Top Banner

of 15

Carina Conductivas Mixtas

Apr 14, 2018

Download

Documents

marihesa138
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 7/27/2019 Carina Conductivas Mixtas

    1/15

    D106856 A

    Page 1 of 15

    Otologics Fully Implantable Hearing Systems

    Surgical Guide

    Carina with MET V

    For Conductive & Mixed Loss A lications

    01232007

  • 7/27/2019 Carina Conductivas Mixtas

    2/15

    D106856 A

    Page 2 of 15

    This Instruction document outlines the basic steps required to implant the Carina Fully ImplantableHearing Device with the MET V transducer for treatment of conductive and mixed hearing losses. Itis not intended to contain all information relating to the Otologics implant and its components.Refer to Otologics Surgical Training Manual D104096 for additional information.

    Step 1 Planning Capsule Placement and Making the Incision

    Prior to surgery an implant model or capsule template is used toidentify the optimal implant placement and location of the incision.

    See dotted line in Figure 1.

    The implant capsule and coil are placed in a relatively flat regionof the head. The capsule is placed in a bone bed so that it does not

    protrude more than ~2mm above the surface of the skull . Bending

    of coil more than 10 relative to capsule may cause damage, and

    often results in longer charging times for the patient.

    The optimal position of the implant capsule typically lies betweenlines of 30 and 90 degree angle relative to the horizontal with thesilicone charging coil superior to the titanium capsule body, asshown in figure.

    The microphone must be firmly anchored to bone, and away

    from muscle. Proper position is directly posterior to ear canal.Deviations from this position may result in feedback and other

    performance issues for the patient.

    CAUTION: Do not allow any of the implant

    components to touch each other.

  • 7/27/2019 Carina Conductivas Mixtas

    3/15

    D106856 A

    Page 3 of 15

    Step 2 Assembling the Prosthesis Guide

    The Bending Guide with Prosthesis Template provides important guidance in drilling the mastoidectomy. Assemble the Bending Guide withProsthesis Bending Template as shown below. Set the compression assemble to the middle of its range, to ensure that the real transducer willhave some forward and backward adjustability available.

    It is recommended that 5mm of the prosthesis template extendbeyond the mock transducer, which approximates the length of theMET V transducer with full length prosthesis. If the anatomy orplacement of the bracket requires less than 5mm, this implies that

    the prosthesis will need to be cut to match the prosthesis template.

    Important: Do not extend the prosthesis template more

    than 5mm or less than 1mm beyond the Bending Guide,as this simulates a transducer length that cannot be

    achieved.

    Example of Prosthesisbent and crimped tomatch.

    This represents the length of aMET V with the longest possibleprosthesis.

    5mm

    Set compression

    assembly of BendingGuide to the middle of its

    range as shown

    Tighten Locking Colletto Secure Prosthesis

    Template

    Bendable andExtendable tip

    Bending Guide assembledwith Prosthesis Templateand Locking Collet

    Locking ColletHolds template in position

    Prosthesis Template

    Formable wire with ball tip

    Transducer BoneMounting Bracket

    Locking Tool

  • 7/27/2019 Carina Conductivas Mixtas

    4/15

    D106856 A

    Page 4 of 15

    Step 3 Drill the Mastoidectomy and Facial Recess

    Overview: The surgical approach for placement of the MET V

    transducer is typically through the facial recess. From this approach,the prosthetic transducer tip can be connected to the desired middleear anatomy; stapes superstructure; stapes footplate; oval window;round window; or other ossicular structure.

    Procedure: Drill a mastoidectomy to expose the facial recess. A fullmastoidectomy is not usually needed.

    Use the Prosthesis Guide to determine the appropriatesize and location of the mastoidectomy.

    The mastoidectomy only needs to be large enough to accommodatethe transducer and bracket. Open the facial recess to gain access tothe middle ear.

