Why nice Neotech
        • Why Infant Radiant Warmers are Used
        • MODES OF HEAT LOSS
        • Slide 8
        • Effects of Far IR in Infant body
        • Types of heaters
        • Slide 11
        • Benefits of Heaterhellip
        • Slide 13
        • Slide 14
        • Slide 15
        • How to operate
        • Control Panel Operation
        • Operating Modes
        • Slide 19
        • Slide 20
        • Alarms
        • Heater Module
        • Bed Platform Operation
        • Side Panel Operation
        • X-ray Procedure
        • Comparison of nice 2010 Infant Radiant Warmers
        • How nice Infant Radiant Warmers fare over competitors
        • Slide 28
        • Slide 29
        • Cleaning Procedure
        • Preventive Maintenance
        • Troubleshooting
        • System Fault Infant Radiant Warmer
        • Slide 34
        • Slide 35
        • Certifications and Approvals
        • Contact Us

          Why nice Neotech

          User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

          Why Infant Radiant Warmers are Used

          Assist Thermoregulation

          Body warming by providing Far IR radiant heat

          Ensure baby is stabilized after birth

          INFANT RADIANT WARMERS

          MODES OF HEAT LOSS

          Conduction Heat Loss Direct heat transfer from skin to object

          Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

          Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

          (i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

          Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

          Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

          Electromagnetic Spectrum

          Infrared Spectrum

          Effects of Far IR in Infant body

          Far Infrared Rays are invisible waves of energy

          They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

          Types of heaters Quartz

          A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

          Far IR Infrared

          Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

          Ceramic Heaters

          Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

          Tubular Heater

          A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

          Types of heaters(Contrsquodhelliphellip)

          Benefits of Heaterhellip

          QUARTZ

          bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

          CERAM

          IC

          bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

          TUBULAR

          bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

          Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

          A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

          Designed to perfection for your baby

          nice 2010 series Infant Radiant Warmers

          nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

          nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

          nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

          bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

          Features

          How to operate

          Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

          Control Panel Operation

          1 Skin Temperature Display 15

          Manual ModeIndicator

          2 Sensor Failure Indicator 16

          Skin Mode Indicator

          3 Heater Failure Indicator 17

          Pre Warm Mode Indicator

          4 System Failure Indicator 18

          Increase Key

          5 PowerFailure Indicator 19

          Mute Key

          6 Skin Temperature gt 39degC Indicator

          20

          Heater OutputIndicator

          7 Lamp Indicator 21

          Set TemperatureManualAPGAR Timer

          8 Lock Key 22

          APGAR Timer Indicator

          9 Lock Indicator 23

          Mute ON Indicator

          10

          Lamp Key 24

          Manual Timer Indicator

          11

          APGAR Kay 25

          Skin Temperature Low Indicator

          12

          Decrease Key 26

          Skin Temperature High Indicator

          13

          Mode Key 27

          Alarm Indicator

          14

          Safe Mode Indicator

          Operating Modes

          Manual Mode Operation

          1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

          1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

          1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

          1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

          Skin Mode Operation

          Note The skin temperature probe must be properly attached before choosing Skin mode operation

          1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

          1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

          APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

          Operating Modes(contrsquodhellip)

          Operating Modes(contrsquodhellip)

          Pre-Warm Mode

          bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

          Safe Mode

          bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

          Note All the above mentioned modes can only be accessible when the keypad lock is off condition

          AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

          Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

          Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

          Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

          Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

          Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

          Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

          Heater Module

          Parabolic design

          Even Distribution of heat over bed surface

          Extensive safety

          measures(6 Layer safety for thermal

          cut- off)

          Far IR heating ensures

          stimulation of cells for growth

          The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

          Bed Platform Operation

          bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

          bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

          Side Panel Operation

          bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

          X-ray Procedure

          bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

          bullThe heater housing rotates to the left or right for X-ray procedures

          bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

          bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

          Comparison of nice 2010 Infant Radiant Warmers

          nice 2010 BC nice 2010 B nice 2010

          Skin Mode Y Y Y

          Manual Mode Y Y Y

          Safe Mode Y Y Y

          Pre-warm Mode Y Y Y

          X Ray Facility Y Y N

          Bed Y Y N

          Pivoting Warmer ModuleY Y Y

          Apgar Timer Y Y Y

          Height Adjustment N N Y

          Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

          How nice Infant Radiant Warmers fare over competitors

          nice 2010 series warmers

          User-friendly easy to access all parameters

          Long life quartz heater has a long life compared to other heaters

          6 Layer safety for the protection of infant from burning

          6 Layer Safety Features

          Alarms if infant temperature greater or less than Set Temperature

          Thermal cut off if infant temperature exceeds 39 degC

          Heater output reduces to 30 after 15 mins in Manual mode

          Heater output reduces to 30 if electronics senses no variable feedback

          Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

          Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

          TROUBLESHOOTING

          AND MAINTENANCE

          Cleaning Procedure

          1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

          2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

          3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

          Cleaning Solutions for Mattress

          Generic Formulations Maximum concentration level

          Hydrogen Peroxide 6

          Quaternary Compound 05 aqueous solution

          Quaternary Ammonium 02

          Lodophors 50mglitre

          Preventive Maintenance

          1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

          Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

          1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

          TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

          Sl No Problem Cause Remedy

          1 Power failure LED is ON with audio alarm

          Power FailureCheck the unit is Plugged in to Main supply

          Check the mains are switched ON

          Else Contact nice Neotech

          2System Failure LED is ON with audio alarm

          System Failure Contact nice Neotech

          General System Failure

          System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

          1 No Display on the Control PanelCheck AC Supply Voltage

          Check if Power Cord is disconnected

          Check Fuse in the Rear side of Pillar

          Contact nice Neotech

          2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

          Set Temperature may be less than Actual Temperature

          Check Heater Connection

          Contact nice Neotech

          3 Equipment Working But Observation Lamp amp Heater output not working

          Check Link cord connected between Control Module toHeater Module

          Check Fuse in the Rear side of Heater Module

          Check Transformer Voltage - 12V AC

          Replace Lamp

          4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

          Check Mute ON

          Contact nice Neotech

          5 Patient Temperature Display shows ldquoS-Ordquo

          with Continuous Alarm(S-O Sensor Open)

          Check if skin Probe is Disconnected from theEquipment

          If skin probe is defective replace new probe

          Contact nice Neotech

          7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

          Check Heater Connection

          Contact nice Neotech

          8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

          Check Environment Temperature above 39deg C

          Check Infant Fever

          9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

          Temperature

          Check Environment Temperature is 1degC above SetTemperature

          Contact nice Neotech

          10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

          Temperature

          Probe may be wet

          Contact nice Neotech

          13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

          Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

          Low Supply Voltage

          14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

          Percentage Heater Output is Set at High in ManualMode

          15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

          Check for Strong Air conditioning draft

          16Temperature Display does not display InfantsTemperature accurately

          Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

          17Baby skin Temperature not correlate with display Temperature

          Check sensitive portion of the Probe is fixed with babyskin properly

          Check Probe is fixed to the lower abdomen of baby

          18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

          19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

          Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

          System Fault Infant Radiant Warmer(Contrsquodhellip)

          Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

          Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

          components Do not autoclave or gas sterilize the skin temperature probe Do not

          immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

          Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

          should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

          Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

          and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

          equipment performance

          Certifications and Approvals

          Electrical Safety IEC 60601-1 IEC 60601-1-4

          Particular product IEC 60601-2-21

          EMC Safety IEC 60601-1-2

          Contact Us

          nice Neotech Medical Systems Private Limited

          No 85 Krishna Industrial Estate

          Mettukuppam Vannagaram

          Chennai ndash 600095 Tamil Nadu

          INDIA

          Tel +91-44-2476259424764608

          Fax +91-44-24766920

          Web wwwniceneotechcom

          E-mail infoniceneotechcom marketingniceneotechcom

          Overseas Enquiry overseasniceneotechcom

          Toll Free +91-1800-425-2594

          • nice NeOumltech Medical Systems Pvt Ltd
          • Presentation Flow
          • Company INFO
          • Timeline-nice Neotech
          • Why nice Neotech
          • Why Infant Radiant Warmers are Used
          • MODES OF HEAT LOSS
          • Slide 8
          • Effects of Far IR in Infant body
          • Types of heaters
          • Slide 11
          • Benefits of Heaterhellip
          • Slide 13
          • Slide 14
          • Slide 15
          • How to operate
          • Control Panel Operation
          • Operating Modes
          • Slide 19
          • Slide 20
          • Alarms
          • Heater Module
          • Bed Platform Operation
          • Side Panel Operation
          • X-ray Procedure
          • Comparison of nice 2010 Infant Radiant Warmers
          • How nice Infant Radiant Warmers fare over competitors
          • Slide 28
          • Slide 29
          • Cleaning Procedure
          • Preventive Maintenance
          • Troubleshooting
          • System Fault Infant Radiant Warmer
          • Slide 34
          • Slide 35
          • Certifications and Approvals
          • Contact Us

            Why Infant Radiant Warmers are Used

            Assist Thermoregulation

            Body warming by providing Far IR radiant heat

            Ensure baby is stabilized after birth

            INFANT RADIANT WARMERS

            MODES OF HEAT LOSS

            Conduction Heat Loss Direct heat transfer from skin to object

            Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

            Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

            (i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

            Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

            Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

            Electromagnetic Spectrum

            Infrared Spectrum

            Effects of Far IR in Infant body

            Far Infrared Rays are invisible waves of energy

            They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

            Types of heaters Quartz

            A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

            Far IR Infrared

            Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

            Ceramic Heaters

            Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

            Tubular Heater

            A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

            Types of heaters(Contrsquodhelliphellip)

            Benefits of Heaterhellip

            QUARTZ

            bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

            CERAM

            IC

            bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

            TUBULAR

            bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

            Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

            A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

            Designed to perfection for your baby

            nice 2010 series Infant Radiant Warmers

            nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

            nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

            nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

            bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

            Features

            How to operate

            Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

            Control Panel Operation

            1 Skin Temperature Display 15

            Manual ModeIndicator

            2 Sensor Failure Indicator 16

            Skin Mode Indicator

            3 Heater Failure Indicator 17

            Pre Warm Mode Indicator

            4 System Failure Indicator 18

            Increase Key

            5 PowerFailure Indicator 19

            Mute Key

            6 Skin Temperature gt 39degC Indicator

            20

            Heater OutputIndicator

            7 Lamp Indicator 21

            Set TemperatureManualAPGAR Timer

            8 Lock Key 22

            APGAR Timer Indicator

            9 Lock Indicator 23

            Mute ON Indicator

            10

            Lamp Key 24

            Manual Timer Indicator

            11

            APGAR Kay 25

            Skin Temperature Low Indicator

            12

            Decrease Key 26

            Skin Temperature High Indicator

            13

            Mode Key 27

            Alarm Indicator

            14

            Safe Mode Indicator

            Operating Modes

            Manual Mode Operation

            1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

            1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

            1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

            1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

            Skin Mode Operation

            Note The skin temperature probe must be properly attached before choosing Skin mode operation

            1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

            1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

            APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

            Operating Modes(contrsquodhellip)

            Operating Modes(contrsquodhellip)

            Pre-Warm Mode

            bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

            Safe Mode

            bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

            Note All the above mentioned modes can only be accessible when the keypad lock is off condition

            AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

            Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

            Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

            Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

            Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

            Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

            Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

            Heater Module

            Parabolic design

            Even Distribution of heat over bed surface

            Extensive safety

            measures(6 Layer safety for thermal

            cut- off)

            Far IR heating ensures

            stimulation of cells for growth

            The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

            Bed Platform Operation

            bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

            bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

            Side Panel Operation

            bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

            X-ray Procedure

            bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

            bullThe heater housing rotates to the left or right for X-ray procedures

            bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

            bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

            Comparison of nice 2010 Infant Radiant Warmers

            nice 2010 BC nice 2010 B nice 2010

            Skin Mode Y Y Y

            Manual Mode Y Y Y

            Safe Mode Y Y Y

            Pre-warm Mode Y Y Y

            X Ray Facility Y Y N

            Bed Y Y N

            Pivoting Warmer ModuleY Y Y

            Apgar Timer Y Y Y

            Height Adjustment N N Y

            Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

            How nice Infant Radiant Warmers fare over competitors

            nice 2010 series warmers

            User-friendly easy to access all parameters

            Long life quartz heater has a long life compared to other heaters

            6 Layer safety for the protection of infant from burning

            6 Layer Safety Features

            Alarms if infant temperature greater or less than Set Temperature

            Thermal cut off if infant temperature exceeds 39 degC

            Heater output reduces to 30 after 15 mins in Manual mode

            Heater output reduces to 30 if electronics senses no variable feedback

            Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

            Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

            TROUBLESHOOTING

            AND MAINTENANCE

            Cleaning Procedure

            1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

            2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

            3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

            Cleaning Solutions for Mattress

            Generic Formulations Maximum concentration level

            Hydrogen Peroxide 6

            Quaternary Compound 05 aqueous solution

            Quaternary Ammonium 02

            Lodophors 50mglitre

            Preventive Maintenance

            1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

            Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

            1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

            TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

            Sl No Problem Cause Remedy

            1 Power failure LED is ON with audio alarm

            Power FailureCheck the unit is Plugged in to Main supply

            Check the mains are switched ON

            Else Contact nice Neotech

            2System Failure LED is ON with audio alarm

            System Failure Contact nice Neotech

            General System Failure

            System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

            1 No Display on the Control PanelCheck AC Supply Voltage

            Check if Power Cord is disconnected

            Check Fuse in the Rear side of Pillar

            Contact nice Neotech

            2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

            Set Temperature may be less than Actual Temperature

            Check Heater Connection

            Contact nice Neotech

            3 Equipment Working But Observation Lamp amp Heater output not working

            Check Link cord connected between Control Module toHeater Module

            Check Fuse in the Rear side of Heater Module

            Check Transformer Voltage - 12V AC

            Replace Lamp

            4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

            Check Mute ON

            Contact nice Neotech

            5 Patient Temperature Display shows ldquoS-Ordquo

            with Continuous Alarm(S-O Sensor Open)

            Check if skin Probe is Disconnected from theEquipment

            If skin probe is defective replace new probe

            Contact nice Neotech

            7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

            Check Heater Connection

            Contact nice Neotech

            8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

            Check Environment Temperature above 39deg C

            Check Infant Fever

            9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

            Temperature

            Check Environment Temperature is 1degC above SetTemperature

            Contact nice Neotech

            10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

            Temperature

            Probe may be wet

            Contact nice Neotech

            13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

            Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

            Low Supply Voltage

            14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

            Percentage Heater Output is Set at High in ManualMode

            15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

            Check for Strong Air conditioning draft

            16Temperature Display does not display InfantsTemperature accurately

            Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

            17Baby skin Temperature not correlate with display Temperature

            Check sensitive portion of the Probe is fixed with babyskin properly

            Check Probe is fixed to the lower abdomen of baby

            18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

            19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

            Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

            System Fault Infant Radiant Warmer(Contrsquodhellip)

            Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

            Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

            components Do not autoclave or gas sterilize the skin temperature probe Do not

            immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

            Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

            should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

            Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

            and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

            equipment performance

            Certifications and Approvals

            Electrical Safety IEC 60601-1 IEC 60601-1-4

            Particular product IEC 60601-2-21

            EMC Safety IEC 60601-1-2

            Contact Us

            nice Neotech Medical Systems Private Limited

            No 85 Krishna Industrial Estate

            Mettukuppam Vannagaram

            Chennai ndash 600095 Tamil Nadu

            INDIA

            Tel +91-44-2476259424764608

            Fax +91-44-24766920

            Web wwwniceneotechcom

            E-mail infoniceneotechcom marketingniceneotechcom

            Overseas Enquiry overseasniceneotechcom

            Toll Free +91-1800-425-2594

            • nice NeOumltech Medical Systems Pvt Ltd
            • Presentation Flow
            • Company INFO
            • Timeline-nice Neotech
            • Why nice Neotech
            • Why Infant Radiant Warmers are Used
            • MODES OF HEAT LOSS
            • Slide 8
            • Effects of Far IR in Infant body
            • Types of heaters
            • Slide 11
            • Benefits of Heaterhellip
            • Slide 13
            • Slide 14
            • Slide 15
            • How to operate
            • Control Panel Operation
            • Operating Modes
            • Slide 19
            • Slide 20
            • Alarms
            • Heater Module
            • Bed Platform Operation
            • Side Panel Operation
            • X-ray Procedure
            • Comparison of nice 2010 Infant Radiant Warmers
            • How nice Infant Radiant Warmers fare over competitors
            • Slide 28
            • Slide 29
            • Cleaning Procedure
            • Preventive Maintenance
            • Troubleshooting
            • System Fault Infant Radiant Warmer
            • Slide 34
            • Slide 35
            • Certifications and Approvals
            • Contact Us

              MODES OF HEAT LOSS

              Conduction Heat Loss Direct heat transfer from skin to object

              Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

              Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

              (i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

              Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

              Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

              Electromagnetic Spectrum

              Infrared Spectrum

              Effects of Far IR in Infant body

              Far Infrared Rays are invisible waves of energy

              They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

              Types of heaters Quartz

              A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

              Far IR Infrared

              Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

              Ceramic Heaters

              Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

              Tubular Heater

              A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

              Types of heaters(Contrsquodhelliphellip)

              Benefits of Heaterhellip

              QUARTZ

              bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

              CERAM

              IC

              bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

              TUBULAR

              bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

              Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

              A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

              Designed to perfection for your baby

              nice 2010 series Infant Radiant Warmers

              nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

              nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

              nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

              bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

              Features

              How to operate

              Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

              Control Panel Operation

              1 Skin Temperature Display 15

              Manual ModeIndicator

              2 Sensor Failure Indicator 16

              Skin Mode Indicator

              3 Heater Failure Indicator 17

              Pre Warm Mode Indicator

              4 System Failure Indicator 18

              Increase Key

              5 PowerFailure Indicator 19

              Mute Key

              6 Skin Temperature gt 39degC Indicator

              20

              Heater OutputIndicator

              7 Lamp Indicator 21

              Set TemperatureManualAPGAR Timer

              8 Lock Key 22

              APGAR Timer Indicator

              9 Lock Indicator 23

              Mute ON Indicator

              10

              Lamp Key 24

              Manual Timer Indicator

              11

              APGAR Kay 25

              Skin Temperature Low Indicator

              12

              Decrease Key 26

              Skin Temperature High Indicator

              13

              Mode Key 27

              Alarm Indicator

              14

              Safe Mode Indicator

              Operating Modes

              Manual Mode Operation

              1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

              1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

              1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

              1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

              Skin Mode Operation

              Note The skin temperature probe must be properly attached before choosing Skin mode operation

              1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

              1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

              APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

              Operating Modes(contrsquodhellip)

              Operating Modes(contrsquodhellip)

              Pre-Warm Mode

              bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

              Safe Mode

              bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

              Note All the above mentioned modes can only be accessible when the keypad lock is off condition

              AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

              Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

              Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

              Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

              Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

              Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

              Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

              Heater Module

              Parabolic design

              Even Distribution of heat over bed surface

              Extensive safety

              measures(6 Layer safety for thermal

              cut- off)

              Far IR heating ensures

              stimulation of cells for growth

              The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

              Bed Platform Operation

              bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

              bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

              Side Panel Operation

              bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

              X-ray Procedure

              bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

              bullThe heater housing rotates to the left or right for X-ray procedures

              bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

              bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

              Comparison of nice 2010 Infant Radiant Warmers

              nice 2010 BC nice 2010 B nice 2010

              Skin Mode Y Y Y

              Manual Mode Y Y Y

              Safe Mode Y Y Y

              Pre-warm Mode Y Y Y

              X Ray Facility Y Y N

              Bed Y Y N

              Pivoting Warmer ModuleY Y Y

              Apgar Timer Y Y Y

              Height Adjustment N N Y

              Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

              How nice Infant Radiant Warmers fare over competitors

              nice 2010 series warmers

              User-friendly easy to access all parameters

              Long life quartz heater has a long life compared to other heaters

              6 Layer safety for the protection of infant from burning

              6 Layer Safety Features

              Alarms if infant temperature greater or less than Set Temperature

              Thermal cut off if infant temperature exceeds 39 degC

              Heater output reduces to 30 after 15 mins in Manual mode

              Heater output reduces to 30 if electronics senses no variable feedback

              Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

              Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

              TROUBLESHOOTING

              AND MAINTENANCE

              Cleaning Procedure

              1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

              2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

              3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

              Cleaning Solutions for Mattress

              Generic Formulations Maximum concentration level

              Hydrogen Peroxide 6

              Quaternary Compound 05 aqueous solution

              Quaternary Ammonium 02

              Lodophors 50mglitre

              Preventive Maintenance

              1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

              Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

              1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

              TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

              Sl No Problem Cause Remedy

              1 Power failure LED is ON with audio alarm

              Power FailureCheck the unit is Plugged in to Main supply

              Check the mains are switched ON

              Else Contact nice Neotech

              2System Failure LED is ON with audio alarm

              System Failure Contact nice Neotech

              General System Failure

              System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

              1 No Display on the Control PanelCheck AC Supply Voltage

              Check if Power Cord is disconnected

              Check Fuse in the Rear side of Pillar

              Contact nice Neotech

              2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

              Set Temperature may be less than Actual Temperature

              Check Heater Connection

              Contact nice Neotech

              3 Equipment Working But Observation Lamp amp Heater output not working

              Check Link cord connected between Control Module toHeater Module

              Check Fuse in the Rear side of Heater Module

              Check Transformer Voltage - 12V AC

              Replace Lamp

              4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

              Check Mute ON

              Contact nice Neotech

              5 Patient Temperature Display shows ldquoS-Ordquo

              with Continuous Alarm(S-O Sensor Open)

              Check if skin Probe is Disconnected from theEquipment

              If skin probe is defective replace new probe

              Contact nice Neotech

              7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

              Check Heater Connection

              Contact nice Neotech

              8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

              Check Environment Temperature above 39deg C

              Check Infant Fever

              9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

              Temperature

              Check Environment Temperature is 1degC above SetTemperature

              Contact nice Neotech

              10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

              Temperature

              Probe may be wet

              Contact nice Neotech

              13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

              Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

              Low Supply Voltage

              14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

              Percentage Heater Output is Set at High in ManualMode

              15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

              Check for Strong Air conditioning draft

              16Temperature Display does not display InfantsTemperature accurately

              Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

              17Baby skin Temperature not correlate with display Temperature

              Check sensitive portion of the Probe is fixed with babyskin properly

              Check Probe is fixed to the lower abdomen of baby

              18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

              19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

              Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

              System Fault Infant Radiant Warmer(Contrsquodhellip)

              Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

              Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

              components Do not autoclave or gas sterilize the skin temperature probe Do not

              immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

              Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

              should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

              Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

              and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

              equipment performance

              Certifications and Approvals

              Electrical Safety IEC 60601-1 IEC 60601-1-4

              Particular product IEC 60601-2-21

              EMC Safety IEC 60601-1-2

              Contact Us

              nice Neotech Medical Systems Private Limited

              No 85 Krishna Industrial Estate

              Mettukuppam Vannagaram

              Chennai ndash 600095 Tamil Nadu

              INDIA

              Tel +91-44-2476259424764608

              Fax +91-44-24766920

              Web wwwniceneotechcom

              E-mail infoniceneotechcom marketingniceneotechcom

              Overseas Enquiry overseasniceneotechcom

              Toll Free +91-1800-425-2594

              • nice NeOumltech Medical Systems Pvt Ltd
              • Presentation Flow
              • Company INFO
              • Timeline-nice Neotech
              • Why nice Neotech
              • Why Infant Radiant Warmers are Used
              • MODES OF HEAT LOSS
              • Slide 8
              • Effects of Far IR in Infant body
              • Types of heaters
              • Slide 11
              • Benefits of Heaterhellip
              • Slide 13
              • Slide 14
              • Slide 15
              • How to operate
              • Control Panel Operation
              • Operating Modes
              • Slide 19
              • Slide 20
              • Alarms
              • Heater Module
              • Bed Platform Operation
              • Side Panel Operation
              • X-ray Procedure
              • Comparison of nice 2010 Infant Radiant Warmers
              • How nice Infant Radiant Warmers fare over competitors
              • Slide 28
              • Slide 29
              • Cleaning Procedure
              • Preventive Maintenance
              • Troubleshooting
              • System Fault Infant Radiant Warmer
              • Slide 34
              • Slide 35
              • Certifications and Approvals
              • Contact Us

                Electromagnetic Spectrum

                Infrared Spectrum

                Effects of Far IR in Infant body

                Far Infrared Rays are invisible waves of energy

                They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

                Types of heaters Quartz

                A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

                Far IR Infrared

                Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

                Ceramic Heaters

                Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

                Tubular Heater

                A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

                Types of heaters(Contrsquodhelliphellip)

