M295 M295 M295 M295 E 02.4.CF.2 T-RCP Remote Control Pad Instructions
71

T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Mar 27, 2018

Download

Documents

truongkhuong
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

M295 M295 M295 M295 EEEE 02.4.CF.2

T-RCP Remote Control Pad

Instructions

Page 2: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

- 1 -

Contents

Introduction............................................................................................................................................................. 31 General Description........................................................................................................................... 32 Appearance of Remote Control Pad.................................................................................................. 43 Connection of Remote Control Pad and HUB Controller ................................................................... 54 Startup and Shutdown ....................................................................................................................... 65 Screen Configuration......................................................................................................................... 7

(1) Summary................................................................................................................................. 7(2) Layout of OPERATION Screen ............................................................................................... 7(3) Configuration of SETTING MENU Screens ............................................................................. 8(4) Configuration of MAINTENANCE MENU Screens................................................................. 11

I Preparations ................................................................................................................................................... 121 Connecting and Setting Electrical Components............................................................................... 12

(1) Connecting Electrical Components to HUB Controller ........................................................... 12(2) Turning ON the EXTERNAL Switch of Dia-Illumination Lamp Power Supply......................... 13(3) Placing Uniblitz Shutter Controller into N.C. Mode ................................................................ 13

2 Setting Attachment Information ....................................................................................................... 14(1) Objective ............................................................................................................................... 14(2) Condenser Module ................................................................................................................ 15(3) Filter Block ............................................................................................................................ 16(4) Excitation Filter...................................................................................................................... 17(5) Barrier Filter .......................................................................................................................... 18

3 Setting Uniblitz Shutter Controller Connection................................................................................. 194 Setting Shutter Key Function ........................................................................................................... 205 Setting Foot Switch Function ........................................................................................................... 216 Indicating Z-Axis Position ................................................................................................................ 22

II Basic Operation .............................................................................................................................................. 231 Selecting Objective Lens ................................................................................................................. 232 Selecting Filter Block ....................................................................................................................... 243 Selecting Optical Path ..................................................................................................................... 254 Selecting Condenser Module........................................................................................................... 255 Selecting Excitation Filter ................................................................................................................ 266 Selecting Barrier Filter ..................................................................................................................... 267 Inserting/Removing Analyzer........................................................................................................... 278 Opening/Closing Shutter ................................................................................................................. 279 Operating Dia-Illumination Lamp (12V100W lamp only) .................................................................. 27

(1) Turning Remote Control Operation ON/OFF ......................................................................... 27(2) Turning Lamp ON/OFF.......................................................................................................... 28(3) Adjusting Lamp Voltage......................................................................................................... 28

10 Resetting Z-Axis Position to Zero .................................................................................................... 2911 Outputting External Trigger.............................................................................................................. 29

III Additional Functions ....................................................................................................................................... 301 Interlocking Objective and Condenser Module Selection................................................................. 302 Interlock Filter Block and Excitation/Barrier Filters Selection ........................................................... 313 Restricting Rotation of Revolving Nosepiece ................................................................................... 32

(1) Set Rotation Restriction......................................................................................................... 32(2) Cancel Rotation Restriction ................................................................................................... 33

IV Other Operations ............................................................................................................................................ 341 Operating LCD Backlight ................................................................................................................. 34

(1) Turning Backlight ON/OFF .................................................................................................... 34(2) Adjusting Brightness of Backlight .......................................................................................... 34

Page 3: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

- 2 -

2 Adjusting LCD Contrast ................................................................................................................... 35V Operation and Indication Functions ................................................................................................................ 36

1 Changing Operation Mode of Fine Focus Knob............................................................................... 362 Setting Indication on Operation Screen ........................................................................................... 373 Setting Buzzer ON/OFF................................................................................................................... 384 Setting Remote Control ON/OFF..................................................................................................... 395 Turning ON/OFF LED with LCD ON/OFF Key ................................................................................. 406 Setting Communications Delimiter of External Device ..................................................................... 417 Registering Data for Optical Members............................................................................................. 42

(1) Objective ............................................................................................................................... 42(2) Condenser Module ................................................................................................................ 44(3) Filter Block ............................................................................................................................ 45(4) Dichroic Mirror ....................................................................................................................... 47(5) Excitation Filter...................................................................................................................... 48(6) Barrier Filter .......................................................................................................................... 49

VI Maintenance Menu ......................................................................................................................................... 501 Indicating Program Version and Checksum..................................................................................... 502 Testing Electrical Component Operations ....................................................................................... 51

(1) All-Test Mode ........................................................................................................................ 51(2) Individual Test Mode ............................................................................................................. 53

3 Testing Hardware of Remote Control Pad ....................................................................................... 54(1) Key Input Test ....................................................................................................................... 54(2) LED ON/OFF Test................................................................................................................. 55(3) LCD Indication Test............................................................................................................... 55

4 Initializing Focus Motion Potentiometer ........................................................................................... 565 Resetting Remote Control Pad Settings to Default Values .............................................................. 576 Resetting HUB Controller Settings to Default Values....................................................................... 587 Reset Control Data to Default Values .............................................................................................. 59

(1) All Components ..................................................................................................................... 59(2) Revolving Nosepiece............................................................................................................. 60(3) Condenser Module ................................................................................................................ 61(4) Optical Path........................................................................................................................... 62(5) Filter Block ............................................................................................................................ 63(6) Excitation Filter...................................................................................................................... 64(7) Barrier Filter .......................................................................................................................... 65(8) Auxiliary Filter........................................................................................................................ 66

VII Troubleshooting .............................................................................................................................................. 671 Remote control pad doesn’t start up. (TITLE screen appears, and the screen freezes.).................. 672 Electrical component doesn’t operate.............................................................................................. 673 Revolving nosepiece doesn’t rotate at a certain position of the fine focus knob............................... 674 Dia-illumination lamp cannot be turned ON or OFF. ........................................................................ 675 Voltage of dia-illumination lamp cannot be adjusted. ....................................................................... 686 Opening/closing status of Uniblitz shutter doesn’t coincide with LED indication. ............................. 687 Objective information is not indicated. ............................................................................................. 688 Condenser module information is not indicated. .............................................................................. 689 Filter block information is not indicated............................................................................................ 6810 Excitation filter information is not indicated. ..................................................................................... 6911 Barrier filter information is not indicated........................................................................................... 6912 Dia-illumination lamp voltage is not indicated. ................................................................................. 6913 Z-axis coordinates are not indicated. ............................................................................................... 6914 Buzzer doesn’t sound when a key is pressed. ................................................................................. 69

Page 4: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

1 General Description

- 3 -

Introduction

1 General Description

The remote control pad is connected to the HUB controller and used for remote control of the accessory electricalcomponents of the ECLIPSE TE2000 series.

The remote control pad enables control of the electrical components with key operations on the pad. The remotecontrol pad also indicates a variety of information using the LED and displays on the LCD, such as electricalcomponent positions and information on the attached optical members.

Page 5: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

2 Appearance of Remote Control Pad

- 4 -

2 Appearance of Remote Control Pad

1

2

34

56

7

8 9

1011

12

1314

15

16

17

18

No. Parts Function whenOPERATION screen is indicated

Functions when settingmenu/maintenance menu is indicated

1 Liquid crystal display (LCD)

OPERATION screen: Indicateselectrical component positions andinformation on the attached opticalmembers

Setting menu: Allows setting ofvarious functions/parameters of theelectrical components.

2 OBJECTIVE key Selects an objective lens

3 SHUTTER A key Opens/closes the shutter allocated toShutter A key

4 SHUTTER B key Opens/closes the shutter allocated toShutter B key

5 EXCITER key Selects an excitation filter6 BARRIER key Selects a barrier filter

Enters alphanumeric characters.

7 EPI-FILTER key Selects a filter block8 Z-AXI. RESET key Resets the z-axis position to zero9 CAMERA EXP. key Outputs an external trigger

10 LCD MODE key Changes to another screen Changes to another screen.11 LCD ON/OFF key Turns ON/OFF the LCD backlight

12 LCD Bright/Dark key Adjusts the brightness of the LCDbacklight

13 DIA LAMP REMOTE key Turns ON/OFF the remote controloperation of the dia-illumination lamp

14 DIA LAMP ON/OFF key Turns ON/OFF the dia-illuminationlamp

15 DIA LAMP ADJ. key Adjusts the voltage of the dia-illumination lamp PREV/NEXT keys (Change selection.)

16 CONDENSER key Selects a condenser module BS key (Deletes entered letters), CLRkey (All clear)

17 ANALYZER IN/OUT key Inserts/removes the analyzer

18 LIGHT PATH key Selects an optical pathCursor key (Selects an item.)Return key (Confirms selection.)

Page 6: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

3 Connection of Remote Control Pad and HUB Controller

- 5 -

3 Connection of Remote Control Pad and HUB Controller

Connect the HUB connection cable fixed to the remote control pad to REMOTE connector of the HUB controller asshown below.

Hub connection cable

[Rear panel of remote control pad]

REMOTE connector

[Connector panel of HUB controller]

Page 7: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

4 Startup and Shutdown

- 6 -

4 Startup and Shutdown

1 Startup1) Turn ON power for the dia-illumination lamp. When

using Uniblitz shutter, also turn ON power for theUniblitz shutter controller.

2) Turn ON the power to the HUB controller.

3) Power is supplied to the remote control pad from theHUB controller.

AC

[Side panel of HUB controller]

The TITLE screen appears on the LCD of the remote control pad and changes to the OPERATION screen in 2 or 3seconds.

