SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\Localizer
TA: 5.8 s PAT: Off Voxel size: 1.1×1.0×5.0 mm Rel. SNR: 1.00
SIEMENS: gre
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 1 Dist. factor 20 % Position L0.0
A50.0 H0.0 Orientation Sagittal Phase enc. dir. A >> P
Rotation 0.00 degSlice group 2 Slices 1 Dist. factor 20 % Position
Isocenter Orientation Transversal Phase enc. dir. A >> P
Rotation 0.00 degSlice group 3 Slices 1 Dist. factor 20 % Position
Isocenter Orientation Coronal Phase enc. dir. R >> L Rotation
0.00 degPhase oversampling 0 %FoV read 250 mmFoV phase 100.0 %Slice
thickness 5.0 mmTR 6.2 msTE 2.67 msAverages 1Concatenations 3Filter
Elliptical filterCoil elements A32
ContrastTD 0 msMTC OffMagn. preparation NoneFlip angle 10 degFat
suppr. NoneWater suppr. NoneSWI Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Short termReconstruction MagnitudeMeasurements
1Multiple series Each measurement
ResolutionBase resolution 256
Phase resolution 90 %Phase partial Fourier OffInterpolation
On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode None- - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - -
Image Filter OffDistortion Corr. OffPrescan Normalize
OffNormalize OffB1 filter OffRaw filter OffElliptical filter OnMode
Inplane
GeometryMulti-slice mode SequentialSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Saturation mode StandardSpecial sat. None
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
-
Table position HTable position 0 mmInline Composing Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Tim CT mode Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HSave uncombined OffCoil Combine
Mode Adaptive CombineAutoAlign ---Auto Coil Select Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode Tune upAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume Position Isocenter
Orientation Transversal Rotation 0.00 deg R >> L 350 mm A
>> P 263 mm F >> H 350 mm
Physio1st Signal/Mode NoneSegments 1
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Tagging NoneDark blood Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Resp. control Off
InlineSubtract OffLiver registration OffStd-Dev-Sag Off
1/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Std-Dev-Cor OffStd-Dev-Tra OffStd-Dev-Time OffMIP-Sag OffMIP-Cor
OffMIP-Tra OffMIP-Time OffSave original images On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Wash - In OffWash - Out OffTTP OffPEI OffMIP - time Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
-
MapIt NoneContrasts 1
SequenceIntroduction OnDimension 2DPhase stabilisation
OffAsymmetric echo AllowedBandwidth 320 Hz/PxFlow comp. No
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
RF pulse type NormalGradient mode NormalExcitation Slice-sel.RF
spoiling On
2/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\AAHScout_32ch
TA: 8.9 s PAT: 4 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00
SIEMENS: AALScout
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OnLoad images to graphicsegments
On
Auto open inline display OffStart measurement withoutfurther
preparation
Off
Wait for user to start OffStart measurements single
RoutineSlab group 1 Slabs 1 Dist. factor 20 % Position L0.0
A50.0 H0.0 Orientation Sagittal Phase enc. dir. A >> P
Rotation 0 degAutoAlign HeadPhase oversampling 0 %Slice
oversampling 0.0 %Slices per slab 128FoV read 260 mmFoV phase 100.0
%Slice thickness 1.6 mmTR 2.90 msTE 1.2 msAverages 1Concatenations
1Filter NoneCoil elements A32
ContrastFlip angle 9.0 deg
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Short termReconstruction MagnitudeMeasurements
1
ResolutionBase resolution 160Phase resolution 100 %Slice
resolution 69 %Phase partial Fourier 6/8Slice partial Fourier
6/8
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 24Accel. factor
3D 2Ref. lines 3D 24Reference scan mode Integrated
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Image Filter OffDistortion Corr. OffPrescan Normalize
OffNormalize OffB1 filter OffRaw filter OffElliptical filter
Off
GeometryMulti-slice mode SequentialSeries Ascending
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode REFMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HSave uncombined OffCoil Combine