    (NOTE: If a larger mastoidectomy has been drilled in a previousprocedure, a special bone bracket may be necessary to span the

    larger opening.)

    Left Ear

  • 7/27/2019 Carina Conductivas Mixtas

    5/15

    D106856 A

    Page 5 of 15

    Step 4 Positioning and Fastening the Mounting Bracket

    Preliminary Placement Place the mounting bracket so

    that the Prosthesis Template can reach the desired middle

    ear anatomy. Bend the wire of the Prosthesis Template as

    needed to achieve an effective approach to the desired

    contact point.

    Perform additional drilling as necessary to allow proper

    placement of the bone bracket and Bending Guide.

    Important: Do not extend the Prosthesis Templatemore than 5mm or less than 1mm beyond the mocktransducer, as this simulates a transducer length thatcannot be achieved.

    The mounting bracket legs should be bent to allow flush

    placement against the cortex.

    Fastening: Once the bracket legs have been bent to

    conform to the cortex, and the Prosthesis Template is able

    to contact the desired anatomy, drill a hole in each of the

    bracket legs. Screw down all four legs with at least onescrew each.

    Final Adjustments to Prosthesis Template: After

    fastening the bracket to the skull, final adjustment to the

    Prosthesis Template may be made by loosening the

    Locking Collet and/or carefully bending the prosthesis tip

    with an instrument.

    After all adjustments are complete:

    Tighten the Locking Collet to maintain the position ofthe prosthesis template

    Loosen the Locking Tool.Carefully remove the entire Bending Guide assembly

    from the mounting bracket.

    Always maintain a hold on the Bending guide as it is

    loosened so that it does not slip and contact the

    ossicles.

    Loosen the Locking Tool and move the BendingGuide as shown in the diagram to alter the angle ofthe Prosthesis Template. This will also allow forwardand backward adjustment of the Bending Guide.

    If necessary, loosen the Locking Collet for additionalforward or backward adjustment of the ProsthesisTemplate.

    Left Ear

    Left Ear

  • 7/27/2019 Carina Conductivas Mixtas

    6/15

    D106856 A

    Page 6 of 15

    Step 5 Transducer Health Verification

    Transducer Health

    After removing the MET V transducer from the sterile package, connect the sterile TLA IS-1 clip cable and start the TLA

    software.IMPORTANT: The tip of the transducer is delicate, handle carefully and avoid contact with tip.

    Run Initialization to ensure that the transducer is healthy Typically, initialization results in a Green reading.

    NOTE: If TLA gives a yellow Initialization with MET V, look for 2 other things to know the transducer is healthy:

    1. The Initialization values displayed on the right side of the software are in the 150-1300 range.2. A normal Inductance value of 16.5 to 13 mH is measured after clicking Run

    In determining transducer health, the Inductance values is just as important as Initialization. Once the Run button has been

    pressed, TLA should indicate an Inductance value of 13-16.5mH.

    Connect cable to transducer lead.

    Initialize TLA

    TLA Initialization is a

    measurement of transducerimpedance. Most of thevalues are typically be in the

    range 150-400, with a peak

    value of 600-1300 identifiedby the green indicator.

  • 7/27/2019 Carina Conductivas Mixtas

    7/15

    D106856 A

    Page 7 of 15

    Transducer Monitoring

    Run TLA and monitor the Inductance and Impedance values to monitor transducer health

    and ensure that transducer is being handled within safe limits.

    IMPORTANT: During handling, crimping, and placement, the TLA Inductance

    value should remain above 10mH most of the time and should not drop below 7mH.

    An inductance value below 7mH indicates that the transducer is being handled

    outside its operational limits. If Inductance reading drops below 7mH, reduce the

    contact/loading of the transducer tip and ensure that Inductance returns to >12mH.

    Carefully place the transducer into the crimper as shown bellow. Push the transducer forward until it stops, so that the middle of

    the transducer tip is between the crimping pins. MAKE SURE THE TRANSDUCER LEAD IS NOT UNDERNEATH THE

    HANDLE, TO AVOID DAMAING THE LEAD WHEN CRIMPING.