T-RCP REMOTE CONTROL PAD

V1.00

[TITLE screen] [OPERATION screen]

Note:When you use the remote control pad for the first time, all the optical members, such as the objectives, willbe set by default to “Not attached” and the Operation screen will appear as shown below.

“---------” indicates that the opticalmember is “not attached.”

If an electric component is not connected to the HUB controller, the indication of the component will be blank.If the electrical component is not connected to the HUB controller correctly, the HUB controller cannot detectthe component and the indication of the component will also be blank.If no electrical component is connected to the HUB controller, the screen will appear as shown below.

2 Shutdown1) Turn OFF the power to the HUB controller.

2) Turn OFF power for the dia-illumination lamp and theUniblitz shutter controller.

Power switch

Connect the specified AC adapter

Page 8: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

5 Screen Configuration

- 7 -

5 Screen Configuration

(1) Summary

Power ON

MODE key

MODE key

Is MODE keypressed?

MODE keyNo

Yes

T-RCP REMOTE CONTROL PAD

V1.00

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Note:If power is turned ON while the MODE key is held down, the MAINTENANCE MENU screen will appear.

(2) Layout of OPERATION Screen

Objective lensinformation

P. 14

Condenser moduleinformationP. 15

Excitation filterinformationP. 17

Barrier filterinformationP. 18

Filter block informationP. 16

Z-axis positionP. 21

Dia-Illuminationlamp voltage

P. 28

Note:If the electric components are not connected to the HUB controller, information for the unconnectedcomponents will not appear.

Page 9: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

5 Screen Configuration

- 8 -

(3) Configuration of SETTING MENU Screens

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

OBJECTIVESeries      Mag.

1[P Fluor   ][DL       4x      ]

2[P Fluor   ][DL       10x     ]

3[P Fluor   ][ELWD DM  20x     ]

4[P Fluor   ][ELWD DM  40x     ]

5[P Fluor   ][ELWD DLL 60x     ]

6[P Fluor   ][DLL      100x    ]

 Edit  Type       N.A.  W.D.       Dry         0.13 17.10

CONDENSERName

1[BF        ]

2[PhL       ]

3[Ph1       ]

4[Ph2       ]

5[Ph3       ]

 Edit

FILTERName

1[UV-1A     ] EX : EX365/10

2[V-1A      ] DM : DM400

3[BV-1A     ] BA : BA400

4[B-1A      ]

5[G-1A      ]

6[DIA       ]

 Edit

EXCITER

Name1[EX465-495 ]5[――――――――――]

2[EX460-500 ]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[DIA       ]

 Edit

BARRIER

Name1[BA515-555 ]5[――――――――――]

2[BA510     ]6[――――――――――]

3[BA510-560 ]7[――――――――――]

4[――――――――――]8[DIA       ]

 Edit

COMBINATION(OBJ.>CONDENSER)

Objective              Condenser1.P Fluor DL       4x  [2]:PhL

2.P Fluor DL       10x [3]:Ph1

3.P Fluor ELWD DM  20x [3]:Ph1

4.P Fluor ELWD DM  40x [4]:Ph2

5.P Fluor ELWD DLL 60x [4]:Ph2

6.P Fluor DLL      100x[5]:Ph3

COMBINATION(FILTER>EX/BA)

Filter   EX         BA1.UV-1A  [-]:―――――――[-]:―――――――

2.V-1A   [-]:―――――――[-]:―――――――

3.BV-1A  [-]:―――――――[-]:―――――――

4.B-1A   [1]:465-495[1]:515-555

5.G-1A   [-]:―――――――[-]:―――――――

6.DIA    [8]:DIA    [8]:DIA

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

*1

*3

*4

*5

*6

*2

MODE key

[ ] key

Page 10: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

5 Screen Configuration

- 9 -

OBJECTIVESeries      Mag.

1[P Fluor   ][DL       4x      ]

2[P Fluor   ][DL       10x     ]

3[P Fluor   ][ELWD DM  20x     ]

4[P Fluor   ][ELWD DM  40x     ]

5[P Fluor   ][ELWD DLL 60x     ]

6[P Fluor   ][DLL      100x    ]

 Edit  Type       N.A.  W.D.       Dry        0.13  17.10

CONDENSERName

1[BF        ]

2[PhL       ]

3[Ph1       ]

4[Ph2       ]

5[Ph3       ]

 Edit

FILTERName

1[UV-1A     ] EX : EX365/10

2[V-1A      ] DM : DM400

3[BV-1A     ] BA : BA400

4[B-1A      ]

5[G-1A      ]

6[DIA       ]

 Edit

EXCITER

Name1[EX465-495 ]5[――――――――――]

2[EX460-500 ]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[DIA       ]

 Edit

BARRIER

Name1[BA515-555 ]5[――――――――――]

2[BA510     ]6[――――――――――]

3[BA510-560 ]7[――――――――――]

4[――――――――――]8[DIA       ]

 Edit

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[――――――――――]

Mag.    :[0  ]x

N.A.    :[0    ]

W.D.    :[0    ]

Type    :[――――――――――]

EDIT CONDENSER DATA

Location:[Other1]

Name    :[          ]

EDIT FILTER DATA

Location:[Other1]

Name    :[          ]

EX      :[――――――――――]Edit

DM      :[――――――――――]Edit

BA      :[――――――――――]Edit

EDIT DM DATA

Location:[Other1]

Name    :[          ]

EDIT EX DATA

Location:[Other1]

Name    :[          ]

EDIT BA DATA

Location:[Other1]

Name    :[          ]

EDIT BA DATA

Location:[Other1]

Name    :[          ]

EDIT EX DATA

Location:[Other1]

Name    :[          ]

*1

*2

*3

*4

*5

Page 11: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

5 Screen Configuration

- 10 -

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

ROTATION STOPPER

Current setting:OFF

ON     OFF

FINE FOCUS KNOB

[100]μm/rot.

OTHER UNIT

UNIBlitze Shutter :[OFF]

REMOTE CONTROL PAD

DisplayObjective    :[ON ]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[OFF]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

SHUTTER KEY

Shutter A    Shutter B[          ][          ]

FOOT SWITCH

Unused

Revolving nosepiece

Condenser cassette

Filter block

Exciter filter

Barrier filter

AUX filter

COMMUNICATION DELIMITERFOR EXTERNAL UNIT

CR     LF

MODE key

*6

[ ] key

Page 12: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

Introduction

5 Screen Configuration

- 11 -

(4) Configuration of MAINTENANCE MENU Screens

PAD DATA

Initialize PAD data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes

and press ENTER to start.

HUB DATA

Initialize HUB data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start.

TEST1           1/5→ ALL TEST

OBJECTIVE              1             [  ]              2             [  ]              3             [  ]              4             [  ]              5             [  ]              6             [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select item  and press enter to start.

VERSION

T-RCP REMOTE CONTROL PADV1.00 Sum=1234

T-HUBC HUB CONTROLLERMAIN-cpu: V1.00 Sum=5678SUB-cpu : V1.00 Sum=ABCD

CONTROL DATA

Initialize all control data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX filter

Control Data

Initialize Control data               for AUX filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start.

・・・・・・

[ ] key

MODE key

Page 13: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

1 Connecting and Setting Electrical Components

- 12 -

IPreparations

1 Connecting and Setting Electrical Components

(1) Connecting Electrical Components to HUB Controller

Make sure that the power switches for the microscope and the HUB controller are OFF.

Connect the electrical components to be used to the HUB controller.

Revolving nosepiece(Motorized nosepiece)

Analyzer(Motorized analyzer)

Excitation filter changeover(Ex filter wheel)

Episcopic illuminationshutter(Motorized Epi-fl attachment)

Barrier filter changeover(BA filter wheel)

Filter block changeover(Motorized cassetteholder)

Condenser modulechangeover (Motorizedcondenser turret)

Reserve connector(not used)

PCExternal device(XY stage, etc.)

Uniblitz shuttercontroller

Remote control pad

Dia-Illumination lamp power supply(TS2-PS100W power supply)* TE-PS30 and TE-PSE30 powersupplies cannot be connected.

Foot switch, etc.

[Connector panel of HUB controller]

Page 14: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

1 Connecting and Setting Electrical Components

- 13 -

I

(2) Turning ON the EXTERNAL Switch of Dia-Illumination Lamp Power Supply

EXTERNAL switch

[Rear panel of dia-illumination lamp power supply]

Note:When the EXTERNAL switch is OFF, you cannot operate the dia-illumination lamp on the remote control pad.

A 30-W power supply cannot be connected to the HUB controller.

(3) Placing Uniblitz Shutter Controller into N.C. Mode

Be sure to set the controller to the N.C. mode when using the Uniblitz shutter.

Turn the switch of the channel used to N.C.

UNIBLITZ

[Front panel of Uniblitz shutter controller]

Note:N.C. stands for Normal Close.

If the Uniblitz shutter controller is set to the N.C. mode, when power is turned on, the controller starts up with theshutter closed.

Page 15: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

2 Setting Attachment Information

- 14 -

2 Setting Attachment Information

Start up the remote control pad and set the optical members attached to the electrical components in use.

The LCD indicates information for the set optical members.

(1) Objective

Set the information on objectives attached to the revolving nosepiece (motorized nosepiece).

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Objective” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Series” for the objective using the PREV or NEXTkey.

Note:The numbers at the left of the screen represent theaddresses of the revolving nosepiece.

OBJECTIVESeries      Mag.

