Page 1: Connex® Spot Monitor (CSM) 1.x Software Version

Connex® Spot Monitor (CSM) 1.22 Software Version

Clinical In-Service Presentation

MC13845

Page 2: Connex® Spot Monitor (CSM) 1.x Software Version

Table of Contents

• Device Overview & Intended Use• Mobile Stand Options

– MS3 Classic Mobile Stand– Accessory Power Management (APM) Stand– Mobile Work Station (MWS) Stand

• CSM Exterior• Power On• Screen Lock Mode• Clinician Login Screen • Home Screen

– Device Status Area – Content Area – Navigation Area

• Profiles – Switching Profiles– Spot Profile– Intervals Profile

• Clinician Login using Clinician Symbol• Patient Identification• Non-Invasive Blood Pressure (NIBP)

– Start/Stop– Results– Technology– Proper Technique– Interval BPs– Interval BP Set Up– Automatic Intervals – Program Intervals

• Pulse Oximetry (SpO2 ) and Pulse Rate (PR)– Technique, Storage, and Technology– Intervals

• Temperature– Technology– Probe Covers (SureTemp® Plus)– Probe Covers (Braun)– Braun Technique– Braun Interferences– Braun Contraindications and Tips– SureTemp® Plus Oral– SureTemp® Plus Oral Interferences– SureTemp® Plus Axillary– SureTemp® Plus Rectal

• Alarm Management– Adjusting Alarms– Silencing Alarms

• Custom Modifiers and Manual Entry• Manual Parameters• Early Warning Scores (EWS)

– Summary Screen• Saving/Clearing Vital Signs• Review Tab

– Multiple Patient View– Single Patient View

• Clinician Sign Out• Power Down/Sleep• Cleaning

– Compatible Wipes and Solutions• Support Resources

2MC13845

Page 3: Connex® Spot Monitor (CSM) 1.x Software Version

Device Overview & Intended Use

Use to obtain vital signs and other parameters

• Adults (>12 yrs)

• Pediatric (29 days-12 yrs)

• Neonate (<28 days)

Note: To change patient type, touch “Patient” tab and select Patient Type from the dropdown list

Return to Table of Contents 3MC13845

Page 4: Connex® Spot Monitor (CSM) 1.x Software Version

Mobile Stand Options

Connex Spot Monitor (CSM) can be mounted on one of three mobile stands:

1. MS3 Classic Mobile Stand

2. APM Mobile Stand

3. MWS Stand

Both have WhisperDrive™ wheels for smooth, quiet maneuverability

1

2

Return to Table of Contents MC13845

3

Page 5: Connex® Spot Monitor (CSM) 1.x Software Version

Accessory basket

SpO2 cable wrap

SpO2 sensor hanger

MS3 Classic Mobile Stand

Note: one of three mobile stand options

Return to Table of Contents MC13845

Page 6: Connex® Spot Monitor (CSM) 1.x Software Version

Accessory Power Management (APM) Mobile Stand

6

Night light

Accessory compartments

SpO2 sensor and cable holder

Cleaning wipes holder

Work surface/integrated extended battery

Note: one of three mobile stand optionsReturn to Table of Contents MC13845

Page 7: Connex® Spot Monitor (CSM) 1.x Software Version

Mobile Work Station (MWS) Stand

7

Accessory compartments

SpO2 sensor and cable holder

Cleaning wipes holder

Work surface/integrated extended battery

Note: one of three mobile stand options

Return to Table of Contents MC13845

Page 8: Connex® Spot Monitor (CSM) 1.x Software Version

CSM Exterior

Blood Pressure Tubing

Power Cord SpO2 Cable

Return to Table of Contents MC13845

Page 9: Connex® Spot Monitor (CSM) 1.x Software Version

Power On

Power On

Press Power Button

Return to Table of Contents MC13845

Page 10: Connex® Spot Monitor (CSM) 1.x Software Version

Screen Lock Mode

Unlock Screen:

1. Touch the “Lock” icon

2. Follow Clinician login process

10

• Protects patient-identifiable information during use

• Increases visibility of vitals

• Appears after a configurable amount of time

Note: (if applicable to your facility) Screen Lock mode initiated by configurable period of inactivity or via a manual lock decision