Mode Sum of SquaresAuto Coil Select Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode Tune upAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume Position Isocenter
Orientation Transversal Rotation 0.00 deg R >> L 350 mm A
>> P 263 mm F >> H 350 mm
InlineTime to center 4.5 s
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
MapIt NoneContrasts 1
SequenceIntroduction OnDimension 3DAsymmetric echo WeakBandwidth
550 Hz/Px
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
RF pulse type FastGradient mode NormalExcitation Non-sel.RF
spoiling On
3/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\Localizer_aligned
TA: 0:23 PAT: Off Voxel size: 1.1×1.0×5.0 mm Rel. SNR: 1.00
SIEMENS: gre
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OnLoad images to graphicsegments
On
Auto open inline display OffStart measurement withoutfurther
preparation
Off
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 5 Dist. factor 400 % Position
Isocenter Orientation Sagittal Phase enc. dir. A >> P
Rotation 0.00 degSlice group 2 Slices 5 Dist. factor 600 % Position
Isocenter Orientation Transversal Phase enc. dir. A >> P
Rotation 0.00 degSlice group 3 Slices 5 Dist. factor 300 % Position
Isocenter Orientation Coronal Phase enc. dir. R >> L Rotation
0.00 degPhase oversampling 0 %FoV read 250 mmFoV phase 100.0 %Slice
thickness 5.0 mmTR 6.2 msTE 2.67 msAverages 1Concatenations
15Filter Elliptical filterCoil elements A32
ContrastTD 0 msMTC OffMagn. preparation NoneFlip angle 10 degFat
suppr. NoneWater suppr. NoneSWI Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Short termReconstruction MagnitudeMeasurements
1Multiple series Each measurement
ResolutionBase resolution 256
Phase resolution 90 %Phase partial Fourier OffInterpolation
On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode None- - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - -
Image Filter OffDistortion Corr. OffPrescan Normalize
OffNormalize OffB1 filter OffRaw filter OffElliptical filter OnMode
Inplane
GeometryMulti-slice mode SequentialSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Saturation mode StandardSpecial sat. None
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
-
Table position HTable position 0 mmInline Composing Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Tim CT mode Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HSave uncombined OffCoil Combine
Mode Adaptive CombineAutoAlign Head > BrainAuto Coil Select
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode Tune upAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume Position Isocenter
Orientation Transversal Rotation 0.00 deg R >> L 350 mm A
>> P 263 mm F >> H 350 mm
Physio1st Signal/Mode NoneSegments 1
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Tagging NoneDark blood Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Resp. control Off
InlineSubtract OffLiver registration OffStd-Dev-Sag Off
4/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Std-Dev-Cor OffStd-Dev-Tra OffStd-Dev-Time OffMIP-Sag OffMIP-Cor
OffMIP-Tra OffMIP-Time OffSave original images On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Wash - In OffWash - Out OffTTP OffPEI OffMIP - time Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
-
MapIt NoneContrasts 1
SequenceIntroduction OnDimension 2DPhase stabilisation
OffAsymmetric echo AllowedBandwidth 320 Hz/PxFlow comp. No
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
RF pulse type NormalGradient mode NormalExcitation Slice-sel.RF
spoiling On
5/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\AFI
TA: 3:01 PAT: Off Voxel size: 3.0×2.4×5.0 mm Rel. SNR: 1.00
USER: eja_gre
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OffInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlab group 1 Slabs 1 Dist. factor 20 % Position L0.0
A48.2 F15.8 Orientation Transversal Phase enc. dir. R >> L
Rotation 90.00 degPhase oversampling 0 %Slice oversampling 0.0
%Slices per slab 40FoV read 256 mmFoV phase 71.7 %Slice thickness
5.00 mmTR 65 msTE 3.10 msAverages 1Concatenations 1Filter NoneCoil
elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 45 degFat suppr.