                Benefits of Heaterhellip

                QUARTZ

                bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

                CERAM

                IC

                bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

                TUBULAR

                bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

                Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

                A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

                Designed to perfection for your baby

                nice 2010 series Infant Radiant Warmers

                nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                Features

                How to operate

                Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                Control Panel Operation

                1 Skin Temperature Display 15

                Manual ModeIndicator

                2 Sensor Failure Indicator 16

                Skin Mode Indicator

                3 Heater Failure Indicator 17

                Pre Warm Mode Indicator

                4 System Failure Indicator 18

                Increase Key

                5 PowerFailure Indicator 19

                Mute Key

                6 Skin Temperature gt 39degC Indicator

                20

                Heater OutputIndicator

                7 Lamp Indicator 21

                Set TemperatureManualAPGAR Timer

                8 Lock Key 22

                APGAR Timer Indicator

                9 Lock Indicator 23

                Mute ON Indicator

                10

                Lamp Key 24

                Manual Timer Indicator

                11

                APGAR Kay 25

                Skin Temperature Low Indicator

                12

                Decrease Key 26

                Skin Temperature High Indicator

                13

                Mode Key 27

                Alarm Indicator

                14

                Safe Mode Indicator

                Operating Modes

                Manual Mode Operation

                1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                Skin Mode Operation

                Note The skin temperature probe must be properly attached before choosing Skin mode operation

                1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                Operating Modes(contrsquodhellip)

                Operating Modes(contrsquodhellip)

                Pre-Warm Mode

                bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                Safe Mode

                bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                Heater Module

                Parabolic design

                Even Distribution of heat over bed surface

                Extensive safety

                measures(6 Layer safety for thermal

                cut- off)

                Far IR heating ensures

                stimulation of cells for growth

                The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                Bed Platform Operation

                bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                Side Panel Operation

                bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                X-ray Procedure

                bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                bullThe heater housing rotates to the left or right for X-ray procedures

                bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                Comparison of nice 2010 Infant Radiant Warmers

                nice 2010 BC nice 2010 B nice 2010

                Skin Mode Y Y Y

                Manual Mode Y Y Y

                Safe Mode Y Y Y

                Pre-warm Mode Y Y Y

                X Ray Facility Y Y N

                Bed Y Y N

                Pivoting Warmer ModuleY Y Y

                Apgar Timer Y Y Y

                Height Adjustment N N Y

                Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                How nice Infant Radiant Warmers fare over competitors

                nice 2010 series warmers

                User-friendly easy to access all parameters

                Long life quartz heater has a long life compared to other heaters

                6 Layer safety for the protection of infant from burning

                6 Layer Safety Features

                Alarms if infant temperature greater or less than Set Temperature

                Thermal cut off if infant temperature exceeds 39 degC

                Heater output reduces to 30 after 15 mins in Manual mode

                Heater output reduces to 30 if electronics senses no variable feedback

                Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                TROUBLESHOOTING

                AND MAINTENANCE

                Cleaning Procedure

                1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                Cleaning Solutions for Mattress

                Generic Formulations Maximum concentration level

                Hydrogen Peroxide 6

                Quaternary Compound 05 aqueous solution

                Quaternary Ammonium 02

                Lodophors 50mglitre

                Preventive Maintenance

                1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                Sl No Problem Cause Remedy

                1 Power failure LED is ON with audio alarm

                Power FailureCheck the unit is Plugged in to Main supply

                Check the mains are switched ON

                Else Contact nice Neotech

                2System Failure LED is ON with audio alarm

                System Failure Contact nice Neotech

                General System Failure

                System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                1 No Display on the Control PanelCheck AC Supply Voltage

                Check if Power Cord is disconnected

                Check Fuse in the Rear side of Pillar

                Contact nice Neotech

                2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                Set Temperature may be less than Actual Temperature

                Check Heater Connection

                Contact nice Neotech

                3 Equipment Working But Observation Lamp amp Heater output not working

                Check Link cord connected between Control Module toHeater Module

                Check Fuse in the Rear side of Heater Module

                Check Transformer Voltage - 12V AC

                Replace Lamp

                4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                Check Mute ON

                Contact nice Neotech

                5 Patient Temperature Display shows ldquoS-Ordquo

                with Continuous Alarm(S-O Sensor Open)

                Check if skin Probe is Disconnected from theEquipment

                If skin probe is defective replace new probe

                Contact nice Neotech

                7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                Check Heater Connection

                Contact nice Neotech

                8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                Check Environment Temperature above 39deg C

                Check Infant Fever

                9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                Temperature

                Check Environment Temperature is 1degC above SetTemperature

                Contact nice Neotech

                10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                Temperature

                Probe may be wet

                Contact nice Neotech

                13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                Low Supply Voltage

                14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                Percentage Heater Output is Set at High in ManualMode

                15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                Check for Strong Air conditioning draft

                16Temperature Display does not display InfantsTemperature accurately

                Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                17Baby skin Temperature not correlate with display Temperature

                Check sensitive portion of the Probe is fixed with babyskin properly

                Check Probe is fixed to the lower abdomen of baby

                18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                System Fault Infant Radiant Warmer(Contrsquodhellip)

                Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                components Do not autoclave or gas sterilize the skin temperature probe Do not

                immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                equipment performance

                Certifications and Approvals

                Electrical Safety IEC 60601-1 IEC 60601-1-4

                Particular product IEC 60601-2-21

                EMC Safety IEC 60601-1-2

                Contact Us

                nice Neotech Medical Systems Private Limited

                No 85 Krishna Industrial Estate

                Mettukuppam Vannagaram

                Chennai ndash 600095 Tamil Nadu

                INDIA

                Tel +91-44-2476259424764608

                Fax +91-44-24766920

                Web wwwniceneotechcom

                E-mail infoniceneotechcom marketingniceneotechcom

                Overseas Enquiry overseasniceneotechcom

                Toll Free +91-1800-425-2594

                • nice NeOumltech Medical Systems Pvt Ltd
                • Presentation Flow
                • Company INFO
                • Timeline-nice Neotech
                • Why nice Neotech
                • Why Infant Radiant Warmers are Used
                • MODES OF HEAT LOSS
                • Slide 8
                • Effects of Far IR in Infant body
                • Types of heaters
                • Slide 11
                • Benefits of Heaterhellip
                • Slide 13
                • Slide 14
                • Slide 15
                • How to operate
                • Control Panel Operation
                • Operating Modes
                • Slide 19
                • Slide 20
                • Alarms
                • Heater Module
                • Bed Platform Operation
                • Side Panel Operation
                • X-ray Procedure
                • Comparison of nice 2010 Infant Radiant Warmers
                • How nice Infant Radiant Warmers fare over competitors
                • Slide 28
                • Slide 29
                • Cleaning Procedure
                • Preventive Maintenance
                • Troubleshooting
                • System Fault Infant Radiant Warmer
                • Slide 34
                • Slide 35
                • Certifications and Approvals
                • Contact Us

                  Effects of Far IR in Infant body

                  Far Infrared Rays are invisible waves of energy

                  They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

                  Types of heaters Quartz

                  A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

                  Far IR Infrared

                  Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

                  Ceramic Heaters

                  Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

                  Tubular Heater

                  A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

                  Types of heaters(Contrsquodhelliphellip)

                  Benefits of Heaterhellip

                  QUARTZ

                  bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

                  CERAM

                  IC

                  bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

                  TUBULAR

                  bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

                  Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

                  A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

                  Designed to perfection for your baby

                  nice 2010 series Infant Radiant Warmers

                  nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                  nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                  nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                  bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                  Features

                  How to operate

                  Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                  Control Panel Operation

                  1 Skin Temperature Display 15

                  Manual ModeIndicator

                  2 Sensor Failure Indicator 16

                  Skin Mode Indicator

                  3 Heater Failure Indicator 17

                  Pre Warm Mode Indicator

                  4 System Failure Indicator 18

                  Increase Key

                  5 PowerFailure Indicator 19

                  Mute Key

                  6 Skin Temperature gt 39degC Indicator

                  20

                  Heater OutputIndicator

                  7 Lamp Indicator 21

                  Set TemperatureManualAPGAR Timer

                  8 Lock Key 22

                  APGAR Timer Indicator

                  9 Lock Indicator 23

                  Mute ON Indicator

                  10

                  Lamp Key 24

                  Manual Timer Indicator

                  11

                  APGAR Kay 25

                  Skin Temperature Low Indicator

                  12

                  Decrease Key 26

                  Skin Temperature High Indicator

                  13

                  Mode Key 27

                  Alarm Indicator

                  14

                  Safe Mode Indicator

                  Operating Modes

                  Manual Mode Operation

                  1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                  1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                  1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                  1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                  Skin Mode Operation

                  Note The skin temperature probe must be properly attached before choosing Skin mode operation

                  1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                  1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                  APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                  Operating Modes(contrsquodhellip)

                  Operating Modes(contrsquodhellip)

                  Pre-Warm Mode

                  bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                  Safe Mode

                  bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                  Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                  AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                  Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                  Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                  Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                  Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                  Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                  Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                  Heater Module

                  Parabolic design

                  Even Distribution of heat over bed surface

                  Extensive safety

                  measures(6 Layer safety for thermal

                  cut- off)

                  Far IR heating ensures

                  stimulation of cells for growth

                  The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                  Bed Platform Operation

                  bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                  bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                  Side Panel Operation

                  bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                  X-ray Procedure

                  bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                  bullThe heater housing rotates to the left or right for X-ray procedures

                  bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                  bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                  Comparison of nice 2010 Infant Radiant Warmers

                  nice 2010 BC nice 2010 B nice 2010

                  Skin Mode Y Y Y

                  Manual Mode Y Y Y

                  Safe Mode Y Y Y

                  Pre-warm Mode Y Y Y

                  X Ray Facility Y Y N

                  Bed Y Y N

                  Pivoting Warmer ModuleY Y Y

                  Apgar Timer Y Y Y

                  Height Adjustment N N Y

                  Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                  How nice Infant Radiant Warmers fare over competitors

                  nice 2010 series warmers

                  User-friendly easy to access all parameters

                  Long life quartz heater has a long life compared to other heaters

                  6 Layer safety for the protection of infant from burning

                  6 Layer Safety Features

                  Alarms if infant temperature greater or less than Set Temperature

                  Thermal cut off if infant temperature exceeds 39 degC

                  Heater output reduces to 30 after 15 mins in Manual mode

                  Heater output reduces to 30 if electronics senses no variable feedback

                  Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                  Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                  TROUBLESHOOTING

                  AND MAINTENANCE

                  Cleaning Procedure

                  1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                  2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                  3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                  Cleaning Solutions for Mattress

                  Generic Formulations Maximum concentration level

                  Hydrogen Peroxide 6

                  Quaternary Compound 05 aqueous solution

                  Quaternary Ammonium 02

                  Lodophors 50mglitre

                  Preventive Maintenance

                  1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                  Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                  1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                  TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                  Sl No Problem Cause Remedy

                  1 Power failure LED is ON with audio alarm

                  Power FailureCheck the unit is Plugged in to Main supply

                  Check the mains are switched ON

                  Else Contact nice Neotech

                  2System Failure LED is ON with audio alarm

                  System Failure Contact nice Neotech

                  General System Failure

                  System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                  1 No Display on the Control PanelCheck AC Supply Voltage

                  Check if Power Cord is disconnected

                  Check Fuse in the Rear side of Pillar

                  Contact nice Neotech

                  2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                  Set Temperature may be less than Actual Temperature

                  Check Heater Connection

                  Contact nice Neotech

                  3 Equipment Working But Observation Lamp amp Heater output not working

                  Check Link cord connected between Control Module toHeater Module

                  Check Fuse in the Rear side of Heater Module

                  Check Transformer Voltage - 12V AC

                  Replace Lamp

                  4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                  Check Mute ON

                  Contact nice Neotech

                  5 Patient Temperature Display shows ldquoS-Ordquo

                  with Continuous Alarm(S-O Sensor Open)

                  Check if skin Probe is Disconnected from theEquipment

                  If skin probe is defective replace new probe

                  Contact nice Neotech

                  7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                  Check Heater Connection

                  Contact nice Neotech

                  8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                  Check Environment Temperature above 39deg C

                  Check Infant Fever

                  9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                  Temperature

                  Check Environment Temperature is 1degC above SetTemperature

                  Contact nice Neotech

                  10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                  Temperature

                  Probe may be wet

                  Contact nice Neotech

                  13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                  Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                  Low Supply Voltage

                  14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                  Percentage Heater Output is Set at High in ManualMode

                  15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                  Check for Strong Air conditioning draft

                  16Temperature Display does not display InfantsTemperature accurately

                  Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                  17Baby skin Temperature not correlate with display Temperature

                  Check sensitive portion of the Probe is fixed with babyskin properly

                  Check Probe is fixed to the lower abdomen of baby

                  18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                  19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                  Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                  System Fault Infant Radiant Warmer(Contrsquodhellip)

                  Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                  Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                  components Do not autoclave or gas sterilize the skin temperature probe Do not

                  immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                  Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                  should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                  Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                  and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                  equipment performance

                  Certifications and Approvals

                  Electrical Safety IEC 60601-1 IEC 60601-1-4

                  Particular product IEC 60601-2-21

                  EMC Safety IEC 60601-1-2

                  Contact Us

                  nice Neotech Medical Systems Private Limited

                  No 85 Krishna Industrial Estate

                  Mettukuppam Vannagaram

                  Chennai ndash 600095 Tamil Nadu

                  INDIA

                  Tel +91-44-2476259424764608

                  Fax +91-44-24766920

                  Web wwwniceneotechcom

                  E-mail infoniceneotechcom marketingniceneotechcom

                  Overseas Enquiry overseasniceneotechcom

                  Toll Free +91-1800-425-2594

                  • nice NeOumltech Medical Systems Pvt Ltd
                  • Presentation Flow
                  • Company INFO
                  • Timeline-nice Neotech
                  • Why nice Neotech
                  • Why Infant Radiant Warmers are Used
                  • MODES OF HEAT LOSS
                  • Slide 8
                  • Effects of Far IR in Infant body
                  • Types of heaters
                  • Slide 11
                  • Benefits of Heaterhellip
                  • Slide 13
                  • Slide 14
                  • Slide 15
                  • How to operate
                  • Control Panel Operation
                  • Operating Modes
                  • Slide 19
                  • Slide 20
                  • Alarms
                  • Heater Module
                  • Bed Platform Operation
                  • Side Panel Operation
                  • X-ray Procedure
                  • Comparison of nice 2010 Infant Radiant Warmers
                  • How nice Infant Radiant Warmers fare over competitors
                  • Slide 28
                  • Slide 29
                  • Cleaning Procedure
                  • Preventive Maintenance
                  • Troubleshooting
                  • System Fault Infant Radiant Warmer
                  • Slide 34
                  • Slide 35
                  • Certifications and Approvals
                  • Contact Us

                    Types of heaters Quartz

                    A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

                    Far IR Infrared

                    Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

                    Ceramic Heaters

                    Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

                    Tubular Heater

                    A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

                    Types of heaters(Contrsquodhelliphellip)

                    Benefits of Heaterhellip

                    QUARTZ

                    bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

                    CERAM

                    IC

                    bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

                    TUBULAR

                    bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

                    Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

                    A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

                    Designed to perfection for your baby

                    nice 2010 series Infant Radiant Warmers

                    nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                    nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                    nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                    bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                    Features

                    How to operate

                    Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                    Control Panel Operation

                    1 Skin Temperature Display 15

                    Manual ModeIndicator

                    2 Sensor Failure Indicator 16

                    Skin Mode Indicator

                    3 Heater Failure Indicator 17

                    Pre Warm Mode Indicator

                    4 System Failure Indicator 18

                    Increase Key

                    5 PowerFailure Indicator 19

                    Mute Key

                    6 Skin Temperature gt 39degC Indicator

                    20

                    Heater OutputIndicator

                    7 Lamp Indicator 21

                    Set TemperatureManualAPGAR Timer

                    8 Lock Key 22

                    APGAR Timer Indicator

                    9 Lock Indicator 23

                    Mute ON Indicator

                    10

                    Lamp Key 24

                    Manual Timer Indicator

                    11

                    APGAR Kay 25

                    Skin Temperature Low Indicator

                    12

                    Decrease Key 26

                    Skin Temperature High Indicator

                    13

                    Mode Key 27

                    Alarm Indicator

                    14

                    Safe Mode Indicator

                    Operating Modes

                    Manual Mode Operation

                    1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                    1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                    1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                    1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                    Skin Mode Operation

                    Note The skin temperature probe must be properly attached before choosing Skin mode operation

                    1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                    1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                    APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                    Operating Modes(contrsquodhellip)

                    Operating Modes(contrsquodhellip)

                    Pre-Warm Mode

                    bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                    Safe Mode

                    bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                    Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                    AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                    Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                    Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                    Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                    Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                    Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                    Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                    Heater Module

                    Parabolic design

                    Even Distribution of heat over bed surface

                    Extensive safety

                    measures(6 Layer safety for thermal

                    cut- off)

                    Far IR heating ensures

                    stimulation of cells for growth

                    The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                    Bed Platform Operation

                    bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                    bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                    Side Panel Operation

                    bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                    X-ray Procedure

                    bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                    bullThe heater housing rotates to the left or right for X-ray procedures

                    bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                    bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                    Comparison of nice 2010 Infant Radiant Warmers

                    nice 2010 BC nice 2010 B nice 2010

                    Skin Mode Y Y Y

                    Manual Mode Y Y Y

                    Safe Mode Y Y Y

                    Pre-warm Mode Y Y Y

                    X Ray Facility Y Y N

                    Bed Y Y N

                    Pivoting Warmer ModuleY Y Y

                    Apgar Timer Y Y Y

                    Height Adjustment N N Y

                    Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                    How nice Infant Radiant Warmers fare over competitors

                    nice 2010 series warmers

                    User-friendly easy to access all parameters

                    Long life quartz heater has a long life compared to other heaters

                    6 Layer safety for the protection of infant from burning

                    6 Layer Safety Features

                    Alarms if infant temperature greater or less than Set Temperature

                    Thermal cut off if infant temperature exceeds 39 degC

                    Heater output reduces to 30 after 15 mins in Manual mode

                    Heater output reduces to 30 if electronics senses no variable feedback

                    Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                    Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                    TROUBLESHOOTING

                    AND MAINTENANCE

                    Cleaning Procedure

                    1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                    2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                    3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                    Cleaning Solutions for Mattress

                    Generic Formulations Maximum concentration level

                    Hydrogen Peroxide 6

                    Quaternary Compound 05 aqueous solution

                    Quaternary Ammonium 02

                    Lodophors 50mglitre

                    Preventive Maintenance

                    1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                    Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                    1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                    TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                    Sl No Problem Cause Remedy

                    1 Power failure LED is ON with audio alarm

                    Power FailureCheck the unit is Plugged in to Main supply

                    Check the mains are switched ON

                    Else Contact nice Neotech

                    2System Failure LED is ON with audio alarm

                    System Failure Contact nice Neotech

                    General System Failure

                    System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                    1 No Display on the Control PanelCheck AC Supply Voltage

                    Check if Power Cord is disconnected

                    Check Fuse in the Rear side of Pillar

                    Contact nice Neotech

                    2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                    Set Temperature may be less than Actual Temperature

                    Check Heater Connection

                    Contact nice Neotech

                    3 Equipment Working But Observation Lamp amp Heater output not working

                    Check Link cord connected between Control Module toHeater Module

                    Check Fuse in the Rear side of Heater Module

                    Check Transformer Voltage - 12V AC

                    Replace Lamp

                    4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                    Check Mute ON

                    Contact nice Neotech

                    5 Patient Temperature Display shows ldquoS-Ordquo

                    with Continuous Alarm(S-O Sensor Open)

                    Check if skin Probe is Disconnected from theEquipment

                    If skin probe is defective replace new probe

                    Contact nice Neotech

                    7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                    Check Heater Connection

                    Contact nice Neotech

                    8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                    Check Environment Temperature above 39deg C

                    Check Infant Fever

                    9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                    Temperature

                    Check Environment Temperature is 1degC above SetTemperature

                    Contact nice Neotech

                    10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                    Temperature

                    Probe may be wet

                    Contact nice Neotech

                    13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                    Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                    Low Supply Voltage

                    14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                    Percentage Heater Output is Set at High in ManualMode

                    15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                    Check for Strong Air conditioning draft

                    16Temperature Display does not display InfantsTemperature accurately

                    Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                    17Baby skin Temperature not correlate with display Temperature

                    Check sensitive portion of the Probe is fixed with babyskin properly

                    Check Probe is fixed to the lower abdomen of baby

                    18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                    19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                    Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                    System Fault Infant Radiant Warmer(Contrsquodhellip)

                    Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                    Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                    components Do not autoclave or gas sterilize the skin temperature probe Do not

                    immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                    Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                    should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                    Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                    and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                    equipment performance

                    Certifications and Approvals

                    Electrical Safety IEC 60601-1 IEC 60601-1-4

                    Particular product IEC 60601-2-21

                    EMC Safety IEC 60601-1-2

                    Contact Us

                    nice Neotech Medical Systems Private Limited

                    No 85 Krishna Industrial Estate

                    Mettukuppam Vannagaram

                    Chennai ndash 600095 Tamil Nadu

                    INDIA

                    Tel +91-44-2476259424764608

                    Fax +91-44-24766920

                    Web wwwniceneotechcom

                    E-mail infoniceneotechcom marketingniceneotechcom

                    Overseas Enquiry overseasniceneotechcom

                    Toll Free +91-1800-425-2594

                    • nice NeOumltech Medical Systems Pvt Ltd
                    • Presentation Flow
                    • Company INFO
                    • Timeline-nice Neotech
                    • Why nice Neotech
                    • Why Infant Radiant Warmers are Used
                    • MODES OF HEAT LOSS
                    • Slide 8
                    • Effects of Far IR in Infant body
                    • Types of heaters
                    • Slide 11
                    • Benefits of Heaterhellip
                    • Slide 13
                    • Slide 14
                    • Slide 15
                    • How to operate
                    • Control Panel Operation
                    • Operating Modes
                    • Slide 19
                    • Slide 20
                    • Alarms
                    • Heater Module
                    • Bed Platform Operation
                    • Side Panel Operation
                    • X-ray Procedure
                    • Comparison of nice 2010 Infant Radiant Warmers
                    • How nice Infant Radiant Warmers fare over competitors
                    • Slide 28
                    • Slide 29
                    • Cleaning Procedure
                    • Preventive Maintenance
                    • Troubleshooting
                    • System Fault Infant Radiant Warmer
                    • Slide 34
                    • Slide 35
                    • Certifications and Approvals
                    • Contact Us

                      Ceramic Heaters

                      Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

                      Tubular Heater

                      A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

                      Types of heaters(Contrsquodhelliphellip)

                      Benefits of Heaterhellip

                      QUARTZ

                      bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

                      CERAM

                      IC

                      bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

                      TUBULAR

                      bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

                      Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

                      A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

                      Designed to perfection for your baby

                      nice 2010 series Infant Radiant Warmers

                      nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                      nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                      nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                      bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                      Features

                      How to operate

                      Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                      Control Panel Operation

                      1 Skin Temperature Display 15

                      Manual ModeIndicator

                      2 Sensor Failure Indicator 16

                      Skin Mode Indicator

                      3 Heater Failure Indicator 17

                      Pre Warm Mode Indicator

                      4 System Failure Indicator 18

                      Increase Key

                      5 PowerFailure Indicator 19

                      Mute Key

                      6 Skin Temperature gt 39degC Indicator

                      20

                      Heater OutputIndicator

                      7 Lamp Indicator 21

                      Set TemperatureManualAPGAR Timer

                      8 Lock Key 22

                      APGAR Timer Indicator

                      9 Lock Indicator 23

                      Mute ON Indicator

                      10

                      Lamp Key 24

                      Manual Timer Indicator

                      11

                      APGAR Kay 25

                      Skin Temperature Low Indicator

                      12

                      Decrease Key 26

                      Skin Temperature High Indicator

                      13

                      Mode Key 27

                      Alarm Indicator

                      14

                      Safe Mode Indicator

                      Operating Modes

                      Manual Mode Operation

                      1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                      1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                      1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                      1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                      Skin Mode Operation

                      Note The skin temperature probe must be properly attached before choosing Skin mode operation

                      1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                      1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                      APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                      Operating Modes(contrsquodhellip)

                      Operating Modes(contrsquodhellip)

                      Pre-Warm Mode

                      bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                      Safe Mode

                      bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                      Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                      AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                      Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                      Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                      Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                      Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                      Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                      Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                      Heater Module

                      Parabolic design

                      Even Distribution of heat over bed surface

                      Extensive safety

                      measures(6 Layer safety for thermal

                      cut- off)