    SLIDE TRANSDUCERFULLY FORWARD

    Inductance

  • 7/27/2019 Carina Conductivas Mixtas

    8/15

    D106856 A

    Page 8 of 15

    Step 6 Prosthesis Shaping, Cutting, and CrimpingOnce the Prosthesis Guide has been bent to achieve contact with the desired anatomy, remove the Prosthesis Guide from the bone

    bracket and place the entire assembly complete into the base of the crimper tool as shown. Ensure that the transducer portion of

    the laser guide is all the way forward as the arrow indicates.

    Slide

    forward as

    shown.

    Place prosthesis in hole on end ofcrimper tool.Continue to hold with tweezers donot let go at any time. Use one pair tohold the base of the prosthesis at the

    edge of the crimper block, and theother to manipulate and cut to length.Cutting line providescorrect final length.

    Replicate bends onoverhanging portion

    onl

    NOTE: The edge of

    the crimper

    represents the end of

    the MET V

    transducer

    Important: Crimper blockdoes not hold prosthesissecurely. A firm grip mustbe maintained on theportion that is to be used.

  • 7/27/2019 Carina Conductivas Mixtas

    9/15

    D106856 A

    Page 9 of 15

    Place the MET V Attachment prosthesis in the tip of the MET V Transducer. Rotate and align MET V Transducer attachment so

    that it matches the Prosthesis Guide as shown in diagram. Slide the MET V Transducer Attachment wire all the way into the

    transducer tip as far as it will go (2mm). The TLA readings will respond during this process, however they must return to the

    range of a healthy transducer once the crimp is complete, as detailed in the following steps.

    It is critical to the reliability of the crimp joint, that the MET V Transducer Attachment wire is inserted fully.

    While the MET V Transducer Attachment wire is inserted the full 2mm, the crimp is made by pushing down firmly on the handle

    once and only once, until it hits the stop.

    NOTE: Cut in a rockingmotion with a roundedblade. DO NOT use aslicing motion.

    The discarded portion of the prosthesiswire will remain inside the fixture in thisslot.

    Transfer cut prosthesis to the tip of thetransducer in the same orientation as the

    bendin tem late uide.

  • 7/27/2019 Carina Conductivas Mixtas

    10/15

    D106856 A

    Page 10 of 15

    You may now release the crimper handle and the MET V Attachment. Verify attachment with a gentle tug.

    Verify TLA readings indicate a healthy transducer, i.e. >12mH Inductance; >500 Ohms Impedance.

    The MET V Transducer may now be removed from the crimper. Press the ejector button as shown below to raise the transducer.

    Gently grab the case of the transducer (DO NOT grab the tip of the transducer OR the prosthesis tip!) and lift it up and out.

    Insert MET V Attachment fully.Monitor TLA Inductance.

    While gently holding MET Vattachment, firmly press

    the crimper handle, all theway down, until it hits the

    stop.

    DO THIS ONLY ONCE

  • 7/27/2019 Carina Conductivas Mixtas

    11/15

    D106856 A

    Page 11 of 15

    Step

    7Mounting the MET V

    Transducer

    Place the locking ring firmly onto the Locking Tool.

    Slide the Locking Tool onto the Insertion Instrument

    Joystick. Grasp the MET V transducer with the

    thumbscrew and joystick assembly.

    Again, the TLA software should be monitoring thetransducer during these steps. Guide the transducer

    into the mounting bracket, taking care to preserve the

    shape of the prosthesis as much as possible. Place the

    transducer such that the prosthesis attachment is near

    the target anatomical site. Tighten the locking ring to

    secure the transducer.

