Top Banner
Chema Diagnostica Via Campania 2/4 I-60030 Monsano | An Italy T +39 0731 605064 | F +39 0731 605672 www.chema.com | [email protected] Hitachi 911-912 instructions rev 410.0.3 - 2014-11-28
47

Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Feb 01, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Chema DiagnosticaVia Campania 2/4I-60030 Monsano | An ItalyT +39 0731 605064 | F +39 0731 605672www.chema.com | [email protected]

Hitachi 911-912 instructionsrev 410.0.3 - 2014-11-28

Page 2: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use
Page 3: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 3 of 47

Warning!

These programs should be used only as guideline.

To check the reagent correct setting on the

instrument, use only good quality control sera

and work accordingly to with good laboratory

practice. Instrumental factor must be checked.

Chema Diagnostica will not be responsible for

instrumental bad programming.

Hitachi 911-912 instructionsrev 410.0.2 - 2010-08-01

Page 4: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 4 of 47

Application: ALBUMIN- Codes BC 0100 / 0500 / 1000 / 1500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: ALB APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 600

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 2 DECREASE 2 INCREASE 5

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 3.8 - 5.1 CLASS 1 TECHNICAL RANGE: 0 - 6.0

Page 5: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 5 of 47

Application: ALKALINEPHOSPHATASEDGKCFL - Codes AL F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: ALP APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 20 - 30

SAMPLE VOL: NOMINAL 6 DECREASE 3 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 64 - 306 CLASS 1 TECHNICAL RANGE: 0 - 3000

Page 6: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 6 of 47

Application: ALKALINEPHOSPHATASEIFCCFL - Codes AF F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: ALP APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 20 - 30

SAMPLE VOL: NOMINAL 6 DECREASE 3 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 35 - 104 CLASS 1 TECHNICAL RANGE: 0 - 3000

Page 7: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 7 of 47

Application: AMYLASEEPSFL - Codes EA F080 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: AMY APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 14 - 25

SAMPLE VOL: NOMINAL 10 DECREASE 3 INCREASE 20

R1 VOLUME: 240 R2 VOLUME: 60 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 10 - 220 CLASS 1 TECHNICAL RANGE: 0 - 1500

Page 8: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 8 of 47

Application: AMYLASEFL - Codes AM F060 / F120 / F245 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

NOTE: CNP-G3 method is not always available among calibrator values. If it fails, use PNP-G7 value or EPS, and program the parameters (*) as indicated below:

EXPECTED VALUE: 0 - 220 CLASS 1 TECHNICAL RANGE: 0 - 5000

PROGRAM

TEST: AMY APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 7 - 14

SAMPLE VOL: NOMINAL 7 DECREASE 3 INCREASE 20

R1 VOLUME: 280 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 90 CLASS 1 TECHNICAL RANGE: 0 - 3000

Page 9: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 9 of 47

Application: PANCREATICISOAMYLASEFL - Code PA F080 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: IAMY APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 24 - 30

SAMPLE VOL: NOMINAL 6 DECREASE 3 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 10 - 220 CLASS 1 TECHNICAL RANGE: 0 - 2500

Page 10: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 10 of 4710 of 4747

Application: BICARBONATEFL - Code BR F060 / F245 / F400 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CO2 APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - DEC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 22 - 29 CLASS 1 TECHNICAL RANGE: 0 - 50

Page 11: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 11 of 4711 of 4747

Application: BILIRUBINTOTALFL- Code DT F125 / F500 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: BILT APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 505

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 12 DECREASE 5 INCREASE 25

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 1.2 CLASS 1 TECHNICAL RANGE: 0 - 20

Page 12: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 12 of 4712 of 4747

Application: BILIRUBINDIRECTFL- Code DD F125 / F500 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: BILD APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 12 DECREASE 5 INCREASE 25

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 0.2 CLASS 1 TECHNICAL RANGE: 0 - 13

Page 13: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 13 of 4713 of 4747

Application: CALCIUM - Code CA 0305 / 0505 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CA APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 570

ASSAY: 1POINT TIME: 10 POINT: 25

SAMPLE VOL: NOMINAL 5 DECREASE 3 INCREASE 12

R1 VOLUME: 150 R2 VOLUME: 150 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 9.2 - 11.0 CLASS 1 TECHNICAL RANGE: 0 - 20

Page 14: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 14 of 4714 of 4747

Application: CALCIUMASX - Code CA 0100 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: AX APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 660

ASSAY: 1POINT TIME: 10 POINT: 20

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 9.2 - 11.0 CLASS 1 TECHNICAL RANGE: 0 - 20

