Antibody-Modified Conduits for Extracorporeal Selective Cytokine Filtration in Sepsis by Morgan K. Moroi Submitted to the Department of Mechanical Engineering in partial fulfillment of the requirements for the degree of Bachelor of Science in Engineering as Recommended by the Department of Mechanical Engineering at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY June 2016 @ Massachusetts Institute of Technology 2016. All rights reserved. Author .... Signature redacted 6' Department of Mechanical Engineering May 6, 2016 Cetfidby.Signature redacted Certified by... ....... ....... Robert S. Langer David H. Koch Institute Professor Thesis Supervisor Accepted bySignature redacted by... - -- - tu r -- e - c e d ............. Anette Hosoi MASSACHUSETTS INSTITUTE Professor of Mechanical Engineering OF TECHNOLOGY Undergraduate Officer JUL 08 2016 LIBRARIES
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
by
Morgan K. Moroi
Submitted to the Department of Mechanical Engineering in partial
fulfillment of the requirements for the degree of
Bachelor of Science in Engineering as Recommended by the Department
of Mechanical Engineering
at the
Author .... Signature redacted
Cetfidby.Signature redactedCertified by... ....... .......
Thesis Supervisor
Accepted bySignature redactedby... - -- - tu r -- e - c e d
............. Anette Hosoi
MASSACHUSETTS INSTITUTE Professor of Mechanical Engineering OF
TECHNOLOGY Undergraduate Officer
JUL 08 2016
Cytokine Filtration in Sepsis
Morgan K. Moroi
Submitted to the Department of Mechanical Engineering on May 6,
2016, in partial fulfillment of the
requirements for the degree of Bachelor of Science in Engineering
as Recommended by the Department of
Mechanical Engineering
Abstract
Sepsis kills millions of people worldwide each year and occurs when
microorganisms enter the bloodstream of an infected host. The
presence of microorganisms in the bloodstream triggers the body to
produce many inflammatory proteins, known as cytokines, that cause
damage to blood vessels and vital organs. This leads to cap- illary
leak, failing organs, and often death. We have developed a novel
approach to modulate the inflammatory response, using
antibody-modified conduits (AMCs) to filter harmful cytokines
selectively from the circulation and in a time-specific manner.
Here, we characterize variables that affect AMC performance to
determine optimal AMC conditions for later use downstream.
Thesis Supervisor: Robert S. Langer Title: David H. Koch Institute
Professor
3
4
Acknowledgments
The author gratefully acknowledges Dr. J. Brian McAlvin, Dr. Daniel
S. Kohane,
and Dr. Robert S. Langer for their guidance throughout the
completion of this work.
The author would also like to thank all family and friends for
their continual support.
5
6
Contents
2.1 M aterials . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . 17
2.2.1 Synthesis and Dialysis of Furan-Functionalized Antibody . . .
18
2.2.2 Formation of Antibody-Modified Conduits . . . . . . . . . . .
18
2.3 In Vitro Demonstration of Selective Cytokine Filtration . . . .
. . . . 21
2.4 Influence of Flow Rate on Elimination Kinetics . . . . . . . .
. . . . 22
2.5 Influence of Antibody Surface Packing Density on Elimination
Kinetics 23
2.6 Influence of Surface Area to Volume Relationships on
Elimination Ki-
netics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . 23
2.7 Influence of PEG Spacer Length on Elimination Kinetics . . . .
. . . 23
2.8 Determination of Cytokine Binding Capacity . . . . . . . . . .
. . . . 24
2.9 Statistics . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . 24
3 Results 25
3.1 In Vitro Demonstration of Selective Cytokine Filtration . . . .
. . . . 25
3.2 Influence of Flow Rate on Elimination Kinetics . . . . . . . .
. . . . 26
3.3 Influence of Antibody Surface Packing Density on Elimination
Kinetics 27
3.4 Influence of Surface Area to Volume Relationships on
Elimination Ki-
netics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . 28
3.5 Influence of PEG Spacer Length on Elimination Kinetics . . . .
. . . 29
7
4 Discussion 31
5 Conclusion 35
6 References 37
1-1 Pathophysiology of the systemic inflammatory response in sepsis
. . . 14
2-1 Schema of formation of an antibody-modified conduit . . . . . .
. . . 20
2-2 Experimental setup for testing selective TNF-a filtration with
infliximab-
m odified conduits . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . 21
3-1 Type of cytokine filtered was determined by the selection of
AMC
surface antibody . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 26
3-2 TNF-a filtration by infliximab-modified conduits, where the BSA
TNF-
a solution was circulated through AMCs at varying flow rates . . .
. 27
3-3 TNF-a filtration by infliximab-modified conduits prepared with
vary-
ing infliximab concentrations . . . . . . . . . . . . . . . . . . .
. . . . 28
area to volume ratios . . . . . . . . . . . . . . . . . . . . . . .