    Left Ear

  • 7/27/2019 Carina Conductivas Mixtas

    12/15

    D106856 A

    Page 12 of 15

    The transducer and prosthesis can be advanced by using the micro-adjust. Additional shaping of the prosthesis may be necessary,

    however the end of the MET V transducer is delicate and care must be taken when performing manipulations of the prosthesis

    attachment. If possible, stabilize the transducer tip with one instrument while manipulating the prosthesis with another. Monitor

    TLA software to guide final placement of transducer, as described below. TheTLA software inductance value is used to ensure

    that manipulations of the prosthesis do not over-stress the transducer.

    Final Placement ofProsthesis TipStep 8

    Placement of MET V transducer on stapes capitulum

    Left Ear

    Placement of MET V transducer on

    round window

    Left Ear

  • 7/27/2019 Carina Conductivas Mixtas

    13/15

    D106856 A

    Page 13 of 15

    Understanding TLA: The TLA software should be used to monitor the transducer throughout final placement. Keep in mind that

    TLA measures pressure on the tip of the transducer. With the MET V prosthesis tip, there may not be as much pressure on the

    transducer as with standard incus placement; therefore, TLA loading guidelines are different for MET V.

    Proper Visualization of Prosthesis is Critical!

    Transducer Health: Use TLA

    Inductance as a monitor of

    transducer health. The Inductance

    may change very little, or it may

    drop 2 mH or more, depending on

    how much pressure is exerted back

    through the prosthesis to the

    transducer. Final loading is

    typically within 2mH of the initial

    Inductance value measured with

    transducer in free air. Final

    inductance MUST be >12mH.

    Transducer Contact: Use TLA

    Impedance as an indicator of

    transducer contact. The TLA

    Impedance value will often drop by

    several hundred Ohms when the

    transducer tip makes contact.

    However, it is not necessary to

    achieve 12mH.

  • 7/27/2019 Carina Conductivas Mixtas

    14/15

    D106856 A

    Page 14 of 15

    Is monopolar only forbidden with electronics or also with the transducer???

    Step

    9

    Securing the Electronics,SAFI Testing, and Closing

    the Incision

    IMPORTANT: Place the microphone posterior to thepinna, away from any musculature. Muscle movementcan cause unwanted noise for the microphone.

    Although the implant capsule and coil have been designedfor long term implantation, care must be taken wheninserting the coil into a tissue pocket. Excessivemanipulation or bending of the coil can damage it. Inaddition, the coil should be placed such that it is not bentmore than 10relative to the implant capsule.

    Place the implant electronics capsule in the bone bed andsecure with two bone screws. Place the microphone in thebone bed pocket and secure with two screws. Rememberthat no implant component may touch any other, as this willreduce performance and my cause feedback. Thetransducer and microphone leads should be free of tensionand sharp bends.

    Verify that the skin flap is no thicker than 6mm. Thin downas necessary to ensure that the charger andprogramming coil will properlycommunicate with the implant once skin iscovering the implant. Take care to avoiddamage to the electronics or lead duringsuturing.

    CAUTION: Once the electronics havebeen implanted, only bi-polarcauterization is permissible. The use ofmono-polar cauterizing equipment maydamage the electronics.

    Right Ear

  • 7/27/2019 Carina Conductivas Mixtas

    15/15

    D106856 A

    Test the system with the SAFI software for

    final functionality confirmation before

    suturing. With MET V, it is acceptable to

    have a resonance peak on the SAFI

    Impedance test, as shown. This peak may

    even extend above the upper red line.

    Typically, if there is a peak in the SAFI

    Impedance data, this peak will correspond to

    the frequency which was selected by the TLA

    software (green or yellow bar) during

    Initialization, and the value will be similar tothe final TLA Impedance value measured.

    SAFI Impedance and TLA impedance are

    similar measurements, and the SAFI

    Impedance chart will usually have similar

    values to the TLA Initialization bands.

    Similar ValueSAFI Peak value similar to

    TLA final loading value

    Similar FrequencySAFI Peak frequency similar

    to TLA Initializationfrequency