NoneWater suppr. NoneSWI Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Short termReconstruction MagnitudeMeasurements
1Multiple series Off
ResolutionBase resolution 106Phase resolution 80 %Slice
resolution 100 %Phase partial Fourier 6/8Slice partial Fourier
6/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode None- - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - -
Image Filter OffDistortion Corr. OffPrescan Normalize
OffNormalize OffB1 filter OffRaw filter Off
Elliptical filter Off
GeometryMulti-slice mode SequentialSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Saturation mode StandardSpecial sat. None
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
-
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HSave uncombined OffCoil Combine
Mode Sum of SquaresAutoAlign ---Auto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode Tune upAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.0 A23.9 H7.4
! Orientation T > C-20.2 ! Rotation 0.00 deg ! R >> L 124
mm ! A >> P 147 mm ! F >> H 71 mm
Physio1st Signal/Mode NoneSegments 1
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Tagging NoneDark blood Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Resp. control Off
InlineSubtract OffLiver registration OffStd-Dev-Sag
OffStd-Dev-Cor OffStd-Dev-Tra OffStd-Dev-Time OffMIP-Sag OffMIP-Cor
OffMIP-Tra OffMIP-Time OffSave original images On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Wash - In OffWash - Out OffTTP OffPEI OffMIP - time Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
6/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
MapIt NoneContrasts 1
SequenceIntroduction OffDimension 3DElliptical scanning OffPhase
stabilisation OffReordering LinearAsymmetric echo OffBandwidth 300
Hz/PxFlow comp. No
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
RF pulse type NormalGradient mode FastExcitation Slab-sel.RF
spoiling On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
B0/B1 mapping Dual-TR B1TR Offset 40000 usRel. RO spoiler mom.
10.00Rel. 3D spoiler mom. 40.00Dual-TR spoiler ratio 0.170Dummy
scan duration 2000 msFFT scale factor 1.00Trigger per meas. Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
TX/RX Nucleus 1HTX/RX delta frequency 0 HzTX Nucleus NoneTX
delta frequency 0 Hz
7/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\SpinEchoFieldMap_PA
TA: 1:26 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_se
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.51 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 3000 msTE 60 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
3Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode Interleaved
Series Interleaved- - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.01 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] Active
8/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Motion correction OffSpatial filter Off
SequenceIntroduction OffBandwidth 1924 Hz/PxFree echo spacing
OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 4480 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 3 1/FoVzBlip scheme 0MB
kernel size 5MB knockout band 0RF pulse shape 1EPI noise scans 0EPI
full reference scan 0Single-band images OnMB LeakBlock kernel OffMB
RF phase scramble OffTime-shifted MB RF OffSENSE1 coil combine
OffLog physiology to file OffInvert RO/PE polarity OnSave reduced
raw data OnReadout slice trace OffDisable ramp sampling OffOnline
multi-band recon. OnlineFFT scale factor 1.00GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 2Multiplier 1Step [1] 1Step [2] 0
9/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\SpinEchoFieldMap_AP
TA: 1:26 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_se
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
Off
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.51 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 3000 msTE 60 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
3Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode Interleaved
Series Interleaved- - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.01 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] Active
10/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Motion correction OffSpatial filter Off
SequenceIntroduction OffBandwidth 1924 Hz/PxFree echo spacing
OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 4480 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 3 1/FoVzBlip scheme 0MB
kernel size 5MB knockout band 0RF pulse shape 1EPI noise scans 0EPI
full reference scan 0Single-band images OnMB LeakBlock kernel OffMB
RF phase scramble OffTime-shifted MB RF OffSENSE1 coil combine
OffLog physiology to file OffInvert RO/PE polarity OffSave reduced
raw data OnReadout slice trace OffDisable ramp sampling OffOnline
multi-band recon. OnlineFFT scale factor 1.00GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 2Multiplier 1Step [1] 1Step [2] 0
11/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\rfMRI_REST_AP
TA: 6:20 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_bold
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OnStart measurement withoutfurther
preparation
On
Wait for user to start OnStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.52 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 1000 msTE 22.2 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 45 degFat suppr.
Fat sat.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
300Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] ActiveMotion correction OffSpatial
filter Off
12/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
SequenceIntroduction OffBandwidth 1924 Hz/PxFlow comp. NoFree
echo spacing OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5760 usSlice multiplier 1Multi-band PE
shift 3 1/FoVzBlip scheme 0MB kernel size 5MB knockout band 0No. of
interleaved TEs 0RF pulse shape 1EPI noise scans 0EPI full
reference scan 0Single-band images OnMB LeakBlock kernel OffMB RF
phase scramble OnSENSE1 coil combine OffLog physiology to file
OffInvert RO/PE polarity OffSave reduced raw data OnReadout slice
trace OffDisable ramp sampling OffPF omits higher k-space OffOnline
multi-band recon. OnlineFFT scale factor 0.60GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 0Multiplier 0
13/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\rfMRI_REST_PA
TA: 6:20 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_bold
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.52 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 1000 msTE 22.2 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 45 degFat suppr.
