BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option Workflow Overview The following figure shows the steps for daily workflow using BD FACSDiva software. Before starting your daily workflow, ensure that your lab’s software administrator has performed all the necessary tasks to set up the software for your use. This guide shows a workflow that uses application settings. Start Up System Check Performance Set Up Experiment Record Data Analyze Data Shut Down System t t t t t Starting Up the System Turn on the cytometer main power. Start up the computer, start BD FACSDiva software, and log in. Check fluid levels in the Cytometer window. Select Cytometer > Fluidics Startup if automatic cleaning is disabled. Check the flow cell for air bubbles. Check that laser warmup has finished, indicated by a ready status. Checking Cytometer Performance Select Cytometer > CST. Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer Performance Report. Close the Cytometer Setup and Tracking window. 23-9548-00 This guide contains instructions for using BD FACSDiva software version 6.0 and later with BD FACSCanto and BD FACSCanto II systems equipped with the BD FACS Loader option. Verify the Cytometer Configuration and bead Lot ID. Clear the checkbox. If needed, select a different configuration or bead lot ID. 1 1 2 3 4 5 6 1 2 3 4
6

BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

May 02, 2018

Download

Documents

lamduong
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

BD FACSDiva Software Quick Reference Guidefor BD FACSCanto Systems with Loader Option

Workflow OverviewThe following figure shows the steps for daily workflow using BD FACSDiva software.

Before starting your daily workflow, ensure that your lab’ssoftware administrator has performed all the necessarytasks to set up the software for your use. This guide showsa workflow that uses application settings.

Start UpSystem

CheckPerformance

Set UpExperiment

RecordData

AnalyzeData

Shut DownSystem

t tt t t

Starting Up the System

Turn on the cytometer main power.

Start up the computer, start BD FACSDiva software, and log in.

Check fluid levels in the Cytometer window.

Select Cytometer > Fluidics Startup if automatic cleaning is disabled.

Check the flow cell for air bubbles.

Check that laser warmup has finished, indicated by a ready status.

Checking Cytometer Performance

Select Cytometer > CST.

Place a tube of the BD™ Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads.

View the Cytometer Performance Report.

Close the Cytometer Setup and Tracking window.

23-9548-00

This guide contains instructions for using BD FACSDiva™ software version 6.0 and later with BD FACSCanto and BD FACSCanto II systems equipped with the BD FACS™ Loader option.

Verify theCytometerConfigurationand beadLot ID.

Clear the checkbox.

If needed, selecta differentconfiguration or bead lot ID.

1

1

2

3

4

5

6

1

2

3

4

Page 2: BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

Setting Up the Experiment

Select Edit > User Preferences and verify that selected preferences are appropriate.

Create an experiment in the Browser. Right-click in the Browser. Select Application Settings > Apply.

Select Experiment > Compensation Setup > Create Compensation Controls.

Copy and paste plots from the Unstained Control normal worksheet to a Global Worksheet.

Specify Carousel Setup settings.

23-9548-00

Create label-specificcontrols as needed.

Select an application setting.

Click Overwriteif necessary.

Select Edit > Select Alland then Edit > Copyto copy plots from the normal workseet.

Select Edit > Pasteto paste the plots toa global worksheet.

Specify acarousel ID.

Specify run settings.

1

2

3

2

4

5

6

Page 3: BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

Verify that the current tube pointer is set to the Unstained Control tube and that a global worksheet is displayed.

Install compensation control tubes in the carousel in the same order as listed in the Browser and install the carousel on the Loader.

Verify that the cytometer is configured for automatic loading and Click .

Click and then click .

View recorded data in the normal worksheets and gate the positive populations.

Select Experiment > Compensation Setup > Calculate Compensation.

23-9548-00

Select a tube to load.

Adjust the P1 gate,right-click, andselect Apply toAll CompensationControls.

Adjust the P2 gatesto fit the positivepopulations.

Rename thecompensation setup.

Verify that theFSC, SSC, andthreshold settingsare appropriate.

10

11

12

3

7

9

8

Page 4: BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

Recording Specimen Data

Create Browser elements.

Specify Carousel Setup settings.

Make entries in the Experiment Layout.

Install the carousel on the Loader and click .

23-9548-00

Use the Browser toolbarto add elements.

Specify acarousel ID.

Specify runsettings and tubepressurization error handling settings.

Specify reagents labels,keywords, and acquisition criteria as needed.

4

1

3

2

4

Page 5: BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

Specify tubes to run.

Analyzing Data

Create plots, gates, and statistics needed for analysis.

Perform quality control of the analysis.

23-9548-00

Create new global worksheets.

Customize plots using the Plot Inspector.

Create custom text and graphics.

Verify that gates are setappropriately for all samples.

Use the population hierarchy to verifyparent/childrelationships.

Check at the specimenlevel to select all tubesunder the specimen.

5

5

1

2

Page 6: BD FACSDiva Software Quick Reference Guide for BD ... · ™Place a tube of the BD Cytometer Setup and Tracking beads in position 1 on a carousel and run the beads. View the Cytometer

23-9548-00

Do one of the following to print or export the results.

• SelectFile>Printtoprinttheactiveworksheet.

• SelectFile>Exporttoexportselectedelements.

• Right-clickaspecimenorexperimentandselectBatchAnalysis(usingaglobalworksheet).

Shutting Down the System

Select Carousel > Clean.

Install the carousel with the appropriate cleaning tubes and perform the cleaning cycle.

Perform a fluidics shutdown.

Empty waste and refill fluids if prompted to do so.

Turn off the cytometer main power and shut down the computer.

Specify where to save the PDF and exported statistics files.

Select to print, save as a PDF,or export the statistics as needed.

Select cleaningtubes and time.

6

3

1

4

3

2

5