                      Far IR heating ensures

                      stimulation of cells for growth

                      The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                      Bed Platform Operation

                      bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                      bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                      Side Panel Operation

                      bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                      X-ray Procedure

                      bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                      bullThe heater housing rotates to the left or right for X-ray procedures

                      bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                      bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                      Comparison of nice 2010 Infant Radiant Warmers

                      nice 2010 BC nice 2010 B nice 2010

                      Skin Mode Y Y Y

                      Manual Mode Y Y Y

                      Safe Mode Y Y Y

                      Pre-warm Mode Y Y Y

                      X Ray Facility Y Y N

                      Bed Y Y N

                      Pivoting Warmer ModuleY Y Y

                      Apgar Timer Y Y Y

                      Height Adjustment N N Y

                      Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                      How nice Infant Radiant Warmers fare over competitors

                      nice 2010 series warmers

                      User-friendly easy to access all parameters

                      Long life quartz heater has a long life compared to other heaters

                      6 Layer safety for the protection of infant from burning

                      6 Layer Safety Features

                      Alarms if infant temperature greater or less than Set Temperature

                      Thermal cut off if infant temperature exceeds 39 degC

                      Heater output reduces to 30 after 15 mins in Manual mode

                      Heater output reduces to 30 if electronics senses no variable feedback

                      Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                      Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                      TROUBLESHOOTING

                      AND MAINTENANCE

                      Cleaning Procedure

                      1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                      2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                      3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                      Cleaning Solutions for Mattress

                      Generic Formulations Maximum concentration level

                      Hydrogen Peroxide 6

                      Quaternary Compound 05 aqueous solution

                      Quaternary Ammonium 02

                      Lodophors 50mglitre

                      Preventive Maintenance

                      1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                      Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                      1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                      TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                      Sl No Problem Cause Remedy

                      1 Power failure LED is ON with audio alarm

                      Power FailureCheck the unit is Plugged in to Main supply

                      Check the mains are switched ON

                      Else Contact nice Neotech

                      2System Failure LED is ON with audio alarm

                      System Failure Contact nice Neotech

                      General System Failure

                      System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                      1 No Display on the Control PanelCheck AC Supply Voltage

                      Check if Power Cord is disconnected

                      Check Fuse in the Rear side of Pillar

                      Contact nice Neotech

                      2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                      Set Temperature may be less than Actual Temperature

                      Check Heater Connection

                      Contact nice Neotech

                      3 Equipment Working But Observation Lamp amp Heater output not working

                      Check Link cord connected between Control Module toHeater Module

                      Check Fuse in the Rear side of Heater Module

                      Check Transformer Voltage - 12V AC

                      Replace Lamp

                      4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                      Check Mute ON

                      Contact nice Neotech

                      5 Patient Temperature Display shows ldquoS-Ordquo

                      with Continuous Alarm(S-O Sensor Open)

                      Check if skin Probe is Disconnected from theEquipment

                      If skin probe is defective replace new probe

                      Contact nice Neotech

                      7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                      Check Heater Connection

                      Contact nice Neotech

                      8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                      Check Environment Temperature above 39deg C

                      Check Infant Fever

                      9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                      Temperature

                      Check Environment Temperature is 1degC above SetTemperature

                      Contact nice Neotech

                      10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                      Temperature

                      Probe may be wet

                      Contact nice Neotech

                      13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                      Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                      Low Supply Voltage

                      14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                      Percentage Heater Output is Set at High in ManualMode

                      15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                      Check for Strong Air conditioning draft

                      16Temperature Display does not display InfantsTemperature accurately

                      Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                      17Baby skin Temperature not correlate with display Temperature

                      Check sensitive portion of the Probe is fixed with babyskin properly

                      Check Probe is fixed to the lower abdomen of baby

                      18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                      19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                      Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                      System Fault Infant Radiant Warmer(Contrsquodhellip)

                      Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                      Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                      components Do not autoclave or gas sterilize the skin temperature probe Do not

                      immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                      Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                      should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                      Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                      and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                      equipment performance

                      Certifications and Approvals

                      Electrical Safety IEC 60601-1 IEC 60601-1-4

                      Particular product IEC 60601-2-21

                      EMC Safety IEC 60601-1-2

                      Contact Us

                      nice Neotech Medical Systems Private Limited

                      No 85 Krishna Industrial Estate

                      Mettukuppam Vannagaram

                      Chennai ndash 600095 Tamil Nadu

                      INDIA

                      Tel +91-44-2476259424764608

                      Fax +91-44-24766920

                      Web wwwniceneotechcom

                      E-mail infoniceneotechcom marketingniceneotechcom

                      Overseas Enquiry overseasniceneotechcom

                      Toll Free +91-1800-425-2594

                      • nice NeOumltech Medical Systems Pvt Ltd
                      • Presentation Flow
                      • Company INFO
                      • Timeline-nice Neotech
                      • Why nice Neotech
                      • Why Infant Radiant Warmers are Used
                      • MODES OF HEAT LOSS
                      • Slide 8
                      • Effects of Far IR in Infant body
                      • Types of heaters
                      • Slide 11
                      • Benefits of Heaterhellip
                      • Slide 13
                      • Slide 14
                      • Slide 15
                      • How to operate
                      • Control Panel Operation
                      • Operating Modes
                      • Slide 19
                      • Slide 20
                      • Alarms
                      • Heater Module
                      • Bed Platform Operation
                      • Side Panel Operation
                      • X-ray Procedure
                      • Comparison of nice 2010 Infant Radiant Warmers
                      • How nice Infant Radiant Warmers fare over competitors
                      • Slide 28
                      • Slide 29
                      • Cleaning Procedure
                      • Preventive Maintenance
                      • Troubleshooting
                      • System Fault Infant Radiant Warmer
                      • Slide 34
                      • Slide 35
                      • Certifications and Approvals
                      • Contact Us

                        Benefits of Heaterhellip

                        QUARTZ

                        bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

                        CERAM

                        IC

                        bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

                        TUBULAR

                        bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

                        Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

                        A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

                        Designed to perfection for your baby

                        nice 2010 series Infant Radiant Warmers

                        nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                        nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                        nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                        bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                        Features

                        How to operate

                        Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                        Control Panel Operation

                        1 Skin Temperature Display 15

                        Manual ModeIndicator

                        2 Sensor Failure Indicator 16

                        Skin Mode Indicator

                        3 Heater Failure Indicator 17

                        Pre Warm Mode Indicator

                        4 System Failure Indicator 18

                        Increase Key

                        5 PowerFailure Indicator 19

                        Mute Key

                        6 Skin Temperature gt 39degC Indicator

                        20

                        Heater OutputIndicator

                        7 Lamp Indicator 21

                        Set TemperatureManualAPGAR Timer

                        8 Lock Key 22

                        APGAR Timer Indicator

                        9 Lock Indicator 23

                        Mute ON Indicator

                        10

                        Lamp Key 24

                        Manual Timer Indicator

                        11

                        APGAR Kay 25

                        Skin Temperature Low Indicator

                        12

                        Decrease Key 26

                        Skin Temperature High Indicator

                        13

                        Mode Key 27

                        Alarm Indicator

                        14

                        Safe Mode Indicator

                        Operating Modes

                        Manual Mode Operation

                        1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                        1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                        1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                        1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                        Skin Mode Operation

                        Note The skin temperature probe must be properly attached before choosing Skin mode operation

                        1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                        1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                        APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                        Operating Modes(contrsquodhellip)

                        Operating Modes(contrsquodhellip)

                        Pre-Warm Mode

                        bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                        Safe Mode

                        bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                        Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                        AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                        Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                        Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                        Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                        Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                        Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                        Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                        Heater Module

                        Parabolic design

                        Even Distribution of heat over bed surface

                        Extensive safety

                        measures(6 Layer safety for thermal

                        cut- off)

                        Far IR heating ensures

                        stimulation of cells for growth

                        The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                        Bed Platform Operation

                        bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                        bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                        Side Panel Operation

                        bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                        X-ray Procedure

                        bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                        bullThe heater housing rotates to the left or right for X-ray procedures

                        bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                        bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                        Comparison of nice 2010 Infant Radiant Warmers

                        nice 2010 BC nice 2010 B nice 2010

                        Skin Mode Y Y Y

                        Manual Mode Y Y Y

                        Safe Mode Y Y Y

                        Pre-warm Mode Y Y Y

                        X Ray Facility Y Y N

                        Bed Y Y N

                        Pivoting Warmer ModuleY Y Y

                        Apgar Timer Y Y Y

                        Height Adjustment N N Y

                        Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                        How nice Infant Radiant Warmers fare over competitors

                        nice 2010 series warmers

                        User-friendly easy to access all parameters

                        Long life quartz heater has a long life compared to other heaters

                        6 Layer safety for the protection of infant from burning

                        6 Layer Safety Features

                        Alarms if infant temperature greater or less than Set Temperature

                        Thermal cut off if infant temperature exceeds 39 degC

                        Heater output reduces to 30 after 15 mins in Manual mode

                        Heater output reduces to 30 if electronics senses no variable feedback

                        Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                        Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                        TROUBLESHOOTING

                        AND MAINTENANCE

                        Cleaning Procedure

                        1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                        2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                        3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                        Cleaning Solutions for Mattress

                        Generic Formulations Maximum concentration level

                        Hydrogen Peroxide 6

                        Quaternary Compound 05 aqueous solution

                        Quaternary Ammonium 02

                        Lodophors 50mglitre

                        Preventive Maintenance

                        1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                        Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                        1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                        TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                        Sl No Problem Cause Remedy

                        1 Power failure LED is ON with audio alarm

                        Power FailureCheck the unit is Plugged in to Main supply

                        Check the mains are switched ON

                        Else Contact nice Neotech

                        2System Failure LED is ON with audio alarm

                        System Failure Contact nice Neotech

                        General System Failure

                        System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                        1 No Display on the Control PanelCheck AC Supply Voltage

                        Check if Power Cord is disconnected

                        Check Fuse in the Rear side of Pillar

                        Contact nice Neotech

                        2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                        Set Temperature may be less than Actual Temperature

                        Check Heater Connection

                        Contact nice Neotech

                        3 Equipment Working But Observation Lamp amp Heater output not working

                        Check Link cord connected between Control Module toHeater Module

                        Check Fuse in the Rear side of Heater Module

                        Check Transformer Voltage - 12V AC

                        Replace Lamp

                        4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                        Check Mute ON

                        Contact nice Neotech

                        5 Patient Temperature Display shows ldquoS-Ordquo

                        with Continuous Alarm(S-O Sensor Open)

                        Check if skin Probe is Disconnected from theEquipment

                        If skin probe is defective replace new probe

                        Contact nice Neotech

                        7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                        Check Heater Connection

                        Contact nice Neotech

                        8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                        Check Environment Temperature above 39deg C

                        Check Infant Fever

                        9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                        Temperature

                        Check Environment Temperature is 1degC above SetTemperature

                        Contact nice Neotech

                        10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                        Temperature

                        Probe may be wet

                        Contact nice Neotech

                        13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                        Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                        Low Supply Voltage

                        14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                        Percentage Heater Output is Set at High in ManualMode

                        15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                        Check for Strong Air conditioning draft

                        16Temperature Display does not display InfantsTemperature accurately

                        Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                        17Baby skin Temperature not correlate with display Temperature

                        Check sensitive portion of the Probe is fixed with babyskin properly

                        Check Probe is fixed to the lower abdomen of baby

                        18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                        19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                        Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                        System Fault Infant Radiant Warmer(Contrsquodhellip)

                        Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                        Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                        components Do not autoclave or gas sterilize the skin temperature probe Do not

                        immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                        Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                        should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                        Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                        and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                        equipment performance

                        Certifications and Approvals

                        Electrical Safety IEC 60601-1 IEC 60601-1-4

                        Particular product IEC 60601-2-21

                        EMC Safety IEC 60601-1-2

                        Contact Us

                        nice Neotech Medical Systems Private Limited

                        No 85 Krishna Industrial Estate

                        Mettukuppam Vannagaram

                        Chennai ndash 600095 Tamil Nadu

                        INDIA

                        Tel +91-44-2476259424764608

                        Fax +91-44-24766920

                        Web wwwniceneotechcom

                        E-mail infoniceneotechcom marketingniceneotechcom

                        Overseas Enquiry overseasniceneotechcom

                        Toll Free +91-1800-425-2594

                        • nice NeOumltech Medical Systems Pvt Ltd
                        • Presentation Flow
                        • Company INFO
                        • Timeline-nice Neotech
                        • Why nice Neotech
                        • Why Infant Radiant Warmers are Used
                        • MODES OF HEAT LOSS
                        • Slide 8
                        • Effects of Far IR in Infant body
                        • Types of heaters
                        • Slide 11
                        • Benefits of Heaterhellip
                        • Slide 13
                        • Slide 14
                        • Slide 15
                        • How to operate
                        • Control Panel Operation
                        • Operating Modes
                        • Slide 19
                        • Slide 20
                        • Alarms
                        • Heater Module
                        • Bed Platform Operation
                        • Side Panel Operation
                        • X-ray Procedure
                        • Comparison of nice 2010 Infant Radiant Warmers
                        • How nice Infant Radiant Warmers fare over competitors
                        • Slide 28
                        • Slide 29
                        • Cleaning Procedure
                        • Preventive Maintenance
                        • Troubleshooting
                        • System Fault Infant Radiant Warmer
                        • Slide 34
                        • Slide 35
                        • Certifications and Approvals
                        • Contact Us

                          A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

                          Designed to perfection for your baby

                          nice 2010 series Infant Radiant Warmers

                          nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                          nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                          nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                          bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                          Features

                          How to operate

                          Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                          Control Panel Operation

                          1 Skin Temperature Display 15

                          Manual ModeIndicator

                          2 Sensor Failure Indicator 16

                          Skin Mode Indicator

                          3 Heater Failure Indicator 17

                          Pre Warm Mode Indicator

                          4 System Failure Indicator 18

                          Increase Key

                          5 PowerFailure Indicator 19

                          Mute Key

                          6 Skin Temperature gt 39degC Indicator

                          20

                          Heater OutputIndicator

                          7 Lamp Indicator 21

                          Set TemperatureManualAPGAR Timer

                          8 Lock Key 22

                          APGAR Timer Indicator

                          9 Lock Indicator 23

                          Mute ON Indicator

                          10

                          Lamp Key 24

                          Manual Timer Indicator

                          11

                          APGAR Kay 25

                          Skin Temperature Low Indicator

                          12

                          Decrease Key 26

                          Skin Temperature High Indicator

                          13

                          Mode Key 27

                          Alarm Indicator

                          14

                          Safe Mode Indicator

                          Operating Modes

                          Manual Mode Operation

                          1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                          1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                          1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                          1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                          Skin Mode Operation

                          Note The skin temperature probe must be properly attached before choosing Skin mode operation

                          1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                          1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                          APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                          Operating Modes(contrsquodhellip)

                          Operating Modes(contrsquodhellip)

                          Pre-Warm Mode

                          bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                          Safe Mode

                          bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                          Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                          AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                          Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                          Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                          Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                          Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                          Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                          Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                          Heater Module

                          Parabolic design

                          Even Distribution of heat over bed surface

                          Extensive safety

                          measures(6 Layer safety for thermal

                          cut- off)

                          Far IR heating ensures

                          stimulation of cells for growth

                          The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                          Bed Platform Operation

                          bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                          bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                          Side Panel Operation

                          bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                          X-ray Procedure

                          bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                          bullThe heater housing rotates to the left or right for X-ray procedures

                          bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                          bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                          Comparison of nice 2010 Infant Radiant Warmers

                          nice 2010 BC nice 2010 B nice 2010

                          Skin Mode Y Y Y

                          Manual Mode Y Y Y

                          Safe Mode Y Y Y

                          Pre-warm Mode Y Y Y

                          X Ray Facility Y Y N

                          Bed Y Y N

                          Pivoting Warmer ModuleY Y Y

                          Apgar Timer Y Y Y

                          Height Adjustment N N Y

                          Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                          How nice Infant Radiant Warmers fare over competitors

                          nice 2010 series warmers

                          User-friendly easy to access all parameters

                          Long life quartz heater has a long life compared to other heaters

                          6 Layer safety for the protection of infant from burning

                          6 Layer Safety Features

                          Alarms if infant temperature greater or less than Set Temperature

                          Thermal cut off if infant temperature exceeds 39 degC

                          Heater output reduces to 30 after 15 mins in Manual mode

                          Heater output reduces to 30 if electronics senses no variable feedback

                          Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                          Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                          TROUBLESHOOTING

                          AND MAINTENANCE

                          Cleaning Procedure

                          1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                          2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                          3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                          Cleaning Solutions for Mattress

                          Generic Formulations Maximum concentration level

                          Hydrogen Peroxide 6

                          Quaternary Compound 05 aqueous solution

                          Quaternary Ammonium 02

                          Lodophors 50mglitre

                          Preventive Maintenance

                          1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                          Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                          1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                          TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                          Sl No Problem Cause Remedy

                          1 Power failure LED is ON with audio alarm

                          Power FailureCheck the unit is Plugged in to Main supply

                          Check the mains are switched ON

                          Else Contact nice Neotech

                          2System Failure LED is ON with audio alarm

                          System Failure Contact nice Neotech

                          General System Failure

                          System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                          1 No Display on the Control PanelCheck AC Supply Voltage

                          Check if Power Cord is disconnected

                          Check Fuse in the Rear side of Pillar

                          Contact nice Neotech

                          2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                          Set Temperature may be less than Actual Temperature

                          Check Heater Connection

                          Contact nice Neotech

                          3 Equipment Working But Observation Lamp amp Heater output not working

                          Check Link cord connected between Control Module toHeater Module

                          Check Fuse in the Rear side of Heater Module

                          Check Transformer Voltage - 12V AC

                          Replace Lamp

                          4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                          Check Mute ON

                          Contact nice Neotech

                          5 Patient Temperature Display shows ldquoS-Ordquo

                          with Continuous Alarm(S-O Sensor Open)

                          Check if skin Probe is Disconnected from theEquipment

                          If skin probe is defective replace new probe

                          Contact nice Neotech

                          7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                          Check Heater Connection

                          Contact nice Neotech

                          8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                          Check Environment Temperature above 39deg C

                          Check Infant Fever

                          9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                          Temperature

                          Check Environment Temperature is 1degC above SetTemperature

                          Contact nice Neotech

                          10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                          Temperature

                          Probe may be wet

                          Contact nice Neotech

                          13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                          Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                          Low Supply Voltage

                          14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                          Percentage Heater Output is Set at High in ManualMode

                          15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                          Check for Strong Air conditioning draft

                          16Temperature Display does not display InfantsTemperature accurately

                          Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                          17Baby skin Temperature not correlate with display Temperature

                          Check sensitive portion of the Probe is fixed with babyskin properly

                          Check Probe is fixed to the lower abdomen of baby

                          18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                          19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                          Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                          System Fault Infant Radiant Warmer(Contrsquodhellip)

                          Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                          Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                          components Do not autoclave or gas sterilize the skin temperature probe Do not

                          immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                          Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                          should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                          Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                          and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                          equipment performance

                          Certifications and Approvals

                          Electrical Safety IEC 60601-1 IEC 60601-1-4

                          Particular product IEC 60601-2-21

                          EMC Safety IEC 60601-1-2

                          Contact Us

                          nice Neotech Medical Systems Private Limited

                          No 85 Krishna Industrial Estate

                          Mettukuppam Vannagaram

                          Chennai ndash 600095 Tamil Nadu

                          INDIA

                          Tel +91-44-2476259424764608

                          Fax +91-44-24766920

                          Web wwwniceneotechcom

                          E-mail infoniceneotechcom marketingniceneotechcom

                          Overseas Enquiry overseasniceneotechcom

                          Toll Free +91-1800-425-2594

                          • nice NeOumltech Medical Systems Pvt Ltd
                          • Presentation Flow
                          • Company INFO
                          • Timeline-nice Neotech
                          • Why nice Neotech
                          • Why Infant Radiant Warmers are Used
                          • MODES OF HEAT LOSS
                          • Slide 8
                          • Effects of Far IR in Infant body
                          • Types of heaters
                          • Slide 11
                          • Benefits of Heaterhellip
                          • Slide 13
                          • Slide 14
                          • Slide 15
                          • How to operate
                          • Control Panel Operation
                          • Operating Modes
                          • Slide 19
                          • Slide 20
                          • Alarms
                          • Heater Module
                          • Bed Platform Operation
                          • Side Panel Operation
                          • X-ray Procedure
                          • Comparison of nice 2010 Infant Radiant Warmers
                          • How nice Infant Radiant Warmers fare over competitors
                          • Slide 28
                          • Slide 29
                          • Cleaning Procedure
                          • Preventive Maintenance
                          • Troubleshooting
                          • System Fault Infant Radiant Warmer
                          • Slide 34
                          • Slide 35
                          • Certifications and Approvals
                          • Contact Us

                            nice 2010 series Infant Radiant Warmers

                            nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

                            nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

                            nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

                            bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                            Features

                            How to operate

                            Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                            Control Panel Operation

                            1 Skin Temperature Display 15

                            Manual ModeIndicator

                            2 Sensor Failure Indicator 16

                            Skin Mode Indicator

                            3 Heater Failure Indicator 17

                            Pre Warm Mode Indicator

                            4 System Failure Indicator 18

                            Increase Key

                            5 PowerFailure Indicator 19

                            Mute Key

                            6 Skin Temperature gt 39degC Indicator

                            20

                            Heater OutputIndicator

                            7 Lamp Indicator 21

                            Set TemperatureManualAPGAR Timer

                            8 Lock Key 22

                            APGAR Timer Indicator

                            9 Lock Indicator 23

                            Mute ON Indicator

                            10

                            Lamp Key 24

                            Manual Timer Indicator

                            11

                            APGAR Kay 25

                            Skin Temperature Low Indicator

                            12

                            Decrease Key 26

                            Skin Temperature High Indicator

                            13

                            Mode Key 27

                            Alarm Indicator

                            14

                            Safe Mode Indicator

                            Operating Modes

                            Manual Mode Operation

                            1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                            1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                            1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                            1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                            Skin Mode Operation

                            Note The skin temperature probe must be properly attached before choosing Skin mode operation

                            1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                            1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                            APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                            Operating Modes(contrsquodhellip)

                            Operating Modes(contrsquodhellip)

                            Pre-Warm Mode

                            bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                            Safe Mode

                            bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                            Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                            AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                            Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                            Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                            Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                            Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                            Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                            Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                            Heater Module

                            Parabolic design

                            Even Distribution of heat over bed surface

                            Extensive safety

                            measures(6 Layer safety for thermal

                            cut- off)

                            Far IR heating ensures

                            stimulation of cells for growth

                            The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                            Bed Platform Operation

                            bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                            bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                            Side Panel Operation

                            bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                            X-ray Procedure

                            bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                            bullThe heater housing rotates to the left or right for X-ray procedures

                            bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                            bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                            Comparison of nice 2010 Infant Radiant Warmers

                            nice 2010 BC nice 2010 B nice 2010

                            Skin Mode Y Y Y

                            Manual Mode Y Y Y

                            Safe Mode Y Y Y

                            Pre-warm Mode Y Y Y

                            X Ray Facility Y Y N

                            Bed Y Y N

                            Pivoting Warmer ModuleY Y Y

                            Apgar Timer Y Y Y

                            Height Adjustment N N Y

                            Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                            How nice Infant Radiant Warmers fare over competitors

                            nice 2010 series warmers

                            User-friendly easy to access all parameters

                            Long life quartz heater has a long life compared to other heaters

                            6 Layer safety for the protection of infant from burning

                            6 Layer Safety Features

                            Alarms if infant temperature greater or less than Set Temperature

                            Thermal cut off if infant temperature exceeds 39 degC

                            Heater output reduces to 30 after 15 mins in Manual mode

                            Heater output reduces to 30 if electronics senses no variable feedback

                            Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                            Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                            TROUBLESHOOTING

                            AND MAINTENANCE

                            Cleaning Procedure

                            1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                            2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                            3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                            Cleaning Solutions for Mattress

                            Generic Formulations Maximum concentration level

                            Hydrogen Peroxide 6

                            Quaternary Compound 05 aqueous solution

                            Quaternary Ammonium 02

                            Lodophors 50mglitre

                            Preventive Maintenance

                            1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                            Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                            1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                            TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                            Sl No Problem Cause Remedy

                            1 Power failure LED is ON with audio alarm

                            Power FailureCheck the unit is Plugged in to Main supply

                            Check the mains are switched ON

                            Else Contact nice Neotech

                            2System Failure LED is ON with audio alarm

                            System Failure Contact nice Neotech

                            General System Failure

                            System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                            1 No Display on the Control PanelCheck AC Supply Voltage

                            Check if Power Cord is disconnected

                            Check Fuse in the Rear side of Pillar

                            Contact nice Neotech

                            2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                            Set Temperature may be less than Actual Temperature