Fat sat.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
300Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] ActiveMotion correction OffSpatial
filter Off
14/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
SequenceIntroduction OffBandwidth 1924 Hz/PxFlow comp. NoFree
echo spacing OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5760 usSlice multiplier 1Multi-band PE
shift 3 1/FoVzBlip scheme 0MB kernel size 5MB knockout band 0No. of
interleaved TEs 0RF pulse shape 1EPI noise scans 0EPI full
reference scan 0Single-band images OnMB LeakBlock kernel OffMB RF
phase scramble OnSENSE1 coil combine OffLog physiology to file
OffInvert RO/PE polarity OnSave reduced raw data OnReadout slice
trace OffDisable ramp sampling OffPF omits higher k-space OffOnline
multi-band recon. OnlineFFT scale factor 0.60GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 0Multiplier 0
15/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\DWI_dir58_AP
TA: 6:25 PAT: 3 Voxel size: 1.3×1.3×1.3 mm Rel. SNR: 1.00 USER:
eja_ep2d_diff
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 112 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation 0.00 degPhase oversampling 0 %FoV read 210 mmFoV phase
100.0 %Slice thickness 1.25 mmTR 5100 msTE 66.2 msAverages
1Multi-band accel. factor 2Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
1Delay in TR 0 msMultiple series Off
ResolutionBase resolution 168Phase resolution 100 %Phase partial
Fourier 6/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 3Ref. lines PE 66Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
Geometry
Multi-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
DiffDiffusion mode FreeDiff. weightings 1b-value 2000 s/mm²Diff.
weighted images OnTrace weighted images OffAverage ADC maps
OffIndividual ADC maps OffFA maps OffMosaic OnTensor OffNoise level
40Diff. directions 58
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SequenceIntroduction OffBandwidth 1566 Hz/PxFree echo spacing
OffEcho spacing 0.74 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 168Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5120 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 0 1/FoV
16/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
MB RF pulse shift 1920 uszBlip scheme 0MB kernel size 0MB
knockout band 0RF pulse shape 1EPI noise scans 0EPI full reference
scan 0Diffusion Scheme MonopolarSingle-band images OnMB LeakBlock
kernel OffMB RF phase scramble OffTime-shifted MB RF OnSENSE1 coil
combine OnLog physiology to file OffInvert RO/PE polarity OffSave
reduced raw data OnReadout slice trace OffDisable ramp sampling
OffUser defined diff. delta OffOnline multi-band recon. OnlineFFT
scale factor 1.00Diff. spoil factor 3.0GRE iPAT ref. FA 12.0
degSend B1 shim trigger Never
17/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\DWI_dir58_PA
TA: 6:25 PAT: 3 Voxel size: 1.3×1.3×1.3 mm Rel. SNR: 1.00 USER:
eja_ep2d_diff
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
Off
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 112 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation 0.00 degPhase oversampling 0 %FoV read 210 mmFoV phase
100.0 %Slice thickness 1.25 mmTR 5100 msTE 66.2 msAverages
1Multi-band accel. factor 2Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
1Delay in TR 0 msMultiple series Off
ResolutionBase resolution 168Phase resolution 100 %Phase partial
Fourier 6/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 3Ref. lines PE 66Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
Geometry
Multi-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
DiffDiffusion mode FreeDiff. weightings 1b-value 2000 s/mm²Diff.