. . . . 29
3-5 TNF-a filtration by infliximab-modified conduits with varying
PEG
spacer lengths......... ............................... 30
lution) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . 30
2.1 PDMS tube specifications used to prepare antibody-modified
conduits. 19
11
12
Introduction
Septic shock affects millions of people worldwide every year and
remains the lead-
ing cause of death in critically ill patients in the United States
[1-4]. Sepsis occurs
when microorganisms (most commonly bacteria) enter the bloodstream
and cause a
dynamic, unregulated systemic inflammatory response [2] (Figure
1-1). The early
stage of sepsis is dominated by a surge of pro-inflammatory
cytokines, such as tumor
necrosis factor alpha (TNF-a) [5], interleukin 1 beta (IL-1#)
[5,6], interleukin 6 (IL-6)
[5], and vascular endothelial growth factor (VEGF) [7]. As sepsis
persists, a decrease
in pro-inflammatory cytokines and an increase in anti-inflammatory
cytokines occurs,
resulting in immunoparalysis [2,5]. Recovery from this
immunosuppressive state may
be associated with positive clinical outcomes [8].
13
0 to E
Figure 1-1: Pathophysiology of the systemic inflammatory response
in sepsis.
Investigators have tried for decades to modulate the immune
response during
sepsis in an effort to restore homeostasis [5]. Unfortunately, most
attempts, including
corticosteroids [5,9], antiendotoxin antibodies [5,10], tumor
necrosis factor antagonists
[5,11], interleukin-1-receptor antagonists [5,6], and ibuprofen
[12], have largely proven
unsuccessful. Additionally, efforts to non-specifically remove both
pro-inflammatory
and anti-inflammatory cytokines, by dialysis [13] or therapeutic
plasma exchange
(TPE) [14], have also failed to improved survival. One reason for
these failures may
be because the complex pathophysiology of sepsis changes over time.
For example,
many targeted immune therapies that are beneficial during the
pro-inflammatory
surge exacerbate the immunosuppression that is present in the later
phase of sepsis.
We presume that previous attempts to modulate the immune system
have been
unsuccessful due to prolonged inimunosuppression (via
anti-inflammatory therapies)
or non-s)ecific removal of cytokines that would have been better
left in the circu-
lation (via dialysis or TPE). Furthermore, we hypothesize that
successful immune
14
a time-specific manner. Pro-inflammatory cytokines should only be
removed during
the pro-inflammatory burst to prevent long-lasting immune
suppression. Similarly,
anti-inflammatory cytokines should only be removed during
immunoparalysis to aid
in the restoration of homeostasis.
To accomplish this, silicone tubes were individually surface
modified with antibod-
ies against a particular cytokine. Known as an antibody-modified
conduit (AMC),
each tube is capable of removing one type of cytokine from blood
that extracorpo-
really circulates through it. If installed in parallel and for
patient-specific durations
of time, AMCs may prevent multi-organ injury and mortality by
allowing for spe-
cific temporally controlled manipulations of the cytokine cascade
in sepsis. Here,
we specifically focus on the characterization of variables that
affect AMC filtration
performance.
15
16
with N-hydroxysuccinimide ester and maleimide moieties
(NHS-PEGa-Maleimide,
where n is the molecular weight of the PEG spacer) was purchased
from Rapp
Polymere (Tuebingen, Germany). 2-Furoic acid,
N-(3-Dimethylaminopropyl)-N '-
ethylcarbodiimide hydrochloride (EDC), and N-Hydroxysuccinimide
(NHS) were pur-
chased from Sigma-Aldrich (St. Louis, MO). Infliximab (Janssen
Biotech, Inc., Hor-
sham, PA) in distilled water (10 mg/mL) was donated by Boston
Childrens Hospital
(Boston, MA). Bevacizumab (Genentech, Inc., San Francisco, CA) in
distilled water
(25 mg/mL) was donated by Boston Childrens Hospital (Boston, MA).
Vascular en-
dothelial growth factor A (VEGF-A) and tumor necrosis factor alpha
(TNF-a) were
purchased from R&D Systems Inc. (Minneapolis, MN).
17
2.2.1 Synthesis and Dialysis of Furan-Functionalized Anti-
body
Furan with N-hydroxysuccinimide ester (NHS-furan) was synthesized
by dissolving 1
molar equivalent of furoic acid, 3 molar equivalents of EDC, and 3
molar equivalents
of NHS in dichloromethane (DCM). The aqueous solution was
continuously stirred for
13 hours at room temperature and purified using a CombiFlash@ Rf
150 purification
system (Teledyne Isco, Inc., Lincoln, NE) with a 40-gram high
performance silica
column (Teledyne Isco, Inc., Lincoln, NE). Elution fractions
containing the product
were combined and the liquid was evaporated using a Rotavapor@
(Buchi Corpora-
tion, New Castle, DE) set to rotate at 45 'C. The final product was
lyophilized (SP
Scientific VirTis Lyophilizer, Warminster, PA) for 16 hours.
NHS-furan was dissolved in dimethyl sulfoxide (DMSO) (45 mg/mL) and
added to
an aqueous antibody solution (10 - 25 mg/mL) to form a 10:1 molar
mixture of NHS-
furan to antibody. The mixture was rotated (VWR@ Tube Rotator,
Radnor, PA)
for 2 hours at room temperature to encourage the formation of
furan-functionalized
antibody.
The resulting solution was placed into the lumen of a
Slide-A-LyzerT M dialysis
cassette with a 10 kDa molecular weight cutoff (Thermo Scientific,
Rockford, IL).