1[P Fluor   ][―――――――――――――――――]

2[――――――――――][―――――――――――――――――]

3[――――――――――][―――――――――――――――――]

4[――――――――――][―――――――――――――――――]

5[――――――――――][―――――――――――――――――]

6[――――――――――][―――――――――――――――――]

 Edit  Type       N.A.  W.D.       Dry         0.13 17.10

3 Move the cursor using the key or the key.

Select “Mag.” for the objective using the PREV or NEXTkey.

OBJECTIVESeries      Mag.

1[P Fluor   ][DL       4x      ]

2[――――――――――][―――――――――――――――――]

3[――――――――――][―――――――――――――――――]

4[――――――――――][―――――――――――――――――]

5[――――――――――][―――――――――――――――――]

6[――――――――――][―――――――――――――――――]

 Edit  Type       N.A.  W.D.       Dry         0.13 17.10

4 Move to the next address of the revolving nosepiece usingthe key or the key. Repeat steps 2 and 3 above.

OBJECTIVESeries      Mag.

1[P Fluor   ][DL       4x      ]

2[――――――――――][―――――――――――――――――]

3[――――――――――][―――――――――――――――――]

4[――――――――――][―――――――――――――――――]

5[――――――――――][―――――――――――――――――]

6[――――――――――][―――――――――――――――――]

 Edit  Type       N.A.  W.D.       ―――――――――― ――――― ―――――

5 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If the revolving nosepiece is not connected, you cannot set the information on objectives.

“---------” indicates that no objective is attached or registered at the address.

If an attached objective is not listed among the presets, register it to the list of optional objectives.

For details, see V-7 (2) Objective in section V-7 (Registering Data for Optical Members).

Page 16: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

2 Setting Attachment Information

- 15 -

I

(2) Condenser Module

Set the information on condenser modules attached to the condenser module changeover (motorized condenserturret).

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Condenser” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Name” for the condenser module using the PREVor NEXT key.

If the condenser module is marked “A,” select “BF.”Note:

The numbers at the left of the screen represent theaddresses of the condenser module changeover.

CONDENSERName

1[BF        ]

2[――――――――――]

3[――――――――――]

4[――――――――――]

5[――――――――――]

 Edit

3 Move to the next address of the condenser modulechangeover using the key or the key. Repeat step 2above.

CONDENSERName

1[BF        ]

2[――――――――――]

3[――――――――――]

4[――――――――――]

5[――――――――――]

 Edit

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If the condenser module changeover is not connected, you cannot set the information on condenser modules.

“---------” indicates that no condenser module is attached or registered at the address.

If an attached condenser module is not listed in the presets, register it to the list of optional condenser modules.

For details, see V-7 (2) Condenser Module in section V-7 (Registering Data for Optical Members).

Page 17: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

2 Setting Attachment Information

- 16 -

(3) Filter Block

Set the information on filter blocks attached to the filter block changeover (motorized cassette holder).

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Filter” using the key or the key and then pressthe [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Name” for the filter block using the PREV or NEXTkey.

Note:The numbers at the left of the screen represent theaddresses of the filter block changeover.

FILTERName

1[UV-1A     ] EX : EX365/10

2[――――――――――] DM : DM400

3[――――――――――] BA : BA400

4[――――――――――]

5[――――――――――]

6[――――――――――]

 Edit

3 Move to the next address of the filter block changeoverusing the key or the key. Repeat step 2 above.

FILTERName

1[UV-1A     ] EX : ――――――――――

2[――――――――――] DM : ――――――――――

3[――――――――――] BA : ――――――――――

4[――――――――――]

5[――――――――――]

6[――――――――――]

 Edit

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If the filter block changeover is not connected, you cannot set the information on filter blocks.

“---------” indicates that no filter block is attached or registered at the address.

If a filter block attached is not in the presets, register it to the list of optional filter blocks. For details, see V-7 (3)Filter Block in section V-7 (Registering Data for Optical Members).

Page 18: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

2 Setting Attachment Information

- 17 -

I

(4) Excitation Filter

Set the information on excitation filters attached to the excitation filter changeover (EX filter wheel).

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Exciter” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Name” for the excitation filter using the PREV orNEXT key.

Note:The numbers at the left of the screen represent theaddresses of the excitation filter changeover.

EXCITER

Name1[EX465-495 ]5[――――――――――]

2[――――――――――]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[――――――――――]

 Edit

3 Move to the next address of the excitation filter changeoverusing the key or the key. Repeat step 2 above.

EXCITER

Name1[EX465-495 ]5[――――――――――]

2[――――――――――]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[――――――――――]

 Edit

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If the excitation filter changeover is not connected, you cannot set the information on excitation filters.

“---------” indicates that no excitation filter is attached or registered at the address.

If an attached excitation filter is not listed in the presets, register it to the list of optional excitation filters. For details,see V-7 (5) Excitation Filter in section V-7 (Registering Data for Optical Members).

If any of the openings in the changeover is empty (i.e., in the DIA position) when switching from one excitation filterto another, a bright light may enter the observation side, depending on the layout of the optical members,temporarily affecting your vision. To prevent this, we recommend that you attach excitation filters to thechangeover, as shown below.

Sample layout of excitation filters

12

3

45

6

7

8

Set Address 8 to DIA (empty). Attach excitation filters from Address 1 forward,leaving no empty openings in-between.

Put a light shielding plate on the openingswith no filter attached.

Excitation filter changeover(EX filter wheel)

[Excitation filter attaching side]

Page 19: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

2 Setting Attachment Information

- 18 -

(5) Barrier Filter

Set the information on barrier filters attached to the barrier filter changeover (BA filter wheel).

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Barrier” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Name” for the barrier filter using the PREV or NEXTkey.

Note:The numbers at the left of the screen represent theaddresses of the barrier filter changeover.

BARRIER

Name1[BA515-555 ]5[――――――――――]

2[――――――――――]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[――――――――――]

 Edit

3 Move to the next address of the barrier filter changeoverusing the key or the key. Repeat step 2 above.

BARRIER

Name1[BA515-555 ]5[――――――――――]

2[――――――――――]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[――――――――――]

 Edit

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If the barrier filter changeover is not connected, you cannot set the information on barrier filters.

“---------” indicates that no barrier filter is attached or registered at the address.

If a barrier filter attached is not in the presets, register it to the list of optional barrier filters. For details, see V-7 (6)Barrier Filter in section V-7 (Registering Data for Optical Members).

Page 20: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

3 Setting Uniblitz Shutter Controller Connection

- 19 -

I

3 Setting Uniblitz Shutter Controller Connection

The HUB controller doesn’t automatically detect the Uniblitz shutter controller, unlike other electrical components.Be sure to set the connection of the Uniblitz shutter controller when using the Uniblitz shutter.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and then pressthe [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Other Unit” using the key or the key and thenpress the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Select “ON” using the PREV or NEXT key. OTHER UNIT

UNIBlitze Shutter :[ON ]

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 21: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

4 Setting Shutter Key Function

- 20 -

4 Setting Shutter Key Function

Allocate the shutters to the Shutter A and Shutter B keys.

1 Press the MODE key, and the SETTING MENU screen willappear on

Select “More” using the key or the key and then pressthe [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Shutter Key” using the key or the key andthen press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Select a shutter to be operated by the SHUTTER A keyusing the PREV or NEXT key.

SHUTTER KEY

Shutter A     Shutter B[Epi shutter] [OFF        ]

4 Move the cursor using the key or the key.

Select a shutter to be operated by the SHUTTER B keyusing the PREV or NEXT key.

SHUTTER KEY

Shutter A     Shutter B[Epi shutter] [UNI CH.#1  ]

5 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If the Nikon motorized Epi-Fl attachment is not connected and the Uniblitz shutter controller connection is set to“OFF,” you cannot set the shutter key function.For details on the Uniblitz shutter controller connection setting, see section I-3 (Setting Uniblitz Shutter ControllerConnection).

You can’t allocate the same shutter to both the Shutter A and the Shutter B keys.

Page 22: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

5 Setting Foot Switch Function

- 21 -

I

5 Setting Foot Switch Function

Allocate an electrical component to the foot switch.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and then pressthe [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Foot Switch” using the key or the key andthen press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the cursor using the key or the key and selectan electrical component.

FOOT SWITCH

Unused

Revolving nosepiece

Condenser cassette

Filter block

Exciter filter

Barrier filter

AUX filter

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:If a component is not connected, you cannot select the component.

If the selected electrical component is disconnected while the OPERATION screen is displayed on the LCD, thecomponent is automatically set to “Unused.”

Page 23: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

I Preparations

6 Indicating Z-Axis Position

- 22 -

6 Indicating Z-Axis Position

Indication of the z-axis position is set to “OFF” by default.

To indicate the z-axis position, edit the setting using the following procedure. (If TE2000-U or TE2000-S is used,the Z-Axis Position cannot be indicated.)

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and then pressthe [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Remote Control Pad” using the key or the keyand then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Select “Z-Position” using the key or the key andselect “ON” using the PREV or NEXT key.

REMOTE CONTROL PAD

DisplayObjective    :[ON ]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[ON ]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 24: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

1 Selecting Objective

- 23 -

II

IIBasic Operation

1 Selecting Objective

Press the OBJECTIVE key to select an objective.

Each OBJECTIVE key corresponds to one of theobjectives; you can select an objective by simply pressingthe key corresponding to the objective.

(3)

Screen indication:

(1) The series name of the objective in the optical path isindicated.

(2) The magnification of the objective in the optical pathis highlighted.