weighted images OnTrace weighted images OffAverage ADC maps
OffIndividual ADC maps OffFA maps OffMosaic OnTensor OffNoise level
40Diff. directions 58
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SequenceIntroduction OffBandwidth 1566 Hz/PxFree echo spacing
OffEcho spacing 0.74 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 168Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5120 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 0 1/FoV
18/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
MB RF pulse shift 1920 uszBlip scheme 0MB kernel size 0MB
knockout band 0RF pulse shape 1EPI noise scans 0EPI full reference
scan 0Diffusion Scheme MonopolarSingle-band images OnMB LeakBlock
kernel OffMB RF phase scramble OffTime-shifted MB RF OnSENSE1 coil
combine OnLog physiology to file OffInvert RO/PE polarity OnSave
reduced raw data OnReadout slice trace OffDisable ramp sampling
OffUser defined diff. delta OffOnline multi-band recon. OnlineFFT
scale factor 1.00Diff. spoil factor 3.0GRE iPAT ref. FA 12.0
degSend B1 shim trigger Never
19/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\SpinEchoFieldMap_PA
TA: 1:26 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_se
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.51 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 3000 msTE 60 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
3Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode Interleaved
Series Interleaved- - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.01 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] Active
20/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Motion correction OffSpatial filter Off
SequenceIntroduction OffBandwidth 1924 Hz/PxFree echo spacing
OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 4480 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 3 1/FoVzBlip scheme 0MB
kernel size 5MB knockout band 0RF pulse shape 1EPI noise scans 0EPI
full reference scan 0Single-band images OnMB LeakBlock kernel OffMB
RF phase scramble OffTime-shifted MB RF OffSENSE1 coil combine
OffLog physiology to file OffInvert RO/PE polarity OnSave reduced
raw data OnReadout slice trace OffDisable ramp sampling OffOnline
multi-band recon. OnlineFFT scale factor 1.00GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 2Multiplier 1Step [1] 1Step [2] 0
21/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\SpinEchoFieldMap_AP
TA: 1:26 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_se
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
Off
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.51 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 3000 msTE 60 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
3Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode Interleaved
Series Interleaved- - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.01 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] Active
22/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Motion correction OffSpatial filter Off
SequenceIntroduction OffBandwidth 1924 Hz/PxFree echo spacing
OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 4480 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 3 1/FoVzBlip scheme 0MB
kernel size 5MB knockout band 0RF pulse shape 1EPI noise scans 0EPI
full reference scan 0Single-band images OnMB LeakBlock kernel OffMB
RF phase scramble OffTime-shifted MB RF OffSENSE1 coil combine
OffLog physiology to file OffInvert RO/PE polarity OffSave reduced
raw data OnReadout slice trace OffDisable ramp sampling OffOnline
multi-band recon. OnlineFFT scale factor 1.00GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 2Multiplier 1Step [1] 1Step [2] 0
23/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\rfMRI_REST_PA
TA: 6:20 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_bold
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.52 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 1000 msTE 22.2 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 45 degFat suppr.
Fat sat.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
300Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] ActiveMotion correction OffSpatial
filter Off
24/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
SequenceIntroduction OffBandwidth 1924 Hz/PxFlow comp. NoFree
echo spacing OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5760 usSlice multiplier 1Multi-band PE
shift 3 1/FoVzBlip scheme 0MB kernel size 5MB knockout band 0No. of
interleaved TEs 0RF pulse shape 1EPI noise scans 0EPI full
reference scan 0Single-band images OnMB LeakBlock kernel OffMB RF
phase scramble OnSENSE1 coil combine OffLog physiology to file
OffInvert RO/PE polarity OnSave reduced raw data OnReadout slice
trace OffDisable ramp sampling OffPF omits higher k-space OffOnline
multi-band recon. OnlineFFT scale factor 0.60GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 0Multiplier 0
25/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\rfMRI_REST_AP
TA: 6:20 PAT: 2 Voxel size: 1.6×1.6×1.6 mm Rel. SNR: 1.00 USER:
eja_ep2d_bold
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 85 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation -0.52 degPhase oversampling 0 %FoV read 208 mmFoV phase
100.0 %Slice thickness 1.60 mmTR 1000 msTE 22.2 msMulti-band accel.
factor 5Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 45 degFat suppr.