                            Check Heater Connection

                            Contact nice Neotech

                            3 Equipment Working But Observation Lamp amp Heater output not working

                            Check Link cord connected between Control Module toHeater Module

                            Check Fuse in the Rear side of Heater Module

                            Check Transformer Voltage - 12V AC

                            Replace Lamp

                            4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                            Check Mute ON

                            Contact nice Neotech

                            5 Patient Temperature Display shows ldquoS-Ordquo

                            with Continuous Alarm(S-O Sensor Open)

                            Check if skin Probe is Disconnected from theEquipment

                            If skin probe is defective replace new probe

                            Contact nice Neotech

                            7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                            Check Heater Connection

                            Contact nice Neotech

                            8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                            Check Environment Temperature above 39deg C

                            Check Infant Fever

                            9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                            Temperature

                            Check Environment Temperature is 1degC above SetTemperature

                            Contact nice Neotech

                            10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                            Temperature

                            Probe may be wet

                            Contact nice Neotech

                            13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                            Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                            Low Supply Voltage

                            14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                            Percentage Heater Output is Set at High in ManualMode

                            15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                            Check for Strong Air conditioning draft

                            16Temperature Display does not display InfantsTemperature accurately

                            Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                            17Baby skin Temperature not correlate with display Temperature

                            Check sensitive portion of the Probe is fixed with babyskin properly

                            Check Probe is fixed to the lower abdomen of baby

                            18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                            19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                            Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                            System Fault Infant Radiant Warmer(Contrsquodhellip)

                            Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                            Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                            components Do not autoclave or gas sterilize the skin temperature probe Do not

                            immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                            Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                            should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                            Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                            and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                            equipment performance

                            Certifications and Approvals

                            Electrical Safety IEC 60601-1 IEC 60601-1-4

                            Particular product IEC 60601-2-21

                            EMC Safety IEC 60601-1-2

                            Contact Us

                            nice Neotech Medical Systems Private Limited

                            No 85 Krishna Industrial Estate

                            Mettukuppam Vannagaram

                            Chennai ndash 600095 Tamil Nadu

                            INDIA

                            Tel +91-44-2476259424764608

                            Fax +91-44-24766920

                            Web wwwniceneotechcom

                            E-mail infoniceneotechcom marketingniceneotechcom

                            Overseas Enquiry overseasniceneotechcom

                            Toll Free +91-1800-425-2594

                            • nice NeOumltech Medical Systems Pvt Ltd
                            • Presentation Flow
                            • Company INFO
                            • Timeline-nice Neotech
                            • Why nice Neotech
                            • Why Infant Radiant Warmers are Used
                            • MODES OF HEAT LOSS
                            • Slide 8
                            • Effects of Far IR in Infant body
                            • Types of heaters
                            • Slide 11
                            • Benefits of Heaterhellip
                            • Slide 13
                            • Slide 14
                            • Slide 15
                            • How to operate
                            • Control Panel Operation
                            • Operating Modes
                            • Slide 19
                            • Slide 20
                            • Alarms
                            • Heater Module
                            • Bed Platform Operation
                            • Side Panel Operation
                            • X-ray Procedure
                            • Comparison of nice 2010 Infant Radiant Warmers
                            • How nice Infant Radiant Warmers fare over competitors
                            • Slide 28
                            • Slide 29
                            • Cleaning Procedure
                            • Preventive Maintenance
                            • Troubleshooting
                            • System Fault Infant Radiant Warmer
                            • Slide 34
                            • Slide 35
                            • Certifications and Approvals
                            • Contact Us

                              bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

                              Features

                              How to operate

                              Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                              Control Panel Operation

                              1 Skin Temperature Display 15

                              Manual ModeIndicator

                              2 Sensor Failure Indicator 16

                              Skin Mode Indicator

                              3 Heater Failure Indicator 17

                              Pre Warm Mode Indicator

                              4 System Failure Indicator 18

                              Increase Key

                              5 PowerFailure Indicator 19

                              Mute Key

                              6 Skin Temperature gt 39degC Indicator

                              20

                              Heater OutputIndicator

                              7 Lamp Indicator 21

                              Set TemperatureManualAPGAR Timer

                              8 Lock Key 22

                              APGAR Timer Indicator

                              9 Lock Indicator 23

                              Mute ON Indicator

                              10

                              Lamp Key 24

                              Manual Timer Indicator

                              11

                              APGAR Kay 25

                              Skin Temperature Low Indicator

                              12

                              Decrease Key 26

                              Skin Temperature High Indicator

                              13

                              Mode Key 27

                              Alarm Indicator

                              14

                              Safe Mode Indicator

                              Operating Modes

                              Manual Mode Operation

                              1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                              1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                              1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                              1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                              Skin Mode Operation

                              Note The skin temperature probe must be properly attached before choosing Skin mode operation

                              1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                              1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                              APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                              Operating Modes(contrsquodhellip)

                              Operating Modes(contrsquodhellip)

                              Pre-Warm Mode

                              bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                              Safe Mode

                              bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                              Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                              AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                              Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                              Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                              Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                              Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                              Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                              Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                              Heater Module

                              Parabolic design

                              Even Distribution of heat over bed surface

                              Extensive safety

                              measures(6 Layer safety for thermal

                              cut- off)

                              Far IR heating ensures

                              stimulation of cells for growth

                              The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                              Bed Platform Operation

                              bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                              bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                              Side Panel Operation

                              bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                              X-ray Procedure

                              bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                              bullThe heater housing rotates to the left or right for X-ray procedures

                              bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                              bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                              Comparison of nice 2010 Infant Radiant Warmers

                              nice 2010 BC nice 2010 B nice 2010

                              Skin Mode Y Y Y

                              Manual Mode Y Y Y

                              Safe Mode Y Y Y

                              Pre-warm Mode Y Y Y

                              X Ray Facility Y Y N

                              Bed Y Y N

                              Pivoting Warmer ModuleY Y Y

                              Apgar Timer Y Y Y

                              Height Adjustment N N Y

                              Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                              How nice Infant Radiant Warmers fare over competitors

                              nice 2010 series warmers

                              User-friendly easy to access all parameters

                              Long life quartz heater has a long life compared to other heaters

                              6 Layer safety for the protection of infant from burning

                              6 Layer Safety Features

                              Alarms if infant temperature greater or less than Set Temperature

                              Thermal cut off if infant temperature exceeds 39 degC

                              Heater output reduces to 30 after 15 mins in Manual mode

                              Heater output reduces to 30 if electronics senses no variable feedback

                              Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                              Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                              TROUBLESHOOTING

                              AND MAINTENANCE

                              Cleaning Procedure

                              1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                              2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                              3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                              Cleaning Solutions for Mattress

                              Generic Formulations Maximum concentration level

                              Hydrogen Peroxide 6

                              Quaternary Compound 05 aqueous solution

                              Quaternary Ammonium 02

                              Lodophors 50mglitre

                              Preventive Maintenance

                              1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                              Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                              1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                              TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                              Sl No Problem Cause Remedy

                              1 Power failure LED is ON with audio alarm

                              Power FailureCheck the unit is Plugged in to Main supply

                              Check the mains are switched ON

                              Else Contact nice Neotech

                              2System Failure LED is ON with audio alarm

                              System Failure Contact nice Neotech

                              General System Failure

                              System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                              1 No Display on the Control PanelCheck AC Supply Voltage

                              Check if Power Cord is disconnected

                              Check Fuse in the Rear side of Pillar

                              Contact nice Neotech

                              2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                              Set Temperature may be less than Actual Temperature

                              Check Heater Connection

                              Contact nice Neotech

                              3 Equipment Working But Observation Lamp amp Heater output not working

                              Check Link cord connected between Control Module toHeater Module

                              Check Fuse in the Rear side of Heater Module

                              Check Transformer Voltage - 12V AC

                              Replace Lamp

                              4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                              Check Mute ON

                              Contact nice Neotech

                              5 Patient Temperature Display shows ldquoS-Ordquo

                              with Continuous Alarm(S-O Sensor Open)

                              Check if skin Probe is Disconnected from theEquipment

                              If skin probe is defective replace new probe

                              Contact nice Neotech

                              7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                              Check Heater Connection

                              Contact nice Neotech

                              8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                              Check Environment Temperature above 39deg C

                              Check Infant Fever

                              9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                              Temperature

                              Check Environment Temperature is 1degC above SetTemperature

                              Contact nice Neotech

                              10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                              Temperature

                              Probe may be wet

                              Contact nice Neotech

                              13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                              Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                              Low Supply Voltage

                              14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                              Percentage Heater Output is Set at High in ManualMode

                              15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                              Check for Strong Air conditioning draft

                              16Temperature Display does not display InfantsTemperature accurately

                              Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                              17Baby skin Temperature not correlate with display Temperature

                              Check sensitive portion of the Probe is fixed with babyskin properly

                              Check Probe is fixed to the lower abdomen of baby

                              18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                              19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                              Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                              System Fault Infant Radiant Warmer(Contrsquodhellip)

                              Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                              Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                              components Do not autoclave or gas sterilize the skin temperature probe Do not

                              immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                              Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                              should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                              Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                              and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                              equipment performance

                              Certifications and Approvals

                              Electrical Safety IEC 60601-1 IEC 60601-1-4

                              Particular product IEC 60601-2-21

                              EMC Safety IEC 60601-1-2

                              Contact Us

                              nice Neotech Medical Systems Private Limited

                              No 85 Krishna Industrial Estate

                              Mettukuppam Vannagaram

                              Chennai ndash 600095 Tamil Nadu

                              INDIA

                              Tel +91-44-2476259424764608

                              Fax +91-44-24766920

                              Web wwwniceneotechcom

                              E-mail infoniceneotechcom marketingniceneotechcom

                              Overseas Enquiry overseasniceneotechcom

                              Toll Free +91-1800-425-2594

                              • nice NeOumltech Medical Systems Pvt Ltd
                              • Presentation Flow
                              • Company INFO
                              • Timeline-nice Neotech
                              • Why nice Neotech
                              • Why Infant Radiant Warmers are Used
                              • MODES OF HEAT LOSS
                              • Slide 8
                              • Effects of Far IR in Infant body
                              • Types of heaters
                              • Slide 11
                              • Benefits of Heaterhellip
                              • Slide 13
                              • Slide 14
                              • Slide 15
                              • How to operate
                              • Control Panel Operation
                              • Operating Modes
                              • Slide 19
                              • Slide 20
                              • Alarms
                              • Heater Module
                              • Bed Platform Operation
                              • Side Panel Operation
                              • X-ray Procedure
                              • Comparison of nice 2010 Infant Radiant Warmers
                              • How nice Infant Radiant Warmers fare over competitors
                              • Slide 28
                              • Slide 29
                              • Cleaning Procedure
                              • Preventive Maintenance
                              • Troubleshooting
                              • System Fault Infant Radiant Warmer
                              • Slide 34
                              • Slide 35
                              • Certifications and Approvals
                              • Contact Us

                                How to operate

                                Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

                                Control Panel Operation

                                1 Skin Temperature Display 15

                                Manual ModeIndicator

                                2 Sensor Failure Indicator 16

                                Skin Mode Indicator

                                3 Heater Failure Indicator 17

                                Pre Warm Mode Indicator

                                4 System Failure Indicator 18

                                Increase Key

                                5 PowerFailure Indicator 19

                                Mute Key

                                6 Skin Temperature gt 39degC Indicator

                                20

                                Heater OutputIndicator

                                7 Lamp Indicator 21

                                Set TemperatureManualAPGAR Timer

                                8 Lock Key 22

                                APGAR Timer Indicator

                                9 Lock Indicator 23

                                Mute ON Indicator

                                10

                                Lamp Key 24

                                Manual Timer Indicator

                                11

                                APGAR Kay 25

                                Skin Temperature Low Indicator

                                12

                                Decrease Key 26

                                Skin Temperature High Indicator

                                13

                                Mode Key 27

                                Alarm Indicator

                                14

                                Safe Mode Indicator

                                Operating Modes

                                Manual Mode Operation

                                1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                                1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                                1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                                1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                                Skin Mode Operation

                                Note The skin temperature probe must be properly attached before choosing Skin mode operation

                                1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                                1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                                APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                                Operating Modes(contrsquodhellip)

                                Operating Modes(contrsquodhellip)

                                Pre-Warm Mode

                                bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                                Safe Mode

                                bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                                Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                                AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                                Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                                Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                                Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                                Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                                Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                                Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                                Heater Module

                                Parabolic design

                                Even Distribution of heat over bed surface

                                Extensive safety

                                measures(6 Layer safety for thermal

                                cut- off)

                                Far IR heating ensures

                                stimulation of cells for growth

                                The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                Bed Platform Operation

                                bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                Side Panel Operation

                                bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                X-ray Procedure

                                bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                bullThe heater housing rotates to the left or right for X-ray procedures

                                bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                Comparison of nice 2010 Infant Radiant Warmers

                                nice 2010 BC nice 2010 B nice 2010

                                Skin Mode Y Y Y

                                Manual Mode Y Y Y

                                Safe Mode Y Y Y

                                Pre-warm Mode Y Y Y

                                X Ray Facility Y Y N

                                Bed Y Y N

                                Pivoting Warmer ModuleY Y Y

                                Apgar Timer Y Y Y

                                Height Adjustment N N Y

                                Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                How nice Infant Radiant Warmers fare over competitors

                                nice 2010 series warmers

                                User-friendly easy to access all parameters

                                Long life quartz heater has a long life compared to other heaters

                                6 Layer safety for the protection of infant from burning

                                6 Layer Safety Features

                                Alarms if infant temperature greater or less than Set Temperature

                                Thermal cut off if infant temperature exceeds 39 degC

                                Heater output reduces to 30 after 15 mins in Manual mode

                                Heater output reduces to 30 if electronics senses no variable feedback

                                Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                TROUBLESHOOTING

                                AND MAINTENANCE

                                Cleaning Procedure

                                1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                Cleaning Solutions for Mattress

                                Generic Formulations Maximum concentration level

                                Hydrogen Peroxide 6

                                Quaternary Compound 05 aqueous solution

                                Quaternary Ammonium 02

                                Lodophors 50mglitre

                                Preventive Maintenance

                                1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                Sl No Problem Cause Remedy

                                1 Power failure LED is ON with audio alarm

                                Power FailureCheck the unit is Plugged in to Main supply

                                Check the mains are switched ON

                                Else Contact nice Neotech

                                2System Failure LED is ON with audio alarm

                                System Failure Contact nice Neotech

                                General System Failure

                                System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                1 No Display on the Control PanelCheck AC Supply Voltage

                                Check if Power Cord is disconnected

                                Check Fuse in the Rear side of Pillar

                                Contact nice Neotech

                                2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                Set Temperature may be less than Actual Temperature

                                Check Heater Connection

                                Contact nice Neotech

                                3 Equipment Working But Observation Lamp amp Heater output not working

                                Check Link cord connected between Control Module toHeater Module

                                Check Fuse in the Rear side of Heater Module

                                Check Transformer Voltage - 12V AC

                                Replace Lamp

                                4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                Check Mute ON

                                Contact nice Neotech

                                5 Patient Temperature Display shows ldquoS-Ordquo

                                with Continuous Alarm(S-O Sensor Open)

                                Check if skin Probe is Disconnected from theEquipment

                                If skin probe is defective replace new probe

                                Contact nice Neotech

                                7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                Check Heater Connection

                                Contact nice Neotech

                                8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                Check Environment Temperature above 39deg C

                                Check Infant Fever

                                9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                Temperature

                                Check Environment Temperature is 1degC above SetTemperature

                                Contact nice Neotech

                                10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                Temperature

                                Probe may be wet

                                Contact nice Neotech

                                13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                Low Supply Voltage

                                14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                Percentage Heater Output is Set at High in ManualMode

                                15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                Check for Strong Air conditioning draft

                                16Temperature Display does not display InfantsTemperature accurately

                                Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                17Baby skin Temperature not correlate with display Temperature

                                Check sensitive portion of the Probe is fixed with babyskin properly

                                Check Probe is fixed to the lower abdomen of baby

                                18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                System Fault Infant Radiant Warmer(Contrsquodhellip)

                                Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                components Do not autoclave or gas sterilize the skin temperature probe Do not

                                immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                equipment performance

                                Certifications and Approvals

                                Electrical Safety IEC 60601-1 IEC 60601-1-4

                                Particular product IEC 60601-2-21

                                EMC Safety IEC 60601-1-2

                                Contact Us

                                nice Neotech Medical Systems Private Limited

                                No 85 Krishna Industrial Estate

                                Mettukuppam Vannagaram

                                Chennai ndash 600095 Tamil Nadu

                                INDIA

                                Tel +91-44-2476259424764608

                                Fax +91-44-24766920

                                Web wwwniceneotechcom

                                E-mail infoniceneotechcom marketingniceneotechcom

                                Overseas Enquiry overseasniceneotechcom

                                Toll Free +91-1800-425-2594

                                • nice NeOumltech Medical Systems Pvt Ltd
                                • Presentation Flow
                                • Company INFO
                                • Timeline-nice Neotech
                                • Why nice Neotech
                                • Why Infant Radiant Warmers are Used
                                • MODES OF HEAT LOSS
                                • Slide 8
                                • Effects of Far IR in Infant body
                                • Types of heaters
                                • Slide 11
                                • Benefits of Heaterhellip
                                • Slide 13
                                • Slide 14
                                • Slide 15
                                • How to operate
                                • Control Panel Operation
                                • Operating Modes
                                • Slide 19
                                • Slide 20
                                • Alarms
                                • Heater Module
                                • Bed Platform Operation
                                • Side Panel Operation
                                • X-ray Procedure
                                • Comparison of nice 2010 Infant Radiant Warmers
                                • How nice Infant Radiant Warmers fare over competitors
                                • Slide 28
                                • Slide 29
                                • Cleaning Procedure
                                • Preventive Maintenance
                                • Troubleshooting
                                • System Fault Infant Radiant Warmer
                                • Slide 34
                                • Slide 35
                                • Certifications and Approvals
                                • Contact Us

                                  Control Panel Operation

                                  1 Skin Temperature Display 15

                                  Manual ModeIndicator

                                  2 Sensor Failure Indicator 16

                                  Skin Mode Indicator

                                  3 Heater Failure Indicator 17

                                  Pre Warm Mode Indicator

                                  4 System Failure Indicator 18

                                  Increase Key

                                  5 PowerFailure Indicator 19

                                  Mute Key

                                  6 Skin Temperature gt 39degC Indicator

                                  20

                                  Heater OutputIndicator

                                  7 Lamp Indicator 21

                                  Set TemperatureManualAPGAR Timer

                                  8 Lock Key 22

                                  APGAR Timer Indicator

                                  9 Lock Indicator 23

                                  Mute ON Indicator

                                  10

                                  Lamp Key 24

                                  Manual Timer Indicator

                                  11

                                  APGAR Kay 25

                                  Skin Temperature Low Indicator

                                  12

                                  Decrease Key 26

                                  Skin Temperature High Indicator

                                  13

                                  Mode Key 27

                                  Alarm Indicator

                                  14

                                  Safe Mode Indicator

                                  Operating Modes

                                  Manual Mode Operation

                                  1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                                  1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                                  1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                                  1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                                  Skin Mode Operation

                                  Note The skin temperature probe must be properly attached before choosing Skin mode operation

                                  1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                                  1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                                  APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                                  Operating Modes(contrsquodhellip)

                                  Operating Modes(contrsquodhellip)

                                  Pre-Warm Mode

                                  bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                                  Safe Mode

                                  bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                                  Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                                  AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                                  Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                                  Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                                  Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                                  Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                                  Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                                  Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                                  Heater Module

                                  Parabolic design

                                  Even Distribution of heat over bed surface

                                  Extensive safety

                                  measures(6 Layer safety for thermal

                                  cut- off)

                                  Far IR heating ensures

                                  stimulation of cells for growth

                                  The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                  Bed Platform Operation

                                  bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                  bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                  Side Panel Operation

                                  bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                  X-ray Procedure

                                  bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                  bullThe heater housing rotates to the left or right for X-ray procedures

                                  bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                  bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                  Comparison of nice 2010 Infant Radiant Warmers

                                  nice 2010 BC nice 2010 B nice 2010

                                  Skin Mode Y Y Y

                                  Manual Mode Y Y Y

                                  Safe Mode Y Y Y

                                  Pre-warm Mode Y Y Y

                                  X Ray Facility Y Y N

                                  Bed Y Y N

                                  Pivoting Warmer ModuleY Y Y

                                  Apgar Timer Y Y Y

                                  Height Adjustment N N Y

                                  Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                  How nice Infant Radiant Warmers fare over competitors

                                  nice 2010 series warmers

                                  User-friendly easy to access all parameters

                                  Long life quartz heater has a long life compared to other heaters

                                  6 Layer safety for the protection of infant from burning

                                  6 Layer Safety Features

                                  Alarms if infant temperature greater or less than Set Temperature

                                  Thermal cut off if infant temperature exceeds 39 degC

                                  Heater output reduces to 30 after 15 mins in Manual mode

                                  Heater output reduces to 30 if electronics senses no variable feedback

                                  Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                  Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                  TROUBLESHOOTING

                                  AND MAINTENANCE

                                  Cleaning Procedure

                                  1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                  2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                  3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                  Cleaning Solutions for Mattress

                                  Generic Formulations Maximum concentration level

                                  Hydrogen Peroxide 6

                                  Quaternary Compound 05 aqueous solution

                                  Quaternary Ammonium 02

                                  Lodophors 50mglitre

                                  Preventive Maintenance

                                  1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                  Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                  1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                  TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                  Sl No Problem Cause Remedy

                                  1 Power failure LED is ON with audio alarm

                                  Power FailureCheck the unit is Plugged in to Main supply

                                  Check the mains are switched ON

                                  Else Contact nice Neotech

                                  2System Failure LED is ON with audio alarm

                                  System Failure Contact nice Neotech

                                  General System Failure

                                  System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                  1 No Display on the Control PanelCheck AC Supply Voltage

                                  Check if Power Cord is disconnected

                                  Check Fuse in the Rear side of Pillar

                                  Contact nice Neotech

                                  2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                  Set Temperature may be less than Actual Temperature

                                  Check Heater Connection

                                  Contact nice Neotech

                                  3 Equipment Working But Observation Lamp amp Heater output not working

                                  Check Link cord connected between Control Module toHeater Module

                                  Check Fuse in the Rear side of Heater Module

                                  Check Transformer Voltage - 12V AC

                                  Replace Lamp

                                  4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                  Check Mute ON

                                  Contact nice Neotech

                                  5 Patient Temperature Display shows ldquoS-Ordquo

                                  with Continuous Alarm(S-O Sensor Open)

                                  Check if skin Probe is Disconnected from theEquipment

                                  If skin probe is defective replace new probe

                                  Contact nice Neotech

                                  7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                  Check Heater Connection

                                  Contact nice Neotech

                                  8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                  Check Environment Temperature above 39deg C

                                  Check Infant Fever

                                  9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                  Temperature

                                  Check Environment Temperature is 1degC above SetTemperature

                                  Contact nice Neotech

                                  10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                  Temperature

                                  Probe may be wet

                                  Contact nice Neotech

                                  13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                  Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                  Low Supply Voltage

                                  14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                  Percentage Heater Output is Set at High in ManualMode

                                  15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                  Check for Strong Air conditioning draft

                                  16Temperature Display does not display InfantsTemperature accurately

                                  Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                  17Baby skin Temperature not correlate with display Temperature

                                  Check sensitive portion of the Probe is fixed with babyskin properly

                                  Check Probe is fixed to the lower abdomen of baby

                                  18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                  19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                  Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                  System Fault Infant Radiant Warmer(Contrsquodhellip)

                                  Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                  Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                  components Do not autoclave or gas sterilize the skin temperature probe Do not

                                  immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                  Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                  should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                  Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                  and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                  equipment performance