The dialysis bag was placed into 4L 1X phosphate buffered saline
(PBS) (pH 7.4)
and incubated at 2 - 8 'C with continuous stirring. The PBS buffer
was exchanged
every 8 hours for 24 hours, and the final dialyzed solution was
stored at 2 - 8 *C until
use.
2.2.2 Formation of Antibody-Modified Conduits
PDMS tubes were cut to length based on the desired surface area to
volume ratio and
plasma oxidized (Harrick Plasma Cleaner, PDC-001, Ithaca, NY) at
1.5 mbar for the
corresponding duration (Table 2.1) (Figure 2-1). Tubes with the
largest surface area
18
to volume ratio (i.e. long length, small diameter) were plasma
oxidized for a longer
duration of time to ensure complete diffusion of plasma throughout
the tube.
Table 2.1: PDMS tube specifications used to prepare
antibody-modified conduits. Surface Area to Inner Surface
Plasma
. .Length Srae Volume Volume Ratio Diameter Area Oxidation
(mm- 1 ) (mm) (mm) (mm 2 ) (mm Duration (min)
2.50 1.6 875 4400 1760 20 1.25 3.2 219 2200 1760 2
0.834 4.8 97 1467 1760 2 0.625 6.4 55 1100 1760 2 0.416 9.6 24 733
1760 2
Oxidized PDMS tubes were connected in series and 5% (v/v) APTMS in
acetone
was circulated through the tubes at room temperature for 1 hour
using a peristaltic
pump (Watson-Marlow, Inc., 120U/DV, Wilmington, MA) set at 40.0 rpm
(Figure 2-
1). The resulting aminopropyl silane coated tubes were flushed with
ethanol followed
by forced air to remove any residual solution. NHS-PEG,-Maleimide
in 1X PBS (pH
8.5) (2 mg/mL) was then circulated at room temperature for 1 hour
at 160.0 rpm
to produce Maleimide-PEG, modified tubes (Figure 2-1). Tubes were
then flushed
with 10 mM 2-(N-morpholino)ethanesulfonic acid (MES) (pH 5.5)
followed by forced
air. To quench residual amines, 200 mM acetic anhydride in methanol
(MeOH) was
circulated at room temperature for 2 hours at 160.0 rpm (Figure
2-1). Tubes were then
flushed with 10 mM MES (pH 5.5) followed by forced air. The
furan-functionalized
antibody solution was diluted with 10 mM MES (pH 5.5) to an
antibody concentration
of 0.5 mg/mL and circulated at 37 'C for 13 hours at 160.0 rpm
(Figure 2-1). Before
use, antibody-modified conduits were flushed with sterile IX PBS
(pH 7.4).
Antibody-modified conduits were stored in sterile IX PBS (pH 7.4)
at 2 - 8 'C
until use for a maximum of 2 weeks.
19
H3Cnmo' P'CH
Unmodified PDMS
pH 5.5
HO' I 1 0 'OH
Oxidized PDMS
0 0
i' ' Si Si
with acetic anhydride)
H 3CO
NH 2 NH 2 NH 2
HO' I 10 OH
Aminopropyl Silane Coated Surface
Acetic Anhydride in MeOH 0 0
ace ation
Figure 2-1: Schema of formation of an antibody-modified conduit.
Conjugation of antibodies onto the inner surface of silicone tubes
was accomplished through four sequential reactions. The first step
was to activate the silicone surface by oxygen plasma oxidation to
create reactive silanol species (Si-OH). The anchoring molecule
(APTMS) was then covalently grafted to the Si-OH moieties. Next, a
spacer molecule (NHS-PEG,-Maleimiide) was attached to the primary
amine of the APTMS anchor. In the final step, the terminal
naleimide on the PEG spacer was reacted with the furan moieties
that were added to the antibodies.
20
Filtration
Antibody-modified conduits with a surface area to volume ratio of
0.834 (Table 2.1)
were prepared with NHS-PEG10-Maleimide and a 0.5 mg/mL infiiximab
solution and
tested to determine if selective TNF-a filtration could be achieved
(PEG10 , 10 kDa
molecular weight PEG) (infiiximab is a monoclonal antibody that
binds TNF-a).
Solutions of 53 (w/v) BSA in sterile lX PBS (pH 7.4) were enriched
with 2000
pg/mL TNF-a and 2000 pg/mL VEGF-A and circulated through
infiiximab-modified
conduits using the experimental setup in Figure 2-2. The cytokine
concentrations
were chosen to be an order of magnitude greater than what would be
encountered
clinically during an episode of sepsis to sufficiently challenge
AMCs.
Unm>dified Pump-Compa111>le -----4"'- ·.ll
Tube
Conduit
Figure 2-2: Experimental setup for testing selective TNF-a
filtration with infiiximab modified conduits. An
infiiximab-modified conduit was connected in series with two other
tubes: (1) an unmodified tube (Cole-Parmer Instrument Company, PVC,
ID 3/16, OD 7 /16 , Vernon Hills, IL) compatible with the gear pump
system (Cole Parmer Instrument Company, Vernon Hills, IL) , and
(2) an unmodified tube (Sigma Aldrich , PVC, ID 3/16, OD 1/8, St.