LED INDICATION:

(3) The LED of the OBJECTIVE key corresponding tothe objective in the optical path is lit green.

(2) (1)

Note:If the revolving nosepiece is not connected, the objective information will not be indicated on the LCD and the LED ofOBJECTIVE key will not light up.

Page 25: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

2 Selecting Filter Block

- 24 -

2 Selecting Filter Block

Press the FILTER key to select a filter block.

Each FILTER key corresponds to one of the filter blocks;you can select a filter block by simply pressing the keycorresponding to the block.

(4)

Screen indication:

(1) The name of the filter block in the optical path isindicated.

(2) The excitation method of the filter block is indicated.If the filter block is in the optical path, the excitationmethod is highlighted.

If no filter block is attached, “---------“ is indicated. Ifa registered filter block is attached, “O” (= Other) isindicated.

(3) The wavelengths of the excitation filter, dichroicmirror, and barrier filter, which make up the filterblock, are indicated.

LED INDICATION:

(4) The LED of the FILTER key corresponding to thefilter block in the optical path is lit green.

(1)(2) (3)

Note:If the filter block changeover is not connected, the filter block information will not be indicated on the LCD and theLED of FILTER key will not light up.

Page 26: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

3 Selecting Optical Path

- 25 -

II

3 Selecting Optical Path

Press the LIGHT PATH key to select an optical path.

Each LIGHT PATH key corresponds to one of the opticalpaths; you can select an optical path by simply pressingthe key corresponding to the path.

LED INDICATION:

(1) The LED of the LIGHT PATH key corresponding tothe currently selected optical path is lit green.

(1)

Note:If TE2000-U or TE2000-S is used, you cannot use the LIGHT PATH keys to select an optical path. Use optical pathswitchover dial on the microscope instead.

4 Selecting Condenser Module

Press the CONDENSER key to turn the condenser modulechangeover clockwise or counterclockwise to select acondenser module.

Screen indication:

(1) The currently selected address in the condensermodule changeover is indicated.

(2) The name of the condenser module in the opticalpath is indicated.

(1) (2)

Note:If the condenser module changeover is not connected, the condenser module information will not be indicated onthe LCD.

Page 27: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

5 Selecting Excitation Filter

- 26 -

5 Selecting Excitation Filter

Press the EXCITER key to turn the excitation filterchangeover clockwise or counterclockwise to select anexcitation filter.

Screen indication:

(1) The currently selected address in the excitation filterchangeover is indicated.

(2) The wavelength of the excitation filter in the opticalpath is indicated. (1) (2)

Note:If the excitation filter changeover is not connected, the excitation filter information will not be indicated on the LCD.

If any of the openings in the changeover is empty (i.e., in the DIA position) when switching over one excitation filterto another, a bright light may enter the observation side, depending on the layout of the optical members. The lightmay temporarily impair your vision or the object may be unnecessarily exposed to light.

To prevent this, we recommend that you close the episcopic illumination shutter when switching over excitationfilters.

6 Selecting Barrier Filter

Press the BARRIER key to turn the barrier filterchangeover clockwise or counterclockwise to select abarrier filter.

Screen indication:

(1) The currently selected address in the barrier filterchangeover is indicated.

(2) The wavelength of the barrier filter in the optical pathis indicated.

(1) (2)

Note:If the barrier filter changeover is not connected, the barrier filter information will not be indicated on the LCD.

Page 28: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

7 Inserting/Removing Analyzer

- 27 -

II

7 Inserting/Removing Analyzer

Press the ANALYZER IN/OUT key to insert and removethe analyzer.

LED INDICATION:

IN OUTGreen Orange

Note:If the analyzer is not connected, the LED will not light up.

8 Opening/Closing Shutter

Press the SHUTTER A key to open and close the shutterallocated to SHUTTER A.

Press the SHUTTER B key to open and close the shutterallocated to SHUTTER B.

LED INDICATION:

OPEN CLOSEGreen Orange

Note:If the allocation for the SHUTTER A key and the SHUTTER B key is set to “OFF,” the LED will not light up.

For details on shutter allocation, see section I-4 (Setting Shutter Key Function).

9 Operating Dia-Illumination Lamp (12V100W lamp only)

(1) Turning Remote Control Operation ON/OFF

Remote control operation refers to operating the dia-illumination lamp of the microscope from the remote controlpad.

Press the DIA LAMP REMOTE key to turn remote controloperation ON or OFF.

LED INDICATION:

Operation ON Operation OFFGreen Orange

Note:When EXTERNAL switch of the dia-illumination lamp power supply is OFF, you cannot operate the lamp on theremote control pad nor turn the remote control operation ON or OFF. The LED will not light up.

Page 29: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

9 Operating Dia-Illumination Lamp (12V100W lamp only)

- 28 -

(2) Turning Lamp ON/OFF

You can turn the lamp ON or OFF as long as the LED ofthe REMOTE key is green.

Press the DIA LAMP ON/OFF key to turn the lamp ON orOFF.

LED INDICATION:

Lamp ON Lamp OFFGreen Orange

Note:When the LED of the DIA LAMP REMOTE key is orange or is unlit, the LED of the DIA LAMP ON/OFF key will notlight up.

(3) Adjusting Lamp Voltage

You can adjust the lamp voltage as long as the LED of theREMOTE key is green.

Press the DIA LAMP ADJ. key to adjust the lamp voltage.

Screen indication:

(1) The lamp voltage is indicated with a resolution of0.5 v.

For example, if the voltage is 9 v or more and below9.5 v, it will be indicated as 9 v.

Indication examples:

0 v

Below 3 v

3 vbelow 3.5 v

12 v or more

(1)

Note:When the LED of the DIA LAMP REMOTE key is orange or is unlit, the lamp voltage will not be indicated.

Adjusting the lamp voltage will also alter the color cast of the light. When color fidelity is important - for example,when taking a color photo, - use an ND filter to adjust brightness.

Page 30: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

II Basic Operation

10 Resetting Z-Axis Position to Zero

- 29 -

II

10 Resetting Z-Axis Position to Zero

The current z-axis position can be indicated as 0.00 µm.

Press the Z-AXI. RESET key to set the indication of the z-axis position to zero.

Screen indication:

(1) The z-axis position is indicated.

(1)

Note:When the indication of the z-axis position is OFF, you cannot reset the indication to zero.

If TE2000-U or TE2000-S is used, the Z-axis position cannot be indicated.

11 Outputting External Trigger

You can click the shutter of the 35-mm camera connectedto the HUB controller.

Press the CAMERA EXP. key to output an external trigger.

Note:For information about connection to a camera, refer to 22, “Connection to the EXP Connector,” in “VI. Assembly” inthe “Motorized Units for ECLIPSE TE2000-E/TE2000-U/TE2000-S”manual.

Page 31: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

III Additional Functions

1 Interlocking Objective and Condenser Module Selection

- 30 -

IIIAdditional Functions

1 Interlocking Objective and Condenser Module Selection

If combinations of objectives and condenser modules are set, then when an objective is selected, the condensermodule corresponding to the objective is automatically selected.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Combination (OBJ>Condenser)” using the keyor the key and then press the [ ] key .

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select the address in the condenser module changeover tointerlock with Address 1 of the revolving nosepiece usingthe PREV or NEXT key.

COMBINATION(OBJ.>CONDENSER)

Objective              Condenser1.P Fluor DL       4x  [2]:PhL

2.P Fluor DL       10x [-]:―――――

3.P Fluor ELWD DM  20x [-]:―――――

4.P Fluor ELWD DM  40x [-]:―――――

5.P Fluor ELWD DLL 60x [-]:―――――

6.P Fluor DLL      100x[-]:―――――

3 Move to the next address of the revolving nosepiece usingthe key or the key and repeat step 2 above.

COMBINATION(OBJ.>CONDENSER)

Objective              Condenser1.P Fluor DL       4x  [2]:PhL

2.P Fluor DL       10x [-]:―――――

3.P Fluor ELWD DM  20x [-]:―――――

4.P Fluor ELWD DM  40x [-]:―――――

5.P Fluor ELWD DLL 60x [-]:―――――

6.P Fluor DLL      100x[-]:―――――

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:“-” indicates that the address is not interlocked.

Unless both the revolving nosepiece and the condenser module changeover are connected, you cannot set thecombinations.

Page 32: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

III Additional Functions

2 Interlock Filter Block and Excitation/Barrier Filters Selection

- 31 -

III

2 Interlock Filter Block and Excitation/Barrier Filters Selection

If combinations of filter block and excitation/barrier filters are set, then when a filter block is selected, theexcitation/barrier filters corresponding to the block are automatically selected.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Combination (FILTER>EX/BA)” using the key orthe key and then press the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select the address in the excitation filter changeover tointerlock with Address 1 of the filter block changeover usingthe PREV or NEXT key.

COMBINATION(FILTER>EX/BA)

Filter   EX         BA1.UV-1A  [8]:DIA    [-]:―――――――

2.V-1A   [-]:―――――――[-]:―――――――

3.BV-1A  [-]:―――――――[-]:―――――――

4.B-1A   [-]:―――――――[-]:―――――――

5.G-1A   [-]:―――――――[-]:―――――――

6.DIA    [-]:―――――――[-]:―――――――

3 Move the cursor using the key or the key.

Select the address in the barrier filter changeover tointerlock with Address 1 of the filter block changeover usingthe PREV or NEXT key.