Return to Table of Contents MC13845

Page 11: Connex® Spot Monitor (CSM) 1.x Software Version

Optional clinician login screen restricts access to protected health information

Note: Not applicable to all facilities

Clinician Login Screen

11

To sign in using clinician login screen:

1. Scan clinician barcode or touch Clinician ID box to type in ID

2. Touch Password box to type in password (if applicable)

3. Touch Sign in to activate Profile selection area

4. Select desired profile

Return to Table of Contents

4

1 2 3

MC13845

Page 12: Connex® Spot Monitor (CSM) 1.x Software Version

Home Screen

Home screen divided into three sections:

12

Device Status Area

Content Area

Navigation Area

Return to Table of Contents MC13845

Page 13: Connex® Spot Monitor (CSM) 1.x Software Version

Home Screen: Device Status Area

Provides device status information, including:• Clinician ID• Device location (i.e. nursing unit)• Time• Connection status (i.e. Ethernet, Wireless, USB, or Bluetooth®)• Profile Selection (i.e. Spot, Office, or Intervals)• Battery status• Alarms and Messages

13Return to Table of Contents MC13845

Page 14: Connex® Spot Monitor (CSM) 1.x Software Version

Home Screen: Content Area

Displays:

• Patient information (if applicable)

• Vital signs

• Modifiers (if applicable)

• Manual parameters (if applicable)

• Early Warning Scores (if applicable)

• Clear and Next buttons

14Return to Table of Contents MC13845

Page 15: Connex® Spot Monitor (CSM) 1.x Software Version

Home Screen: Navigation Area

Displays:

• Navigation tabs for a variety of information

• Tabs available depend on which profile is selected

15Return to Table of Contents MC13845

Page 16: Connex® Spot Monitor (CSM) 1.x Software Version

To change Profiles:

1. Touch current profile shown in Device Status Area

2. Choose desired profile

• Your choices may include:

− Spot − Intervals − AND/OR Office

Note: Available profiles depend on facility’s chosen settings

Note: If profile changed while acquiring patient measurements or while unsaved patient measurements are displayed, measurements are deleted

16

Profiles: Switching Profiles

Return to Table of Contents

1

2

MC13845

Page 17: Connex® Spot Monitor (CSM) 1.x Software Version

Features:• No alarms

• No BP intervals

• Early Warning Scores (if applicable)

• Custom Modifiers and Manual Parameters (if applicable)

Profiles: Spot Profile

17

Note: Take SpO2 sensor OFF prior to saving readings

Return to Table of Contents MC13845

Page 18: Connex® Spot Monitor (CSM) 1.x Software Version

Note: SpO2 sensor stays ON when touching SAVE

Features:• Interval BPs

• Parameter alarms

• Early Warning Scores (if applicable)

• Custom Modifiers and Manual Parameters (if applicable)

Profiles: Intervals Profile

18Return to Table of Contents MC13845

Page 19: Connex® Spot Monitor (CSM) 1.x Software Version

Clinician Login using Clinician Symbol

To sign in using clinician symbol:

1. Touch the clinician symbol

Return to Table of Contents MC13845

Page 20: Connex® Spot Monitor (CSM) 1.x Software Version

Note: Clinician ID will appear next to clinician symbol once clinician has successfully signed in

Clinician Login using Clinician Symbol

To sign in using clinician symbol:

1. Touch clinician symbol

2. Scan clinician ID using barcode scanner ortouch Clinician ID box to manually enter ID and password (if applicable)

3. Touch Sign in to return to Home screen

2

3

Return to Table of Contents MC13845

Page 21: Connex® Spot Monitor (CSM) 1.x Software Version

Patient Identification

• Identify a patient before taking vital signs (VS) by scanning patient’s armband from Home screen

• Patient name/ID will appear in patient ID section of Content Area

Return to Table of Contents MC13845

Page 22: Connex® Spot Monitor (CSM) 1.x Software Version

Stop NIBP:

• Touch to cancel BP in progress

Non-Invasive Blood Pressure (NIBP): Start/Stop

Start NIBP:

• Touch to initiate BP

Return to Table of Contents MC13845

Page 23: Connex® Spot Monitor (CSM) 1.x Software Version

NIBP: Results

• Static (one-time) pulse rate generated from cuff’s BP reading unless SpO2

probe used on patient− Displays SOURCE: NIBP

• Touch NIBP cell to toggle between SYS/DIA reading and MAP

Return to Table of Contents MC13845

Page 24: Connex® Spot Monitor (CSM) 1.x Software Version

NIBP: Technology

With a two-lumen BP hose, the CSM utilizes SureBP® Technology

• SureBP® captures BP on cuff inflation– Reading takes approx. 15 seconds!

– Reverts to Step Deflation method (BP on cuff deflation):• In times of extreme movement• Artifact• Neonate mode• If programmed to do so via Advanced Settings

Return to Table of Contents MC13845

Page 25: Connex® Spot Monitor (CSM) 1.x Software Version

NIBP: Proper Technique

Cuff on bare arm

Align Artery Index Marker with brachial artery

Properly sized cuff

No more than two fingers between cuff and arm

Allow patient to sit quietly for a few minutes

Back supported, legs uncrossed

Arm supported at heart level

Patient sitting quietly and not moving

For more information on Blood Pressure technique, go to:https://www.youtube.com/watch?v=ScbFRb1dKc4&index=21&list=PLkXnwokwhxrU8d-LnfwYy2Lsjzqa2z7kX

If using Welch Allyn FlexiPort® cuff, place port to outside of patient’s body

Return to Table of Contents MC13845

Page 26: Connex® Spot Monitor (CSM) 1.x Software Version

Initiating Interval BPs:

1. Touch symbol

• To use interval BPs the device must be in Intervals Profile– See Switching Profiles for instructions

• Depending on facility’s chosen settings, the device may require a patient and/or a clinician logged in before Intervals may be initiated

NIBP: Interval BPs

Return to Table of Contents MC13845

Page 27: Connex® Spot Monitor (CSM) 1.x Software Version

Initiating Interval BPs:

1. Touch symbol

2. Choose automatic, program, or STAT

3. Adjustintervals/program*

4. Ensure selected

5. Touch• To use interval BPs the device must be in Intervals Profile

– See Switching Profiles for instructions

• Depending on facility’s chosen settings, the device may require a patient and/or a clinician logged in before Intervals may be initiated

• *Note: See following slides for detailed instructions on adjusting intervals

NIBP: Interval BPs

Return to Table of Contents MC13845

Page 28: Connex® Spot Monitor (CSM) 1.x Software Version

NIBP: Interval BPs

After pressing Start Intervals:

• First BP starts immediately

• Intervals symbol now displays count-down until next scheduled NIBP

• Automatically acquired VS save to Review Tab

• Taking extra sets of vitals does not alter intervals schedule

− Must press Save to chart extra sets

28

To Stop Intervals, touch and then Stop Intervals button

Return to Table of Contents MC13845

Page 29: Connex® Spot Monitor (CSM) 1.x Software Version

Switching between automatic, program, and STAT:

- Touch Interval BP type button

- Toggle between options until desired type selected

NIBP: Interval BP Set Up

29Return to Table of Contents MC13845

Page 30: Connex® Spot Monitor (CSM) 1.x Software Version

• Takes NIBP automatically at set frequency

– Frequency ranges from 1 – 240 minutes

• To set frequency:

– Touch white tile to use keypad, or touch arrows

NIBP: Automatic Intervals

30

Ex. Currently set to take NIBP automatically every 15 min

Return to Table of Contents MC13845

Page 31: Connex® Spot Monitor (CSM) 1.x Software Version

• Takes NIBP automatically based on protocols at your facility– Up to 5 Programs can be entered, 6th will always

remain customizable by end user

• To begin:– Set up new Interval Program or choose existing

Program

31

NIBP: Program Intervals

Return to Table of Contents MC13845

Page 32: Connex® Spot Monitor (CSM) 1.x Software Version

For new Interval Program:

1. Touch Program 6

2. Touch line 1 to set its frequency and duration

• Takes NIBP automatically based on protocols at your facility– Up to 5 Programs can be entered, 6th will always

remain customizable by end user

• To begin:– Set up new Interval Program or choose existing

Program

32

NIBP: Program Intervals

Return to Table of Contents MC13845

Page 33: Connex® Spot Monitor (CSM) 1.x Software Version

For new Interval Program:

1. Touch Program 6

2. Touch line 1 to set its frequency and duration

3. Keypad appears. Type in Interval (ex. Q15min) then touch Frequency box andtype in (ex. X8)

4. Press OK to confirm and select next line

• Takes NIBP automatically based on protocols at your facility– Up to 5 Programs can be entered, 6th will always

remain customizable by end user

• To begin:– Set up new Interval Program or choose existing

Program

33

NIBP: Program Intervals

Ex. Program 1q 15 min x 8, thenq 30 min x 4, then q 60 min x 4, then q 120 min x2

Return to Table of Contents MC13845

Page 34: Connex® Spot Monitor (CSM) 1.x Software Version

For new Interval Program:

1. Touch Program 6

2. Touch line 1 to set its frequency and duration

3. Keypad appears. Type in Interval (ex. Q15min) then touch Frequency box andtype in (ex. X8)

4. Press OK to confirm and select next line

• Takes NIBP automatically based on protocols at your facility– Up to 5 Programs can be entered, 6th will always

remain customizable by end user

• To begin:– Set up new Interval Program or choose existing

Program

34

NIBP: Program Intervals

Ex. Program 1q 15 min x 8, thenq 30 min x 4, then q 60 min x 4, then q 120 min x2

Return to Table of Contents MC13845

Page 35: Connex® Spot Monitor (CSM) 1.x Software Version

• Touch SpO2 cell to toggle between large number reading and pleth waveform

Pulse Oximetry (SpO2 ) and Pulse Rate (PR)

Place pulse ox on patient’s finger to obtain SpO2 and PR

• Pulse Rate derived from pulse ox when in use– Displays SOURCE: SpO2

Return to Table of Contents MC13845

Page 36: Connex® Spot Monitor (CSM) 1.x Software Version

SpO2 and PR: Technique, Storage, and Technology

To obtain most accurate SpO2 and PR, ensure:

Clean sensor (use alcohol pad)

Warm, perfused extremity

No nail polish or artificial nails

Placement on opposite extremity of NIBP cuff

When not in use, place pulse ox back

in holster

Nonin ®PureSat®

Technology options

Nellcor®

Masimo®

Return to Table of Contents MC13845

Page 37: Connex® Spot Monitor (CSM) 1.x Software Version

Initiating Interval SpO2 :

1. Touch on Home screen

2. Choose automatic, program, or STAT

3. Adjust intervals as desired

4. Ensure selected and

deselected

5. Touch

Interval SpO2 allows clinician to take interval O2 sats without also taking interval BPs.

SpO2: Intervals

35Return to Table of Contents MC13845

Page 38: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature

38

Home Screen displays:

• Temperature in facility’s chosen unit of measure

• Converted temperature underneath

• Method used to obtain temperature

Return to Table of Contents MC13845

Page 39: Connex® Spot Monitor (CSM) 1.x Software Version

39

Temperature: Technology

Braun ThermoScan®

PRO 6000 Tympanic

Two options for thermometry:

SureTemp® Plus Oral/Axillary

Return to Table of Contents MC13845

Page 40: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: Probe Covers (SureTemp)

• Only use covers designed specifically for Welch Allyn thermometers

– Using other brands or types can cause damageand/or inaccurate readings

• All probe covers are single use only

Return to Table of Contents MC13845

Page 41: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: Probe Covers (Braun)

• Only use covers designed specifically for Braun thermometers

– Using other brands or types can cause damageand/or inaccurate readings

• All probe covers are single use only

Return to Table of Contents MC13845

Page 42: Connex® Spot Monitor (CSM) 1.x Software Version

42

Remove thermometer from cradle

Push probe tip into box to apply new probe cover

Wait for ready light and beep

Place probe snugly in ear canal

Direct probe at opposite temple

Press and release Measure button

Long beep and steady green ExacTempTM light signal end

Press Probe Cover Ejector button to remove cover

Temperature: Braun Technique

Return to Table of Contents MC13845

Page 43: Connex® Spot Monitor (CSM) 1.x Software Version

43

For persons wearing hearing aids or in extreme ambient temperatures, remove the hearing aid or the individual from the situation and wait 20 minutes prior to taking a temperature.