Louis , MO) to return solution to the reservoir. The unmodified
pump-compatible tube was clamped within the pump head, and the free
ends of the modified conduit and unmodified tube were placed in a
reservoir. Reservoirs were made from 50 mL conical tubes, placed on
a 37 °C hot plate (IKA@ Works, Inc. , RCT basic , Wilmington, NC),
and covered with Parafilm@ M (Brand GMBH, Wertheim, Germany) to
avoid evaporation.
21
Reservoirs were filled with 20 mL of BSA cytokine solution and a
0.5 mL sample
was taken at time zero. The cytokine solution was circulated at 20
mL/min and addi-
tional 0.5 mL samples were obtained at 0.5, 1, 2, 3, and 4 hours.
The reservoir volume
and circulation rate were chosen in anticipation of in vivo
experiments (rat model of
sepsis), where the rat circulating blood volume is 20-25 mL and the
extracorporeal
circulation rate will be 20 mL/min. Samples were stored at -20 'C
until analysis.
VEGF-A and TNF-a concentrations in the samples were determined
using VEGF-A
and TNF-a enzyme-linked immunosorbent assay (ELISA) kits (R&D
Systems, Inc.,
Minneapolis, MN).
This experiment was repeated with bevacizumab-functionalized AMCs
with a sur-
face area to volume ratio of 0.834, using a 0.5 mg/mL bevacizumab
solution (beva-
cizumab is a monoclonal antibody that binds VEGF-A).
2.4 Influence of Flow Rate on Elimination Kinetics
Infliximab-modified conduits with a surface area to volume ratio of
0.834 (Table 2.1)
were prepared with NHS-PEGiO-Maleimide and tested to characterize
how flow rate
affects cytokine filtration. Solutions of 5% (w/v) BSA in sterile
1X PBS (pH 7.4)
were enriched with 2000 pg/mL TNF-a and circulated through
infliximab-modified
conduits using the experimental setup in Figure 2-2. Experiments
were performed
with flow rates of 10, 20, 30 and 40 mL/min to determine the
influence of flow rate
on cytokine elimination kinetics. TNF-a concentration was measured
by ELISA at
0, 0.5, 1, 2, 3, and 4 hours.
The rate constant, k (min-1 ), was determined for each flow rate as
a measure
of how fast cytokines were cleared from the solution. Assuming a
one phase decay
model, each rate constant was determined using a nonlinear curve
fit.
22
on Elimination Kinetics
Antibody-modified conduits with a surface area to volume ratio of
0.834 (Table 2.1)
were prepared with NHS-PEGIO-Maleimide, followed by the addition of
infliximab us-
ing circulating antibody concentrations of 0.05, 0.5, 1.0, 1.5 and
2.0 mg/mL. Solutions
of 5% (w/v) BSA in sterile 1X PBS (pH 7.4) were enriched with 2000
pg/mL TNF-a
and circulated through infliximab-modified conduits at a flow rate
of 20 mL/min, us-
ing the experimental setup in Figure 2-2. Samples were obtained at
0, 0.5, 1, 2, 3 and
4 hours, and TNF-a concentrations were determined by ELISA. The
rate constant,
k, was determined for each antibody concentration used.
2.6 Influence of Surface Area to Volume Relation-
ships on Elimination Kinetics
Infliximab-modified conduits with various surface area to volume
relationships were
prepared with NHS-PEGIO-Maleimide (Table 2.1). Solutions of 5%
(w/v) BSA in
sterile IX PBS (pH 7.4) were enriched with 3000 pg/mL TNF-a and
circulated at 20
mL/min through infliximab-modified conduits, using the experimental
setup in Figure
2-2. Samples were obtained at 0, 0.5, 1, 2, 3 and 4 hours, and
TNF-a concentrations
were measured by ELISA. The rate constant, k, was determined for
each surface area
to volume ratio.
tion Kinetics
Infliximab-modified conduits with a surface area to volume ratio of
2.50 (Table 2.1)
were prepared with either NHS-PEGiO-Maleimide or NHS-PEG
3-Maleimide. Solu-
tions of 5% (w/v) BSA in sterile IX PBS (pH 7.4) were enriched with
80,000 pg/mL
23
TNF-a and circulated through infliximab-modified conduits at a flow
rate of 20
mL/min, using the experimental setup in Figure 2-2. Samples were
obtained at
0, 0.5, 1, 2, 3, and 4 hours, and TNF-a concentrations were
measured by ELISA. The
rate constant, k, was determined for each PEG spacer length.
2.8 Determination of Cytokine Binding Capacity
Infliximab-modified conduits were prepared with the highest
performing characteris-
tics (2.50 surface area to volume ratio, NHS-PEGiO-Maleimide, 1
mg/mL infliximab
solution) to determine the maximum cytokine binding capacity for
AMCs. Solutions
of 5% (w/v) BSA in sterile 1X PBS (pH 7.4) were enriched with
80,000 pg/mL TNF-
a to saturate AMCs and circulated through infliximab-modified
conduits, using the
experimental setup in Figure 2-2. Samples were obtained at 0, 0.5,
1, 2, 3, 4, 5, 6, 7,
and 8 hours, and TNF-a concentrations were measured by ELISA.