COMBINATION(FILTER>EX/BA)

Filter   EX         BA1.UV-1A  [8]:DIA    [-]:―――――――

2.V-1A   [-]:―――――――[-]:―――――――

3.BV-1A  [-]:―――――――[-]:―――――――

4.B-1A   [-]:―――――――[-]:―――――――

5.G-1A   [-]:―――――――[-]:―――――――

6.DIA    [-]:―――――――[-]:―――――――

4 Move to the next address of the filter block changeoverusing the key or the key and repeat steps 2 and 3above.

COMBINATION(FILTER>EX/BA)

Filter   EX         BA1.UV-1A  [8]:DIA    [8]:DIA

2.V-1A   [-]:―――――――[-]:―――――――

3.BV-1A  [-]:―――――――[-]:―――――――

4.B-1A   [-]:―――――――[-]:―――――――

5.G-1A   [-]:―――――――[-]:―――――――

6.DIA    [-]:―――――――[-]:―――――――

5 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:“-” indicates that the address is not interlocked.

Unless the filter block changeover is connected and either the excitation filter changeover and barrier filterchangeover is connected (or both are connected), you cannot set the combinations.

Page 33: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

III Additional Functions

3 Restricting Rotation of Revolving Nosepiece

- 32 -

3 Restricting Rotation of Revolving Nosepiece

You can restrict the rotation of the revolving nosepiece with the OBJECTIVE key, so that the revolving nosepiecewill not turn from the set z-axis position and upward. The upper limit setting is effective for all the addresses of therevolving nosepiece.

(1) Set Rotation Restriction

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Rotation Stopper” using the key or the keyand then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the z-axis position to the position to restrict therevolving nosepiece rotation (the position at which youwant the revolving nosepiece to stop turning).

4 Move the cursor to “ON” using the key or the key andthen press the [ ] key.

ROTATION STOPPER

Current setting:OFF

ON     OFF

5 A confirmation message appears.

Press the [ ] key.

To cancel, move the cursor to “NO” using the key or the key and then press the [ ] key.

ROTATION STOPPER

Current setting:ON

ON     OFF

Rotation stopper ON ?

Yes     No

6 When the setting is accepted, the “Complete!” messageappears and remains displayed for about one second.

ROTATION STOPPER

Current setting:ON

ON     OFF

Complete!

7 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 34: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

III Additional Functions

3 Restricting Rotation of Revolving Nosepiece

- 33 -

III

(2) Cancel Rotation Restriction

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Rotation Stopper” using the key or the keyand then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the cursor to “OFF” using the key or the keyand then press the [ ] key.

ROTATION STOPPER

Current setting:ON

ON     OFF

4 A confirmation message appears.

Press the [ ] key.

To cancel, move the cursor to “NO” using the key or the key and then press the [ ] key.

ROTATION STOPPER

Current setting:ON

ON     OFF

Rotation stopper OFF ?

Yes     No

5 When the cancellation is accepted, a “Complete!” messageappears and remains displayed for about one second.

ROTATION STOPPER

Current setting:OFF

ON     OFF

Complete!

6 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 35: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

IV Other Operations

1 Operating LCD Backlight

- 34 -

IVOther Operations

1 Operating LCD Backlight

(1) Turning Backlight ON/OFF

Press the LCD ON/OFF key to turn the backlight ON orOFF.

LED INDICATION:

ON OFFGreen Orange

(2) Adjusting Brightness of Backlight

Press the DIA Bright/Dark key to adjust the brightness ofthe backlight.

Page 36: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

IV Other Operations

2 Adjusting LCD Contrast

- 35 -

IV

2 Adjusting LCD Contrast

1 To adjust the contrast of the LCD, first remove the rearpanel of the remote control pad, then set DIP switch No. 9to ON.

2 Press the DIA Bright/Dark key to adjust the contrast, justlike adjusting backlight brightness.

3 After completing the adjustment, set DIP switch No. 9 toOFF and reattach the rear panel to the remote control pad.

Page 37: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

1 Changing Operation Mode of Fine Focus Knob

- 36 -

VOperation and Indication Functions

1 Changing Operation Mode of Fine Focus Knob

Change the operation mode of the fine focus knob used for up/down focus motion.

[Operation Mode and the travel distance]

Coarse = 100 µm/rot. Middle = 50 µm/rot. Fine = 25 µm/rot.

The fine focus knob is set to the coarse mode by default.

Note:

The travel distance per rotation of the fine focus knob in each operation mode is a rough guide; precise accuracy isnot guaranteed. The travel distance will vary, depending on the speed at which the fine focus knob is turned.When the TE2000-U or the TE2000-S is used, the up/down focus motion cannot be controlled electrically and youcannot change the operation mode.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Fine Focus Knob” using the key or the keyand then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Select a travel distance (operation mode) using the PREVor NEXT key.

FINE FOCUS KNOB

[ 25]μm/rot.

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 38: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

2 Setting Indication on Operation Screen

- 37 -

V

2 Setting Indication on Operation Screen

The OPERATION screen indicates a variety of information, such as the current electrical component positions andinformation on the currently attached optical members.

This function turns indication of such information ON or OFF.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Remote Control Pad” using the key or the key and then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Turn indication of the information for the objective ON orOFF using the PREV or NEXT key.

REMOTE CONTROL PAD

DisplayObjective    :[ON ]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[ON ]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

4 Select an item using the key or the key and turnindication ON or OFF using the PREV or NEXT key.Repeat this step to set all of the items for which you wantto edit settings.

REMOTE CONTROL PAD

DisplayObjective    :[OFF]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[ON ]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

5 Press the MODE key to end settings and to return to theOPERATION screen.

Page 39: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

3 Setting Buzzer ON/OFF

- 38 -

3 Setting Buzzer ON/OFF

Turn ON or OFF the buzzer confirmation that sounds when a key is pressed.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Remote Control Pad” using the key or the key and then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the cursor to “Buzzer” using the key or the key.

Turn the buzzer ON or OFF using the PREV or NEXT key.

REMOTE CONTROL PAD

DisplayObjective    :[ON ]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[ON ]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Note:Even if the buzzer is set to “ON.”, when the remote control is set to “OFF”, the buzzer won’t sound when a key ispressed on the OPERATION screen.

Page 40: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

4 Setting Remote Control ON/OFF

- 39 -

V

4 Setting Remote Control ON/OFF

Turn remote-control operation of the electrical components using the remote control pad ON or OFF.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Remote Control Pad” using the key or the key and then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the cursor to “Remote Control” using the key orthe key.

Turn remote-control operation ON or OFF using the PREVor NEXT key.

REMOTE CONTROL PAD

DisplayObjective    :[ON ]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[ON ]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 41: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

5 Turning ON/OFF LED with LCD ON/OFF Key

- 40 -

5 Turning ON/OFF LED with LCD ON/OFF Key

Enable the LCD ON/OFF key to turn the LED ON or OFF.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Remote Control Pad” using the key or the key and then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the cursor to “LCD” using the key or the key.

Select “OFF” using the PREV or NEXT key.

If “OFF” is selected, when the LCD backlight is turned offwith LCD ON/OFF key, the LED is also turned OFF. The“LCD” is set to “ON” by default.

REMOTE CONTROL PAD

DisplayObjective    :[ON ]Condenser    :[ON ]Filter       :[ON ]Exciter      :[ON ]Barrier      :[ON ]

 DIA Lamp     :[ON ]Z-Position   :[ON ]

Buzzer        :[ON ]

Remote Control:[ON ]

LCD           :[OFF]

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 42: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

6 Setting Communications Delimiter of External Device

- 41 -

V

6 Setting Communications Delimiter of External Device

Set the delimiter for communications between the HUB controller and an external device.

[Delimiter]

CR LF

CR is selected by default.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “More” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Communication Delimiter” using the key or the key and then press the [ ] key.

MORERotation Stopper

Fine Focus Knob

Other Unit

Remote Control Pad

Shutter Key

Foot Switch

Communication Delimiter

3 Move the cursor using the key or the key to select“CR” or “LF.”

Note:Check the communications delimiter with the instructionmanual of the external device.

COMMUNICATION DELIMITERFOR EXTERNAL UNIT

CR     LF

4 Press the MODE key to end the setting and to return to theOPERATION screen.

Page 43: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 42 -

7 Registering Data for Optical Members

(1) Objective

You can register optional data for up to nine objectives.

If an attached objective is not listed in the presets, register the objective.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Objective” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Edit” using the key or the key and then pressthe [ ] key.

OBJECTIVESeries   Mag.

1[P Fluor   ][DL       4x      ]

2[P Fluor   ][DL       10x     ]

3[P Fluor   ][ELWD DM  20x     ]

4[P Fluor   ][ELWD DM  40x     ]

5[P Fluor   ][ELWD DLL 60x     ]

6[P Fluor   ][DLL      100x    ]

 Edit

3 Select one location out of nine locations to register theoptional data using the PREV or NEXT key.

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[――――――――――]

Mag.    :[0  ]x

N.A.    :[0    ]

W.D.    :[0    ]

Type    :[――――――――――]

4 Move the cursor to “Series” using the key or the key.

Select the series of the objective using the PREV or NEXTkey.

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[Other     ]

Mag.    :[1  ]x

N.A.    :[0。001]

W.D.    :[0。01 ]

Type    :[dry       ]

5 Move the cursor to “Mag.” using the key or the key.

Select the magnification of the objective using the PREVor NEXT key.

If no entry is made in “Series” (indicated as “--------”), thecursor will not move to “Mag.”