                                  Certifications and Approvals

                                  Electrical Safety IEC 60601-1 IEC 60601-1-4

                                  Particular product IEC 60601-2-21

                                  EMC Safety IEC 60601-1-2

                                  Contact Us

                                  nice Neotech Medical Systems Private Limited

                                  No 85 Krishna Industrial Estate

                                  Mettukuppam Vannagaram

                                  Chennai ndash 600095 Tamil Nadu

                                  INDIA

                                  Tel +91-44-2476259424764608

                                  Fax +91-44-24766920

                                  Web wwwniceneotechcom

                                  E-mail infoniceneotechcom marketingniceneotechcom

                                  Overseas Enquiry overseasniceneotechcom

                                  Toll Free +91-1800-425-2594

                                  • nice NeOumltech Medical Systems Pvt Ltd
                                  • Presentation Flow
                                  • Company INFO
                                  • Timeline-nice Neotech
                                  • Why nice Neotech
                                  • Why Infant Radiant Warmers are Used
                                  • MODES OF HEAT LOSS
                                  • Slide 8
                                  • Effects of Far IR in Infant body
                                  • Types of heaters
                                  • Slide 11
                                  • Benefits of Heaterhellip
                                  • Slide 13
                                  • Slide 14
                                  • Slide 15
                                  • How to operate
                                  • Control Panel Operation
                                  • Operating Modes
                                  • Slide 19
                                  • Slide 20
                                  • Alarms
                                  • Heater Module
                                  • Bed Platform Operation
                                  • Side Panel Operation
                                  • X-ray Procedure
                                  • Comparison of nice 2010 Infant Radiant Warmers
                                  • How nice Infant Radiant Warmers fare over competitors
                                  • Slide 28
                                  • Slide 29
                                  • Cleaning Procedure
                                  • Preventive Maintenance
                                  • Troubleshooting
                                  • System Fault Infant Radiant Warmer
                                  • Slide 34
                                  • Slide 35
                                  • Certifications and Approvals
                                  • Contact Us

                                    Operating Modes

                                    Manual Mode Operation

                                    1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

                                    1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

                                    1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

                                    1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

                                    Skin Mode Operation

                                    Note The skin temperature probe must be properly attached before choosing Skin mode operation

                                    1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                                    1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                                    APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                                    Operating Modes(contrsquodhellip)

                                    Operating Modes(contrsquodhellip)

                                    Pre-Warm Mode

                                    bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                                    Safe Mode

                                    bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                                    Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                                    AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                                    Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                                    Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                                    Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                                    Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                                    Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                                    Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                                    Heater Module

                                    Parabolic design

                                    Even Distribution of heat over bed surface

                                    Extensive safety

                                    measures(6 Layer safety for thermal

                                    cut- off)

                                    Far IR heating ensures

                                    stimulation of cells for growth

                                    The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                    Bed Platform Operation

                                    bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                    bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                    Side Panel Operation

                                    bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                    X-ray Procedure

                                    bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                    bullThe heater housing rotates to the left or right for X-ray procedures

                                    bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                    bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                    Comparison of nice 2010 Infant Radiant Warmers

                                    nice 2010 BC nice 2010 B nice 2010

                                    Skin Mode Y Y Y

                                    Manual Mode Y Y Y

                                    Safe Mode Y Y Y

                                    Pre-warm Mode Y Y Y

                                    X Ray Facility Y Y N

                                    Bed Y Y N

                                    Pivoting Warmer ModuleY Y Y

                                    Apgar Timer Y Y Y

                                    Height Adjustment N N Y

                                    Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                    How nice Infant Radiant Warmers fare over competitors

                                    nice 2010 series warmers

                                    User-friendly easy to access all parameters

                                    Long life quartz heater has a long life compared to other heaters

                                    6 Layer safety for the protection of infant from burning

                                    6 Layer Safety Features

                                    Alarms if infant temperature greater or less than Set Temperature

                                    Thermal cut off if infant temperature exceeds 39 degC

                                    Heater output reduces to 30 after 15 mins in Manual mode

                                    Heater output reduces to 30 if electronics senses no variable feedback

                                    Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                    Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                    TROUBLESHOOTING

                                    AND MAINTENANCE

                                    Cleaning Procedure

                                    1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                    2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                    3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                    Cleaning Solutions for Mattress

                                    Generic Formulations Maximum concentration level

                                    Hydrogen Peroxide 6

                                    Quaternary Compound 05 aqueous solution

                                    Quaternary Ammonium 02

                                    Lodophors 50mglitre

                                    Preventive Maintenance

                                    1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                    Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                    1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                    TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                    Sl No Problem Cause Remedy

                                    1 Power failure LED is ON with audio alarm

                                    Power FailureCheck the unit is Plugged in to Main supply

                                    Check the mains are switched ON

                                    Else Contact nice Neotech

                                    2System Failure LED is ON with audio alarm

                                    System Failure Contact nice Neotech

                                    General System Failure

                                    System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                    1 No Display on the Control PanelCheck AC Supply Voltage

                                    Check if Power Cord is disconnected

                                    Check Fuse in the Rear side of Pillar

                                    Contact nice Neotech

                                    2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                    Set Temperature may be less than Actual Temperature

                                    Check Heater Connection

                                    Contact nice Neotech

                                    3 Equipment Working But Observation Lamp amp Heater output not working

                                    Check Link cord connected between Control Module toHeater Module

                                    Check Fuse in the Rear side of Heater Module

                                    Check Transformer Voltage - 12V AC

                                    Replace Lamp

                                    4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                    Check Mute ON

                                    Contact nice Neotech

                                    5 Patient Temperature Display shows ldquoS-Ordquo

                                    with Continuous Alarm(S-O Sensor Open)

                                    Check if skin Probe is Disconnected from theEquipment

                                    If skin probe is defective replace new probe

                                    Contact nice Neotech

                                    7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                    Check Heater Connection

                                    Contact nice Neotech

                                    8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                    Check Environment Temperature above 39deg C

                                    Check Infant Fever

                                    9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                    Temperature

                                    Check Environment Temperature is 1degC above SetTemperature

                                    Contact nice Neotech

                                    10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                    Temperature

                                    Probe may be wet

                                    Contact nice Neotech

                                    13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                    Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                    Low Supply Voltage

                                    14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                    Percentage Heater Output is Set at High in ManualMode

                                    15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                    Check for Strong Air conditioning draft

                                    16Temperature Display does not display InfantsTemperature accurately

                                    Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                    17Baby skin Temperature not correlate with display Temperature

                                    Check sensitive portion of the Probe is fixed with babyskin properly

                                    Check Probe is fixed to the lower abdomen of baby

                                    18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                    19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                    Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                    System Fault Infant Radiant Warmer(Contrsquodhellip)

                                    Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                    Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                    components Do not autoclave or gas sterilize the skin temperature probe Do not

                                    immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                    Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                    should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                    Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                    and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                    equipment performance

                                    Certifications and Approvals

                                    Electrical Safety IEC 60601-1 IEC 60601-1-4

                                    Particular product IEC 60601-2-21

                                    EMC Safety IEC 60601-1-2

                                    Contact Us

                                    nice Neotech Medical Systems Private Limited

                                    No 85 Krishna Industrial Estate

                                    Mettukuppam Vannagaram

                                    Chennai ndash 600095 Tamil Nadu

                                    INDIA

                                    Tel +91-44-2476259424764608

                                    Fax +91-44-24766920

                                    Web wwwniceneotechcom

                                    E-mail infoniceneotechcom marketingniceneotechcom

                                    Overseas Enquiry overseasniceneotechcom

                                    Toll Free +91-1800-425-2594

                                    • nice NeOumltech Medical Systems Pvt Ltd
                                    • Presentation Flow
                                    • Company INFO
                                    • Timeline-nice Neotech
                                    • Why nice Neotech
                                    • Why Infant Radiant Warmers are Used
                                    • MODES OF HEAT LOSS
                                    • Slide 8
                                    • Effects of Far IR in Infant body
                                    • Types of heaters
                                    • Slide 11
                                    • Benefits of Heaterhellip
                                    • Slide 13
                                    • Slide 14
                                    • Slide 15
                                    • How to operate
                                    • Control Panel Operation
                                    • Operating Modes
                                    • Slide 19
                                    • Slide 20
                                    • Alarms
                                    • Heater Module
                                    • Bed Platform Operation
                                    • Side Panel Operation
                                    • X-ray Procedure
                                    • Comparison of nice 2010 Infant Radiant Warmers
                                    • How nice Infant Radiant Warmers fare over competitors
                                    • Slide 28
                                    • Slide 29
                                    • Cleaning Procedure
                                    • Preventive Maintenance
                                    • Troubleshooting
                                    • System Fault Infant Radiant Warmer
                                    • Slide 34
                                    • Slide 35
                                    • Certifications and Approvals
                                    • Contact Us

                                      Skin Mode Operation

                                      Note The skin temperature probe must be properly attached before choosing Skin mode operation

                                      1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

                                      1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

                                      APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

                                      Operating Modes(contrsquodhellip)

                                      Operating Modes(contrsquodhellip)

                                      Pre-Warm Mode

                                      bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                                      Safe Mode

                                      bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                                      Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                                      AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                                      Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                                      Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                                      Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                                      Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                                      Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                                      Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                                      Heater Module

                                      Parabolic design

                                      Even Distribution of heat over bed surface

                                      Extensive safety

                                      measures(6 Layer safety for thermal

                                      cut- off)

                                      Far IR heating ensures

                                      stimulation of cells for growth

                                      The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                      Bed Platform Operation

                                      bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                      bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                      Side Panel Operation

                                      bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                      X-ray Procedure

                                      bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                      bullThe heater housing rotates to the left or right for X-ray procedures

                                      bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                      bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                      Comparison of nice 2010 Infant Radiant Warmers

                                      nice 2010 BC nice 2010 B nice 2010

                                      Skin Mode Y Y Y

                                      Manual Mode Y Y Y

                                      Safe Mode Y Y Y

                                      Pre-warm Mode Y Y Y

                                      X Ray Facility Y Y N

                                      Bed Y Y N

                                      Pivoting Warmer ModuleY Y Y

                                      Apgar Timer Y Y Y

                                      Height Adjustment N N Y

                                      Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                      How nice Infant Radiant Warmers fare over competitors

                                      nice 2010 series warmers

                                      User-friendly easy to access all parameters

                                      Long life quartz heater has a long life compared to other heaters

                                      6 Layer safety for the protection of infant from burning

                                      6 Layer Safety Features

                                      Alarms if infant temperature greater or less than Set Temperature

                                      Thermal cut off if infant temperature exceeds 39 degC

                                      Heater output reduces to 30 after 15 mins in Manual mode

                                      Heater output reduces to 30 if electronics senses no variable feedback

                                      Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                      Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                      TROUBLESHOOTING

                                      AND MAINTENANCE

                                      Cleaning Procedure

                                      1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                      2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                      3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                      Cleaning Solutions for Mattress

                                      Generic Formulations Maximum concentration level

                                      Hydrogen Peroxide 6

                                      Quaternary Compound 05 aqueous solution

                                      Quaternary Ammonium 02

                                      Lodophors 50mglitre

                                      Preventive Maintenance

                                      1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                      Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                      1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                      TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                      Sl No Problem Cause Remedy

                                      1 Power failure LED is ON with audio alarm

                                      Power FailureCheck the unit is Plugged in to Main supply

                                      Check the mains are switched ON

                                      Else Contact nice Neotech

                                      2System Failure LED is ON with audio alarm

                                      System Failure Contact nice Neotech

                                      General System Failure

                                      System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                      1 No Display on the Control PanelCheck AC Supply Voltage

                                      Check if Power Cord is disconnected

                                      Check Fuse in the Rear side of Pillar

                                      Contact nice Neotech

                                      2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                      Set Temperature may be less than Actual Temperature

                                      Check Heater Connection

                                      Contact nice Neotech

                                      3 Equipment Working But Observation Lamp amp Heater output not working

                                      Check Link cord connected between Control Module toHeater Module

                                      Check Fuse in the Rear side of Heater Module

                                      Check Transformer Voltage - 12V AC

                                      Replace Lamp

                                      4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                      Check Mute ON

                                      Contact nice Neotech

                                      5 Patient Temperature Display shows ldquoS-Ordquo

                                      with Continuous Alarm(S-O Sensor Open)

                                      Check if skin Probe is Disconnected from theEquipment

                                      If skin probe is defective replace new probe

                                      Contact nice Neotech

                                      7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                      Check Heater Connection

                                      Contact nice Neotech

                                      8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                      Check Environment Temperature above 39deg C

                                      Check Infant Fever

                                      9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                      Temperature

                                      Check Environment Temperature is 1degC above SetTemperature

                                      Contact nice Neotech

                                      10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                      Temperature

                                      Probe may be wet

                                      Contact nice Neotech

                                      13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                      Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                      Low Supply Voltage

                                      14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                      Percentage Heater Output is Set at High in ManualMode

                                      15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                      Check for Strong Air conditioning draft

                                      16Temperature Display does not display InfantsTemperature accurately

                                      Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                      17Baby skin Temperature not correlate with display Temperature

                                      Check sensitive portion of the Probe is fixed with babyskin properly

                                      Check Probe is fixed to the lower abdomen of baby

                                      18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                      19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                      Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                      System Fault Infant Radiant Warmer(Contrsquodhellip)

                                      Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                      Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                      components Do not autoclave or gas sterilize the skin temperature probe Do not

                                      immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                      Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                      should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                      Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                      and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                      equipment performance

                                      Certifications and Approvals

                                      Electrical Safety IEC 60601-1 IEC 60601-1-4

                                      Particular product IEC 60601-2-21

                                      EMC Safety IEC 60601-1-2

                                      Contact Us

                                      nice Neotech Medical Systems Private Limited

                                      No 85 Krishna Industrial Estate

                                      Mettukuppam Vannagaram

                                      Chennai ndash 600095 Tamil Nadu

                                      INDIA

                                      Tel +91-44-2476259424764608

                                      Fax +91-44-24766920

                                      Web wwwniceneotechcom

                                      E-mail infoniceneotechcom marketingniceneotechcom

                                      Overseas Enquiry overseasniceneotechcom

                                      Toll Free +91-1800-425-2594

                                      • nice NeOumltech Medical Systems Pvt Ltd
                                      • Presentation Flow
                                      • Company INFO
                                      • Timeline-nice Neotech
                                      • Why nice Neotech
                                      • Why Infant Radiant Warmers are Used
                                      • MODES OF HEAT LOSS
                                      • Slide 8
                                      • Effects of Far IR in Infant body
                                      • Types of heaters
                                      • Slide 11
                                      • Benefits of Heaterhellip
                                      • Slide 13
                                      • Slide 14
                                      • Slide 15
                                      • How to operate
                                      • Control Panel Operation
                                      • Operating Modes
                                      • Slide 19
                                      • Slide 20
                                      • Alarms
                                      • Heater Module
                                      • Bed Platform Operation
                                      • Side Panel Operation
                                      • X-ray Procedure
                                      • Comparison of nice 2010 Infant Radiant Warmers
                                      • How nice Infant Radiant Warmers fare over competitors
                                      • Slide 28
                                      • Slide 29
                                      • Cleaning Procedure
                                      • Preventive Maintenance
                                      • Troubleshooting
                                      • System Fault Infant Radiant Warmer
                                      • Slide 34
                                      • Slide 35
                                      • Certifications and Approvals
                                      • Contact Us

                                        Operating Modes(contrsquodhellip)

                                        Pre-Warm Mode

                                        bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

                                        Safe Mode

                                        bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

                                        Note All the above mentioned modes can only be accessible when the keypad lock is off condition

                                        AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                                        Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                                        Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                                        Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                                        Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                                        Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                                        Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                                        Heater Module

                                        Parabolic design

                                        Even Distribution of heat over bed surface

                                        Extensive safety

                                        measures(6 Layer safety for thermal

                                        cut- off)

                                        Far IR heating ensures

                                        stimulation of cells for growth

                                        The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                        Bed Platform Operation

                                        bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                        bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                        Side Panel Operation

                                        bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                        X-ray Procedure

                                        bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                        bullThe heater housing rotates to the left or right for X-ray procedures

                                        bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                        bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                        Comparison of nice 2010 Infant Radiant Warmers

                                        nice 2010 BC nice 2010 B nice 2010

                                        Skin Mode Y Y Y

                                        Manual Mode Y Y Y

                                        Safe Mode Y Y Y

                                        Pre-warm Mode Y Y Y

                                        X Ray Facility Y Y N

                                        Bed Y Y N

                                        Pivoting Warmer ModuleY Y Y

                                        Apgar Timer Y Y Y

                                        Height Adjustment N N Y

                                        Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                        How nice Infant Radiant Warmers fare over competitors

                                        nice 2010 series warmers

                                        User-friendly easy to access all parameters

                                        Long life quartz heater has a long life compared to other heaters

                                        6 Layer safety for the protection of infant from burning

                                        6 Layer Safety Features

                                        Alarms if infant temperature greater or less than Set Temperature

                                        Thermal cut off if infant temperature exceeds 39 degC

                                        Heater output reduces to 30 after 15 mins in Manual mode

                                        Heater output reduces to 30 if electronics senses no variable feedback

                                        Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                        Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                        TROUBLESHOOTING

                                        AND MAINTENANCE

                                        Cleaning Procedure

                                        1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                        2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                        3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                        Cleaning Solutions for Mattress

                                        Generic Formulations Maximum concentration level

                                        Hydrogen Peroxide 6

                                        Quaternary Compound 05 aqueous solution

                                        Quaternary Ammonium 02

                                        Lodophors 50mglitre

                                        Preventive Maintenance

                                        1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                        Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                        1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                        TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                        Sl No Problem Cause Remedy

                                        1 Power failure LED is ON with audio alarm

                                        Power FailureCheck the unit is Plugged in to Main supply

                                        Check the mains are switched ON

                                        Else Contact nice Neotech

                                        2System Failure LED is ON with audio alarm

                                        System Failure Contact nice Neotech

                                        General System Failure

                                        System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                        1 No Display on the Control PanelCheck AC Supply Voltage

                                        Check if Power Cord is disconnected

                                        Check Fuse in the Rear side of Pillar

                                        Contact nice Neotech

                                        2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                        Set Temperature may be less than Actual Temperature

                                        Check Heater Connection

                                        Contact nice Neotech

                                        3 Equipment Working But Observation Lamp amp Heater output not working

                                        Check Link cord connected between Control Module toHeater Module

                                        Check Fuse in the Rear side of Heater Module

                                        Check Transformer Voltage - 12V AC

                                        Replace Lamp

                                        4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                        Check Mute ON

                                        Contact nice Neotech

                                        5 Patient Temperature Display shows ldquoS-Ordquo

                                        with Continuous Alarm(S-O Sensor Open)

                                        Check if skin Probe is Disconnected from theEquipment

                                        If skin probe is defective replace new probe

                                        Contact nice Neotech

                                        7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                        Check Heater Connection

                                        Contact nice Neotech

                                        8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                        Check Environment Temperature above 39deg C

                                        Check Infant Fever

                                        9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                        Temperature

                                        Check Environment Temperature is 1degC above SetTemperature

                                        Contact nice Neotech

                                        10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                        Temperature

                                        Probe may be wet

                                        Contact nice Neotech

                                        13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                        Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                        Low Supply Voltage

                                        14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                        Percentage Heater Output is Set at High in ManualMode

                                        15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                        Check for Strong Air conditioning draft

                                        16Temperature Display does not display InfantsTemperature accurately

                                        Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                        17Baby skin Temperature not correlate with display Temperature

                                        Check sensitive portion of the Probe is fixed with babyskin properly

                                        Check Probe is fixed to the lower abdomen of baby

                                        18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                        19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                        Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                        System Fault Infant Radiant Warmer(Contrsquodhellip)

                                        Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                        Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                        components Do not autoclave or gas sterilize the skin temperature probe Do not

                                        immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                        Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                        should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                        Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                        and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                        equipment performance

                                        Certifications and Approvals

                                        Electrical Safety IEC 60601-1 IEC 60601-1-4

                                        Particular product IEC 60601-2-21

                                        EMC Safety IEC 60601-1-2

                                        Contact Us

                                        nice Neotech Medical Systems Private Limited

                                        No 85 Krishna Industrial Estate

                                        Mettukuppam Vannagaram

                                        Chennai ndash 600095 Tamil Nadu

                                        INDIA

                                        Tel +91-44-2476259424764608

                                        Fax +91-44-24766920

                                        Web wwwniceneotechcom

                                        E-mail infoniceneotechcom marketingniceneotechcom

                                        Overseas Enquiry overseasniceneotechcom

                                        Toll Free +91-1800-425-2594

                                        • nice NeOumltech Medical Systems Pvt Ltd
                                        • Presentation Flow
                                        • Company INFO
                                        • Timeline-nice Neotech
                                        • Why nice Neotech
                                        • Why Infant Radiant Warmers are Used
                                        • MODES OF HEAT LOSS
                                        • Slide 8
                                        • Effects of Far IR in Infant body
                                        • Types of heaters
                                        • Slide 11
                                        • Benefits of Heaterhellip
                                        • Slide 13
                                        • Slide 14
                                        • Slide 15
                                        • How to operate
                                        • Control Panel Operation
                                        • Operating Modes
                                        • Slide 19
                                        • Slide 20
                                        • Alarms
                                        • Heater Module
                                        • Bed Platform Operation
                                        • Side Panel Operation
                                        • X-ray Procedure
                                        • Comparison of nice 2010 Infant Radiant Warmers
                                        • How nice Infant Radiant Warmers fare over competitors
                                        • Slide 28
                                        • Slide 29
                                        • Cleaning Procedure
                                        • Preventive Maintenance
                                        • Troubleshooting
                                        • System Fault Infant Radiant Warmer
                                        • Slide 34
                                        • Slide 35
                                        • Certifications and Approvals
                                        • Contact Us

                                          AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

                                          Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

                                          Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

                                          Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

                                          Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

                                          Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

                                          Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

                                          Heater Module

                                          Parabolic design

                                          Even Distribution of heat over bed surface

                                          Extensive safety

                                          measures(6 Layer safety for thermal

                                          cut- off)

                                          Far IR heating ensures

                                          stimulation of cells for growth

                                          The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                          Bed Platform Operation

                                          bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                          bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                          Side Panel Operation

                                          bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                          X-ray Procedure

                                          bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                          bullThe heater housing rotates to the left or right for X-ray procedures

                                          bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                          bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                          Comparison of nice 2010 Infant Radiant Warmers

                                          nice 2010 BC nice 2010 B nice 2010

                                          Skin Mode Y Y Y

                                          Manual Mode Y Y Y

                                          Safe Mode Y Y Y

                                          Pre-warm Mode Y Y Y

                                          X Ray Facility Y Y N

                                          Bed Y Y N

                                          Pivoting Warmer ModuleY Y Y

                                          Apgar Timer Y Y Y

                                          Height Adjustment N N Y

                                          Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                          How nice Infant Radiant Warmers fare over competitors

                                          nice 2010 series warmers

                                          User-friendly easy to access all parameters

                                          Long life quartz heater has a long life compared to other heaters

                                          6 Layer safety for the protection of infant from burning

                                          6 Layer Safety Features

                                          Alarms if infant temperature greater or less than Set Temperature

                                          Thermal cut off if infant temperature exceeds 39 degC

                                          Heater output reduces to 30 after 15 mins in Manual mode

                                          Heater output reduces to 30 if electronics senses no variable feedback

                                          Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                          Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                          TROUBLESHOOTING

                                          AND MAINTENANCE

                                          Cleaning Procedure

                                          1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                          2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                          3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                          Cleaning Solutions for Mattress

                                          Generic Formulations Maximum concentration level

                                          Hydrogen Peroxide 6

                                          Quaternary Compound 05 aqueous solution

                                          Quaternary Ammonium 02

                                          Lodophors 50mglitre

                                          Preventive Maintenance

                                          1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                          Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                          1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                          TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                          Sl No Problem Cause Remedy

                                          1 Power failure LED is ON with audio alarm

                                          Power FailureCheck the unit is Plugged in to Main supply

                                          Check the mains are switched ON

                                          Else Contact nice Neotech

                                          2System Failure LED is ON with audio alarm

                                          System Failure Contact nice Neotech

                                          General System Failure

                                          System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                          1 No Display on the Control PanelCheck AC Supply Voltage

                                          Check if Power Cord is disconnected

                                          Check Fuse in the Rear side of Pillar

                                          Contact nice Neotech

                                          2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                          Set Temperature may be less than Actual Temperature

                                          Check Heater Connection

                                          Contact nice Neotech

                                          3 Equipment Working But Observation Lamp amp Heater output not working

                                          Check Link cord connected between Control Module toHeater Module

                                          Check Fuse in the Rear side of Heater Module

                                          Check Transformer Voltage - 12V AC

                                          Replace Lamp

                                          4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                          Check Mute ON

                                          Contact nice Neotech

                                          5 Patient Temperature Display shows ldquoS-Ordquo

                                          with Continuous Alarm(S-O Sensor Open)

                                          Check if skin Probe is Disconnected from theEquipment

                                          If skin probe is defective replace new probe

                                          Contact nice Neotech

                                          7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                          Check Heater Connection

                                          Contact nice Neotech

                                          8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                          Check Environment Temperature above 39deg C

                                          Check Infant Fever

                                          9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                          Temperature

                                          Check Environment Temperature is 1degC above SetTemperature

                                          Contact nice Neotech

                                          10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                          Temperature

                                          Probe may be wet

                                          Contact nice Neotech

                                          13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                          Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                          Low Supply Voltage

                                          14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                          Percentage Heater Output is Set at High in ManualMode

                                          15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                          Check for Strong Air conditioning draft

                                          16Temperature Display does not display InfantsTemperature accurately

                                          Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                          17Baby skin Temperature not correlate with display Temperature