Page 15: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 15 of 4715 of 4747

Application: CHOLESTEROLFL - Codes CT F100 / F400 / 100F / 150F CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CHOL APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 505

ASSAY: 1POINT TIME: 10 POINT: 25

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 140 - 200 CLASS 1 TECHNICAL RANGE: 0 - 700

Page 16: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 16 of 4716 of 4747

Application: HDLdirectFL- Code HD F080 / F245 / F400 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B CALIBRATOR - RECONSTITUTE AS INDICATED IN INSERT SHEET Storage: REFRIGERATE AT 2-8°C Stability: 60 DAYS OPEN BOTTLES - CALIBRATOR: AS INDICATED IN INSERT SHEET

PROGRAM

TEST: HDLC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 600

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 270 R2 VOLUME: 0 R3 VOLUME: 90 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 35 - 55 CLASS 1 TECHNICAL RANGE: 0 - 220

Page 17: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 17 of 4717 of 4747

Application: LDLdirectFL- Code DL F080 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B CALIBRATOR - RECONSTITUTE AS INDICATED IN INSERT SHEET Storage: REFRIGERATE AT 2-8°C Stability: 60 DAYS OPEN BOTTLES - CALIBRATOR: AS INDICATED IN INSERT SHEET

PROGRAM

TEST: LDLC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 600

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 270 R2 VOLUME: 0 R3 VOLUME: 90 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 76 - 218 CLASS 1 TECHNICAL RANGE: 0 - 400

Page 18: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 18 of 4718 of 4747

Application: CHOLINESTERASEFL(DGKC)- Codes CH F096 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CHE APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 25 - 32

SAMPLE VOL: NOMINAL 5 DECREASE 3 INCREASE 10

R1 VOLUME: 250 R2 VOLUME: 0 R3 VOLUME: 50 R4 VOLUME: 0

ABS LIMIT: 6500 - DEC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 5600 - 11200 CLASS 1 TECHNICAL RANGE: 0 - 25000

Page 19: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 19 of 4719 of 4747

Application: CK-NACFL - Codes CK F060 / F120 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CPK APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: RATE A TIME: 10 POINT: 23 - 31

SAMPLE VOL: NOMINAL 12 DECREASE 5 INCREASE 20

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 190 CLASS 1 TECHNICAL RANGE: 0 - 2000

Page 20: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 20 of 4720 of 4747

Application: CK-MBFL - Codes MB F060 / F120 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CKMB APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: RATE A TIME: 10 POINT: 23 - 31

SAMPLE VOL: NOMINAL 12 DECREASE 5 INCREASE 20

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 24 CLASS 1 TECHNICAL RANGE: 0 - 2000

Page 21: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 21 of 4721 of 4747

Application: COPPER- Code CU 0100 CH Preparation: R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET Storage: DO NOT REFRIGERATE! Stability: 2 WEEKS

PROGRAM

TEST: CU APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 600

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 20 DECREASE 10 INCREASE 30

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 80 - 140 CLASS 1 TECHNICAL RANGE: 0 - 500

Page 22: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 22 of 4722 of 4747

Application: CREATININE- Codes CR 0500 / 1000 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CREA APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 570 - 505

ASSAY: 2 POINT RATE TIME: 10 POINT: 19 - 23

SAMPLE VOL: NOMINAL 25 DECREASE 15 INCREASE 50

R1 VOLUME: 125 R2 VOLUME: 0 R3 VOLUME: 125 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0.6 - 1.2 CLASS 1 TECHNICAL RANGE: 0 - 20

Page 23: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 23 of 4723 of 4747

Application: CREATININE-EFL- Codes CE F125 / F375 / F600 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CE APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 5 DECREASE 2 INCREASE 15

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0.6 - 1.2 CLASS 1 TECHNICAL RANGE: 0 - 50

Page 24: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 24 of 4724 of 4747

Application: CRPFL- Code RP 0090 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: CRP APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 340

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 11 DECREASE 5 INCREASE 20

R1 VOLUME: 200 R2 VOLUME: 0 R3 VOLUME: 25 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LOGIT-LOG (5p) (POINT: 6 - SPAN: 6 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 5 CLASS 1 TECHNICAL RANGE: 0 - 200

Page 25: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 25 of 4725 of 4747

Application: GAMMA-GTFL - Codes GT F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: GGT APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 415

ASSAY: RATE A TIME: 10 POINT: 19 - 30

SAMPLE VOL: NOMINAL 25 DECREASE 10 INCREASE 40

R1 VOLUME: 200 R2 VOLUME: 0 R3 VOLUME: 50 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 11 - 50 CLASS 1 TECHNICAL RANGE: 0 - 800