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[Other     ]

Mag.    :[10 ]x

N.A.    :[0。001]

W.D.    :[0。01 ]

Type    :[dry       ]

6 Move the cursor to “N.A.” using the key or the key.

Enter the numerical aperture using the numerical keys andthe key or the key within the range between 0.001and 2.000.

If the entry exceeds the upper limit or falls below the lowerlimit, the N.A. is automatically set to the upper or lowerlimit.

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[Other     ]

Mag.    :[10 ]x

N.A.    :[0.25 ]

W.D.    :[0。01 ]

Type    :[dry       ]

Page 44: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 43 -

V

7 Move the cursor to “W.D.” using the key or the key.

Using the numerical keys and the key or the key,enter the working distance within the range between 0.01and 40.00.

If the entry exceeds the upper limit or falls below the lowerlimit, the W.D. is automatically set to the upper or lowerlimit.

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[Other     ]

Mag.    :[10 ]x

N.A.    :[0.25 ]

W.D.    :[6.2  ]

Type    :[dry       ]

8 Move the cursor to “Type” using the key or the key.

Select the type of objective using the PREV or NEXT key.

EDIT OBJECTIVE DATA

Location:[Other1]

Series  :[Other     ]

Mag.    :[10 ]x

N.A.    :[0.25 ]

W.D.    :[6.2  ]

Type    :[dry       ]

9 To continue registration of optional data, move the cursorto “Location” using the key or the key and repeatsteps 3 - 8 above.

EDIT OBJECTIVE DATA

Location:[Other2]

Series  :[――――――――――]

Mag.    :[0  ]x

N.A.    :[0    ]

W.D.    :[0    ]

Type    :[――――――――――]

10 Press the MODE key to end the setting and to return to theOPERATION screen.

To delete registered data

1) Move the cursor to “Location” using the key or the key and select data to delete using PREV or NEXTkey.

2) Press CLR key.

Page 45: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 44 -

(2) Condenser Module

You can register optional data for up to nine condenser modules.

If an attached condenser module is not listed in the presets, register the condenser module.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Condenser” using the key or the key andthen press the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Edit” using the key or the key and then pressthe [ ] key.

CONDENSERName

1[BF        ]

2[PhL       ]

3[Ph1       ]

4[Ph2       ]

5[Ph3       ]

 Edit

3 Select one location out of nine locations to register theoptional data using the PREV or NEXT key.

EDIT CONDENSER DATA

Location:[Other1]

Name    :[          ]

4 Move the cursor to “Name” using the key or the key.

Using the numerical/alphabetical keys and the key orthe key, type in the name of the condenser module.

You can type in up to 10 characters.

EDIT CONDENSER DATA

Location:[Other1]

Name    :[ABC       ]

5 To continue registration of optional data, move the cursorto “Location” using the key or the key and repeatsteps 3 and 4 above.

EDIT CONDENSER DATA

Location:[Other2]

Name    :[          ]

6 Press the MODE key to end the setting and to return to theOPERATION screen.

To delete registered data

1) Move the cursor to “Location” using the key or the key and select data to delete using PREV or NEXTkey.

2) Press CLR key.

Page 46: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 45 -

V

(3) Filter Block

You can register optional data for up to nine filter blocks.

If an attached filter block is not listed in the presets, register the filter block.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Filter” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Edit” using the key or the key and then pressthe [ ] key.

FILTERName

1[UV-1A     ] EX : EX365/10

2[V-1A      ] DM : DM400

3[BV-1A     ] BA : BA400

4[B-1A      ]

5[G-1A      ]

6[DIA       ]

 Edit

3 Select one location out of nine locations to register theoptional data using the PREV or NEXT key.

EDIT FILTER DATA

Location:[Other1]

Name    :[          ]

EX      :[――――――――――]Edit

DM      :[――――――――――]Edit

BA      :[――――――――――]Edit

4 Move the cursor to “Name” using the key or the key.

Type in the name of the filter block using thenumerical/alphabetical keys and the key or the key.

You can type in up to 10 characters.

EDIT FILTER DATA

Location:[Other1]

Name    :[ABC       ]

EX      :[――――――――――]Edit

DM      :[――――――――――]Edit

BA      :[――――――――――]Edit

5 Move the cursor to “EX” using the key or the key.

Select an excitation filter of the filter block using the PREVor NEXT key.

If the excitation filter is not listed in the presets, register theoptional data for the excitation filter.

To register the optional data, move to “Edit” using the key or the key and then press the key.

For details, see steps 3 forward for V-7 (5) ExcitationFilter.

EDIT FILTER DATA

Location:[Other1]

Name    :[ABC       ]

EX      :[EX365/10  ]Edit

DM      :[――――――――――]Edit

BA      :[――――――――――]Edit

6 Move the cursor to “DM” using the key or the key.

Select a dichroic mirror for the filter block using the PREVor NEXT key. If the dichroic mirror is not listed in thepresets, register the optional data for the dichroic mirror.

To register the optional data, move to “Edit” using the key or the key and then press the key.

For details, see V-7 (4) Dichroic Mirror.

EDIT FILTER DATA

Location:[Other1]

Name    :[ABC       ]

EX      :[EX365/10  ]Edit

DM      :[DM400     ]Edit

BA      :[――――――――――]Edit

Page 47: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 46 -

7 Move the cursor to “BA” using the key or the key.

Select a barrier filter for the filter block using the PREV orNEXT key. If the barrier filter is not listed in the presets,register the optional data for the barrier filter.

To register the optional data, move to “Edit” using the key or the key and then press the key.

For details, see steps from 3 forward of V-7 (6) BarrierFilter.

EDIT FILTER DATA

Location:[Other1]

Name    :[ABC       ]

EX      :[EX365/10  ]Edit

DM      :[DM400     ]Edit

BA      :[BA400     ]Edit

8 To continue registration of optional data, move the cursorto “Location” using the key or the key and repeatsteps 3 - 7 above.

EDIT FILTER DATA

Location:[Other2]

Name    :[          ]

EX      :[――――――――――]Edit

DM      :[――――――――――]Edit

BA      :[――――――――――]Edit

9 Press the MODE key to end the setting and to return to theOPERATION screen.

To delete registered data

1) Move the cursor to “Location” using the key or the key and select data to delete using PREV or NEXTkey.

2) Press CLR key.

Page 48: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 47 -

V

(4) Dichroic Mirror

You can register optional data for up to nine dichroic mirrors.

If an attached dichroic mirror is not listed in the presets, register the dichroic mirror.

1 Select “Edit” using the key or the key and then pressthe [ ] key.

EDIT FILTER DATA

Location:[Other1]

Name    :[ABC       ]

EX      :[EX365/10  ]Edit

DM      :[DM400     ]Edit

BA      :[――――――――――]Edit

2 Select one location out of nine locations to register theoptional data using the PREV or NEXT key.

EDIT DM DATA

Location:[Other1]

Name    :[          ]

3 Move the cursor to “Name” using the key or the key .

Using the numerical/alphabetical keys and the key orthe key, type in the wavelength of the dichroic mirror.You can type in up to 10 characters.

EDIT DM DATA

Location:[Other1]

Name    :[DM410     ]

4 To continue registration of optional data, move the cursorto “Location” using the key or the key and repeatsteps 2 and 3 above.

EDIT DM DATA

Location:[Other2]

Name    :[          ]

5 Press the MODE key to end the setting and to return to theOPERATION screen.

To delete registered data

1) Move the cursor to “Location” using the key or the key and select data to delete using PREV or NEXTkey.

2) Press CLR key.

Page 49: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 48 -

(5) Excitation Filter

You can register optional data for up to nine excitation filters.

If an attached excitation filter is not listed in the presets, register the excitation filter.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Exciter” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Edit” using the key or the key and then pressthe [ ] key.

EXCITER

Name1[EX465-495 ]5[――――――――――]

2[EX460-500 ]6[――――――――――]

3[――――――――――]7[――――――――――]

4[――――――――――]8[DIA       ]

 Edit

3 Select a location to register the optional data using thePREV or NEXT key.

EDIT EX DATA

Location:[Other1]

Name    :[          ]

4 Move the cursor to “Name” using the key or the key .

Using the numerical/alphabetical keys and the key orthe key, type in the wavelength of the excitation filter.You can type in up to 10 characters.

EDIT EX DATA

Location:[Other1]

Name    :[EX350-380 ]

5 To continue registration of optional data, move the cursorto “Location” using the key or the key and repeatsteps 3 and 4 above.

EDIT EX DATA

Location:[Other2]

Name    :[          ]

6 Press the MODE key to end the setting and to return to theOPERATION screen.

To delete registered data

1) Move the cursor to “Location” using the key or the key and select data to delete using PREV or NEXTkey.

2) Press CLR key.

Page 50: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

V Operation and Indication Functions

7 Registering Data for Optical Members

- 49 -

V

(6) Barrier Filter

You can register optional data for up to nine barrier filters.

If an attached barrier filter is not listed in the presets, register the barrier filter.

1 Press the MODE key, and the SETTING MENU screen willappear on the LCD.

Select “Barrier” using the key or the key and thenpress the [ ] key.

SETTING MENUObjective

Condenser

Filter

Exciter

Barrier

Combination(Obj.>Condenser)

Combination(Filter>EX/BA)

More

2 Select “Edit” using the key or the key and then pressthe [ ] key.

BARRIER

Name1[BA515-555 ]5[――――――――――]

2[BA510     ]6[――――――――――]

3[BA510-560 ]7[――――――――――]

4[――――――――――]8[DIA       ]

 Edit

3 Select one location out of nine locations to register theoptional data using the PREV or NEXT key.

EDIT BA DATA

Location:[Other1]

Name    :[          ]

4 Move the cursor to “Name” using the key or the key.

Using the numerical/alphabetical keys and the key orthe key, type in the wavelength of the barrier filter.You can type in up to 10 characters.