The following external factors may influence ear temperatures:

Used probe cover

Wet/dirty/damaged lens

Lying on a pillow

Hearing aid*

The following external factors DO NOT influence ear temperatures:

Ambient temperature

Moderate cerumen (ear wax)

Otitis media (ear infection)

Tympanostomy tubes

Temperature: Braun Interferences

Return to Table of Contents MC13845

Page 44: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: Braun Contraindications and Tips

44

Complete ear canal occlusion due to cerumen (ear wax) can result in lower temperature readings

Use untreated ear if ear drops/medications have been placed in ear canal

The following are contraindications to the use of tympanic thermometry:

Patient exhibiting symptoms of chronic or acute inflammatory condition of external ear canal

Certain facial or ear deformities

Blood/drainage in external ear canal

Return to Table of Contents MC13845

Page 45: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: SureTemp® Plus Oral

Oral Temperatures:

1. Pull probe from well, apply Welch Allyn probe cover

2. Ensure Oral mode displayed

3. Place probe in sublingual pocket

4. Thermometer beeps when reading complete (4-6 sec)

5. Press top of probe to release cover, then return probe to well

Return to Table of Contents MC13845

Page 46: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: SureTemp® Plus Oral Interferences

46

The following interfere with obtaining an accurate oral temperature for at least 20 minutes

Eating food Chewing gum or mints

Ingesting hot or cold liquids

Brushing teeth/mouth care

Smoking Performing strenuous activity

Return to Table of Contents MC13845

Page 47: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: SureTemp® Plus Axillary

Axillary Temperature:1. Pull probe from well, apply

Welch Allyn probe cover

2. Press Mode twice to select Adult Axillary

3. Place probe vertically in axilla against bare skin

– Think: “Align with the spine”

4. Thermometer will beep when reading complete (12-15 sec)

5. Press top of probe to release cover, then return probe to well

Return to Table of Contents MC13845

Page 48: Connex® Spot Monitor (CSM) 1.x Software Version

Temperature: SureTemp® Plus Rectal

Rectal Temperature:

1. Insert red probe and well into device

2. Pull probe from well, apply Welch Allyn probe cover

3. If needed, apply small amount of lubrication

4. Place probe 3/8” into rectum

– Insert to line on tip of thermometer for adults, less for peds.

5. Thermometer will beep when reading complete (10-15 sec)

6. Press top of probe to release cover, then return probe to well

Return to Table of Contents MC13845

Page 49: Connex® Spot Monitor (CSM) 1.x Software Version

Adjusting alarms:1. Touch bell in desired

parameter tile

Alarm Management: Adjusting Alarms

• Higher and lower alarm parameters displayed on Home screen when using Intervals profile only.

Return to Table of Contents MC13845

Page 50: Connex® Spot Monitor (CSM) 1.x Software Version

Adjusting alarms:1. Touch bell in desired

parameter tile

2. Touch number to be changed and type in desired alarm limit, then press OK• Arrows may also be

used to make incremental changes

3. Touch to turn on/off alarm*• While off, no visual or

audible alarm signals will occur for that parameter

4. Press Home tab when complete to return to Home screen and confirm changes

• Higher and lower alarm parameters displayed on Home screen when using Intervals profile only.