Cytokine binding capacity was calculated using the following
equation:
Capacity = ([Cytokine]initiai - [Cytokine]saturation) x
Volumecirculating fluid (2.1)
2.9 Statistics
Statistical analyses were performed with Prism (Graph-Pad Software
Inc., San Diego,
CA). All data were presented as mean SD. In Sections 3.2 - 3.4,
one-way analysis
of variance (ANOVA) was performed to evaluate the significance of
the data and
calculate the P value. In Section 3.5, unpaired Student t-tests
were performed to
evaluate the significance of the data and calculate the P
value.
24
Filtration
AMCs performed selective cytokine filtration by only clearing the
cytokine they were
designed to filter. AMCs modified with infliximab removed 82% of
the TNF-a present
at the starting concentration after 4 hours, while VEGF-A levels
remained unaffected
(Figure 3-1). AMCs modified with bevacizumab removed 78% of the
VEGF-A present
at the starting concentration after 4 hours, while TNF-a levels
remained unaffected
(Figure 3-1).
TNF-a (infliximab conduit)
TNF-a (bevacizumab conduit)
VEGF-A (infliximab conduit)
VEGF-A (bevacizumab conduit)
Figure 3-1: Type of cytokine filtered was determined by the
selection of AMC surface antibody. Conduits modified with
infliximab (monoclonal antibody against TNF-o) removed TNF-o from
the circulation. Conduits modified with bevacizumab (mono- clonal
antibody against VEGF-A) removed VEGF-A. Data were presented as
mean
SD.
9
3.2 Influence of Flow Rate on Elimination Kinetics
Rate of TNF-a elimination was directly proportional to flow rate
(Figure 3-2). Curve
fitting analysis revealed that, TNF-a elimination followed a single
phase, exponential
decay with strong correlation (R 2 > 0.90 for all groups)
(Figure 3-2A). The differ-
ences in k values were determined to be statistically significant
(P = 0.0004, one-way
ANOVA) (Figure 3-2B). Higher flow rates were correlated with higher
k values, sug-
gesting that faster and more favorable kinetics were achieved at
higher flow rates
(Figure 3-2B).
i20 30 mUmin 0.8 -- -.. ** 20 mUmin
0
z ~.. II0.0-
0 1 2 3 4 0 10 20 30 40 Time (hours) Flow Rate (mUmin)
Figure 3-2: TNF-n filtration by infliximab-modified conduits, where
the BSA TNF- a solution was circulated through AMCs at varying flow
rates. All conduits had a surface area to volume ratio of 0.834
(Table 2.1) and were prepared with NHS-PEGio- Maleimide and a 0.5
ing/mL infliximab solution. (A) TNF-a elimination profiles for each
group. (B) Exponential decay rate constants (k) for each group.
Data were
presented as mean SD.
on Elimination Kinetics
Rate of TNF-a elimination was affected by antibody surface packing
density (Figure
3-3). Curve fitting analysis revealed that TNF-a elimination
followed a single phase,
exponential decay (R2 = 0.64, 0.77, 0.44, 0.93, 0.89, 0.94 for 2.0
ng/mL, 1.5 ng/mL,
1.0 mg/mL, 0.5 mg/mL, 0.1 mg/mL, 0.05 mg/mL, respectively) (Figure
3-3A). The
differences in k values were determined to be statistically
significant (P = 0.0104,
one-way ANOVA) (Figure 3-3B). AMC performance (i.e. rate of TNF-a
elimina-
tion) was directly proportional to infliximab concentration used to
prepare AMCs
for concentrations up to 1.0 mg/mL (Figure 3-3B). AMCs prepared
with infliximab
concentrations of 1.0 ing/mL or higher had similar k values (Figure
3-3B).
27
A B -J -- 2.0 mg/mL
30- 1.5 mg/mL '-'250 0 1.0 mg/mL = 200 0.5 mg/mL r
C 150 -- * 0.1mg/mL 1 0.05 mg/mL .V
0 100 500 *
0 1 2 3 4 0.0 0.5 1.0 1.5 2.0 Time (hours) Infliximab Concentration
(mg/mL)
Figure 3-3: TNF-o filtration by infliximab-modified conduits with a
surface area to volume ratio of 0.834 (Table 2.1). AMCs were
prepared with NHS-PEGio-Maleimnide. Surface antibody addition was
performed with solutions of varying infliximab con- centration. (A)
TNF-o elimination profiles for each group. (B) Exponential decay
rate constants (k) for each group. Data were presented as mean
SD.
3.4 Influence of Surface Area to Volume Relation-
ships on Elimination Kinetics
Rate of TNF-o elimination was directly proportional to the ratio of
AMC surface
area to volume (Figure 3-4). Curve fitting analysis revealed that
TNF-a elimination
followed a single phase, exponential decay with strong correlation
(R' > 0.93 for all
groups) (Figure 3-4A). The differences in k values were determined
to be statistically
significant (P < 0.0001, one-way ANOVA) (Figure 3-4B). Higher
surface area to
volume ratios were correlated with higher k values., suggesting
that faster and more
favorable kinetics were achieved by AMCs with higher surface area
to volume ratios
(Figure 3-4B).
4.
3-
1. *
0.0 0.5 1.0 1.5 2.0 2.5 3.0 Surface Area to Volume Ratio
(1/mm)
Figure 3-4: TNF-i filtration by infliximab-modified conduits with
varying surface area to volume ratios. All conduits were prepared
with NHS-PEG1 0 -Maleimide and a 0.5 mig/niL infliximab solution.