Fat sat.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
300Delay in TR 0 msMultiple series Off
ResolutionBase resolution 130Phase resolution 100 %Phase partial
Fourier 7/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 2Ref. lines PE 96Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
GeometryMulti-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
BOLDGLM Statistics OffDynamic t-maps OffStarting ignore meas
0Ignore after transition 0Model transition states OnTemp. highpass
filter OnThreshold 4.00Paradigm size 20Meas[1] BaselineMeas[2]
BaselineMeas[3] BaselineMeas[4] BaselineMeas[5] BaselineMeas[6]
BaselineMeas[7] BaselineMeas[8] BaselineMeas[9] BaselineMeas[10]
BaselineMeas[11] ActiveMeas[12] ActiveMeas[13] ActiveMeas[14]
ActiveMeas[15] ActiveMeas[16] ActiveMeas[17] ActiveMeas[18]
ActiveMeas[19] ActiveMeas[20] ActiveMotion correction OffSpatial
filter Off
26/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
SequenceIntroduction OffBandwidth 1924 Hz/PxFlow comp. NoFree
echo spacing OffEcho spacing 0.64 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 130Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5760 usSlice multiplier 1Multi-band PE
shift 3 1/FoVzBlip scheme 0MB kernel size 5MB knockout band 0No. of
interleaved TEs 0RF pulse shape 1EPI noise scans 0EPI full
reference scan 0Single-band images OnMB LeakBlock kernel OffMB RF
phase scramble OnSENSE1 coil combine OffLog physiology to file
OffInvert RO/PE polarity OffSave reduced raw data OnReadout slice
trace OffDisable ramp sampling OffPF omits higher k-space OffOnline
multi-band recon. OnlineFFT scale factor 0.60GRE iPAT ref. FA 12.0
degSend B1 shim trigger NeverTriggering scheme StandardStarting
ignore meas 0Paradigm size 0Multiplier 0
27/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\DWI_dir57_AP
TA: 6:19 PAT: 3 Voxel size: 1.3×1.3×1.3 mm Rel. SNR: 1.00 USER:
eja_ep2d_diff
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 112 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation 0.00 degPhase oversampling 0 %FoV read 210 mmFoV phase
100.0 %Slice thickness 1.25 mmTR 5100 msTE 66.2 msAverages
1Multi-band accel. factor 2Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
1Delay in TR 0 msMultiple series Off
ResolutionBase resolution 168Phase resolution 100 %Phase partial
Fourier 6/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 3Ref. lines PE 66Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
Geometry
Multi-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
DiffDiffusion mode FreeDiff. weightings 1b-value 2000 s/mm²Diff.
weighted images OnTrace weighted images OffAverage ADC maps
OffIndividual ADC maps OffFA maps OffMosaic OnTensor OffNoise level
40Diff. directions 57
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SequenceIntroduction OffBandwidth 1566 Hz/PxFree echo spacing
OffEcho spacing 0.74 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 168Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5120 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 0 1/FoV
28/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
MB RF pulse shift 1920 uszBlip scheme 0MB kernel size 0MB
knockout band 0RF pulse shape 1EPI noise scans 0EPI full reference
scan 0Diffusion Scheme MonopolarSingle-band images OnMB LeakBlock
kernel OffMB RF phase scramble OffTime-shifted MB RF OnSENSE1 coil
combine OnLog physiology to file OffInvert RO/PE polarity OffSave
reduced raw data OnReadout slice trace OffDisable ramp sampling
OffUser defined diff. delta OffOnline multi-band recon. OnlineFFT
scale factor 1.00Diff. spoil factor 3.0GRE iPAT ref. FA 12.0
degSend B1 shim trigger Never
29/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\DWI_dir57_PA
TA: 6:19 PAT: 3 Voxel size: 1.3×1.3×1.3 mm Rel. SNR: 1.00 USER:
eja_ep2d_diff
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
Off
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 112 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation 0.00 degPhase oversampling 0 %FoV read 210 mmFoV phase
100.0 %Slice thickness 1.25 mmTR 5100 msTE 66.2 msAverages
1Multi-band accel. factor 2Filter NoneCoil elements A32
ContrastMTC OffMagn. preparation NoneFlip angle 90 degRefocus
flip angle 180 degFat suppr. NoneGrad. rev. fat suppr. Disabled
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Long termReconstruction MagnitudeMeasurements
1Delay in TR 0 msMultiple series Off
ResolutionBase resolution 168Phase resolution 100 %Phase partial
Fourier 6/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
PAT mode GRAPPAAccel. factor PE 3Ref. lines PE 66Reference scan
mode GRE
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Distortion Corr. OffPrescan Normalize OffRaw filter OnElliptical
filter OffHamming Off
Geometry
Multi-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position HTable position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HCoil Combine Mode Sum of
SquaresAutoAlign Head > BrainAuto Coil Select Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Physio1st Signal/Mode None
DiffDiffusion mode FreeDiff. weightings 1b-value 2000 s/mm²Diff.