*Note: Ability to turn on/off alarms dependent upon facility’s chosen settings

Alarm Management: Adjusting Alarms

Return to Table of Contents MC13845

2

4

2

Page 51: Connex® Spot Monitor (CSM) 1.x Software Version

Alarm Management: Silencing Alarms

To silence an alarm:• Touch the triangle once or touch the highlighted parameter to silence for

one minute– Depending on facility’s settings, tapping the triangle twice will silence

it for longer than one minute

• Alarm will disappear when patient condition resolved or alarm limit adjusted accordingly

Return to Table of Contents MC13845

Page 52: Connex® Spot Monitor (CSM) 1.x Software Version

To enter Modifiers from Spot or Intervals profile:

1. Take all desired VS

2. Press and hold a particular vital signs tileEx. Press and hold NIBP to enter blood pressure modifiers, such as patient position or cuff site

Custom Modifiers

52

Note: Availability of custom modifier options will depend on facility’s chosen device configurations.

Return to Table of Contents

2

MC13845

Page 53: Connex® Spot Monitor (CSM) 1.x Software Version

• Individual parameter modifiers screen displays:

– keypad for manual entry (if applicable) on the left • Utilized when another method or device is used to obtain VS

– options for extra information that adds to, or “modifies” that specific VS on the right

• Availability of modifiers depends on the facility’s chosen configurations

Custom Modifiers and Manual Entry

53Return to Table of Contents

To Manually Enter VS and/or Enter Modifiers:

1. To manually enter a VS, touch text box and type in desired value.

2. To manually enter modifiers:

a. Touch text boxesto type in numeric value)

b. Touch dropdown menus to select appropriate choice • Shown: Type of BP device –

Wrist, Manual, Automatic, Other

c. Use side scroll bar to view additional modifiers (if applicable)

3. Press OK to confirm return to Home screen

2a

2b

2c

3

MC13845

Page 54: Connex® Spot Monitor (CSM) 1.x Software Version

After required parameters and other desired information entered:

1. Touch manual parameter ribbon

2. Enter desired information

3. Press OK to return to Home screen

• Additional parameters include extra information that may be entered, but do not specifically relate to one of the four VS taken on the CSM

• Available parameters depend on facility’s chosen configuration

• An asterisk (*) beside a manual parameter indicates it is required by the facility

Manual Parameters

54Return to Table of Contents MC13845

Page 55: Connex® Spot Monitor (CSM) 1.x Software Version

Calculating EWS:

1. After all VS taken and Manual Parameters/ Modifiers entered, press EWS button (*if applicable)

• Calculate Early Warning Scores at point-of-care using CSM

• Scoring system customized by facility

• Clinician chooses one of up to three different protocols per set of vitals taken

• Note: EWS only available in Spot and Intervals Profiles

Early Warning Scores (EWS)

Return to Table of Contents MC13845

Page 56: Connex® Spot Monitor (CSM) 1.x Software Version

Calculating EWS:

1. After all VS taken and Manual Parameters/ Modifiers entered, press EWS button (*if applicable)

2. Choose desired EWS program (if applicable)

3. Enter required information to complete EWS protocol

4. Press Next to continue to enter information and/or return to Home

• Calculate Early Warning Scores at point-of-care using CSM

• Scoring system customized by facility

• Clinician chooses one of up to three different protocols per set of vitals taken

• Note: EWS only available in Spot and Intervals Profiles

EWS

Return to Table of Contents

2

3

4

MC13845

Page 57: Connex® Spot Monitor (CSM) 1.x Software Version

Home screen now displays:

• All VS taken

• Modifiers (if applicable) next to corresponding VS

• Manual parameters (if applicable)

• EWS

• To clear all vital signs from the screen and save to device’s memory (Review tab), press

• To clear all vital signs without saving, press

Saving/Clearing Vital Signs

Return to Table of Contents MC13845

Page 58: Connex® Spot Monitor (CSM) 1.x Software Version

Once all EWS information is taken/entered, pressing save brings up a summary screen displaying:• Color-coded individual parameter scores • Required response• Total calculated (aggregate) score

Note: colors, required response, and scoring customized by facility

EWS: Summary Screen

Return to Table of Contents

Press Next to confirm information and return to Home Screen

MC13845

Page 59: Connex® Spot Monitor (CSM) 1.x Software Version

Review saved vitals:• Touch Review tab

Review Tab

59Return to Table of Contents MC13845

Page 60: Connex® Spot Monitor (CSM) 1.x Software Version

Review saved vitals:• Touch Review tab

Send vitals*:• Select check box next to

desired set(s)