Page 26: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 26 of 4726 of 4747

Application: GLUCOSEFL - Codes GL F400 / 100F / 150F CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: GLUC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 505

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 60 - 110 CLASS 1 TECHNICAL RANGE: 0 - 500

Page 27: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 27 of 4727 of 4747

Application: GLUCOSEUVFL - Code GL F601 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: GLUC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 60 - 110 CLASS 1 TECHNICAL RANGE: 0 - 700

Page 28: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 28 of 4728 of 4747

Application: GOT/ASTFL - Codes GO F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: GOT APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: RATE A TIME: 10 POINT: 20 - 30

SAMPLE VOL: NOMINAL 25 DECREASE 10 INCREASE 40

R1 VOLUME: 200 R2 VOLUME: 0 R3 VOLUME: 50 R4 VOLUME: 0

ABS LIMIT: 6500 - DEC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 40 CLASS 1 TECHNICAL RANGE: 0 - 440

Page 29: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 29 of 4729 of 4747

Application: GPT/ALTFL - Codes GP F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: GPT APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: RATE A TIME: 10 POINT: 20 - 30

SAMPLE VOL: NOMINAL 25 DECREASE 10 INCREASE 40

R1 VOLUME: 200 R2 VOLUME: 0 R3 VOLUME: 50 R4 VOLUME: 0

ABS LIMIT: 6500 - DEC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 40 CLASS 1 TECHNICAL RANGE: 0 - 440

Page 30: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 30 of 4730 of 4747

Application: IgAFL- Codes GA 0050 / 0100 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: IGA APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 600

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 2 DECREASE 2 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LOGIT-LOG (5p) (POINT: 6 - SPAN: 6 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 4 CLASS 1 TECHNICAL RANGE: 0 - 6

Page 31: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 31 of 4731 of 4747

Application: IgGFL- Codes GG 0050 / 0100 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: IGG APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 600

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LOGIT-LOG (5p) (POINT: 6 - SPAN: 6 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 16 CLASS 1 TECHNICAL RANGE: 0 - 27

Page 32: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 32 of 4732 of 4747

Application: IgMFL- Codes GM 0050 / 0100 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: IGM APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 340

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LOGIT-LOG (5p) (POINT: 6 - SPAN: 6 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 2.5 CLASS 1 TECHNICAL RANGE: 0 - 3.6

Page 33: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 33 of 4733 of 4747

Application: IRONFZ - Codes FE F245 / F400 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B (preparation: read insert sheet) Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: FE APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 570

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 40 DECREASE 20 INCREASE 50

R1 VOLUME: 160 R2 VOLUME: 0 R3 VOLUME: 40 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 59 - 158 CLASS 1 TECHNICAL RANGE: 0 - 1000

Page 34: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 34 of 4734 of 4747

Application: LDHFL - Codes LD F060 / F120 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: LDH APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: RATE A TIME: 10 POINT: 19 - 30

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 10

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 75 R4 VOLUME: 0

ABS LIMIT: 6500 - DEC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 225 - 450 CLASS 1 TECHNICAL RANGE: 0 - 5000

Page 35: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 35 of 4735 of 4747

Application: LIPASEFL - Codes LP F060 / F125 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: LIP APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 570

ASSAY: RATE A TIME: 10 POINT: 20 - 28

SAMPLE VOL: NOMINAL 4 DECREASE 2 INCREASE 10

R1 VOLUME: 200 R2 VOLUME: 0 R3 VOLUME: 50 R4 VOLUME: 0

ABS LIMIT: 15000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 0 - 60 CLASS 1 TECHNICAL RANGE: 0 - 300

Page 36: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 36 of 4736 of 4747

Application: MAGNESIUM- Codes MG 0200 / 0500 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: MG APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 505

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 10

R1 VOLUME: 150 R2 VOLUME: 0 R3 VOLUME: 150 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 1.3 - 2.1 CLASS 1 TECHNICAL RANGE: 0 - 8.0

Page 37: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 37 of 4737 of 4747

Application: MAGNESIUMXL- Codes MX 0300 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: MG APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 1POINT TIME: 10 POINT: 25

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 6

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 1.30 - 2.10 CLASS 1 TECHNICAL RANGE: 0 - 6

Page 38: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 38 of 4738 of 4747

Application: PHOSPHORUS- Code PH 0100 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN INSERT SHEET

PROGRAM

TEST: PHOS APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 10

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 2.3 - 4.7 CLASS 1 TECHNICAL RANGE: 0 - 20