(A) TNF-o elimination profiles for each group. (B) Exponential
decay rate constants (k) for each group. Data were presented as
mean
SD.
tion Kinetics
AMC perfornanice was enhanced by increasing PEG spacer length
(Figure 3-5). Curve
fitting analysis revealed that TNF-a elimination followed a single
phase. exponential
decay with strong correlation (B2 = 0.73, 0.99 for 3 kDa PEG. 10
kDa PEG. respec-
tively) (Figure 3-5A). The difference in k values was determined to
be statistically
significant (P = 0.0422, unpaired Student t-test) (Figure 3-5B).
AMCs prepared with
NHS-PEG1 O-Maleimide (PEGio, 10 kDa PEG) had higher k values,
suggesting that
faster and more favorable kinetics were achieved by AMCs with a
longer PEG spacer
length (Figure 3-5B).
A- B E10000 - 3 kDa PEG 2.
.- e- 10 kDa PEG C 8000 . 1. 0
6000 --.
.... . ...... ..... 4000- \ %
20000
0 1 2 3 4 0 5 10 Time (hours) PEG Spacer Length (kDa)
Figure 3-5: TNF-o filtration by inflixinab-iodified conduits.
Conduits with a surface area to volune ratio of 2.50 were prepared
with varying PEG spacer lengths and a 0.5 nig/mL inflixiinab
solution. (A) TNF-i elimination profiles for each group. (B)
Exponential decay rate constants (k) for each group. Data were
presented as mean
SD.
Ideal AICs (NHS-PEG o-Maleimide. 1.0 mg/nL antibody loading
concentration,
2.50 surface area to voluie ratio) performed selective cytokine
filtration until satu-
ration (Figure 3-6). Using Equation 2.1, the binding capacity for
an ideal AMC was
found to be 394 47 pg/nun 2.
2 '1000000.
0'* 0 1 2 3 4 5 6 7 8
Time (hours)
Figure 3-6: TNF-o, filtration by ideal infliximab-modified conduits
(2.50 surface area to volume ratio, NHS-PEGIO-Maleinide, 1 ng/niL
infliximab solution). Cytokine binding capacity for this ideal type
of AMC was determined to be 394 47 pg/1i11112. Data were presented
as mean SD.
30
Discussion
The purpose of this thesis was to demonstrate the concept of
selective blood filtra-
tion, using cytokines as the model target. Ultimately, these data
will be used to
develop a novel therapeutic approach to the treatment of sepsis,
where circulating
blood is purified by selective removal of the signaling molecules
that drive systemic
inflammation. Here, we successfully characterized the variables
that influence AMC
performance.
Ideally, AMCs must fulfill the following criteria: they must (1)
only bind the
targeted cytokine, and (2) be able to reduce clinically relevant
cytokine levels in a
timely manner and maintain them at zero. Objective (1) was
fulfilled by circulating
two types of cytokines through an AMC and observing only the
removal of the type
of cytokine targeted by the antibody (Section 3.1). Objective (2)
was fulfilled by
characterizing AMC variables and designing an ideal AMC to have
quick elimination
kinetics and a large cytokine binding capacity (Sections 3.2 -
3.6).
AMCs were determined to be capable of selective cytokine
filtration. Conduits
cleared targeted cytokines, whereas non-targeted cytokines remained
unaffected and
in the circulating fluid. This proof of concept provides confidence
that conjugating
any type of antibody to the inner surface of silicone circuits will
allow for the rapid
removal of specific cytokines. By applying this concept and
expanding this technology
to target other key cytokines involved in sepsis, we hope to be
able to successfully
create a platform for immune modulation via selective cytokine
filtration.
31
Before moving forward with expansion and further testing, it was
important to
characterize variables that affected AMC performance so that the
highest performing
AMCs could be used for future experimentation. These variables
included: fluid flow
rate, antibody loading concentration, tubing dimension, and PEG
spacer length.
In varying flow rates, we observed that higher flow rates were
correlated with
faster elimination kinetics. By analogy, solute clearance with
traditional hemodialysis
is enhanced by increasing the flow rate of blood [16]. Given that
the highest flow rate
tested (40 mL/min) achieved the best kinetics, a flow rate of 40
mL/min will be
used in future tests. 40 mL/min is equivalent to 100 mL/kg/min for
a 400 gram rat,
and this normalized flow rate is in range with those achieved
during extracorporeal
membrane oxygenation (ECMO) in children and infants (75 - 150
mL/kg/min) [17].
In varying antibody loading concentration, we discovered that AMC
performance
was enhanced up to an antibody loading concentration of 1.0 mg/mL.
Above 1.0
mg/mL, the kinetics remained the same. We hypothesize that this
occurred because
either: (1) the AMCs became saturated with antibodies when
circulated with con-
centrations of 1.0 mg/mL or higher, or (2) the AMCs became
oversaturated with an-
tibodies when circulated with concentrations greater than 1.0
mg/mL, increasing the
extent of steric hindrance and preventing antibodies from favorably
binding cytokines
[18]. In the future, AMCs will be prepared with an antibody loading
concentration
of 1.0 mg/mL.