• Press Send

• Envelope indicates vitals successfully sent

– VS may clear from Review when successfully sent instead, depending on hospital’s chosen settings

*Note: Sending only available if connected to EMR

Delete vitals:• Select check box next to

desired set(s)

• Press Delete

• Review tab holds up to 400 sets of vitals for 24 hours– Can view multiple patients or single patient

• Displays Date/Time and vitals saved

• If CSM connected to Electronic Medical Record (EMR), and patient armband scanned, patient name or ID number will display under Patient column (depending on facility’s chosen settings)

Review Tab: Multiple Patient View

60Return to Table of Contents MC13845

Page 61: Connex® Spot Monitor (CSM) 1.x Software Version

Review Tab: Single Patient View

61

• Select patient’s name to see single patient view

Return to Table of Contents MC13845

Page 62: Connex® Spot Monitor (CSM) 1.x Software Version

Review Tab: Single Patient View

62

• Select patient’s name to see single patient view

• Displays:

– All VS taken on selected patient

– Measurements that triggered an alarm are highlighted

Note: Only available if patient ID being used

Return to Table of Contents MC13845

Page 63: Connex® Spot Monitor (CSM) 1.x Software Version

Review Tab: Single Patient View

63

• Readings cannot be sent or deleted from single patient view

• Touch to return to multi-patient view

• Touch Home tab to return to Home screen

Return to Table of Contents MC13845

Page 64: Connex® Spot Monitor (CSM) 1.x Software Version

64

To log clinician out:

1. Touch Clinician symbol to open Clinician menu

2. Touch

Note: Powering down CSM will also clear clinician ID

Clinician Sign Out

Return to Table of Contents

1

2

MC13845

Page 65: Connex® Spot Monitor (CSM) 1.x Software Version

Touch Power Button

Power Down/Sleep

Return to Table of Contents MC13845

Page 66: Connex® Spot Monitor (CSM) 1.x Software Version

Power Down/Sleep

Power Down:Touch to shut down

Sleep:Touch to turn off screen

Cancel:Touch to return to

Home

Return to Table of Contents MC13845

Page 67: Connex® Spot Monitor (CSM) 1.x Software Version

CSM and mobile stand:

• Clean on a routine basis according to your facility’s protocols and standards

Accessories:

• See Directions for Use (DFU) for cleaning instructions

Do not submerge or spray anything directly onto device

Cleaning

Return to Table of Contents MC13845

Page 68: Connex® Spot Monitor (CSM) 1.x Software Version

Cleaning: Compatible Wipes and Solutions

Wipes

Accel® INTERVention™ Germicidal Cleaner

Accel® TB Oxiver® TB

CaviWipes® Sani-Cloth® Plus

Cleancide® Sani-Cloth® Bleach

Clinell® Universal Wipes Super Sani-Cloth®

Clorox® HealthCare Bleach Tuffie5 Cleaning Wipes

Solutions (applied to a clean cloth)

Super HDQ® L10 - Dilution rate of ½ oz per gallon of water (1:256)

Virex® II (256) - Dilution rate of ½ ozper gallon of water (1:256)

10% bleach solution (.5% - 1% sodium hypochlorite)

70% isopropyl alcohol solution

Return to Table of Contents MC13845

Page 69: Connex® Spot Monitor (CSM) 1.x Software Version

Customer Care:

1-800-535-6663

Website:

www.welchallyn.com

Directions for Use:https://www.welchallyn.com/content/dam/welchallyn/documents/sap-documents/LIT/80019/80019224LITPDF.pdf

Quick Reference Guide:http://www.welchallyn.com/content/dam/welchallyn/documents/sap-documents/LIT/80019/80019624LITPDF.pdf

Blood Pressure Accuracy and Variability – Quick Reference:http://www.welchallyn.com/content/dam/welchallyn/documents/upload-docs/Training-and-Use/Quick-Reference-Guide/MC4978_Blood-Pressure-Accuracy-and-Variability_Quick-Reference-Guide.pdf

Support Resources