Page 39: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 39 of 4739 of 4747

Application: PROTEINSHS- Code HS 0100 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: HS APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 600

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 2 DECREASE 2 INCREASE 10

R1 VOLUME: 350 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 28 - 141 CLASS 1 TECHNICAL RANGE: 0 - 500

Page 40: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 40 of 4740 of 4747

Application: TOTALPROTEINS- Codes TP 0100 / 0500 / 1000 / 1500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: PROT APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 1POINT TIME: 10 POINT: 31 - 0

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 10

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 6.3 - 8.4 CLASS 1 TECHNICAL RANGE: 0 - 12.0

Page 41: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 41 of 4741 of 4747

Application: TRANSFERRINFL- Codes TF 0050 / 0100 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: TF APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 505

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 2 DECREASE 2 INCREASE 10

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LOGIT-LOG (5p) (POINT: 6 - SPAN: 6 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 1 - 4 CLASS 1 TECHNICAL RANGE: 0 - 8.5

Page 42: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 42 of 4742 of 4747

Application: TRIGLYCERIDESFL - Code TR F100 / F400 / 100F / 150F CH Preparation: R1 - LIQUID READY TO USE SINGLE RAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: TRIG APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 1POINT TIME: 10 POINT: 20

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 10 - 190 CLASS 1 TECHNICAL RANGE: 0 - 1000

Page 43: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 43 of 4743 of 4747

Application: UREAUVFL - Codes AZ F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: UREA APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 340

ASSAY: 2POINTS RATE TIME: 10 POINT: 19 - 23

SAMPLE VOL: NOMINAL 3 DECREASE 2 INCREASE 8

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 6500 - DEC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 10 - 50 CLASS 1 TECHNICAL RANGE: 0 - 300

Page 44: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 44 of 4744 of 4747

Application: UREACOLORFL - Codes UC F400 / 100F CH Preparation: R1 - MIX REAGENT A1 + REAGENT A2 AS INDICATED IN INSERT SHEET R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: UC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 600

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 2 DECREASE 2 INCREASE 10

R1 VOLUME: 200 R2 VOLUME: 0 R3 VOLUME: 200 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 10 - 50 CLASS 1 TECHNICAL RANGE: 0 - 300

Page 45: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 45 of 4745 of 4747

Application: URICACIDTFL - Codes AU F100 / F250 / F402 CH Preparation: R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET Storage: REFRIGERATE AT 2-8°C Stability: 90 DAYS

PROGRAM

TEST: URIC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 1POINT TIME: 10 POINT: 20

SAMPLE VOL: NOMINAL 8 DECREASE 3 INCREASE 10

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 3.5 - 7.7 CLASS 1 TECHNICAL RANGE: 0 - 30

Page 46: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 46 of 4746 of 4747

Application: URICACIDAOXFL - Codes AX F100 / F250 / F600 / 100F CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PROGRAM

TEST: AOX APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 546

ASSAY: 2 POINT END TIME: 10 POINT: 15 - 31

SAMPLE VOL: NOMINAL 12 DECREASE 10 INCREASE 30

R1 VOLUME: 240 R2 VOLUME: 0 R3 VOLUME: 60 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 3.5 - 7.7 CLASS 1 TECHNICAL RANGE: 0 - 35

Page 47: Hitachi 911-912 instructions - Chema 911-912.pdf · Analyzer: Hitachi 911/912 Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912 IUS.410.0.3 Directions for use

Analyzer: Hitachi911/912

Chema Diagnostica - Guidelines for automatic analyzers - Hitachi 911/912

IUS.410.0.3 Directions for use Edition 2014-11-28 page 47 of 4747 of 4747

Application: ZINC - Code ZN 0125 CH Preparation: R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET Storage: REFRIGERATE AT 2-8°C Stability: 2 WEEKS

PROGRAM

TEST: ZINC APP. CODE: 301 - 400 (choose in this interval)

WAVELENGHTS: 700 - 600

ASSAY: 1POINT TIME: 10 POINT: 31

SAMPLE VOL: NOMINAL 15 DECREASE 10 INCREASE 30

R1 VOLUME: 300 R2 VOLUME: 0 R3 VOLUME: 0 R4 VOLUME: 0

ABS LIMIT: 10000 - INC PROZONE LIMIT: 0 - LOWER

CALIB METHOD: LINEAR (POINT: 2 - SPAN: 2 - WEIGHT: 0)

SD LIMIT: 0.250 DUPLICATE LIMIT: 5% 10 Abs SENSITIVITY LIMIT: EXPECTED VALUE: 70 - 115 CLASS 1 TECHNICAL RANGE: 0 - 1000