                                          Check sensitive portion of the Probe is fixed with babyskin properly

                                          Check Probe is fixed to the lower abdomen of baby

                                          18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                          19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                          Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                          System Fault Infant Radiant Warmer(Contrsquodhellip)

                                          Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                          Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                          components Do not autoclave or gas sterilize the skin temperature probe Do not

                                          immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                          Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                          should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                          Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                          and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                          equipment performance

                                          Certifications and Approvals

                                          Electrical Safety IEC 60601-1 IEC 60601-1-4

                                          Particular product IEC 60601-2-21

                                          EMC Safety IEC 60601-1-2

                                          Contact Us

                                          nice Neotech Medical Systems Private Limited

                                          No 85 Krishna Industrial Estate

                                          Mettukuppam Vannagaram

                                          Chennai ndash 600095 Tamil Nadu

                                          INDIA

                                          Tel +91-44-2476259424764608

                                          Fax +91-44-24766920

                                          Web wwwniceneotechcom

                                          E-mail infoniceneotechcom marketingniceneotechcom

                                          Overseas Enquiry overseasniceneotechcom

                                          Toll Free +91-1800-425-2594

                                          • nice NeOumltech Medical Systems Pvt Ltd
                                          • Presentation Flow
                                          • Company INFO
                                          • Timeline-nice Neotech
                                          • Why nice Neotech
                                          • Why Infant Radiant Warmers are Used
                                          • MODES OF HEAT LOSS
                                          • Slide 8
                                          • Effects of Far IR in Infant body
                                          • Types of heaters
                                          • Slide 11
                                          • Benefits of Heaterhellip
                                          • Slide 13
                                          • Slide 14
                                          • Slide 15
                                          • How to operate
                                          • Control Panel Operation
                                          • Operating Modes
                                          • Slide 19
                                          • Slide 20
                                          • Alarms
                                          • Heater Module
                                          • Bed Platform Operation
                                          • Side Panel Operation
                                          • X-ray Procedure
                                          • Comparison of nice 2010 Infant Radiant Warmers
                                          • How nice Infant Radiant Warmers fare over competitors
                                          • Slide 28
                                          • Slide 29
                                          • Cleaning Procedure
                                          • Preventive Maintenance
                                          • Troubleshooting
                                          • System Fault Infant Radiant Warmer
                                          • Slide 34
                                          • Slide 35
                                          • Certifications and Approvals
                                          • Contact Us

                                            Heater Module

                                            Parabolic design

                                            Even Distribution of heat over bed surface

                                            Extensive safety

                                            measures(6 Layer safety for thermal

                                            cut- off)

                                            Far IR heating ensures

                                            stimulation of cells for growth

                                            The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

                                            Bed Platform Operation

                                            bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                            bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                            Side Panel Operation

                                            bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                            X-ray Procedure

                                            bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                            bullThe heater housing rotates to the left or right for X-ray procedures

                                            bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                            bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                            Comparison of nice 2010 Infant Radiant Warmers

                                            nice 2010 BC nice 2010 B nice 2010

                                            Skin Mode Y Y Y

                                            Manual Mode Y Y Y

                                            Safe Mode Y Y Y

                                            Pre-warm Mode Y Y Y

                                            X Ray Facility Y Y N

                                            Bed Y Y N

                                            Pivoting Warmer ModuleY Y Y

                                            Apgar Timer Y Y Y

                                            Height Adjustment N N Y

                                            Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                            How nice Infant Radiant Warmers fare over competitors

                                            nice 2010 series warmers

                                            User-friendly easy to access all parameters

                                            Long life quartz heater has a long life compared to other heaters

                                            6 Layer safety for the protection of infant from burning

                                            6 Layer Safety Features

                                            Alarms if infant temperature greater or less than Set Temperature

                                            Thermal cut off if infant temperature exceeds 39 degC

                                            Heater output reduces to 30 after 15 mins in Manual mode

                                            Heater output reduces to 30 if electronics senses no variable feedback

                                            Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                            Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                            TROUBLESHOOTING

                                            AND MAINTENANCE

                                            Cleaning Procedure

                                            1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                            2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                            3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                            Cleaning Solutions for Mattress

                                            Generic Formulations Maximum concentration level

                                            Hydrogen Peroxide 6

                                            Quaternary Compound 05 aqueous solution

                                            Quaternary Ammonium 02

                                            Lodophors 50mglitre

                                            Preventive Maintenance

                                            1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                            Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                            1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                            TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                            Sl No Problem Cause Remedy

                                            1 Power failure LED is ON with audio alarm

                                            Power FailureCheck the unit is Plugged in to Main supply

                                            Check the mains are switched ON

                                            Else Contact nice Neotech

                                            2System Failure LED is ON with audio alarm

                                            System Failure Contact nice Neotech

                                            General System Failure

                                            System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                            1 No Display on the Control PanelCheck AC Supply Voltage

                                            Check if Power Cord is disconnected

                                            Check Fuse in the Rear side of Pillar

                                            Contact nice Neotech

                                            2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                            Set Temperature may be less than Actual Temperature

                                            Check Heater Connection

                                            Contact nice Neotech

                                            3 Equipment Working But Observation Lamp amp Heater output not working

                                            Check Link cord connected between Control Module toHeater Module

                                            Check Fuse in the Rear side of Heater Module

                                            Check Transformer Voltage - 12V AC

                                            Replace Lamp

                                            4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                            Check Mute ON

                                            Contact nice Neotech

                                            5 Patient Temperature Display shows ldquoS-Ordquo

                                            with Continuous Alarm(S-O Sensor Open)

                                            Check if skin Probe is Disconnected from theEquipment

                                            If skin probe is defective replace new probe

                                            Contact nice Neotech

                                            7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                            Check Heater Connection

                                            Contact nice Neotech

                                            8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                            Check Environment Temperature above 39deg C

                                            Check Infant Fever

                                            9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                            Temperature

                                            Check Environment Temperature is 1degC above SetTemperature

                                            Contact nice Neotech

                                            10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                            Temperature

                                            Probe may be wet

                                            Contact nice Neotech

                                            13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                            Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                            Low Supply Voltage

                                            14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                            Percentage Heater Output is Set at High in ManualMode

                                            15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                            Check for Strong Air conditioning draft

                                            16Temperature Display does not display InfantsTemperature accurately

                                            Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                            17Baby skin Temperature not correlate with display Temperature

                                            Check sensitive portion of the Probe is fixed with babyskin properly

                                            Check Probe is fixed to the lower abdomen of baby

                                            18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                            19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                            Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                            System Fault Infant Radiant Warmer(Contrsquodhellip)

                                            Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                            Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                            components Do not autoclave or gas sterilize the skin temperature probe Do not

                                            immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                            Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                            should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                            Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                            and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                            equipment performance

                                            Certifications and Approvals

                                            Electrical Safety IEC 60601-1 IEC 60601-1-4

                                            Particular product IEC 60601-2-21

                                            EMC Safety IEC 60601-1-2

                                            Contact Us

                                            nice Neotech Medical Systems Private Limited

                                            No 85 Krishna Industrial Estate

                                            Mettukuppam Vannagaram

                                            Chennai ndash 600095 Tamil Nadu

                                            INDIA

                                            Tel +91-44-2476259424764608

                                            Fax +91-44-24766920

                                            Web wwwniceneotechcom

                                            E-mail infoniceneotechcom marketingniceneotechcom

                                            Overseas Enquiry overseasniceneotechcom

                                            Toll Free +91-1800-425-2594

                                            • nice NeOumltech Medical Systems Pvt Ltd
                                            • Presentation Flow
                                            • Company INFO
                                            • Timeline-nice Neotech
                                            • Why nice Neotech
                                            • Why Infant Radiant Warmers are Used
                                            • MODES OF HEAT LOSS
                                            • Slide 8
                                            • Effects of Far IR in Infant body
                                            • Types of heaters
                                            • Slide 11
                                            • Benefits of Heaterhellip
                                            • Slide 13
                                            • Slide 14
                                            • Slide 15
                                            • How to operate
                                            • Control Panel Operation
                                            • Operating Modes
                                            • Slide 19
                                            • Slide 20
                                            • Alarms
                                            • Heater Module
                                            • Bed Platform Operation
                                            • Side Panel Operation
                                            • X-ray Procedure
                                            • Comparison of nice 2010 Infant Radiant Warmers
                                            • How nice Infant Radiant Warmers fare over competitors
                                            • Slide 28
                                            • Slide 29
                                            • Cleaning Procedure
                                            • Preventive Maintenance
                                            • Troubleshooting
                                            • System Fault Infant Radiant Warmer
                                            • Slide 34
                                            • Slide 35
                                            • Certifications and Approvals
                                            • Contact Us

                                              Bed Platform Operation

                                              bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

                                              bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

                                              Side Panel Operation

                                              bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                              X-ray Procedure

                                              bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                              bullThe heater housing rotates to the left or right for X-ray procedures

                                              bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                              bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                              Comparison of nice 2010 Infant Radiant Warmers

                                              nice 2010 BC nice 2010 B nice 2010

                                              Skin Mode Y Y Y

                                              Manual Mode Y Y Y

                                              Safe Mode Y Y Y

                                              Pre-warm Mode Y Y Y

                                              X Ray Facility Y Y N

                                              Bed Y Y N

                                              Pivoting Warmer ModuleY Y Y

                                              Apgar Timer Y Y Y

                                              Height Adjustment N N Y

                                              Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                              How nice Infant Radiant Warmers fare over competitors

                                              nice 2010 series warmers

                                              User-friendly easy to access all parameters

                                              Long life quartz heater has a long life compared to other heaters

                                              6 Layer safety for the protection of infant from burning

                                              6 Layer Safety Features

                                              Alarms if infant temperature greater or less than Set Temperature

                                              Thermal cut off if infant temperature exceeds 39 degC

                                              Heater output reduces to 30 after 15 mins in Manual mode

                                              Heater output reduces to 30 if electronics senses no variable feedback

                                              Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                              Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                              TROUBLESHOOTING

                                              AND MAINTENANCE

                                              Cleaning Procedure

                                              1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                              2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                              3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                              Cleaning Solutions for Mattress

                                              Generic Formulations Maximum concentration level

                                              Hydrogen Peroxide 6

                                              Quaternary Compound 05 aqueous solution

                                              Quaternary Ammonium 02

                                              Lodophors 50mglitre

                                              Preventive Maintenance

                                              1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                              Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                              1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                              TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                              Sl No Problem Cause Remedy

                                              1 Power failure LED is ON with audio alarm

                                              Power FailureCheck the unit is Plugged in to Main supply

                                              Check the mains are switched ON

                                              Else Contact nice Neotech

                                              2System Failure LED is ON with audio alarm

                                              System Failure Contact nice Neotech

                                              General System Failure

                                              System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                              1 No Display on the Control PanelCheck AC Supply Voltage

                                              Check if Power Cord is disconnected

                                              Check Fuse in the Rear side of Pillar

                                              Contact nice Neotech

                                              2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                              Set Temperature may be less than Actual Temperature

                                              Check Heater Connection

                                              Contact nice Neotech

                                              3 Equipment Working But Observation Lamp amp Heater output not working

                                              Check Link cord connected between Control Module toHeater Module

                                              Check Fuse in the Rear side of Heater Module

                                              Check Transformer Voltage - 12V AC

                                              Replace Lamp

                                              4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                              Check Mute ON

                                              Contact nice Neotech

                                              5 Patient Temperature Display shows ldquoS-Ordquo

                                              with Continuous Alarm(S-O Sensor Open)

                                              Check if skin Probe is Disconnected from theEquipment

                                              If skin probe is defective replace new probe

                                              Contact nice Neotech

                                              7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                              Check Heater Connection

                                              Contact nice Neotech

                                              8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                              Check Environment Temperature above 39deg C

                                              Check Infant Fever

                                              9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                              Temperature

                                              Check Environment Temperature is 1degC above SetTemperature

                                              Contact nice Neotech

                                              10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                              Temperature

                                              Probe may be wet

                                              Contact nice Neotech

                                              13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                              Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                              Low Supply Voltage

                                              14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                              Percentage Heater Output is Set at High in ManualMode

                                              15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                              Check for Strong Air conditioning draft

                                              16Temperature Display does not display InfantsTemperature accurately

                                              Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                              17Baby skin Temperature not correlate with display Temperature

                                              Check sensitive portion of the Probe is fixed with babyskin properly

                                              Check Probe is fixed to the lower abdomen of baby

                                              18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                              19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                              Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                              System Fault Infant Radiant Warmer(Contrsquodhellip)

                                              Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                              Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                              components Do not autoclave or gas sterilize the skin temperature probe Do not

                                              immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                              Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                              should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                              Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                              and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                              equipment performance

                                              Certifications and Approvals

                                              Electrical Safety IEC 60601-1 IEC 60601-1-4

                                              Particular product IEC 60601-2-21

                                              EMC Safety IEC 60601-1-2

                                              Contact Us

                                              nice Neotech Medical Systems Private Limited

                                              No 85 Krishna Industrial Estate

                                              Mettukuppam Vannagaram

                                              Chennai ndash 600095 Tamil Nadu

                                              INDIA

                                              Tel +91-44-2476259424764608

                                              Fax +91-44-24766920

                                              Web wwwniceneotechcom

                                              E-mail infoniceneotechcom marketingniceneotechcom

                                              Overseas Enquiry overseasniceneotechcom

                                              Toll Free +91-1800-425-2594

                                              • nice NeOumltech Medical Systems Pvt Ltd
                                              • Presentation Flow
                                              • Company INFO
                                              • Timeline-nice Neotech
                                              • Why nice Neotech
                                              • Why Infant Radiant Warmers are Used
                                              • MODES OF HEAT LOSS
                                              • Slide 8
                                              • Effects of Far IR in Infant body
                                              • Types of heaters
                                              • Slide 11
                                              • Benefits of Heaterhellip
                                              • Slide 13
                                              • Slide 14
                                              • Slide 15
                                              • How to operate
                                              • Control Panel Operation
                                              • Operating Modes
                                              • Slide 19
                                              • Slide 20
                                              • Alarms
                                              • Heater Module
                                              • Bed Platform Operation
                                              • Side Panel Operation
                                              • X-ray Procedure
                                              • Comparison of nice 2010 Infant Radiant Warmers
                                              • How nice Infant Radiant Warmers fare over competitors
                                              • Slide 28
                                              • Slide 29
                                              • Cleaning Procedure
                                              • Preventive Maintenance
                                              • Troubleshooting
                                              • System Fault Infant Radiant Warmer
                                              • Slide 34
                                              • Slide 35
                                              • Certifications and Approvals
                                              • Contact Us

                                                Side Panel Operation

                                                bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

                                                X-ray Procedure

                                                bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                                bullThe heater housing rotates to the left or right for X-ray procedures

                                                bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                                bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                                Comparison of nice 2010 Infant Radiant Warmers

                                                nice 2010 BC nice 2010 B nice 2010

                                                Skin Mode Y Y Y

                                                Manual Mode Y Y Y

                                                Safe Mode Y Y Y

                                                Pre-warm Mode Y Y Y

                                                X Ray Facility Y Y N

                                                Bed Y Y N

                                                Pivoting Warmer ModuleY Y Y

                                                Apgar Timer Y Y Y

                                                Height Adjustment N N Y

                                                Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                                How nice Infant Radiant Warmers fare over competitors

                                                nice 2010 series warmers

                                                User-friendly easy to access all parameters

                                                Long life quartz heater has a long life compared to other heaters

                                                6 Layer safety for the protection of infant from burning

                                                6 Layer Safety Features

                                                Alarms if infant temperature greater or less than Set Temperature

                                                Thermal cut off if infant temperature exceeds 39 degC

                                                Heater output reduces to 30 after 15 mins in Manual mode

                                                Heater output reduces to 30 if electronics senses no variable feedback

                                                Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                                Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                                TROUBLESHOOTING

                                                AND MAINTENANCE

                                                Cleaning Procedure

                                                1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                Cleaning Solutions for Mattress

                                                Generic Formulations Maximum concentration level

                                                Hydrogen Peroxide 6

                                                Quaternary Compound 05 aqueous solution

                                                Quaternary Ammonium 02

                                                Lodophors 50mglitre

                                                Preventive Maintenance

                                                1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                Sl No Problem Cause Remedy

                                                1 Power failure LED is ON with audio alarm

                                                Power FailureCheck the unit is Plugged in to Main supply

                                                Check the mains are switched ON

                                                Else Contact nice Neotech

                                                2System Failure LED is ON with audio alarm

                                                System Failure Contact nice Neotech

                                                General System Failure

                                                System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                1 No Display on the Control PanelCheck AC Supply Voltage

                                                Check if Power Cord is disconnected

                                                Check Fuse in the Rear side of Pillar

                                                Contact nice Neotech

                                                2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                Set Temperature may be less than Actual Temperature

                                                Check Heater Connection

                                                Contact nice Neotech

                                                3 Equipment Working But Observation Lamp amp Heater output not working

                                                Check Link cord connected between Control Module toHeater Module

                                                Check Fuse in the Rear side of Heater Module

                                                Check Transformer Voltage - 12V AC

                                                Replace Lamp

                                                4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                Check Mute ON

                                                Contact nice Neotech

                                                5 Patient Temperature Display shows ldquoS-Ordquo

                                                with Continuous Alarm(S-O Sensor Open)

                                                Check if skin Probe is Disconnected from theEquipment

                                                If skin probe is defective replace new probe

                                                Contact nice Neotech

                                                7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                Check Heater Connection

                                                Contact nice Neotech

                                                8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                Check Environment Temperature above 39deg C

                                                Check Infant Fever

                                                9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                Temperature

                                                Check Environment Temperature is 1degC above SetTemperature

                                                Contact nice Neotech

                                                10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                Temperature

                                                Probe may be wet

                                                Contact nice Neotech

                                                13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                Low Supply Voltage

                                                14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                Percentage Heater Output is Set at High in ManualMode

                                                15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                Check for Strong Air conditioning draft

                                                16Temperature Display does not display InfantsTemperature accurately

                                                Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                17Baby skin Temperature not correlate with display Temperature

                                                Check sensitive portion of the Probe is fixed with babyskin properly

                                                Check Probe is fixed to the lower abdomen of baby

                                                18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                equipment performance

                                                Certifications and Approvals

                                                Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                Particular product IEC 60601-2-21

                                                EMC Safety IEC 60601-1-2

                                                Contact Us

                                                nice Neotech Medical Systems Private Limited

                                                No 85 Krishna Industrial Estate

                                                Mettukuppam Vannagaram

                                                Chennai ndash 600095 Tamil Nadu

                                                INDIA

                                                Tel +91-44-2476259424764608

                                                Fax +91-44-24766920

                                                Web wwwniceneotechcom

                                                E-mail infoniceneotechcom marketingniceneotechcom

                                                Overseas Enquiry overseasniceneotechcom

                                                Toll Free +91-1800-425-2594

                                                • nice NeOumltech Medical Systems Pvt Ltd
                                                • Presentation Flow
                                                • Company INFO
                                                • Timeline-nice Neotech
                                                • Why nice Neotech
                                                • Why Infant Radiant Warmers are Used
                                                • MODES OF HEAT LOSS
                                                • Slide 8
                                                • Effects of Far IR in Infant body
                                                • Types of heaters
                                                • Slide 11
                                                • Benefits of Heaterhellip
                                                • Slide 13
                                                • Slide 14
                                                • Slide 15
                                                • How to operate
                                                • Control Panel Operation
                                                • Operating Modes
                                                • Slide 19
                                                • Slide 20
                                                • Alarms
                                                • Heater Module
                                                • Bed Platform Operation
                                                • Side Panel Operation
                                                • X-ray Procedure
                                                • Comparison of nice 2010 Infant Radiant Warmers
                                                • How nice Infant Radiant Warmers fare over competitors
                                                • Slide 28
                                                • Slide 29
                                                • Cleaning Procedure
                                                • Preventive Maintenance
                                                • Troubleshooting
                                                • System Fault Infant Radiant Warmer
                                                • Slide 34
                                                • Slide 35
                                                • Certifications and Approvals
                                                • Contact Us

                                                  X-ray Procedure

                                                  bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

                                                  bullThe heater housing rotates to the left or right for X-ray procedures

                                                  bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

                                                  bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

                                                  Comparison of nice 2010 Infant Radiant Warmers

                                                  nice 2010 BC nice 2010 B nice 2010

                                                  Skin Mode Y Y Y

                                                  Manual Mode Y Y Y

                                                  Safe Mode Y Y Y

                                                  Pre-warm Mode Y Y Y

                                                  X Ray Facility Y Y N

                                                  Bed Y Y N

                                                  Pivoting Warmer ModuleY Y Y

                                                  Apgar Timer Y Y Y

                                                  Height Adjustment N N Y

                                                  Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                                  How nice Infant Radiant Warmers fare over competitors

                                                  nice 2010 series warmers

                                                  User-friendly easy to access all parameters

                                                  Long life quartz heater has a long life compared to other heaters

                                                  6 Layer safety for the protection of infant from burning

                                                  6 Layer Safety Features

                                                  Alarms if infant temperature greater or less than Set Temperature

                                                  Thermal cut off if infant temperature exceeds 39 degC

                                                  Heater output reduces to 30 after 15 mins in Manual mode

                                                  Heater output reduces to 30 if electronics senses no variable feedback

                                                  Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                                  Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                                  TROUBLESHOOTING

                                                  AND MAINTENANCE

                                                  Cleaning Procedure

                                                  1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                  2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                  3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                  Cleaning Solutions for Mattress

                                                  Generic Formulations Maximum concentration level

                                                  Hydrogen Peroxide 6

                                                  Quaternary Compound 05 aqueous solution

                                                  Quaternary Ammonium 02

                                                  Lodophors 50mglitre

                                                  Preventive Maintenance

                                                  1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                  Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                  1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                  TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                  Sl No Problem Cause Remedy

                                                  1 Power failure LED is ON with audio alarm

                                                  Power FailureCheck the unit is Plugged in to Main supply

                                                  Check the mains are switched ON

                                                  Else Contact nice Neotech

                                                  2System Failure LED is ON with audio alarm

                                                  System Failure Contact nice Neotech

                                                  General System Failure

                                                  System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                  1 No Display on the Control PanelCheck AC Supply Voltage

                                                  Check if Power Cord is disconnected

                                                  Check Fuse in the Rear side of Pillar

                                                  Contact nice Neotech

                                                  2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                  Set Temperature may be less than Actual Temperature

                                                  Check Heater Connection

                                                  Contact nice Neotech

                                                  3 Equipment Working But Observation Lamp amp Heater output not working

                                                  Check Link cord connected between Control Module toHeater Module

                                                  Check Fuse in the Rear side of Heater Module

                                                  Check Transformer Voltage - 12V AC

                                                  Replace Lamp

                                                  4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                  Check Mute ON

                                                  Contact nice Neotech

                                                  5 Patient Temperature Display shows ldquoS-Ordquo

                                                  with Continuous Alarm(S-O Sensor Open)

                                                  Check if skin Probe is Disconnected from theEquipment

                                                  If skin probe is defective replace new probe

                                                  Contact nice Neotech

                                                  7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                  Check Heater Connection

                                                  Contact nice Neotech

                                                  8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                  Check Environment Temperature above 39deg C

                                                  Check Infant Fever

                                                  9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                  Temperature

                                                  Check Environment Temperature is 1degC above SetTemperature

                                                  Contact nice Neotech

                                                  10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                  Temperature

                                                  Probe may be wet

                                                  Contact nice Neotech

                                                  13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                  Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                  Low Supply Voltage

                                                  14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                  Percentage Heater Output is Set at High in ManualMode

                                                  15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                  Check for Strong Air conditioning draft

                                                  16Temperature Display does not display InfantsTemperature accurately

                                                  Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                  17Baby skin Temperature not correlate with display Temperature

                                                  Check sensitive portion of the Probe is fixed with babyskin properly

                                                  Check Probe is fixed to the lower abdomen of baby

                                                  18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                  19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                  Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                  System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                  Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                  Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                  components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                  immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                  Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                  should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                  Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                  and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                  equipment performance

                                                  Certifications and Approvals

                                                  Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                  Particular product IEC 60601-2-21