In varying AMC dimensions, we observed that AMCs with higher
surface area to
volume ratios had faster elimination kinetics. This is most likely
because AMCs with
a large surface area to volume ratio possess a large blood-circuit
interface, allowing
their surface antibodies to be exposed to a greater proportion of
the circulating fluid.
Given this information, future AMCs will be prepared with silicone
tubing with a
surface area to volume ratio of 2.50. It is important to note,
however, that changing
the surface area to volume ratio will also change the fluid
dynamics of the system.
The resistance to blood flow is directly proportional to the length
and inversely pro-
portional to the radius to the fourth power. This could possibly
cause hemolysis due
to increased shear forces, as increased resistance has been shown
to cause hemolysis in
32
extracorporeal circuits [19]. Future in vitro studies will need to
be conducted, using
blood as the circulating fluid, to determine whether AMCs with an
aspect ratio of
2.50 cause hemolysis.
In varying PEG spacer length, we found that AMCs prepared with a 10
kDa PEG
spacer outperformed those prepared with a 3 kDa PEG spacer. The use
of spacers to
distance immobilized antibodies from a surface has been shown to
introduce greater
freedom of antibody motion [20], reduce steric interference
[20,21], and improve sur-
face antibody orientation [21]. Additionally, it is known that the
use of PEG spacers
increases antibody-antigen interactions by up to 90% and is
dependent upon spacer
length [21]. Thus, it is not surprising that the longer spacer
displayed faster elimina-
tion kinetics. Moving forward, AMCs will be prepared with
NHS-PEGiO-Maleimide
(PEG10 , 10 kDa PEG).
Given these findings, ideal tubes (NHS-PEGio-Maleimide, 1.0 mg/mL
antibody
loading concentration, 2.50 surface area to volume ratio, 40 mL/min
flow rate) were
tested, and the total cytokine binding capacity was calculated. The
total binding
capacity was determined to be 394 47 pg/mm 2 . This capacity will
be helpful in
estimating the number of AMCs needed for use in various in vivo
models.
We now know that selective cytokine filtration with AMCs yields
rapid elimination
of a specific cytokine with the capacity to sustain elimination for
prolonged periods.
Moving forward, we plan to modulate cytokine concentrations in
vivo. While we
have demonstrated proof of concept, a crucial test of this
technology will be to use
AMCs, singly or in combination, to regulate cytokine levels in rats
rendered septic.
In testing our device in vivo, we hope to validate that customized
AMC combinations
can provide precise temporal control over circulating cytokine
concentrations, thereby
preventing organ failures and death.
Also, it is important to note that while we chose sepsis as the
model disease for this
technology, the underlying mechanism being addressed is the
systemic inflammatory
response syndrome (SIRS), which is a common mechanism for a variety
of critical
illnesses (i.e. sepsis [2], ARDS [22], pancreatitis [23], trauma
[23], burns [23]). Thus,
this device, by modulating single or multiple cytokines in a
temporally controlled
33
manner, has the potential to aid in the critical dissection of the
impact of events in the
systemic inflammatory response and become a useful tool for
scientific investigation
as well as therapeutic intervention.
34
In summary, antibody-modified conduits (AMCs) were capable of
selective and timely
cytokine filtration. By characterizing variables that affect AMC
performance, ideal
AMC conditions (NHS-PEGio-Maleimide, 1.0 mg/mL antibody loading
concentra-
tion, 2.50 surface area to volume ratio, 40 mL/min flow rate) were
determined for
later use downstream. Next steps include in vivo testing to
demonstrate that AMCs
can modulate cytokine levels, singly or in combination, in a
temporally controlled
manner in animals rendered septic.
35
36
References
[1] Angus, D.C., Linde-Zwirble, W.T., Lidicker, J., Clermont, G.,
Carcillo, J. and
Pinsky, M.R., 2001, "Epidemiology of severe sepsis in the United
States: analysis
of incidence, outcome, and associated costs of care," Critical Care
Medicine, 29(7),
1303-1310.
[2] Hotchkiss, R.S. and Karl, I.E., 2003, "The pathophysiology and
treatment of
sepsis," New England Journal of Medicine, 348(2), 138-150.
[3] Martin, G.S., Mannino, D.M., Eaton, S. and Moss, M., 2003, "The
epidemiology
of sepsis in the United States from 1979 through 2000," New England
Journal of
Medicine, 348(16), 1546-1554.
[4] Proulx, F., Fayon, M., Farrell, C.A., Lacroix, J. and Gauthier,
M., 1996, "Epi-
demiology of sepsis and multiple organ dysfunction syndrome in
children," Chest,
109(4), 1033-1037.
[5] Hall, M.W. and Muszynski, J.A., 2009, "Immune modulation in
sepsis," Journal
of Pediatric Infectious Disease, 4(2), 127-136.
[6] Fisher, C.J., Slotman, G.J., Opalm S.M., Pribble, J.P., Bone,
R.C., Emmanuel,
G., Ng, D., Bloedow, D.C. and Catalano, M.A., 1994, "Initial
evaluation of human
recombinant interleukin-1 receptor antagonist in the treatment of
sepsis syndrome: a
randomized, open-label, placebo-controlled multicenter trial,"
Critical Care Medicine,
22(1), 12-21.