weighted images OnTrace weighted images OffAverage ADC maps
OffIndividual ADC maps OffFA maps OffMosaic OnTensor OffNoise level
40Diff. directions 57
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SequenceIntroduction OffBandwidth 1566 Hz/PxFree echo spacing
OffEcho spacing 0.74 ms
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
SIR accel. factor 1EPI factor 168Gradient mode NormalRF spoiling
Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Excite pulse duration 5120 usRefocus pulse duration 10240
usSlice multiplier 1Multi-band PE shift 0 1/FoV
30/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
MB RF pulse shift 1920 uszBlip scheme 0MB kernel size 0MB
knockout band 0RF pulse shape 1EPI noise scans 0EPI full reference
scan 0Diffusion Scheme MonopolarSingle-band images OnMB LeakBlock
kernel OffMB RF phase scramble OffTime-shifted MB RF OnSENSE1 coil
combine OnLog physiology to file OffInvert RO/PE polarity OnSave
reduced raw data OnReadout slice trace OffDisable ramp sampling
OffUser defined diff. delta OffOnline multi-band recon. OnlineFFT
scale factor 1.00Diff. spoil factor 3.0GRE iPAT ref. FA 12.0
degSend B1 shim trigger Never
31/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
\\USER\HCP\LifeSpan\Session 1\FieldMap_diffusion
TA: 1:02 Voxel size: 2.5×1.3×2.5 mm Rel. SNR: 1.00 USER:
gre_field_mapping
PropertiesPrio Recon OffBefore measurementAfter measurementLoad
to viewer OnInline movie OffAuto store images OnLoad to stamp
segments OffLoad images to graphicsegments
Off
Auto open inline display OffStart measurement withoutfurther
preparation
On
Wait for user to start OffStart measurements single
RoutineSlice group 1 Slices 56 Dist. factor 0 % Position L0.0
P12.0 H13.0 Orientation T > C-20.0 Phase enc. dir. A >> P
Rotation 0.00 degPhase oversampling 0 %FoV read 210 mmFoV phase
100.0 %Slice thickness 2.5 mmTR 475.0 msTE 1 4.08 msTE 2 5.1
msAverages 1Concatenations 1Filter NoneCoil elements A32
ContrastMTC OffFlip angle 35 degFat suppr. None
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Averaging mode Short termReconstruction Magn./PhaseMeasurements
1Multiple series Off
ResolutionBase resolution 168Phase resolution 50 %Phase partial
Fourier 6/8Interpolation Off
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Image Filter OffDistortion Corr. OffPrescan Normalize
OffNormalize OffB1 filter OffRaw filter OffElliptical filter
Off
GeometryMulti-slice mode InterleavedSeries Interleaved
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Special sat. None- - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - -
Table position H
Table position 0 mmInline Composing Off
SystemV32 OffA32 On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Positioning mode FIXMSMA S - C - TSagittal R >> LCoronal A
>> PTransversal F >> HSave uncombined OffCoil Combine
Mode Adaptive CombineAutoAlign Head > BrainAuto Coil Select
Default
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
Shim mode AdvancedAdjust with body coil OffConfirm freq.
adjustment OffAssume Silicone Off? Ref. amplitude 1H 0.000
VAdjustment Tolerance AutoAdjust volume ! Position L0.7 P13.2 H9.5
! Orientation T > C-15.6 > S0.3 ! Rotation 0.00 deg ! R
>> L 130 mm ! A >> P 170 mm ! F >> H 120 mm
Composing
SequenceIntroduction OnDimension 2DAsymmetric echo OffContrasts
2Bandwidth 607 Hz/PxFlow comp. Yes
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - -
RF pulse type NormalGradient mode NormalRF spoiling On
32/+
-
SIEMENS MAGNETOM Investigational_Device_7T syngo MR B17
Table of contents
\\USER HCP
LifeSpan Session 1
Localizer AAHScout_32ch Localizer_aligned AFI
SpinEchoFieldMap_PA SpinEchoFieldMap_AP rfMRI_REST_AP rfMRI_REST_PA
DWI_dir58_AP DWI_dir58_PA SpinEchoFieldMap_PA SpinEchoFieldMap_AP
rfMRI_REST_PA rfMRI_REST_AP DWI_dir57_AP DWI_dir57_PA
FieldMap_diffusion
33/-