EDIT BA DATA

Location:[Other1]

Name    :[BA410     ]

5 To continue registration of optional data, move the cursorto “Location” using the key or the key and repeatsteps 3 and 4 above.

EDIT BA DATA

Location:[Other2]

Name    :[          ]

6 Press the MODE key to end the setting and to return to theOPERATION screen.

To delete registered data

1) Move the cursor to “Location” using the key or the key and select data to delete using PREV or NEXTkey.

2) Press CLR key.

Page 51: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

1 Indicating Program Version and Checksum

- 50 -

VIMaintenance Menu

1 Indicating Program Version and Checksum

Indicate the program version of the remote control pad and the HUB controller and the checksum on the LCD.

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

2 Press the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

3 The program versions and the checksum appear on theLCD.

VERSION

T-RCP REMOTE CONTROL PADV1.00 Sum1234h

T-HUBC HUB CONTROLLERMAIN-cpu: V1.00 Sum5678hSUB-cpu : V1.00 SumABCDh

4 Press the MODE key to return to the OPERATION screen.

Page 52: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

2 Testing Electrical Component Operations

- 51 -

VI

2 Testing Electrical Component Operations

Test the operations of the electrical components connected to the HUB controller.

There are two test modes.

All-test mode: Test all the electrical components connected.

Individual test mode: Select a component and test the component.

(1) All-Test Mode

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, pressthe MODE key.

Select “TEST 1” using the key or the key and releasethe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 On the TEST 1 screen, the arrow “→” is pointed at “All-Test.” Press the [ ] key.

“→” moves to “OBJECTIVE” and the test will begin.

TEST1           1/5→ ALL TEST

OBJECTIVE              1            [   ]              2            [   ]              3            [   ]              4            [   ]              5            [   ]              6            [   ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select item  and press enter to start.

3 The test results are indicated in [ ] as “OK” or “NG.” TEST1           1/5 ALL TEST

→ OBJECTIVE              1            [OK ]              2            [   ]              3            [   ]              4            [   ]              5            [   ]              6            [   ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

Page 53: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

2 Testing Electrical Component Operations

- 52 -

4 After the test of one component is completed, “→” movesto the next component and the test of the component willbegin.The components not connected to the HUB controller willbe skipped.

TEST1           2/5

 EPI-FILTER              1             [OK]              2             [OK]              3             [OK]              4             [OK]              5             [OK]              6             [OK]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

TEST1           3/5

 LIGHT PATH              1             [OK]              2             [OK]              3             [OK]              4             [OK]              5             [OK]  CONDENSER              Right         [OK]              Left          [OK]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

TEST1           4/5

 EXCITER              Right         [  ]              Left          [  ]  BARRIER              Right         [  ]              Left          [  ]

  ANALYZER                  [OK]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

5 When all the components have been tested, “Complete!”appears and remains displayed for about one second.

TEST1           5/5

  DIA LAMP              REMOTE        [OK]              ON/OFF        [OK]              ADJ-up        [OK]              ADJ-down      [OK]→ Shutter              A             [OK]              B             [OK]

  CAMERA EXP.               [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

6 Press the MODE key to return to the OPERATION screen.

How to interrupt the all-test

Press the [ ] key to interrupt the test.

When the test is interrupted, the following message appears after the test of the currently tested component iscompleted.

Press enter to restart.

To resume the test, press the [ ] key again.

How to end the all-test midway

Press the [ ] key to interrupt the test.

Press the MODE key.

How to check test results again

Move “→” using the key or the key to change screens.

This operation is also available while the test is interrupted.

Page 54: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

2 Testing Electrical Component Operations

- 53 -

VI

(2) Individual Test Mode

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “TEST 1” using the key or the key and pressthe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 Move “→” using the key or the key to select anelectrical component to test.

Press the [ ] key.

The test of the selected component begins.If a component is not connected to the HUB controller, youcannot select the component.

TEST1           1/5 ALL TEST

→ OBJECTIVE              1             [  ]              2             [  ]              3             [  ]              4             [  ]              5             [  ]              6             [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

TEST1           2/5

 EPI-FILTER              1             [  ]              2             [  ]              3             [  ]              4             [  ]              5             [  ]              6             [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

TEST1           3/5

 LIGHT PATH              1             [  ]              2             [  ]              3             [  ]              4             [  ]              5             [  ]  CONDENSER              Right         [  ]              Left          [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

TEST1           4/5

 EXCITER              Right         [  ]              Left          [  ]  BARRIER              Right         [  ]              Left          [  ]

  ANALYZER                  [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

TEST1           5/5

  DIA LAMP              REMOTE        [  ]              ON/OFF        [  ]              ADJ-up        [  ]              ADJ-down      [  ] Shutter

              A             [  ]              B             [  ]

  CAMERA EXP.               [  ]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running ALL TEST...  Press enter to pause.

3 When the test is completed, “Complete!” appears andremains displayed for about one second.

TEST1           1/5 ALL TEST

→ OBJECTIVE              1             [OK]              2             [OK]              3             [OK]              4             [OK]              5             [OK]              6             [OK]

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

4 Press the MODE key to return to the OPERATION screen.

How to check test results again

Move “→” using the key or the key to change screens.

Page 55: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

3 Testing Hardware of Remote Control Pad

- 54 -

3 Testing Hardware of Remote Control Pad

The following tests are available to evaluate the hardware of the remote control pad.

Key input test Checks to see if the input of each key is read correctlyLED ON/OFF test Checks to see if all the LEDs are controlled correctlyLCD indication test Checks to see if the LCD indication is controlled correctly and tests the quality of the LCD

(1) Key Input Test

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “TEST 2” using the key or the key and pressthe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 Move “→” using the key or the key to select “KEY.”

Press the [ ] key.

3 Press the keys (except the [ ] key) one by one.

When the input for a key is read correctly, the key positionin the key layout on the screen will be indicated in black.

4 Press the [ ] key to end the key input test.

5 Press the MODE key to return to OPERATION screen.

Page 56: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

3 Testing Hardware of Remote Control Pad

- 55 -

VI

(2) LED ON/OFF Test

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “TEST 2” using the key or the key and pressthe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 Move “→” using the key or the key to select “LED.”

Press the [ ] key. The LED ON/OFF test will begin.

All the LEDs light green and then go out one by one inintervals of about 0.5 seconds, except that the ANALYZERIN/OUT key and LCD ON/OFF key light green, turn toorange, and then go out.

3 Press the [ ] key to end the LED ON/OFF test.

4 Press the MODE key to return to OPERATION screen.

(3) LCD Indication Test

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “TEST 2” using the key or the key and pressthe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 Move “→” using the key or the key to select “LCD.”

Press the [ ] key. The LCD indication test will begin.

The whole screen turns black and then white repeatedly atintervals of about 3 seconds.

3 Press the [ ] key to end the LCD indication test.

4 Press the MODE key to return to OPERATION screen.

Page 57: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

4 Initializing Focus Motion Potentiometer

- 56 -

4 Initializing Focus Motion Potentiometer

The potentiometer can be initialized at each reference point to detect absolute position of the focusing mechanism.

Use this function after changing the combination of the microscope and HUB controller.

Initialization takes about 3 minutes to be completed. During initialization, never touch the coarse or fine focusknobs.

1 Remove the rear panel of the remote control pad and setDIP switch No.8 to ON.

2 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “TEST 2” using the key or the key and pressthe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

3 After the TEST 2 screen appears, press the Z-AXI. RESETkey.

Initialization of the focus motion potentiometer begins.

The initialization takes about 3 minutes at maximum.

4 When the initialization is completed, a “Complete!”message appears and remains displayed for about onesecond.Also, the HUB controller sounds the buzzer.

Note:The buzzer sounds as follows:

Upon normal completion: a peepUpon failure of normal completion: four peepings

5 Turn off the power to the HUB controller.

Set DIP switch No.8 to OFF and reattach the rear panel tothe remote control pad.

Use the coarse focus knob to lower the focusingmechanism 2 to 3 mm, then turn on the HUB controller.

Turning the coarse focus knob halfway lowers the focusingmechanism about 2.5 mm.

Note:After initialization, the focusing mechanism islocated at the highest position.

Page 58: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

5 Resetting Remote Control Pad Settings to Default Values

- 57 -

VI

5 Resetting Remote Control Pad Settings to Default Values

Use this function to reset the settings stored in the memory of the remote control pad back to the default values(See Table 1).

The settings of the LCD brightness and contrast are also reset to their default values.

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “PAD Data” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the[ ] key.

If you select “No” and press the [ ] key, the display willreturn to the MAINTENANCE MENU screen.

PAD DATA

Initialize PAD data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press enter to start.

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

PAD DATA

Initialize PAD data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

[Table 1] Settings Stored in Remote Control Pad Memory and Default Settings

Menu option Default settingSETTING MENU → More → Remote Control PadDisplay Objective ON

Condenser ONFilter ONExciter ONBarrier ONDIA Lamp ON

LCD ON

Page 59: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

6 Resetting HUB Controller Settings to Default Values

- 58 -

6 Resetting HUB Controller Settings to Default Values

Use this function to reset the data stored in the memory of the HUB controller back to the default values. Forexample, the memory stores information on the attached objectives, registered optional data of optical members,and information on the interlocking of the revolving nosepiece and the condenser module changeover (See Table2).