                                                  EMC Safety IEC 60601-1-2

                                                  Contact Us

                                                  nice Neotech Medical Systems Private Limited

                                                  No 85 Krishna Industrial Estate

                                                  Mettukuppam Vannagaram

                                                  Chennai ndash 600095 Tamil Nadu

                                                  INDIA

                                                  Tel +91-44-2476259424764608

                                                  Fax +91-44-24766920

                                                  Web wwwniceneotechcom

                                                  E-mail infoniceneotechcom marketingniceneotechcom

                                                  Overseas Enquiry overseasniceneotechcom

                                                  Toll Free +91-1800-425-2594

                                                  • nice NeOumltech Medical Systems Pvt Ltd
                                                  • Presentation Flow
                                                  • Company INFO
                                                  • Timeline-nice Neotech
                                                  • Why nice Neotech
                                                  • Why Infant Radiant Warmers are Used
                                                  • MODES OF HEAT LOSS
                                                  • Slide 8
                                                  • Effects of Far IR in Infant body
                                                  • Types of heaters
                                                  • Slide 11
                                                  • Benefits of Heaterhellip
                                                  • Slide 13
                                                  • Slide 14
                                                  • Slide 15
                                                  • How to operate
                                                  • Control Panel Operation
                                                  • Operating Modes
                                                  • Slide 19
                                                  • Slide 20
                                                  • Alarms
                                                  • Heater Module
                                                  • Bed Platform Operation
                                                  • Side Panel Operation
                                                  • X-ray Procedure
                                                  • Comparison of nice 2010 Infant Radiant Warmers
                                                  • How nice Infant Radiant Warmers fare over competitors
                                                  • Slide 28
                                                  • Slide 29
                                                  • Cleaning Procedure
                                                  • Preventive Maintenance
                                                  • Troubleshooting
                                                  • System Fault Infant Radiant Warmer
                                                  • Slide 34
                                                  • Slide 35
                                                  • Certifications and Approvals
                                                  • Contact Us

                                                    Comparison of nice 2010 Infant Radiant Warmers

                                                    nice 2010 BC nice 2010 B nice 2010

                                                    Skin Mode Y Y Y

                                                    Manual Mode Y Y Y

                                                    Safe Mode Y Y Y

                                                    Pre-warm Mode Y Y Y

                                                    X Ray Facility Y Y N

                                                    Bed Y Y N

                                                    Pivoting Warmer ModuleY Y Y

                                                    Apgar Timer Y Y Y

                                                    Height Adjustment N N Y

                                                    Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

                                                    How nice Infant Radiant Warmers fare over competitors

                                                    nice 2010 series warmers

                                                    User-friendly easy to access all parameters

                                                    Long life quartz heater has a long life compared to other heaters

                                                    6 Layer safety for the protection of infant from burning

                                                    6 Layer Safety Features

                                                    Alarms if infant temperature greater or less than Set Temperature

                                                    Thermal cut off if infant temperature exceeds 39 degC

                                                    Heater output reduces to 30 after 15 mins in Manual mode

                                                    Heater output reduces to 30 if electronics senses no variable feedback

                                                    Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                                    Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                                    TROUBLESHOOTING

                                                    AND MAINTENANCE

                                                    Cleaning Procedure

                                                    1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                    2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                    3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                    Cleaning Solutions for Mattress

                                                    Generic Formulations Maximum concentration level

                                                    Hydrogen Peroxide 6

                                                    Quaternary Compound 05 aqueous solution

                                                    Quaternary Ammonium 02

                                                    Lodophors 50mglitre

                                                    Preventive Maintenance

                                                    1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                    Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                    1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                    TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                    Sl No Problem Cause Remedy

                                                    1 Power failure LED is ON with audio alarm

                                                    Power FailureCheck the unit is Plugged in to Main supply

                                                    Check the mains are switched ON

                                                    Else Contact nice Neotech

                                                    2System Failure LED is ON with audio alarm

                                                    System Failure Contact nice Neotech

                                                    General System Failure

                                                    System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                    1 No Display on the Control PanelCheck AC Supply Voltage

                                                    Check if Power Cord is disconnected

                                                    Check Fuse in the Rear side of Pillar

                                                    Contact nice Neotech

                                                    2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                    Set Temperature may be less than Actual Temperature

                                                    Check Heater Connection

                                                    Contact nice Neotech

                                                    3 Equipment Working But Observation Lamp amp Heater output not working

                                                    Check Link cord connected between Control Module toHeater Module

                                                    Check Fuse in the Rear side of Heater Module

                                                    Check Transformer Voltage - 12V AC

                                                    Replace Lamp

                                                    4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                    Check Mute ON

                                                    Contact nice Neotech

                                                    5 Patient Temperature Display shows ldquoS-Ordquo

                                                    with Continuous Alarm(S-O Sensor Open)

                                                    Check if skin Probe is Disconnected from theEquipment

                                                    If skin probe is defective replace new probe

                                                    Contact nice Neotech

                                                    7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                    Check Heater Connection

                                                    Contact nice Neotech

                                                    8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                    Check Environment Temperature above 39deg C

                                                    Check Infant Fever

                                                    9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                    Temperature

                                                    Check Environment Temperature is 1degC above SetTemperature

                                                    Contact nice Neotech

                                                    10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                    Temperature

                                                    Probe may be wet

                                                    Contact nice Neotech

                                                    13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                    Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                    Low Supply Voltage

                                                    14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                    Percentage Heater Output is Set at High in ManualMode

                                                    15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                    Check for Strong Air conditioning draft

                                                    16Temperature Display does not display InfantsTemperature accurately

                                                    Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                    17Baby skin Temperature not correlate with display Temperature

                                                    Check sensitive portion of the Probe is fixed with babyskin properly

                                                    Check Probe is fixed to the lower abdomen of baby

                                                    18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                    19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                    Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                    System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                    Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                    Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                    components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                    immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                    Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                    should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                    Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                    and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                    equipment performance

                                                    Certifications and Approvals

                                                    Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                    Particular product IEC 60601-2-21

                                                    EMC Safety IEC 60601-1-2

                                                    Contact Us

                                                    nice Neotech Medical Systems Private Limited

                                                    No 85 Krishna Industrial Estate

                                                    Mettukuppam Vannagaram

                                                    Chennai ndash 600095 Tamil Nadu

                                                    INDIA

                                                    Tel +91-44-2476259424764608

                                                    Fax +91-44-24766920

                                                    Web wwwniceneotechcom

                                                    E-mail infoniceneotechcom marketingniceneotechcom

                                                    Overseas Enquiry overseasniceneotechcom

                                                    Toll Free +91-1800-425-2594

                                                    • nice NeOumltech Medical Systems Pvt Ltd
                                                    • Presentation Flow
                                                    • Company INFO
                                                    • Timeline-nice Neotech
                                                    • Why nice Neotech
                                                    • Why Infant Radiant Warmers are Used
                                                    • MODES OF HEAT LOSS
                                                    • Slide 8
                                                    • Effects of Far IR in Infant body
                                                    • Types of heaters
                                                    • Slide 11
                                                    • Benefits of Heaterhellip
                                                    • Slide 13
                                                    • Slide 14
                                                    • Slide 15
                                                    • How to operate
                                                    • Control Panel Operation
                                                    • Operating Modes
                                                    • Slide 19
                                                    • Slide 20
                                                    • Alarms
                                                    • Heater Module
                                                    • Bed Platform Operation
                                                    • Side Panel Operation
                                                    • X-ray Procedure
                                                    • Comparison of nice 2010 Infant Radiant Warmers
                                                    • How nice Infant Radiant Warmers fare over competitors
                                                    • Slide 28
                                                    • Slide 29
                                                    • Cleaning Procedure
                                                    • Preventive Maintenance
                                                    • Troubleshooting
                                                    • System Fault Infant Radiant Warmer
                                                    • Slide 34
                                                    • Slide 35
                                                    • Certifications and Approvals
                                                    • Contact Us

                                                      How nice Infant Radiant Warmers fare over competitors

                                                      nice 2010 series warmers

                                                      User-friendly easy to access all parameters

                                                      Long life quartz heater has a long life compared to other heaters

                                                      6 Layer safety for the protection of infant from burning

                                                      6 Layer Safety Features

                                                      Alarms if infant temperature greater or less than Set Temperature

                                                      Thermal cut off if infant temperature exceeds 39 degC

                                                      Heater output reduces to 30 after 15 mins in Manual mode

                                                      Heater output reduces to 30 if electronics senses no variable feedback

                                                      Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                                      Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                                      TROUBLESHOOTING

                                                      AND MAINTENANCE

                                                      Cleaning Procedure

                                                      1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                      2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                      3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                      Cleaning Solutions for Mattress

                                                      Generic Formulations Maximum concentration level

                                                      Hydrogen Peroxide 6

                                                      Quaternary Compound 05 aqueous solution

                                                      Quaternary Ammonium 02

                                                      Lodophors 50mglitre

                                                      Preventive Maintenance

                                                      1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                      Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                      1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                      TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                      Sl No Problem Cause Remedy

                                                      1 Power failure LED is ON with audio alarm

                                                      Power FailureCheck the unit is Plugged in to Main supply

                                                      Check the mains are switched ON

                                                      Else Contact nice Neotech

                                                      2System Failure LED is ON with audio alarm

                                                      System Failure Contact nice Neotech

                                                      General System Failure

                                                      System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                      1 No Display on the Control PanelCheck AC Supply Voltage

                                                      Check if Power Cord is disconnected

                                                      Check Fuse in the Rear side of Pillar

                                                      Contact nice Neotech

                                                      2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                      Set Temperature may be less than Actual Temperature

                                                      Check Heater Connection

                                                      Contact nice Neotech

                                                      3 Equipment Working But Observation Lamp amp Heater output not working

                                                      Check Link cord connected between Control Module toHeater Module

                                                      Check Fuse in the Rear side of Heater Module

                                                      Check Transformer Voltage - 12V AC

                                                      Replace Lamp

                                                      4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                      Check Mute ON

                                                      Contact nice Neotech

                                                      5 Patient Temperature Display shows ldquoS-Ordquo

                                                      with Continuous Alarm(S-O Sensor Open)

                                                      Check if skin Probe is Disconnected from theEquipment

                                                      If skin probe is defective replace new probe

                                                      Contact nice Neotech

                                                      7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                      Check Heater Connection

                                                      Contact nice Neotech

                                                      8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                      Check Environment Temperature above 39deg C

                                                      Check Infant Fever

                                                      9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                      Temperature

                                                      Check Environment Temperature is 1degC above SetTemperature

                                                      Contact nice Neotech

                                                      10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                      Temperature

                                                      Probe may be wet

                                                      Contact nice Neotech

                                                      13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                      Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                      Low Supply Voltage

                                                      14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                      Percentage Heater Output is Set at High in ManualMode

                                                      15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                      Check for Strong Air conditioning draft

                                                      16Temperature Display does not display InfantsTemperature accurately

                                                      Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                      17Baby skin Temperature not correlate with display Temperature

                                                      Check sensitive portion of the Probe is fixed with babyskin properly

                                                      Check Probe is fixed to the lower abdomen of baby

                                                      18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                      19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                      Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                      System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                      Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                      Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                      components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                      immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                      Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                      should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                      Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                      and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                      equipment performance

                                                      Certifications and Approvals

                                                      Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                      Particular product IEC 60601-2-21

                                                      EMC Safety IEC 60601-1-2

                                                      Contact Us

                                                      nice Neotech Medical Systems Private Limited

                                                      No 85 Krishna Industrial Estate

                                                      Mettukuppam Vannagaram

                                                      Chennai ndash 600095 Tamil Nadu

                                                      INDIA

                                                      Tel +91-44-2476259424764608

                                                      Fax +91-44-24766920

                                                      Web wwwniceneotechcom

                                                      E-mail infoniceneotechcom marketingniceneotechcom

                                                      Overseas Enquiry overseasniceneotechcom

                                                      Toll Free +91-1800-425-2594

                                                      • nice NeOumltech Medical Systems Pvt Ltd
                                                      • Presentation Flow
                                                      • Company INFO
                                                      • Timeline-nice Neotech
                                                      • Why nice Neotech
                                                      • Why Infant Radiant Warmers are Used
                                                      • MODES OF HEAT LOSS
                                                      • Slide 8
                                                      • Effects of Far IR in Infant body
                                                      • Types of heaters
                                                      • Slide 11
                                                      • Benefits of Heaterhellip
                                                      • Slide 13
                                                      • Slide 14
                                                      • Slide 15
                                                      • How to operate
                                                      • Control Panel Operation
                                                      • Operating Modes
                                                      • Slide 19
                                                      • Slide 20
                                                      • Alarms
                                                      • Heater Module
                                                      • Bed Platform Operation
                                                      • Side Panel Operation
                                                      • X-ray Procedure
                                                      • Comparison of nice 2010 Infant Radiant Warmers
                                                      • How nice Infant Radiant Warmers fare over competitors
                                                      • Slide 28
                                                      • Slide 29
                                                      • Cleaning Procedure
                                                      • Preventive Maintenance
                                                      • Troubleshooting
                                                      • System Fault Infant Radiant Warmer
                                                      • Slide 34
                                                      • Slide 35
                                                      • Certifications and Approvals
                                                      • Contact Us

                                                        6 Layer Safety Features

                                                        Alarms if infant temperature greater or less than Set Temperature

                                                        Thermal cut off if infant temperature exceeds 39 degC

                                                        Heater output reduces to 30 after 15 mins in Manual mode

                                                        Heater output reduces to 30 if electronics senses no variable feedback

                                                        Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

                                                        Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

                                                        TROUBLESHOOTING

                                                        AND MAINTENANCE

                                                        Cleaning Procedure

                                                        1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                        2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                        3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                        Cleaning Solutions for Mattress

                                                        Generic Formulations Maximum concentration level

                                                        Hydrogen Peroxide 6

                                                        Quaternary Compound 05 aqueous solution

                                                        Quaternary Ammonium 02

                                                        Lodophors 50mglitre

                                                        Preventive Maintenance

                                                        1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                        Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                        1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                        TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                        Sl No Problem Cause Remedy

                                                        1 Power failure LED is ON with audio alarm

                                                        Power FailureCheck the unit is Plugged in to Main supply

                                                        Check the mains are switched ON

                                                        Else Contact nice Neotech

                                                        2System Failure LED is ON with audio alarm

                                                        System Failure Contact nice Neotech

                                                        General System Failure

                                                        System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                        1 No Display on the Control PanelCheck AC Supply Voltage

                                                        Check if Power Cord is disconnected

                                                        Check Fuse in the Rear side of Pillar

                                                        Contact nice Neotech

                                                        2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                        Set Temperature may be less than Actual Temperature

                                                        Check Heater Connection

                                                        Contact nice Neotech

                                                        3 Equipment Working But Observation Lamp amp Heater output not working

                                                        Check Link cord connected between Control Module toHeater Module

                                                        Check Fuse in the Rear side of Heater Module

                                                        Check Transformer Voltage - 12V AC

                                                        Replace Lamp

                                                        4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                        Check Mute ON

                                                        Contact nice Neotech

                                                        5 Patient Temperature Display shows ldquoS-Ordquo

                                                        with Continuous Alarm(S-O Sensor Open)

                                                        Check if skin Probe is Disconnected from theEquipment

                                                        If skin probe is defective replace new probe

                                                        Contact nice Neotech

                                                        7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                        Check Heater Connection

                                                        Contact nice Neotech

                                                        8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                        Check Environment Temperature above 39deg C

                                                        Check Infant Fever

                                                        9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                        Temperature

                                                        Check Environment Temperature is 1degC above SetTemperature

                                                        Contact nice Neotech

                                                        10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                        Temperature

                                                        Probe may be wet

                                                        Contact nice Neotech

                                                        13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                        Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                        Low Supply Voltage

                                                        14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                        Percentage Heater Output is Set at High in ManualMode

                                                        15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                        Check for Strong Air conditioning draft

                                                        16Temperature Display does not display InfantsTemperature accurately

                                                        Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                        17Baby skin Temperature not correlate with display Temperature

                                                        Check sensitive portion of the Probe is fixed with babyskin properly

                                                        Check Probe is fixed to the lower abdomen of baby

                                                        18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                        19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                        Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                        System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                        Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                        Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                        components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                        immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                        Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                        should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                        Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                        and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                        equipment performance

                                                        Certifications and Approvals

                                                        Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                        Particular product IEC 60601-2-21

                                                        EMC Safety IEC 60601-1-2

                                                        Contact Us

                                                        nice Neotech Medical Systems Private Limited

                                                        No 85 Krishna Industrial Estate

                                                        Mettukuppam Vannagaram

                                                        Chennai ndash 600095 Tamil Nadu

                                                        INDIA

                                                        Tel +91-44-2476259424764608

                                                        Fax +91-44-24766920

                                                        Web wwwniceneotechcom

                                                        E-mail infoniceneotechcom marketingniceneotechcom

                                                        Overseas Enquiry overseasniceneotechcom

                                                        Toll Free +91-1800-425-2594

                                                        • nice NeOumltech Medical Systems Pvt Ltd
                                                        • Presentation Flow
                                                        • Company INFO
                                                        • Timeline-nice Neotech
                                                        • Why nice Neotech
                                                        • Why Infant Radiant Warmers are Used
                                                        • MODES OF HEAT LOSS
                                                        • Slide 8
                                                        • Effects of Far IR in Infant body
                                                        • Types of heaters
                                                        • Slide 11
                                                        • Benefits of Heaterhellip
                                                        • Slide 13
                                                        • Slide 14
                                                        • Slide 15
                                                        • How to operate
                                                        • Control Panel Operation
                                                        • Operating Modes
                                                        • Slide 19
                                                        • Slide 20
                                                        • Alarms
                                                        • Heater Module
                                                        • Bed Platform Operation
                                                        • Side Panel Operation
                                                        • X-ray Procedure
                                                        • Comparison of nice 2010 Infant Radiant Warmers
                                                        • How nice Infant Radiant Warmers fare over competitors
                                                        • Slide 28
                                                        • Slide 29
                                                        • Cleaning Procedure
                                                        • Preventive Maintenance
                                                        • Troubleshooting
                                                        • System Fault Infant Radiant Warmer
                                                        • Slide 34
                                                        • Slide 35
                                                        • Certifications and Approvals
                                                        • Contact Us

                                                          TROUBLESHOOTING

                                                          AND MAINTENANCE

                                                          Cleaning Procedure

                                                          1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                          2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                          3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                          Cleaning Solutions for Mattress

                                                          Generic Formulations Maximum concentration level

                                                          Hydrogen Peroxide 6

                                                          Quaternary Compound 05 aqueous solution

                                                          Quaternary Ammonium 02

                                                          Lodophors 50mglitre

                                                          Preventive Maintenance

                                                          1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                          Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                          1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                          TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                          Sl No Problem Cause Remedy

                                                          1 Power failure LED is ON with audio alarm

                                                          Power FailureCheck the unit is Plugged in to Main supply

                                                          Check the mains are switched ON

                                                          Else Contact nice Neotech

                                                          2System Failure LED is ON with audio alarm

                                                          System Failure Contact nice Neotech

                                                          General System Failure

                                                          System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                          1 No Display on the Control PanelCheck AC Supply Voltage

                                                          Check if Power Cord is disconnected

                                                          Check Fuse in the Rear side of Pillar

                                                          Contact nice Neotech

                                                          2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                          Set Temperature may be less than Actual Temperature

                                                          Check Heater Connection

                                                          Contact nice Neotech

                                                          3 Equipment Working But Observation Lamp amp Heater output not working

                                                          Check Link cord connected between Control Module toHeater Module

                                                          Check Fuse in the Rear side of Heater Module

                                                          Check Transformer Voltage - 12V AC

                                                          Replace Lamp

                                                          4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                          Check Mute ON

                                                          Contact nice Neotech

                                                          5 Patient Temperature Display shows ldquoS-Ordquo

                                                          with Continuous Alarm(S-O Sensor Open)

                                                          Check if skin Probe is Disconnected from theEquipment

                                                          If skin probe is defective replace new probe

                                                          Contact nice Neotech

                                                          7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                          Check Heater Connection

                                                          Contact nice Neotech

                                                          8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                          Check Environment Temperature above 39deg C

                                                          Check Infant Fever

                                                          9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                          Temperature

                                                          Check Environment Temperature is 1degC above SetTemperature

                                                          Contact nice Neotech

                                                          10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                          Temperature

                                                          Probe may be wet

                                                          Contact nice Neotech

                                                          13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                          Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                          Low Supply Voltage

                                                          14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                          Percentage Heater Output is Set at High in ManualMode

                                                          15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                          Check for Strong Air conditioning draft

                                                          16Temperature Display does not display InfantsTemperature accurately

                                                          Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                          17Baby skin Temperature not correlate with display Temperature

                                                          Check sensitive portion of the Probe is fixed with babyskin properly

                                                          Check Probe is fixed to the lower abdomen of baby

                                                          18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                          19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                          Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                          System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                          Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                          Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                          components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                          immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                          Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                          should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                          Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                          and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                          equipment performance

                                                          Certifications and Approvals

                                                          Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                          Particular product IEC 60601-2-21

                                                          EMC Safety IEC 60601-1-2

                                                          Contact Us

                                                          nice Neotech Medical Systems Private Limited

                                                          No 85 Krishna Industrial Estate

                                                          Mettukuppam Vannagaram

                                                          Chennai ndash 600095 Tamil Nadu

                                                          INDIA

                                                          Tel +91-44-2476259424764608

                                                          Fax +91-44-24766920

                                                          Web wwwniceneotechcom

                                                          E-mail infoniceneotechcom marketingniceneotechcom

                                                          Overseas Enquiry overseasniceneotechcom

                                                          Toll Free +91-1800-425-2594

                                                          • nice NeOumltech Medical Systems Pvt Ltd
                                                          • Presentation Flow
                                                          • Company INFO
                                                          • Timeline-nice Neotech
                                                          • Why nice Neotech
                                                          • Why Infant Radiant Warmers are Used
                                                          • MODES OF HEAT LOSS
                                                          • Slide 8
                                                          • Effects of Far IR in Infant body
                                                          • Types of heaters
                                                          • Slide 11
                                                          • Benefits of Heaterhellip
                                                          • Slide 13
                                                          • Slide 14
                                                          • Slide 15
                                                          • How to operate
                                                          • Control Panel Operation
                                                          • Operating Modes
                                                          • Slide 19
                                                          • Slide 20
                                                          • Alarms
                                                          • Heater Module
                                                          • Bed Platform Operation
                                                          • Side Panel Operation
                                                          • X-ray Procedure
                                                          • Comparison of nice 2010 Infant Radiant Warmers
                                                          • How nice Infant Radiant Warmers fare over competitors
                                                          • Slide 28
                                                          • Slide 29
                                                          • Cleaning Procedure
                                                          • Preventive Maintenance
                                                          • Troubleshooting
                                                          • System Fault Infant Radiant Warmer
                                                          • Slide 34
                                                          • Slide 35
                                                          • Certifications and Approvals
                                                          • Contact Us

                                                            Cleaning Procedure

                                                            1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

                                                            2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

                                                            3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

                                                            Cleaning Solutions for Mattress

                                                            Generic Formulations Maximum concentration level

                                                            Hydrogen Peroxide 6

                                                            Quaternary Compound 05 aqueous solution

                                                            Quaternary Ammonium 02

                                                            Lodophors 50mglitre

                                                            Preventive Maintenance

                                                            1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                            Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                            1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                            TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                            Sl No Problem Cause Remedy

                                                            1 Power failure LED is ON with audio alarm

                                                            Power FailureCheck the unit is Plugged in to Main supply

                                                            Check the mains are switched ON

                                                            Else Contact nice Neotech

                                                            2System Failure LED is ON with audio alarm

                                                            System Failure Contact nice Neotech

                                                            General System Failure

                                                            System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                            1 No Display on the Control PanelCheck AC Supply Voltage

                                                            Check if Power Cord is disconnected

                                                            Check Fuse in the Rear side of Pillar

                                                            Contact nice Neotech

                                                            2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                            Set Temperature may be less than Actual Temperature

                                                            Check Heater Connection

                                                            Contact nice Neotech

                                                            3 Equipment Working But Observation Lamp amp Heater output not working

                                                            Check Link cord connected between Control Module toHeater Module

                                                            Check Fuse in the Rear side of Heater Module

                                                            Check Transformer Voltage - 12V AC

                                                            Replace Lamp

                                                            4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                            Check Mute ON

                                                            Contact nice Neotech

                                                            5 Patient Temperature Display shows ldquoS-Ordquo

                                                            with Continuous Alarm(S-O Sensor Open)

                                                            Check if skin Probe is Disconnected from theEquipment

                                                            If skin probe is defective replace new probe

                                                            Contact nice Neotech

                                                            7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                            Check Heater Connection

                                                            Contact nice Neotech

                                                            8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                            Check Environment Temperature above 39deg C

                                                            Check Infant Fever

                                                            9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                            Temperature

                                                            Check Environment Temperature is 1degC above SetTemperature

                                                            Contact nice Neotech

                                                            10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                            Temperature

                                                            Probe may be wet

                                                            Contact nice Neotech

                                                            13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                            Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                            Low Supply Voltage

                                                            14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                            Percentage Heater Output is Set at High in ManualMode

                                                            15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                            Check for Strong Air conditioning draft

                                                            16Temperature Display does not display InfantsTemperature accurately

                                                            Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                            17Baby skin Temperature not correlate with display Temperature