[7] Karlsson, S., Pettila, V., Tenhunen, J., Lund, V., Hovilehto,
S. and Ruokonen, E.,
37
2008, "Vascular endothelial growth factor in severe sepsis and
septic shock," Anes-
thesia and Analgesia, 106(6), 1820-1826.
[8] Frazier, W.J. and Hall, M.W., 2008, "Immunoparalysis and
Adverse Outcomes
from Critical Illness," Pediatric clinics of North America, 55(3),
647-668.
[9] Bone, R.C., Fisher, C.J., Clemmer, T.P., Slotman, G.J., Metz
C.A. and Balk, R.A.,
1987, "A controlled clinical trial of high-dose methylprednisolone
in the treatment of
severe sepsis and septic shock," New England Journal of Medicine,
317(11), 653-658.
[10] Ziegler, E.J., Fisher, C.J., Sprung, C.L., Straube, R.C.,
Sadoff, J.C., Foulke,
G.E., Wortel, C.H., Fink, M.P., Dellinger, R.P. and Teng, N.N.,
1991, "Treatment
of gram-negative bacteremia and septic shock with HA-1A human
monoclonal anti-
body against endotoxin: a randomized, double-blind,
placebo-controlled trial," New
England Journal of Medicine, 324(7), 429-436.
[11] Abraham, E., Wunderink, R., Silverman, H., Perl, T.M.,
Nasraway, S., Levy,
H., Bone, R., Wenzel, R.P. Balk, R. and Allred, R., 1995, "Efficacy
and safety of
monoclonal antibody to human tumor necrosis factor alpha in
patients with sepsis
syndrome: a randomized, controlled, double-blind, multicenter
clinical trial," JAMA,
273(12), 934-941.
[12] Bernard, G.R., Wheeler, A.P., Russell, J.A., Schien, R.,
Summer, W.R., Stein-
berg, K.P. Fulkerson, W.J., Wright, P.E., Christman, B.W., Dupont,
W.D., Higgins,
S.B. and Swindell, B.B., 1997, "The effects of ibuprofen on the
physiology and survival
of patients with sepsis," New England Journal of Medicine, 336(13),
912-918.
[13] Payen, D., Mateo, J., Cavaillon, J.M., Fraisse, F., Floriot,
C. and Vicaut, E.,
2009, "Impact of continuous venovenous hemofiltration on organ
failure during the
early phase of severe sepsis: a randomized controlled trial,"
Critical Care Medicine,
37(3), 803-810.
[14] Qu, L., Kiss, J.E., Dargo, G. and Carcillo, J.A., 2011,
"Outcomes of previously
healthy pediatric patients with fulminant sepsis-induced
multisystem organ failure
receiving therapeutic plasma exchange," Journal of Clinical
Apheresis, 26(4), 208-
213.
[15] Damas, P., Ledoux, D., Nys, M., Vrindts, Y., De Groote, D.,
Franchimont, P.
38
and Lamy, M., 1992, "Cytokine serum level during severe sepsis in
human IL-6 as a
marker of severity," Annals of Surgery, 215(4), 356-362.
[16] Hauk, M., Kuhlmann, M.K., Riegel, W. and Kohler, H., 2000, "In
vivo effects of
dialysate flow rate on Kt/V in maintenance hemodialysis patients,"
American Journal
of Kidney Diseases, 35(1), 105-111.
[17] Laquier, L., Horton, S.B., McMullan, D.M. and Bartlett, R.H.,
2013, "Extra-
corporeal Membrane Oxygenation Circuitry," Pediatric Critical Care
Medicine, 14,
S7-12.
[18] Xu, H., Lu, J.R. and Williams, D.E., 2006, "Effect of Surface
Packing Density
of Interfacially Adsorbed Monoclonal Antibody on the Binding of
Hormonal Antigen
Human Chorionic Gonadotrophin," Journal of Physical Chemistry,
110(4), 1907-
1014.
[19] Polaschegg, H.D., 2009, "Red blood cell damage from
extracorporeal circulation
in hemodialysis," Seminars in Dialysis, 22(5), 524-531.
[20] Weimer, B.C., Walsh, M.K. and Wang, X., 2000, "Influence of a
poly-ethylene
glycol spacer on antigen capture by immobilized antibodies,"
Journal of Biochemical
and Biophysical Methods, 45(2), 211-219.
[21] Cao, T., Wang, A., Liang, X., Tang, H., Auner, G.W., Salley,
S.O. and Ng,
K.Y., 2007, "Investigation of spacer length effect on immobilized
Escherichia coli pili-
antibody molecular recognition by AFM," Biotechnology and
Bioengineering, 98(6),
1109-1122.
[22] Trager, K., Schutz, C., Fischer, G., Schroder, J., Skrabal,
C., Liebold, A. and
Reinelt, H., 2016, "Cytokine Reduction in the Setting of an
ARDS-Associated In-
flammatory Response with Multiple Organ Failure," Case Reports in
Critical Care.
[23] Fontaine, M., Lepape, A., Piriou, V., Venet, F. and Friggeri,
A., 2016, "Innate
danger signals in acute injury: from bench to bedside,"
Anaesthesia, Critical Care,
and Pain Medicine.