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “HUB Data” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

HUB DATA

Initialize HUB data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press enter to start.

3 When the following message appears, turn on the poweragain.

“Please re-switch on a power supply.”

HUB DATA

Initialize HUB data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Running...

[Table 2] Settings Stored in HUB Controller Memory and Default Settings

Menu option Default settingSETTING MENU

Addresses 16 Not attachedObjectiveOptional data Not registeredAddresses 15 Not attachedCondenserOptional data Not registeredAddresses 16 Not attachedOptional data (filter block) Not registered

Filter

Optional data (dichroic mirror) Not registeredAddresses 18 Not attachedExciterOptional data (excitation filter) Not registeredAddresses 18 Not attachedBarrierOptional data (barrier filter) Not registered

Combination(OBJ.>... Addresses 16 Not interlockedCombination(Filter. Addresses 16 Not interlocked

Page 60: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 59 -

VI

7 Reset Control Data to Default Values

Use this function to reset the control data stored in the memory of the HUB controller to their default values. Usethis function when an electrical component doesn’t work properly or has been replaced with a new component.

The control data of the electrical components are initialized either in batches or individually. After replacing anelectrical component, initialize the control data of the replaced component only.

(1) All Components

When the control data of all the electrical components are initialized in batches, the settings in the table below willalso be initialized.

[Table 3]

Menu option Default settingSETTING MENU → MoreRotation Stopper OFFFine Focus Knob Coarse = 100 µmOther Unit OFF

Display Z-Position OFFBuzzer ON

Remote Control Pad

Remote Control ONShutter A OFFShutter KeyShutter B OFF

Foot Switch UnusedCommunication... CR

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “All Control Data” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the[ ] key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize all control data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

Page 61: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 60 -

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize all control data ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

(2) Revolving Nosepiece

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “Revolving Nosepiece” using the key or the key and press the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the[ ] key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data  for revolving nosepiece ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data  for revolving nosepiece ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 62: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 61 -

VI

(3) Condenser Module

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “Condenser Module (Cassette)” using the key orthe key and press the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the[ ] key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data   for condenser cassette ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data   for condenser cassette ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 63: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 62 -

(4) Optical Path

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “Light Path” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the[ ] key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data           for light path ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data           for light path ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 64: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 63 -

VI

(5) Filter Block

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “Filter Block” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the [ ]key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data         for filter block ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data         for filter block ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 65: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 64 -

(6) Excitation Filter

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “Exciter Filter” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data       for exciter filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data       for exciter filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 66: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 65 -

VI

(7) Barrier Filter

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “Barrier Filter” using the key or the key andpress the [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data       for barrier filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data       for barrier filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 67: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VI Maintenance Menu

7 Reset Control Data to Default Values

- 66 -

(8) Auxiliary Filter

1 Turn on the power to the HUB controller while holdingdown the MODE key.

After the MAINTENANCE MENU screen appears, releasethe MODE key.

Select “AUX Filter” using the key or the key and pressthe [ ] key.

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

2 A confirmation message appears.

Select “Yes” using the key or the key and press the key.

If you select “No” and press the key, the display willreturn to the MAINTENANCE MENU screen.

CONTROL DATA

Initialize control data           for AUX filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Select Yes  and press ENTER to start。

3 When the initialization is completed, “Complete!” appearsand remains displayed for about one second. TheMAINTENANCE MENU screen then appears.

CONTROL DATA

Initialize control data           for AUX filter ?

Yes   No

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Complete!

MAINTENANCE MENU

1.Version2.TEST13.TEST24.PAD Data5.HUB Data6.Control Data

a.All Control Datab.Revoiving Nosepiecec.Condenser Cassetted.Light Pathe.Filter Blockf.Exciter Filterg.Barrier Filterh.AUX Filter

Page 68: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VII Troubleshooting

1 Remote control pad doesn’t start up. (Title screen appears, and the screen freezes.)

- 67 -

VII

VIITroubleshooting

1 Remote control pad doesn’t start up. (TITLE screen appears,and the screen freezes.)

Cause Action

The remote control pad is not connected to theHUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the remotecontrol pad from the HUB controller and connect it again.See “Introduction-3 Connection of Remote Control Pad and HUBController.”

2 Electrical component doesn’t operate.

Cause Action

The electrical component is not connected to theHUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the electricalcomponent from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

The remote control operation is set to “OFF.”Set the remote control operation to “ON.”See “V-4 Setting Remote Control ON/OFF.”

3 Revolving nosepiece doesn’t rotate at a certain position ofthe fine focus knob.

Cause Action

The revolving nosepiece rotation is restricted.Cancel the rotation restriction.See “III-3 Restricting Rotation of Revolving nosepiece.”

4 Dia-illumination lamp cannot be turned ON or OFF.

Cause Action

The dia-illumination lamp power supply is notconnected to the HUB controller correctly.

Turn OFF the power of the HUB controller. Disconnect the dia-illuminationlamp power supply from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

EXTERNAL switch of the dia-illumination lamppower supply is OFF.

Turn ON the EXTERNAL switch of the dia-illumination lamp power supply.See “I-1 (2) Turning ON EXTERNAL Switch of Dia-Illumination LampPower Supply.”

The remote control operation is set to “OFF.”Set the remote control operation to “ON.”See “V-4 Setting Remote Control ON/OFF.”

Page 69: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VII Troubleshooting

5 Voltage of dia-illumination lamp cannot be adjusted.

- 68 -

5 Voltage of dia-illumination lamp cannot be adjusted.

Cause Action

The dia-illumination lamp power supply is notconnected to the HUB controller correctly.

Turn OFF the power of the HUB controller. Disconnect the dia-illumination lamp power supply from the HUB controller and connect itagain.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

EXTERNAL switch of the dia-illumination lamppower supply is OFF.

Turn ON the EXTERNAL switch of the dia-illumination lamp power supply.See “I-1 (2) Turning ON EXTERNAL Switch of Dia-Illumination LampPower Supply.”

The remote control operation is set to “OFF.”Set the remote control operation to “ON.”See “V-4 Setting Remote Control ON/OFF.”

The dia-illumination lamp is OFF.Turn ON the dia-illumination lamp.See “II-9 (1) Turning Lamp ON/OFF.”

6 Opening/closing status of Uniblitz shutter doesn’t coincidewith LED indication.

Cause ActionThe Uniblitz shutter controller is not in the N.C.mode.

Place the Uniblitz shutter controller into N.C. mode.See “I-1 (3) Placing Uniblitz Shutter Controller into N.C. Mode.”

7 Objective information is not indicated.

Cause Action

The revolving nosepiece is not connected to theHUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the revolvingnosepiece from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

Indication of information for the objective is OFF.Turn ON the indication function for the information on the objective.See “V-2 Setting Indication on Operation Screen.”

8 Condenser module information is not indicated.

Cause Action

The condenser module changeover is notconnected to the HUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the condensermodule changeover from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

Indication of the information for the condensermodule is OFF.

Turn ON the indication function for the information on the condensermodule.See “V-2 Setting Indication on Operation Screen.”

9 Filter block information is not indicated.

Cause Action

The filter block changeover is not connected to theHUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the filter blockchangeover from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

Indication of the information for the filter block isOFF.

Turn ON the indication function for the information for the filter block.See “V-2 Setting Indication on Operation Screen.”

Page 70: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510

VII Troubleshooting

10 Excitation filter information is not indicated.

- 69 -

VII

10 Excitation filter information is not indicated.

Cause Action

The excitation filter changeover is not connected tothe HUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the excitationfilter changeover from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

Indication of the information for the excitation filteris OFF.

Turn ON the indication function for the information on the excitation filter.See “V-2 Setting Indication on Operation Screen.”

11 Barrier filter information is not indicated.

Cause Action

The barrier filter changeover is not connected to theHUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the barrier filterchangeover from the HUB controller and connect it again.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

Indication of the information for the barrier filter isOFF.

Turn ON the indication function for the information on the barrier filter.See “V-2 Setting Indication on Operation Screen.”

12 Dia-illumination lamp voltage is not indicated.

Cause Action

The dia-illumination lamp power supply is notconnected to the HUB controller correctly.

Turn OFF the power to the HUB controller. Disconnect the dia-illumination lamp power supply from the HUB controller and connect itagain.See “I-1 (1) Connecting Electrical Components to HUB Controller.”

The EXTERNAL switch of the dia-illumination lamppower supply is OFF.

Turn ON the EXTERNAL switch of the dia-illumination lamp power supply.See “I-1 (2) Turning ON EXTERNAL Switch of Dia-Illumination LampPower Supply.”

The remote control operation is set to “OFF.”Set the remote control operation to “ON.”See “V-4 Setting Remote Control ON/OFF.”

13 Z-axis coordinates are not indicated.

Cause ActionThe HUB controller is not connected to themicroscope correctly.

Turn OFF the power to the HUB controller.Check that the HUB controller is correctly connected to the microscope.

Indication of z-axis coordinates is OFF.Turn ON the indication function for z-axis coordinates.See “V-2 Setting Indication on Operation Screen.”

14 Buzzer doesn’t sound when a key is pressed.

Cause Action

The buzzer is OFF.Turn the buzzer ON.See “V-3 Setting Buzzer ON/OFF.”

Page 71: T-RCP Remote Control Pad Instructions · PDF file2 Appearance of Remote Control Pad.....4 3 Connection of Remote Control Pad and HUB Controller ... 2[BA510