                                                            Check sensitive portion of the Probe is fixed with babyskin properly

                                                            Check Probe is fixed to the lower abdomen of baby

                                                            18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                            19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                            Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                            System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                            Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                            Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                            components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                            immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                            Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                            should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                            Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                            and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                            equipment performance

                                                            Certifications and Approvals

                                                            Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                            Particular product IEC 60601-2-21

                                                            EMC Safety IEC 60601-1-2

                                                            Contact Us

                                                            nice Neotech Medical Systems Private Limited

                                                            No 85 Krishna Industrial Estate

                                                            Mettukuppam Vannagaram

                                                            Chennai ndash 600095 Tamil Nadu

                                                            INDIA

                                                            Tel +91-44-2476259424764608

                                                            Fax +91-44-24766920

                                                            Web wwwniceneotechcom

                                                            E-mail infoniceneotechcom marketingniceneotechcom

                                                            Overseas Enquiry overseasniceneotechcom

                                                            Toll Free +91-1800-425-2594

                                                            • nice NeOumltech Medical Systems Pvt Ltd
                                                            • Presentation Flow
                                                            • Company INFO
                                                            • Timeline-nice Neotech
                                                            • Why nice Neotech
                                                            • Why Infant Radiant Warmers are Used
                                                            • MODES OF HEAT LOSS
                                                            • Slide 8
                                                            • Effects of Far IR in Infant body
                                                            • Types of heaters
                                                            • Slide 11
                                                            • Benefits of Heaterhellip
                                                            • Slide 13
                                                            • Slide 14
                                                            • Slide 15
                                                            • How to operate
                                                            • Control Panel Operation
                                                            • Operating Modes
                                                            • Slide 19
                                                            • Slide 20
                                                            • Alarms
                                                            • Heater Module
                                                            • Bed Platform Operation
                                                            • Side Panel Operation
                                                            • X-ray Procedure
                                                            • Comparison of nice 2010 Infant Radiant Warmers
                                                            • How nice Infant Radiant Warmers fare over competitors
                                                            • Slide 28
                                                            • Slide 29
                                                            • Cleaning Procedure
                                                            • Preventive Maintenance
                                                            • Troubleshooting
                                                            • System Fault Infant Radiant Warmer
                                                            • Slide 34
                                                            • Slide 35
                                                            • Certifications and Approvals
                                                            • Contact Us

                                                              Preventive Maintenance

                                                              1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

                                                              Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

                                                              1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

                                                              TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                              Sl No Problem Cause Remedy

                                                              1 Power failure LED is ON with audio alarm

                                                              Power FailureCheck the unit is Plugged in to Main supply

                                                              Check the mains are switched ON

                                                              Else Contact nice Neotech

                                                              2System Failure LED is ON with audio alarm

                                                              System Failure Contact nice Neotech

                                                              General System Failure

                                                              System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                              1 No Display on the Control PanelCheck AC Supply Voltage

                                                              Check if Power Cord is disconnected

                                                              Check Fuse in the Rear side of Pillar

                                                              Contact nice Neotech

                                                              2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                              Set Temperature may be less than Actual Temperature

                                                              Check Heater Connection

                                                              Contact nice Neotech

                                                              3 Equipment Working But Observation Lamp amp Heater output not working

                                                              Check Link cord connected between Control Module toHeater Module

                                                              Check Fuse in the Rear side of Heater Module

                                                              Check Transformer Voltage - 12V AC

                                                              Replace Lamp

                                                              4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                              Check Mute ON

                                                              Contact nice Neotech

                                                              5 Patient Temperature Display shows ldquoS-Ordquo

                                                              with Continuous Alarm(S-O Sensor Open)

                                                              Check if skin Probe is Disconnected from theEquipment

                                                              If skin probe is defective replace new probe

                                                              Contact nice Neotech

                                                              7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                              Check Heater Connection

                                                              Contact nice Neotech

                                                              8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                              Check Environment Temperature above 39deg C

                                                              Check Infant Fever

                                                              9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                              Temperature

                                                              Check Environment Temperature is 1degC above SetTemperature

                                                              Contact nice Neotech

                                                              10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                              Temperature

                                                              Probe may be wet

                                                              Contact nice Neotech

                                                              13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                              Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                              Low Supply Voltage

                                                              14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                              Percentage Heater Output is Set at High in ManualMode

                                                              15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                              Check for Strong Air conditioning draft

                                                              16Temperature Display does not display InfantsTemperature accurately

                                                              Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                              17Baby skin Temperature not correlate with display Temperature

                                                              Check sensitive portion of the Probe is fixed with babyskin properly

                                                              Check Probe is fixed to the lower abdomen of baby

                                                              18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                              19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                              Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                              System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                              Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                              Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                              components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                              immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                              Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                              should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                              Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                              and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                              equipment performance

                                                              Certifications and Approvals

                                                              Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                              Particular product IEC 60601-2-21

                                                              EMC Safety IEC 60601-1-2

                                                              Contact Us

                                                              nice Neotech Medical Systems Private Limited

                                                              No 85 Krishna Industrial Estate

                                                              Mettukuppam Vannagaram

                                                              Chennai ndash 600095 Tamil Nadu

                                                              INDIA

                                                              Tel +91-44-2476259424764608

                                                              Fax +91-44-24766920

                                                              Web wwwniceneotechcom

                                                              E-mail infoniceneotechcom marketingniceneotechcom

                                                              Overseas Enquiry overseasniceneotechcom

                                                              Toll Free +91-1800-425-2594

                                                              • nice NeOumltech Medical Systems Pvt Ltd
                                                              • Presentation Flow
                                                              • Company INFO
                                                              • Timeline-nice Neotech
                                                              • Why nice Neotech
                                                              • Why Infant Radiant Warmers are Used
                                                              • MODES OF HEAT LOSS
                                                              • Slide 8
                                                              • Effects of Far IR in Infant body
                                                              • Types of heaters
                                                              • Slide 11
                                                              • Benefits of Heaterhellip
                                                              • Slide 13
                                                              • Slide 14
                                                              • Slide 15
                                                              • How to operate
                                                              • Control Panel Operation
                                                              • Operating Modes
                                                              • Slide 19
                                                              • Slide 20
                                                              • Alarms
                                                              • Heater Module
                                                              • Bed Platform Operation
                                                              • Side Panel Operation
                                                              • X-ray Procedure
                                                              • Comparison of nice 2010 Infant Radiant Warmers
                                                              • How nice Infant Radiant Warmers fare over competitors
                                                              • Slide 28
                                                              • Slide 29
                                                              • Cleaning Procedure
                                                              • Preventive Maintenance
                                                              • Troubleshooting
                                                              • System Fault Infant Radiant Warmer
                                                              • Slide 34
                                                              • Slide 35
                                                              • Certifications and Approvals
                                                              • Contact Us

                                                                TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

                                                                Sl No Problem Cause Remedy

                                                                1 Power failure LED is ON with audio alarm

                                                                Power FailureCheck the unit is Plugged in to Main supply

                                                                Check the mains are switched ON

                                                                Else Contact nice Neotech

                                                                2System Failure LED is ON with audio alarm

                                                                System Failure Contact nice Neotech

                                                                General System Failure

                                                                System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                                1 No Display on the Control PanelCheck AC Supply Voltage

                                                                Check if Power Cord is disconnected

                                                                Check Fuse in the Rear side of Pillar

                                                                Contact nice Neotech

                                                                2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                                Set Temperature may be less than Actual Temperature

                                                                Check Heater Connection

                                                                Contact nice Neotech

                                                                3 Equipment Working But Observation Lamp amp Heater output not working

                                                                Check Link cord connected between Control Module toHeater Module

                                                                Check Fuse in the Rear side of Heater Module

                                                                Check Transformer Voltage - 12V AC

                                                                Replace Lamp

                                                                4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                                Check Mute ON

                                                                Contact nice Neotech

                                                                5 Patient Temperature Display shows ldquoS-Ordquo

                                                                with Continuous Alarm(S-O Sensor Open)

                                                                Check if skin Probe is Disconnected from theEquipment

                                                                If skin probe is defective replace new probe

                                                                Contact nice Neotech

                                                                7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                                Check Heater Connection

                                                                Contact nice Neotech

                                                                8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                                Check Environment Temperature above 39deg C

                                                                Check Infant Fever

                                                                9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                                Temperature

                                                                Check Environment Temperature is 1degC above SetTemperature

                                                                Contact nice Neotech

                                                                10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                                Temperature

                                                                Probe may be wet

                                                                Contact nice Neotech

                                                                13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                                Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                                Low Supply Voltage

                                                                14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                                Percentage Heater Output is Set at High in ManualMode

                                                                15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                                Check for Strong Air conditioning draft

                                                                16Temperature Display does not display InfantsTemperature accurately

                                                                Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                                17Baby skin Temperature not correlate with display Temperature

                                                                Check sensitive portion of the Probe is fixed with babyskin properly

                                                                Check Probe is fixed to the lower abdomen of baby

                                                                18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                                19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                                Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                                System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                                Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                                Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                                components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                                immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                                Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                                should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                                Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                                and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                                equipment performance

                                                                Certifications and Approvals

                                                                Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                                Particular product IEC 60601-2-21

                                                                EMC Safety IEC 60601-1-2

                                                                Contact Us

                                                                nice Neotech Medical Systems Private Limited

                                                                No 85 Krishna Industrial Estate

                                                                Mettukuppam Vannagaram

                                                                Chennai ndash 600095 Tamil Nadu

                                                                INDIA

                                                                Tel +91-44-2476259424764608

                                                                Fax +91-44-24766920

                                                                Web wwwniceneotechcom

                                                                E-mail infoniceneotechcom marketingniceneotechcom

                                                                Overseas Enquiry overseasniceneotechcom

                                                                Toll Free +91-1800-425-2594

                                                                • nice NeOumltech Medical Systems Pvt Ltd
                                                                • Presentation Flow
                                                                • Company INFO
                                                                • Timeline-nice Neotech
                                                                • Why nice Neotech
                                                                • Why Infant Radiant Warmers are Used
                                                                • MODES OF HEAT LOSS
                                                                • Slide 8
                                                                • Effects of Far IR in Infant body
                                                                • Types of heaters
                                                                • Slide 11
                                                                • Benefits of Heaterhellip
                                                                • Slide 13
                                                                • Slide 14
                                                                • Slide 15
                                                                • How to operate
                                                                • Control Panel Operation
                                                                • Operating Modes
                                                                • Slide 19
                                                                • Slide 20
                                                                • Alarms
                                                                • Heater Module
                                                                • Bed Platform Operation
                                                                • Side Panel Operation
                                                                • X-ray Procedure
                                                                • Comparison of nice 2010 Infant Radiant Warmers
                                                                • How nice Infant Radiant Warmers fare over competitors
                                                                • Slide 28
                                                                • Slide 29
                                                                • Cleaning Procedure
                                                                • Preventive Maintenance
                                                                • Troubleshooting
                                                                • System Fault Infant Radiant Warmer
                                                                • Slide 34
                                                                • Slide 35
                                                                • Certifications and Approvals
                                                                • Contact Us

                                                                  System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

                                                                  1 No Display on the Control PanelCheck AC Supply Voltage

                                                                  Check if Power Cord is disconnected

                                                                  Check Fuse in the Rear side of Pillar

                                                                  Contact nice Neotech

                                                                  2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

                                                                  Set Temperature may be less than Actual Temperature

                                                                  Check Heater Connection

                                                                  Contact nice Neotech

                                                                  3 Equipment Working But Observation Lamp amp Heater output not working

                                                                  Check Link cord connected between Control Module toHeater Module

                                                                  Check Fuse in the Rear side of Heater Module

                                                                  Check Transformer Voltage - 12V AC

                                                                  Replace Lamp

                                                                  4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

                                                                  Check Mute ON

                                                                  Contact nice Neotech

                                                                  5 Patient Temperature Display shows ldquoS-Ordquo

                                                                  with Continuous Alarm(S-O Sensor Open)

                                                                  Check if skin Probe is Disconnected from theEquipment

                                                                  If skin probe is defective replace new probe

                                                                  Contact nice Neotech

                                                                  7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

                                                                  Check Heater Connection

                                                                  Contact nice Neotech

                                                                  8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

                                                                  Check Environment Temperature above 39deg C

                                                                  Check Infant Fever

                                                                  9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

                                                                  Temperature

                                                                  Check Environment Temperature is 1degC above SetTemperature

                                                                  Contact nice Neotech

                                                                  10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

                                                                  Temperature

                                                                  Probe may be wet

                                                                  Contact nice Neotech

                                                                  13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                                  Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                                  Low Supply Voltage

                                                                  14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                                  Percentage Heater Output is Set at High in ManualMode

                                                                  15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                                  Check for Strong Air conditioning draft

                                                                  16Temperature Display does not display InfantsTemperature accurately

                                                                  Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                                  17Baby skin Temperature not correlate with display Temperature

                                                                  Check sensitive portion of the Probe is fixed with babyskin properly

                                                                  Check Probe is fixed to the lower abdomen of baby

                                                                  18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                                  19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                                  Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                                  System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                                  Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                                  Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                                  components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                                  immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                                  Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                                  should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                                  Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                                  and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                                  equipment performance

                                                                  Certifications and Approvals

                                                                  Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                                  Particular product IEC 60601-2-21

                                                                  EMC Safety IEC 60601-1-2

                                                                  Contact Us

                                                                  nice Neotech Medical Systems Private Limited

                                                                  No 85 Krishna Industrial Estate

                                                                  Mettukuppam Vannagaram

                                                                  Chennai ndash 600095 Tamil Nadu

                                                                  INDIA

                                                                  Tel +91-44-2476259424764608

                                                                  Fax +91-44-24766920

                                                                  Web wwwniceneotechcom

                                                                  E-mail infoniceneotechcom marketingniceneotechcom

                                                                  Overseas Enquiry overseasniceneotechcom

                                                                  Toll Free +91-1800-425-2594

                                                                  • nice NeOumltech Medical Systems Pvt Ltd
                                                                  • Presentation Flow
                                                                  • Company INFO
                                                                  • Timeline-nice Neotech
                                                                  • Why nice Neotech
                                                                  • Why Infant Radiant Warmers are Used
                                                                  • MODES OF HEAT LOSS
                                                                  • Slide 8
                                                                  • Effects of Far IR in Infant body
                                                                  • Types of heaters
                                                                  • Slide 11
                                                                  • Benefits of Heaterhellip
                                                                  • Slide 13
                                                                  • Slide 14
                                                                  • Slide 15
                                                                  • How to operate
                                                                  • Control Panel Operation
                                                                  • Operating Modes
                                                                  • Slide 19
                                                                  • Slide 20
                                                                  • Alarms
                                                                  • Heater Module
                                                                  • Bed Platform Operation
                                                                  • Side Panel Operation
                                                                  • X-ray Procedure
                                                                  • Comparison of nice 2010 Infant Radiant Warmers
                                                                  • How nice Infant Radiant Warmers fare over competitors
                                                                  • Slide 28
                                                                  • Slide 29
                                                                  • Cleaning Procedure
                                                                  • Preventive Maintenance
                                                                  • Troubleshooting
                                                                  • System Fault Infant Radiant Warmer
                                                                  • Slide 34
                                                                  • Slide 35
                                                                  • Certifications and Approvals
                                                                  • Contact Us

                                                                    Temperature

                                                                    Probe may be wet

                                                                    Contact nice Neotech

                                                                    13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

                                                                    Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

                                                                    Low Supply Voltage

                                                                    14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

                                                                    Percentage Heater Output is Set at High in ManualMode

                                                                    15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

                                                                    Check for Strong Air conditioning draft

                                                                    16Temperature Display does not display InfantsTemperature accurately

                                                                    Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

                                                                    17Baby skin Temperature not correlate with display Temperature

                                                                    Check sensitive portion of the Probe is fixed with babyskin properly

                                                                    Check Probe is fixed to the lower abdomen of baby

                                                                    18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

                                                                    19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

                                                                    Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

                                                                    System Fault Infant Radiant Warmer(Contrsquodhellip)

                                                                    Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                                    Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                                    components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                                    immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                                    Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                                    should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                                    Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                                    and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                                    equipment performance

                                                                    Certifications and Approvals

                                                                    Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                                    Particular product IEC 60601-2-21

                                                                    EMC Safety IEC 60601-1-2

                                                                    Contact Us

                                                                    nice Neotech Medical Systems Private Limited

                                                                    No 85 Krishna Industrial Estate

                                                                    Mettukuppam Vannagaram

                                                                    Chennai ndash 600095 Tamil Nadu

                                                                    INDIA

                                                                    Tel +91-44-2476259424764608

                                                                    Fax +91-44-24766920

                                                                    Web wwwniceneotechcom

                                                                    E-mail infoniceneotechcom marketingniceneotechcom

                                                                    Overseas Enquiry overseasniceneotechcom

                                                                    Toll Free +91-1800-425-2594

                                                                    • nice NeOumltech Medical Systems Pvt Ltd
                                                                    • Presentation Flow
                                                                    • Company INFO
                                                                    • Timeline-nice Neotech
                                                                    • Why nice Neotech
                                                                    • Why Infant Radiant Warmers are Used
                                                                    • MODES OF HEAT LOSS
                                                                    • Slide 8
                                                                    • Effects of Far IR in Infant body
                                                                    • Types of heaters
                                                                    • Slide 11
                                                                    • Benefits of Heaterhellip
                                                                    • Slide 13
                                                                    • Slide 14
                                                                    • Slide 15
                                                                    • How to operate
                                                                    • Control Panel Operation
                                                                    • Operating Modes
                                                                    • Slide 19
                                                                    • Slide 20
                                                                    • Alarms
                                                                    • Heater Module
                                                                    • Bed Platform Operation
                                                                    • Side Panel Operation
                                                                    • X-ray Procedure
                                                                    • Comparison of nice 2010 Infant Radiant Warmers
                                                                    • How nice Infant Radiant Warmers fare over competitors
                                                                    • Slide 28
                                                                    • Slide 29
                                                                    • Cleaning Procedure
                                                                    • Preventive Maintenance
                                                                    • Troubleshooting
                                                                    • System Fault Infant Radiant Warmer
                                                                    • Slide 34
                                                                    • Slide 35
                                                                    • Certifications and Approvals
                                                                    • Contact Us

                                                                      Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

                                                                      Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

                                                                      components Do not autoclave or gas sterilize the skin temperature probe Do not

                                                                      immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

                                                                      Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

                                                                      should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

                                                                      Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

                                                                      and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

                                                                      equipment performance

                                                                      Certifications and Approvals

                                                                      Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                                      Particular product IEC 60601-2-21

                                                                      EMC Safety IEC 60601-1-2

                                                                      Contact Us

                                                                      nice Neotech Medical Systems Private Limited

                                                                      No 85 Krishna Industrial Estate

                                                                      Mettukuppam Vannagaram

                                                                      Chennai ndash 600095 Tamil Nadu

                                                                      INDIA

                                                                      Tel +91-44-2476259424764608

                                                                      Fax +91-44-24766920

                                                                      Web wwwniceneotechcom

                                                                      E-mail infoniceneotechcom marketingniceneotechcom

                                                                      Overseas Enquiry overseasniceneotechcom

                                                                      Toll Free +91-1800-425-2594

                                                                      • nice NeOumltech Medical Systems Pvt Ltd
                                                                      • Presentation Flow
                                                                      • Company INFO
                                                                      • Timeline-nice Neotech
                                                                      • Why nice Neotech
                                                                      • Why Infant Radiant Warmers are Used
                                                                      • MODES OF HEAT LOSS
                                                                      • Slide 8
                                                                      • Effects of Far IR in Infant body
                                                                      • Types of heaters
                                                                      • Slide 11
                                                                      • Benefits of Heaterhellip
                                                                      • Slide 13
                                                                      • Slide 14
                                                                      • Slide 15
                                                                      • How to operate
                                                                      • Control Panel Operation
                                                                      • Operating Modes
                                                                      • Slide 19
                                                                      • Slide 20
                                                                      • Alarms
                                                                      • Heater Module
                                                                      • Bed Platform Operation
                                                                      • Side Panel Operation
                                                                      • X-ray Procedure
                                                                      • Comparison of nice 2010 Infant Radiant Warmers
                                                                      • How nice Infant Radiant Warmers fare over competitors
                                                                      • Slide 28
                                                                      • Slide 29
                                                                      • Cleaning Procedure
                                                                      • Preventive Maintenance
                                                                      • Troubleshooting
                                                                      • System Fault Infant Radiant Warmer
                                                                      • Slide 34
                                                                      • Slide 35
                                                                      • Certifications and Approvals
                                                                      • Contact Us

                                                                        Certifications and Approvals

                                                                        Electrical Safety IEC 60601-1 IEC 60601-1-4

                                                                        Particular product IEC 60601-2-21

                                                                        EMC Safety IEC 60601-1-2

                                                                        Contact Us

                                                                        nice Neotech Medical Systems Private Limited

                                                                        No 85 Krishna Industrial Estate

                                                                        Mettukuppam Vannagaram

                                                                        Chennai ndash 600095 Tamil Nadu

                                                                        INDIA

                                                                        Tel +91-44-2476259424764608

                                                                        Fax +91-44-24766920

                                                                        Web wwwniceneotechcom

                                                                        E-mail infoniceneotechcom marketingniceneotechcom

                                                                        Overseas Enquiry overseasniceneotechcom

                                                                        Toll Free +91-1800-425-2594

                                                                        • nice NeOumltech Medical Systems Pvt Ltd
                                                                        • Presentation Flow
                                                                        • Company INFO
                                                                        • Timeline-nice Neotech
                                                                        • Why nice Neotech
                                                                        • Why Infant Radiant Warmers are Used
                                                                        • MODES OF HEAT LOSS
                                                                        • Slide 8
                                                                        • Effects of Far IR in Infant body
                                                                        • Types of heaters
                                                                        • Slide 11
                                                                        • Benefits of Heaterhellip
                                                                        • Slide 13
                                                                        • Slide 14
                                                                        • Slide 15
                                                                        • How to operate
                                                                        • Control Panel Operation
                                                                        • Operating Modes
                                                                        • Slide 19
                                                                        • Slide 20
                                                                        • Alarms
                                                                        • Heater Module
                                                                        • Bed Platform Operation
                                                                        • Side Panel Operation
                                                                        • X-ray Procedure
                                                                        • Comparison of nice 2010 Infant Radiant Warmers
                                                                        • How nice Infant Radiant Warmers fare over competitors
                                                                        • Slide 28
                                                                        • Slide 29
                                                                        • Cleaning Procedure
                                                                        • Preventive Maintenance
                                                                        • Troubleshooting
                                                                        • System Fault Infant Radiant Warmer
                                                                        • Slide 34
                                                                        • Slide 35
                                                                        • Certifications and Approvals
                                                                        • Contact Us

                                                                          Contact Us

                                                                          nice Neotech Medical Systems Private Limited

                                                                          No 85 Krishna Industrial Estate

                                                                          Mettukuppam Vannagaram

                                                                          Chennai ndash 600095 Tamil Nadu

                                                                          INDIA

                                                                          Tel +91-44-2476259424764608

                                                                          Fax +91-44-24766920

                                                                          Web wwwniceneotechcom

                                                                          E-mail infoniceneotechcom marketingniceneotechcom

                                                                          Overseas Enquiry overseasniceneotechcom

                                                                          Toll Free +91-1800-425-2594

                                                                          • nice NeOumltech Medical Systems Pvt Ltd
                                                                          • Presentation Flow
                                                                          • Company INFO
                                                                          • Timeline-nice Neotech
                                                                          • Why nice Neotech
                                                                          • Why Infant Radiant Warmers are Used
                                                                          • MODES OF HEAT LOSS
                                                                          • Slide 8
                                                                          • Effects of Far IR in Infant body
                                                                          • Types of heaters
                                                                          • Slide 11
                                                                          • Benefits of Heaterhellip
                                                                          • Slide 13
                                                                          • Slide 14
                                                                          • Slide 15
                                                                          • How to operate
                                                                          • Control Panel Operation
                                                                          • Operating Modes
                                                                          • Slide 19
                                                                          • Slide 20
                                                                          • Alarms
                                                                          • Heater Module
                                                                          • Bed Platform Operation
                                                                          • Side Panel Operation
                                                                          • X-ray Procedure
                                                                          • Comparison of nice 2010 Infant Radiant Warmers
                                                                          • How nice Infant Radiant Warmers fare over competitors
                                                                          • Slide 28
                                                                          • Slide 29
                                                                          • Cleaning Procedure
                                                                          • Preventive Maintenance
                                                                          • Troubleshooting
                                                                          • System Fault Infant Radiant Warmer
                                                                          • Slide 34
                                                                          • Slide 35
                                                                          • Certifications and Approvals
                                                                          • Contact Us